(Circulation. 2001;103:2159.)
© 2001 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Division of Cardiology, Department of Medicine, Cedars-Sinai Medical Center, Los Angeles, Calif.
Correspondence to Charles D. Swerdlow, MD, 8635 W Third Street, Ste 1190W, Los Angeles, CA 90048. E-mail swerdlow{at}ucla.edu
| Abstract |
|---|
|
|
|---|
Methods and
ResultsImplantable
cardioverter-defibrillators were implanted in 80 patients if T-wave
shocks did not induce VF and the baseline-rhythm R wave was
7 mV. The
T-wave shock was 10 J in the first 45 patients (group A) and 15 J in
the last 35 patients (group B). After inductionless implantations, the
first VF shock was programmed to the T-wave shock plus 5 J. If T-wave
shocks induced VF, the ULV was measured and the first shock was
programmed to the ULV+5 J. Inductionless implantations were
performed in 58 patients (72%), 28 in group A (62%) and 30 in group B
(86%; P=0.04). If T-wave
scanning had been done at 15 J in group A patients, inductionless
implantations could have been performed in 84% of them. At 3 months,
VF was induced twice during
electrophysiological study in 75 patients
(94%). All VFs were detected in
4.7 s and were terminated by the
first shock. During follow-up, 197 of 198 appropriate first shocks for
rapid ventricular tachycardia or VF (99%) were
successful in patients who had inductionless implantations (95%
confidence intervals, 97% to 100%).
ConclusionInductionless implantations can be performed in >80% of implantable cardioverter-defibrillator recipients using a vulnerability safety margin based on a T-wave scan at 15 J.
Key Words: defibrillation defibrillators, implantable shock
| Introduction |
|---|
|
|
|---|
The ULV is the weakest shock at which VF is not induced when a shock is delivered during the vulnerable phase of the cardiac cycle. The ULV hypothesis of defibrillation links the ULV to the minimum shock strength that defibrillates reliably.6 The ULV approximates the shock strength that defibrillates with a 90% probability of success.7 Shocks programmed to the ULV+3 J or ULV+5 J defibrillate induced VF consistently, both acutely7 and chronically.8 Shocks at the ULV+5 J convert spontaneous ICD-detected VF reliably.9
This studys primary end points were detection and defibrillation of induced VF 3 months after implantation. Secondary end points were the detection and defibrillation of spontaneous, rapid ventricular tachycardia (VT) and VF.
| Methods |
|---|
|
|
|---|
|
Implant Procedure
Technique
An active-can ICD (Medtronic model 7227, 7229, 7271, or 7273) and a dual-coil, integrated-bipolar
ventricular defibrillation lead with a sensing
interelectrode distance of 12 mm (Medtronic model 6942 or 6945) were inserted through a left pectoral incision, as
described previously.10 The
tip of the ventricular lead was placed at the right
ventricular apex if the R wave was
7 mV and the pacing
threshold was <1.0 V. Otherwise, the lead was placed at the closest
position to the apex that had an R wave
7 mV and a pacing threshold
<1.0 V. The value of 7 mV was selected on the basis of 2 assumptions.
(1) The minimum chronic R wave amplitude necessary for reliable sensing
of VF at a sensitivity of 0.3 mV is 3 to 4
mV,4 11 and (2)
the amplitude of the chronic R wave decreases by
40% from
implantation.12 If the sinus
rhythm R wave was <7 mV, VF was induced to testing
detection.
Vulnerability Testing
If the rhythm R wave was
7 mV, the T wave was
scanned at 10 J in the first 45 patients (group A) and at 15 J in the
last 35 patients (group B). The shock strength was liberalized to 15 J
in group B on the basis of an analysis of the results of
3-month electrophysiological studies in the
first 40 patients. Clinical variables shown in
Table 1
did not differ significantly between
groups.
The main features of the T-wave scan have been reported previously.9 10 Pacing was performed via the right ventricular electrodes. The baseline drive (S1) was delivered at a cycle length of 500 ms. All 12-lead electrocardiographic leads were recorded simultaneously, and the lead with the latest-peaking, monophasic T wave was selected. The first T-wave shock was delivered after 8 S1s at the peak of this T wave. The right ventricle was the anode for the first phase of biphasic shocks. If VF was not induced, subsequent shocks were delivered 20 ms and 40 ms before and 20 ms after the peak. There was a 1-minute recovery period between shocks that did not induce VF. If VF was not induced by these 4 shocks, the ICD was implanted.
If any of the T-wave shocks induced VF, defibrillation was
performed with a shock 10 J greater than the T-wave shock. If this
shock failed, a 200-J external shock was delivered. There was a
4-minute recovery period after each VF episode. The T-wave shock was
then increased by 5 J and the T-wave scan was repeated. This process
was repeated until none of the 4 shocks induced VF. The ULV was
recorded as the shock strength that did not induce
VF.6 7 8 9 10
The lead was repositioned at least once if the ULV was
25
J.
ICD Programming
Up to 3 detection zones were programmed: 1 zone (VF)
was programmed in 8 patients, 2 zones (VF and VT) were programmed in 18
patients, and 3 zones (VF, VT, and fast VT) were programmed in 54
patients. The VF detection interval was 284±19 ms, and the fast
VT detection interval was 323±11 ms. This latter value is comparable
to the programmed VF detection interval in studies of 2-zone
ICDs.13 The minimum
autoadjusting sensitivity was 0.3 mV. The first fast VT therapy was 1
trial of antitachycardia pacing. The second fast VT therapy
(first shock) and first VF shock were set to 5 J greater than the
lowest tested shock strength that did not induce
VF.9 This was 15 J in group A
patients in whom VF was not induced, 20 J in group B patients in whom
VF was not induced, and the ULV+5 J in patients in whom VF was induced.
The second shock was 10 J greater than the first shock, and subsequent
shocks were 30 to 35 J.
Chronic
Electrophysiological Study
Patients were studied 3 months postoperatively, as
described previously.8
Detection of VF and efficacy of the first programmed shock were tested
twice in trials separated by 4 minutes. VF was induced by a 2-J T-wave
shock. The time for detection of VF was measured manually at 25
mm/s on stored electrograms from the T-wave shock to the VF detection
marker. Because detection amplifiers were saturated for
0.5 s after
the T-wave shock, the measured detection time slightly overestimates
the true detection time.
Follow-Up
Patients were followed from the date of implantation
until the date of study closure or reprogramming of the first VF shock.
The first shock strength was increased by 10 J if patients had acute
myocardial infarction, exacerbation of heart failure by >1 New York
Heart Association class, or unsuccessful first
shocks.8 9 The
cardiac rhythms detected in the fast VT and VF zones were classified
using previously reported, prospective criteria for stored detection
intervals9 14 and
electrograms.9 15
Each appropriate first VF shock was analyzed using stored
postshock intervals and electrograms. VT and fast VT shocks were not
analyzed.
Statistical Analysis
Data are presented as mean±1SD. Basic
comparative statistics were calculated using the 2-tailed, paired
t test, or
2 test.
P<0.05 was used to reject the
null hypothesis for single
comparisons.
| Results |
|---|
|
|
|---|
|
VF at Implantation
VF was induced to assess sensing in 2 patients who had
R waves
7 mV, despite lead repositioning. T-wave scans did not induce
VF in these patients.
The ULV was determined at implantation because vulnerability testing induced VF in 20 patients. Of 16 group A patients, the ULV was 15 J in 10 patients, 20 J in 5 patients, and 25 J in 1 patient. If T-wave scanning had been done at 15 J in group A, inductionless implantations would have been performed in 38 of the 45 patients (84%), which is similar to the percentage of patients in group B who had inductionless implantations. The ULV was 20 J in all 4 group B patients in whom it was measured.
The ventricular lead was repositioned in
16 patients because of pacing thresholds >1.0 V, in 18 patients
because the R wave was <7 mV, and in 2 patients because of ULVs
25
J.
Acute Versus Chronic R Waves
Table 2
shows the amplitudes of baseline-rhythm R waves
measured at implantation by direct connection to the electrodes with
those measured 3 months later by telemetry from the ICD. Chronic R
waves were lower than acute R waves for both active-fixation and
passive-fixation leads
(P<0.001). For the 78 patients
with R waves
7 mV at implantation, the lowest chronic R wave was 4.1
mV on a passive lead. Overall, chronic R-waves <5 mV were present
in 3 of 43 active leads (7%) and in 3 of 37 passive leads (8%,
P=0.8).
|
Chronic
Electrophysiological Study
An electrophysiological
study was performed at 87±9 days in 75 of 80 patients (94%),
including 55 of the 58 patients (95%) who underwent inductionless
implantations. In these 55 patients, induced VF was detected in
4.7 s
(Figure 2
). The mean detection time was 3.7±0.5 s.
The interval between the T-wave shock and the first VF
electrogram was 0.4±0.1 s.
|
All first shocks were successful; these included 110 shocks
in the 55 patients who underwent inductionless implantations (95%
confidence intervals, 97% to 100%), 4 shocks programmed based on
vulnerability safety margins in 2 patients in whom VF was induced to
assess sensing, and 36 shocks in 18 patients programmed to the ULV+5
J. The charge time was 2.8±0.3 s to 15 J and 3.9±0.3 s to 20
J. The high-voltage lead impedance was 37±8
versus 39±7
at
implantation
(P=0.07).
Follow-Up
The duration of follow-up was 17±4 months in group A
and 8±4 months in group B. Total follow-up for patients who had
inductionless implantations was 616 months (476 months for the 28
patients in group A and 240 months for the 30 patients in group B). One
patient died of heart failure and one had a cardiac transplant. Six
other patients were censored at the time first shocks were reprogrammed
because of exacerbation of heart failure by 2 functional classes (4
patients), failed first shock and exacerbation of heart failure by 2
classes (1 patient), and acute myocardial infarction (1
patient).
First-Shock Efficacy for Spontaneous
VT/VF
In patients who had inductionless implantations, there
were 84 appropriate first shocks in the VF zone
(Figure 3
) and 114 in the fast VT zone. Details are
summarized in
Table 3
. Overall, 197 of 198 first shocks (99%) were
successful (95% confidence intervals, 97% to 100%). The only
ineffective shock occurred during treatment with
intravenous amiodarone and lidocaine for VT storm
associated with exacerbation of heart failure:
antitachycardia pacing accelerated monomorphic VT in the VT
zone to polymorphic VT.
|
|
In patients in whom VF was induced at implantation, there were 18 appropriate first shocks in the VF zone and 21 in the fast VT zone, all of which were successful. (95% confidence intervals, 92% to 100%).
Detection Times and Syncope
The time to detection of the 84 appropriate
arrhythmias in the VF zone was 4.8±0.5 s (range, 2.9 to
6.1 s). The longest detection time was caused by VT on the
boundary of the VT and VF zones, not undersensing. One patient with a
ULV of 20 J had syncope before a first shock after
antitachycardia pacing accelerated VT into the VF zone. The
patient who had a failed first shock had syncope before the successful
second shock.
| Discussion |
|---|
|
|
|---|
Previous Implantation and Programming
Strategies
Strategies based on defibrillation safety margin,
defibrillation threshold, and ULV and have been
compared.9 The defibrillation
safety margin strategy limits fibrillation-defibrillation testing to
the minimum necessary to confirm a safety margin between the ICDs
maximum output and the shock strength required for consistent
defibrillation.2 9 16
Patient-specific strategies based on the defibrillation
threshold2 and
ULV9 permit programming
shocks that are strong enough to defibrillate but not strong enough to
cause unnecessary17 or
prolonged18 myocardial
depression and unnecessary conduction
block.19 They prevent
avoidable battery depletion if sustained or self-terminating
arrhythmias cause capacitors to charge frequently, and they may
reduce the risk of syncope during capacitor charging. However, the
defibrillation threshold method requires multiple
fibrillation-defibrillation episodes, and the ULV method requires 1 VF
episode as well as multiple T-wave
shocks.9
Vulnerability Safety-Margin Strategy
The principal goal of ULV and defibrillation threshold
implantation strategies is to identify patients who have low
defibrillation energy requirements. In the present study, these
patients were identified accurately by 4 T-wave shocks without inducing
VF. This vulnerability safety-margin strategy resulted in a first-shock
success rate as good
as9 13 or better
than20 those reported for
other methods of implant testing.
Why Is VF Induced at ICD Implantation?
VF is induced to ensure the detection of VF, an
adequate defibrillation safety margin, and appropriate function of the
ICD generator. ICD generator failures are extraordinarily
rare.
When Should Detection of VF Be Tested at
Implantation?
Detection of VF depends both on the electrodes
sensing bipole and the ICDs sensing circuit and detection algorithm.
Interactions of ICDs with implanted electronic
devices11 21 22
require testing VF sensing. Otherwise, a baseline-rhythm R wave
5 mV
ensures rapid detection of VF at a nominal sensitivity of 0.3
mV.4 11
Undersensing VF can occur rarely, despite an adequate R wave, if the
amplitude of the VF signal varies
rapidly.23 Implant testing
may not exclude this remote possibility because VF electrogram patterns
have limited
reproducibility.24
Redetection of VF may fail after an unsuccessful shock, but this has
been reported only with closely-spaced (6 mm), integrated bipolar
leads.23 25
The present study shows that the detection of VF is
reliable with the lead, sensing circuit, and detection algorithm used
if the baseline R wave is
7 mV. Because R-wave amplitude may decrease
as leads mature, postimplantation testing may be more useful than
implantation testing.
When Should Defibrillation Be Tested at
Implantation?
More than 95% of patients treated with active-can,
left-pectoral ICDs pass defibrillation implantation
criteria.26 However, in up
to 11% of patients, the shock polarity, lead position, or generator
position was changed after failed defibrillation to meet these
criteria.26 Thus, a measure
of defibrillation efficacy at implantation is necessary to ensure an
acceptable defibrillation safety margin. Vulnerability safety-margin
testing provides such a measure without inducing VF in most patients.
If T-wave scanning had been performed at 15 J in all patients in the
present study, VF would have been induced in only 12 patients
(15%), 2 of whom required lead repositioning.
Implantation Morbidity Shocks Versus VF
Some complications of implantation testing, such as
intractable VF,3 cerebral
hypoperfusion,2 and
myocardial ischemia,1
are directly related to VF or circulatory arrest rather than to shocks.
Prudence dictates that VF and circulatory arrest be induced only when
they provide a specific benefit. Patients at higher than usual risk for
VF induction include those in whom external defibrillation is
unreliable, those with high-grade coronary stenoses,
and those with recent coronary interventions.
Because shocks in sinus rhythm cause transient, energy-dependent myocardial depression,18 both shocks and VF contribute to the myocardial depression caused by fibrillation-defibrillation testing.3 Thus, prudence also dictates that the number and strength of shocks be minimized. The vulnerability safety-margin method both minimizes VF and limits the number and strength of shocks.
Limitations
(1) Patients who are candidates for inductionless
implantations using the vulnerability safety-margin strategy cannot be
identified preoperatively. (2) Reliable detection of VF if the baseline
R wave is
7 mV may not apply to all ICD systems. (3) Reported shock
efficacy for spontaneous arrhythmias applies to clinically
stable patients. (4) In this study, as in previous
studies,9 13 most
spontaneous arrhythmias in the VF zone were monomorphic VT. The
number of spontaneous episodes of VF or polymorphic VT is low in
the present study. (5) This study does not establish vulnerability
safety-margin testing as the optimal method of minimizing shocks and VF
at implantation.
Clinical Implications
Vulnerability safety-margin testing permits
inductionless implantations in >80% of patients if the T wave is
scanned at 15 J. Patient-specific first shocks programmed by this
method are below maximum output in >90% of patients and provide
clinical results comparable to the best previously reported methods.
The optimal shock strength for T-wave scanning depends on the tradeoff
between the value of avoiding VF and that of programming the lowest
effective shock strength.
| Acknowledgments |
|---|
Received November 14, 2000; revision received February 1, 2001; accepted February 12, 2001.
| References |
|---|
|
|
|---|
2. Singer I, Lang D. Defibrillation threshold: clinical utility and therapeutic implications. Pacing Clin Electrophysiol.. 1992;15:932949.[Medline] [Order article via Infotrieve]
3. Vlay SC. Defibrillation threshold testing: necessary but evil? Am Heart J. 1989;117:499504.[Medline] [Order article via Infotrieve]
4. Panotopoulos P, Krum D, Axtell K, et al. Ventricular fibrillation sensing and detection by implantable defibrillators: is one better than the others? A prospective, comparative study. J Cardiovasc Electrophysiol. In press.
5. Michelson BI, Igel DA, Wilkoff BL. Adequacy of implantable cardioverter-defibrillator lead placement for tachyarrhythmia detection by sinus rhythm electrogram amplitude. Am J Cardiol. 1995;76:11621166.[Medline] [Order article via Infotrieve]
6. Chen PS, Swerdlow CD, Hwang C, et al. Current concepts of ventricular defibrillation. J Cardiovasc Electrophysiol. 1998;9:553562.[Medline] [Order article via Infotrieve]
7. Swerdlow C, Ahern T, Kass R, et al. Upper limit of vulnerability is a good estimator of shock strength associated with 90% probability of successful defibrillation in humans with transvenous implantable cardioverter defibrillators. J Am Coll Cardiol. 1996;27:11121117.[Abstract]
8. Martin DJ, Chen PS, Hwang C, et al. Upper limit of vulnerability predicts chronic defibrillation threshold for transvenous implantable defibrillators. J Cardiovasc Electrophysiol. 1997;8:241248.[Medline] [Order article via Infotrieve]
9.
Swerdlow CD, Peter
CT, Kass RM, et al. Programming of implantable
cardioverter-defibrillators on the basis of the upper limit of
vulnerability. Circulation. 1997;95:14971504.
10. Swerdlow CD, Martin DJ, Kass RM, et al. The zone of vulnerability to T wave shocks in humans. J Cardiovasc Electrophysiol. 1997;8:145154.[Medline] [Order article via Infotrieve]
11. Glikson M, Luria D, Friedman PA, et al. Are routine arrhythmia inductions necessary in patients with pectoral implantable cardioverter defibrillators? J Cardiovasc Electrophysiol. 2000;11:127135.[Medline] [Order article via Infotrieve]
12. Hurst T, Waldecker B. One year experience with a new, steroid eluting, tined defibrillation lead: results of a prospective, randomised, multicentre comparison with a conventional, screw-in defibrillation lead. Eur J Pacing Electrophysiol. 1998;8:1116.
13. Neuzner J, Liebrich A, Jung J, et al. Safety and efficacy of implantable defibrillator therapy with programmed shock energy at twice the augmented step-down defibrillation threshold: results of the prospective, randomized, multicenter low-energy Endotak trial. Am J Cardiol. 1999;83:34D39D.[Medline] [Order article via Infotrieve]
14.
Raitt MH, Dolack
GL, Kudenchuk PJ, et al. Ventricular arrhythmias
detected after transvenous defibrillator implantation in patients with
a clinical history of only ventricular fibrillation:
implications for use of implantable defibrillator.
Circulation. 1995;91:19962001.
15. Marchlinski FE, Callans DJ, Gottlieb CD, et al. Benefits and lessons learned from stored electrogram information in implantable defibrillators. J Cardiovasc Electrophysiol. 1995;6:832851.[Medline] [Order article via Infotrieve]
16. Lang DJ, Swanson DK. Safety margin for defibrillation. In: Alt E, Klein H, Griffin JC, eds. The Implantable Cardioverter/Defibrillator. Berlin: Springer-Verlag; 1992:272279.
17.
Jones JL, Milne
KB. Dysfunction and safety factor strength-duration curves for biphasic
defibrillator waveforms. Am J
Physiol. 1994;266:H263H271.
18. Tokano T, Bach D, Chang J, et al. Effect of ventricular shock strength on cardiac hemodynamics. J Cardiovasc Electrophysiol. 1998;9:791797.[Medline] [Order article via Infotrieve]
19.
Yabe S, Smith WM,
Daubert JP. Conduction disturbances caused by high current
density electric fields. Circ
Res. 1990;66:11901203.
20.
Venditti FJ Jr,
John RM, Hull M, et al. Circadian variation in defibrillation energy
requirements. Circulation. 1996;94:16071612.
21. Glikson M, Trusty JM, Grice SK, et al. A stepwise testing protocol for modern implantable cardioverter-defibrillator systems to prevent pacemaker-implantable cardioverter-defibrillator interactions. Am J Cardiol. 1999;83:360366.[Medline] [Order article via Infotrieve]
22. Glotzer TV, Gordon M, Sparta M, et al. Electromagnetic interference from a muscle stimulation device causing discharge of an implantable cardioverter defibrillator: epicardial bipolar and endocardial bipolar sensing circuits are compared. Pacing Clin Electrophysiol.. 1998;21:19961998.[Medline] [Order article via Infotrieve]
23.
Natale A, Sra J,
Axtell K, et al. Undetected ventricular fibrillation in
transvenous implantable cardioverter-defibrillators: prospective
comparison of different lead system-device combinations.
Circulation. 1996;93:9198.
24. Taneja T, Goldberger J, Parker MA, et al. Reproducibility of ventricular fibrillation characteristics in patients undergoing implantable cardioverter defibrillator implantation. J Cardiovasc Electrophysiol. 1997;8:12091217.[Medline] [Order article via Infotrieve]
25.
Jung M, Manz M,
Moosdorf R, et al. Failure of an implantable cardioverter-defibrillator
to redetect ventricular fibrillation in patients with a
nonthoracotomy lead system.
Circulation. 1992;86:12171222.
26. Bardy G, Yee R, Jung W. Multicenter experience with a pectoral unipolar implantable defibrillator. J Am Coll Cardiol. 1996;28:400410.[Abstract]
This article has been cited by other articles:
![]() |
B. Lemke, T. Lawo, M. Zarse, A. Lubinski, U. Kreutzer, J. Mueller, A. Schuchert, S. Mitzenheim, D. Danilovic, T. Deneke, et al. Patient-tailored implantable cardioverter defibrillator testing using the upper limit of vulnerability: the TULIP protocol Europace, August 1, 2008; 10(8): 907 - 913. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Day, R. N. Doshi, P. Belott, U. Birgersdotter-Green, M. Behboodikhah, P. Ott, K. A. Glatter, S. Tobias, H. Frumin, B. K. Lee, et al. Inductionless or Limited Shock Testing Is Possible in Most Patients With Implantable Cardioverter- Defibrillators/Cardiac Resynchronization Therapy Defibrillators: Results of the Multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations) Circulation, May 8, 2007; 115(18): 2382 - 2389. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow Reappraisal of implant testing of implantable cardioverter defibrillators J. Am. Coll. Cardiol., July 7, 2004; 44(1): 92 - 94. [Full Text] [PDF] |
||||
![]() |
C. Swerdlow, K. Shivkumar, and J. Zhang Determination of the Upper Limit of Vulnerability Using Implantable Cardioverter-Defibrillator Electrograms Circulation, June 24, 2003; 107(24): 3028 - 3033. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |