Circulation, Vol 61, 704-709, Copyright © 1980 by American Heart Association
M Hirai, T Ohta and S Shionoya
In the present study, we investigated the technique and reliability of
measuring digital blood pressure by means of a rigid, airtight container (a
bladder-free cuff) and a photocell placed on the digit. As cuff pressure
increased, the position of digits in the cuff was maintained in the hand by
fixing one or two fingers inside the cuff and, in the foot, by using a
supporting stick to keep the distance from the cuff to the bed constant. An
airtight seal inside the cuff was achieved using a thin rubber sleeve and
adhesive tape. Blood pressures determined by the bladder-free cuff
correlated well with those obtained by the conventional cuff with bladder.
By using the bladder-free cuff, blood pressure could be measured at the
distal phalanx of the fingers and in all toes.
ARTICLES
Development of a bladder-free cuff for measuring the blood pressure of the fingers and toes
This article has been cited by other articles:
![]() |
M. Hirai, S. Kawai, T. Ohta, T. Seko, and S. Shionoya Measurement of Blood Pressure in All Toes in Arterial Occlusive Disease of the Leg Angiology, June 1, 1982; 33(6): 418 - 426. [Abstract] [PDF] |
||||
|