(Circulation. 2000;102:523.)
© 2000 American Heart Association, Inc.
Clinical Investigation and Reports |
From Stanford University Medical Center, Stanford, Calif (A.O., M.H., J.A.M., S.N.O., P.G.Y., P.J.F.); University of Texas Health Science Center, San Antonio (S.R.B.); Beth Israel Hospital, Boston, Mass (D.S.B); Yale University Hospital, New Haven, Conn (M.W.C.); The New York Hospital, New York, NY (E.D.); Georgetown University Medical Center, Washington, DC (D.J.D.); Washington Hospital Center, Washington, DC (M.B.L.); Lenox Hill Hospital, New York, NY (J.W.M.); Lubbock Medical Center, Lubbock, Tex (P.A.O.); University of Florida, Gainesville (C.J.P.); William Beaumont Hospital, Royal Oak, Mich (R.D.S.); Maimonides Medical Center, Brooklyn, NY (J.S.); The Sanger Clinic, Charlotte, NC (C.A.S.); University of Texas, Houston, (R.W.S.); the Scripps Clinic, La Jolla, Calif (P.S.T.); and Duke University Medical Center, Durham, NC (J.P.Z.).
Correspondence to Peter J. Fitzgerald, MD, PhD, Center for Research in Cardiovascular Interventions, Stanford University Medical Center, 300 Pasteur Dr, Room H3554, Stanford, CA 94305-5246. E-mail peter_fitzgerald{at}cvmed.stanford.edu
| Abstract |
|---|
|
|
|---|
Methods and ResultsThe CRUISE (Can Routine Ultrasound Influence Stent Expansion) study, a multicenter study IVUS substudy of the Stent Anti-thrombotic Regimen Study, was designed to assess the impact of IVUS on stent deployment in the high-pressure era. Nine centers were prospectively assigned to stent deployment with the use of ultrasound guidance and 7 centers to angiographic guidance alone with documentary (blinded) IVUS at the conclusion of the procedure. A total of 525 patients were enrolled with completed quantitative coronary angiography, quantitative coronary ultrasound, and clinical events adjudicated at 9 months for 499 patients. The IVUS-guided group had a larger minimal lumen diameter (2.9±0.4 versus 2.7±0.5 mm, P<0.001) by quantitative coronary angiography and a larger minimal stent area (7.78±1.72 versus 7.06±2.13 mm2, P<0.001) by quantitative coronary ultrasound. Target vessel revascularization, defined as clinically driven repeat interventional or surgical therapy of the index vessel at 9 month-follow-up, occurred significantly less frequently in the IVUS-guided group (8.5% versus 15.3%, P<0.05; relative reduction of 44%).
ConclusionsThese data suggest that ultrasound guidance of stent implantation may result in more effective stent expansion compared with angiographic guidance alone.
Key Words: stents coronary disease ultrasonics angiography restenosis
| Introduction |
|---|
|
|
|---|
The role of IVUS in the current, high-pressure era of stenting has not been clearly defined. Several studies have shown that IVUS is more accurate than angiography in determining in-stent dimensions and is better able to detect subtle findings such as incomplete apposition of the stent to the vessel wall and dissections at the stent margins.4 5 6 7 8 9 Recently, several single-center studies have demonstrated that the IVUS measurement of minimal stent area (MSA) is the single most powerful predictor of long-term patency and clinical outcome.10 11 12 13 No previous study, however, has directly addressed whether IVUS-guided stenting leads to improved results than stenting with angiographic guidance alone.
The CRUISE study (Can Routine Ultrasound Influence Stent Expansion) was designed to compare IVUS-guided versus angiographic-guided stenting in a prospective patient cohort undergoing stent deployment in the high-pressure era. CRUISE was a substudy of the Stent Anti-thrombotic Regimen Study (STARS),14 in which the use of IVUS or angiography was assigned on a center-by-center basis. The primary end point of the study was the postprocedure minimal stent dimensions as determined by angiography and intravascular ultrasound. The secondary end point was major cardiac events (death, Q-wave myocardial infarction (MI), and target vessel revascularization, TVR) at 9 months as a function of IVUS versus angiographic guidance.
| Methods |
|---|
|
|
|---|
For the CRUISE substudy, 16 centers were selected on the basis of experience with IVUS imaging in previous trials. To avoid influencing the primary randomization of the STARS trial to these angiographic protocols, the use of IVUS was assigned on a center-by-center basis (ie, each center performed either IVUS or angiographic optimization). In the angiographic-guidance centers, a blinded (documentary) IVUS examination was performed after final stent optimization. Seven centers participated in the IVUS-documentary group and 9 centers were designated as the IVUS-guided group. The centers and respective principal investigators are listed in the Appendix.
The primary end point of the CRUISE trial was the postprocedure measure of minimal stent dimensions by angiography (diameter) and ultrasound (area, minimal diameter). The secondary end point was independently adjudicated major cardiac events (death, MI, and TVR) determined by follow-up at 9 months.
Patient Selection
Inclusion criteria included symptomatic
ischemic heart disease, new lesions or restenotic
lesions of the native coronary circulation, and planned stent
implantation with up to 2 stents deployed per patient. Only
Palmaz-Schatz balloon-expandable stents were deployed. Patients were
excluded from the study if any of the following conditions occurred:
(1) the patient required revascularization of
lesions other than the stented lesion; (2) use of aspirin, ticlopidine,
or cumarin was contraindicated; (3) the presence of a left main
coronary artery lesion; (4) an MI within the past 7 days; (5)
occurrence of a stroke/transient ischemic neurological attack
within the past 3 months.
Stent Implantation Procedure
In the IVUS-documentary group (angiographic guidance), the
protocol included standard high-pressure stent deployment until
angiographic success was achieved (defined as diameter stenosis
<10% by visual assessment). After the final balloon inflation, a
blinded IVUS was performed within the target segment. This included a
slow manual or motorized pullback from
10 mm beyond the stent,
through the stent, and then through the proximal 10-mm reference
segment. Because the IVUS examination was blinded, there was no
operator response to the IVUS information.
In the IVUS-guided group, after angiographic success, iterative IVUS was performed within the target segment, and the operators used this information to optimize stent deployment. For example, the operators could perform additional inflations with the use of higher pressure, larger balloons, and/or additional stents based on iterative IVUS pullbacks.
Postprocedure Medication Protocol
After stent implantation, with or without IVUS guidance,
patients were randomly assigned to one of the following antithrombotic
regimens: aspirin alone (long-term aspirin, 325 mg), aspirin and
ticlopidine (long-term aspirin, 325 mg, and 250 BID ticlopidine for 1
month), or aspirin and cumarin (long-term aspirin, 325 mg, and adequate
cumarin to maintain INR between 2.0 to 2.5 for 4 weeks).
Quantitative Coronary Angiography
All cineangiograms were analyzed by the
Washington Hospital Center Angiographic Core Laboratory (Washington,
DC) by analysts blinded to the guidance method and postprocedure
medication protocol. Cine frames were selected from 2 views before
intervention, after stent deployment, and after final balloon
inflation. These frames were digitized and analyzed with the
use of an automated edge-detection algorithm (CAAS-II). Image
calibration was performed with the use of contrast-filled catheters as
the reference standard. The minimal lumen diameter inside and outside
the stent and reference diameter were used to calculate the percent
diameter stenosis before intervention, after stent deployment,
and after the final balloon dilatation.
IVUS Imaging Protocol
A commercially available system (CVIS/Boston Scientific Corp)
was used for all IVUS studies. The catheter contains a single-element,
30-MHz transducer mounted on the tip of a flexible shaft and rotating
at 1800 rpm within a 2.9F, rapid-exchange/common distal lumen imaging
sheath, or within a 3.2F, short monorail imaging sheath. IVUS imaging
was performed after achievement of angiographic success, and images
were recorded on half-inch, high-resolution Super-VHS (S-VHS)
videotape for off-line quantitative analysis.
IVUS Analysis
All ultrasound images were reviewed and evaluated for both
qualitative and quantitative parameters by an independent
core laboratory at the Center for Research in
Cardiovascular Interventions, Stanford University
Medical Center (Stanford, Calif). The images were digitized to perform
morphometric analysis with commercially available planimetry
software (TapeMeasure, Indec Systems, Inc). Quantitative
parameters consisted of stent and reference lumen
cross-sectional areas; stent and reference lumen minimal diameters; and
stent and reference vessel cross-sectional areas. The vessel area was
defined as the area within the media/adventitial border (that is,
including lumen, plaque, and media). Plaque area was calculated as
vessel area minus lumen or stent area.
IVUS measurements were performed at 5 cross sections in the target
segment: the tightest segment within the stent; the proximal and distal
stent edges (measured within
1 mm of the proximal and distal
stent ends); and the proximal and distal reference segments (defined as
the location in the native vessel with the least amount of disease
within 5 mm of the proximal and distal stent edges and before the
emergence of any major side branches) (Figure 1
). For the purposes of quantification,
when 2 stents in a vessel overlapped, they were treated as a single
stented segment. However, 2 nonoverlapping stents in a single vessel
were treated as 2 individually stented segments. In cases in which the
stent was placed in an ostial position, no efforts were made to measure
proximal reference segments.
|
Clinical Follow-Up
Clinical follow-up was obtained at 9 months after stent
implantation for the occurrence of major cardiac events (death, MI, or
TVR). TVR was defined as clinically driven repeat
revascularization of the initially treated target
vessel. Clinical data were independently adjudicated at the
Cardiovascular Data Analysis Center at the Beth
Israel Deaconess Medical Center (Boston, Mass).
Statistical Analysis
All clinical, angiographic, and ultrasound data were submitted
to the Cardiovascular Data Analysis Center.
Quantitative data were presented as a mean value±SD and
qualitative data were presented as frequencies. Continuous
variables were compared by means of paired or unpaired t
tests. Binary variables were examined by use of Fishers exact and
2 tests. Predictors of TVR were examined by
use of multivariate logistic models. Significance was
defined at a threshold of P=0.05. All statistical
analyses were performed with the SAS for Windows version 6.12
(SAS Institute).
| Results |
|---|
|
|
|---|
|
|
Procedure Characteristics
Procedure characteristics are listed in Table 3
. Before IVUS imaging, the average
balloon size and pressure used in the documentary group was similar to
the IVUS-guided group (balloon size 3.34+0.43 versus
3.28+0.47 mm, P=NS; balloon pressure
15.62+2.29 versus 15.91+2.70 atm,
P=NS). However, at the end of the procedure, larger balloon
and higher inflation pressure were used in the IVUS-guided group
(3.88±0.51 versus 3.69±0.59 mm, 18.0±2.58 versus 16.6±3.01
atm, respectively; both P<0.001). In the IVUS-guided group,
36% of the patients received additional therapy based on IVUS
information. In this subgroup, MSA increased from 6.25±1.39
mm2 to 7.14±1.47 mm2.
For these patients, the operators chose to use higher pressure in
59.0% of patients, larger balloon in 33.7%, and an additional stent
in 7.3%.
|
Case Examples
Figure 2
is an example of a
IVUS-documentary case. After high-pressure stent deployment, the
angiogram illustrates adequate stent expansion in the mid left
circumflex artery (Figure 2A
). However, IVUS reveals an MSA of
only 5.41 mm2 (Figure 2B
), which is
underdilated compared with a reference segment lumen area of 10.40
mm2 (Figure 2C
). In contrast, in an
IVUS-guided case shown in Figure 3
, in a
mid right coronary artery, the angiographic image after
high-pressure stent deployment appears adequate (Figure 3A
).
However, on review by IVUS, the operators deemed stent expansion not
acceptable (Figure 3B
) and proceeded to use a 0.5-mm larger
balloon size at 18 atm within the stent, achieving a nearly
3-mm2 improvement in MSA (Figure 3C
).
|
|
Angiographic and IVUS Results
Angiographic and IVUS results are shown in Table 4
. Angiographic core laboratory
analysis showed a significantly larger postprocedure minimal
lumen diameter and a significantly lower residual diameter
stenosis in the IVUS-guided group (2.9±0.4 versus
2.7±0.5 mm; P<0.001, 7.6±10.4% versus 9.8±11.2%;
P<0.05, respectively). IVUS core laboratory
analysis also showed a significantly larger postprocedure
minimal lumen area and minimal lumen diameter in the IVUS-guided group
(7.78±1.72 versus 7.06±2.13 mm2,
2.96±0.55 versus 2.59±0.43 mm, respectively; both
P<0.001).
|
Clinical Outcome at 9 Months
Clinical outcomes at 9 months are shown in Table 5
and illustrated in Figure 4
. There were no differences in the
incidence of death and MI during follow-up. However, the incidence of
TVR was 44% lower in the IVUS-guided group (8.5% versus 15.3%;
P<0.05). Interestingly, in the 64% of the patients not
receiving additional therapy in the IVUS-guided group, TVR occurred in
18 of 173 or 10.4%, which trended lower than the IVUS-documentary
group but did not reach statistical significance (P=0.15).
The potential predictors of TVR at 9-month follow-up were entered into
multivariable regression models including multivessel disease and
history of MI (which had different incidences in the 2 baseline groups;
see Table 1
) are shown in Table 6
.
The only significant predictor of TVR was postprocedure in-stent size.
|
|
|
| Discussion |
|---|
|
|
|---|
There are several possible reasons for the improved acute morphometric outcomes observed in the IVUS-guided cohort. Vessel overlap, tortuosity, and complex lesion morphology may mask subtle changes in luminal dimensions on the angiogram. In stented lesions, the metal of the struts or irregularities in the stented segment may cause QCA overestimation compared with QCU. Hoffmann et al15 showed that in 71 patients who had Palmaz-Schatz stent implantation, QCA overestimated acute lumen gain (QCA-IVUS; 0.33+0.39, P<0.0001). Blasini et al9 showed in 225 patients with Palmaz-Schatz stent implantation after low-pressure dilatation that angiography overestimated the minimal lumen diameter relative to IVUS imaging by a mean of 0.4+0.4 mm. Additionally, several studies have demonstrated that visual angiographic estimation of stenosis severity in nonstented segments results in both overestimation before intervention and underestimation after intervention.16 17 Thus operators using visual angiographic end points in the catheterization laboratory alone may tend to conclude the stenting procedure sooner without appreciating subtleties in stent underexpansion. Finally, in the presence of arterial remodeling with adaptive vessel expansion identified by IVUS, target lesions have been found to safely accommodate larger balloon sizes.18 This provides confidence in the catheterization laboratory that certain lesion subsets may be aggressively treated with balloon sizes that traditionally would be thought too injurious. Thus, iterative IVUS guidance during stent deployment augments the angiographic information and adds lesion-specific information to more efficiently "fine tune" the stent geometry for a particular lesion segment.
Recently, in the high-pressure era, several single-center studies have examined stent geometry and clinical follow-up. de Feyter et al12 showed an inverse relation between MSA as tracked by IVUS after angiographically successful deployment and TVR. In a larger single center study, Moussa et al11 also showed an inverse relation between MSA and angiographic restenosis with 425 consecutive patients (496 lesions). These data were corroborated by a further single-center study by Hoffmann et al,13 who showed that IVUS after interventional stent dimensions was one of the most consistent predictors of angiographic in-stent restenosis.
The ability to optimize stent deployment may not only improve clinical benefits but may translate to an overall cost-saving benefit at 1 year (even with the "up-front" cost of an IVUS catheter). In a recent article by Ellis et al,19 a reduction in an absolute 10% of restenosis was projected to yield an overall savings of nearly a billion dollars in the United States alone. This degree of savings could potentially justify inexpensive "on-board" guidance strategy with IVUS during initial stent placement for the coronary artery.
Even with IVUS-guided stenting, vessel size, plaque composition, and especially plaque burden limit optimal achievement of acute lumen gain. One way to achieve a larger relative lumen gain may be to debulk the lesion (directional coronary atherectomy or rotational coronary atherectomy) before stenting. Several single-center studies have shown excellent results with a combination of debulking followed by a stent.20 21 The Stenting after Optimal Lesion Debulking (SOLD) pilot study, which tested a directional coronary atherectomy before stent deployment protocol, showed an angiographic restenosis rate of 4.9% at 6-month follow-up.22 This technique may not only increase lumen gain but may also allow optimal stent expansion with low deployment pressures-capitalizing on the change in vessel compliance after primary debulking therapy.
Study Limitations
One clear limitation of the present study is the center-based
assignment to IVUS guidance as opposed to a direct randomization. This
strategy was selected because of the economy and efficiency of using a
relatively large trial apparatus (STARS) to provide
additional information without disrupting the primary randomization
scheme. There was no clear difference in prior experience with stenting
or volume of cases performed in the centers assigned for angiographic
or IVUS guidance.
A second limitation was that the optimization strategy was not standardized and was left to the discretion of the individual institutional practice. However, in this study, the incidence of IVUS-guided additional therapy (36% of patients) is consistent with recent studies (Angiography Versus Intravascular Ultrasound Directed Coronary Stent Placement [AVID] 31%, Albiero et al 44%).10 23 It is interesting to note that TVR was less (but the difference did not reach statistical significance) in the IVUS-guided group not receiving additional therapy compared with the IVUS-documentary group. This may represent bias (ie, the group assigned to IVUS may be better "stenters" as a result of their continued IVUS experience) or underscore the value of IVUS in recognizing severely underdeployed stents in the setting of adequate angiographic representation. Although this study cannot address this issue, it is an important consideration in the design of future IVUS trials.
Finally, the incidence of previous MI and triple-vessel disease was lower in the baseline IVUS-guided group than in the IVUS-documentary group. However, in multivariate regression analysis, these factors were not significant predictors of clinical outcome.
Conclusions
In the CRUISE trial, centers that used IVUS guidance achieved
significantly larger minimal stent dimensions than centers that used
angiographic guidance alone. This difference was associated with a 44%
lower rate of TVR but no difference in mortality or MI.
| Acknowledgments |
|---|
| Footnotes |
|---|
| Appendix 1 |
|---|
|
|
|---|
Received July 9, 1999; revision received February 14, 2000; accepted February 29, 2000.
| References |
|---|
|
|
|---|
2.
Serruys PW, de Jaegere P, Kiemeneij F, et al. A
comparison of balloon-expandable-stent implantation with balloon
angioplasty in patients with coronary artery disease.
N Engl J Med. 1994;331:489495.
3. Colombo A, Ferraro M, Itoh A, et al. Results of coronary stenting for restenosis. J Am Coll Cardiol. 1996;28:830836.[Abstract]
4.
Nakamura S, Colombo A, Gaglione A, et al.
Intracoronary ultrasound observations during stent
implantation. Circulation. 1994;89:20262034.
5. Goldberg SL, Colombo A, Nakamura S, et al. Benefit of intracoronary ultrasound in the deployment of Palmaz-Schatz stents. J Am Coll Cardiol. 1994;24:9961003.[Abstract]
6. Gorge G, Haude M, Ge J, et al. Intravascular ultrasound after low and high inflation pressure coronary artery stent implantation. J Am Coll Cardiol. 1995;26:725730.[Abstract]
7.
Colombo A, Hall P, Nakamura S, et al.
Intracoronary stenting without anticoagulation accomplished
with intravascular ultrasound guidance. Circulation. 1995;91:16761688.
8. Schwarzacher SP, Metz JA, Yock PG, et al. Vessel tearing at the edge of intracoronary stents detected with intravascular ultrasound imaging. Cathet Cardiovasc Diagn. 1997;40:152155.[Medline] [Order article via Infotrieve]
9. Blasini R, Neumann FJ, Schmitt C, et al. Comparison of angiography and intravascular ultrasound for the assessment of lumen size after coronary stent placement: impact of dilation pressures. Cathet Cardiovasc Diagn. 1997;42:113119.[Medline] [Order article via Infotrieve]
10.
Albiero R, Rau T, Schulter M, et al. Comparison of
immediate and intermediate-term results of intravascular ultrasound
versus angiography-guided Palmaz-Schatz stent implantation in matched
lesions. Circulation. 1997;96:29973005.
11. Moussa I, Di Mario C, Moses J, et al. Does the specific intravascular ultrasound criterion used to optimize stent expansion have an impact on the probability of stent restenosis? Am J Cardiol. 1999;83:10121017.[Medline] [Order article via Infotrieve]
12.
de Feyter PJ, Kay P, Disco C, et al. Reference chart
derived from poststent-implantation intravascular ultrasound
predictors of 6-month expected restenosis on quantitative
coronary angiography. Circulation. 1999;100:17771783.
13.
Hoffmann R, Mintz GS, Mehran R, et al. Intravascular
ultrasound predictors of angiographic restenosis in lesions
treated with Palmaz-Schatz stents. J Am Coll Cardiol. 1998;31:4349.
14.
Leon MB, Baim DS, Popma JJ, et al, for the stent
anticoagulation restenosis study investigators. N
Engl J Med. 1998;339:16651671.
15. Hoffmann R, Mintz GS, Popma JJ, et al. Overestimation of acute lumen gain and late lumen loss by quantitative coronary angiography (compared with intravascular ultrasound) in stented lesions. Am J Cardiol. 1997;80:12771281.[Medline] [Order article via Infotrieve]
16. Goldberg RK, Kleiman NS, Minor ST, et al. Comparison of quantitative coronary angiography to visual estimates of lesion severity pre and post PTCA. Am Heart J. 1990;119:178184.[Medline] [Order article via Infotrieve]
17. Bertrand ME, Lablanche JM, Bauters C, et al. Discordant results of visual and quantitative estimates of stenosis severity before and after coronary angioplasty. Cathet Cardiovasc Diagn. 1993;28:16.[Medline] [Order article via Infotrieve]
18.
Stone GW, Hodgson JM, St Goar FG, et al. Improved
procedural results of coronary angioplasty with intravascular
ultrasound-guided balloon sizing: the CLOUT pilot trial.
Circulation. 1997;95:20442052.
19.
Ellis SG, Brown KJ, Ellert R, et al. Cost of cardiac
care in the 3 years after coronary
catheterization in a contained care system: critical
determinants and implications. J Am Coll Cardiol. 1998;31:13061313.
20. Kobayashi Y, Moussa I, Akiyama T, et al. Low restenosis rate in lesions of the left anterior descending coronary artery with stenting following directional coronary atherectomy. Cathet Cardiovasc Diagn. 1998;45:131138.[Medline] [Order article via Infotrieve]
21. Hoffmann R, Mintz GS, Kent KM, et al. Comparative early and 9-month results of rotational atherectomy, stents, and the combination of both for calcified lesions in large coronary arteries. Am J Cardiol. 1998;81:552577.[Medline] [Order article via Infotrieve]
22.
Moussa I, Moses J, Di Mario C, et al. Stenting after
optimal lesion debulking (SOLD) registry: angiographic and clinical
outcome. Circulation. 1998;98:16041609.
23. Russo RJ, Attubato MJ, Davidson CJ, et al. Angiography versus intravascular ultrasound-directed stent placement: final results from AVID. Circulation. 1999;100(suppl I):I-234. Abstract.
This article has been cited by other articles:
![]() |
S. Kaul Man Must Measure: Except for Cardiologists! J. Am. Coll. Cardiol. Img., September 1, 2009; 2(9): 1111 - 1113. [Full Text] [PDF] |
||||
![]() |
G. W. Stone and G. S. Mintz Unprotected Left Main Intervention: The Light at the End of the Tunnel? Circ Cardiovasc Interv, June 1, 2009; 2(3): 156 - 158. [Full Text] [PDF] |
||||
![]() |
S.-J. Park, Y.-H. Kim, D.-W. Park, S.-W. Lee, W.-J. Kim, J. Suh, S.-C. Yun, C. W. Lee, M.-K. Hong, J.-H. Lee, et al. Impact of Intravascular Ultrasound Guidance on Long-Term Mortality in Stenting for Unprotected Left Main Coronary Artery Stenosis Circ Cardiovasc Interv, June 1, 2009; 2(3): 167 - 177. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Russo, P. D. Silva, P. S. Teirstein, M. J. Attubato, C. J. Davidson, A. C. DeFranco, P. J. Fitzgerald, S. L. Goldberg, J. B. Hermiller, M. B. Leon, et al. A Randomized Controlled Trial of Angiography Versus Intravascular Ultrasound-Directed Bare-Metal Coronary Stent Placement (The AVID Trial) Circ Cardiovasc Interv, April 1, 2009; 2(2): 113 - 123. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. V. Kelly, M. J. Gillespie, M. G. Cohen, D. P. McLaughlin, E. Magnus Ohman, and G. A. Stouffer The Contrast Media Iohexol Causes Vasoconstriction of the Proximal Left Anterior Descending Coronary Artery: Implications for Appropriate Stent Sizing Angiology, October 1, 2008; 59(5): 574 - 580. [Abstract] [PDF] |
||||
![]() |
P. Roy, D. H. Steinberg, S. J. Sushinsky, T. Okabe, T. L. Pinto Slottow, K. Kaneshige, Z. Xue, L. F. Satler, K. M. Kent, W. O. Suddath, et al. The potential clinical utility of intravascular ultrasound guidance in patients undergoing percutaneous coronary intervention with drug-eluting stents Eur. Heart J., August 1, 2008; 29(15): 1851 - 1857. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Di Mario and P. Barlis Optical Coherence Tomography: A New Tool to Detect Tissue Coverage in Drug-Eluting Stents J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 174 - 175. [Full Text] [PDF] |
||||
![]() |
B. L. van der Hoeven, S.-S. Liem, J. Dijkstra, S. C. Bergheanu, H. Putter, M. L. Antoni, D. E. Atsma, M. Bootsma, K. Zeppenfeld, J. W. Jukema, et al. Stent Malapposition After Sirolimus-Eluting and Bare-Metal Stent Implantation in Patients with ST-Segment Elevation Myocardial Infarction: Acute and 9-Month Intravascular Ultrasound Results of the MISSION! Intervention Study J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 192 - 201. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Romagnoli, G. M. Sangiorgi, J. Cosgrave, E. Guillet, and A. Colombo Drug-eluting stenting the case for post-dilation. J. Am. Coll. Cardiol. Intv., February 1, 2008; 1(1): 22 - 31. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J Russo, P. D Silva, and M. Yeager Coronary artery overexpansion increases neointimal hyperplasia after stent placement in a porcine model Heart, December 1, 2007; 93(12): 1609 - 1615. [Abstract] [Full Text] [PDF] |
||||
![]() |
A K Mitra and D K Agrawal In stent restenosis: bane of the stent era. J. Clin. Pathol., March 1, 2006; 59(3): 232 - 239. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Kruger, K.-C. Koch, I. Kaumanns, M. W. Merx, P. Hanrath, and R. Hoffmann Clinical Significance of Fractional Flow Reserve for Evaluation of Functional Lesion Severity in Stent Restenosis and Native Coronary Arteries Chest, September 1, 2005; 128(3): 1645 - 1649. [Abstract] [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, S. Silber, P. Albertsson, F. F. Aviles, P. G. Camici, A. Colombo, C. Hamm, E. Jorgensen, J. Marco, J.-E. Nordrehaug, et al. Guidelines for Percutaneous Coronary Interventions: The Task Force for Percutaneous Coronary Interventions of the European Society of Cardiology Eur. Heart J., April 2, 2005; 26(8): 804 - 847. [Full Text] [PDF] |
||||
![]() |
T. H. Lee, D. H. Kim, B.-H. Lee, H. J. Kim, C. H. Choi, K. P. Park, D. S. Jung, S. Kim, and T. Y. Moon Preliminary Results of Endovascular Stent-Assisted Angioplasty for Symptomatic Middle Cerebral Artery Stenosis AJNR Am. J. Neuroradiol., January 1, 2005; 26(1): 166 - 174. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Sahara, H. Kirigaya, Y. Oikawa, J. Yajima, K. Nagashima, H. Hara, K. Ogasawara, and T. Aizawa Soft plaque detected on intravascular ultrasound is the strongest predictor of in-stent restenosis: an intravascular ultrasound study Eur. Heart J., November 2, 2004; 25(22): 2026 - 2033. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Orford, A. Lerman, and D. R. Holmes Routine intravascular ultrasound guidance of percutaneous coronary intervention: A critical reappraisal J. Am. Coll. Cardiol., April 21, 2004; 43(8): 1335 - 1342. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Sahara, H. Kirigaya, Y. Oikawa, J. Yajima, K. Ogasawara, H. Satoh, K. Nagashima, H. Hara, Y. Nakatsu, and T. Aizawa Arterial remodeling patterns before intervention predict diffuse in-stent restenosis: An intravascular ultrasound study J. Am. Coll. Cardiol., November 19, 2003; 42(10): 1731 - 1738. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. J. Weissman Vascular remodeling: do we really need yet another study? J. Am. Coll. Cardiol., September 3, 2003; 42(5): 811 - 813. [Full Text] [PDF] |
||||
![]() |
A L Gaster, U Slothuus Skjoldborg, J Larsen, L Korsholm, C von Birgelen, S Jensen, P Thayssen, K E Pedersen, and T H Haghfelt Continued improvement of clinical outcome and cost effectiveness following intravascular ultrasound guided PCI: insights from a prospective, randomised study Heart, September 1, 2003; 89(9): 1043 - 1049. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Cheneau, L. Leborgne, G. S. Mintz, J.-i. Kotani, A. D. Pichard, L. F. Satler, D. Canos, M. Castagna, N. J. Weissman, and R. Waksman Predictors of Subacute Stent Thrombosis: Results of a Systematic Intravascular Ultrasound Study Circulation, July 8, 2003; 108(1): 43 - 47. [Abstract] [Full Text] [PDF] |
||||
![]() |
B E Bouma, G J Tearney, H Yabushita, M Shishkov, C R Kauffman, D DeJoseph Gauthier, B D MacNeill, S L Houser, H T Aretz, E F Halpern, et al. Evaluation of intracoronary stenting by intravascular optical coherence tomography Heart, March 1, 2003; 89(3): 317 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. V. Oemrawsingh, G. S. Mintz, M. J. Schalij, A. H. Zwinderman, J. W. Jukema, and E. E.v.d. Wall Intravascular Ultrasound Guidance Improves Angiographic and Clinical Outcome of Stent Implantation for Long Coronary Artery Stenoses: Final Results of a Randomized Comparison With Angiographic Guidance (TULIP Study) Circulation, January 7, 2003; 107(1): 62 - 67. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Spanos, G. Stankovic, J. Tobis, and A. Colombo The challenge of in-stent restenosis: insights from intravascular ultrasound Eur. Heart J., January 2, 2003; 24(2): 138 - 150. [Full Text] [PDF] |
||||
![]() |
N. J. Goswami, J. M. Moody Jr, and S. R. Bailey Percutaneous Mechanical Reperfusion During Acute Myocardial Infarction J Intensive Care Med, July 1, 2002; 17(4): 162 - 173. [Abstract] [PDF] |
||||
![]() |
N. H.J. Pijls, V. Klauss, U. Siebert, E. Powers, K. Takazawa, W. F. Fearon, J. Escaned, Y. Tsurumi, T. Akasaka, H. Samady, et al. Coronary Pressure Measurement After Stenting Predicts Adverse Events at Follow-Up: A Multicenter Registry Circulation, June 25, 2002; 105(25): 2950 - 2954. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Dangas and F. Kuepper Restenosis: Repeat Narrowing of a Coronary Artery: Prevention and Treatment Circulation, June 4, 2002; 105(22): 2586 - 2587. [Full Text] [PDF] |
||||
![]() |
M. Voskuil, R. A. M. van Liebergen, M. Albertal, E. Boersma, J. G. P. Tijssen, P. W. Serruys, J. J. Piek, and the DEBATE II Investigators Coronary hemodynamics of stent implantation after suboptimal and optimal balloon angioplasty J. Am. Coll. Cardiol., May 1, 2002; 39(9): 1513 - 1517. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Briguori, J. Tobis, T. Nishida, M. Vaghetti, R. Albiero, C. Di Mario, and A. Colombo Discrepancy between angiography and intravascular ultrasound when analysing small coronary arteries Eur. Heart J., February 1, 2002; 23(3): 247 - 254. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Kern The meaning of suboptimal coronary flow reserve after coronary balloon angioplasty Eur. Heart J., January 2, 2002; 23(2): 99 - 100. [Full Text] [PDF] |
||||
![]() |
N.G. Uren, S.P. Schwarzacher, J.A. Metz, D.P. Lee, Y. Honda, A.C. Yeung, P.J. Fitzgerald, and P.G. Yock Predictors and outcomes of stent thrombosis. An intravascular ultrasound registry Eur. Heart J., January 2, 2002; 23(2): 124 - 132. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Takano, L. A. Yeatman, J. R. Higgins, J. W. Currier, E. Ascencio, K. A. Kopelson, and J. M. Tobis Optimizing stent expansion with new stent delivery systems J. Am. Coll. Cardiol., November 15, 2001; 38(6): 1622 - 1627. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Hoffmann, P. Haager, G.S. Mintz, G. Kerckhoff, R. Schwarz, A. Franke, J. Vom Dahl, and P. Hanrath The impact of high pressure vs low pressure stent implantation on intimal hyperplasia and follow-up lumen dimensions; results of a randomized trial Eur. Heart J., November 1, 2001; 22(21): 2015 - 2024. [Abstract] [PDF] |
||||
![]() |
W. F. Fearon, J. Luna, H. Samady, E. R. Powers, T. Feldman, N. Dib, E. M. Tuzcu, M. W. Cleman, T. M. Chou, D. J. Cohen, et al. Fractional Flow Reserve Compared With Intravascular Ultrasound Guidance for Optimizing Stent Deployment Circulation, October 16, 2001; 104(16): 1917 - 1922. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Kern and B. Meier Evaluation of the Culprit Plaque and the Physiological Significance of Coronary Atherosclerotic Narrowings Circulation, June 26, 2001; 103(25): 3142 - 3149. [Full Text] [PDF] |
||||
![]() |
H. Schuhlen, A. Kastrati, J.u. Pache, J. Dirschinger, and A. Schomig Incidence of thrombotic occlusion and major adverse cardiac events between two and four weeks after coronary stent placement: analysis of 5,678 patients with a four-week ticlopidine regimen J. Am. Coll. Cardiol., June 15, 2001; 37(8): 2066 - 2073. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Smith Jr, J. T. Dove, A. K. Jacobs, J. Ward Kennedy, D. Kereiakes, M. J. Kern, R. E. Kuntz, J. J. Popma, H. V. Schaff, D. O. Williams, et al. ACC/AHA guidelines for percutaneous coronary intervention (revision of the 1993 PTCA guidelines): A report of the American College of Cardiology/ American Heart Association Task Force on practice guidelines (Committee to revise the 1993 guidelines for percutaneous transluminal coronary angioplasty) endorsed by the Society for Cardiac Angiography and Interventions J. Am. Coll. Cardiol., June 15, 2001; 37(8): 2239 - 2239. [Full Text] [PDF] |
||||
![]() |
A L MCLEOD, D B NORTHRIDGE, and N G UREN Ultrasound guided stenting Heart, June 1, 2001; 85(6): 605 - 606. [Full Text] |
||||
![]() |
F. Schiele Intravascular Ultrasound Guidance for Stent Implantation Circulation, May 29, 2001; 103 (21): e110 - e110. [Full Text] [PDF] |
||||
![]() |
H. Okura, Y. Morino, A. Oshima, M. Hayase, M. R. Ward, J. J. Popma, R. E. Kuntz, H. N. Bonneau, P. G. Yock, and P. J. Fitzgerald Preintervention arterial remodeling affects clinical outcome following stenting: an intravascular ultrasound study J. Am. Coll. Cardiol., March 15, 2001; 37(4): 1031 - 1035. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Nissen and P. Yock Intravascular Ultrasound : Novel Pathophysiological Insights and Current Clinical Applications Circulation, January 30, 2001; 103(4): 604 - 616. [Abstract] [Full Text] [PDF] |
||||
![]() |
IVUS Improves Stent Expansion Journal Watch Cardiology, October 13, 2000; 2000(1013): 3 - 3. [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2000 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |