(Circulation. 1999;100:203-208.)
© 1999 American Heart Association, Inc.
Basic Science Reports |
From the Division of Cardiology (S.A.S., T.T., F.M.) and the Department of Biomedical Engineering (J.-U.A.K., C.C.), University of Michigan, Ann Arbor.
Correspondence to S. Adam Strickberger, MD, University of Michigan Medical Center, 1500 E Medical Center Dr, Box 0022, Ann Arbor, MI 48109.
| Abstract |
|---|
|
|
|---|
Methods and ResultsTen dogs were anesthetized and underwent a thoracotomy. The heart was covered with a polyvinyl chloride membrane. The thorax above the membrane was perfused with degassed water, which functioned as a coupling medium for the ultrasound. A 7.0-MHz diagnostic ultrasound probe was affixed to a spherically focused 1.4-MHz high-intensity focused ultrasound transducer with a 1.1x8.3-mm focal zone 63.5 mm from the ablation transducer. The diagnostic ultrasound probe was calibrated such that the location of the focal zone of the ablation transducer was identifiable on the 2-dimensional ultrasound image. Target sites were identified with the diagnostic ultrasound. The maximum ultrasound intensity for ablation (2.8 kW/cm2) was delivered to the AV junction only during electrical diastole and for a total of 30 seconds. Complete AV block was achieved in each of the 10 dogs with 6.5±5.6 (range, 3 to 21) 30-second applications of therapeutic ultrasound. Gross inspection showed that the mean lesion volume was 124±143 mm3, with a depth of 6.7±3.6 mm, a length of 5.7±2.5 mm, and a width of 4.7±1.8 mm. Four hours after the dogs were killed, histopathological study demonstrated a well-demarcated area of necrosis and early inflammation.
ConclusionsHigh-intensity focused ultrasound produces well-demarcated lesions and appears to be a feasible energy source to create complete AV block within the beating heart without damaging the overlying or underlying cardiac tissue. This energy source may allow for a noninvasive approach to ablation of cardiac arrhythmias.
Key Words: ablation atrioventricular node arrhythmias
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
|
Combined Diagnostic/Ablation Ultrasound System
The diagnostic ultrasound component of the combined
ultrasound system used a conventional 2-dimensional (2D) ultrasound
machine (Diasonics VST Master Series, Diasonics/Vingmed Ultrasound Inc)
with a 7.0-MHz probe that was affixed to the ablation ultrasound
transducer (Figure 1
). The ablation ultrasound transducer was a
spherically focused single piezoelectric element (Etalon). The radius
of curvature was 63.5 mm, and the focal point was 63.5 mm
from the transducer. The operating frequency of the transducer was 1.44
MHz, and the focal zone (-6 dB) was 1.1 mm wide and 8.3 mm
long. In all experiments, the continuous-wave spatial peak temporal
average intensity was 2.8 kW/cm2. This is orders
of magnitude greater than that associated with diagnostic
and transcatheter ablation ultrasound. The spherically
focused ablation transducer exposed tissue to high-intensity focused
ultrasound energy in a cone shape. The input impedance of the ablation
transducer was electrically matched to the 50-
output impedance from
the linear amplifier. The high-intensity focused ultrasound transducer
was driven by a function generator (model 23, Wavetek) and a high-power
amplifier (300-W maximum; AP 400B, ENI). The high-power amplifier and
the function generator produced a gated sinusoidal waveform at 1.44
MHz. The ablation ultrasound transducer was positioned 63.5 mm
from the target by use of a 3-axis mechanical manipulator.
Calibration of the Ablation Ultrasound Transducer to the
Ultrasound Image
A 63.5-mm-long pointer, a length identical to the distance to
the focal point of the ablation ultrasound transducer, was attached to
the ablation ultrasound transducer. The combined ablation and imaging
fixture was placed in a degassed water bath. The location of the tip of
the pointer, relative to the diagnostic ultrasound image,
was noted on the ultrasound imaging monitor. This location was
displayed on the imaging monitor throughout the experiment and defined
the location of the focal point of the ablation transducer within the
2D ultrasound image (Figure 2
). This
allowed the identification of ablation targets by use of the ultrasound
image.
|
Ablation Protocol
The 2D ultrasound images were used to identify the AV junction
(Figure 2
). All target sites were identified directly from the
2D ultrasound image. The calibration mark indicating the focal zone of
the ablation transducer was positioned at the summit of the
interventricular septum. A computer interface (Gateway
2000) was used to control the duration and timing of each energy
application. Because of cardiac motion and the desire to create
well-defined lesions at the AV junction, applications of ablative
energy were gated to the QRS complex and delivered only during
electrical diastole, as defined by the average preablation
heart rate. Mechanical ventilation and diagnostic
ultrasound imaging were continued during applications of ablative
energy. The ablation ultrasound energy was applied for a total of 30
seconds. The total number of applications per target site (119.7±7.7)
required to achieve a total application duration of 30 seconds depended
on the sinus cycle length (384±36 ms). The initial target site was
selected at the summit of the interventricular septum. If
complete AV block did not develop with a 30-second application of
ablative ultrasound energy, a new target slightly below or adjacent to
the previous target site was selected by use of the 2D ultrasound
image, and an additional 30-second application of energy was delivered.
This was repeated until complete AV block was induced.
After completion of the experimental protocol, the animals were
euthanized by an intravenous injection of supersaturated
potassium chloride. The heart was removed, and the lesion in the AV
junction was localized grossly. The 3 dimensions of the lesion were
measured, and the lesion volume was calculated according to the
following equation, which describes an ellipsoid: Lesion
volume=(4/3)
xlength/2xdepth/2xwidth/2.
Microscopic Evaluation
Pathological examination was completed in 4 animals. Lesions
from 3 animals were sent for microscopic evaluation immediately after
euthanasia. One additional animal was killed 4 hours after complete AV
block was created to allow evolution of the lesion.
Representative blocks of tissue encompassing the lesion
and including the epicardium and the surrounding myocardium
were identified by visual inspection. These blocks were fixed in 10%
formalin and embedded in paraffin. Sections were subsequently stained
with hematoxylin and eosin.
| Results |
|---|
|
|
|---|
|
|
Gross and Microscopic Pathology
By gross inspection, each lesion was well demarcated and
surrounded by normal adjacent tissue (Figure 3
). The
myocardium between the ultrasound ablation transducer and
the lesion was unaffected (Figure 3
). Anatomically, the location
of lesions always was consistent with the target sites
identified by 2D ultrasound.
Pathological examination was completed in 4 animals. Microscopic
examination of tissue from the 3 animals killed immediately after
complete heart block was achieved showed a well-demarcated area of very
acute inflammation. The adjacent tissue was
histologically normal. Microscopic examination of
tissue from the 1 animal killed at 4 hours demonstrated a
well-demarcated lesion (Figure 4
). The
myocardium immediately adjacent to the lesion, including
the tissue between the lesion and the ablation ultrasound transducer,
was histologically normal. Within the lesion, necrosis
was clearly evident, with increased eosinophilia of myocyte cytoplasm,
prominent contraction bands, and an incipient leukocytic response
(Figure 4
).
|
| Discussion |
|---|
|
|
|---|
Pathology and Mechanism of Lesion Formation
The lesions created in this study with high-intensity focused
ultrasound were well demarcated. Similar lesions have been observed
when high-intensity focused ultrasound was used to create lesions in
noncardiac tissue.1 5 7 11 15 The pathology of the acute
lesions noted in the present study is similar to that of the
well-demarcated lesions created with radiofrequency
energy.16 17 18 This type of well-demarcated lesion is not
believed to be associated with a risk of proarrhythmia.
However, the effect of time on cardiac lesions created with
high-intensity focused ultrasound energy has not been evaluated.
Imaging of Ablation Targets and Lesions
Intracardiac ultrasound imaging has been used to direct
applications of radiofrequency energy and to observe the development of
lesions generated by radiofrequency catheter
ablation.19 20 In animals subjected to radiofrequency
catheter ablation, the lesion size noted with intracardiac ultrasound
correlates well with the lesion size observed
postmortem.20 In the present study,
diagnostic ultrasound was used to select ablation target
sites and to monitor lesion formation. Lesion formation was associated
with the development of an echo density at the target site. These
preliminary data suggest that combining diagnostic
ultrasound imaging with ultrasound ablation may provide a technique for
identifying target sites and for recognizing when an energy application
has resulted in lesion formation.
Comparison With Transcatheter Ultrasound
Ablation
Ultrasound cardiac ablation has been performed in dogs with
a 7F catheter.21 22 The lesion depth and volume were
similar to those created with radiofrequency catheter
ablation.18 21 22 Catheter-based ultrasound ablation
requires
10 MHz to produce adequate tissue absorption. Approximately
1 W of acoustic power is needed; there is direct tissue contact, and
lesion depth is
8 to 10 mm.21 22 In the
present study, target sites for ablation were selected with
conventional, noninvasive, 2D ultrasound, and tissue contact was not
required. The high-intensity ultrasound used 120 W of total acoustic
power that was focused and used to ablate tissue at a distance of 6.3
cm from the transducer without damaging the intervening tissue.
High-intensity focused ultrasound has been used in an in vitro system
to create lesions of various sizes and shapes, in a variety of
locations, within the complex anatomy of the canine
heart.23 A phased-array therapeutic ultrasound system
composed of tens or hundreds of ultrasound elements may be used to
create lesions at specific target depths up to 15 cm, without
significant heating of the intervening tissue.6
Phased-array systems may be suitable for noninvasive cardiac ablation
because of the ability to control for the position of the target site
by switching between different beam patterns at electronic speed, to
correct for aberrations due to complex inhomogeneous
intervening tissue, such as the ribs and lungs, and the ability to
change the effective aperture dimensions during treatment by adjusting
the driving signals.6 24 25
Previous Studies Using High-Intensity Ultrasound
Previous investigators have evaluated noninvasive high-intensity
focused ultrasound in vivo and in vitro for the treatment of a variety
of disorders.1 2 4 5 9 11 14 26 Catheter-based ultrasound
ablation has been used to dissolve clots in patients with acute
myocardial infarctions, and externally delivered ultrasound has proved
effective for recanalizing thrombosed ileofemoral arteries in
rabbits.26 27 The present study is the first to report
that high-intensity ultrasound energy can be focused from outside the
beating heart to a desired location within this organ without damage to
intervening tissue.
Technological Limitations of Ultrasound Ablation
A spherically focused single-element ultrasound ablation system,
as used in the present study, cannot be used to ablate cardiac
tissue through a closed chest. However, modeling studies of a human
chest phantom demonstrated the feasibility of using a phased-array
ultrasound system for precise beam formation within the thorax and
around the ribs.24 In addition, many
parameters, such as frequency, the ratio between the
aperture size and focal depth, and duration of energy application, can
affect lesion size and volume. These factors need to be more completely
understood to optimize lesion formation. In addition, lesion volume
varied between animals. Differences in biological factors may have
contributed to this variation, or perhaps the variable lesion size
resulted from applications that did not form lesions, from overlap of
successive target sites, or because the focal zone was small.
Conclusions and Implications
These results demonstrate that extracardiac application of
ablative ultrasound energy gated to the cardiac cycle and guided by
diagnostic 2D ultrasound can be accurately aimed at the AV
junction and can create complete AV block in the beating canine heart
without damaging adjacent tissue. The lesions are well-demarcated and
consistent with thermal necrosis. This technology may form the
basis of extracorporeal application of an ablative energy that can
create controlled lesions of various sizes and shapes. Theoretically, a
combined diagnostic and therapeutic phased-array ultrasound
system may allow for noninvasive ablation of cardiac
arrhythmias.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received December 22, 1998; revision received March 18, 1999; accepted March 31, 1999.
| References |
|---|
|
|
|---|
2. Goss SA, Fry FJ. High intensity ultrasonic treatment of tumors. IEEE Ultrasonics Symposium. 1984:743744.
3. Roemer R, Swindell W, Clegg S, Kress R. Simulation of a focused scanned ultrasonic heating of deep-seated tumors: the effects of blood perfusion. IEEE Trans Sonics Ultrasonics. 1984;31:457466.
4. Lele P. Effect of ultrasound on solid mammalian tissues and tumors in vivo. In: Rapacholi M, Gransdolfo M, Rindi A, eds. Ultrasound. New York, NY: Plenum Publishing Corp; 1987.
5. Yang R, Sanghvi N, Rescorla F, Kopecky K, Grosfeld J. Liver cancer ablation with extracorporeal high-intensity focused ultrasound. Eur Urol. 1993;23(suppl 1):1722.
6. Wan J, VanBaren P, Ebbini E, Cain C. Ultrasound surgery: comparison of strategies using phased array systems. IEEE Trans UFFC. 1996;10851098.
7. Sibille A, Prat F, Chapelon JY, Fadil FAE, Henry L, Theillere Y, Ponchon T, Cathignol D. Extracorporeal ablation of liver tissue by focused ultrasound. Oncology. 1993;50:375379.[Medline] [Order article via Infotrieve]
8. Hynynen K, DeYoung D, Kundrat M, Moros E. The effect of blood perfusion rate on the temperature distributions induced by multiple, scanned and focused ultrasonic beams in dogs' kidney in vivo. Int J Hyperthermia. 1989;5:485498.[Medline] [Order article via Infotrieve]
9. Dunn F, Fry FJ. Ultrasonic threshold dosages for the mammalian central nervous system. IEEE Trans Biomed Eng. 1971;18:253256.[Medline] [Order article via Infotrieve]
10. Frizzell LA. Threshold dosages for damage to mammalian liver by high-intensity focused ultrasound. IEEE Trans UFFC. 1988;35:578581.[Medline] [Order article via Infotrieve]
11. ter Haar G, Sinnett D, Rivvens I. High intensity focused ultrasound: a surgical technique for the treatment of discrete liver tumors. Phys Med Biol. 1989;34:17431750.[Medline] [Order article via Infotrieve]
12. Driller J, Lizzi FL. Therapeutic applications of ultrasound. IEEE EMB Mag. 1987;6:3340.
13. Corry P, Jabboury K, Armour E, Kong J. Human cancer treatment with ultrasound. IEEE Trans Sonics Ultrasonics. 1984;31:444456.
14. Fry FJ, Kossof G, Eggleton RC. Threshold ultrasonic dosages for structural changes in mammalian brain. J Acoust Soc Am. 1970;48:14131417.
15.
Vaughan MG, ter Haar GR, Hill CR, Clarke RL, Hopewell
JW. Minimally invasive cancer surgery using focused ultrasound: a
pre-clinical, normal tissue study. Br J Radiol. 1994;67:267274.
16. Haines DE, Watson DD. Tissue heating during radiofrequency catheter ablation: a thermodynamic model and observations in isolated perfused and superfused canine right ventricular free wall. Pacing Clin Electrophysiol. 1989;12:962977.[Medline] [Order article via Infotrieve]
17. Haverkamp W, Hindricks G, Gulker H, Rissel U, Pfennings W, Borggrefe M, Breithardt G. Coagulation of ventricular myocardium using radiofrequency alternating current: biophysical aspects and experimental findings. Pacing Clin Electrophysiol. 1989;12:187195.[Medline] [Order article via Infotrieve]
18.
Langberg JJ, Gallagher M, Strickberger SA, Amirana O.
Temperature-guided radiofrequency catheter ablation with very large
distal electrodes. Circulation. 1993;88:245249.
19.
Chu E, Fitzpatrick AP, Chin MC, Sudhir K, Yock PG, Lesh
MD. Radiofrequency catheter ablation guided by intracardiac
echocardiography. Circulation. 1994;89:13011305.
20. Kalman JM, Jue J, Sudhir K, Fitzgerald P, Yock P, Lesh MD. In vitro quantification of radiofrequency ablation lesion size using intracardiac echocardiography in dogs. Am J Cardiol. 1996;77:217219.[Medline] [Order article via Infotrieve]
21. Zimmer JE, Hynynen K, Marcus F, He DS. The feasibility of using ultrasound for cardiac ablation. IEEE Ultrasonics Symposium. 1993:12031206.
22.
He DS, Zimmer JE, Hynynen K, Marcus FI, Caruso AC,
Lampe LF, Aguirre ML. Application of ultrasound energy for intracardiac
ablation of arrhythmias. Eur Heart J. 1995;16:961966.
23. Strickberger SA, Seip R, Bogun F, Abrams GD, Ebbini E, Morady F, Cain C. Extracardiac application of high intensity focused ultrasound for ablation of ventricular myocardium. Eur J Card Pacing Electrophysiol. 1997;7:6067.
24. Kluiwstra JU, Zhang Y, Van Baren P, Strickberger SA, Ebbini ES, Cain CA. Ultrasound phased arrays for noninvasive myocardial ablation: initial studies. Proc 1994 IEEE Ultrasonics Symp. 1995:16051608.
25. Wang H, Ebbini E, O'Donnell M, Cain CA. Phase aberration correction and motion compensation for ultrasonic hyperthermia phased arrays: experimental results. IEEE Trans UFFC. 1994;41:3443.
26.
Birnbaum Y, Luo H, Nagai T, Fishbein MC,
Peterson TM, Li S, Kricsfeld D, Porter TR, Siegel RJ. Noninvasive in
vivo clot dissolution without a thrombolytic drug:
recanalization of thrombosed iliofemoral arteries
by transcutaneous ultrasound combined with intravenous
infusion of microbubbles. Circulation. 1998;97:130134.
27.
Rosenschein U, Roth A, Rassin T, Basan S,
Laniado S, Miller HI. Analysis of coronary ultrasound
thrombolysis endpoints in acute myocardial infarction
(ACUTE Trial): results of the feasibility phase.
Circulation. 1997;95:14111416.High-intensity focused
ultrasound can be applied externally to ablate internal tissue. No
published study has assessed the feasibility of ablating cardiac tissue
within the beating heart from outside the body with high-intensity
focused ultrasound. The purpose of this study was to determine whether
high-intensity focused ultrasound guided by conventional 2-dimensional
diagnostic ultrasound can be used to create complete AV
block within the beating heart. After undergoing a thoracotomy, 10 dogs
successfully underwent ablation of the AV junction with extracardiac
application of therapeutic ultrasound. Gross lesions were obvious, and
histopathological evaluation demonstrated a well-demarcated area of
necrosis. High-intensity focused ultrasound may provide a technique for
noninvasive cardiac ablation.
This article has been cited by other articles:
![]() |
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr. Surgical Treatment of Atrial Fibrillation Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394. [Full Text] |
||||
![]() |
K. Fujikura, R. Otsuka, A. Kalisz, J. A. Ketterling, Z. Jin, R. R. Sciacca, C. C. Marboe, J. Wang, R. Muratore, E. J. Feleppa, et al. Effects of ultrasonic exposure parameters on myocardial lesions induced by high-intensity focused ultrasound. J. Ultrasound Med., November 1, 2006; 25(11): 1375 - 1386. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. ter Haar High Intensity Ultrasound Surgical Innovation, March 1, 2001; 8(1): 77 - 89. [Abstract] [PDF] |
||||
![]() |
J. E. Burnes, B. Taccardi, and Y. Rudy A Noninvasive Imaging Modality for Cardiac Arrhythmias Circulation, October 24, 2000; 102(17): 2152 - 2158. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |