(Circulation. 2000;102:2823.)
© 2000 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Department of Internal Medicine II (S.A., D.R., T.G., W.G.D., W.M.), Institute of Medical Physics (S.U., M.K., W.A.K.), and Department of Diagnostic Radiology (U.B., W.B.), University of Erlangen-Nürnberg, Germany.
Correspondence to PD Dr Stephan Achenbach, Medizinische Klinik II mit Poliklinik, Universität Erlangen-Nürnberg, Östliche Stadtmauerstr 29, D-91054 Erlangen, Germany. E-mail stephan.achenbach{at}rzmail.uni-erlangen.de
| Abstract |
|---|
|
|
|---|
Methods and ResultsTwenty-five patients in sinus rhythm (significant coronary artery stenoses ruled out by invasive angiography) were studied with a multislice spiral CT (Siemens SOMATOM Volume Zoom). In inspiration (mean breathhold, 37 seconds), a volume data set of the heart was acquired (intravenous contrast agent; 4x1-mm slice thickness; 500-ms rotation; table feed, 1.5 mm/360°). Simultaneous recording of the ECG permitted retrospective reconstruction of contiguous cross sections in intervals of 1 mm at any desired interval of the cardiac cycle. The mean duration of the image reconstruction window was 185 ms. Next to 3-dimensional reconstructions of the heart and coronary arteries, multiplanar reconstructions were rendered to determine the visualized length of the coronary arteries, the contrast-to-noise ratio, and the correlation of coronary artery diameters to quantitative coronary angiography.
ConclusionsThe coronary arteries could be visualized over long segments (left main, 9±4 mm; left anterior descending, 112±34 mm; left circumflex, 80±29 mm; right coronary artery, 116±33 mm). On average, 78±16% of these distances were visualized free of motion artifacts. The mean contrast-to-noise ratio was 9.3±3.3. Coronary artery diameters in multislice spiral CT showed close correlation to quantitative coronary angiography (CT, 3.3±1.0 mm; angiography, 3.2±0.9 mm; mean difference, 0.38 mm; r=0.86). Contrast-enhanced multislice spiral CT permits visualization of the coronary artery lumen. Further studies are necessary to determine whether image quality is sufficient to reliably detect coronary artery stenoses.
Key Words: imaging coronary disease tomography
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
50% diameter
reduction) had been ruled out. All patients had sinus rhythm (heart
rate, 51 to 115 bpm; mean, 77 bpm) and were in a clinically stable
condition. The mean weight of the patients was 80.2 kg (range, 56 to
115 kg). All patients gave written informed consent, and the study
protocol was approved by the institutional review board.
Multislice Spiral CT
Multislice spiral CT (SOMATOM Volume Zoom, Siemens
AG) was performed 1 to 3 days after invasive angiography. Medications
were maintained between the invasive angiogram and the multislice
spiral CT study. All image acquisitions were performed in inspiratory
breathhold preceded by mild hyperventilation. First, a noncontrast
localization scan was performed that yielded an anteroposterior view of
the chest; it was used to position the imaging volume for coronary
artery imaging, which extended from the carina to
1 cm below the
diaphragmal face of the heart. In a second step, a bolus of 20 mL of
contrast agent (Iohexol, Ultravist350, Schering AG) was injected
intravenously at 4 mL/s via an 18-gauge catheter placed in the cubital
vein. After a delay of 10 seconds, a sequence of 10 axial images at the
level of the carina was acquired with an interval of 2 seconds between
subsequent images. From the time interval between contrast agent
injection and acquisition of the image with peak attenuation in the
aortic root, the contrast agent transit time was determined. In a third
step, the volume data set for coronary artery visualization was
acquired in spiral mode, with simultaneous acquisition of 4 parallel
slices with 1.0-mm collimation each. The gantry rotation time was 500
ms, and the patient table was continuously advanced at 3.0 mm/s. The
tube current was 150 mA. During the scan, the patients ECG was
digitized and continuously recorded.
Contrast agent (160 mL) was injected into the cubital vein at 4 mL/s, and the CT scan was initiated with a delay according to the previously determined contrast agent transit time. Using the PC program WinDose (Wellhöfer Dosimetrie/Scanditronix Medical), we estimated the effective radiation dose of the complete investigation to be between 3.9 mSv (men) and 5.8 mSv (women).
Image Reconstruction
The CT raw data were transferred to an offline,
PC-based workstation. Next to ungated image reconstruction with a
conventional spiral CT algorithm (180° Linear Interpolation;
180°LI),29
cross-sectional images with a slice thickness of 1.2 to 1.4 mm were
reconstructed in 1-mm intervals with the new algorithm
180°MCI.27 In
short, the algorithm collects spiral data from successive heart beats
by applying a filter function not only in the z
direction (along the patients longitudinal axis) but also in the
heart phase. The planar data set acquired by this means then undergoes
the normal, filtered backprojection process for image reconstruction in
CT. The width of the image reconstruction window is expressed through
the full-width tenth-maximum of the phase contribution
profile,27 which
strongly depends on the patients heart rate and was between 125 and
243 ms in our study (mean, 185±41 ms; see
Figure 1
). The field of view was 180 mm, with an image
matrix of 512x512 pixels. For each patient, 10 data sets were created.
For the first data set, the image acquisition window was centered
around the peak of the R wave. For the second data set, the center was
positioned at 10% of the R-R interval, and for the subsequent data
sets, the center of the image acquisition window was moved toward the
end of the cardiac cycle in increments of 10%. Because reconstruction
of one cross-sectional image required
2 seconds, 30 to 45 minutes
was necessary to create all 10 data sets. All image data sets were
inspected concerning the presence of motion artifacts, and for each
coronary artery, the data set that contained the fewest motion
artifacts was chosen for further evaluation.
|
Data Evaluation
The reconstructed images were transferred to a
commercially available image processing workstation (NetraMD, ScImage)
and evaluated by a single blinded observer. As measures of image
quality, the visualized length of each coronary artery, the
contrast-to-noise ratio (CNR), and the correlation of coronary artery
diameters in multislice spiral CT and in quantitative x-ray coronary
angiography were determined. For each artery (left main, left anterior
descending, left circumflex, and right coronary artery), a curved
multiplanar reconstruction was rendered as previously
described.30 In the
multiplanar reconstructions, the visualized length of each coronary
artery was measured. In addition, the length of the vessel segments
that were visually judged to be free of motion artifacts was
determined. The CNR was determined for each coronary artery by
measuring the mean CT density within the proximal artery lumen and the
mean CT density in a region of interest in the connective tissue
immediately next to the coronary artery and dividing the respective
difference by the image noise (SD of CT density in a region of interest
placed in the aortic root). Finally, the diameters of the left main and
proximal left anterior descending, left circumflex, and right coronary
artery were determined immediately before or after bifurcation points
that could be identified in both multislice spiral CT and invasive
coronary angiography. Diameters were measured by use of the principle
of "full-width
half-maximum":31
The multiplanar reconstructions were displayed with a window of 1
Hounsfield unit and a level that corresponded to one half of the peak
CT density measured in the proximal coronary artery segment. The length
of the shortest possible line that could be drawn across the coronary
artery lumen was considered to represent the coronary artery
diameter.
In addition, 3-dimensional reconstructions were created with the "volume rendering technique" after manual segmentation of the data sets to remove structures that would obstruct the view onto the coronary arteries, such as the chest wall, pulmonary trunk, and atrial appendages.
Quantitative Coronary Angiography
The x-ray coronary angiograms of all patients were
evaluated by means of QCA. The tip of the angiography catheter was used
for calibration, and the vessel diameters of the coronary segments
corresponding to those used for determination of the artery diameter in
multislice spiral CT were measured in 2 orthogonal planes with
automated contour detection. The mean of the 2 obtained values was
considered to represent the actual vessel diameter at that site.
Diameters were measured at the same time instant in the cardiac cycle
as in multislice spiral CT.
| Results |
|---|
|
|
|---|
Using retrospective ECG gating with the image reconstruction
algorithm 180°MCI, we could suppress cardiac motion artifacts in most
images (see
Figure 2
). The cardiac phase that yielded optimal image
quality varied strongly from patient to patient and between different
coronary arteries. For the left main and left anterior descending
coronary arteries, optimal image quality was usually achieved if the
data acquisition window was centered at 70% or 80% of the cardiac
cycle, whereas for the left circumflex and right coronary arteries, the
most frequent "optimal" position of the data acquisition window was
at 50% of the cardiac cycle (see
the
Table
). In 3-dimensional reconstructions, it was
possible to visualize the anatomy of the heart and coronary arteries
(see
Figures 3
and 4
). The mean visualized length of the left main
coronary artery was 9±4 mm (range, 4 to 17 mm). The mean visualized
length of the left anterior descending coronary was 112±34 mm (range,
52 to 155 mm); left circumflex coronary artery, 80±29 mm (range, 22 to
139 mm); and right coronary artery, 116±33 mm (range, 45 to 176 mm).
Artifacts caused by coronary artery motion were not always fully
suppressed and usually affected the mid right coronary artery and
distal segments of the left anterior descending coronary artery and
left circumflex coronary artery (see
Figure 5
). Although the left main coronary artery was free
of visible motion artifact in all cases, the mean proportion of the
left anterior descending coronary artery that was visualized free of
motion artifact was 79±16% (range, 55% to 100%). The mean
proportion of the left circumflex coronary artery that was visualized
free of motion artifact was 79±17% (range, 41% to 100%), and on
average, the right coronary artery could be visualized free of motion
artifact over 75±14% (range, 54% to 100%) of its entire
length.
|
|
|
|
|
The mean CNR determined for the proximal coronary arteries was 9.3±3.3 (left main, 8.9±4.1; left anterior descending, 9.9±3.4; left circumflex, 9.3±2.7; and right coronary artery, 9.3±2.8).
Coronary artery diameters were compared between multislice
spiral CT and QCA at 100 points (see
Figure 6
). The mean coronary artery diameter in multislice
spiral CT was 3.3±1.0 mm (range, 1.5 to 7.6 mm); the mean diameter in
QCA was 3.2±0.9 mm (range, 1.6 to 6.2 mm). The mean absolute
difference between the CT and angiographic measurement was 0.38±0.35
mm (range, 0.0 to 1.4 mm). There was no significant difference between
diameters in QCA and multislice spiral CT. The slope of the regression
line was 0.87, the intercept was 0.32, and the correlation coefficient
was 0.86.
|
| Discussion |
|---|
|
|
|---|
One major drawback of retrospectively ECG-gated spiral
CT is the fact that radiation is continuously applied, whereas image
reconstruction is restricted to data acquired during only a fraction of
the cardiac cycle. It is doubtful whether prospective triggering
algorithmsyet to be developedwould lead to a substantial reduction
in radiation exposure, because the optimal positioning of the image
reconstruction window used varies from patient to patient and between
the different coronary arteries. Based on our data, only a relatively
small portion of the cardiac cycle (
20%) could potentially be
excluded from data acquisition. Also, coronary artery visualization by
spiral CT requires intravenous injection of iodinated contrast agent,
which may lead to side effects and necessitates perfect timing of image
acquisition to contrast injection. Minor reductions in the amount of
contrast injection could be achieved by exactly adapting the amount of
contrast used to the duration of data acquisition, which depends on the
craniocaudal distance of the volume data set, instead of using a
standard dose of 160 mL as we did in our study protocol. Another
limitation is the relatively long breathhold. At the cost of increased
slice thickness, however, the speed of table movement could be
increased, leading to a shorter overall scan duration required to cover
the volume of the heart. Further investigations are necessary to
determine the optimal balance between breathhold duration and slice
thickness, which translates into image resolution along the patients
longitudinal axis.
In conclusion, multislice spiral CT with retrospective ECG gating is an interesting new approach for noninvasive coronary artery imaging. It could be demonstrated that the coronary arteries can be visualized with high spatial resolution and good overall image quality, although artifacts caused by motion, especially of the right coronary artery, can so far not be sufficiently eliminated. In the future, the accuracy of this approach concerning the detection of coronary artery stenoses will have to be determined, and the method needs to be compared against other evolving and established techniques for coronary artery visualization, such as MRI and EBT.
Received June 9, 2000; revision received July 20, 2000; accepted July 21, 2000.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Nagao, H. Matsuoka, H. Kawakami, H. Higashino, T. Mochizuki, K. Murase, and M. Uemura Quantification of Myocardial Perfusion by Contrast-Enhanced 64-MDCT: Characterization of Ischemic Myocardium Am. J. Roentgenol., July 1, 2008; 191(1): 19 - 25. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Fei, X. Du, Q. Yang, Y. Shen, P. Li, J. Liao, and K. Li 64-MDCT Coronary Angiography: Phantom Study of Effects of Vascular Attenuation on Detection of Coronary Stenosis Am. J. Roentgenol., July 1, 2008; 191(1): 43 - 49. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Voros Can Computed Tomography Angiography of the Coronary Arteries Characterize Atherosclerotic Plaque Composition?: Is the CAT (Scan) Out of the Bag? J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 183 - 185. [Full Text] [PDF] |
||||
![]() |
C. L. Hollingsworth, T. T. Yoshizumi, D. P. Frush, F. P. Chan, G. Toncheva, G. Nguyen, C. R. Lowry, and L. M. Hurwitz Pediatric Cardiac-Gated CT Angiography: Assessment of Radiation Dose Am. J. Roentgenol., July 1, 2007; 189(1): 12 - 18. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Ropers, F.-K. Pohle, A. Kuettner, T. Pflederer, K. Anders, W. G. Daniel, W. Bautz, U. Baum, and S. Achenbach Diagnostic Accuracy of Noninvasive Coronary Angiography in Patients After Bypass Surgery Using 64-Slice Spiral Computed Tomography With 330-ms Gantry Rotation Circulation, November 28, 2006; 114(22): 2334 - 2341. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach Computed Tomography Coronary Angiography J. Am. Coll. Cardiol., November 21, 2006; 48(10): 1919 - 1928. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, R. S. Blumenthal, J. J. Carr, J. G. Goldin, P. Greenland, A. D. Guerci, J. A.C. Lima, D. J. Rader, G. D. Rubin, et al. Assessment of Coronary Artery Disease by Cardiac Computed Tomography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention, Council on Cardiovascular Radiology and Intervention, and Committee on Cardiac Imaging, Council on Clinical Cardiology Circulation, October 17, 2006; 114(16): 1761 - 1791. [Full Text] [PDF] |
||||
![]() |
J. R. Crouse III Thematic review series: Patient-Oriented Research. Imaging atherosclerosis: state of the art J. Lipid Res., August 1, 2006; 47(8): 1677 - 1699. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nikolaou, A. Knez, C. Rist, B. J. Wintersperger, A. Leber, T. Johnson, M. F. Reiser, and C. R. Becker Accuracy of 64-MDCT in the diagnosis of ischemic heart disease. Am. J. Roentgenol., July 1, 2006; 187(1): 111 - 117. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Nikolaou, C. Rist, B. J. Wintersperger, T. F. Jakobs, R. van Gessel, M. A. Kirchin, A. Knez, F. von Ziegler, M. F. Reiser, and C. R. Becker Clinical value of MDCT in the diagnosis of coronary artery disease in patients with a low pretest likelihood of significant disease. Am. J. Roentgenol., June 1, 2006; 186(6): 1659 - 1668. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Zanzonico, L. N. Rothenberg, and H. W. Strauss Radiation Exposure of Computed Tomography and Direct Intracoronary Angiography: Risk Has its Reward J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1846 - 1849. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Coles, M. A. Smail, I. S. Negus, P. Wilde, M. Oberhoff, K. R. Karsch, and A. Baumbach Comparison of Radiation Doses From Multislice Computed Tomography Coronary Angiography and Conventional Diagnostic Angiography J. Am. Coll. Cardiol., May 2, 2006; 47(9): 1840 - 1845. [Abstract] [Full Text] [PDF] |
||||
![]() |
C Peebles Computed tomographic coronary angiography: how many slices do you need? Heart, May 1, 2006; 92(5): 582 - 584. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Hoffmann, F. Moselewski, K. Nieman, I.-K. Jang, M. Ferencik, A. M. Rahman, R. C. Cury, S. Abbara, H. Joneidi-Jafari, S. Achenbach, et al. Noninvasive Assessment of Plaque Morphology and Composition in Culprit and Stable Lesions in Acute Coronary Syndrome and Stable Lesions in Stable Angina by Multidetector Computed Tomography J. Am. Coll. Cardiol., April 18, 2006; 47(8): 1655 - 1662. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A.S. Cordeiro and J. A.C. Lima Atherosclerotic plaque characterization by multidetector row computed tomography angiography. J. Am. Coll. Cardiol., April 18, 2006; 47(8 Suppl): C40 - C47. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jahnke, I. Paetsch, S. Achenbach, B. Schnackenburg, R. Gebker, E. Fleck, and E. Nagel Coronary MR Imaging: Breath-hold Capability and Patterns, Coronary Artery Rest Periods, and {beta}-Blocker Use Radiology, April 1, 2006; 239(1): 71 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hausleiter, T. Meyer, M. Hadamitzky, E. Huber, M. Zankl, S. Martinoff, A. Kastrati, and A. Schomig Radiation Dose Estimates From Cardiac Multislice Computed Tomography in Daily Practice: Impact of Different Scanning Protocols on Effective Dose Estimates Circulation, March 14, 2006; 113(10): 1305 - 1310. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Alkadhi, S. Wildermuth, D. A. Bettex, A. Plass, B. Baumert, S. Leschka, L. M. Desbiolles, B. Marincek, and T. Boehm Mitral Regurgitation: Quantification with 16-Detector Row CT--Initial Experience Radiology, December 21, 2005; (2005) 2381042216. [Abstract] [Full Text] |
||||
![]() |
G. Kohl The Evolution and State-of-the-Art Principles of Multislice Computed Tomography Proceedings of the ATS, December 1, 2005; 2(6): 470 - 476. [Abstract] [Full Text] [PDF] |
||||
![]() |
N E Manghat, G J Morgan-Hughes, A J Marshall, and C A Roobottom Multidetector row computed tomography: imaging congenital coronary artery anomalies in adults Heart, December 1, 2005; 91(12): 1515 - 1522. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Achenbach, D. Ropers, F.-K. Pohle, D. Raaz, J. von Erffa, A. Yilmaz, G. Muschiol, and W. G. Daniel Detection of coronary artery stenoses using multi-detector CT with 16x0.75 collimation and 375 ms rotation Eur. Heart J., October 1, 2005; 26(19): 1978 - 1986. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Kaiser, J. Bremerich, S. Haller, H. P. Brunner-La Rocca, G. Bongartz, M. Pfisterer, and P. Buser Limited diagnostic yield of non-invasive coronary angiography by 16-slice multi-detector spiral computed tomography in routine patients referred for evaluation of coronary artery disease Eur. Heart J., October 1, 2005; 26(19): 1987 - 1992. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Alkadhi, D. Bettex, S. Wildermuth, B. Baumert, A. Plass, J. Grunenfelder, L. Desbiolles, B. Marincek, and T. Boehm Dynamic Cine Imaging of the Mitral Valve with 16-MDCT: A Feasibility Study Am. J. Roentgenol., September 1, 2005; 185(3): 636 - 646. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hacker, T. Jakobs, F. Matthiesen, C. Vollmar, K. Nikolaou, C. Becker, A. Knez, T. Pfluger, M. Reiser, K. Hahn, et al. Comparison of Spiral Multidetector CT Angiography and Myocardial Perfusion Imaging in the Noninvasive Detection of Functionally Relevant Coronary Artery Lesions: First Clinical Experiences J. Nucl. Med., August 1, 2005; 46(8): 1294 - 1300. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Leschka, H. Alkadhi, A. Plass, L. Desbiolles, J. Grunenfelder, B. Marincek, and S. Wildermuth Accuracy of MSCT coronary angiography with 64-slice technology: first experience Eur. Heart J., August 1, 2005; 26(15): 1482 - 1487. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Kim, J. Hur, S. J. Kim, H. S. Kim, B. W. Choi, K. O. Choe, Y. W. Yoon, and H. M. Kwon Two-Phase Reconstruction for the Assessment of Left Ventricular Volume and Function Using Retrospective ECG-Gated MDCT: Comparison with Echocardiography Am. J. Roentgenol., August 1, 2005; 185(2): 319 - 325. [Abstract] [Full Text] [PDF] |
||||