Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 2001;104:2436-2441
doi: 10.1161/hc4501.099317
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Grube, E.
Right arrow Articles by Stertzer, S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Grube, E.
Right arrow Articles by Stertzer, S.
Related Collections
Right arrow Coronary circulation
Right arrow Catheter-based coronary and valvular interventions: other

(Circulation. 2001;104:2436.)
© 2001 American Heart Association, Inc.


Clinical Investigation and Reports

Prevention of Distal Embolization During Coronary Angioplasty in Saphenous Vein Grafts and Native Vessels Using Porous Filter Protection

Eberhard Grube, MD; Ulrich Gerckens, MD; Alan C. Yeung, MD; Sascha Rowold; Nicole Kirchhof, DVM; Jerry Sedgewick, BA; Jay S. Yadav, MD; Simon Stertzer, MD

From the Department of Cardiology/Angiology, Heart Center Siegburg, Siegburg, Germany (E.G., U.G., S.R.); the Division of Cardiovascular Medicine, Stanford University, Stanford, Calif (A.C.Y., S.S.); the University of Minnesota, Minneapolis (N.K., J.S.); and the Departments of Cardiology and Molecular Cardiology and the Joseph J. Jacobs Center for Thrombosis and Vascular Biology, The Cleveland Clinic Foundation, Cleveland, Ohio (J.S.Y.).

Correspondence to Eberhard Grube, MD, Department of Cardiology/Angiology, Heart Center Siegburg, 53721 Siegburg, Germany. E-mail GrubeE{at}aol.com


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background— Although distal embolization and the "no-reflow" phenomenon are well described in saphenous vein graft (SVG) interventions, the frequency, magnitude, and characterization of embolized debris have not been evaluated in routine coronary interventions. A unique embolus protection device described herein provides a means of containing and retrieving plaque material dislodged during percutaneous coronary interventions. This report details the first clinical experience of the effectiveness and safety of an emboli protection system in 11 SVG lesions and 15 native coronary artery lesions.

Methods and Results— The AngioGuard Emboli Capture Guidewire (Cordis) consists of a PTCA wire with an expandable filter at the distal tip. The porous membrane permits normal distal blood flow, while trapping potential emboli by filtration. After crossing the lesion, the filter is expanded, and routine angioplasty is performed over the same wire. Emboli retrieval is achieved by collapsing the filter and retracting the emboli capture wire (ECW). In 26 patients, standard angioplasty was performed over the ECW; 20 of these 26 patients received a stent. Collected debris was sent for histopathological analysis. Plaque debris was retrieved after native coronary and SVG interventions in all cases. The ECW was positioned and retrieved without complications. No major adverse events occurred. Myocardial infarctions and no-reflow were not observed.

Conclusions— The embolization of plaque fragments frequently occurs during coronary and SVG intervention. Distal embolization leading to microvascular obstruction and no-reflow could be successfully minimized by using the ECW.


Key Words: angioplasty • embolism • grafting


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Percutaneous intervention for atherosclerosis has become well established for coronary and peripheral arteries and has recently been extended to the carotid vessels. Embolization has not been considered a prohibitive issue in most native coronary or peripheral intervention. However, the risk of distal embolization has been considered significant in the carotid arteries, in degenerated saphenous vein grafts (SVG), and in thrombotic lesions seen in the acute coronary syndrome.13 In degenerated SVGs, the risk of distal embolization is especially high, with the "no-reflow" phenomenon being reported in up to 31.8% of cases when there is evidence of thrombus and in up to 7.9% of cases with no thrombus present.4 This complication increases the risk of myocardial infarction and in-hospital death.4,5 Accordingly, a guidewire incorporating a porous filter (100-µm pores), the AngioGuard emboli capture guidewire (ECW; Cordis) was developed to prevent embolization in carotid arteries and SVGs.

This study reports the first clinical results with this ECW and establishes the frequency and magnitude of embolization in native coronary artery and SVG interventions and whether the ECW is a plausible method of preventing embolic myocardial infarction and no-reflow.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Device Description
The ECW consists of a core wire (diameter of 0.014 inches), with an integrated wire basket and a polymeric filter at the distal end of the device. The core wire is made of stainless steel. The basket consists of a thin, porous, polymeric membrane supported by a fine metal skeleton (Figure 1). The basket material is made of nickel-titanium, and the filter is made of polyurethane (pore size, 100 µm). The basket diameters vary from 4.0 to 8.0 mm and fit into 3.0 to 7.5 mm coronary arteries or SVGs. The device is inserted into the body with an outer delivery sheath (2.5F) that is used to hold the filter basket in the closed position. The use of the ECW requires a guide catheter lumen diameter of 7F/0.081 inches. After crossing the lesion and reaching the distal site, the outer sheath is then pulled back, allowing basket deployment. The wire has radiopaque marking, allowing exact placement of the basket (Figure 2). During the procedure as the blood passes through the filter, emboli collect in the filter (Figure 3). Once the procedure is completed, a second outer sleeve is inserted to close the filter and remove the device.



View larger version (135K):
[in this window]
[in a new window]
 
Figure 1. Magnified view of the collection basket.



View larger version (150K):
[in this window]
[in a new window]
 
Figure 2. A, View of the ECW deployed during PTCA of a right coronary artery. B, ECW before deployment. C, Expanded filter in vessel.



View larger version (145K):
[in this window]
[in a new window]
 
Figure 3. Magnified view of a filter with captured emboli.

Study Design
In 26 consecutive patients (18 men; mean age, 66±2.7 years), coronary or SVG angioplasty with the support of the ECW was performed. All other aspects of the intervention (eg, balloon angioplasty, stent deployment, or adjunctive pharmacotherapy) were performed at the operator’s discretion. All patients were treated under a protocol approved by the Institutional Ethics Committee of Freiburg, Germany. Patients were admitted to the study with acute infarction (2 patients, 7.7%), unstable coronary syndrome (11 patients, 42.3%), or stable angina (13 patients, 50%). All patients were given aspirin (100 mg/d) and clopidogrel (300 mg loading dose) before stent implantation and a bolus of 7500 IU of heparin. One patient was given tirofiban before the intervention. In 25 patients, the femoral approach was used, and in one patient, right brachial access was chosen. All patients had an ECG before and after the intervention. Serial creatine kinase (CK) and CK-MB levels (upper limit of normal value, 80 U/L for CK and 20 U/L for CK-MB) were drawn before and 6, 12 to 18, and 24 hours after the procedure. TIMI flow was assessed before device placement, before intervention with the filter basket open distal to the lesion, and after completion of the procedure, with and without the ECW in the vessel. The criteria for clinical evidence of distal embolization were occlusion of a distal branch, onset of worsened flow in a distal segment of the target vessel during the procedure, or significant postinterventional rise of CK level, which was defined as CK values more than twice the upper limit, with positive MB.

In the first 20 patients, immediately on withdrawal from the patient, the distal end of the retrieval device was placed in a test tube of 10% neutral buffered formalin and cut proximal to the basket. The test tube containing the nitinol basket, filter, and distal wire tip was sent to the pathology laboratory at the University of Minnesota for analysis.

After fixation, morphometric analysis of the size and number of particles adherent to the filter and free-floating in the fixative was performed using computerized edge detection software. Surface area calculations were performed using the 2 longest orthogonal diameters. At least 20 sections per sample were stained with hematoxylin and eosin. Sections were also stained for van Gieson, Masson’s trichrome stain, Alcian blue, phosphotungstic acid hematoxylin, and van Kossa.

Immunohistochemical staining was performed with mouse monoclonal antibodies (DAKO) to smooth muscle actin, glycoprotein IIb/IIIa (CD41), and neutrophil elastase. For negative controls, the primary antibody was replaced with normal serum. Tissue sections known to express the relevant antigens served as positive controls. The stained specimens were analyzed by a single observer who was blinded to the patient’s clinical diagnosis. The specimens were analyzed for (1) the presence of extracellular matrix with single cells as part of an atherosclerotic plaque, (2) the presence of foam cells, (3) the presence of calcified deposits, (4) the presence of neutrophils, (5) the presence of artifacts, and (6) the presence of thrombus and cellular components of blood, such as platelets.

Data Collection
Clinical report forms were completed; they noted major adverse cardiac events (death, myocardial infarction, emergent CABG) and angiographic adverse events (abrupt closure, perforation, major dissection, or no-reflow). Periprocedural myocardial infarction was defined as total CK values more than twice the upper limit with positive CK-MB. Minor complications (ie, transient ST elevation, pseudoaneurysm at puncture site) occurring during the procedure and hospitalization were also recorded. On the clinical report forms, specific emphasis was placed on the assessment of TIMI flow and percent diameter stenosis before and after the procedure and adverse events. Percent diameter stenosis was calculated by visual caliper using the "worst view" method. The procedure was considered technically successful if there was successful device placement, filter expansion, and retraction. The procedure was considered angiographically successful if postinterventional stenosis was <50% and there were no device-related, clinical, or angiographic adverse events by the end of the hospitalization.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
The patient and lesion characteristics are shown in Tables 1 and 2. The periprocedural CK/CK-MB levels are shown in Table 2, and the individual TIMI flow grade is listed in Tables 3 and 4. In the SVG group, which had a mean graft age of 8.2 years, we included ostial, midgraft, and distal lesions. The final angiographic result showed a residual stenosis of 3.2% in the native coronary arteries and 5.9% in SVGs. Two patients (patients 12 and 15 in Table 3) with evolving myocardial infarction were included in the study to evaluate the feasibility of the ECW in this clinical setting. Except for the 2 patients presenting with acute myocardial infarction, there were no significant rises in the CK levels (as defined above) or in CK-MB levels more than the upper limit of normal value, either in the native coronary patients or in those with degenerated SVG lesions.


View this table:
[in this window]
[in a new window]
 
Table 1. Patient Demographics


View this table:
[in this window]
[in a new window]
 
Table 2. Lesion Characteristics and CK Levels


View this table:
[in this window]
[in a new window]
 
Table 3. TIMI Flow in Native Coronary Arteries


View this table:
[in this window]
[in a new window]
 
Table 4. TIMI Flow in SVGs

The technical success rate was 96.2%; in one patient (patient 6, Table 3) the ECW wire could not pass a distal lesion in the right coronary artery that had a shepherd’s crook configuration. There were no major adverse events related to the device itself. Two patients developed transient ST-segment elevation. One had an occlusive dissection after stent implantation and was treated immediately with a second distal stent. No subsequent CK/CK-MB elevation or Q-waves were observed. The second patient underwent recanalization of an occluded right coronary artery with flow decrease in a side branch during the procedure, presumably due to a completely filled basket, which was placed immediately after the first successful lesion crossing after predilatation using a small balloon catheter. A large thrombus was extracted; this would have caused distal embolization in an unprotected vessel. In this patient, there was a maximal CK/CK-MB of 102/18 U/L at follow-up. Debris was retrieved from all patients, and in 24 of the 26, it was grossly visible (Figure 4).



View larger version (135K):
[in this window]
[in a new window]
 
Figure 4. Example of collected debris.

Native Coronary Arteries
The initial flow in 10 of 13 patients with elective PTCA was TIMI 2 to 3 (Table 3); in 3 patients (patients 1, 3, and 5 in Table 3), the initial distal flow was severely reduced (TIMI I). Normal distal run-off was observed in all patients with the ECW open distal to the lesion. None of these patients developed no-reflow. In those patients with acute myocardial infarction, after recanalization of the target vessel and PTCA of the culprit lesion, an adequate TIMI flow (TIMI II) was finally achieved (Table 3). These patients developed a maximum CK level of up to 1151 U/L (CK-MB, 158 U/L), with an early peak at 6 hours after the intervention, indicating washout phenomenon after infarct artery recanalization. The CK values of these 2 patients were excluded in the averaged CK levels (Table 2). These occluded, infarct-related arteries were crossed easily with the ECW.

Saphenous Vein Grafts
The TIMI flow before, during, and after intervention in the patient cohort with SVGs is shown in Table 4 and Figure 5. Distal blood flow in the SVGs was unchanged after the placement and opening of the porous basket, but 4 of 11 patients exhibited flow decrement after balloon angioplasty and stent placement (2 patients with flow decrease from grade 3 to 2 and 2 patients with TIMI 0). However, after collapsing and extracting the basket, normal blood flow was immediately re-established. Neither of these patients experienced an increase in CK/CK-MB. This phenomenon was only observed in large grafts with complex lesions, indicating significant obstruction of the filter pores with liberated debris. There was no evidence of distal embolization after the procedure in this high-risk patient group. Two patients developed a pseudoaneurysm at the right femoral artery, one of which required subsequent surgery. Neither of these adverse events was related to the use of the device.



View larger version (128K):
[in this window]
[in a new window]
 
Figure 5. PTCA of a SVG.

Morphometric and Histological Analysis
The devices used in the initial 20 patients (15 coronary, 5 SVG) were sent for morphometric and histological analysis. Deposits covered from 5% to 80% of the filter surface area (mean, 38±26%). Particles were recovered from all devices, and the number of particles ranged from 20 to 361, with the mean being 147±111. Mean particle size was 0.10±0.5 mm2, with a range of 0.015 to 20 mm2. The larger "particles" were aggregates of smaller particles that seemed to have been compressed together in the filter basket and capture sheath. The mean embolic burden per patient was 37±36 mm2, with a range of 0.6 to 110 mm2. There was a positive correlation between lesion severity and total embolic load (r2=0.59), but it was not statistically significant. The mean embolic load retrieved from vein grafts was 35.4±30 mm2; from coronary arteries, it was 29.7±37 mm2. Because all but 3 of the patients received stents, meaningful comparisons between the stented and nonstented groups were not possible.

Histologically, the vast majority of particles retrieved from the fixative solution or adherent to the filter or capture sheath were characterized by the presence of abundant amorphous extracellular matrix, with only a few cells. This hypocellular matrix stained positive with Alcian blue, stained strongly for von Willebrand factor, and was negative for CD41, which led to classification as chronic thrombus and atheromatous components. Neutrophils were found in almost all cases; lymphocytes and macrophages were much less common. Foam cells were often seen, but cholesterol clefts and calcified deposits were less common. Occasionally, collagen fibers and smooth muscle cells were present. An intact endothelial cell layer associated with the underlying plaque component was present in one case. Staining for glycoprotein IIb/IIIa was always negative. Acute thrombus and aggregated platelets were very rare.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
Embolization has not been considered a frequent event in coronary intervention. Only in the subgroup of patients with degenerated SVGs has embolization been thought to be clinically important. This study provides the first direct evidence that embolization of arterial plaque components occurs routinely during native coronary artery intervention in both right and left coronary arteries and in vein grafts.

A number of recent clinical and imaging trials have provided circumstantial evidence that microvascular obstruction, presumably due to embolization, is associated with adverse clinical outcomes. In acute myocardial infarction, Ito and colleagues6 showed >25% of patients with TIMI-3 flow had perfusion defects by myocardial contrast echocardiography. MRI perfusion studies have confirmed this finding and demonstrated worsening of clinical outcomes in patients with microvascular obstruction after reperfusion therapy for myocardial infarction.7 The surprising finding of a decrease in TIMI-3 flow in patients with acute myocardial infarction who were treated with stenting compared with angioplasty would be explainable on the basis of increased embolization with the more aggressive interventional technique.8

Along with the aggressiveness of the revascularization technique, diffuseness of atherosclerotic disease is a major predictor of adverse outcomes. Mehran and colleagues9 recently quantified plaque burden using intravascular ultrasound during coronary intervention and correlated it with major adverse cardiac events. The lesion plaque volume was the strongest independent predictor of adverse clinical outcomes. Increased plaque embolization due to larger plaque burdens provides the most cogent explanation for these findings.

Embolization is the major cause of acute complications during SVG intervention. During angioplasty in SVGs, Lefkovits et al4 reported a 10-fold increase in the risk of an adverse acute outcome (death, myocardial infarction, repeat CABG) associated with distal embolization. Within a decade after surgery, half of all bypass SVGs have severe atherosclerotic disease, and within 5 years, {approx}20% are reported to be occluded.1012 Repeat bypass operation is associated with a significant risk for the patient,13 leaving percutaneous angioplasty as the best alternative for many patients. As a result, new devices have been developed to limit distal embolization; these include the intracoronary administration of urokinase,14 extraction coronary atherectomy,15 directional coronary atherectomy,16 laser angioplasty,17,18 ultrasound thrombolysis,19 and AngioJet rapid thrombectomy.20,21 Unfortunately, these techniques have generally failed to reduce distal embolization adequately.

We and others recently published the first results using a balloon occlusion distal embolization protection device in SVGs that was effective in preventing embolization but did not allow angiography or perfusion during the procedure.22,23 The ECW, however, ensures normal blood flow at all times during the procedure. This allows unrestricted angiographic guidance for angioplasty and stent placement. Nevertheless, voluminous debris collected within the filter can temporarily occlude distal flow, as shown in 4 patients in this study. To differentiate between flow obstruction by dissection and flow obstruction by debris collected in the basket, it is helpful to insert the capture sheath partially and re-establish distal flow by partially collapsing the basket. In contradistinction to coronary intervention, glycoprotein IIb/IIIa antagonists do not reduce the incidence of acute events in these patients, suggesting that large embolic burdens can overwhelm effective platelet inhibition.24

The subjects in the present study had a wide range of embolic burdens. There was a trend toward correlation with lesion plaque volume and stenting, but the small sample size rendered these findings nonsignificant. SVGs had substantially larger embolic loads compared with native coronary arteries, but the histological constituents were similar between the 2 groups.

For the initial experience using this new distal protection device, we did not attempt to establish its superiority to conventional procedures. Retrospectively, comparison of the CK levels and TIMI flow of this heterogeneous patient population with those of patients undergoing percutaneous interventions using conventional techniques can be misleading. Certainly from the control arms of the glycoprotein IIb/IIIa inhibitor trials, the incidence of myocardial infarction, as defined by a rise in CK, can be in the range of 10% in native vessel interventions and up to 18% in vein graft interventions.24,25 The goal of our study is to demonstrate that embolization occurs very frequently during routine coronary intervention and may provide a theoretical framework for explaining some of the adverse results shown in a large body of clinical trials and observations. Therefore, further randomized trials are required to evaluate the clinical impact of distal protection devices during percutaneous coronary interventions.


*    Footnotes
 
Dr Yadav is the founder and a major stockholder of Cordis-AngioGuard; Dr Grube is a minor stockholder of Cordis-AngioGuard.

Received July 20, 2001; revision received August 31, 2001; accepted September 5, 2001.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 

  1. MacDonald RG, Feldman RL, Conti CR, et al. Thromboembolic complications of coronary angioplasty. Am J Cardiol. 1984; 54: 916–917.[Medline] [Order article via Infotrieve]
  2. Weyne AE, Heyndricks GR, Vandekerckhove YR, et al. Embolization complicating coronary angioplasty in the presence of an intracoronary thrombus. Clin Cardiol. 1986; 9: 463–465.[Medline] [Order article via Infotrieve]
  3. Cameron J, Buchbinder M, Wexler L, et al. Thromboembolic complications of percutaneous transluminal coronary angioplasty for myocardial infarction. Cathet Cardiovasc Diagn. 1987; 13: 100–106.[Medline] [Order article via Infotrieve]
  4. Lefkovits J, Holmes DR, Califf R, et al. Predictors and sequelae of distal embolization during saphenous vein graft intervention from the CAVEAT-II trial. Circulation. 1995; 92: 734–740.[Abstract/Free Full Text]
  5. Misumi K, Matthews RV, Sun G-W, et al. Reduced distal embolization with transluminal extraction atherectomy compared to balloon angioplasty for saphenous vein graft disease. Cathet Cardiovasc Diagn. 1996; 39: 246–251.[Medline] [Order article via Infotrieve]
  6. Ito H, Maruyama A, Iwakura K, et al. Clinical implications of the ‘no reflow’ phenomenon: a predictor of complications and left ventricular remodeling in reperfused anterior wall myocardial infarction. Circulation. 1996; 93: 223–228.[Abstract/Free Full Text]
  7. Wu K, Zerhouni EA, Judd RM, et al. Prognostic significance of microvascular obstruction by magnetic resonance imaging in patients with acute myocardial infarction. Circulation. 1998; 97: 765–772.[Abstract/Free Full Text]
  8. Stone GW, Brodie BR, Griffin JJ, et al, on behalf of the Primary Angioplasty in Myocardial Infarction (PAMI) Investigators. Improved short-term outcomes of primary coronary stenting compared to primary balloon angioplasty in acute myocardial infarction at experienced centers: the PAMI study group experience. J Interv Cardiol. 1999; 12: 101–108.
  9. Mehran R, Dangas G, Mintz GS, et al. Atherosclerotic plaque burden and CK-MB enzyme elevation after coronary interventions: intravascular ultrasound study of 2256 patients. Circulation. 2000; 101: 604–610.[Abstract/Free Full Text]
  10. Hamby RI, Aintablian A, Handler M. Aortocoronary saphenous vein bypass grafts: long-term patency, morphology and blood flow in patients with patent grafts early after surgery. Circulation. 1979; 60: 901–909.[Free Full Text]
  11. Virmani R, Atkinson JB, Forman MB. Aortocoronary saphenous vein bypass grafts. Cardiovasc Clin. 1988; 18: 41–59.[Medline] [Order article via Infotrieve]
  12. Uricchio JF, Bentivoglio LG. Patency of saphenous vein grafts five or more years after coronary bypass surgery. Am J Med. 1978; 65: 619–626.[Medline] [Order article via Infotrieve]
  13. Cameron A, Kemp HGJ, Green GE. Reoperation for coronary artery disease; 18 years of clinical follow up. Circulation. 1988; 78 (suppl 1): I-158–I-162.
  14. Hartmann J, McKeever L, Teran J. Prolonged infusion of urokinase for re-canalization of chronically occluded aortocoronary bypass grafts. Am J Cardiol. 1988; 61: 189–191.[Medline] [Order article via Infotrieve]
  15. Safian RD, Grines CL, May MA, et al. Clinical and angiographic results of transluminal extraction coronary atherectomy in saphenous vein bypass grafts. Circulation. 1994; 89: 302–312.[Abstract/Free Full Text]
  16. Holmes DRJ, Topol EJ, Califf RM, et al. A multicenter, randomized trial of coronary angioplasty versus directional atherectomy for patients with saphenous vein bypass graft lesions. Circulation. 1995; 91: 1966–1974.[Abstract/Free Full Text]
  17. Bittl JA, Sanborn TA, Yardley DE, et al. Predictors of outcome of percutaneous excimer laser coronary angioplasty of saphenous vein bypass graft lesions. Am J Cardiol. 1994; 74: 144–148.[Medline] [Order article via Infotrieve]
  18. Haude M, Welge D, Koch L, et al. Laserangioplastie und rekanalisation. Herz. 1997; 22: 299–307.[Medline] [Order article via Infotrieve]
  19. Rosenschein U, Gaul G, Erbel R, et al. Percutaneous transluminal therapy of occluded saphenous vein grafts: can the challenge be met with ultrasound thrombolysis? Circulation. 1999; 99: 26–29.[Abstract/Free Full Text]
  20. Ramee SR, Schatz RA, Carrozza JP, et al. Results of the VeGAS 1 Pilot Study of the Possis Coronary AngioJet Thrombectomy Catheter. Circulation. 1996; 94 (suppl): I-619.Abstract.
  21. Hamburger JN, Serruys PW. Treatment of thrombus containing lesions in diseased native coronary arteries and saphenous vein bypass grafts using the AngioJet Rapid Thrombectomy System. Herz. 1997; 22: 318–321.[Medline] [Order article via Infotrieve]
  22. Carlino M, De Gregorio J, Di Mario C, et al. Prevention of distal embolization during saphenous vein graft lesion angioplasty: experience with a new temporary occlusion and aspiration system. Circulation. 1999; 99: 3221–3223.[Abstract/Free Full Text]
  23. Grube E, Gerckens U, Müller R, et al. The SAFE study: multicenter evaluation of a protection catheter system for distal embolization in coronary venous bypass grafts (SVGs). Am J Cardiol. 1999; 84 (suppl 6A): 19.Abstract.
  24. Mak KH, Challapalli R, Eisenberg MJ, et al. Effect of platelet gycoprotein IIb/IIIa receptor inhibition on distal embolization during percutaneous revascularization of aortocoronary saphenous vein grafts: EPIC investigators (Evaluation of IIb/IIIa platelet receptor antagonist 7E3 in preventing ischemic complications). Am J Cardiol. 1997; 80: 985–988.[Medline] [Order article via Infotrieve]
  25. Topol EJ, Mark DB, Lincoff AM, et al. Outcomes at 1 year and economic implications of platelet glycoprotein IIb/IIIa blockade in patients undergoing coronary stenting: results from a multicentre randomised trial: EPISTENT investigators (Evaluation of Platelet IIb/IIIa Inhibitor for Stenting). Lancet. 1999; 354: 2019–2024.[Medline] [Order article via Infotrieve]



This article has been cited by other articles:


Home page
J Am Coll Cardiol IntvHome page
D. J. Kereiakes, M. A. Turco, J. Breall, N. Z. Farhat, R. L. Feldman, B. McLaurin, J. J. Popma, L. Mauri, P. Zimetbaum, J. Massaro, et al.
A Novel Filter-Based Distal Embolic Protection Device for Percutaneous Intervention of Saphenous Vein Graft Lesions: Results of the AMEthyst Randomized Controlled Trial
J. Am. Coll. Cardiol. Intv., June 1, 2008; 1(3): 248 - 257.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
C. Di Mario and G. Ferrante
Embolization: The "Dark Side" of Percutaneous Coronary Interventions
J. Am. Coll. Cardiol. Intv., June 1, 2008; 1(3): 277 - 278.
[Full Text] [PDF]


Home page
Asian Cardiovasc. Thorac. Ann.Home page
J. Kojuri, M. A Ostovan, N. Zamiri, A. Zolghadr Asli, M. A Bani Hashemi, and A. Borhani Haghighi
Procedural Outcome and Midterm Result of Carotid Stenting in High-Risk Patients
Asian Cardiovasc Thorac Ann, April 1, 2008; 16(2): 93 - 96.
[Abstract] [Full Text] [PDF]


Home page
J. Histochem. Cytochem.Home page
Q.-Y. Zhang, J.-B. Ge, J.-Z. Chen, J.-H. Zhu, L.-H. Zhang, C.-P. Lau, and H.-F. Tse
Mast Cell Contributes to Cardiomyocyte Apoptosis after Coronary Microembolization
J. Histochem. Cytochem., May 1, 2006; 54(5): 515 - 523.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
J. Herrmann
Peri-procedural myocardial injury: 2005 update
Eur. Heart J., December 1, 2005; 26(23): 2493 - 2519.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
D. A. Gorog, R. A. Foale, and I. Malik
Distal Myocardial Protection During Percutaneous Coronary Intervention: When and Where?
J. Am. Coll. Cardiol., October 18, 2005; 46(8): 1434 - 1445.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart J SupplHome page
D. Dudek, W. Mielecki, A. Dziewierz, J. Legutko, and J. S. Dubiel
The role of thrombectomy and embolic protection devices
Eur. Heart J. Suppl., October 1, 2005; 7(suppl_I): I15 - I20.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
T. Lefevre, E. Garcia, B. Reimers, I. Lang, C. di Mario, A. Colombo, F.-J. Neumann, M. V. Chavarri, P. Brunel, E. Grube, et al.
X-Sizer for Thrombectomy in Acute Myocardial Infarction Improves ST-Segment Resolution: Results of the X-Sizer in AMI for Negligible Embolization and Optimal ST Resolution (X AMINE ST) Trial
J. Am. Coll. Cardiol., July 19, 2005; 46(2): 246 - 252.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
Authors/Task Force Members, S. Silber, P. Albertsson, F. F. Aviles, P. G. Camici, A. Colombo, C. Hamm, E. Jorgensen, J. Marco, J.-E. Nordrehaug, et al.
Guidelines for Percutaneous Coronary Interventions: The Task Force for Percutaneous Coronary Interventions of the European Society of Cardiology
Eur. Heart J., April 2, 2005; 26(8): 804 - 847.
[Full Text] [PDF]


Home page
StrokeHome page
A. Kastrup, K. Groschel, J. B. Schulz, T. Nagele, and U. Ernemann
Clinical Predictors of Transient Ischemic Attack, Stroke, or Death Within 30 Days of Carotid Angioplasty and Stenting
Stroke, April 1, 2005; 36(4): 787 - 791.
[Abstract] [Full Text] [PDF]


Home page
NEJMHome page
J. S. Yadav, M. H. Wholey, R. E. Kuntz, P. Fayad, B. T. Katzen, G. J. Mishkel, T. K. Bajwa, P. Whitlow, N. E. Strickman, M. R. Jaff, et al.
Protected Carotid-Artery Stenting versus Endarterectomy in High-Risk Patients
N. Engl. J. Med., October 7, 2004; 351(15): 1493 - 1501.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
A. Angelini, P. Rubartelli, F. Mistrorigo, M. Della Barbera, F. Abbadessa, M. Vischi, G. Thiene, and S. Chierchia
Distal Protection With a Filter Device During Coronary Stenting in Patients With Stable and Unstable Angina
Circulation, August 3, 2004; 110(5): 515 - 521.
[Abstract] [Full Text] [PDF]


Home page
J CARDIOVASC PHARMACOL THERHome page
T. Reffemann and R. A. Kloner
Microvascular Alterations After Temporary Coronary Artery Occlusion: The No-Reflow Phenomenon
Journal of Cardiovascular Pharmacology and Therapeutics, July 1, 2004; 9(3): 163 - 172.
[Abstract] [PDF]


Home page
HeartHome page
J. K French and H. D White
Clinical implications of the new definition of myocardial infarction
Heart, January 1, 2004; 90(1): 99 - 106.
[Full Text] [PDF]


Home page
CirculationHome page
S. Lee, Y. Otsuji, S. Minagoe, S. Hamasaki, K. Toyonaga, M. Negishi, M. Tsurugida, H. Toda, and C. Tei
Noninvasive Evaluation of Coronary Reperfusion by Transthoracic Doppler Echocardiography in Patients With Anterior Acute Myocardial Infarction Before Coronary Intervention
Circulation, December 2, 2003; 108(22): 2763 - 2768.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
M. Napodano, G. Pasquetto, S. Sacca, C. Cernetti, V. Scarabeo, P. Pascotto, and B. Reimers
Intracoronary thrombectomy improves myocardial reperfusion in patients undergoing direct angioplasty for acute myocardial infarction
J. Am. Coll. Cardiol., October 15, 2003; 42(8): 1395 - 1402.
[Abstract] [Full Text] [PDF]


Home page
ChestHome page
H.-K. Yip, C.-J. Wu, H.-W. Chang, Y.-K. Hsieh, C.-Y. Fang, S.-M. Chen, and M.-C. Chen
Impact of Tirofiban on Angiographic Morphologic Features of High-Burden Thrombus Formation During Direct Percutaneous Coronary Intervention and Short-term Outcomes
Chest, September 1, 2003; 124(3): 962 - 968.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
G. W. Stone, C. Rogers, J. Hermiller, R. Feldman, P. Hall, R. Haber, A. Masud, P. Cambier, R. P. Caputo, M. Turco, et al.
Randomized Comparison of Distal Protection With a Filter-Based Catheter and a Balloon Occlusion and Aspiration System During Percutaneous Intervention of Diseased Saphenous Vein Aorto-Coronary Bypass Grafts
Circulation, August 5, 2003; 108(5): 548 - 553.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
U. Limbruno, A. Micheli, M. De Carlo, G. Amoroso, R. Rossini, C. Palagi, V. Di Bello, A. S. Petronio, G. Fontanini, and M. Mariani
Mechanical Prevention of Distal Embolization During Primary Angioplasty: Safety, Feasibility, and Impact on Myocardial Reperfusion
Circulation, July 15, 2003; 108(2): 171 - 176.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
W. W. O'Neill
"Watchful waiting"after thrombolysis: It's time for a re-evaluation
J. Am. Coll. Cardiol., July 2, 2003; 42(1): 17 - 19.
[Full Text] [PDF]


Home page
CirculationHome page
M. Roffi, D. Mukherjee, D. P. Chew, D. L. Bhatt, L. Cho, M. A. Robbins, K. M. Ziada, D. M. Brennan, S. G. Ellis, and E. J. Topol
Lack of Benefit From Intravenous Platelet Glycoprotein IIb/IIIa Receptor Inhibition as Adjunctive Treatment for Percutaneous Interventions of Aortocoronary Bypass Grafts: A Pooled Analysis of Five Randomized Clinical Trials
Circulation, December 10, 2002; 106(24): 3063 - 3067.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
G. W. Stone, C. Rogers, S. Ramee, C. White, R. E. Kuntz, J. J. Popma, J. George, S. Almany, and S. Bailey
Distal filter protection during saphenous vein graft stenting: Technical and clinical correlates of efficacy
J. Am. Coll. Cardiol., November 20, 2002; 40(10): 1882 - 1888.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
J.-i. Kotani, S. Nanto, G. S. Mintz, M. Kitakaze, T. Ohara, T. Morozumi, S. Nagata, and M. Hori
Plaque Gruel of Atheromatous Coronary Lesion May Contribute to the No-Reflow Phenomenon in Patients With Acute Coronary Syndrome
Circulation, September 24, 2002; 106(13): 1672 - 1677.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Physiol. Heart Circ. Physiol.Home page
T. Reffelmann and R. A. Kloner
Microvascular reperfusion injury: rapid expansion of anatomic no reflow during reperfusion in the rabbit
Am J Physiol Heart Circ Physiol, September 1, 2002; 283(3): H1099 - H1107.
[Abstract] [Full Text] [PDF]


Home page
Circ. Res.Home page
M. Thielmann, H. Dorge, C. Martin, S. Belosjorow, U. Schwanke, A. van de Sand, I. Konietzka, A. Buchert, A. Kruger, R. Schulz, et al.
Myocardial Dysfunction With Coronary Microembolization: Signal Transduction Through a Sequence of Nitric Oxide, Tumor Necrosis Factor-{alpha}, and Sphingosine
Circ. Res., April 19, 2002; 90(7): 807 - 813.
[Abstract] [Full Text] [PDF]


Home page
Journal Watch CardiologyHome page
Reducing Distal Embolization with Filter Protection
Journal Watch Cardiology, January 18, 2002; 2002(118): 4 - 4.
[Full Text]


Home page
Circ. Res.Home page
M. Thielmann, H. Dorge, C. Martin, S. Belosjorow, U. Schwanke, A. van de Sand, I. Konietzka, A. Buchert, A. Kruger, R. Schulz, et al.
Myocardial Dysfunction With Coronary Microembolization: Signal Transduction Through a Sequence of Nitric Oxide, Tumor Necrosis Factor-{alpha}, and Sphingosine
Circ. Res., April 19, 2002; 90(7): 807 - 813.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow