(Circulation. 1999;100:1684-1689.)
© 1999 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Thoraxcenter, Heartcenter, University Hospital Rotterdam, Dijkzigt (A.J.W., I.P.K., M.S., A.L.G., J.M.R.L., A.d.B., W.J.v.d.G., P.W.S.), and the Daniel den Hoed, Cancer Center (V.L.M.A.C., P.C.L.), Rotterdam, The Netherlands.
Correspondence to Prof Patrick W. Serruys, MD, PhD, Head of the Department of Interventional Cardiology, Thoraxcenter Bd 418, University Hospital Dijkzigt, Dr Molewaterplein 40, 3015 GD Rotterdam, The Netherlands. E-mail Serruys{at}card.azr.nl
| Abstract |
|---|
|
|
|---|
Methods and ResultsThe radioisotope 32P, a ß-particle emitter with a half-life of 14.3 days, was directly embedded into the Isostent. The calculated range of radioactivity was 0.75 to 1.5 µCi. Quantitative coronary angiography measurements were performed before and after the procedure and at 6-month follow-up. A total of 31 radioactive stents were used in 26 patients; 30 (97%) were successfully implanted, and 1 was embolized. Treated lesions were in the left anterior descending coronary artery (n=12), the right coronary artery (n=8), or the left circumflex coronary artery (n=6). Five patients received additional, nonradioactive stents. Treated lesion lengths were 13±4 mm, with a reference diameter of 2.93±0.47 mm. Minimum lumen diameter increased from 0.87±0.28 mm preprocedure to 2.84±0.35 mm postprocedure. No in-hospital adverse cardiac events occurred. All patients received aspirin indefinitely and ticlopidine for 4 weeks. Twenty-three patients (88%) returned for 6-month angiographic follow-up; 17% of them had in-stent restenosis, and 13% had repeat revascularization. No restenosis was observed at the stent edges. Minimum lumen diameter at follow-up averaged 1.85±0.69 mm, which resulted in a late loss of 0.99±0.59 mm and a late loss index of 0.53±0.35. No other major cardiac events occurred during the 6-month follow-up.
ConclusionsThe use of radioactive stents with an activity of 0.75 to 1.5 µCi is safe and feasible.
Key Words: ß-rays angioplasty radioactive isotopes restenosis stents
| Introduction |
|---|
|
|
|---|
Irradiation is used to decrease neointimal proliferation because the actively proliferating cells have an increased sensitivity to the lethal effects of radiation, which inhibits benign hyperplastic reactions such as keloid formation and heterotopic ossification.14 15 Several experimental and clinical trials showed that brachytherapy with a radioactive source after PTCA or stent implantation can reduce restenosis by inhibiting neointimal hyperplasia,16 17 18 19 and several animal studies demonstrated a dose-related reduction of in-stent restenosis with the use of radioactive stents.20 21 22 Furthermore, a dose-dependent delay in the endothelialization of the stent occurred, which increased the chance of subacute thrombosis.20 23
This study evaluated the safety and feasibility of radioactive stent implantation (activity level, 0.75 to 1.5 µCi) in single-lesion, native coronary artery disease.
| Methods |
|---|
|
|
|---|
Radioactive Stent, Dosimetry, and Safety Issues
Two types of stents were implanted in this study: the
Palmaz-Schatz (Cordis Corp, Johnson and Johnson Interventional Systems
Co) and BX stent (Isostent Inc). Phosphorus-32
(32P), a pure ß-emitter with a half-life of
14.3 days, was produced by neutron irradiation of red amorphous
31P for 10 days to achieve a concentration of
20x10-6
32P/31P (100 mCi). The
irradiated phosphorus was then placed into a mass separator, ionized,
and accelerated. A dipole magnet separated the
32P and 31P. Subsequently,
32P was directly implanted into the metal stent
surface.21 The calculated radioactivity of the stents at
implantation was 0.75 to 1.5 µCi, and the dose delivered over 100
days at 1 mm from the stent surface was calculated for each stent.
All personnel were trained in the appropriate handling of radioactive
materials. During implantation, the lucite shield enclosing the stent
and the sheathed introduction system prevented exposure of the operator
to the radiation of the stent. Background measurements of radioactivity
were made by means of a Geiger counter (Model 14c, Ludlum Measurements
Inc). All disposable materials that were in contact with the stent were
immediately disposed of in a plexiglas container, and radioactivity
measurements were made by the radiation technician.
Quantitative Coronary Angiography
Quantitative coronary angiography (QCA) was performed
preprocedure, postprocedure, and at 6-month follow-up. Coronary
angiography was performed after intracoronary administration of
nitrates. The off-line analysis of
2 orthogonal
projections was performed by the CAAS II analysis system
(Pie Medical BV). Calibration of the system was based on dimensions of
the catheters not filled with contrast medium. This method of
analysis has been extensively validated and applied in numerous
clinical trials.24 25 26 The following measurements were
obtained in each projection: minimum lumen diameter (MLD),
reference diameter, percent diameter stenosis (%DS), and
lesion length. Lesion length was user-defined.26
Procedural success was defined as <20% DS as measured by online QCA.
Short-term gain was defined as MLD postprocedure minus MLD
preprocedure. Late loss was defined as MLD postprocedure minus MLD at
follow-up. Late loss index was defined as short-term gain divided by
late loss.27 Restenosis was defined as >50% DS
at follow-up located within the stent or
5 mm from the stent
edges. The latter represents an area where tissue is subjected
both to balloon-induced trauma and to a lower dose of
radiation,21 which may stimulate restenosis. This
edge-effect phenomenon has recently been described in patients and
called the "candy-wrapper effect."28 To quantify an
edge effect, a QCA segmental analysis was performed. At both
postprocedure and follow-up, the treated vessels were first divided
into segments
5 mm in length; then, the mean diameter of the
5-mm segments distal and proximal to the stent edges were calculated
using the CAAS II analysis system. Careful comparison of the
proximal and distal edges was performed postprocedure and at
follow-up.
Procedure and Follow-Up
Patients received 250 mg of aspirin and 10 000 IU of heparin at
the start of the procedure. The activation clotting time was maintained
at >300 s. After balloon predilatation, the radioactive stent was
implanted at a nominal deployment pressure of 8 to 10 atm. If needed,
stent deployment was optimized using shorter postdilatation balloons of
longer diameters to higher pressures (Table 1
). Extreme care was taken to avoid
inflating the balloon outside the edges of the stent. Because of the
poor radiopacity of the Palmaz-Schatz and the BX stents, the best
angiographic view was selected, and images were filmed in a magnified
field (5 inch) with digital zoom enhancement to optimize stent
visualization. All patients received ticlopidine 250 mg BID for 4 weeks
after stent implantation and aspirin 80 mg daily indefinitely. CK and
CK-MB measurements were made, and the ECG was recorded at 6 and 12
to 18 hours postprocedure in all patients.
|
Patients returned for 1- and 6-month clinical follow-up. An ECG was performed at each visit. The 30-day and 6-month clinical end points were death, Q-wave MI (using the Minnesota code criteria29 ), non Q-wave MI (CK-MB rise >2 times normal upper limit), bypass surgery, target segment revascularization, sustained abrupt closure, or subacute thrombosis of the target vessel.
At the 6-month visit, an exercise stress test was performed. Target vessel revascularization was performed on the basis of clinical symptoms and/or evidence of ischemia on exercise testing.
Statistical Analysis
Data are presented as mean±SD. Continuous data were
compared by 2-tailed Student's t test or linear regression
when appropriate.
| Results |
|---|
|
|
|---|
|
Procedural Success
A total of 30 of the 31 stents (97%) were successfully implanted
(26 were BX Isostent and 4 were Palmaz-Schatz) in 26 patients. One
stent (BX) was lost in the peripheral circulation without
clinical sequelae. Eighteen patients were successfully treated with a
single radioactive stent, and 4 required a second radioactive stent to
cover lesions >15 mm. Five patients received additional
nonradioactive stents: 2 due to procedural dissection not covered by
the radioactive stent, 2 because a second radioactive stent was not
available, and 1 because a second radioactive stent became dislodged
when trying to implant it distal to the first radioactive stent. All
procedures were successful, and no complications occurred.
Follow-Up
The mean hospital stay was 1.8 days. All patients were angina-free
at hospital discharge. At 30-day follow-up, no clinical end points had
occurred: 24 patients (92%) were asymptomatic, and 2
patients (8%) had recurrent angina pectoris (AP) of Canadian
Cardiovascular Society Classification (CCS) 1 (n=1) and
CCS 2 (n=1). All 26 patients returned for 6-month clinical follow-up.
Twenty-one (81%) were asymptomatic, and 5 patients (19%)
had AP CCS 1 (n=1), CCS 2 (n=2), CCS 3 (n=1), or CCS 4 (n=1).
Six-month angiographic follow-up was performed in 23 patients (88%). The remaining 3 patients (12%) refused: 2 of them were asymptomatic, and the third had AP CCS 1. Four patients had angiographic restenosis (17%). All restenotic lesions were diffuse (located throughout the entire length of the stent). One of the 4 restenoses occurred in a patient with a single radioactive stent, 1 restenosis was in a patient receiving 2 radioactive stents in combination with a nonradioactive stent, and 2 restenoses were observed in patients receiving a combination of 1 radioactive and 1 nonradioactive stent. In the restenotic patients who received an additional nonradioactive stent, restenosis occurred in both the radioactive and the nonradioactive stent. On QCA, no discernible differences existed between the patterns of proliferation between the Palmaz-Schatz and BX stents. No cases of restenosis at the stent edges were noted. Two of the 4 restenotic patients underwent a re-PTCA. One was referred for bypass surgery for in-stent restenosis in the proximal left anterior descending coronary artery and progression of a previously nonsignificant lesion in the proximal left circumflex artery (main stem equivalent). One was treated medically; this patient was asymptomatic, with a negative stress test. No other clinical end points existed at 6-month follow-up.
QCA Measurements
QCA and procedural data are presented in Table 3
. MLD increased from 0.87±0.28 mm
preprocedure to 2.84±0.35 mm postprocedure
(P<0.0001). MLD at follow-up was 1.85±0.69 mm
(P<0.0001 relative to postprocedure), resulting in a late
loss index of 0.53±0.35. Segmental analysis of the mean
diameter of the 5-mm segments distal and proximal to the stent edges
showed significant changes. The proximal diameter decreased from
3.19±0.42 mm postprocedure to 2.78±0.62 mm at follow-up
(P=0.006). The distal diameter decreased from
2.69±0.49 mm postprocedure to 2.45±0.50 mm at follow-up
(P=0.0167).
|
Radiation Doses
Stent activity level and the cumulative dose over 100 days that
was delivered to a 1 mm depth outside the stent are
presented in Table 3
. No correlation existed between
stent activity or delivered dose and MLD or late loss index at
follow-up. No additional environmental radiation was measured during
the procedure.
| Discussion |
|---|
|
|
|---|
Using a multivariate model constructed from the data of the Benestent trials that was based on similar lesions, vessel size, and short-term result, a predicted restenosis rate of 12% and an MLD at follow-up of 2.05 mm was calculated.30 31 Thus, the actual results achieved are somewhat less favorable; however, in such a small patient cohort, no definite conclusions can be drawn except that the late results are within the acceptable limits for safety and feasibility of this stent. It must be noted that 3 of the 4 patients who had in-stent restenosis had multiple stents implanted, which increases the risk of restenosis; in the group of 18 patients who had a single radioactive stent implanted, only 1 had restenosis. Overall, these 6-month clinical and angiographic results are similar to the published results of nonradioactive stents.10 11
The Milan group was the first to report restenosis within the stent and at the edges of the stent (the candy-wrapper phenomenon); this restenosis was possibly caused by increased balloon injury (barotrauma) and the lower radiation dose at the stent edges.21 28 In the Rotterdam series, particular attention was paid to avoiding balloon injury outside the stent to minimize the edge effect. No cases of edge restenosis were seen in this cohort; however, the proximal and distal mean diameter at the stent edges, measured postprocedure and at follow-up, decreased significantly. Because extreme care was taken to avoid inflating the balloon outside the stent edges, this edge effect may be caused by the lower radiation dose.
Dosimetry
Previous work by Janicki et al32 on the 1.0-µCi
Palmaz-Schatz stent demonstrated the nonuniformity of dosing in areas
adjacent to stent strut wires and those areas between the wires. Models
showed that for a 32P stent of 1.0 µCi that was
15 mm in length, at a distance of 0.1 mm, dose values of 2500
cGy were delivered at the strut wires (peaks) and 800 cGy between the
wires (valleys) over 1 half-life (14.3 days). The nonuniformity of
dosing, reflective of stent geometry, decreased at distances 1 to
2 mm from the stent surface. Although these data provide an
in-vitro analysis of dosing from a radioactive stent, the
actual dose distribution is probably affected by variations in
atherosclerotic plaque morphology and the symmetry of the lesion and
stent expansion. The 2D dosimetry representation of the
Palmaz-Schatz and BX stent were done using the Janicki
model32 (Figures 1
and 2
).
|
|
Currently, dose-finding studies examining restenosis after implantation of 32P BX stents in patients with lesion morphology similar to that described in this study are underway. It is possible that increased doses will decrease in-stent restenosis, as has been described in animal studies.21 22 23 Therefore, a European Dose Response trial has been started with activities ranging from 1.5 to 3, 3 to 6, 6 to 12, and 12 to 20 µCi.
Conclusion
This study reports that the implantation of ß-particleemitting
radioactive stents with an activity of 0.75 to 1.5 µCi is safe and
feasible.
| Acknowledgments |
|---|
Received February 8, 1999; revision received June 28, 1999; accepted July 2, 1999.
| References |
|---|
|
|
|---|
2.
Serruys PW, Luijten HE, Beatt KJ, Geuskens R, de
Feyter PJ, van den Brand M, Reiber JH, ten Katen HJ, van Es GA,
Hugenholtz PG. Incidence of restenosis after successful
coronary angioplasty: a time-related phenomenon: a quantitative
angiographic study in 342 consecutive patients at 1, 2, 3, and 4
months. Circulation. 1988;77:361371.
3. Mintz GS, Pichard AD, Kent KM, Satler LF, Popma JJ, Leon MB. Intravascular ultrasound comparison of restenotic and de novo coronary artery narrowings. Am J Cardiol. 1994;74:12781280.[Medline] [Order article via Infotrieve]
4. Currier JW, Faxon DP. Restenosis after percutaneous transluminal coronary angioplasty: have we been aiming at the wrong target? J Am Coll Cardiol. 1995;25:516520.[Abstract]
5.
Mintz GS, Popma JJ, Pichard AD, Kent KM, Satler LF,
Wong C, Hong MK, Kovach JA, Leon MB. Arterial remodeling
after coronary angioplasty: a serial intravascular ultrasound
study. Circulation. 1996;94:3543.
6. Nobuyoshi M, Kimura T, Ohishi H, Horiuchi H, Nosaka H, Hamasaki N, Yokoi H, Kim K. Restenosis after percutaneous transluminal coronary angioplasty: pathologic observations in 20 patients. J Am Coll Cardiol. 1991;17:433439.[Abstract]
7. MacLeod DC, Strauss BH, de Jong M, Escaned J, Umans VA, van Suylen RJ, Verkerk A, de Feyter PJ, Serruys PW. Proliferation and extracellular matrix synthesis of smooth muscle cells cultured from human coronary atherosclerotic and restenotic lesions. J Am Coll Cardiol. 1994;23:5965.[Abstract]
8. Guarda E, Katwa LC, Campbell SE, Tanner MA, Webel RM, Laughlin H, Jenkins S, Myers PR. Extracellular matrix collagen synthesis and degradation following coronary balloon angioplasty. J Mol Cell Cardiol. 1996;28:699706.[Medline] [Order article via Infotrieve]
9. Hamon M, Bauters C, McFadden EP, Wernert N, Lablanche JM, Dupuis B, Bertrand ME. Restenosis after coronary angioplasty. Eur Heart J. 1995;16(suppl I):3348.
10.
Serruys PW, de Jaegere P, Kiemeneij F, Macaya C, Rutsch
W, Heyndrickx G, Emanuelsson H, Marco J, Legrand V, Materne P. A
comparison of balloon-expandable-stent implantation with balloon
angioplasty in patients with coronary artery disease: Benestent
Study Group. N Engl J Med. 1994;331:489495.
11.
Fischman DL, Leon MB, Baim DS, Schatz RA, Savage MP,
Penn I, Detre K, Veltri L, Ricci D, Nobuyoshi M. A randomized
comparison of coronary-stent placement and balloon angioplasty
in the treatment of coronary artery disease. Stent
Restenosis Study Investigators. N Engl J
Med. 1994;331:496501.
12. Haude M, Erbel R, Issa H, Meyer J. Quantitative analysis of elastic recoil after balloon angioplasty and after intracoronary implantation of balloon-expandable Palmaz-Schatz stents. J Am Coll Cardiol. 1993;21:2634.[Abstract]
13. Dussaillant GR, Mintz GS, Pichard AD, Kent KM, Satler LF, Popma JJ, Wong SC, Leon MB. Small stent size and intimal hyperplasia contribute to restenosis: a volumetric intravascular ultrasound analysis. J Am Coll Cardiol. 1995;26:720724.[Abstract]
14. Kovalic JJ, Perez CA. Radiation therapy following keloidectomy: a 20-year experience. Int J Radiat Oncol Biol Phys. 1989;17:7780.[Medline] [Order article via Infotrieve]
15. Sylvester JE, Greenberg P, Selch MT, Thomas BJ, Amstutz H. The use of postoperative irradiation for the prevention of heterotopic bone formation after total hip replacement. Int J Radiat Oncol Biol Phys. 1988;14:471476.[Medline] [Order article via Infotrieve]
16. Wiedermann JG, Marboe C, Amols H, Schwartz A, Weinberger J. Intracoronary irradiation markedly reduces restenosis after balloon angioplasty in a porcine model. J Am Coll Cardiol. 1994;23:14911498.[Abstract]
17.
Waksman R, Robinson KA, Crocker IR, Gravanis MB,
Cipolla GD, King SB 3rd. Endovascular low-dose irradiation inhibits
neointima formation after coronary artery balloon
injury in swine: a possible role for radiation therapy in
restenosis prevention. Circulation. 1995;91:15331539.
18. Wiedermann JG, Marboe C, Amols H, Schwartz A, Weinberger J. Intracoronary irradiation markedly reduces neointimal proliferation after balloon angioplasty in swine: persistent benefit at 6-month follow-up. J Am Coll Cardiol. 1995;25:14511456.[Abstract]
19. Liermann D, Bottcher HD, Kollath J, Schopohl B, Strassmann G, Strecker EP, Breddin KH. Prophylactic endovascular radiotherapy to prevent intimal hyperplasia after stent implantation in femoropopliteal arteries. Cardiovasc Intervent Radiol. 1994;17:1216.[Medline] [Order article via Infotrieve]
20.
Hehrlein C, Gollan C, Donges K, Metz J, Riessen R,
Fehsenfeld P, von Hodenberg E, Kubler W. Low-dose radioactive
endovascular stents prevent smooth muscle cell proliferation and
neointimal hyperplasia in rabbits. Circulation. 1995;92:15701575.
21.
Hehrlein C, Stintz M, Kinscherf R, Schlosser K, Huttel
E, Friedrich L, Fehsenfeld P, Kubler W. Pure ß-particle-emitting
stents inhibit neointima formation in rabbits.
Circulation. 1996;93:641645.
22.
Carter AJ, Laird JR, Bailey LR, Hoopes TG, Farb A,
Fischell DR, Fischell RE, Fischell TA, Virmani R. Effects of
endovascular radiation from a beta-particle-emitting stent in a porcine
coronary restenosis model: a dose-response study.
Circulation. 1996;94:23642368.
23. Carter AJ, Laird JR. Experimental results with endovascular irradiation via a radioactive stent. Int J Radiat Oncol Biol Phys. 1996;36:797803.[Medline] [Order article via Infotrieve]
24. Haase J, Escaned J, van Swijndregt EM, Ozaki Y, Gronenschild E, Slager CJ, Serruys PW. Experimental validation of geometric and densitometric coronary measurements on the new generation Cardiovascular Angiography Analysis System (CAAS II). Cathet Cardiovasc Diagn. 1993;30:104114.[Medline] [Order article via Infotrieve]
25. Di Mario C, Hermans WR, Rensing BJ, Serruys PW. Calibration using angiographic catheters as scaling devices: importance of filming the catheters not filled with contrast medium. Am J Cardiol. 1992;69:13771378.[Medline] [Order article via Infotrieve]
26. Serruys PW, Foley DP, de Feyter PJ. Quantitative Coronary Angiography in Clinical Practice. Dordrecht: Kluwer Academic Publishers; 1994.
27. Kuntz RE, Gibson CM, Nobuyoshi M, Baim DS. Generalized model of restenosis after conventional balloon angioplasty, stenting and directional atherectomy. J Am Coll Cardiol. 1993;21:1525.[Abstract]
28. Albiero R, Di Mario C, van der Giessen WJ, De Gregorio J, Kobayashi N, Wardeh AJ, Amato A, Coen VLMA, Serruys PW, Colombo A. Procedural results and 30-day clinical outcome after implantation of ß-particle emitting radioactive stents in human coronary arteries. Eur Heart J. 1998;19:457.
29.
Blackburn H, Keys A, Simonson E, Rautaharju P, Punsar
S. The electrocardiogram in population studies: a
classification system. Circulation. 1960;21:11601175.
30. Serruys PW, Kay P, Deshpande NV, de Feyter PJ. Periprocedural QCA following Palmaz-Schatz stent implantation predicts restenosis rate at 6 months: results of a meta-analysis of Benestent I, Benestent II pilot, Benestent II, and MUSIC trials. J Am Coll Cardiol. In press.
31.
Serruys PW, Emanuelsson H, van der Giessen W, Lunn AC,
Kiemeney F, Macaya C, Rutsch W, Heyndrickx G, Suryapranata H, Legrand
V, Goy JJ, Materne P, Bonnier H, Morice MC, Fajadet J, Belardi J,
Colombo A, Garcia E, Ruygrok P, de Jaegere P, Morel MA. Heparin-coated
Palmaz-Schatz stents in human coronary arteries: early outcome
of the Benestent-II Pilot Study. Circulation. 1996;93:412422.
32. Janicki C, Duggan DM, Coffey CW, Fischell DR, Fischell TA. Radiation dose from a phosphorous-32 impregnated wire mesh vascular stent. Med Phys. 1997;24:437445.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
M. N. Vicenzi, T. Meislitzer, B. Heitzinger, M. Halaj, L. A. Fleisher, and H. Metzler Coronary artery stenting and non-cardiac surgery--a prospective outcome study Br. J. Anaesth., June 1, 2006; 96(6): 686 - 693. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Odurny Radiological Investigation and Treatment of the Critically Ischemic Limb--A Review International Journal of Lower Extremity Wounds, March 1, 2002; 1(1): 33 - 42. [Abstract] [PDF] |
||||
![]() |
A.J. Wardeh, R. Albiero, I.P. Kay, A.H.M. Knook, W. Wijns, K. Kozuma, T. Nishida, V. Ferrero, P.C. Levendag, W.J. van der Giessen, et al. Angiographical Follow-Up After Radioactive "Cold Ends" Stent Implantation: A Multicenter Trial Circulation, February 5, 2002; 105(5): 550 - 553. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Hehrlein, J. J. DeVries, A. Arab, S. D. Haller, K. Schloesser, F. O. Tio, and T. A. Fischell Failure of a Novel Balloon-Expandable {gamma}-Emitting (103Pd) Stent to Prevent Edge Effects Circulation, November 6, 2001; 104(19): 2358 - 2362. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. A. Rosenthal, T. J. Bohlmeyer, E. Monnet, C. MacPhail, A. D. Robertson, M. A. Horwitz, J. E.B. Burchenal, and L. D. Horwitz An Iron-Binding Exochelin Prevents Restenosis Due to Coronary Artery Balloon Injury in a Porcine Model Circulation, October 30, 2001; 104(18): 2222 - 2227. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J. van der Giessen, E. Regar, M. S. Harteveld, V. L.M.A. Coen, R. Bhagwandien, A. Au, P. C. Levendag, J. Ligthart, P. W. Serruys, A. den Boer, et al. "Edge Effect" of 32P Radioactive Stents Is Caused by the Combination of Chronic Stent Injury and Radioactive Dose Falloff Circulation, October 30, 2001; 104(18): 2236 - 2241. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. P. Kay, A. J. Wardeh, K. Kozuma, G. Sianos, E. Regar, M. Knook, W. J. van der Giessen, A. Thury, J. M. R. Ligthart, V. M. A. Coen, et al. The pattern of restenosis and vascular remodelling after cold-end adioactive stent implantation Eur. Heart J., August 1, 2001; 22(15): 1311 - 1317. [Abstract] [PDF] |
||||
![]() |
A. Farb, S. Shroff, M. John, W. Sweet, and R. Virmani Late Arterial Responses (6 and 12 Months) After 32P {beta}-Emitting Stent Placement : Sustained Intimal Suppression With Incomplete Healing Circulation, April 10, 2001; 103(14): 1912 - 1919. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Seabra-Gomes Radioactive stents to reduce restenosis: time for an epitaph? Eur. Heart J., April 2, 2001; 22(8): 621 - 623. [PDF] |
||||
![]() |
A.J Wardeh, A.H.M Knook, I.P Kay, M Sabate, V.L.M.A Coen, D.P Foley, J.N Hamburger, P.C Levendag, W.J van der Giessen, and P.W Serruys Clinical and angiographical follow-up after implantation of a 6-12{micro}Ci radioactive stent in patients with coronary artery disease Eur. Heart J., April 2, 2001; 22(8): 669 - 675. [Abstract] [PDF] |
||||
![]() |
I. P. Kay, A. J. Wardeh, K. Kozuma, D. P. Foley, A. H. M. Knook, A. Thury, G. Sianos, W. J. van der Giessen, P. C. Levendag, and P. W. Serruys Radioactive Stents Delay but Do Not Prevent In-Stent Neointimal Hyperplasia Circulation, January 2, 2001; 103(1): 14 - 17. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.J. De Feyter and D. Foley Coronary stent implantation: a panacea for the interventional cardiologist? Eur. Heart J., November 1, 2000; 21(21): 1719 - 1726. [PDF] |
||||
![]() |
I. P. Kay, M. Sabate, M. A. Costa, K. Kozuma, M. Albertal, W. J. van der Giessen, A. J. Wardeh, J. M. R. Ligthart, V. M. A. Coen, P. C. Levendag, et al. Positive Geometric Vascular Remodeling Is Seen After Catheter-Based Radiation Followed by Conventional Stent Implantation but Not After Radioactive Stent Implantation Circulation, September 19, 2000; 102(12): 1434 - 1439. [Abstract] [Full Text] [PDF] |
||||
![]() |