(Circulation. 1999;100:II-206.)
© 1999 American Heart Association, Inc.
Thoracic Transplantation and Ventricular Assist Devices |
From the University of Michigan, Ann Arbor.
Correspondence to Francis D. Pagani, MD, PhD, Assistant Professor of Surgery, Director, Heart Transplant Program, Section of Cardiac Surgery, University of Michigan, Taubman 2120, Box 0344, 1500 E Medical Center Dr, Ann Arbor, MI 48109. E-mail fpagani{at}umich.edu
| Abstract |
|---|
|
|
|---|
Methods and ResultsFrom October 1, 1996, through September 30,
1998, 32 adult patients who presented with refractory
cardiogenic shock (cardiac index <2.0 L ·
min-1 · m-2, with
systolic blood pressure <100 mm Hg and pulmonary
capillary wedge pressure
24 mm Hg and dependent on
2
inotropes with or without intra-aortic balloon pump) were
evaluated and accepted as candidates for mechanical assistance as a
bridge to transplant. Of the 32 patients, 14 (group I) had a cardiac
arrest or severe hemodynamic instability
(systolic blood pressure
75 mm Hg) with evidence of
multiorgan failure (defined as serum creatinine level >3
mg/dL or oliguria; international normalized ratio >1.5 or
transaminases >5 times normal or total bilirubin >3 mg/dL; and
needing mechanical ventilation). Group I patients were placed on ECMO
support; 7 underwent subsequent LVAD implant and 1 was bridged directly
to transplant. Six patients in group I survived to transplant
hospitalization discharge. The remaining 18 patients (group II)
underwent LVAD implant without ECMO support; 12 survived to transplant
hospitalization discharge and 2 remained alive with ongoing LVAD
support and awaited transplant. One-year actuarial survival from the
initiation of circulatory support was 43% in group I and 75% in group
II. One-year actuarial survival from the time of LVAD implant in group
I, conditional on surviving ECMO, was 71% (P=NS
compared with group II).
ConclusionsIn appropriately selected high-risk patients, the rate of LVAD survival after initial ECMO support was not significantly different from the survival rate after LVAD support alone. An initial period of resuscitation with ECMO is an effective strategy to salvage patients with extreme hemodynamic instability and multiorgan injury. Use of LVAD resources is improved by avoiding LVAD implant in a very-high-risk cohort of patients who do not survive ECMO.
Key Words: cardiopulmonary resuscitation heart-assist device transplantation shock survival heart failure
| Introduction |
|---|
|
|
|---|
For such patients, additional options for circulatory support can be provided to achieve immediate resuscitation. Extracorporeal life support (extracorporeal membrane oxygenation [ECMO]) is a well-established technique for providing emergent circulatory support for patients with respiratory failure or cardiogenic shock.6 7 ECMO has several advantages: (1) the percutaneous approach is technically simple, rapid, and can be done at the bedside; (2) it provides both cardiac and pulmonary support for hypoxic patients; (3) it avoids a sternotomy incision; (4) it provides time to evaluate potential transplant candidates; (5) it applies to patients in cardiac arrest; (6) it avoids an apical ventriculotomy in patients with a recent large anteroapical myocardial infarction; and (7) it is relatively less costly than other forms of mechanical circulatory support. Despite these advantages, ECMO support has several disadvantages that limit its applicability as long-term circulatory support for a bridge to heart transplant. These limitations include the high incidence of complications with long-term support (ie, >7 days), limited duration of support, limited rehabilitation potential, significant risk of stroke, and need for anticoagulation.
The application of ECMO for immediate resuscitation followed by implantation of a LVAD may be a useful strategy for improvement of patient survival and use of LVAD resources because it offers the option of mechanical circulatory support to patients who are initially not acceptable candidates for an implantable LVAD. We reviewed our experience with 32 adult patients who presented at the University of Michigan Hospital System from October 1, 1996, to September 30, 1998, with cardiogenic shock that required mechanical circulatory support and who were treated first with ECMO or LVAD as a mechanical bridge to transplant.
| Methods |
|---|
|
|
|---|
24 mm Hg and dependent on
2
inotropes with or without intra-aortic balloon pump), cardiac arrest,
or high risk of eminent death (ie, recurrent life-threatening
ventricular arrhythmia); (2) no existing absolute
contraindications to heart transplantation; and (3) age
65 years.
Patients were divided into 2 groups on the basis of presenting
degree of hemodynamic instability and organ injury.
Group I consisted of patients who presented with cardiac arrest
or severe hemodynamic instability (systolic
blood pressure
75 mm Hg) with evidence of multiple organ injury
(defined as serum creatinine >3 mg/dL or oliguria;
international normalized ratio (INR) >1.5 or transaminases >5 times
normal or total bilirubin >3 mg/dL; and who needed mechanical
ventilation). Group II consisted of patients presenting with
refractory cardiogenic shock with stable hemodynamics
associated with evidence of organ injury in <2 organ systems. Group I
patients were placed on circulatory support with ECMO with the intent
to bridge to an LVAD and subsequently to heart transplant when
clinically appropriate. Group II patients were placed on mechanical
circulatory support with an LVAD (HeartMate; Thermo Cardiosystems,
Inc), also with the intent to bridge to heart transplant. Arterial inflow to institute ECMO support was obtained by right carotid artery cutdown in 7 patients and percutaneous femoral artery cannulation in 7 patients. One patient required arterial reinfusion to profunda artery of the distal extremity through a side port of the arterial inflow cannulae for limb ischemia. Venous drainage was obtained by right internal jugular cutdown in 6 patients and percutaneous femoral vein cannulation in 8 patients.
Atrial septostomy was performed in 5 of 14 (36%) patients placed on ECMO support. Early in the experience, atrial septostomy was performed in 3 patients after the development of pulmonary hemorrhage; 1 patient survived ECMO to LVAD implant. On the basis of this experience, atrial septostomy was performed after initiation of ECMO when the following existed: (1) echocardiographic evidence of left ventricular dilation and (2) pulmonary hypertension (mean pulmonary artery pressure >30 mm Hg) as assessed by Swan-Ganz monitoring. With the use of these guidelines, atrial septostomy was subsequently performed in 2 additional patients: 1 survived to LVAD implant, and the other was removed from ECMO support for preexisting, absolute contraindications to transplant unknown at the time of initiation of circulatory support. No further episodes of pulmonary hemorrhage occurred on ECMO with the use of our current indications for atrial septostomy.
Comparison of means was performed using the independent sample t test and Mann-Whitney test for nonparametric data. Actuarial survival was determined using the Kaplan-Meier method. Survival curves were compared by the log-rank test. Statistical significance was defined as P<0.05 without adjustment for multiple comparisons.
| Results |
|---|
|
|
|---|
|
Of the 14 group I patients, 6 died while on ECMO support, 7 survived to LVAD placement, and 1 with a postinfarct ventricular septal defect was directly bridged from ECMO to heart transplant and survived to discharge. Of the 6 deaths, 3 occurred in patients for whom ECMO support was withdrawn because of absolute contraindications to heart transplant that were not known to be present at the initiation of ECMO support. These included brain death, severe stroke, and active intravenous drug abuse. The 3 remaining deaths were attributed to complications that occurred while the patient was on ECMO support and included pulmonary hemorrhage (n=2) and technical difficulties at initiation of ECMO support (n=1). Of the 7 patients who survived to LVAD placement, 6 were placed on the HeartMate LVAD (5 vented-electric, 1 pneumatic) and 1 was placed on the Abiomed LVAD (Abiomed, Inc). Two patients died after LVAD placement (1 on a HeartMate LVAD and 1 on the Abiomed LVAD). Both patients died from multiorgan failure (postoperative days 2 and 19, respectively). The remaining 5 patients survived to heart transplant and were discharged. The duration of ECMO support and LVAD support (mean±SD) for group I was 6±5 days and 124±97 days, respectively. The duration of LVAD support (mean±SD) for group II was 51±35 days and was significantly (P<0.05) shorter compared with group I.
A significantly higher incidence of right-sided circulatory failure occurred that required mechanical assistance (RVAD or ECMO) after LVAD implant in group I versus group II; 57% (4 of 7 patients) versus 11% (2 of 18 patients), respectively. For group I patients experiencing right-sided circulatory failure, survival after LVAD implant was 50%, compared with 100% for group II.
Three deaths occurred after LVAD placement in group II (2 from technical complications [postoperative days 1 and 3] and 1 from multiorgan failure [postoperative day 15]). Thirteen patients survived to heart transplant, with 1 postoperative death. The remaining 12 patients were discharged and remained alive as of the time this article was written. Also at this time, 2 patients are currently on LVAD support, doing well, and awaiting heart transplant.
Median duration of follow-up for the entire cohort of patients was 4.3
months (group I, 0.6 months; group II, 6.2 months; P<0.05).
Two-year actuarial survival for the entire group of patients from the
time of initiation of circulatory support was 61% (Figure 1
). One-year actuarial survival from the
initiation of any mechanical support was 43% for group I and 75% for
group II (Figure 2
). In a subgroup
analysis of group I patients, 6-month actuarial survival was
29% for patients experiencing cardiac arrest before initiation of ECMO
support versus 57% for those group I patients not in cardiac arrest at
the time of initiation of ECMO support. Of those group I patients (n=7)
who survived ECMO support to LVAD implant, conditional 1-year actuarial
survival was 71%, which was not significantly different than in group
II (Figure 3
).
|
|
|
| Discussion |
|---|
|
|
|---|
ECMO is a well-established technology that provides circulatory support
with the ability to resolve organ injury in patients who present
with cardiac arrest or with severe hemodynamic
instability associated with multiorgan failure.6 7
However, ECMO is not the only alternative to an implantable LVAD for
providing emergency resuscitation to this group of patients. Other
currently available options for emergency circulatory support include
centrifugal pumps, the Abiomed BVS 5000 (Abiomed Inc), and the Thoratec
ventricular assist device (Thoratec Laboratories Corp).
However, ECMO offers some unique advantages in that it is readily
available to rapidly resuscitate in-hospital patients with cardiac
arrest. Fifty percent (Table
) of the patients in the ECMO group in our
report were in cardiac arrest at the time of initiation of ECMO
support. Thus, instituting other forms of extracorporeal mechanical
support in this clinical scenario was not as feasible. In addition, for
our institution, ECMO is less costly than the other systems, does not
require operating room resources, and avoids a sternotomy or
ventriculotomy incision. Furthermore, no data currently suggest that
ECMO is less efficacious for providing emergency circulatory support
than the above-mentioned alternative systems.
The use of ECMO for cardiac failure in adult patients has limitations. One concern is that left ventricular decompression may be inadequate and thus result in pulmonary hypertension, edema, and hemorrhage. To address this concern, we routinely have obtained a transesophageal echocardiogram when pulmonary hypertension persists after initiation of ECMO support to assess filling of the left ventricular cavity. In cases in which the left ventriculum is not adequately decompressed, we have performed an atrial septostomy (36% of cases). The second major concern of ECMO is the limited duration of support and the poor outcomes associated with long-term support in adults.6 7 The use of ECMO as a direct bridge to heart transplantation in adult patients has been avoided at our institution because the duration of support required while patients wait for a donor organ results in an unacceptably low success rate.7 The availability of the implantable LVAD to provide a readily available long-term bridge to transplant has alleviated this last concern.
The strategy of applying initial mechanical circulatory support with
ECMO followed by bridging to an implantable LVAD appears to optimize
patient survival while negating several of the principal limitations of
each technology. Although ECMO provided for prompt resuscitation and
resolution of organ injury and provided time to perform a transplant
evaluation, the implantable LVAD provided a means for lower-risk,
long-term circulatory support and rehabilitation. This strategy
conserved LVAD resources and improved LVAD outcomes. Patients not
surviving the initial period of ECMO support in all likelihood would
not have survived LVAD support. Furthermore, in
20% of patients
placed on ECMO support, clear absolute contraindications to heart
transplant were identified. Thus, costly LVAD implantations in patients
unlikely to survive or unlikely to be transplant candidates were
avoided. Conversely, the availability of the implantable LVAD as an
option for long-term, lower-risk circulatory support prompted more
aggressive consideration and use of ECMO for patients who needed
mechanical circulatory support. At our institution, ECMO is now
routinely considered for patients experiencing in-hospital cardiac
arrest who are potential heart transplant candidates.
The most important finding of this study was that LVAD survival after ECMO support in a group of critically ill patients was not different than in a group of patients undergoing only initial LVAD support. The major presumption in this study is that patients in group I were more ill and that LVAD implant was inappropriate at the time of evaluation or that its use would have been associated with poor outcome. Data are available that support this presumption. First, patients in group I had a higher incidence of multiorgan injury, manifested by increased incidences of intubation, renal insufficiency, and shock liver. Second, if the prognostic screening scale for LVAD outcomes developed by Oz and colleagues5 is applied to the cohort of patients at the time of initial presentation, group I patients had significantly higher values, which implies a predicted worse LVAD outcome than group II patients. Our favorable post-LVAD outcomes in these ECMO patients who underwent subsequent LVAD implantation suggest that ECMO provided sufficient resuscitation and recovery from organ injury to permit LVAD implantation at lower risk. However, despite equivalent survival rates, patients who underwent LVAD implant after ECMO support had significantly increased morbidity, manifested as increased right-side circulatory failure after LVAD implant and increased duration of LVAD support required for rehabilitation. We do not understand the basis for the increased incidence of right-side circulatory failure in the ECMO-LVAD group. This increased incidence could be attributed to a lung injury exacerbated by ECMO or may reflect the overall increased severity of illness and initial degree of hemodynamic insult to this group of patients.
An important consideration in the management of patients in group I is assessment of the degree of organ recovery and timing of the LVAD implant. Due to the finite period that ECMO support can be used without causing significant morbidity in the patient, it is unreasonable to expect complete resolution of all organ injury before the LVAD implant. If the prognostic screening score of Oz et al5 is again applied to group I patients just before the LVAD implant, only modest improvement is shown in the prognostic screening score (3.5±1.6 compared with 4.1±1.2 at the time of presentation), which suggests that incomplete recovery of organ function has occurred. This is particularly true in the case of renal failure, in which it is unlikely that significant renal dysfunction will recover during a short period of ECMO support. In such cases, we have presumed renal function will return once pulsatile flow is established with a LVAD and have not used renal function as a criteria to assess transplant candidacy or timing of LVAD implant. Recovery of pulmonary and liver function is a more important factor considered in the timing of LVAD implant. We have used an INR <1.5 and liver enzymes <5 times normal as important guidelines for the timing of LVAD implant. Total bilirubin is less important an indicator of timing of LVAD implant, given that we frequently observe sustained elevations in bilirubin levels that recover after LVAD implant. We have also observed that pulmonary compliance appears to be an important determinant of outcome: in ECMO patients with a sustained deterioration in pulmonary compliance (<25 cm3/cm H2O), ECMO and LVAD outcome appears poor.
In summary, ECMO to LVAD bridge to heart transplant therapy provides an alternative means of circulatory support for patients who might not otherwise be candidates for an implantable LVAD or who represent a very high-risk for poor outcome. ECMO to LVAD bridge to heart transplant therapy improves utilization of LVAD resources by avoiding LVAD implantation in circumstances for which poor outcomes can be anticipated.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2.
Oz MC, Argenziano M, Catanese KA, Gardocki MT,
Goldstein DJ, Ashton RC, Gelijns AC, Rose EA, Levin HR. Bridge
experience with long-term implantable left ventricular
assist devices. Circulation. 1997;95:18441852.
3. Frazier OH, Macris MP, Myers TJ, Duncan M, Radovancevic B, Parnis SM, Cooley D. Improved survival after extended bridge to cardiac transplantation. Ann Thorac Surg. 1994;57:14161422.[Abstract]
4. Oz MC, Rose EA, Levin HR. Selection criteria for placement of left ventricular assist devices. Am Heart J. 1995;129:173177.[Medline] [Order article via Infotrieve]
5. Oz MC, Goldstein DJ, Pepino P, Weinberg AD, Thompson SM, Catanese KA, Vargo RL, McCarthy P, Rose EA, Levin HR. Screening scale predicts patients successfully receiving long-term implantable left ventricular assist devices. Circulation. 1995;92(suppl II):II-169II-173.
6. Anderson H, Steimle C, Shapiro M, Delius R, Chapman R, Hirschl R, Bartlett R. Extracorporeal life support for adult cardiorespiratory failure. Surgery. 1993;114:161173.[Medline] [Order article via Infotrieve]
7. Kolla S, Lee WA, Hirsch R, Bartlett R. Extracorporeal life support for cardiovascular support in adults. ASAIO J. 1996;42:M809819.[Medline] [Order article via Infotrieve]
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|