(Circulation. 2001;104:1588.)
© 2001 American Heart Association, Inc.
Brief Rapid Communications |
From the Departments of Radiology (X.Y., E.A., J.-M.S, D.W., A.K.), Anesthesia and Critical Care Medicine (D.L.), and Oncology (L.C.), Johns Hopkins Hospital, Baltimore, Md.
Correspondence to Xiaoming Yang, MD, PhD, Johns Hopkins Outpatient Center, Room 4243, 601 North Caroline Street, Baltimore, MD 21287. E-mail xyang{at}mri.jhu.edu
| Abstract |
|---|
|
|
|---|
Methods and Results We produced gadolinium/blue dye and gadolinium/gene-vector media by mixing Magnevist with a trypan-blue or a lentiviral vector carrying a green fluorescent protein (GFP) gene. The gadolinium was used as an imaging marker for magnetic resonance (MR) imaging to visualize vessel wall enhancement, and the blue dye/GFP was used as a tissue stain marker for histology/immunohistochemistry to confirm the success of the transfer. Using Remedy gene delivery catheters, we transferred the gadolinium/blue dye (n=8) or gadolinium/GFP lentivirus (n=4) into the arteries of 12 pigs, that were monitored under high-resolution MR imaging. The results showed, in all 12 pigs, the gadolinium enhancement of the target vessel walls on MR imaging and the blue/GFP staining of the target vessel tissues with histology/immunohistochemistry. This study shows the potential of using MR imaging to dynamically visualize (1) where the gadolinium/genes are delivered; (2) how the target portion is marked; and (3) whether the gene transfer procedure causes complications.
Conclusions We present a technical development that uses high-resolution MR imaging as an in vivo imaging tool to monitor catheter-based vascular gene delivery.
Key Words: cardiovascular diseases gene therapy magnetic resonance imaging
| Introduction |
|---|
|
|
|---|
Gene transfer with delivery catheters is currently performed under x-ray fluoroscopy, which displays, using a contrast medium, only the lumen of the vessel without providing direct imaging information about the vessel wall or atherosclerotic plaques. Therefore, one cannot properly monitor either the interaction between the genes and the atherosclerotic lesion or the existence and distribution of the genes within the target lesion during and after vascular gene transfer. Magnetic resonance (MR) technology offers great potential to fill this gap.
Cardiovascular MR imaging has some prominent advantages, including the ability to image the vessel wall, the ability to make multiple diagnostic evaluations of organ function and morphology, and the ability to provide multiple image planes with no risk of ionizing radiation. In the present article, we present our recent technical development using MR imaging to monitor a catheter-based primary vascular gene delivery procedure in vivo.
| Methods |
|---|
|
|
|---|
Gadolinium/Gene-Vector Medium
We constructed a novel lentiviral vector (EF.GFP) containing the human elongation factor (EF)-1a promoter and the GFP gene (Y. Cui, PhD, unpublished data, 2001). Then, we mixed Magnevist with the GFP lentiviral vector to achieve a gadolinium/GFP lentivirus medium with a net Magnevist concentration of 6%. We previously confirmed that 3% to 6% of Magnevist (14 to 28 mg/mL gadolinium) is an optimum concentration for the MR imaging of balloon inflation.5
Gene Delivery Catheter Positioning
The Remedy gene delivery catheter has the dual capability of high-pressure lesion dilation and low-pressure gene infusion. It was positioned, using an 0.014-inch conventional guidewire, into the target arterial segment under x-ray fluoroscopy. Subsequently, we replaced the conventional guidewire with an 0.014-inch intravascular MR imaging-guidewire (Surgi-Vision).
MR Imaging
All experiments were performed on a 1.5 Tesla MR unit (GE). To image deeply located iliac arteries, we operated the MR imaging guidewire in a receive-only mode, but to image the superficially located femoral arteries, we used a custom-made, 3-cm surface coil. We first inflated the dilation balloon with 3% Magnevist and obtained a coronal scout MR image of the pelvis using a fast spoiled-gradient pulse sequence with a 500/2.1 ms repetition time/echo time, 31.2-kHz bandwidth, 24x24-cm field of view, 256x256 matrix, and 3 mm thickness. We then acquired an axial high-resolution MR image of the target arterial wall across the inflated balloon using (1) a spin-echo sequence (with 150/10 ms repetition/echo time, 16-kHz bandwidth, 6x6 cm field of view, 128x256 matrix, 1 to 3 numbers of excitation, and 3 mm thickness) and (2) a fast spoiled-gradient sequence (with 14.8/4.9 ms repetition/echo time, 15.6-kHz bandwidth, 4x4 field of view, 256x256 matrix, 8 numbers of excitation, and 3 mm thickness). The total scan time for each image was 1 minute. During the infusion of the gadolinium/blue dye/GFP lentivirus medium, the dilation balloon was inflated at 4 atm support pressure, and the medium infusion was constantly maintained using a pump (Harvard).
Histological and Immunohisto-chemical Confirmation
In the pigs infused with gadolinium/blue dye, we immediately harvested both target and control vessels for histopathological examination to confirm the success of the primary transfer. The pigs infused with gadolinium/GFP lentivirus were kept alive for 5 days to allow sufficient GFP expression. Then, we harvested both target and control arteries for immunohistochemical confirmation using a specific monoclonal antibody for GFP (Roche).6
Image Analysis
A region of interest was placed on the chosen portion of the target vessel wall to measure MR signal intensities. Then, we converted the measurements to obtain a signal intensity versus time curve. The results of MR images and histology/immunohistochemistry between the control and target tissues were analyzed blindly and were correlated by 3 investigators.
| Results |
|---|
|
|
|---|
Subsequently, in the remaining 2 pigs, we tested the infusion of the gadolinium/blue dye medium at a higher flow of 20 mL/hour for 15 minutes. Under this condition, most of the gadolinium flowed into the areas outside the target vessels (Figure 1), which, on surgical examination, correlated with the blue dye stain in the muscles adjacent to the target vessels (Figure 1).
|
In the preclinical validation studies, the target vessel walls began to be enhanced immediately after initiating the gadolinium/GFP lentivirus infusion (Figure 2). The average time period for the target vessel walls to maintain peak signal intensity was from 6±2 to minute 20±3 minutes after the initiation of the gadolinium/GFP lentivirus infusion, and the signal intensity dropped to the preinfusion level within 40 to 50 minutes (Figure 3). After the gadolinium/gene transfer, all 4 pigs recovered. There were no clinical signs of ischemia in the treated extremities during the 5 days after the procedure.
|
|
The subsequent laboratory examination showed, in all 12 pigs, the blue dye/GFP staining of the target vessel tissues with histology/immunohistochemistry. The expression of GFP, as determined in situ by immunohistochemistry, indicated a functional in vivo gene transfer and expression with our approach. Immunohistochemistry showed that GFP was expressed through all layers of the target vessel wall (Figure 2).
| Discussion |
|---|
|
|
|---|
Our study represents the first encouraging evidence that catheter-based vascular gene delivery can be monitored by MR imaging in vivo. It shows the potential of using MR imaging to visualize dynamically (1) where the gadolinium/genes are delivered, (2) how satisfactorily the target portion of the vessel wall is marked, and (3) whether the gene transfer procedure causes complications, such as perforation, as shown in Figure 1. During catheter-based gene transfer in an atherosclerotic vessel, calcified and unevenly-thickened vessel wall can be easily perforated, which would need to be detected immediately. Our technical development should help clinicians to determine the reasons for a failed vascular gene transfer procedure and promptly decide on further management by either repeating the same gene therapeutic procedure or choosing an alternative treatment.
This study focused only on technical development using an easily prepared mixture of gadolinium contrast with gene vectors. Further work is required to use this new technique to quantify gene delivery/MR imaging protocols using different parameters (such as applied transmural pressure, viral solution volume and concentration, and gene delivery flow rate and infusion time) with various gene delivery catheters. With continuous improvement in the high-resolution imaging capability of MR technology, it will be possible to monitor details regarding the extent of gadolinium/gene transmural distribution. This technique shows the ability to confirm the success of the primary gadolinium/gene delivery but does not give sufficient insight into biological expressivity.
We present a technical development using high-resolution MR imaging as an in vivo imaging tool to monitor catheter-based primary vascular gene delivery. We think that this work opens up an exciting avenue for the future efficient management of cardiovascular ischemic disorders using MR imaging-based vascular gene/drug therapy.
| Acknowledgments |
|---|
Received June 21, 2001; revision received August 8, 2001; accepted August 9, 2001.
| References |
|---|
|
|
|---|
2. Isner J. Manipulating angiogenesis against vascular disease. Hosp Pract. . 1999; 34: 6974.
3. Thomas J, Kuo M, Chawla M, et al. Vascular gene therapy. Radio Graphics. . 1998; 18: 13731394.[Abstract]
4. Kandarpa K. Catheter-based techniques for endovascular local drug delivery. J Vasc Interv Radiol Suppl. . 2000; 11: 419423.
5.
Yang X, Atalar E, Intravascular MR-guided balloon angioplasty using an MR imaging-guidewire: An in vivo feasibility study. Radiology. . 2000; 217: 501506.
6.
Yang X, Liu H, Li D, et al. Digital optical imaging of green fluorescent proteins for tracking vascular gene expression: feasibility study in rabbit and human cell models. Radiology. . 2001; 219: 171175.
7.
Weissleder R. Molecular imaging: exploring the next frontier. Radiology. . 1999; 212: 609614.
This article has been cited by other articles:
![]() |
X. Yang Nano- and Microparticle-based Imaging of Cardiovascular Interventions: Overview Radiology, May 1, 2007; 243(2): 340 - 347. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Du, B. Qiu, X. Zhan, A. Kolmakova, F. Gao, L. V. Hofmann, L. Cheng, S. Chatterjee, and X. Yang Radiofrequency-enhanced Vascular Gene Transduction and Expression for Intravascular MR Imaging-guided Therapy: Feasibility Study in Pigs Radiology, September 1, 2005; 236(3): 939 - 944. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. A. Krombach, J. G. Pfeffer, S. Kinzel, M. Katoh, R. W. Gunther, and A. Buecker MR-guided Percutaneous Intramyocardial Injection with an MR-compatible Catheter: Feasibility and Changes in T1 Values after Injection of Extracellular Contrast Medium in Pigs Radiology, May 1, 2005; 235(2): 487 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Feng, C. L. Dumoulin, S. Dashnaw, R. D. Darrow, R. L. DeLaPaz, P. L. Bishop, and J. Pile-Spellman Feasibility of Stent Placement in Carotid Arteries with Real-time MR Imaging Guidance in Pigs Radiology, February 1, 2005; 234(2): 558 - 562. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Saeed, R. Lee, A. Martin, O. Weber, G. A. Krombach, S. Schalla, M. Lee, D. Saloner, and C. B. Higgins Transendocardial Delivery of Extracellular Myocardial Markers by Using Combination X-ray/MR Fluoroscopic Guidance: Feasibility Study in Dogs Radiology, June 1, 2004; 231(3): 689 - 696. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Olin, J. A. Kaufman, D. A. Bluemke, R. O. Bonow, M. D. Gerhard, M. R. Jaff, G. D. Rubin, and W. Hall Atherosclerotic Vascular Disease Conference: Writing Group IV: Imaging Circulation, June 1, 2004; 109(21): 2626 - 2633. [Full Text] [PDF] |
||||
![]() |
R. C. Susil, A. Krieger, J. A. Derbyshire, A. Tanacs, L. L. Whitcomb, G. Fichtinger, and E. Atalar System for MR Image-guided Prostate Interventions: Canine Study Radiology, September 1, 2003; 228(3): 886 - 894. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Paturneau-Jouas, E. Parzy, G. Vidal, P. G. Carlier, C. Wary, J.-T. Vilquin, E. de Kerviler, K. Schwartz, and A. Leroy-Willig Electrotransfer at MR Imaging: Tool for Optimization of Gene Transfer Protocols--Feasibility Study in Mice Radiology, September 1, 2003; 228(3): 768 - 775. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Du, Y. Yang, C. Le Visage, H. H. Chen, R. DeJong, B. Qiu, D. Wang, K. W. Leong, U. M. Hamper, and X. Yang In Vivo US Monitoring of Catheter-based Vascular Delivery of Gene Microspheres in Pigs: Feasibility Radiology, August 1, 2003; 228(2): 555 - 559. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Yang Imaging of Vascular Gene Therapy Radiology, July 1, 2003; 228(1): 36 - 49. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Thompson, B. A. Nasseri, J. Makower, S. Houser, M. McGarry, T. Lamson, I. Pomerantseva, J. Y. Chang, H. K. Gold, J. P. Vacanti, et al. Percutaneous transvenous cellular cardiomyoplasty: A novel nonsurgical approach for myocardial cell transplantation J. Am. Coll. Cardiol., June 4, 2003; 41(11): 1964 - 1971. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P. Choudhury, V. Fuster, J. J. Badimon, E. A. Fisher, and Z. A. Fayad MRI and Characterization of Atherosclerotic Plaque: Emerging Applications and Molecular Imaging Arterioscler Thromb Vasc Biol, July 1, 2002; 22(7): 1065 - 1074. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |