(Circulation. 2001;104:1999-2002.)
© 2001 American Heart Association, Inc.
Brief Rapid Communications |
From Lenox Hill Heart and Vascular Institute, New York, NY.
Correspondence to Gary S. Roubin, MD, PhD, 130 East 77th Street, New York, NY 10021. E-mail Groubin{at}lenoxhill.net
Received July 23, 2001; revision received September 7, 2001; accepted September 10, 2001.
| Abstract |
|---|
|
|
|---|
Methods and Results Using transcranial Doppler, we compared the frequency of MES during CAS in 2 groups: 39 patients without distal protection and 37 who used the distal-balloon protection system (GuardWire). There were no significant differences in the clinical or angiographic characteristics between the 2 groups. Three phases with increased MES counts were identified during unprotected CAS; these were stent deployment, predilation, and postdilation (75±57, 32±36, and 27±25 METS, respectively). The distal-balloon protection significantly reduced the frequency of MES during CAS (MES counts: 164±108 in the control versus 68±83 in the protection group; P=0.002), particularly during these 3 phases. MES in the protection group were detected predominantly during sheath placement, guidewire manipulation, and distal-balloon deflation.
Conclusion Three phases with increased MES counts were identified during unprotected CAS (eg, stent deployment, predilation, and postdilation). The distal-balloon protection system significantly reduced the frequency of MES during CAS, particularly during these 3 phases.
Key Words: carotid arteries angioplasty stents revascularization embolism
| Introduction |
|---|
|
|
|---|
5% of the cases.13,5 Efforts are ongoing to optimize the outcome of this treatment by developing strategies that eliminate the risk of embolic events.5 One such strategy is distal-balloon protection.6 By occluding the distal carotid artery with a balloon, embolic material released during the intervention is contained within the proximal blood column, which is subsequently aspirated before the protective balloon is deflated. Transcranial Doppler studies have documented the prevalence of microembolization during unprotected CAS.7,8 Subclinical microemboli produce characteristic Doppler signals (microembolic signals, MES) that, in a variety of clinical conditions, correlate with increased risk for embolic neurological events.810 Thus, MES may serve as a surrogate marker of embolic events during CAS and facilitate the assessment of various distal-protection strategies. The higher frequency of MES compared with the clinical events in this setting allows such an assessment in a smaller number of patients.7,8 In this study, we evaluated the effect of distal-balloon protection on the frequency of the Doppler-detected MES during CAS.
| Methods |
|---|
|
|
|---|
60% or
80% diameter obstruction, respectively, as determined by quantitative angiography (Phillips Integris) using the methodology from the North American Symptomatic Carotid Endarterectomy Trial (NASCET).11 Consecutive patients enrolled into either of the 2 protocols at our institution who had an optimal Doppler insonation window and who were cooperative with the transcranial Doppler monitoring procedure were enrolled in the present study.
Stenting Protocol
All procedures were performed in accordance with the guidelines of the Institutional Review Board. CAS was performed using the standard technique, as previously described.12 All patients were given 75 mg of clopidogrel (Plavix, Pfizer Inc) once a day and 325 mg of aspirin twice a day for 7 days before the procedure. A bolus of heparin (4000 U) was administered during the procedure. Glycoprotein IIb/IIIa antagonists were administered at the operators discretion. The GuardWire system was applied as previously described using a single-stage distal-balloon inflation with only-aspiration technique.6
Transcranial Doppler Protocol
Transcranial Doppler signals, using a multichannel system (EME-Pioneer-4040, Nicollet-Vascular), were recorded from the ipsilateral middle cerebral artery via the transtemporal window. Monitoring was started before the procedure and continued for 10 minutes after the procedure whenever possible. Standard settings were used, with 2 sample volumes of 10-mm in diameter, each at a depth of 46 to 56 mm (4 to 6 mm apart). Data were stored on a hard disk using a coding system and were analyzed off-line on a later day by an investigator who was blinded to patient information. MES were identified according to the recommended guidelines.13 The procedure was divided into the following phases: (1) sheath placement, (2) wiring, (3) predilation, (4) stent deployment, (5) postdilation, (6) aspiration phase (protection group only), and (7) angiography. These phases were defined from the time the respective catheter was introduced into the guiding sheath until it was removed.
Statistical Analysis
MES counts are expressed as mean±SD. By convention, 1 second of microembolic shower signal is considered 10 MES.7 All MES recorded during contrast injections were excluded from the analysis because they may indicate microbubbles, which may be less hazardous than solids.14 Data were compared using
2 for categorical and Students t test for continuous variables.
| Results |
|---|
|
|
|---|
|
Table 2 summarizes the angiographic and procedural data. There were no significant differences in angiographic lesion characteristics or residual stenosis between the 2 groups. A bolus of a glycoprotein IIb/IIIa antagonist (Aggrastat, Pfizer Inc.) was administered to 4 patients in the control group and 1 patient in the protection group. Self-expanding stents were used in all patients. The Wallstent (Boston Scientific) was used exclusively in the protection group but in only 49% of patients in the control group.
|
Two procedural, nondisabling embolic strokes occurred, one in each group. Both improved completely within 3 days. One patient in the protection group died 2 days after CAS from a subarachnoidal hemorrhage complicating hyperperfusion syndrome that developed immediately after intervention. His procedure was technically uncomplicated and notable for its low MES count.15 Baseline head MRI showed multiple lacunar infarctions but no acute ischemic changes. At 30 days, there have been no additional neurological events and no deaths in either group.
Microembolic Signals
MES were detected during all procedural phases in both groups (Table 3). Three phases with increased MES counts were identified in the control group; these were stent deployment, predilation, and postdilation (Figure 1). The GuardWire significantly reduced the total MES count compared with the control group (68±83 versus 164±108 MES; P=0.002), particularly during these 3 phases (Table 3). The combined MES count during the unprotected phases of the procedure (sheath placement and wire manipulation) was 30±24 in control group, and 34±41 in the protection group (P=NS). In the protection group, MES occurred predominantly during sheath placement, wire manipulation, and distal balloon deflation (Table 3). During the postprocedural observation period (mean, 9±5 minutes), no MES were detected in any of the patients.
|
|
On further analysis of the control group, there was no significant difference in the MES counts between the patients who had a Wallstent versus those who had other self-expanding stents (156±93 versus 170±101 MES, respectively; P=NS). A significant difference was absent throughout the various steps, including the stenting phase (67±55 versus 78±63 MES; P=NS). Both subgroups were matched in their clinical and angiographic characteristics.
| Discussion |
|---|
|
|
|---|
Several factors may have contributed to the MES observed in the protection group. These include (1) a nonocclusive distal balloon; (2) embolization via the external carotid artery (Figure 2), which questions the suitability of the saline-flushing maneuver into the internal carotid artery before the distal-balloon deflation, diverting potential emboli into the external carotid artery6; and (3) trapped emboli within the "shadow-zone" of the aspiration catheter that were inaccessible for removal.
|
Although the evidence that MES represent microemboli is well established, their use as a surrogate marker of embolic events during CAS may have a potential limitation.7,8 Although high MES counts correlate with an increased risk for embolic stroke in many clinical conditions, such a correlation during CAS has yet to be established. Because of the small number of clinical events, this correlation could not be examined in the present series. In addition, due to the logistics of enrolling consecutive patients into CAFE-USA when the technique became available at our institution, a randomized design of this study was not possible. However, the two groups were well matched in their clinical and angiographic characteristics. Subgroup analysis of the control group suggests that the difference in the stent type between the two groups did not seem to impact MES counts. Nevertheless, these data provide an important insight into the embolization profile during CAS and into the performance of the distal-balloon protection.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Y Iguchi, K Kimura, K Kobayashi, Y Ueno, K Shibazaki, and T Inoue Microembolic signals at 48 hours after stroke onset contribute to new ischaemia within a week J. Neurol. Neurosurg. Psychiatry, March 1, 2008; 79(3): 253 - 259. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. K. Eskandari Preventable Complications of Carotid Stenting Perspectives in Vascular Surgery and Endovascular Therapy, March 1, 2008; 20(1): 17 - 25. [Abstract] [PDF] |
||||
![]() |
E. Mahmud, J. J. Cavendish, and A. Salami Current Treatment of Peripheral Arterial Disease: Role of Percutaneous Interventional Therapies J. Am. Coll. Cardiol., August 7, 2007; 50(6): 473 - 490. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Maynar, S. Baldi, R. Rostagno, T. Zander, M. Rabellino, R. Llorens, J. Alvarez, and F. Barajas Carotid Stenting without Use of Balloon Angioplasty and Distal Protection Devices: Preliminary Experience in 100 Cases AJNR Am. J. Neuroradiol., August 1, 2007; 28(7): 1378 - 1383. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Clark, S. Lessio, M. O'Donoghue, C. Tsalamandris, R. Schainfeld, and K. Rosenfield Mechanisms and Predictors of Carotid Artery Stent Restenosis: A Serial Intravascular Ultrasound Study J. Am. Coll. Cardiol., June 20, 2006; 47(12): 2390 - 2396. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Roubin, S. Iyer, A. Halkin, J. Vitek, and C. Brennan Realizing the Potential of Carotid Artery Stenting: Proposed Paradigms for Patient Selection and Procedural Technique Circulation, April 25, 2006; 113(16): 2021 - 2030. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. J. White Catheter-Based Therapy for Atherosclerotic Renal Artery Stenosis Circulation, March 21, 2006; 113(11): 1464 - 1473. [Full Text] [PDF] |
||||
![]() |
J. C. Lin, R. R. Kolvenbach, and L. Pinter Protected Carotid Artery Stenting and Angioplasty via Transfemoral versus Transcervical Approaches Vascular and Endovascular Surgery, November 1, 2005; 39(6): 499 - 503. [Abstract] [PDF] |
||||
![]() |
D. A. Gorog, R. A. Foale, and I. Malik Distal Myocardial Protection During Percutaneous Coronary Intervention: When and Where? J. Am. Coll. Cardiol., October 18, 2005; 46(8): 1434 - 1445. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. I. Bluth, J. A. Vos, and R. G. A. Ackerstaff Value of US in Selecting Patients for Carotid Angioplasty and Stent Placement * Drs Vos and Ackerstaff respond: Radiology, October 1, 2005; 237(1): 374 - 375. [Full Text] [PDF] |
||||
![]() |
H. D. Aronow, M. Shishehbor, D. A. Davis, I. L. Katzan, D. L. Bhatt, C. T. Bajzer, A. Abou-Chebl, K. W. Derk, P. L. Whitlow, and J. S. Yadav Leukocyte Count Predicts Microembolic Doppler Signals During Carotid Stenting: A Link Between Inflammation and Embolization Stroke, September 1, 2005; 36(9): 1910 - 1914. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Orlandi, S. Fanucchi, S. Gallerini, C. Sonnoli, M. Cosottini, M. Puglioli, F. Sartucci, and L. Murri Impaired Clearance of Microemboli and Cerebrovascular Symptoms During Carotid Stenting Procedures Arch Neurol, August 1, 2005; 62(8): 1208 - 1211. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. P. Lownie, D. M. Pelz, D. H. Lee, S. Men, I. Gulka, and P. Kalapos Efficacy of Treatment of Severe Carotid Bifurcation Stenosis By Using Self-Expanding Stents without Deliberate Use of Angioplasty Balloons AJNR Am. J. Neuroradiol., May 1, 2005; 26(5): 1241 - 1248. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Orlandi, S. Gallerini, M. Cosottini, L. Murri, L. K. Sztriha, E. Voros, P. Szikra, and L. Vecsei Postprocedural Emboli in Carotid Artery Stenting: Where Do They Come From? * Response: Stroke, May 1, 2005; 36(5): 928 - 929. [Full Text] [PDF] |
||||
![]() |
J. A. Vos, J. C. van den Berg, S. M. P. G. Ernst, M. J. Suttorp, T. T. C. Overtoom, H. W. Mauser, O. J. M. Vogels, H. P. M. van Heesewijk, F. L. Moll, Y. van der Graaf, et al. Carotid Angioplasty and Stent Placement: Comparison of Transcranial Doppler US Data and Clinical Outcome with and without Filtering Cerebral Protection Devices in 509 Patients Radiology, February 1, 2005; 234(2): 493 - 499. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. Sztriha, E. Voros, K. Sas, R. Szentgyorgyi, A. Pocsik, P. Barzo, P. Szikra, A. Makai, A. Szolics, P. Elek, et al. Favorable Early Outcome of Carotid Artery Stenting Without Protection Devices Stroke, December 1, 2004; 35(12): 2862 - 2866. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schmidt, K.-W. Diederich, S. Scheinert, S. Braunlich, T. Olenburger, G. Biamino, G. Schuler, and D. Scheinert Effect of two different neuroprotection systems on microembolization during carotid artery stenting J. Am. Coll. Cardiol., November 16, 2004; 44(10): 1966 - 1969. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Zahn, B. Mark, N. Niedermaier, U. Zeymer, P. Limbourg, T. Ischinger, K. Haerten, K. E. Hauptmann, E.-R. v. Leitner, W. Kasper, et al. Embolic protection devices for carotid artery stenting: better results than stenting without protection? Eur. Heart J., September 1, 2004; 25(17): 1550 - 1558. [Abstract] [Full Text] |
||||
![]() |
H. E. Rodriguez and L. Leon Current Status of Carotid Angioplasty and Stenting Perspectives in Vascular Surgery and Endovascular Therapy, June 1, 2004; 16(2): 111 - 119. [Abstract] [PDF] |
||||
![]() |
D. P. Slovut and J. W. Olin Fibromuscular Dysplasia N. Engl. J. Med., April 29, 2004; 350(18): 1862 - 1871. [Full Text] [PDF] |
||||
![]() |
J. J. Ferguson If There's Smoke, Is There Fire? Circulation, March 30, 2004; 109(12): 1442 - 1444. [Full Text] [PDF] |
||||
![]() |
J. D. Barr, J. J. Connors III, D. Sacks, J. C. Wojak, G. J. Becker, J. F. Cardella, B. Chopko, J. E. Dion, A. J. Fox, R. T. Higashida, et al. Quality Improvement Guidelines for the Performance of Cervical Carotid Angioplasty and Stent Placement: Developed by a Collaborative Panel of the American Society of Interventional and Therapeutic Neuroradiology, the American Society of Neuroradiology, and the Society of Interventional Radiology AJNR Am. J. Neuroradiol., November 1, 2003; 24(10): 2020 - 2034. [Full Text] [PDF] |
||||
![]() |
M. Schluter, T. Tubler, J. C. Steffens, D. G. Mathey, and J. Schofer Focal ischemia of the brain after neuroprotected carotid artery stenting J. Am. Coll. Cardiol., September 17, 2003; 42(6): 1007 - 1013. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Yadav Protecting the brain:how do we measure success? J. Am. Coll. Cardiol., September 17, 2003; 42(6): 1014 - 1016. [Full Text] [PDF] |
||||
![]() |
G. W. Stone, C. Rogers, J. Hermiller, R. Feldman, P. Hall, R. Haber, A. Masud, P. Cambier, R. P. Caputo, M. Turco, et al. Randomized Comparison of Distal Protection With a Filter-Based Catheter and a Balloon Occlusion and Aspiration System During Percutaneous Intervention of Diseased Saphenous Vein Aorto-Coronary Bypass Grafts Circulation, August 5, 2003; 108(5): 548 - 553. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Liistro and C. Di Mario Carotid artery stenting Heart, August 1, 2003; 89(8): 944 - 948. [Full Text] [PDF] |
||||
![]() |
B. Eckert and H. Zeumer Editorial Comment--Carotid Artery Stenting With or Without Protection Devices? Strong Opinions, Poor Evidence! Stroke, August 1, 2003; 34(8): 1941 - 1943. [Full Text] [PDF] |
||||
![]() |
A. Kastrup, K. Groschel, H. Krapf, B. R. Brehm, J. Dichgans, and J. B. Schulz Early Outcome of Carotid Angioplasty and Stenting With and Without Cerebral Protection Devices: A Systematic Review of the Literature Stroke, March 1, 2003; 34(3): 813 - 819. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Macdonald and P. A Gaines Current concepts of mechanical cerebral protection during precutaneous carotid intervention Vascular Medicine, February 1, 2003; 8(1): 25 - 32. [Abstract] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2001 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |