(Circulation. 2001;104:1779.)
© 2001 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Thoraxcenter, Department of Cardiology, Erasmus Medical Center, Rotterdam, Netherlands.
Correspondence to Ad den Boer, Department of Cardiology, Thoraxcenter BD 428, PO Box 2040, 3000 GR Rotterdam, Netherlands. E-mail denboer{at}card.azr.nl
| Abstract |
|---|
|
|
|---|
Methods and Results The entrance point, beam size, and dose at the skin of the patient were calculated by use of the geometrical settings of gantry, investigation table, and x-ray beam and an ionization chamber. The data are displayed graphically. Three hundred twenty-two sequential cardiac investigations in adult patients were analyzed. The mean peak entrance dose per investigation was 0.475 Gy to a mean skin area of 8.2 cm2. The cumulative KERMA-area product per investigation was 52.2 Gy/cm2 (25.4 to 99.2 Gy/cm2), and the mean entrance beam size at the skin was 49.2 cm2. Twenty-eight percent of the patients (90/322) received a maximum dose of <1 Gy to a small skin area (
6 cm2), and 13.5% of the patients (42/322) received a maximum dose of >2 Gy.
Conclusions Monitoring of the dose distribution at the skin will alert the operator to the development of high-dose areas; by use of other gantry settings with nonoverlapping entrance fields, different generator settings, and extra collimation, skin lesion can be avoided.
Key Words: radiography catheterization dosage
| Introduction |
|---|
|
|
|---|
Newer x-ray units have an integrated ionization chamber that allows monitoring of the kinetic energy released in matter (KERMA; the absorbed dose in air)area product, but the entrance dose and dose distribution at the skin are not indicated. Therefore, there is a need to monitor and to quantify the skin radiation dose.2130
We have developed a system that automatically measures and monitors the accumulated skin radiation dose and allows detection of high-dose areas in real time.
In this study, the feasibility and usefulness of this technique have been investigated by analysis of the frequency distribution of high-dose areas in consecutive patients undergoing percutaneous intracoronary procedures.
| Methods |
|---|
|
|
|---|
Procedures
We developed a mathematical model to calculate the x-ray entrance field and the location of the irradiated areas of the skin. To be able to measure the x-ray dose for each 1 cm2 of irradiated surface, the following assumptions are made:
The following parameters for radiation monitoring are measured:
The entrance beam location at the skin was derived from the distance from the x-ray focal spot to the patient in the transverse plane by use of the gantry rotation, focal spot position, and patient position. This distance was corrected for the gantry angulation in the longitudinal plane. The entrance beam size was calculated from the collimator beam dimensions.
The radiation dose at the skin was calculated in grays by use of the entrance beam dimensions and the KAP, measured with an ionization chamber in the collimator.
The theoretical mathematical model is not presented in this article but is available from the authors. The model is used in a monitor system developed by the Siemens Co, the prototype of which is used to retrieve the data.
Graphical Display
Figure 1 shows a 3D graph with the dose distribution at a 50x50-cm dorsal aspect of the skin after an interventional procedure involving the LAD, including implantation of 2 stents. This graph was generated with our prototype system during a pilot study. The graphical display of the system used for data accumulation during our study shows a skin area of 90x90 cm.
|
Figure 2 is a plot of skin dose depicted in a 2D plane centered over the backbone; the skin surface is virtually split at the sternum and shown laid flat. This graph is automatically displayed on the x-ray monitor after radiation of any part of the skin exceeds a level of 1 Gy. The irradiated skin area is continuously displayed and updated, even when fluoroscopy is not used. This enables the physician to change gantry settings or use extra collimating to avoid possible overlapping with previously irradiated skin parts without using radiation.
|
Entrance Beam Dimensions
Both units have automated collimation that limits the beam area to the selected field of view of the image intensifier, independent of changes in the source-to-image distance (SID tracking). The area of the beam at the skin, however, does change with the source-to-image-intensifier distance. The maximum possible beam areas are given in Table 1.
|
Dose Settings of the X-Ray Equipment
The dose settings for the x-ray equipment for the pulsed fluoroscopy mode (PFM) and the digital cine mode (DCM) during the study period are shown in Table 2. The highest possible kilovoltage peak (kVp) during PFM was 110 kVp, and during DCM, 125 kVp. The frequency of PFM was adjusted to 12.5 pulses per second, and the extra beam filter used was 0.2 mm Cu.
|
The scattered radiation grid had a ratio of 11 and 40 line pairs per centimeter and a focal spot distance of 950 mm and was carbon fiber filled and covered. The radiation measurement was performed with an ionization chamber built into the collimator, and the KAP is shown.
Statistics
Unless otherwise stated, numerical data are presented as median with interquartile range.
| Results |
|---|
|
|
|---|
|
Radiation Time
The mean fluoroscopy time was 17.8 seconds (range 8.8 to 33.1 seconds), with pulsed fluoroscopy of 12.5 pulses per second. The mean DCM time was 97.9 seconds (range 69.4 to 151.1 seconds), with a frequency of 12.5 frames per second.
Irradiated Skin Surface
The radiation-exposed skin area in 95% of all investigations was restricted to a size of 50x50 cm, including the irradiated area at the groin for catheter introduction and left lateral gantry positioning. This is shown in Figure 1. In the remaining 5% of the cases, the exposed skin area could be visualized by showing a 90x90-cm area.
Peak Entrance Dose
The distribution of peak entrance skin doses to the patients in our series is shown in Figure 4. The mean value was 0.475 Gy, with a mean area of 8.2 cm2 exposed. Dose values >1 Gy occurred in the majority of the patients (52%), whereas parts receiving a dose >4 Gy occurred in only 1.2%. Because our system alerted us in real time when a high-dose area was developing, we were able to avoid excessive dose to that area by changing the gantry settings. It was not necessary to interrupt any procedure as a result of a high-dose "hot spot."
|
Correlation Between the Measured KAP, Patient Obesity, and Fluoroscopy Time
The correlation between the measured KAP and the obesity of the patient, the Quetelet index (kg/m2), was low (R=0.15). Higher doses were correlated with gantry projections that required penetration of thick, highly absorbent body masses. There was a strong relationship between the KAP and fluoroscopy time (R=0.78) (Figure 5).
|
Correlation Between Measured KAP and Peak Entrance Dose
There was a high correlation between measured KAP and peak entrance dose (R=0.89) (Figure 6). We also compared the correlation between KAP and the high-dose areas in the 134 diagnostic and the 188 interventional procedures. The correlation between the interventional procedures (R=0.90) was higher than the correlation between diagnostic procedures (R=0.35) because of more frequent changing of gantry settings and less overlapping of entrance fields in diagnostic cases.
|
Validation of the Measured KAP
The ionization chamber readings were verified regularly with other measuring devices; the accuracy of the KAP readings was within a 5% range. The values that appear on the monitor reflect the free-in-air KERMA, measured in grays. The dose to the tissue is actually 30% to 40% greater than the value displayed during monitoring, because of backscattered radiation and the KERMA-totissue-dose conversion factor.
| Discussion |
|---|
|
|
|---|
Our method allows for skin dose management and makes it possible to keep the dose at a certain skin area as low as reasonable for the procedure. This is important because many patients have more than one procedure, and skin dose accumulates.
It should be kept in mind that a high dose may produce unacceptable skin damage and should be avoided2630 (Table 3). Real-time dose monitoring can prevent these adverse effects because it allows selection of other x-ray techniques, such as extra beam filtering, selection of lower frame rates, and if available, documenting of low-dose fluoroscopic runs during the procedure.1518 Extra beam collimation is not often used but is a highly effective way to decrease both patient and operator radiation exposure without loss in image quality.
|
Finally, the use of other gantry settings allows distribution of the radiation dose over different skin areas and prevents development of a high dose. In our study, only 1.2% of the patients received a dose >4 Gy at some parts of the skin. It should be noted, however, that when equipment without dose-saving measures is used, levels are much higher (up to 7 times).15
Limitations
The monitoring system used in this study assumed a patient with a thorax circumference of 90 cm. Obviously, this does not apply to all patients, and ideally, a monitoring system should be tailored to the size of the patient. The actual size and location of irradiated skin parts may vary from one individual to another; future refinements to model the dimensions of individual patients more accurately would improve the accuracy of the skin dose distribution.
Clinical Recommendations
Our data show that the likelihood of high-dose areas may occur in the following circumstances: fluoroscopy time >60 minutes; gantry positioning unchanged throughout the procedure; and the irradiation occurring through a highly attenuating (eg, bone), thick body mass, requiring a high radiation quality during fluoroscopy and cinematography (>110 kVp). High-dose areas most likely occur when the left superior oblique projection is used.
Received May 7, 2001; revision received July 12, 2001; accepted July 31, 2001.
| References |
|---|
|
|
|---|
2. Calkins H, Niklason L, Sousa J, et al. Radiation exposure during radiofrequency catheter ablation of accessory atrioventricular connections. Circulation. 1991; 6: 23762382.
3. Cascade PN, Peterson LE, Wajszczuk WJ, et al. Radiation exposure to patients undergoing percutaneous transluminal coronary angioplasty. Am J Cardiol. 1987; 9: 996997.
4.
Huda W, Peters KR. Radiation-induced temporary epilation after a neuroradiologically guided embolization procedure. Radiology. 1994; 193: 642644.
5. Kovoor P, Rieciardello M, Collins L, et al. Risk to patients from radiation associated with radiofrequency ablation for supraventricular tachycardia. Circulation. 1998; 15: 15341540.
6. Malkinson FD. Radiation injury to skin following fluoroscopically guided procedures. Arch Dermatol. 1996; 6: 695696.
7. Geise RA, Peters NE, Dunnigan A, et al. Radiation dose during pediatric radiofrequency catheter ablation procedures. Pacing Clin Electrophysiol. 1996; 11: 16051611.
8. Karpinnen J, Parviainen T, Servomaa A, et al. Radiation risk and exposure of radiologists and patients during coronary angiography and percutaneous transluminal coronary angioplasty (PTCA). Radiat Prot Dosimetry. 1995; 57: 481485.[Abstract]
9. Sovik E, Klow NE, Hellesnes J, et al. Radiation induced skin injury after percutaneous transluminal coronary angioplasty: case report. Acta Radiol. 1996; 3: 305306.
10. Nahass GT. Acute radiodermatitis after radiofrequency catheter ablation. J Am Acad Dermatol. 1997; 5: 881884.
11. Lichtenstein DA, Klapholz L, Vardy DA, et al. Chronic radiodermatitis following cardiac catheterization. Arch Dermatol. 1996; 6: 663667.
12. DIncan M, Roger H. Radiodermatitis following cardiac catheterization. Arch Dermatol. 1997; 2: 242243.
13. Shope TB. Radiation-induced skin injuries from fluoroscopy. Radiographics. 1996; 5: 11951199.
14. Knautz MA, Abele DC, Reynolds TL. Radiodermatitis after transjugular intrahepatic portosystemic shunt. South Med J. 1997; 3: 352356.
15. Den Boer A, de Feijter PJ, Hummel WA, et al. Reduction of radiation exposure while maintaining high-quality fluoroscopic images during interventional cardiology using novel x-ray tube technology with extra beam filtering. Circulation. 1994; 6: 27102714.
16. Pitney MR, Allan RM, Giles RW, et al. Modifying fluoroscopic views reduces operator radiation exposure during coronary angioplasty. J Am Coll Cardiol. 1994; 7: 16601663.
17. Food and Drug Administration. Public Health Advisory. Avoidance of Serious X-ray Induced Skin Injuries to Patients During Fluoroscopically-Guided Procedures. Rockville, Md: Center for Devices and Radiological Health, FDA: 1994, September 30.
18. Food and Drug Administration. Important Information for Physicians and Other Healthcare Professionals: Recording Information in the Patients Medical Record That Identifies the Potential for Serious X-ray Induced Skin Injuries Following Fluoroscopic Guided Procedures. Rockville, Md: Center for Devices and Radiological Health, FDA: 1995, September 15.
19. Balter S. Guidelines for personnel radiation monitoring in the cardiac catheterization laboratory. Laboratory Performance Standards Committee of the Society for Cardiac Angiography and Interventions. Cathet Cardiovasc Diagn. 1993; 4: 277279.
20. Vano E, Arranz L, Sastre JM at al. Dosimetric and radiation protection considerations based on some cases of patient skin injuries in interventional cardiology. Br J Radiol. 1998; 845: 510516.
21. Sellers TD, Olley MC, DiLorenzo DR, et al. Radiation exposure during radiofrequency catheter ablation of accessory pathways. Pacing Clin Electrophysiol. 1992; 15: 582584.
22. Bakalyar DM, Castellani MD, Safian RD. Radiation exposure to patients undergoing percutaneous transluminal coronary angioplasty. Cathet Cardiovasc Diagn. 1997; 2: 121125.
23. Meier B. Radiation exposure in the cardiac catheterization laboratory: an issue or a non-issue? Cathet Cardiovasc Diagn. 1997; 4: 352.
24. Watson RM. Radiation exposure: clueless in the cath lab, or sayonara ALARA. Cathet Cardiovasc Diagn. 1997; 2: 126127.
25. Zorzetto M, Bernardi G, Morrocutti G, et al. Radiation exposure to patients and operators during diagnostic catheterization and coronary angioplasty. Cathet Cardiovasc Diagn. 1997; 4: 348351.
26. Radiation-induced skin injuries from fluoroscopy. Food and Drug Administration Medical Bulletin. 1994; 2426.
27. Pattee PL, Johns PC, Chambers RJ. Radiation risk to patient from percutaneous transluminal coronary angioplasty. J Am Coll Cardiol. 1993; 4: 10441051.
28. Wagner LK, Eifel PJ, Geise RA. Potential biological effects following high x-ray dose interventional procedures. J Vasc Interv Radiol. 1994; 1: 7184.
29. Faulkner K, Love HG, Sweeny JK, et al. Radiation dose and somatic risk to patients during cardiac radiological procedures. Br J Radiol. 1986; 700: 359363.
30. Goldschmidt H, Sherwin WK. Reactions to ionizing radiation. J Am Acad Dermatol. 1980; 6: 551579.
This article has been cited by other articles:
![]() |
V. Tsapaki, S. Kottou, S. Korniotis, N. Nikolaki, S. Rammos, and S. C. Apostolopoulou RADIATION DOSES IN PAEDIATRIC INTERVENTIONAL CARDIOLOGY PROCEDURES Radiat Prot Dosimetry, January 2, 2009; (2009) ncn313v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Chida, Y. Kagaya, H. Saito, Y. Takai, S. Takahashi, S. Yamada, M. Kohzuki, and M. Zuguchi Total Entrance Skin Dose: An Effective Indicator of Maximum Radiation Dose to the Skin During Percutaneous Coronary Intervention Am. J. Roentgenol., October 1, 2007; 189(4): W224 - W227. [Abstract] [Full Text] [PDF] |
||||
![]() |
R E Morrell and A T Rogers A mathematical model for patient skin dose assessment in cardiac catheterization procedures. Br. J. Radiol., September 1, 2006; 79(945): 756 - 761. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Suzuki, S. Furui, H. Kohtake, T. Takeshita, M. Suzuki, K. Kozuma, Y. Yamamoto, and T. Isshiki Radiosensitive Functional Dye: Clinical Application for Estimation of Patient Skin Dose Radiology, May 1, 2006; 239(2): 541 - 546. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Chida, H. Saito, H. Otani, M. Kohzuki, S. Takahashi, S. Yamada, K. Shirato, and M. Zuguchi Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. Am. J. Roentgenol., March 1, 2006; 186(3): 774 - 778. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. S. Storm, D. L. Miller, L. J. Hoover, J. D. Georgia, and T. Bivens Radiation Doses from Venous Access Procedures Radiology, March 1, 2006; 238(3): 1044 - 1050. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Vano, L. Gonzalez, E. Guibelalde, P. Aviles, J. M. Fernandez, C. Prieto, and C. Galvan Evaluation of risk of deterministic effects in fluoroscopically guided procedures Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 190 - 194. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Padovani, G. Bernardi, E. Quai, M. Signor, H. S. Toh, G. Morocutti, and L. Spedicato Retrospective evaluation of occurrence of skin injuries in interventional cardiac procedures Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 247 - 250. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Geleijns and J. Wondergem X-ray imaging and the skin: radiation biology, patient dosimetry and observed effects Radiat Prot Dosimetry, May 17, 2005; 114(1-3): 121 - 125. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Bacher, E. Bogaert, R. Lapere, D. De Wolf, and H. Thierens Patient-Specific Dose and Radiation Risk Estimation in Pediatric Cardiac Catheterization Circulation, January 4, 2005; 111(1): 83 - 89. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Guibelalde, L Gonzalez, and E Vano Suitability of resin-coated photographic paper for skin dose measurement during fluoroscopically-guided X-ray procedures Br. J. Radiol., October 1, 2004; 77(922): 871 - 875. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Kuon, J Birkel, M Schmitt, and J B Dahm Radiation exposure benefit of a lead cap in invasive cardiology Heart, October 1, 2003; 89(10): 1205 - 1210. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Kuon, C Glaser, and J B Dahm Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures Br. J. Radiol., June 1, 2003; 76(906): 406 - 413. [Abstract] [Full Text] [PDF] |
||||
![]() |
E Vano, C Prieto, J M Fernandez, L Gonzalez, M Sabate, and C Galvan Skin dose and dose-area product values in patients undergoing intracoronary brachytherapy Br. J. Radiol., January 1, 2003; 76(901): 32 - 38. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |