(Circulation. 1996;93:2114-2120.)
© 1996 American Heart Association, Inc.
Articles |
From the Catheterization Laboratory, Thoraxcenter, Erasmus University, Rotterdam, Netherlands.
Background The new, less shortening, self-expanding Wallstent is characterized by longitudinal flexibility, a protective membrane, a low profile, and a customized range of diameters (3.5 to 6.0 mm). The recent modification of the braiding angle of the Wallstent has resulted in a new device with less shortening on expansion and a concomitant reduction in radial force. We hypothesized that the enforced mechanical remodeling produced by the selection of an oversized Wallstent might result in improved accommodation of subsequent reactive intimal hyperplasia and prevention of chronic recoil of the vessel.
Methods and Results To prove this hypothesis, we recently implanted 44 new, less shortening Wallstents in 35 native coronary arteries in 35 patients with acute or threatened closure after balloon angioplasty, according to a strategy of oversizing of Wallstent diameter and complete coverage of the lesion length. The initial and 6-month follow-up angiograms were analyzed with a computer-based quantitative coronary angiography (QCA) system. Acute gain (minimal luminal diameter [MLD] post minus MLD pre) and late loss (MLD post minus MLD at follow-up) were examined. Stent deployment was successful in 44 of 44 attempts (100%). Nominal stent diameter used was 1.40 mm larger than the maximal vessel diameter. One patient (3%) with a dilated but unstented lesion proximal to the stented segment sustained a subacute occlusion on day 1 associated with myocardial infarction. Event-free survival at 30 days after stent implantation was 97% (34 of 35 patients). Of the 34 patients eligible for 6-month angiographic follow-up, 3 who were asymptomatic declined repeat angiography. MLD (and percent diameter stenosis [% DS]) changed from 0.83±0.50 mm (72%) pre through 3.06±0.48 mm (15%) post to 2.27±0.74 mm (28%) at follow-up. Acute gain was 2.23±0.63 mm, and late loss was 0.78±0.61 mm. Angiographic restenosis (>50% DS) was observed in 5 of 31 patients (16%) at 6 months, all of whom underwent repeat angioplasty. Thus, the overall event-free survival at 6-month follow-up was 83% (29 of 35 patients).
Conclusions The oversized Wallstent implantation with complete coverage of the lesion length conveyed a favorable 6-month clinical and angiographic outcome. The large acute gain obtained by the Wallstent afforded greater accommodation of the subsequent late loss. The enforced mechanical remodeling by oversized new Wallstents may result in prevention of acute and chronic recoil of the vessel wall and subsequently a lower restenosis rate at follow-up.
Key Words: arteries stents coronary disease restenosis angiography angioplasty
This article has been cited by other articles:
![]() |
J. Berkefeld, B. Turowski, A. Dietz, H. Lanfermann, M. Sitzer, T. Schmitz-Rixen, H. Steinmetz, and F. E. Zanella Recanalization Results after Carotid Stent Placement AJNR Am. J. Neuroradiol., January 1, 2002; 23(1): 113 - 120. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.W. Serruys, D.P. Foley, M. Pieper, J.A. Kleijne, and P.J. de Feyter on behalf of the TRAPIST investigators The TRAPIST Study. A multicentre randomized placebo controlled clinical trial of trapidil for prevention of restenosis after coronary stenting, measured by 3-D intravascular ultrasound Eur. Heart J., October 2, 2001; 22(20): 1938 - 1947. [Abstract] [PDF] |
||||
![]() |
J. J. Wentzel, R. Krams, J. C. H. Schuurbiers, J. A. Oomen, J. Kloet, W. J. van der Giessen, P. W. Serruys, and C. J. Slager Relationship Between Neointimal Thickness and Shear Stress After Wallstent Implantation in Human Coronary Arteries Circulation, April 3, 2001; 103(13): 1740 - 1745. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Kornowski, B. Bhargava, D. M. S. Fuchs, A. J. Lansky, L. F. Satler, A. D. Pichard, M. K. Hong, K. M. Kent, R. Mehran, G. W. Stone, et al. Procedural results and late clinical outcomes after percutaneous interventions using long (>=25 mm) versus short (<20 mm) stents J. Am. Coll. Cardiol., March 1, 2000; 35(3): 612 - 618. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Holmes Jr., J. Hirshfeld Jr., D. Faxon, R. Vlietstra, A. Jacobs, S. B. King III, T. M. Bashore, N. D. Bridges, C. B. Higgins, L. F. Hiratzka, et al. ACC expert consensus document on coronary artery stents: Document of the American College of Cardiology J. Am. Coll. Cardiol., November 1, 1998; 32(5): 1471 - 1482. [Full Text] [PDF] |
||||
![]() |
Y. Ozaki, A. G. Violaris, T. Kobayashi, D. Keane, E. Camenzind, C. Di Mario, P. de Feyter, J. R. T. C. Roelandt, and P. W. Serruys Comparison of Coronary Luminal Quantification Obtained From Intracoronary Ultrasound and Both Geometric and Videodensitometric Quantitative Angiography Before and After Balloon Angioplasty and Directional Atherectomy Circulation, July 15, 1997; 96(2): 491 - 499. [Abstract] [Full Text] |
||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |