The quantity of blood measured by electrical impedance plethysmography is defined by its resistive effect in parallel to the resistance of other tissue of the segment. By substitution of this parallel resistive value, together with data relative to the resistivity of blood and the length of the segment in the formula for the volume of an electrical conductor, we are able to derive the volume of the pulse in cubic centimeters. It follows that the volume displaced from the venous reservoir and the rate of refilling of the venous reservoir of an extremity may also be determined quantitatively.
© 1950 American Heart Association, Inc.
Electrical Impedance Plethysmography
A Physical and Physiologic Approach to Peripheral Vascular Study
This article has been cited by other articles:
![]() |
N. Okuda, N. Okuda, M. Yamada, Shoudao Feng, Y. Asai, and T. Fujinami Diagnosis of Arteriosclerosis Obliterans by Impedance Technique with Special Reference to Relative Blood Flow of the Lower Extremity Angiology, September 1, 1989; 40(9): 808 - 813. [Abstract] [PDF] |
||||
![]() |
J. Linhart, I. Prerovsky, and M. Cachovan Segmental Pulse Wave Velocity in the Lower Limbs in Man Angiology, May 1, 1968; 19(5): 277 - 287. [PDF] |
||||
![]() |
S. Figar Some Basic Deficiencies of the Plethysmographic Method and Possibilities of Avoiding Them Angiology, April 1, 1959; 10(2): 120 - 125. [PDF] |
||||
![]() |
F. A. Simeone, J. J. Cranley, A. M. Grass, R. R. Linton, and R. B. Lynn An Oscillographic Plethysmograph Using a New Type of Transducer Science, October 3, 1952; 116(3014): 355 - 357. [PDF] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1950 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |