| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
(Circulation. 2000;101:2696.)
© 2000 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Division of Cardiology (J.S., M.C.J., T.F.L.), Section of Nuclear Medicine (S.K., K.S., A.B., G.v.S.), and MR Center (S.R.), Department of Radiology, University Hospital Zurich, Zurich, Switzerland, and Division of Cardiology (T.D., W.W.P.) and Department of Radiology (H.A., C.B.H.), University of California, San Francisco.
Correspondence to J. Schwitter, MD, Cardiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland. E-mail karscz{at}usz.unizh.ch
| Abstract |
|---|
|
|
|---|
Methods and ResultsSixteen healthy volunteers (age, 22 to 32 years) were examined with MR and PET in random order within 1 to 2 days. At rest and during hyperemia (dipyridamole 0.56 mg/kg), CSF was measured by a cine PC-MR technique (temporal resolution, 40 ms; spatial resolution, 1.25x0.8 mm2), and myocardial blood flow (MBF) was measured by [13N]NH3 PET. PET and MR agreed closely for coronary flow reserve (CFR; mean difference, 2.2±14.7%; Bland-Altman method). CSF divided by either total left ventricular mass or an estimate of drained myocardium (LVMdrain) correlated highly with PET flow data (r=0.93 and 0.95, respectively) and with measures of oxygen demand, ie, heart rate, afterload-corrected fiber shortening, and peak systolic stress determined by MR (overall correlation coefficients, 0.81 and 0.87, respectively, multivariate analysis). CSF/LVMdrain did not differ significantly from PET-derived MBF (difference, 3.6±16.6%). In orthotopic heart transplant recipients (n=9), CFR was reduced and blood supply-demand relationships at rest were shifted toward higher flows (P<0.0001).
ConclusionsThis integrated MR approach allows comprehensive assessment of autoregulated and hyperemic coronary flow and is suitable for serial measurements in patients. In transplanted hearts, elevated resting flow is the major cause of reduced CFR.
Key Words: magnetic resonance imaging tomography circulation transplantation
| Introduction |
|---|
|
|
|---|
Phase-contrast magnetic resonance (PC-MR) techniques have emerged recently as an attractive tool for noninvasive quantification of blood flow.7 8 9 In disease states that affect most of or the entire coronary circulation, flow measurements might be most sensitive for detection of coronary flow alterations when performed in the coronary sinus (CS), which drains a large portion of the left ventricular (LV) myocardium.10 11 With this in mind, we optimized a PC-MR sequence for this vessel to quantify flow at rest and during drug-induced hyperemia. The present study was designed to assess the accuracy of CFR determinations by PC-MR technique in comparison with PET.12 Furthermore, we hypothesized that MBF (mL · min-1 · g-1) can be quantified from PC-MR data if CS flow is related to the amount of myocardium drained by the CS and its tributaries. In addition, various determinants of oxygen demand, ie, indexes of contractility and ventricular loading, were derived from the MR data to characterize blood supply-demand relationships in normal hearts and in hypertrophied hearts of orthotopic heart transplant (OHT) recipients.13 14
| Methods |
|---|
|
|
|---|
In addition, 10 male OHT recipients were examined by MR imaging 11±4 months after transplantation. The control group (matched to donors ages and sexes) comprised 15 healthy male subjects (8 also had a PET study). In all OHT recipients, coronary angiography (performed within 6 months of the MR study) and endomyocardial biopsies (performed within 12±16 days of the MR study) revealed absence of transplant coronary artery disease and rejection (ISHT grading 2), respectively. Immunosuppressive therapy included cyclosporine (CsA), prednisone, and mycophenolate mofetil in all patients plus azathioprine in 5 patients. CsA was withheld for 12 hours before the study; all other drugs were continued. Before the MR examination, blood was drawn for CsA determination (Sandoz Pharmaceutical). Study protocols were approved by the local ethics committees, and written informed consent was obtained from all subjects. All study participants refrained from ingesting caffeinated beverages or food 24 hours before examinations.
MR Examination
Data Acquisition
Each subject was imaged in the supine position in a 1.5-T system
(Horizon Echospeed, GE Medical Systems) with a 2-coil phased array
(GP-Flex Coils, GE) as a radiofrequency receiver. For LV volume and
mass measurements, short-axis gradient echo images covering the entire
LV were acquired.13 On a localizer depicting the CS in
plane, the site of flow measurement was defined by an imaging plane
transecting the center of both the left atrium and the aortic valve,
thereby cutting the CS perpendicularly (Figure 1a
). Imaging parameters were
as follows: repetition time/echo time, 20/7.4 ms; flip angle, 15°;
retrospective ECG gating; 2 excitations averaged; respiratory
compensation (respiratory ordered-phase encoding); field of view,
20x20 cm2 with a matrix of 256x160 (linearly
interpolated before display yielding a spatial resolution of
0.8x0.8 mm2); and velocity encoding, ±0.80
and ±1.40 m/s for baseline and hyperemic conditions,
respectively. CS flow measurements were performed twice during baseline
conditions (at an interval of 5 minutes to determine short-term
reproducibility) and were repeated 4 minutes after
dipyridamole administration (0.56 mg/kg IV over 4
minutes, Perfusor VI).
|
Data Analysis
CS blood flow (mL/min) was calculated by summing the flow per
cardiac phase over the cardiac cycle and multiplying by the mean HR
during the measurement. Flow per cardiac phase equals mean velocity at
that phase times vessel area determined manually on the magnitude
images (whereby border pixels were included in the area measurement).
Phase offset errors were minimized for each cardiac phase by
subtracting phase shifts averaged from 2 regions of stationary tissue
(chest wall and skeletal muscle of the back). CFR was defined as the
ratio of hyperemic to baseline CS flow. For calculation of
CFRnorm, baseline flows were corrected for
differences in rate-pressure product (RPP) (see below).
LV mass and volumes were calculated from Simpsons
rule.13 In a first approach, a measure of MBF (mL ·
min-1 · g-1) was
obtained by dividing CS flow by total LV mass. Because the
inferior third of the interventricular septum
is almost exclusively drained by the middle cardiac
vein10 11 that enters the CS distal to the site of MR flow
measurement (see Figure 1a
) and the inferoposterior wall of the
heart is typically drained by 1 to 3 posterior
veins,10 11 15 LV mass that drains into the CS
(LVMdrain) was defined as total LV mass minus the
inferior third of the interventricular septum
minus the inferoposterior segment (extending between the site of flow
measurement and the inferior insertion of the
interventricular septum; Figure 1b
).
Indexes of Contractility and Loading
As an index of LV contractility, midwall
circumferential fiber shortening (cFS) was calculated from the MR
data14 by applying a 2-shell cylindrical
model.16 cFS was negatively correlated with
circumferential end-systolic wall stress
(cSES) (see Results section), and the slope of
the regression equation was used to correct cFS for differences in
cSES (cFSnorm) by
calculating cFS at the mean cSES of all
volunteers. Systolic wall thickening and LV ejection fraction
were calculated as described earlier.13 14
In the formulas for calculations of end-systolic circumferential17 and meridional18 LV wall stresses, cuff systolic BP was substituted for end-systolic BP as proposed by Reichek at al.18 For estimation of peak systolic stress indexes, LV dimensions and wall thicknesses were obtained at a point one third of ejection17 by linear interpolation between end diastole and end systole. Furthermore, cuff systolic BP was substituted for systolic LV pressure.
PET Examination
Dynamic PET measurements were performed with a whole-body PET
scanner (Advance, GE Medical Systems). Images were reconstructed with
filtered back projection (Hanning filter; cutoff, 5 mm
transaxial, 8.5 mm axial) and a 128x128-pixel output matrix.
Beginning with the intravenous bolus administration of 700
to 800 MBq [13N]NH3,
serial images were acquired for 15 minutes. After a delay time of
50
minutes to allow for 13N decay (physical
half-life, 9.9 minutes), hyperemia was induced by
dipyridamole (same regimen as for the MR study), and
the flow measurement was repeated. Finally, a 20-minute transmission
scan was acquired for attenuation correction.
On reformatted short-axis views, 8 regions of interest (ROIs) were
placed in 8 slices each as demonstrated in Figure 1c
. From all
64 ROIs, regional myocardial tissue time-activity curves were obtained.
The arterial input function was derived from an ROI in the
LV blood pool. As a direct estimator of MBF,
K1 (mL ·
min-1 · g-1) was
calculated from the time-activity curves with a previously validated
2-compartment model (K1,
k2, spillover
correction).12 In analogy to the
LVMdrain, as defined on the MR data set, MBF
values of ROIs 1 through 3, 7, and 8 of all 8 slices were averaged (see
Figure 1c
). This analysis was performed for baseline and
hyperemic flow states.
Both MR and PET baseline flows correlated with the RPP (see Results section), and the slope of the regression equation was used to correct flows for differences in the RPP by calculating their values at the mean RPP of all subjects (pooled MR and PET data).
Statistical Analysis
Values are given as mean±SD. Limits of agreement between PET-
and MR-derived flow and flow reserve measurements are reported as
mean±2 SD of differences calculated from paired PET and MR
measurements.19 Additionally, a repeated-measures ANOVA
was performed (2 within factors: MR versus PET and baseline versus
hyperemia), followed by paired t tests and
Bonferronis correction (P<0.05/4 is significant because
of 4 comparisons). Correlations between determinants of oxygen demand
and MBF data were sought through univariate and stepwise
linear regression analysis. Intraobserver and interobserver
variabilities and short-term reproducibility of MR flow measurements
are reported as mean±2 SD of differences of paired
analyses.19 Comparisons between OHT
recipients and control subjects were performed with unpaired Students
t tests. We considered P<0.05 to be
statistically significant.
| Results |
|---|
|
|
|---|
|
|
Comparison of MR and PET results
The hemodynamic data for the PET-MR comparison are
given in Table 1
. Figure 4
demonstrates the close agreement of the
2 techniques for measurements of CFRnorm (mean
difference, 2.2%; limits of agreement, -27.2% to 31.6%).
|
|
CS flow in these healthy volunteers increased from 68.8±16.1 mL
· min-1 at baseline to 290.2±70.2 mL ·
min-1 during hyperemia
(P<0.0001). MBF calculated as CS flow divided by total LV
mass correlated highly with PET flow data; however, the slope was
considerably less than unity (Figure 5a
and Table 1
). If CS flow was divided by
LVMdrain, the slope approached unity (Figure 5b
and Table 1
). Resting MBF correlated with RPP (slope,
7x10-5 mL ·
g-1 ·
mm Hg-1, r=0.61,
P<0.001; pooled PET and MR data). The difference for
RPP-corrected MR and PET flow data at baseline was 3.4±12.8% and was
similar for hyperemic condition (3.9±20.2%, P=NS
versus baseline; Figure 6
).
|
|
For the intraobserver variability of CS flow and CFR, the mean difference (with 95% CI in parentheses) was -0.3% (-19.7% to 19.1%) and 2.4% (-15.6% to 18.4%), respectively. For the interobserver variability, the results were 5.0% (-9.8% to 19.8%) and -1.6% (-19% to 15.8%), respectively. The short-term reproducibility for baseline CS flow measurements was 1.3% (-29.7% to 32.3%).
Determinants of Coronary Flow in Healthy
Volunteers
LV anatomical and functional data of the study population are
given in Table 2
. The negative
correlation between cFS and cSES (slope, -0.07,
r=-0.76, P<0.001) was used for calculation of
cFSnorm. Both measures of MBF (CS flow divided by
total LV mass and by LVMdrain) were positively
correlated with HR (r=0.76, P<0.001 and
r=0.77, P<0.001, respectively),
cFSnorm (r=0.54, P<0.05
and r=0.70, P<0.005, respectively), and
mSpeak (r=0.51, P<0.05 and
r=0.58, P<0.03, respectively;
univariate analysis; Figure 7a
through 7c). No correlations were
found for systolic BP, cSES, and
mSES. In a stepwise regression analysis,
HR and cFSnorm were identified as independent
predictors of baseline MBF (CS flow/LVMdrain) and
together explained 67% (adjusted R2)
of autoregulated MBF (r=0.85, P<0.0001; Figure 7d
).
|
|
Blood Supply-Demand Relationship in OHT Recipients
OHT recipients exhibited a marked concentric LV
hypertrophy, along with a reduction in LV systolic
loading and reduced indexes of contractility (Table 2
). Nevertheless, MBF (CSF/LVMtotal) per
heart beat was increased in these patients (0.011±0.001 versus
0.008±0.002 mL · g-1 ·
beat-1 in control subjects,
P<0.002). Because HR and cFSnorm were
identified as independent predictors of blood flow in normal hearts
(see above), this blood supply-demand relationship
(represented by the equation
MBF=0.0086xHR+0.029xcFSnorm-0.667) was applied
to control and transplanted hearts, yielding
MBFpredicted (MBF predicted by HR and
cFSnorm). In the control hearts,
MBFpredicted closely matched
MBFmeasured (slope, 0.997; intercept, -0.00363
mL ·
min-1 · g-1; mean
difference, 0.001±0.047 mL ·
min-1 · g-1;
P=0.92 versus 0; Figure 8
).
However, in transplanted hearts, MBFmeasured was
substantially higher than MBFpredicted
(difference, 0.378±0.123 mL ·
min-1 · g-1,
P<0.0001 versus 0; Figure 8
). This difference,
reflecting the degree of blood supply-demand imbalance, was positively
correlated (r=0.72, P=0.026) with blood CsA
(trough levels on the day of MR examination). No relations were found
for azathioprine dose, prednisone dose (mg/kg), type of
antihypertensive treatment, LV mass, and time after transplantation.
Similar results were obtained for CSF divided by
LVMdrain, with MBFmeasured
higher than MBFpredicted (1.23±0.16 versus
0.87±0.21 mL ·
min-1 · g-1,
P<0.002) and the difference correlating with CsA levels
(r=0.74, P=0.024).
|
CFR was reduced in transplanted patients compared with control subjects
(Table 3
). Further analysis
revealed a weak negative correlation between hyperemic MBF and
LV mass index (r=-0.56, P<0.005; n=24).
|
| Discussion |
|---|
|
|
|---|
CFR Measurements by Means of Cine PC-MR Technique
PET with [13N]NH3 is
well established as an accurate method for blood flow
quantification.12 20 The agreement between PC-MR and PET
measurements of CFR was excellent, with a difference of 2.2±14.7%
over a wide range of flow reserves, and the limit of agreement of
±29.4% is acceptable if one considers the reproducibility of PET for
CFR measurements of approximately ±30% (estimated from the method of
Nagamachi et al21 ).
PC-MR techniques accurately measured blood flow in the coronary arteries7 8 and CS.9 Extensive in-plane motion of the CS during the cardiac cycle and the highly pulsatile flow pattern necessitate high temporal resolution of flow data acquisition.7 In the present study, this was achieved by acquisition of k lines without segmentation. Furthermore, recent improvements in scanner hardware and software allowed us to shorten the acquisition time to assess coronary hemodynamics not only during autoregulation but also during pharmacological stimulation. Prolonged breath holds may affect the hemodynamics of the right atrium and hence could change flow dynamics in the CS. This source of error was avoided in the present study by the nonbreath hold approach. Furthermore, in hemodynamically compromised patients, data acquisition without breath holding might be advantageous.
Assessment of Myocardial Blood Supply-Demand Relationship in
Control Hearts
Blood supply to the heart was assessed by measuring CS outflow.
Both estimates of MBF, ie, CS flow divided by total LV mass or
LVMdrain, correlated highly with PET flow data
(r=0.93 and r=0.95, respectively), indicating
that both measures of MBF are useful for assessing blood supply.
Further, indexes of contractility and LV
systolic loading derived from the MR data set correlated with
both measures of MBF, indicating that the presented MR
technique provides all components necessary to establish blood
supply-demand relationships noninvasively.
It is advantageous to relate CS flow to LVMdrain because it allows direct quantitative comparison between MR and PET data. The limits of agreement for the MR and PET measurements of MBF were -29.6% and 36.8%, which are in the range for the reproducibility of PET measurements of ±25% to ±32.4% (calculated from the methods of Nagamachi et al21 and Czernin et al,5 respectively).
Uncoupling of Myocardial Blood Supply and Demand in
Transplanted Hearts
In the present study, resting MBF (CS
flow/LVMdrain) of OHT recipients was 1.23 mL
· min-1 · g-1,
which is in agreement with results of PET studies ranging from 0.94 to
1.63 mL ·
min-1 · g-1.22 23 24 25
In the present study, MBF of OHT recipients was significantly
higher than in control subjects as demonstrated in several PET
studies.22 23 24 25 Despite this increase in resting MBF,
indexes of contractility and systolic loading
were reduced in these concentrically hypertrophied hearts. Accordingly,
predicted MBF that reflects MBF dictated by oxygen demand was
significantly less than measured MBF, supporting the notion that
coupling between oxygen demand and blood supply is altered in human
denervated myocardium. Furthermore, this study provides
evidence that supply-demand uncoupling is modulated by CsA. Similarly,
in rat hearts, coronary blood flow was increased during CsA
treatment while contractile function was depressed.26 One
may speculate whether CsA-induced alterations in mitochondrial calcium
handling26 27 could be responsible for this
observation.
Study Limitations
Regional assessment of myocardial flow and flow reserve is not
achieved by the presented MR approach. However, progression of
coronary artery disease is not limited to individual
coronary arteries but is a generalized disease process of the
coronary vasculature. Therefore, serial measurements of global
CFR by the proposed technique might be useful for monitoring the
progression of coronary artery disease and assessing the
effects of drug interventions in these patients. Another disadvantage
of the presented MR technique is its inability to assess flow
and flow reserve in different myocardial layers, a limitation, however,
that is shared by most other noninvasive techniques aimed at perfusion
measurements in the heart. Whereas CFR measurements by means of the
proposed MR technique are not affected by the anatomic variability of
the coronary venous system, absolute flows (mL ·
min-1 · g-1) depend
on estimated mass of drained myocardium. Accordingly,
alterations in absolute coronary flows may be inferred from
appropriately sized patient populations rather than from a single
measurement.
Implications and Conclusions
PC-MR imaging, as presented in this study, could become a
valuable research tool thanks to its noninvasiveness and an integrated
evaluation of coronary flow and LV function. To the best of our
knowledge, this is the first report to demonstrate in humans the
capability of an integrated MR approach to quantitatively measure MBF
under resting and hyperemic conditions and to
simultaneously provide quantitative estimates of major
determinants of myocardial oxygen demand. Thus, the technique may be
ideal for studying coronary hemodynamics in
generalized diseases of the LV, such as hypertensive heart disease,
valvular heart disease, cardiomyopathies,
and others, and for assessment of therapeutic interventions in these
disease states. In OHT recipients, the presented data suggest
an uncoupling between blood supply and demand under resting conditions.
Further studies are warranted to clarify the role of cardiac
denervation and immunosuppressive therapy in this setting.
| Acknowledgments |
|---|
Received June 11, 1999; revision received December 29, 1999; accepted January 4, 2000.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. Schwitter Extending the Frontiers of Cardiac Magnetic Resonance Circulation, July 8, 2008; 118(2): 109 - 112. [Full Text] [PDF] |
||||
![]() |
H. A.J. Struijker-Boudier, A. E. Rosei, P. Bruneval, P. G. Camici, F. Christ, D. Henrion, B. I. Levy, A. Pries, and J.-L. Vanoverschelde Evaluation of the microcirculation in hypertension and cardiovascular disease Eur. Heart J., December 1, 2007; 28(23): 2834 - 2840. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Kopterides, I. Mavrou, E. Kostadima, G. Gutierrez, L. S. Chawla, and D. H. Roberts Central or Mixed Venous Oxygen Saturation? Chest, August 1, 2005; 128(2): 1073 - 1075. [Full Text] [PDF] |
||||
![]() |
J. H. Mieres, L. J. Shaw, A. Arai, M. J. Budoff, S. D. Flamm, W. G. Hundley, T. H. Marwick, L. Mosca, A. R. Patel, M. A. Quinones, et al. Role of Noninvasive Testing in the Clinical Evaluation of Women With Suspected Coronary Artery Disease: Consensus Statement From the Cardiac Imaging Committee, Council on Clinical Cardiology, and the Cardiovascular Imaging and Intervention Committee, Council on Cardiovascular Radiology and Intervention, American Heart Association Circulation, February 8, 2005; 111(5): 682 - 696. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Croteau, F. Benard, M. Bentourkia, J. Rousseau, M. Paquette, and R. Lecomte Quantitative Myocardial Perfusion and Coronary Reserve in Rats with 13N-Ammonia and Small Animal PET: Impact of Anesthesia and Pharmacologic Stress Agents J. Nucl. Med., November 1, 2004; 45(11): 1924 - 1930. [Abstract] [Full Text] [PDF] |
||||
![]() |
T.H. Giang, D. Nanz, R. Coulden, M. Friedrich, M. Graves, N. Al-Saadi, T.F. Luscher, G.K. von Schulthess, and J. Schwitter Detection of coronary artery disease by magnetic resonance myocardial perfusion imaging with various contrast medium doses: first european multi-centre experience Eur. Heart J., September 2, 2004; 25(18): 1657 - 1665. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Hagendorff, A Werner, D Pfeiffer, and H Becher Estimation of vasodilator response by analysis of Doppler intensity kinetics with myocardial contrast echocardiography using an intravenous standardized bolus administration Eur J Echocardiogr, August 1, 2004; 5(4): 272 - 283. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. K. Lund, N. Watzinger, M. Saeed, G. P. Reddy, M. Yang, P. A. Araoz, D. Curatola, M. Bedigian, and C. B. Higgins Chronic Heart Failure: Global Left Ventricular Perfusion and Coronary Flow Reserve with Velocity-encoded Cine MR Imaging: Initial Results Radiology, April 1, 2003; 227(1): 209 - 215. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Lotz, C. Meier, A. Leppert, and M. Galanski Cardiovascular Flow Measurement with Phase-Contrast MR Imaging: Basic Facts and Implementation RadioGraphics, May 1, 2002; 22(3): 651 - 671. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Koskenvuo, J. J. Hartiala, J. Knuuti, H. Sakuma, J. O. Toikka, M. Komu, M. Saraste, and P. Niemi Assessing Coronary Sinus Blood Flow in Patients with Coronary Artery Disease: A Comparison of Phase-Contrast MR Imaging with Positron Emission Tomography Am. J. Roentgenol., November 1, 2001; 177(5): 1161 - 1166. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. K. von Schulthess and J. Schwitter Cardiac MR Imaging: Facts and Fiction Radiology, February 1, 2001; 218(2): 326 - 328. [Full Text] |
||||
![]() |
G. K. Lund, M. F. Wendland, A. Shimakawa, H. Arheden, F. Ståhlberg, C. B. Higgins, and M. Saeed Coronary Sinus Flow Measurement by Means of Velocity-encoded Cine MR Imaging: Validation by Using Flow Probes in Dogs Radiology, November 1, 2000; 217(2): 487 - 493. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |