(Circulation. 1995;91:1247-1252.)
© 1995 American Heart Association, Inc.
Articles |
From the Department of Medicine, University of North Carolina Hospitals, Chapel Hill, NC (J.J.S.); the Department of Electrical Engineering, Duke University (R.A.M.); and the Departments of Medicine and Pathology, Duke University Medical Center, and the Department of Biomedical Engineering, Duke University, Durham, NC (R.E.I.).
Correspondence to Raymond E. Ideker, MD, PhD, University of Alabama at Birmingham, Room G82A, Volker Hall, Box 201, UAB Station, Birmingham, AL 35294-0019.
Background An upper limit to the strength of shocks that induce fibrillation during the vulnerable period, the upper limit of vulnerability (ULV), has been shown to exist in both humans and animals. The purpose of this study was to compare ULV and defibrillation (DF) probability of success curves for a clinically useful nonthoracotomy lead system.
Methods and Results Sixteen pentobarbital-anesthetized pigs were studied. Single-capacitor biphasic waveforms with both phases 5.5 ms in duration were used for ULV and DF testing. A right ventricular catheter electrode served as first-phase cathode and a superior vena cava catheter electrode coupled with a cutaneous R2 patch electrode served as common first-phase anodes. A pacing catheter was placed in the right ventricle to deliver a train of 15 S1 stimuli at a pacing interval of 250 to 300 ms. A ULV shock was delivered on the peak of the T wave as measured from the surface ECG; if ventricular fibrillation was induced, a DF shock was delivered after 10 seconds of fibrillation. Shock voltages were determined by an up-down protocol. Ventricular fibrillation was induced an average of 53 times in each animal. The composite data indicate that below V97, that is, the voltage that leaves the animal in normal sinus rhythm 97% of the time when delivered on the peak of the T wave or the voltage that defibrillates 97% of the time, ULV is lower than DF. ULV and DF became significantly correlated at V80 and maximally correlated at V97. Even at V97, however, ULV and DF differed by more than 100 V in 2 of the 16 animals.
Conclusions ULV approximately equaled DF at V97. This is fortunate because it is clinically important to set the device voltage at the uppermost portion of the probability of success curve. Estimating DF V97 from ULV V97 would reduce the number of fibrillation inductions needed to establish defibrillation shock strength requirements. However, the large difference between ULV V97 and DF in a few animals indicates that further improvement and testing of algorithms for determining ULV V97 must be developed before the technique is used clinically.
Key Words: defibrillation waves fibrillation
This article has been cited by other articles:
![]() |
B. Rodriguez, L. Li, J. C. Eason, I. R. Efimov, and N. A. Trayanova Differences Between Left and Right Ventricular Chamber Geometry Affect Cardiac Vulnerability to Electric Shocks Circ. Res., July 22, 2005; 97(2): 168 - 175. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yashima, Y.-H. Kim, S. Armin, T.-J. Wu, Y. Miyauchi, W. J. Mandel, P.-S. Chen, and H. S. Karagueuzian On the mechanism of the probabilistic nature of ventricular defibrillation threshold Am J Physiol Heart Circ Physiol, January 1, 2003; 284(1): H249 - H255. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. F. Idriss, P. D. Wolf, W. M. Smith, and R. E. Ideker Effect of pacing site on ventricular fibrillation initiation by shocks during the vulnerable period Am J Physiol Heart Circ Physiol, November 1, 1999; 277(5): H2065 - H2082. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. N. Schauerte, K. Ziegert, M. Waldmann, F. A. Schondube, F. Birkenhauer, K. Mischke, M. Grossmann, P. Hanrath, and C. Stellbrink Effect of Biphasic Shock Duration on Defibrillation Threshold With Different Electrode Configurations and Phase 2 Capacitances : Prediction by Upper-Limit-of-Vulnerability Determination Circulation, March 23, 1999; 99(11): 1516 - 1522. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Huang, B. H. KenKnight, G. P. Walcott, D. L. Rollins, W. M. Smith, and R. E. Ideker Effects of Transvenous Electrode Polarity and Waveform Duration on the Relationship Between Defibrillation Threshold and Upper Limit of Vulnerability Circulation, August 19, 1997; 96(4): 1351 - 1359. [Abstract] [Full Text] |
||||
![]() |
S. Behrens, C. Li, P. Kirchhof, F. L. Fabritz, and M. R. Franz Reduced Arrhythmogenicity of Biphasic Versus Monophasic T-Wave Shocks: Implications for Defibrillation Efficacy Circulation, October 15, 1996; 94(8): 1974 - 1980. [Abstract] [Full Text] |
||||
![]() |
R. A. Malkin, S. F. Idriss, R. G. Walker, and R. E. Ideker Effect of Rapid Pacing and T-Wave Scanning on the Relation Between the Defibrillation and Upper-Limit-of-Vulnerability Dose-Response Curves Circulation, September 1, 1995; 92(5): 1291 - 1299. [Abstract] [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |