(Circulation. 1999;100:516-525.)
© 1999 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Mass; Broward General Medical Center, Fort Lauderdale, Fla; the Cleveland Clinic Foundation, Cleveland, Ohio; Ohio State University, Columbus, Ohio; Memorial Hospital, Colorado Springs, Colo; Mayo Clinic and Mayo Foundation, Rochester, Minn; and Spectranetics Corporation, Colorado Springs, Colo.
Correspondence to Laurence M. Epstein, MD, Cardiovascular Division, Beth Israel Deaconess Medical Center, East Campus, 330 Brookline Ave, Boston, MA 02215. E-mail lepstein{at}bidmc.harvard.edu
| Abstract |
|---|
|
|
|---|
Methods and ResultsIn this study, 863 patients underwent extraction of 1285 leads at 52 sites. Patients treated with the 14F device tended to have older leads than the 12F population; the 16F population, which comprised mostly defibrillator patients, were younger, had younger leads, and were more often male than the 12F population. Clinical success (extracting the entire lead or the lead body minus the distal electrode) was observed in 91% to 92% of cases for all device sizes. The overall complication rate was 3.6%, with 0.8% perioperative mortality. Incidence of complications was independent of laser sheath size.
ConclusionsThe 14F and 16F laser sheaths offer an extraction option for larger long-term transvenous pacemaker and defibrillator leads that is as safe and effective as the 12F laser sheath.
Key Words: heart-assist device pacemakers lasers defibrillation
| Introduction |
|---|
|
|
|---|
Although transvenous pacemaker leads have decreased in diameter over time, there are still a significant number of older, larger leads in the general population. In addition, the results of recently completed trials have expanded the indication for and may dramatically increase the implantation of ICDs. Physicians will increasingly be faced with the prospect of removing these larger transvenous leads. In this study, we report on the initial experience with larger, 14F and 16F, laser sheaths for extraction of chronic transvenous leads.
| Methods |
|---|
|
|
|---|
1-year implantation
duration were eligible for this prospective registry. Inclusion
criteria required that a lead be accessible from the subclavian,
internal or external jugular, or cephalic vein; the patient be able to
give informed written consent; mandatory or necessary indications for
lead removal existed (as described in Reference 11 ); all necessary
extraction equipment be present; and cardiothoracic surgical backup
be available. Exclusion criteria included the inability to use
fluoroscopy, a recent history of pulmonary embolus, an
unacceptable risk for emergent thoracotomy, no lead in the generator
pocket, and a lead too large for the largest laser sheath.
The 40-cm working section of the laser sheath consisted of thin inner
and outer polymer walls between which a layer of optical fibers had
been spirally wrapped.4 At the distal end of the sheath,
the fibers presented a single circumferential ring of light
sandwiched between the inner and outer walls of the tip. At the
proximal end of the sheath, the fibers passed through a connecting
cable to the XeCl excimer laser (CVX-300, Spectranetics Inc). The
excimer laser emitted 135-ns pulses of ultraviolet light (308-nm
wavelength) at a repetition rate of 40 Hz. The fluence (output energy
per unit area of fiber) at the distal tip of the device was set to 60
mJ/mm2. The laser-tissue interaction consisted of
a combination of photochemolysis and photothermal ablation, which
caused the layer of tissue immediately in contact with the device tip
to disintegrate into particles typically 5 µm in
diameter.5 Because the penetration depth of 308-nm light
in vascular tissue is
100 µm, the laser light is completely
absorbed by the tissue immediately in contact with the tip. This
produces controlled and precise removal of only the encapsulating
fibrous tissue directly surrounding the lead body that is in contact
with the tip of the laser sheath. The internal and external diameters
of the 14F and 16F laser sheaths are 10.2/14.5F and 12.5/16.8F,
respectively.
In a typical procedure, the pocket was opened, and the generator (pacemaker or ICD) was removed if present. The lead(s) was then dissected free from the scar tissue in the pocket, and a locking stylet was placed in the lead if possible. In most cases, a polymer outer sheath was preloaded over the laser sheath before the stylet and lead were threaded through the assembly. The laser and outer sheaths were passed over the lead body until the first binding site was reached. Excimer laser energy (5-second bursts) combined with gentle advancement pressure on the laser sheath and withdrawal traction on the locking stylet resulted in ablation of the encapsulating tissue and allowed the sheaths to advance to the next binding site. The sheath assembly was advanced over the lead until the lead was freed from all binding tissue or until the sheath tip reached a point a few millimeters from the heart wall. In the latter case, the outer sheath was advanced, and countertraction was applied to remove the lead. Anticoagulation if present was stopped before the procedure, and intraoperative heparin was not administered because of the risk of potential bleeding complications.
The primary end point was complete removal of the lead via the implant vein. The secondary end point was partial lead removal, leaving behind only the electrode with or without a short segment of conductor coil. Clinical success resulted when either the primary or secondary end point was reached. Procedure failure was defined as failure to extract the lead, the occurrence of a complication, or abandonment of a superior approach for a femoral or thoracic approach.
Because no conventional telescoping sheaths were available to remove the largest of the leads in this study, a randomized trial of larger laser sheaths versus conventional techniques, as in the Pacemaker Lead Extraction with the Excimer Sheath (PLEXES) trial,3 was not undertaken. After completion of the PLEXES trial, all subsequent patients undergoing extraction with all sizes of laser sheath were prospectively tracked in a registry. Registry patients treated with the 12F device were used as a control group to evaluate registry experience with the larger sheaths.
All subjects were enrolled after granting informed written consent at
sites with institutional review board approval of the study. Patient
rights were safeguarded according to the Declaration of Helsinki. Data
were recorded on paper forms and forwarded to the data coordination
center for computerized statistical analysis (SAS/Stat, SAS
Institute). A short follow-up questionnaire was completed
1 month
after the procedure after either a clinical visit or telephone contact
with the patient. Results are presented as mean±SD. During
data analysis, means of continuous variables were compared
by use of a t test at 95% CIs, whereas differences in
categorical variables were significant if a
2 test yielded P<0.05.
| Results |
|---|
|
|
|---|
|
The complete and partial success rates were similar for all 3 groups
(Table 2
). In the 14F group, the complete
and partial success rates were 86% and 5%, for an overall clinical
success rate of 91%. For the 16F group, the results were 90% and 2%,
for an overall clinical success rate of 92%. Finally, in the 12F
group, 89% of leads were completely removed and 3% were partially
removed, for an overall clinical success rate of 92%. In some cases,
initial sheath size did not allow successful extraction, and
"upsizing" to a larger sheath was required. A 14F sheath was used
for 22 leads that were initially treated with a 12F sheath.
Additionally, there were 3 instances of upsizing from a 12F to a 16F
sheath and 12 instances of upsizing from a 14F to a 16F sheath. These
cases were classified according to the largest laser sheath used. A
fluoroscopic image from a typical extraction can be seen in Figure 1
. The 16F sheath had been advanced over
the proximal portion of the distal shocking coil of an ICD lead.
Investigators were not asked to provide information on alternative
approaches when laser extraction was unsuccessful. Although this
information would have been valuable, it is unavailable.
|
|
Although a high success rate was achieved, use of the laser did not
eliminate the significant complications associated with transvenous
lead extraction. Eleven (4%), 10 (6%), and 11 (3%) patients suffered
significant complications associated with death in 4, 1, and 2 patients
in the 14F, 16F, and 12F groups, respectively (Table 2
). Cardiac
tamponade was the most common complication. occurring in 13 patients.
Additional acute complications included hemothorax caused by vascular
perforation above the pericardial reflection, air embolus, myocardial
avulsion, and vascular avulsion. The most common vascular injury was
perforation of the superior vena cava (5 total), which occurred during
attempts to reimplant new leads after extraction. Complication rates
for atrial (2.8%) and ventricular (3.2%) leads were
similar (P=0.79). In addition, there did not appear to be a
significantly increased risk of complication in the cases in which
"upsizing" of the laser sheath was required (5.4% with versus 3%
without upsizing, P=0.71), although the number of
observations is small, giving low statistical power to this
comparison.
Follow-up forms were received for 82% of patients overall, as shown in
Table 3
. The most common complications
observed at follow-up were vein thrombosis and arm swelling (7 patients
total). Pocket hematoma was observed in 2 patients, and pericardial
effusion without sequelae was seen in another 2 patients. A total of 7
patients had died at the time of follow-up, all from comorbid
conditions. Because the follow-up was short, late infections would not
be captured by this study. There were no significant differences
between groups in follow-up observations.
|
| Discussion |
|---|
|
|
|---|
There were no significant differences in success or complications for the 3 sheath sizes used in this study. The differences in patient and lead characteristics between the groups are due to the size of various leads. Larger ICD leads require the 16F sheath for extraction. Therefore, differences between this group and the others would be expected to reflect the differences between pacemaker and ICD patient populations. ICD patients are more often male, are more likely have ventricular leads, and are younger than the mean for pacemaker patients. Because transvenous ICD leads were introduced in the early 1990s, immediately before this study, the implant duration is shorter than for pacing leads. In the 14F group, the implant duration was significantly longer than for the other groups, probably because the older transvenous pacing leads were significantly larger than current leads (see Appendix). In many cases, these older leads will not fit within a 12F sheath and required the 14F sheath for extraction. A higher percentage of leads in the 12F group had active fixation because of the concomitant recall of Telectronics Accufix leads during the registry period. Despite these differences, the results were uniform with regard to efficacy and safety, suggesting that the laser sheath can be applied to a wide range of transvenous leads requiring extraction.
The need for larger laser sheaths was not limited to larger leads. In
some cases, the advancement of the sheath resulted in "snowplowing"
of the encapsulating fibrous tissue or lead insulation, leading to an
inability to advance the sheath. In rare cases, calcified fibrosis
prevented sheath advancement. The larger diameter of the 14F and 16F
sheaths can allow sheath advancement around the obstruction, with
subsequent successful extraction. An example of this technique is shown
in Figure 2
, in which a 25-year-old
biaxial transvenous lead was extracted with the 16F laser sheath. Note
the large piece of calcified fibrosis adherent to the lead remnant.
|
Regardless of technique, transvenous lead extraction carries a small but significant risk of serious complications. Cardiac perforation or tear and vascular damage are the most common serious complications. In this study, the overall acute complication rate was 3.6% per patient, or 2.5% per lead; total perioperative mortality was 0.8%. These frequencies are very close to expectations for lead removal.7 8 A significant number of complications were not related to the extraction procedure directly but were associated with lead reimplantation. It is unclear whether the laser sheath makes the vascular wall more susceptible to perforation. However, this complication can be avoided by using care and a long peel-away introducer to implant a new lead into the channel vacated by the extracted lead. Given the risk of life-threatening complications, lead extractions should be performed only with appropriate equipment and personnel required to address all potential situations, including thoracotomy, sternotomy, and cardiopulmonary bypass.
Previous Studies
In a recently reported randomized trial, the 12F laser sheath was
successful in completely removing 230 of 244 (94%) of leads and
partially removing an additional 6 (2.5%), for an overall clinical
success of 96.5%.3 Clinical success in the present
study was slightly lower at 91% and 92% for the 14F and 16F leads,
respectively. In addition, the success rate for the 12F registry was
92%. The lower overall success rate may reflect several factors. The
randomized trial included only 9 investigators experienced in lead
extraction, whereas the present study included 52 sites with
investigators having a wide range of experience. All patients meeting
the inclusion criteria for lead explant at participating sites were
randomized in the previous trial, including the leads easily removed
with merely a few moments of gentle traction. Such patients were not
included in this registry because the laser was not used; this biases
the registry toward more difficult cases. Mean implant duration in the
randomized study was 65 months, significantly shorter than for the 12F
and 14F groups reported here. Longer implant duration is typically
associated with lower explant success.7 Lead populations
also differed between the randomized trial and this
registry.
Study Limitations
At participating sites, it was not mandatory to use the laser
sheath on patients presenting for lead explant. Patients treated
with other means or referred to surgery were not followed up in this
study. It is not possible to determine from the data collected how
representative of the total pacing and ICD population
this registry might be. Limited follow-up was performed in this study.
Observation of unanticipated longer-term sequelae of laser sheath usage
remains an unanswered question. This study was undertaken to define the
value of a lead extraction tool. Additional study is required to
reassess the clinical indications for lead removal and to help balance
the risk of lead removal and its benefits.
Conclusions
Investigators chose laser sheath size according to the size of the
lead selected for explant (Appendix)
Accordingly,
lead descriptions and patient populations varied slightly for each
laser sheath size. Despite these differences, the rate of clinical
success and the incidence of complications were independent of laser
sheath size. The 14F and 16F laser sheaths are as safe and effective as
the 12F device. Because complications can be life-threatening, prompt
surgical backup is mandatory for lead extraction procedures.
|
|
|
|
|
Received December 11, 1998; revision received May 5, 1999; accepted May 5, 1999.
| References |
|---|
|
|
|---|
2. Byrd CL. Extracting chronically implanted pacemaker leads using the Spectranetics excimer laser: initial clinical experience. Pacing Clin Electrophysiol. 1996;19(suppl II):567. Abstract.
3.
Wilkoff BL, Byrd CL, Love CJ, Hayes DL, Sellers
TD, Schaerf RH, Parsonnet V, Epstein LM, Sorrentino RA, Reiser C.
Pacemaker lead extraction with the laser sheath: results of the Pacing
Lead Extraction With the Excimer Sheath (PLEXES) trial. J Am
Coll Cardiol. 1999;33:16711676.
4. Reiser C, Taylor K, Lippincott R. Large laser sheaths for pacing and defibrillator lead removal. Lasers Surg Med. 1998;22:4245.[Medline] [Order article via Infotrieve]
5. Cross FW, Bowker TJ. The physical properties of tissue ablation with excimer lasers. Med Instrum. 1987;21:226230.[Medline] [Order article via Infotrieve]
6. Wilkoff BL, Byrd CL, Sellers TD, Schaerf RH, Reiser C. Transvenous lead extraction: PLEXES trial results for larger laser sheaths. Circulation. 1997;96(suppl I):I-695. Abstract.
7. Smith HJ, Fearnot NE, Byrd CL, Wilkoff BL, Love CJ, Sellers TD. Five-years experience with intravascular lead extraction. Pacing Clin Electrophysiol. 1994;17:20162020.[Medline] [Order article via Infotrieve]
8. Wilkoff BL, Byrd CL, Love CJ, Sellers TD, Reeves RC, Kutalek SP, Turk KT, Crevey BJ, Young RA, Van Zandt HJ. Risks of intravascular extraction of chronic pacemaker and ICD leads: a multicenter analysis of 1895 patients. Pacing Clin Electrophysiol. 1998;21(suppl II):826. Abstract.
This article has been cited by other articles:
![]() |
G. M. Calvagna, R. Evola, G. Scardace, and S. Valsecchi Single-operator experience with a mechanical approach for removal of pacing and implantable defibrillator leads Europace, November 1, 2009; 11(11): 1505 - 1509. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Hamid, A. Arujna, S. Khan, A. Ladwiniec, M. McPhail, J. Bostock, M. Mobb, N. Patel, C. Bucknall, and C. A. Rinaldi Extraction of chronic pacemaker and defibrillator leads from the coronary sinus: laser infrequently used but required Europace, February 1, 2009; 11(2): 213 - 215. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. G. Bongiorni, E. Soldati, G. Zucchelli, A. Di Cori, L. Segreti, R. De Lucia, G. Solarino, A. Balbarini, M. Marzilli, and M. Mariani Transvenous removal of pacing and implantable cardiac defibrillating leads using single sheath mechanical dilatation and multiple venous approaches: high success rate and safety in more than 2000 leads Eur. Heart J., December 1, 2008; 29(23): 2886 - 2893. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Camboni, C. G. Wollmann, A. Loher, R. Gradaus, H. H. Scheld, and C. Schmid Explantation of Implantable Defibrillator Leads Using Open Heart Surgery or Percutaneous Techniques Ann. Thorac. Surg., January 1, 2008; 85(1): 50 - 55. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. M. Spotnitz Surgical Implantation of Pacemakers and Automatic Defibrillators Card. Surg. Adult, January 1, 2008; 3(2008): 1395 - 1428. [Full Text] |
||||
![]() |
B. J. Swanton, D. Keane, G. J. Vlahakes, and S. C. Streckenbach Intraoperative Transesophageal Echocardiography in the Early Detection of Acute Tamponade After Laser Extraction of a Defibrillator Lead Anesth. Analg., September 1, 2003; 97(3): 654 - 656. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. M. Spotnitz Pacemakers and Automatic Defibrillators Card. Surg. Adult, January 1, 2003; 2(2003): 1293 - 1326. [Full Text] |
||||
![]() |
M. R. Moon, C. J. Camillo, and M. J. Gleva Laser-assist during extraction of chronically implanted pacemaker and defibrillator leads Ann. Thorac. Surg., June 1, 2002; 73(6): 1893 - 1896. [Abstract] [Full Text] [PDF] |
||||
![]() |
C A Rinaldi, J Bostock, N Patel, and C A Bucknall Determinants of procedural outcome of chronically implanted pacemaker and defibrillator leads using the Excimer laser sheath Heart, February 1, 2002; 87(2): 160 - 161. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |