(Circulation. 2001;103:2624.)
© 2001 American Heart Association, Inc.
Basic Science Reports |
From the University of Pittsburgh School of Medicine (F.S.V., M.C., D.F.), Cardiovascular Institute, Pittsburgh, Pa; POINT Biomedical (E.W.G., G.P.), San Carlos, Calif; and the University of Virginia (J.S.), Charlottesville, Va.
Correspondence to Flordeliza S. Villanueva, MD, University of Pittsburgh School of Medicine, Cardiovascular Institute, 200 Lothrop St, S568 Scaife Hall, Pittsburgh, PA 15213. E-mail villanuevafs{at}msx.upmc.ed
| Abstract |
|---|
|
|
|---|
Methods and ResultsThree left anterior descending (LAD) coronary stenoses of varying severity were created in each of 9 open-chest dogs. MCE was performed by continuous intravenous infusion of a nitrogen-filled bilayer shell microbubble, PB127, during triggered power Doppler imaging at incremental pulsing intervals. MCE and radiolabeled microsphere measurements were made at baseline and during each stenosis, with and without adenosine stress. Videointensities in the LAD and left circumflex (LCx) beds were plotted against pulsing interval and fit to a previously described exponential function modeling microbubble destruction and replenishment, which was used to derive parameters of bubble velocity (ß) and peak plateau videointensity (A). Contrast defects matching the location of radiolabeled microsphere hypoperfusion were clearly seen, without need for image processing. The product of ß and A was linearly related to LAD/LCx flow (r=0.90, P<0.0001) and inversely related to stenosis gradient (r=-0.70, P<0.0001). Endocardial/epicardial flow ratios were visualized and quantifiable.
ConclusionsAs with B-mode harmonics, a model of microbubble destruction/replenishment can be applied to power Doppler data as a means to detect a broad range of stenoses. Image clarity and the lack of attenuation or requirement for background subtraction are additional advantages of this imaging approach. Power Doppler MCE imaging holds promise for the detection of coronary artery disease.
Key Words: echocardiography contrast media coronary disease
| Introduction |
|---|
|
|
|---|
Conventional B-mode ultrasound imaging modified to harmonic mode has been a major innovation in microbubble detection3 4 that has established the potential of MCE for the assessment of myocardial perfusion.5 6 7 This method, however, is limited by the persistent tissue "noise" consequent to the nonlinear backscatter characteristics of myocardium.
Power Doppler is a new imaging approach for detecting microbubbles that may have advantages over B-mode harmonic imaging.8 9 This adaptation of conventional Doppler technology theoretically displays signals that derive uniquely from ultrasound-induced microbubble disruption.10 11 The premise underlying power Doppler MCE is that acoustically stimulated nonlinear microbubble oscillation or microbubble destruction12 13 14 15 16 results in a changing scatter pattern of the incident ultrasound beam that is detected by a Doppler processor as a large change in the returning radiofrequency signal.10 In power Doppler mode, the autocorrelator of the imaging system calculates the amplitude of the changing signal (as opposed to the frequency shift, as in color Doppler) and displays the amplitude of the signal as the image.10 Because of the strong nonlinear scattering properties of microbubbles,16 the returning signal derives predominantly from the microbubbles rather than from myocardial tissue, which is relatively more acoustically quiescent from the standpoint of a Doppler autocorrelator.
Despite the theoretical appeal of power Doppler for MCE, there are sparse data validating quantitative power Doppler imaging against independent measures of myocardial perfusion in vivo. Whether power Doppler can distinguish between nonflow-limiting stenoses of differing severity or whether quantitative models to assess myocardial perfusion can be applied to power Doppler images, as has been shown with B-mode harmonics,17 18 is unclear. Accordingly, this study was performed to test the hypothesis that power Doppler contrast imaging can detect and quantify coronary artery stenoses.
| Methods |
|---|
|
|
|---|
A left lateral thoracotomy was performed, and the heart was suspended in a pericardial cradle. A catheter was placed in the left atrium for radiolabeled microsphere injection. A perivascular balloon occluder (Research Products International Corp) was placed around the proximal left anterior descending artery (LAD), and transit time flow probes (Transonics Inc) were placed on the midportion of the LAD and left circumflex artery (LCx). A 25-gauge catheter was placed in the distal LAD for measurement of poststenotic pressure.
Ultrasound Contrast Agent
The ultrasound contrast agent study used for this
study, PB127 (POINT Biomedical Corporation), is specifically engineered
for power Doppler imaging and consists of nitrogen encapsulated by
a spherical bilayer shell composed of human albumin (outer
shell) and a biodegradable polymer (inner shell). The mean diameter is
4.4±2.1 µm, and concentration is 0.5x109
bubbles/mL; for continuous infusion, 4 mL of the agent was diluted in
50 mL of 5% dextrose water.
Myocardial Contrast
Echocardiography
Open-chest power Doppler imaging was performed
with a commercially available scanner (Harmonic Angio Mode, SONOS 5500,
Agilent Technologies, Inc). Machine settings included the following:
large packet size, medium line density, pulse repetition frequency of
3.7 to 4.2 kHz, and a mechanical index of 0.8 to 1.0. A gray-scale
color map ("S" map) was used, and B-mode harmonic data were
excluded from the digital data set so as to generate a pure Doppler
image. Midpapillary-muscle short-axis images were obtained at
increasing pulsing intervals during continuous infusion of PB127 (30
mL/h) and ECG-gated triggering at a point in the cardiac cycle with the
least cardiac motion.
MCE images were analyzed offline as described previously17 19 : the Doppler component of the image was transferred to a Macintosh computer loaded with customized software. Average pixel intensity was measured in regions of interest drawn over the LAD and LCx beds (transmural, endocardial, and epicardial) in aligned frames selected from different pulsing intervals.
The videointensity measurements were plotted against pulsing interval and fit to a previously described exponential function17 : y=A(1-e-ßt), where y is equal to videointensity measured at varying pulsing intervals, and t, A, and ß are parameters derived from curve fits as described below. Conceptually, in this model, each ultrasound pulse is assumed to be destructive to microbubbles, and during continuous infusion of contrast, the microbubbles replenish the ultrasound beam elevation at a rate proportional to red cell velocity (ß). At subsequent pulses separated by time t, part of the beam is replenished with microbubbles to an extent determined by red cell velocity and the duration of time before the next destructive pulse. At progressively longer intervals between pulses, microbubbles fully replenish the beam, and a peak plateau videointensity, A, is achieved. Using B-mode harmonic imaging, Wei and colleagues17 have shown that the product of Axß correlates with myocardial blood flow.
Myocardial Blood Flow Measurement
Regional myocardial perfusion was measured by
radiolabeled microsphere
technology.20 Approximately
2 to 3x106 11-µm radiolabeled
microspheres (New England Nuclear) were suspended in 3 mL of
0.9% saline solution/0.01% Tween 80 and injected into the left atrium
during 90-second arterial reference sample withdrawal. The
left ventricular slice corresponding to the MCE image was
sectioned into 16 pieces that were further divided into endocardial,
midwall, and epicardial portions. Samples were weighed and
gamma-counted (model 5535, Packard Instruments), corrections for energy
spillover into neighboring windows were made, and average
transmural blood flow (mL · min-1 ·
g-1) was
calculated.7 17
Experimental Protocol
Three stenosis stages were modeled in each
dog. A mild and a moderate nonflow-limiting stenosis and a
severe flow-limiting stenosis were successively created. We
defined a mild stenosis as a transstenotic gradient
(difference between mean aortic pressure and distal LAD pressure) of 5
to 10 mm Hg. To create a moderate nonflow-limiting
stenosis, the occluder was initially inflated until epicardial
flow began to decrease to define the maximal nonflow-limiting
gradient, and then partially deflated to result in a stenosis
with a gradient 5 to 7 mm Hg less than the maximal gradient. To
produce a flow-limiting stenosis, the occluder was inflated
until a 50% to 60% reduction in epicardial flow occurred. MCE was
performed at each stage, with and without adenosine (0.4 mg ·
kg-1 ·
min-1). Radiolabeled microsphere
measurements were made during adenosine infusion at each
stenosis stage and without adenosine during the
moderate stenosis stage. The transstenotic gradient
was continuously monitored, and MCE was performed only after at least 5
minutes of hemodynamic stability had been documented
after each balloon manipulation. At the end of the experiment, the LAD
was totally occluded, India ink was injected to delineate the LAD bed,
the dog was euthanized, and the MCE heart slice was sectioned for
radiolabeled microsphere
analysis.
| Results |
|---|
|
|
|---|
|
The transstenotic gradients across the LAD
stenoses indicated a broad range of stenosis severity
(Table 2
). As expected, there was an increase in gradient
with increasing stenosis, with a further increase at each
stenosis during adenosine infusion. The gradient was
highest during severe stenosis and did not substantially
increase with adenosine, most likely owing to a decrease in
flow at that stage.
|
Relationship Between Power Doppler Data and
Stenosis Severity
Figure 1
illustrates an imaging sequence in a dog with a
moderate nonflow-limiting stenosis before (upper panels) and
during (lower panels) adenosine infusion at similar pulsing
intervals. Despite the presence of a stenosis (resting
transstenotic gradient 19 mm Hg), in the absence of
adenosine, there was homogeneous enhancement of the
LAD and LCx beds and increasing opacification at incremental pulsing
intervals
(Figures 1A
through
1E).
With adenosine, there was a contrast defect in the anteroseptal
and anterior segments (region between arrows) as the LAD bed filled
more slowly relative to the LCx
(Figures 1G
through
1I).
Reflecting delayed microbubble transit, the defect was most prominent
at short pulsing intervals
(Figures 1G
and 1H
).
With increasing pulsing intervals, this defect became more filled in
such that the risk area was much less distinct at a pulsing interval of
99
(Figure 1J
). Other notable features of the images are the
absence of posterior wall attenuation typically seen in
contrast-enhanced short-axis images and the dark appearance of the
myocardium after microbubble destruction
(Figures 1A
and 1F
).
|
Figure 2
graphs the pulsing interval versus videointensity
data for the LAD and LCx beds for the dog in
Figure 1
. The uniform opacification of both territories
before adenosine is represented by the curves in
Figure 2A
, which are similar for both the LAD and LCx. With
adenosine
(Figure 2B
), there was delayed appearance of bubbles in the
LAD bed compared with the LCx (prolongation ofß), corresponding to
the prominent defect seen at the shorter pulsing intervals in
Figures 1G
and 1H
.
|
The location of the MCE perfusion defects paralleled the
spatial distribution of hypoperfusion.
Figure 3
maps and color codes microsphere flow in
the endocardial, midwall, and epicardial layers of each of the 16
myocardial segments during adenosine infusion in the same dog
with moderate stenosis shown in
Figure 1
. The location of the MCE perfusion defect in the
LAD bed (region between arrows) corresponded to the myocardial region
with reduced microsphere flow, and the endocardial/epicardial
transmural flow gradient seen with radiolabeled microspheres
was also visually represented on the MCE image.
|
Figure 4
depicts very mild and severe stenoses at
increasing pulsing intervals from another dog during adenosine
infusion. At baseline
(Figure 4A
), there was progressive myocardial enhancement in
both beds with increasing pulsing interval. With severe
stenosis
(Figure 4C
), the LCx bed filled briskly, whereas there was an
LAD defect (region between arrows) that persisted even at pulsing
intervals as long as 99 beats (not shown). With mild stenosis
(Figure 4B
), LAD bed enhancement was also delayed, but the
defect was much less marked and resolved more quickly at higher pulsing
intervals than with severe stenosis
(Figure 4C
).
|
The videointensity pulsing interval curves for the data
shown in
Figure 4
are depicted in
Figure 5
. There was rapid and uniform opacification of the
LAD and LCx beds at baseline
(Figure 5A
). There was progressive prolongation of ß with
mild stenosis
(Figure 5B
), and with severe stenosis, there was
delayed, incomplete filling of the LAD territory relative to the LCx
(Figure 5C
).
|
Figure 6
plots radiolabeled microsphere-derived
blood flow (LAD normalized to LCx) versus the product of Axß
from all 6 stages in the 9 dogs. There was a significant linear
relationship between normalized LAD flow and the product of Axß
derived from the curve fits to the power Doppler image data. There
was a significant inverse linear relationship between stenosis
severity, expressed as transstenotic gradient, and the
product of Axß (y=-0.02x+1.16,
r=-0.70,
P<0.0001). Furthermore, there
was a significant linear relationship between the normalized (LAD/LCx)
ratio of the product of Axß for the endocardium/epicardium and
LAD/LCx endocardial/epicardial flow ratios calculated with radiolabeled
microspheres (y=0.49x+0.52,
r=0.60,
P<0.0001)
|
| Discussion |
|---|
|
|
|---|
Relationship Between Myocardial Blood Flow and
Power Doppler Measurements
The phenomenon of microbubble destruction in an
ultrasound
field14 15 has
been the basis for an approach to the quantification of myocardial
blood flow with MCE. Using B-mode imaging in a similar animal model to
ours, Wei et
al17 18
demonstrated that after an ultrasound pulse of sufficient power to
destroy microbubbles, the rate of microbubble replenishment and the
peak plateau videointensity during continuous microbubble infusion
correlated with myocardial blood flow.
Whether power Doppler can generate physiologically meaningful data or reproduce the B-mode findings described by Wei et al is an important issue for several reasons. First, imaging modalities designed in theory to be advantageous over the more validated B-mode harmonics should demonstrate similar ability to detect stenoses. Second, reproduction of the previously described relationship between microbubble destruction/replenishment and blood flow by a different imaging technique would lend further support to the principles on which the Wei et al17 model is based and to new concepts in microcirculatory physiology emerging from this model.21
It might have been predicted that power Doppler images would apply to the model derived with B-mode harmonics because the ultrasound-induced microbubble destruction and replenishment that are the basis of the model are physical events common to both techniques. On the other hand, a parallel between B-mode harmonic and power Doppler capabilities for quantifying perfusion was not inevitable for at least 3 reasons. First, with power Doppler, the amplitude of the returning Doppler signal is determined by the magnitude of variance between "packets" of ultrasound pulses. In contrast, with B-mode harmonics, the image derives from harmonic frequency components selected from a spectrum of returning frequencies representing the acoustic backscatter of bubbles and tissue.2 3 Thus, the 2 techniques register different specific features of the same acoustic event. Second, the dynamic range of power Doppler (40 dB) is less than that of B-mode harmonic imaging (60 dB) in the imaging system used in the present study. Third, the 2 techniques may differ in the efficacy of microbubble destruction. Power Doppler delivers its energy in multiple, repetitive packets of ultrasound pulses, which may inherently be more destructive than the B-mode pulse.10
Given these considerations, it has not been possible to
assume that power Doppler would parallel B-mode harmonics for
quantifying stenoses. The major finding of the present
study is that the quantitative data taken from a pure Doppler image
set correlates to myocardial blood flow. In addition, as illustrated in
Figures 1
and 4
, qualitative visual interpretation of
stenosis severity is possible, without the need for background
subtraction. Moreover, these data indicate that it is possible to
distinguish transmural gradients in flow from power
Doppler-generated images.
Masugata and colleagues22 recently reported a study of dogs with flow-limiting stenoses subjected to MCE with power Doppler imaging, and the authors compared videointensity measures with B-mode harmonic imaging. In that study, power Doppler appeared more capable of delineating flow-limiting stenosis than B-mode imaging. The present study concurs with that study in providing data to support the capabilities of power Doppler. However, the present study differs from that of Masugata et al22 and adds to our understanding of power Doppler MCE in several important ways. First, Masugata et al22 evaluated flow-limiting stenoses at rest, whereas we imaged nonflow-limiting stenoses during pharmacological stress. We demonstrated the ability of power Doppler to identify even mild stenoses that are not flow limiting at rest, which more closely simulates clinically encountered scenarios. Second, we applied the exponential model of Wei et al17 to pulsing interval versus videointensity data to derive a correlation between the MCE data and radiolabeled microsphere-derived myocardial blood flow.
Advantages of Power Doppler Imaging
Power Doppler has several theoretical advantages
over B-mode harmonic imaging. Because the measurement of interest in
power Doppler is the change in the returning radiofrequency signal
resulting from the nonlinear behavior and/or dissolution of
microbubbles in an ultrasound field, the image should, under ideal
conditions, be a representation solely of the nonlinear and/or
acoustic destruction
events.9 10 11
The greater the change, the greater the amplitude calculated by the
autocorrelator detector of the system. The nonlinear properties of
myocardial tissue that contribute to persistent tissue background noise
in B-mode harmonic imaging should be suppressed with power Doppler
images. This property can be seen in
Figure 1A
, which demonstrates black myocardium
immediately after bubble destruction.
The phenomenon of posterior wall attenuation in contrast-enhanced short-axis views19 has favored the use of apical imaging planes in which far-field attenuation effects are less likely to obscure the myocardium.5 This approach, however, foregoes useful segmental information imparted by short-axis views. In the present study, bright myocardial opacification was achieved at low doses of PB127, with a striking lack of posterior wall attenuation. It is likely that careful dosage selection for continuous infusion17 helped to minimize such attenuation artifact. It is also possible that properties of the agent itself, including a relatively narrow size distribution that would allow for even tighter regulation of the effective cross-sectional scattering area at a given infusion rate, contributed to the lack of attenuation in the short-axis views. An additional advantage, therefore, of the techniques used in the present study is that the lack of far-field attenuation may restore the utility of short-axis imaging planes.
Study Limitations
Because power Doppler is premised on the
detection of change between successive pulses, it is susceptible to
signal artifacts created by other dynamic events such as cardiac
movement.11 This inherent
limitation was not insurmountable, because imaging at a point in the
cardiac cycle with the least cardiac movement largely overcame any
significant motion artifacts at the pulse-repetition frequencies that
were used.
We cannot be certain that all microbubbles were
destroyed with each ultrasound pulse and therefore whether the
enhancement seen at each pulsing interval represented only
de novo arrival of microbubbles into the imaging plane. The appearance
of dark myocardium
(Figure 1
), as well as our empiric confirmation of a
relationship between radiolabeled microsphere flow and
videointensity parameters derived from a mathematical model
that assumes complete
destruction,17 suggests that
incomplete bubble destruction was not a significant limitation.
Moreover, a multipulse technique such as Doppler should have
further enhanced microbubble destruction.
Generally, a high dynamic range has been advocated for contrast echo to optimize the detection of bubble signals above tissue signals, and power Doppler imaging has a lower dynamic range than B-mode harmonic imaging. Nonetheless, as indicated by our data, this property did not preclude the generation of quantifiable images. Conceivably, the impact of bubble specificity and less background tissue noise inherent to power Doppler superseded any effects of a slightly narrower dynamic range, as compared with B-mode imaging, in which a portion of the broader range is "wasted" by the background tissue signals.
We chose a gray-scale map for presentation of the power Doppler images. Subtle perfusion defects can be more difficult to appreciate because of intrinsic limitations to the visual perception of gray-level differences. Although the minimal background noise in the power Doppler images enabled visualization of defects and rendered quantifiable videointensity curves, color coding of the images, as previously described by us and others,5 7 17 may further facilitate visual analysis of subtle defects.
New methods for detecting microbubbles in tissue continue to evolve, including low-mechanical-index techniques that may offer the possibility of performing real-time perfusion imaging. These techniques are distinct from power Doppler, and the relative merits of one technique over the other remain to be determined.
Conclusions
We have demonstrated that detection and quantification
of a range of coronary stenoses using a model based on
microbubble destruction and replenishment is possible with power
Doppler imaging during pharmacological stress and continuous
infusion of an ultrasound contrast agent. Furthermore, excellent image
quality and videodensitometric data can be achieved without the need
for image processing. Minimal posterior wall attenuation in the
short-axis view permits a more comprehensive analysis of
segmental perfusion than is possible with the currently preferred
apical views. These data provide a basis and rationale for the use of
power Doppler as an approach to perfusion imaging during
MCE.
| Acknowledgments |
|---|
Received November 3, 2000; revision received January 12, 2001; accepted January 23, 2001.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. J. Pacella and F. S. Villanueva Effect of Coronary Stenosis on Adjacent Bed Flow Reserve: Assessment of Microvascular Mechanisms Using Myocardial Contrast Echocardiography Circulation, October 31, 2006; 114(18): 1940 - 1947. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Pacella, M. V. Kameneva, M. Csikari, E. Lu, and F. S. Villanueva A novel hydrodynamic approach to the treatment of coronary artery disease Eur. Heart J., October 1, 2006; 27(19): 2362 - 2369. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Yamada, K Komuro, T Mikami, N Kudo, H Onozuka, K Goto, S Fujii, K Yamamoto, and A Kitabatake Novel quantitative assessment of myocardial perfusion by harmonic power Doppler imaging during myocardial contrast echocardiography Heart, February 1, 2005; 91(2): 183 - 188. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Kondo, K. Ohmori, A. Oshita, H. Takeuchi, J. Yoshida, K. Shinomiya, S. Fuke, T. Suzuki, K. Mizushige, and M. Kohno Leukocyte-Targeted Myocardial Contrast Echocardiography Can Assess the Degree of Acute Allograft Rejection in a Rat Cardiac Transplantation Model Circulation, March 2, 2004; 109(8): 1056 - 1061. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Takagi, K. Ohmori, K. Yukiiri, I. Kondo, Y. Yu, A. Oshita, H. Takeuchi, K. Mizushige, and M. Kohno Quantitative assessment of coronary stenosis by harmonic power Doppler with a simple pulsing sequence and vasodilator stress in patients J. Am. Coll. Cardiol., June 4, 2003; 41(11): 2060 - 2067. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Masugata, N. Fujita, I. Kondo, B. Peters, K. Ohmori, K. Mizushige, M. Kohno, and A. N. DeMaria Assessment of right ventricular perfusion after right coronary artery occlusion by myocardial contrast echocardiography J. Am. Coll. Cardiol., May 21, 2003; 41(10): 1823 - 1830. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Fukuda, T Muro, T Hozumi, H Watanabe, K Shimada, M Yoshiyama, K Takeuchi, and J Yoshikawa Changes in transmural distribution of myocardial perfusion assessed by quantitative intravenous myocardial contrast echocardiography in humans Heart, October 1, 2002; 88(4): 368 - 372. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |