Circulation. 2001;103:2585-2590

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Cohen, M. I.
Right arrow Articles by Rhodes, L. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Cohen, M. I.
Right arrow Articles by Rhodes, L. A.
Related Collections
Right arrow Pacemaker
Right arrow Pediatric and congenital heart disease, including cardiovascular surgery

(Circulation. 2001;103:2585.)
© 2001 American Heart Association, Inc.


Clinical Investigation and Reports

Permanent Epicardial Pacing in Pediatric Patients

Seventeen Years of Experience and 1200 Outpatient Visits

Presented in part at the 73rd Scientific Sessions of the American Heart Association, New Orleans, La, November 12–15, 2000.

Mitchell I. Cohen, MD; David M. Bush, MD, PhD; Victoria L. Vetter, MD; Ronn E. Tanel, MD; Tammy S. Wieand, MS; J. William Gaynor, MD; Larry A. Rhodes, MD

From the Divisions of Cardiology (M.I.C., D.M.B., V.L.V., R.E.T., T.S.W., L.A.R.) and Cardiothoracic Surgery (J.W.G.) and the Departments of Pediatrics (M.I.C., D.M.B., V.L.V., R.E.T., T.S.W., L.A.R.) and Surgery (J.W.G.), The Children’s Hospital of Philadelphia and the University of Pennsylvania School of Medicine, Philadelphia.

Correspondence to Mitchell I. Cohen, MD, Division of Pediatric Cardiology, The Children’s Hospital of Philadelphia, 34th and Civic Center Boulevard, Philadelphia, PA 19104.


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background—The purpose of this study was to evaluate the long-term outcome of all pediatric epicardial pacing leads.

Methods and Results—All epicardial leads and 1239 outpatient visits between January 1, 1983, and June 30, 2000, were retrospectively reviewed. Pacing and sensing thresholds were reviewed at implant, at 1 month, and at subsequent 6-month intervals. Lead failure was defined as the need for replacement or abandonment due to pacing or sensing problems, lead fracture, or phrenic/muscle stimulation. A total of 123 patients underwent 207 epicardial lead (60 atrial/147 ventricular, 40% steroid) implantations (median age at implant was 4.1 years [range 1 day to 21 years]). Congenital heart disease was present in 103 (84%) of the patients. Epicardial leads were followed for 29 months (range 1 to 207 months). The 1-, 2-, and 5-year lead survival was 96%, 90%, and 74%, respectively. Compared with conventional epicardial leads, both atrial and ventricular steroid leads had better stimulation thresholds 1 month after implantation; however, only ventricular steroid leads had improved chronic pacing thresholds (at 2 years: for steroid leads, 1.9 µJ [from 0.26 to 16 µJ]; for nonsteroid leads, 4.7 µJ [from 0.6 to 25 µJ]; P<0.01). Ventricular sensing was significantly better in steroid leads 1 month after lead implantation (at 2 years: for steroid leads, 8 mV [from 4 to 31 mV]; for nonsteroid leads, 4 mV [from 0.7 to 10 mV]; P<0.01). Neither congenital heart disease, lead implantation with a concomitant cardiac operation, age or weight at implantation, nor the chamber paced was predictive of lead failure.

Conclusions—Steroid epicardial leads demonstrated relatively stable acute and chronic pacing and sensing thresholds. In this evaluation of >200 epicardial leads, lead survival was good, with steroid-eluting leads demonstrating results similar to those found with historical conventional endocardial leads.


Key Words: electrocardiography • pacemakers • pediatrics


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Pacing in the pediatric population typically results from bradycardia produced by sinus node dysfunction or atrioventricular (AV) block.1 Permanent epicardial pacing is often required in children because of their small size, congenital cardiac defects with right-to-left shunts, or lack of access to the chamber requiring pacing. In patients with prior cardiac surgery, the epicardium often has fibrosis and adhesions resulting in higher pacing thresholds and exit block.2 3 4 5 6 7 The increased pacing thresholds associated with epicardial leads have reduced lead longevity.3 8 Although endocardial pacing systems have demonstrated acceptable midterm sensing and pacing characteristics,9 concern has been raised about long-term vascular and/or valvular integrity with multiple transvenous leads.10 11

Recent advances in lead technology and surgical approach, such as steroid epicardial leads and left atrial leads, have demonstrated encouraging early sensing and pacing qualities when used in children.12 13 The use of these new implant strategies on long-term lead survival have not been well established. The objective of the present study was to report our long-term experience with a large number of epicardial pacemakers and to identify possible predictors of early and late lead-related complications.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
A complete search of the cardiac surgical and pacemaker databases identified all patients who underwent permanent epicardial pacemaker implantation between January 1, 1983, and June 30, 2000, at The Children’s Hospital of Philadelphia. Patients aged >21 years at the time of pacemaker implantation were excluded so as to primarily evaluate a large pediatric cohort. All hospital records, operative records, and 1239 outpatient pacemaker clinic visits were retrospectively reviewed. (A portion of these patients were presented in an earlier study.14 ) These records were abstracted for data on age, sex, cardiac malformation, presence or absence of congenital heart disease, cardiac surgical procedures (type and number), surgical approach, indication for pacing, mode of pacing, chamber placement, lead fixation method, steroid lead properties, pacemaker complications, and all follow-up information available.

Operative Course
Epicardial leads were implanted through a midline sternotomy, lateral thoracotomy, or subxiphoid approach. The surgical approach was based on the patient’s underlying cardiac anatomy, cardiac position in situ, prior operation(s), and/or concurrent operation at the time of lead placement. The ventricular lead was often fixed on the diaphragmatic ventricular surface. The atrial lead was positioned on either the right or left atrium; whichever afforded the best pacing and sensing thresholds. The surplus of lead was addressed by creating loops of the electrode within the pericardium and the pacemaker pocket. Leads were tunneled with care to avoid sites of potential injury along the rib margins to the generator implant site in the abdomen.

Implant measurements were obtained by using a pacing system analyzer (Medtronic 5311). Measurements included lead impedances at 0.5 ms/5 V, sensed P or R waves (if present), amplitude stimulation threshold (minimum voltage delivered at a fixed pulse width of 0.5 ms that consistently captured the tissue), and the measured current at the threshold voltage and pulse width.

Hospital Course
Acute pacing characteristics were defined as energy threshold (ET), impedance, and sensing at implantation. Hospital charts were reviewed for length of stay, complications, number and duration of antibiotics received, and discharge pacing characteristics. Pacing and sensing thresholds were interrogated in all systems 48 hours before hospital discharge.

Pacemaker Follow-Up
Pacing and sensing thresholds were evaluated at 1 month, at 3 months, and at subsequent 6-month intervals. Outpatient evaluations consisted of real-time telemetry of battery and lead measurements. Sensing and pacing thresholds were determined by using a Medtronic 5300, Pacesetter, Intermedics, or CPI system analyzer. Threshold was determined by decreasing the pulse width until there was failure to capture. Early pacemakers without pulse-width autothreshold testing capabilities used voltage decrement at a fixed pulse width to assess threshold. The threshold was considered to be the lowest programmable pulse width at which there was consistent capture. For patients with slow ventricular escape complexes or absence of atrial activity, sensing tests could not always be performed.

Pacing and sensing thresholds were compared at implant, at hospital discharge, and at the 1-month, 6-month, 1-year, 2-year, 5-year, 10-year, and most recent pacemaker follow-up visits. For patients whose primary pacemaker care was at a referring institution, phone calls were made requesting a recent clinical summary, pacemaker interrogation, and any lead- or generator-related complications.

Lead failure was defined as need for replacement or abandonment based on the following: (1) fracture or insulation break, (2) increasing pacing or sensing thresholds, or (3) phrenic or myopotential stimulation. Infections were classified separately into (1) superficial infection (±positive blood culture) or (2) deep infection necessitating generator removal. Lead data were censored for elective change, death, or orthotopic heart transplantation.

Definitions
ET was defined as the least amount of energy producing consistent capture outside the refractory period and was used to facilitate comparison between acute and chronic pacing thresholds. The formula15 used to calculate ET is as follows: ET (µJ)=[voltage (V)2xpulse duration (ms)x106]/[resistance ({Omega})x1000 ms/s].

Statistical Analysis
Exploratory data analysis was performed by using descriptive measures. Categorical variables were expressed in terms of percentages with standard deviations. Continuous variables were expressed as means with standard deviations if the term was normally distributed; skewed variables were expressed as medians (with ranges). The strength of a statistical association was measured by use of the {chi}2 test for categorical variables. When cell numbers were small, the Fisher exact test was used. Statistical significance of the difference between continuous variables was assessed by the Wilcoxon rank sum test for skewed distribution. Logistic regression or the Cox proportional hazards model was used to assess confounding variables. Lead survival was assessed by using Kaplan Meier analysis (STATA 6.0) with significance based on the log-rank test.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
Patient Data
A total of 123 patients underwent 158 operations for 207 epicardial lead implantations during the 17-year study period. The age at lead implantation ranged from 1 day to 21 years (median 4.1 years). The average weight at lead implantation was 17 kg (range 1.4 to 87 kg). Indications for pacemaker included postoperative AV block in 50 (40%) of the patients, sinus node dysfunction in 39 (32%) of the patients, congenital complete heart block in 23 (19%) of the patients, and other indications, such as bradycardia secondary to antiarrhythmic medication, in 11 (9%) of the patients.

The majority (84%) of the patients had structurally abnormal hearts and had undergone surgical correction(s) of complex cardiac malformations. Of the 103 patients with congenital heart disease, 37 had single ventricle physiology. The remaining 66 patients with congenital heart disease included those with the following: AV canal (14 patients), dextrotransposition of the great arteries (8 patients), ventricular septal defect (VSD) (8 patients), complex left ventricular outflow tract obstruction (7 patients), tetralogy of Fallot (6 patients), levotransposition of the great arteries with VSD (6 patients), double-outlet right ventricle (5 patients), heterotaxy syndrome and AV canal (4 patients), partial anomalous pulmonary venous return (2 patients), truncus arteriosus (1 patient), cardiomyopathy (2 patients), anomalous left coronary artery (1 patient), and atrial septal defect (2 patients).

Pacing Characteristics
The initial pacing mode used was AAI in 6 patients, VVI in 75 patients, and DDD in 42 patients. Of the 207 epicardial leads, 177 were implanted during 138 operations at a remote time from reparative or palliative cardiac surgery. Thirty leads were implanted at the completion of 20 open-heart surgical cases. Over the last 2 decades, numerous leads (Table 1Down) and generators (Table 2Down) were used, reflecting the technological advances.


View this table:
[in this window]
[in a new window]
 
Table 1. Epicardial Leads Used and Chamber Paced


View this table:
[in this window]
[in a new window]
 
Table 2. Implantable Generators Used Between 1983 and 2000

Operative and Perioperative Course
Epicardial leads were implanted by a subxiphoid approach (14%, 4 atrial/25 ventricular), a lateral thoracotomy (29%, 22 atrial/38 ventricular), or a sternotomy (57%, 34 atrial/84 ventricular). Fifty-eight patients had a chest tube placed after pacemaker implantation for a period of 1.4±1.7 days. The average length of stay after an isolated pacemaker implantation was 3 days (range 1 day to 69 days). Antibiotics were used in all patients after lead implantation for an average of 48 hours. Four patients developed a superficial cellulitis with negative blood cultures before discharge and received intravenous antibiotics for 7 days. Two other patients with congenital heart disease had a fever and positive blood culture after generator implantation. Antibiotics were given for 4 weeks, with prompt resolution of the bacteremia allowing for continued use of the pacing system. One patient developed a significant pacemaker pocket infection necessitating removal of the leads and generator.

There were 3 hospital deaths after epicardial pacemaker implantation. Two patients with prematurity, congenital complete heart block, and hydrops fetalis died secondary to lung disease of prematurity. One 6-month-old with trisomy 21 underwent pacemaker implantation for third-degree AV block 10 days after repair of an AV canal defect and suffered a cardiac arrest 3 days later.

Implant Data
The atrial ETs were 1.4 µJ (from 0.01 to 10.6 µJ) (steroid, 1.13 µJ [from 0.2 to 6.5 µJ]; nonsteroid, 2.2 µJ [from 0.01 to 8.3 µJ]; P=NS). Atrial sensing at implant was obtainable from 44 leads (2.9 mV [from 0.6 to 7.9 mV]). The atrial lead impedance was 335 {Omega} (from 223 to 748 {Omega}). No significant difference was observed in the acute ventricular stimulation thresholds between steroid (0.9 µJ [from 0.05 to 6.6 µJ]) and nonsteroid (1.1 µJ [from 0.05 to 25 µJ]) leads. Acute R-wave sensing was performed in 125 leads (11 mV [from 2 to 25 mV]). The ventricular lead impedance was 403 {Omega} (from 163 to 1000 {Omega}). No significant differences in implant stimulation thresholds were observed between atrial and ventricular epicardial leads.

Follow-Up Data
Of the 120 patients discharged after epicardial pacemaker implantation, 9 patients relocated and were lost to follow-up. Early readmission (<30 days) occurred in 13 patients. The major indication for early readmission was superficial cellulitis (n=7). All 7 patients had negative blood cultures, responded to intravenous antibiotics, and did not require removal of the lead(s) and/or generator. Three patients were readmitted with postpericardiotomy syndrome. Two other patients were readmitted with pleural effusions, and 1 Fontan patient had an exacerbation of protein-losing enteropathy after a thoracotomy for pacemaker implantation.

Late pacemaker-related readmission (>30 days) occurred in 6 patients (superficial cellulitis [n=3] and deep pacemaker pocket infection necessitating removal of leads and/or generator [n=3]). There were 5 late deaths in patients with pacemakers. Two Fontan patients died at 7 and 9 years, respectively, from pacemaker implantation secondary to low cardiac output and ventricular dysfunction. One Fontan patient with known atrial flutter died suddenly from a presumed arrhythmic death. There were 2 late deaths that may have been related to the pacemaker system. A newborn with double-outlet right ventricle, VSD, and PS developed complete heart block after a Damus-Kaye-Stansel procedure. A VVI pacemaker was implanted 10 days later. The child, who had been doing well at home, died suddenly 5 weeks later. Another newborn with transposition of the great arteries (levotransposition), VSD, PS, and congenital heart block died suddenly 7 weeks after VSD closure, left ventricle–to–pulmonary artery conduit, and pacemaker implantation. In both situations, nonsteroid-eluting epicardial ventricular leads were used.

The average length of follow-up was 29 months (range 1 to 207 months) for all epicardial lead and 19 months (range 3 to 61 months) for steroid-eluting leads. Forty-two leads had follow-up periods exceeding 5 years. Twenty-six patients had at least 1 reoperation for generator replacement with use of the original epicardial leads. An elective lead change at the time of reparative cardiac surgery occurred in 3 patients. Four other patients whose battery life had ended were changed to an endocardial system.

Lead Pacing and Sensing Characteristics
Atrial stimulation thresholds were significantly better for steroid-eluting leads at 1 month (steroid, 1.7 µJ [from 0.2 to 16.1 µJ]; nonsteroid, 4.1 µJ [from 1.2 to 18.1 µJ]; P=0.02). Atrial ETs remained relatively constant over 5 years (1.5 µJ [from 1.1 to 8.5 µJ]) without any further significant difference between steroid- and nonsteroid-eluting leads (Figure 1Down). The sensed P-wave amplitudes at 1 month and 2 years were 2.5±1.8 mV (n=17) and 3.2±2.5 mV (n=17), respectively (P=NS). There were no observable differences in atrial sensing between steroid and nonsteroid leads at any follow-up interval. Similarly, no acute or chronic differences in atrial lead impedance were observed between steroid and nonsteroid leads (at 1 month, 339±82 {Omega}; at 1 year, 364±82 {Omega}; and at 2 years, 372±87 {Omega}). No acute or chronic difference in stimulation thresholds was observed between left atrial (discharge ET [n=14], 1.7 µJ [from 0.6 to 10 µJ]; 2-year ET [n=11], 2.3 µJ [from 0.39 to 12 µJ]) and right atrial (discharge ET [n=27], 1.9 µJ [from 0.03 to 14 µJ]; 2-year ET [n=21], 1.5 µJ [from 0.4 to 6 µJ]) epicardial leads (P=NS).



View larger version (35K):
[in this window]
[in a new window]
 
Figure 1. Atrial ETs stratified by lead type. Nonsteroid-eluting leads (n=23, solid bars) and steroid-eluting leads (n=32, open bars) are shown.

The ventricular stimulation thresholds were significantly better for steroid-eluting leads at acute follow-up (at 1 month: steroid, 2.4 µJ [from 0.05 to 12.7 µJ]; nonsteroid, 6.1 µJ [from 0.4 to 27 µJ]; P<0.01) and chronic follow-up (at 2 years: steroid, 1.9 µJ [from 0.26 to 16 µJ]; nonsteroid, 4.7 µJ [from 0.6 to 25 µJ]; P<0.01) (Figure 2Down). The intrinsic R waves were greater in the steroid-eluting leads at 1 month after implant (steroid, 8 mV [from 4 to 31 mV]; nonsteroid, 4 mV [from 0.7 to 10 mV]; P=0.02) without a chronic difference (at 2-year follow-up, 5.6 mV [from 2.8 to 31 mV]). Ventricular impedances were relatively constant over the acute and chronic period (at 1 month, 356±84 {Omega}; at 1 year, 382±94 {Omega}; and at 2 years, 389±97 {Omega}). There were no significant differences in acute or chronic atrial and/or ventricular pacing or sensing between leads implanted in children with congenital heart disease and those with structurally normal hearts.



View larger version (39K):
[in this window]
[in a new window]
 
Figure 2. Ventricular ETs stratified by lead type. Nonsteroid-eluting leads (n=89, solid bars) and steroid-eluting leads (n=48, open bars) are shown.

Lead Survival
Epicardial lead failure occurred 16% (in 34 of 207 implantations) of the time (Table 3Down). The mean time to lead failure was 2.4±2.3 years. Increasing threshold was the most common cause of lead failure. Only 2 (2.4%) steroid-eluting leads had to be abandoned for exit block. The 1-, 2-, and 5-year epicardial lead survival was 96%, 90%, and 74%, respectively. No differences in lead survival were noted between atrial and ventricular epicardial leads (Figure 3Down). The 5-year freedom from survival for steroid-eluting leads was 83%, whereas for nonsteroid-eluting leads, the survival was 73% (Figure 4Down). The surgical approach significantly correlated with lead failure. None of the 29 subxiphoid-implanted leads failed during the present study (20 months [range 1 to 162 months]) (Figure 5Down). ETs >=3.0 µJ at hospital discharge predicted early lead failure (relative risk 2.8 [95% CI 1.2 to 6.6], P=0.02). There was no significant difference in the failure rate between epicardial leads that were redone and the remaining cohort of epicardial leads used. Neither congenital heart disease, lead implantation with a concomitant cardiac operation, age or weight at implantation, nor the chamber paced was predictive of lead failure.


View this table:
[in this window]
[in a new window]
 
Table 3. Cause of Lead Failure or Abandonment



View larger version (19K):
[in this window]
[in a new window]
 
Figure 3. Freedom from lead failure stratified by paced chamber: for atrial lead (solid line) at 1 year, 96% (CI 84.8% to 99%); at 2 years, 90.8% (CI 77 to 96.5%); at 3 years, 83.3% (CI 66.6% to 92.6%); at 5 years, 72% (CI 48.6% to 86.1%); and at 10 years, 72% (CI 48.6% to 86.1%); for ventricular lead (dotted line) at 1 year, 94.4% (CI 88.5% to 97.3%); at 2 years, 88.6% (CI 80.6% to 93.5%); at 3 years, 81.1% (CI 71.3% to 87.8%); at 5 years, 73.8% (CI 62.7% to 82%); and at 10 years, 69.2% (CI 54.7% to 79.8%).



View larger version (22K):
[in this window]
[in a new window]
 
Figure 4. Freedom from lead failure stratified by lead type for all epicardial leads: for steroid-eluting leads (solid line) at 1 year, 98.8% (CI 91.9% to 99.8%); at 2 years, 83.3% (CI 65.6% to 92.4%); at 3 years, 83.3% (CI 65.6% to 92.4%); and at 5 years, 83.3% (CI 65.6% to 92.4%); for nonsteroid-eluting leads (dotted lined) at 1 year, 92.6% (CI 85.8% to 96.3%); at 2 years, 90.4% (CI 82.9% to 94.8%); at 3 years, 81.4% (CI 72% to 87.9%); at 5 years, 72.7% (CI 62.1% to 80.8%); and at 10 years, 80.5% (CI 63.2% to 90.3%).



View larger version (23K):
[in this window]
[in a new window]
 
Figure 5. Freedom from lead failure stratified by surgical approach: for subxiphoid approach (n=29, solid line) at 1, 2, 3, 5, and 10 years, 100%; for thoracotomy (n=60, dashed line) at 1 year, 94.1% (CI 87.2% to 97.3%); at 2 years, 88.7% (CI 79.8% to 93.8%); at 3 years, 78% (CI 66.7% to 85.9%); at 5 years, 68.6% (CI 55.7% to 78.5%); and at 10 years, 62.4% (CI 44.9% to 79.8%); and sternotomy (n=118, dotted line) at 1 year, 93.9% (CI 82.2% to 98%); at 2 years, 86.0% (CI 71.1% to 93.5%); at 3 years, 82.9% (CI 67.0% to 91.6%); at 5 years, 75.9% (CI 57.6% to 86.9%); and at 10 years, 75.9% (CI 57.6% to 86.9%).


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
Although epicardial pacing is often required in small children, in patients with residual right-to-left shunts, and in patients with chambers that cannot be accessed by the transvenous route, most children can have either an endocardial or epicardial system. Although endocardial pacing requires less extensive surgery than does epicardial lead implantation, there is concern about vascular obstruction, AV valve integrity, and the limitations of lead accommodation with somatic growth.11 16 17 However, these risks have generally been outweighed in children by the higher acute and chronic stimulation thresholds of conventional epicardial leads, resulting in premature battery depletion and the need for subsequent operations.5 8 Recent advances in epicardial leads and surgical approaches have demonstrated improved early pacing and sensing thresholds.12 13 18 19 Prior studies evaluating epicardial leads in children have had low patient numbers or have involved series that predated both the introduction of lithium iodide batteries and steroid-eluting leads.8 12 20 21 In the present study, we report our long-term experience with >200 epicardial pacing leads in children during a 17-year enrollment.

The excessive pacing thresholds and high incidence of exit block with conventional epicardial leads presumably arise from a combination of epicardial fibrosis, scar formation, and/or pericardial adhesions after cardiac surgery. Conventional nonsteroid epicardial leads are associated with a 45% risk of exit block when implantation thresholds exceed 0.9 V at 0.5 ms.5 The addition of dexamethasone to the lead delivery system reduces the inflammatory response and the formation of the fibrous capsule.22 23 In this series of 82 steroid-eluting leads, we report a low incidence (2.4%) of exit block comparable to previously published smaller epicardial series.24

Atrial ETs were stable for both steroid- and nonsteroid-eluting leads at implantation through 5-year follow-up. The only significant drop in ET occurred 1 month after implantation in the steroid-eluting leads. This is similar to the study of Johns et al,12 who reported a significant decrease in atrial pulse width threshold 1 week after implantation, with little change thereafter. In contrast to the atrial leads, the nonsteroid ventricular stimulation thresholds were significantly worse than the steroid-eluting leads at acute and 2-year follow-up. The lack of significance at 5-year follow-up between steroid- and nonsteroid-eluting leads likely reflects some selection bias, in that leads with very high pacing thresholds were removed and replaced with lower ET leads. The relative consistency of our ventricular steroid-eluting leads regarding pacing thresholds over time is similar to previously published smaller series.24 25

The absence of any significant long-term improvement in atrial pacing thresholds with steroid-eluting leads compared with ventricular leads is somewhat interesting. Most of the patients in this cohort had structural heart disease and had undergone numerous cardiac operations. It is possible that the atria in these children have significant "scarring" that cannot be overcome by the simple addition of dexamethasone. Greater benefit may be achieved by reducing the pacing lead–epicardial interface and minimizing the battery drain.

Before the routine use of steroid leads, the 5-year epicardial lead survival was 40% to 70%.8 26 The overall 5-year epicardial lead survival in the present study was 74% and compares reasonably well with reported conventional endocardial systems.9 20 Although improved lead survival may be achieved with steroid endocardial leads, no long-term lead comparisons have been performed between steroid epicardial and endocardial leads. Unique to the present study was the observation of zero lead failure with subxiphoid leads. Avoidance of a thoracotomy minimizes the daily traction imposed on epicardial leads by respiration and arm movement. However, it is generally more difficult to implant atrial leads from a subxiphoid incision, and this usually requires a limited sternotomy or thoracotomy. The observation that discharge ETs >=3.0 µJ predicted lead failure reflects our early experience when intraoperative lead assessment was not routinely performed. In the last 5 years, our approach has been to test all pacing leads in the operating room.

Because surgical repair of complex cardiac malformations is being performed at younger ages, a growing percentage of children is expected to require pacemakers. The primary goal of pacemaker implantation is to achieve the lowest possible ET that can safely pace and sense the tissue appropriately. The present study reported acceptable pacing and sensing thresholds and a low lead-failure rate in a large cohort of epicardial leads implanted over 17 years. Steroid leads, especially ventricular leads, significantly reduced battery drain and the potential for subsequent surgery. Most children requiring a pacemaker will need one for the rest of their lives. As pediatric cardiologists, we should consider not only the immediate result of a particular lead/generator but also (and more important) how best to achieve a lifetime of pacing. Although it is technically feasible to implant endocardial leads in most children weighing {approx}15 kg, use of epicardial leads when the child is older and the innominate/cephalic veins are more developed may be a more prudent approach and may minimize long-term vascular and/or valvular injury.

Received December 14, 2000; revision received March 6, 2001; accepted March 8, 2001.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 

  1. Silka MJ, Manwill JR, Kron J, et al. Bradycardia-mediated tachyarrhythmias in congenital heart disease and responses to chronic pacing at physiologic rates. Am J Cardiol. 1990;65:488–493.[Medline] [Order article via Infotrieve]
  2. Villafane J, Austin E. Cardiac pacing problems in infants and small children: result of a 4-year prospective study. South Med J. 1993;86:784–788.[Medline] [Order article via Infotrieve]
  3. Williams WG, Hesslein PS, Kormos R. Exit block in children with pacemakers. Pacing Clin Electrophysiol. 1986;4:478–489.
  4. Serwer GA, Mericle JM, Armstrong BE. Epicardial ventricular pacemaker longevity in children. Am J Cardiol. 1988;61:104–106.[Medline] [Order article via Infotrieve]
  5. Beder SD, Kuehl KS, Hopkins RA, et al. Precipitous exit block with epicardial steroid-eluting leads. Pacing Clin Electrophysiol. 1997;20(pt 2):2954–2957.
  6. Henglein D, Gillette PC, Shannon C, et al. Long-term follow-up of pulse-width threshold of transvenous and myo-epicardial leads. Pacing Clin Electrophysiol. 1984;7:203–214.[Medline] [Order article via Infotrieve]
  7. Kugler J, Monsour W, Blodgett C. Comparison of two myoepicardial pacemaker leads: follow-up in 80 children, adolescents, and young adults. Pacing Clin Electrophysiol. 1988;11:2216–2222.[Medline] [Order article via Infotrieve]
  8. Sachweh JS, Vazquez-Jimenez JF, Schöndube FA, et al. Twenty years experience with pediatric pacing: epicardial and transvenous stimulation. Eur J Cardiothorac Surg. 2000;17:455–461.[Abstract/Free Full Text]
  9. Lau YR, Gillette PC, Buckles DS, et al. Actuarial survival of transvenous pacing leads in a pediatric population. Pacing Clin Electrophysiol. 1993;16(pt 1):1363–1367.
  10. Gillette PC, Zeigler V, Bradham GB, et al. Pediatric transvenous pacing: a concern for venous thrombosis? Pacing Clin Electrophysiol. 1988;11:1935–1939.[Medline] [Order article via Infotrieve]
  11. Figa FH, McCrindle BW, Bigras JL, et al. Risk factors for venous obstruction in children with transvenous pacing leads. Pacing Clin Electrophysiol. 1997;20(pt 1):1902–1909.
  12. Johns JA, Fish FA, Burger JD, et al. Steroid-eluting epicardial pacing leads in pediatric patients: encouraging early results. J Am Coll Cardiol. 1992;20:395–401.[Abstract]
  13. Ramesh VA, Gaynor JW, Shah MJ, et al. Comparison of left and right atrial epicardial pacing in patients with congenital heart disease. Ann Thorac Surg. 1999;68:2314–2319.[Abstract/Free Full Text]
  14. Cohen MI, Vetter VL, Wernovsky G, et al. Epicardial pacemaker implantation and follow-up in patients with single ventricle after the Fontan operation. J Thorac Cardiovasc Surg. 2001;121:804–811.[Abstract/Free Full Text]
  15. Hamilton RM, Chiu C, Gow RM, et al. A comparison of two stab-on unipolar epicardial pacing leads in children. Pacing Clin Electrophysiol. 1997;20(pt 1):631–636.
  16. Angeli S. Superior vena cava syndrome following pacemaker insertion post atrial septal defect repair. Am Heart J. 1990;120:433–435.[Medline] [Order article via Infotrieve]
  17. Old WD, Paulsen W, Lewis S, et al. Pacemaker lead induced tricuspid stenosis: diagnosed by Doppler echocardiography. Am Heart J. 1990;117:1105–1107.
  18. Bauersfeld U, Nowak B, Molinari L, et al. Low-energy epicardial pacing in children: the benefit of autocapture. Ann Thorac Surg. 1999;68:1380–1383.[Abstract/Free Full Text]
  19. Schmid FX, Nowak B, Kampmann C, et al. Cardiac pacing in premature infants and neonates: steroid eluting leads and automatic output adaptation. Ann Thorac Surg. 1999;67:1400–1402.[Abstract/Free Full Text]
  20. Esperer HD, Singer II, Riede FT, et al. Permanent epicardial and transvenous single- and dual-chamber cardiac pacing in children. Thorac Cardiovasc Surgeon. 1993;41:21–27.[Medline] [Order article via Infotrieve]
  21. Villain E, Martelli H, Bonnet D, et al. Characteristics and results of epicardial pacing in neonates and infants. Pacing Clin Electrophysiol. 2000;23:2052–2056.[Medline] [Order article via Infotrieve]
  22. Mond HG, Stokes KB. The electrode-tissue interface: the revolutionary role of steroid elution. Pacing Clin Electrophysiol. 1992;15:95–107.[Medline] [Order article via Infotrieve]
  23. Radovsky AS, Van Fleet JF. Effects of dexamethasone elution on tissue reaction around stimulating electrodes of endocardial pacing leads in dogs. Am Heart J. 1989;117:1288–1298.[Medline] [Order article via Infotrieve]
  24. Cutler NG, Karpawich PP, Cavitt D, et al. Steroid-eluting epicardial pacing thresholds: 6-year experience of pacing thresholds in a growing pediatric population. Pacing Clin Electrophysiol. 1997;20(pt 1):2943–2948.
  25. Karpawich PP, Hakimi M, Arcineigas E, et al. Improved chronic epicardial pacing in children: steroid contribution to porous platinized electrodes. Pacing Clin Electrophysiol. 1992;15:1151–1157.[Medline] [Order article via Infotrieve]
  26. Kratz JM, Gillette PC, Crawford FA, et al. Atrioventricular pacing in congenital heart disease. Ann Thorac Surg. 1992;54:485–489. [Abstract]



This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al.
ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons
J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62.
[Full Text] [PDF]


Home page
CirculationHome page
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al.
ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons
Circulation, May 27, 2008; 117(21): e350 - e408.
[Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
J. Odim, B. Suckow, B. Saedi, H. Laks, and K. Shannon
Equivalent Performance of Epicardial Versus Endocardial Permanent Pacing in Children: A Single Institution and Manufacturer Experience
Ann. Thorac. Surg., April 1, 2008; 85(4): 1412 - 1416.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
H. Murayama, M. Maeda, H. Sakurai, A. Usui, and Y. Ueda
Predictors affecting durability of epicardial pacemaker leads in pediatric patients.
J. Thorac. Cardiovasc. Surg., February 1, 2008; 135(2): 361 - 366.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
M. Tomaske, P. Harpes, A. Dodge-Khatami, N. Amacker, and U. Bauersfeld
Do daily threshold trend fluctuations of epicardial leads correlate with pacing and sensing characteristics in paediatric patients?
Europace, August 1, 2007; 9(8): 662 - 668.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
E. P. Walsh
Interventional Electrophysiology in Patients With Congenital Heart Disease
Circulation, June 26, 2007; 115(25): 3224 - 3234.
[Full Text] [PDF]


Home page
EuropaceHome page
M. S. Silvetti, F. Drago, A. De Santis, G. Grutter, L. Rava, L. Monti, and R. Fruhwirth
Single-centre experience on endocardial and epicardial pacemaker system function in neonates and infants
Europace, June 1, 2007; 9(6): 426 - 431.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
N. C. Aellig, C. Balmer, A. Dodge-Khatami, M. Rahn, R. Pretre, and U. Bauersfeld
Long-Term Follow-Up After Pacemaker Implantation in Neonates and Infants
Ann. Thorac. Surg., April 1, 2007; 83(4): 1420 - 1423.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
M. S. Silvetti, F. Drago, G. Grutter, A. De Santis, V. Di Ciommo, and L. Rava
Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patients
Europace, July 1, 2006; 8(7): 530 - 536.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
M.-J. Hsieh, K.-H. Yeh, O. S. Satish, and C.-C. Wang
Permanent pacing using a coronary sinus lead in a patient with univentricular physiology: an extended application of biventricular pacing technology.
Europace, February 1, 2006; 8(2): 147 - 150.
[Abstract] [Full Text] [PDF]


Home page
Eur. J. Cardiothorac. Surg.Home page
S. Gabor, G. Prenner, A. Wasler, M. Schweiger, K. H. Tscheliessnigg, and F. M. Smolle-Juttner
A simplified technique for implantation of left ventricular epicardial leads for biventricular resynchronisation using video-assisted thoracoscopy (VATS)
Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 797 - 800.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
J. L. Navia, F. A. Atik, R. A. Grimm, M. Garcia, P. R. Vega, U. Myhre, R. C. Starling, B. L. Wilkoff, D. Martin, P. L. Houghtaling, et al.
Minimally Invasive Left Ventricular Epicardial Lead Placement: Surgical Techniques for Heart Failure Resynchronization Therapy
Ann. Thorac. Surg., May 1, 2005; 79(5): 1536 - 1544.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
T. Bove and K. Francois
Invited commentary
Ann. Thorac. Surg., July 1, 2004; 78(1): 202 - 203.
[Full Text] [PDF]


Home page
Eur Heart JHome page
A. Puglisi, M. Lunati, A. G.M Marullo, S. Bianchi, M. Feccia, F. Sgreccia, I. Vicini, S. Valsecchi, F. Musumeci, and E. Vitali
Limited thoracotomy as a second choice alternative to transvenous implant for cardiac resynchronisation therapy delivery
Eur. Heart J., June 2, 2004; 25(12): 1063 - 1069.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
F. Walker, S. C. Siu, S. Woods, D. A. Cameron, G. D. Webb, and L. Harris
Long-Term outcomes of cardiac pacing in adults with congenital heart disease
J. Am. Coll. Cardiol., May 19, 2004; 43(10): 1894 - 1901.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
J. D. R. Thomson, M. E. Blackburn, C. Van Doorn, A. Nicholls, and K. G. Watterson
Pacing activity, patient and lead survival over 20 years of permanent epicardial pacing in children
Ann. Thorac. Surg., April 1, 2004; 77(4): 1366 - 1370.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
M. I. Cohen, K. Buck, R. E. Tanel, V. L. Vetter, L. A. Rhodes, J. Cox, T. Sheldon, and L. Ruetz
Capture management efficacy in children and young adults with endocardial and unipolar epicardial systems,
Europace, January 1, 2004; 6(3): 248 - 255.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
J.-L. Jansens, M. Jottrand, N. Preumont, E. Stoupel, and D. de Canniere
Robotic-enhanced biventricular resynchronization: an alternative to endovenous cardiac resynchronization therapy in chronic heart failure
Ann. Thorac. Surg., August 1, 2003; 76(2): 413 - 417.
[Abstract] [Full Text] [PDF]


Home page
EuropaceHome page
J.-H. Nurnberg, H. Abdul-Khaliq, P. Ewert, and P. E. Lange
Antibradycardia pacing in patients with congenital heart disease: experience with automatic threshold determination and output regulation (AutocaptureTM)
Europace, January 1, 2003; 5(2): 199 - 205.
[Abstract] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
M. I. Cohen, D. M. Bush, J. W. Gaynor, V. L. Vetter, R. E. Tanel, and L. A. Rhodes
Pediatric pacemaker infections: Twenty years of experience
J. Thorac. Cardiovasc. Surg., October 1, 2002; 124(4): 821 - 827.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Cohen, M. I.
Right arrow A