Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 1974;49:165-172

This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by FALSETTI, H. L.
Right arrow Articles by BELL, A. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by FALSETTI, H. L.
Right arrow Articles by BELL, A. C.

(Circulation. 1974;49:165.)
© 1974 American Heart Association, Inc.


Analysis and Correction of Pressure Wave Distortion in Fluid-Filled Catheter Systems

HERMAN L. FALSETTI M.D.1; ROBERT E. MATES PH.D.1; ROBYN J. CARROLL 1; RAMJI L. GUPTA M.S.1; ADAM C. BELL SC.D.1

1 From the Departments of Medicine and Mechanical Engineering, State University of New York at Buffalo, Buffalo, New York.

Dynamic characteristics of a variety of catheters and their dependence on various operating parameters are presented. Careful flushing of the catheter with degassed water or saline considerably improves the performance of these systems. The operating temperature, number of catheter uses, and average transmural pressure difference do not have a significant effect.

The error due to catheter distortion in commonly used left ventricular performance indices (end diastolic pressure, peak systolic pressure, maximum dp/dt and Vmax) was assessed. On the basis of these results, it is shown that the natural frequency of the catheter system must be greater than 40 Hz to produce results accurate to 10%.

The experimental data were used to develop a semi-empirical model for the catheter transducer system. This model was used in the design of an analog compensator to further improve the dynamic characteristics of the catheter transducer system. The compensator requires only two parameter settings, catheter natural frequency and damping ratio.


Key Words: Frequency response • Cardiac catheter • Pressure compensation • LV pressure • Rate of pressure rise

Submitted on June 7, 1973
Accepted on September 4, 1973




This article has been cited by other articles:


Home page
J. Appl. Physiol.Home page
F. Akhavein, E. J. St.-Michel, E. Seifert, and C. V. Rohlicek
Decreased left ventricular function, myocarditis, and coronary arteriolar medial thickening following monocrotaline administration in adult rats
J Appl Physiol, July 1, 2007; 103(1): 287 - 295.
[Abstract] [Full Text] [PDF]