Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 2001;104:461-466
doi: 10.1161/hc3001.092038
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Ay, T.
Right arrow Articles by Vanoverschelde, J.-L. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Ay, T.
Right arrow Articles by Vanoverschelde, J.-L. J.
Related Collections
Right arrow Echocardiography

(Circulation. 2001;104:461.)
© 2001 American Heart Association, Inc.


Basic Science Reports

Destruction of Contrast Microbubbles by Ultrasound

Effects on Myocardial Function, Coronary Perfusion Pressure, and Microvascular Integrity

Taniyel Ay, MD; Xavier Havaux, MS; Guy Van Camp, MD; Barbara Campanelli, MD; Giovanna Gisellu; Agnès Pasquet, MD; Jean-François Denef, MD; Jacques A. Melin, MD, PhD; Jean-Louis J. Vanoverschelde, MD, PhD

From the Division of Cardiology, Université Catholique de Louvain, School of Medicine, Brussels, Belgium.

Correspondence to Jean Louis J. Vanoverschelde, Division of Cardiology, Cliniques Universitaires St. Luc, Avenue Hippocrate, 10-2881, B-1200 Brussels, Belgium. E-mail vanoverschelde{at}card.ucl.ac.be


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background— Recent experimental data indicate that ultrasound-induced destruction of ultrasound contrast microbubbles can cause immediate rupture of the microvessels in which these microbubbles are located.

Methods and Results— To examine the functional and morphological significance of these findings in the heart, isolated rabbit hearts were perfused retrogradely with buffer containing ultrasound contrast agents and were insonated at increasing levels of acoustic energy with a broadband transducer emitting at 1.8 MHz and receiving at 3.6 MHz and operated in the triggered mode (1 Hz). At the end of each experiment, the hearts were fixed in glutaraldehyde and examined with light microscopy. Neither exposure to ultrasound alone or to contrast alone affected left ventricular developed pressure. By contrast, simultaneous exposure to contrast and ultrasound resulted in a reversible, transient mechanical index (MI)-dependent decrease in left ventricular developed pressure (to 83±5% of baseline at an MI of 1.6) and a transient MI-dependent increase in coronary perfusion pressure (to 120±6% of baseline at an MI of 1.6). Myocardial lactate release also showed significant increases with increasing MIs. Macroscopically, areas of intramural hemorrhage were identified over the beam elevation in hearts exposed to both contrast and high-MI ultrasound. Light microscopy revealed the presence of capillary ruptures, erythrocyte extravasation, and endothelial cell damage. The mean percentage of capillaries ruptured at an MI of 1.6 was 3.6±1.4%.

Conclusions— Simultaneous exposure of isolated rabbit hearts to ultrasound and contrast agents results in an MI-dependent, transient depression of left ventricular contractile function, a rise in coronary perfusion pressure, an increase in lactate production, and limited capillary ruptures.


Key Words: contrast media • echocardiography


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Myocardial contrast echocardiography (MCE) is a relatively new technique that uses ultrasound contrast agents to produce myocardial opacification and assess myocardial perfusion. The ultrasound contrast agents used to produce myocardial opacification typically consist of gas microbubbles <10 µm in diameter that cross the pulmonary capillary bed and persist long enough to reach the myocardium. Because of their acoustic properties, these microbubbles considerably enhance the backscattering capabilities of blood, thereby allowing the myocardial capillary bed to be imaged with ultrasound.

It is now well established that contrast microbubbles can be destroyed by acoustic pulses used in the frequency and intensity range of diagnostic cardiac ultrasound.1,2 The process by which ultrasound destroys microbubbles has been characterized previously.3 Because microbubbles are compressible, they alternately contract and expand in the acoustic field, a phenomenon often referred to as cavitation. At low acoustic pressure, microbubbles usually grow and shrink rhythmically and symmetrically around their equilibrium size. This phenomenon is known as stable or noninertial cavitation. At higher acoustic pressure, however, the expansion and contraction of the microbubbles usually become unequal and markedly exaggerated, leading to their destruction. This second form of activity is known as inertial cavitation. There is accumulating evidence that the process of cavitation, particularly inertial cavitation, may produce capillary damage to organs that contain air, such as the lungs,4 or that are exposed to ultrasound contrast agents, such as has been demonstrated recently in rat spinotrapezius muscle.2,5 The purpose of the present study was to determine whether capillary damage also occurs in hearts exposed to ultrasound contrast agents and to evaluate the potential consequences thereof on left ventricular (LV) contractile performance.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Isolated Hearts
Hearts were removed from 1.5- to 2.0-kg New Zealand White rabbits, perfused retrogradely at 37°C with nonrecirculating erythrocyte-enriched Krebs-Henseleit buffer, and paced at 180 bpm as described previously.6 Red blood cell concentration was adjusted to obtain a hematocrit of 25%. The perfusate was equilibrated with a mixture of room air and 5% CO2 to maintain perfusate PO2 and pH within the physiological range. LV pressure was measured with a compliant latex balloon placed in the LV cavity via the left atrium and connected to a P23Db pressure gauge transducer (Gould Laboratories).

Myocardial Contrast Echocardiography
PESDA (perfluorocarbon-enhanced sonicated dextrose albumin), a second-generation contrast agent, consisting of decafluorobutane-filled albumin microbubbles with a mean diameter of 4.2±0.5 µm and a mean concentration of 109 · mL-1, was used in the present study.7 Experiments were also conducted with the following manufactured contrast agents: Sonozoid (Nycomed Amersham), Optison (Mallinckrodt), and Levovist (Shering AG). Imaging was performed in the triggered mode (1 Hz) with an HP Sonos 5500 system (Hewlett Packard) equipped with a broadband transducer emitting at 1.8 MHz and receiving at 3.6 MHz. The tip of the transducer was carefully positioned between 4 and 5 cm from the epicardial surface, a distance corresponding to the focal distance of the transducer in the z axis.

Experimental Protocol
After initial isolation and surgical preparation, hearts were perfused at a constant flow rate (1.5 to 2 mL · min-1 · g-1) and allowed to equilibrate for 30 minutes. LV end-diastolic pressure was initially set at 4 to 8 mm Hg by adjustment of the volume of the LV balloon. Once a stable level was attained, no further volume changes were made. At the end of the equilibration period and before the infusion of microbubbles was started, the hearts were exposed to triggered (1 Hz) ultrasound emissions for 5 minutes, with a mechanical index (MI) of 1.6. A solution of ultrasound contrast agents containing {approx}8x107 microbubbles · mL-1 was then infused at a constant rate of 1 mL/min. After 5 minutes of infusion, the hearts were again exposed to 5 consecutive triggered ultrasound sequences, each consisting of a 3-minute insonation period followed by a 2-minute recovery period.

The functional effects of acoustic pressure amplitude variations were evaluated in different groups of hearts. A first group of 6 hearts was exposed to ultrasound at an MI of 1.6 without any contrast agent. A second group of hearts (n=7) was exposed to ultrasound at increasing MIs from 0.6 to 1.6 in the presence of PESDA. The combined bioeffects of ultrasound and PESDA on myocardial function, coronary perfusion pressure, and microvascular integrity were further evaluated in 4 additional groups of hearts (n=6 each) in which the hearts were exposed to a single MI of 0.2, 0.6, 1.0, or 1.6. Finally, 3 groups of hearts were exposed to ultrasound at an MI of 1.6 in the presence of either Sonozoid, Optison, or Levovist.

Collection and Analysis of Samples
Coronary venous effluent was collected through the pulmonary artery, and PO2, PCO2, and pH were monitored regularly by means of a commercially available pH-blood gas analyzer. Glucose, troponin I, and lactate concentrations were determined in samples of aortic perfusate and pulmonary artery effluent at baseline and just before and at the end of the first ultrasound exposure, as well as just before and at the end of the last ultrasound exposure.

Morphological Analysis
After completion of each study, hearts were retrogradely perfused with erythrocyte-free Krebs-Henseleit buffer and fixed with a mixture of 3.5% formaldehyde-2.5% glutaraldehyde in phosphate buffer at pH 7.4. In some hearts, the fixation step was preceded by the perfusion of a PBS solution containing 100 IU/mL of the type VI peroxidase enzyme (Sigma). The hearts were then removed from the cannula and sliced into serial 1-mm-thick short-axis slices. One-cubic-millimeter samples of myocardium that was exposed (red zone) or not (white zone) to ultrasound were analyzed under the light microscope to quantify endothelial and capillary alterations. Morphometric assessment was performed on semithick sections stained with toluidine blue. Each section was examined to measure the percentage of capillaries and arterioles presenting with endothelial damage or rupture. A total of 1000 capillaries were examined for each zone of the sample. In hearts perfused with the peroxidase enzyme, the distribution of peroxidase activity was examined on semithick sections. For this purpose, sections were air-dried, washed with toluene and methanol (4 steps of 5 minutes each), and treated with PBS at pH 7.4. They were then incubated with 3-amino-9-ethylcarbazole (Dako A/S) for 10 minutes, rinsed with tap water, and mounted with Faramount (Dako A/S). Areas of peroxidase activity were quantified by planimetry.

Statistical Analysis
Values are mean±1 SD. Significance of differences was determined with an ANOVA for repeated measurements. Individual comparisons between groups were performed post hoc with the Bonferroni test. All tests were 2 sided, and a P value <0.05 was considered indicative of a statistically significant difference.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
Effects of Ultrasound and PESDA on LV Developed Pressure
Typical changes in LV developed pressure (LVDP) measured during ultrasound exposure in a heart perfused in the presence or absence of PESDA are shown in Figure 1. As shown in Figure 2, ultrasound exposure in the presence of PESDA resulted in a progressive MI-dependent decrease in LVDP. The largest decrease in LVDP was found during exposure to an MI of 1.6. At this acoustic pressure, LVDP decreased from 86±12 to 71±12 mm Hg (83±5% of baseline). Figure 2 also shows the percent decrease in LVDP in hearts exposed to 5 consecutive ultrasound emissions at the same MI. Five groups of hearts were studied: control hearts exposed to an MI of 1.6 in the absence of any contrast agent and 4 groups of hearts receiving PESDA and exposed to MIs of 0.2, 0.6, 1.0, or 1.6, respectively. No significant decrease in LVDP was noted in hearts exposed to ultrasound alone or in those receiving PESDA and exposed to an MI <=0.6. In the other 2 groups, a transient decrease in LVDP occurred during each exposure to ultrasound. The decrease in LVDP was reproducible between 2 ultrasound exposures and was greater at an MI of 1.6 than at an MI of 1.0. The maximal decrease in LVDP reached during exposure to an MI of 1.6 was 22±6% of baseline.



View larger version (30K):
[in this window]
[in a new window]
 
Figure 1. Representative intraventricular pressure tracings obtained in heart exposed to ultrasound at MI of 1.6 in absence of PESDA (left), at MI of 0.6 in presence of PESDA (middle), and again at MI of 1.6 in presence of PESDA (right). Ultrasound exposure in presence of PESDA (arrows) resulted in progressive, albeit transient, fall in LVDP.



View larger version (20K):
[in this window]
[in a new window]
 
Figure 2. Top, Amplitude of ultrasound-induced decrease in LVDP in hearts exposed to increasing MIs from 0.6 to 1.6. Bottom, Relative decrease in LVDP in hearts exposed to single MI of 0.2, 0.6, 1.0, or 1.6. ***P<0.001 vs ultrasound alone (US) or baseline (C). **P<0.01.

Effects of Ultrasound and PESDA on Coronary Perfusion Pressure
Figure 3 shows the percent changes in coronary perfusion pressure in hearts exposed to 5 consecutive ultrasound emissions at the same MI. As for LVDP, no significant changes in coronary perfusion pressure were observed in hearts exposed to ultrasound alone or in those receiving PESDA and exposed to an MI <=0.6. In the other 2 groups of hearts, a transient but significant increase in coronary perfusion pressure was noted during each ultrasound exposure. The maximal increase in coronary perfusion pressure reached during exposure to an MI of 1.6 was 20±6%. Simultaneous measurements of myocardial lactate production indicated that the increase in coronary perfusion pressure during ultrasound exposure at MIs of 1.0 and 1.6 was always accompanied by significant release of lactate in the coronary effluent (Figure 3, bottom). There were no significant changes in the other biologic parameters measured.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 3. Relative increase in coronary perfusion pressure (top) in hearts exposed to single MI of 0.2, 0.6, 1.0, or 1.6. and progressive increase in myocardial lactate production (bottom) in those exposed to single MI of 0.6, 1.0, or 1.6. **P<0.01; ***P<0.001 vs MI=0.6 and ultrasound alone (US). *P<0.05.

Effects of Ultrasound and PESDA on Microvascular Integrity
Figure 4 shows the macroscopic appearance of a heart exposed to ultrasound alone and that of a heart exposed to PESDA and ultrasound at an MI of 1.6. Figure 5 shows the light microscopic appearance of myocardial regions located within or outside the beam elevation in 2 hearts exposed to both PESDA and ultrasound. In regions located outside the beam elevation, no morphological abnormalities were noted. By contrast, in the regions covered by the beam elevation, several capillary rupture sites with extravasation of red blood cells into the interstitial space could be identified. As shown in Figure 6, the percentage of ruptured capillaries increased with the intensity of ultrasound exposure and culminated at 3.6±1.4% at an MI of 1.6. In the hearts perfused with the peroxidase enzyme, peroxidase activity could never be found in the regions located outside the beam elevation or in regions exposed to MIs <0.6 (Figure 5B). By contrast, in the regions covered by the beam elevation and exposed to MIs >=0.6, peroxidase activity was always present and covered 4.4±0.4% of the total muscle area in hearts exposed to an MI of 0.6, 14.1±3.4% in those exposed to an MI of 1.0, and 52.6±5.1% in those exposed to an MI of 1.6 (P<0.01, Figure 5D).



View larger version (89K):
[in this window]
[in a new window]
 
Figure 4. Macroscopic appearance of 2 hearts exposed to MI of 1.6 in absence (A) and presence (B) of PESDA. Definite areas of epicardial hemorrhage can be identified over width of beam elevation (arrows) in heart exposed to both ultrasound and PESDA.



View larger version (113K):
[in this window]
[in a new window]
 
Figure 5. A and B, Light micrographs showing normal microscopic appearance (A) and absence of significant peroxidase activity (B) in myocardial regions located outside beam elevation. C and D, Light micrographs showing 2 sites of capillary rupture (red arrows) and extravasation of red blood cells into interstitial space (C) and extensive distribution of peroxidase activity (D) in myocardium covered by beam elevation and exposed to MI of 1.6.



View larger version (15K):
[in this window]
[in a new window]
 
Figure 6. Percentage of ruptured capillaries in hearts exposed to single MI of 0.2, 0.6, 1.0, or 1.6. ***P<0.001 vs ultrasound alone (US) and MI=0.6.

To rule out the possibility that the observed functional and morphological changes were due to aggregation or destruction of large microbubbles within the coronary arteries, additional experiments were conducted in which PESDA was filtered with a 5-µm filter. The decrease in LVDP in hearts exposed to an MI of 1.6 was similar whether filtered or unfiltered PESDA was used.

Influence of Machine Settings
The influence of the echographic system settings on the observed bioeffects was studied in additional experiments in which the pulse emission frequency or the pulse length was varied. Insonation at an emission frequency of 2.1 MHz induced a significantly lesser decrease in LVDP (from 91±6 to 81±3 mm Hg) than insonation at an emission frequency of 1.8 MHz (from 90±7 to 70±6 mm Hg, P<0.05 versus 2.1 MHz). By contrast, no significant differences were noted between hearts exposed to an MI of 1.6 and pulse lengths of 1, 2, or 4 cycles per pulse.

Role of the Contrast Agent
To test whether the observed bioeffects were specific to the contrast agent used, ie, PESDA, additional experiments were also conducted in which hearts were exposed to Sonozoid, Optison, or Levovist. In these hearts, the effects of ultrasound exposure on LVDP and capillary damage were similar to those observed with PESDA.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
Our data indicate that the interaction between ultrasound and microbubbles in the myocardium induces significant bioeffects, which consist of a transient and reversible decrease in LV contractile performance, a transient and reversible increase in coronary perfusion pressure, an increase in lactate production, and the occurrence of microvascular damage. These effects appear to be independent of each other. Nonetheless, they appear to be related to the acoustic pressure amplitude and to require a threshold acoustic power to occur.

Ultrasound-Induced Microvascular Damage
Because of safety considerations, the possible interaction between ultrasound and tissue has received much attention. It is now well established that in organs containing air or in the presence of strong cavitation nuclei, such as contrast agents, ultrasound exposure may induce significant tissue damage, particularly to the microvasculature.8,9 There is also accumulating evidence that the mechanisms by which tissues are being damaged is the process of inertial cavitation. During this process, microbubbles experience extreme variations in their size, which may culminate in their physical destruction. Locally, inertial cavitation also produces extreme temperature and pressure changes, on the order of several hundreds of Kelvin degrees or megapascals, which by themselves may damage tissue. It may also generate free radicals and cause microstreaming and shear stress secondary to the collapse and/or rapid translation movements of the microbubbles.8,1014

Several investigators have reported on the occurrence of tissue hemorrhage and endothelial cell damage after ultrasound exposure of cultured cells and organs containing air, such as the lungs or the intestine. Dalecki et al4 reported extensive tissue hemorrhage in mice exposed to both ultrasound and Albunex. Recently, Skyba et al2 made similar observations in isolated spinotrapezius muscles from the rat and demonstrated that both microbubble destruction and the amount of capillary ruptures were proportional to the applied MI.

Our results extend these previous observations to the heart. As in the study of Skyba et al,2 we observed significant capillary damage in hearts exposed to both ultrasound and contrast agents, with no significant changes noted in hearts exposed to either ultrasound or contrast alone. In addition, our results indicate that at high MIs, the occurrence of capillary damage is independent of the contrast agent used. This is probably due to the fact that the 4 agents used in the present study are completely destroyed when exposed to MIs such as we used. However, we cannot rule out the possibility that differences could be observed at lower MIs, because initial bubble size and shell characteristics are known factors that influence the energy at which inertial cavitation and hence its bioeffects occur.

Ultrasound-Induced Contractile Dysfunction
Our data also indicate that the combined exposure of hearts to ultrasound and contrast agents may transiently alter contractile performance. Because this phenomenon was paralleled by an increase in coronary perfusion pressure and a rise in lactate production, it was likely related to the occurrence of transient myocardial ischemia. Much like capillary damage, contractile dysfunction was more pronounced when hearts were exposed to higher MIs. Nonetheless, given its transient nature, contractile dysfunction was probably unrelated to capillary damage.

The mechanisms by which ultrasound exposure induced transient ischemic dysfunction in the present study remain uncertain. Previous investigators have reported on the occurrence of reversible contractile dysfunction in dogs subjected to high-energy ultrasound delivered directly in the LV cavity by a high-power ultrasound generator.15 The phenomenon was accentuated under low loading conditions, in the presence of low plasma calcium concentration, and during mild selective {alpha}1-adrenergic stimulation.16 The transient and reversible decrease in LV function during ultrasound exposure was attributed to ultrasound-induced endocardial endothelium dysfunction.17 Other investigators have shown that similar dysfunction could be obtained by selectively damaging the endothelial monolayer of the coronary bed.18 At variance with these earlier reports, in the present study, exposure of hearts to ultrasound alone, even at an MI of 1.6, did not produce any LV contractile dysfunction. This is probably because the energy levels used in the present study were considerably lower than those used in the previous reports. Nonetheless, we did observe significant dysfunction once contrast agents were added during ultrasound exposure. One possible explanation for these findings is that the addition of contrast agents made the endothelial cells more susceptible to the ultrasound energy, causing transient endothelial dysfunction and hence vasoconstriction and ischemia. Alternatively, microbubble aggregation, which has also been shown to occur when microbubbles are exposed to ultrasound,3 may have provoked transient capillary and/or arteriolar plugging, which in turn caused ischemia and transient dysfunction.

Implications for the Use of MCE in Humans
The clinical relevance of our findings must be examined with caution. Although our data clearly demonstrate that significant microvascular damage and transient contractile dysfunction can be induced when hearts are simultaneously exposed to high-energy ultrasound and contrast agents ex vivo, the likelihood that such adverse events will ever occur in vivo, and hence in humans, is extremely small. First, the energy levels that caused capillary damage and transient contractile dysfunction in the present study are unlikely to be reached in vivo. Indeed, the threshold MI that caused these events in our experiments was 0.8. Because the ultrasound energy was administered directly to the heart, without any interposing medium and hence attenuation, it is probable that the energy that was delivered to these hearts was 30% to 40% higher than that which could be delivered in vivo, through the chest wall. Second, even if some damage could be provoked, it would most probably be quite localized and thus unlikely to induce any measurable functional effects. The human LV, which weighs between 150 and 180 g, is considerably larger than that of an isolated rabbit heart, which weighs between 5 and 7 g, 2 or 3 g of which was exposed to ultrasound and damaged (the red area on Figure 4). If such an amount (2 g) was to be damaged in a human heart, it would probably cause no functional harm. Only very subtle abnormalities, such as the occurrence of premature ventricular contractions, could potentially be seen.19 Thus, overall, the risk of adverse reactions is small if not negligible. In this regard, the large clinical experience accumulated so far in several clinical trials is particularly reassuring.2023

It has also been suggested that ultrasound-induced capillary ruptures could be used as a means to enhance local tissue permeability and thereby facilitate the local delivery of drugs or genetic material. In the rat spinotrapezius muscle, Price et al5 recently demonstrated the feasibility of delivering colloidal particles through capillary ruptures caused by ultrasound-induced microbubble destruction. These data were confirmed very recently in rat hearts, in which the galactosidase gene was successfully delivered and expressed after their exposure to both ultrasound and contrast microbubbles.24 The present study confirms the ability of high-energy ultrasound in the presence of contrast microbubbles to increase capillary permeability and to facilitate the delivery of large macromolecules, such as the immunoperoxidase enzyme. Our study does show, however, that this can only be accomplished by using MIs that are unlikely to be reached during conventional diagnostic imaging. Transthoracic transducers emitting at a lower frequency, a higher MI, or both will probably be needed to achieve this goal. Intracardiac probes may be another alternative.

In conclusion, we have shown that exposure of isolated hearts to high-energy ultrasound and ultrasonic contrast agents provokes a transient but reversible contractile dysfunction, which is probably ischemic in origin, as well as limited capillary ruptures, which may provide the unique opportunity to increase tissue permeability and thereby facilitate local drug delivery.


*    Acknowledgments
 
This work was supported in part by grant No. 3-4563-98 from the Fonds National de la Recherche Scientifique et Médicale and by the Action de Recherche Concertée No. 96/01-199. Dr Ay is supported by the Damman Foundation, Louvain-la-Neuve, Belgium.


*    Footnotes
 
Presented in part at the 72nd Scientific Sessions of the American Heart Association, Atlanta, Ga, November 7–10, 1999, and published in abstract form (Circulation. 1999;100[suppl I]:I-205).

Received September 28, 2000; revision received April 4, 2001; accepted April 5, 2001.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 

  1. Wei K, Skyba DM, Firschke C, et al. Interactions between microbubbles and ultrasound: in vitro and in vivo observations. J Am Coll Cardiol. 1997; 29: 1081–1088.[Abstract]
  2. Skyba DM, Price RJ, Linka AZ, et al. Direct in vivo visualization of intravascular destruction of microbubbles by ultrasound and its local effects on tissue. Circulation. 1998; 98: 290–293.[Abstract/Free Full Text]
  3. Dayton PA, Morgan KE, Klibanov AL, et al. Optical and acoustical observations of the effects of ultrasound on contrast agents. IEEE Trans. 1999; 46: 220–232.
  4. Dalecki D, Child SZ, Reaman CH, et al. Age-dependence of ultrasonically-induced lung hemorrhage in mice. Ultrasound Med Biol. 1996; 23: 917–925.
  5. Price RJ, Skyba DM, Kaul S, et al. Delivery of colloidal particles and red blood cells to tissue through microvessel ruptures created by targeted microbubble destruction with ultrasound. Circulation. 1998; 98: 1264–1267.[Abstract/Free Full Text]
  6. Vanoverschelde J-L, Janier MF, Bergmann SR. The relative importance of myocardial energy metabolism compared with ischemic contracture in the determination of ischemic injury in isolated perfused rabbit hearts. Circ Res. 1994; 74: 817–828.[Abstract/Free Full Text]
  7. Porter TR, Xie F. Visually discernible myocardial echocardiographic contrast after intravenous injection of sonicated dextrose albumin microbubbles containing high molecular weight, less soluble gases. J Am Coll Cardiol. 1995; 25: 509–515.[Abstract]
  8. Holland CK, Deng C, Apfel RE. Direct evidence of cavitation in vivo. Ultrasound Med Biol. 1996; 22: 917–925.[Medline] [Order article via Infotrieve]
  9. Holland CK, Apfel E. Thresholds for transient cavitation produced by pulsed ultrasound in a controlled nuclei environment. J Acoust Soc Am. 1990; 88: 2059–2069.[Medline] [Order article via Infotrieve]
  10. Deleted in proof.
  11. WFUMB Safety Symposium, Kloster-banz. Ultrasound: nonthermal issues: cavitation—its nature, detection and measurement. Ultrasound Med Biol. 1998; 24: S11–S21.
  12. Mornstein V. Cavitation-induced risks associated with contrast agents used in ultrasonography. Eur J Ultrasound. 1997; 5: 101–111.
  13. Miller DL, Thomas RM. Frequency dependence of cavitation activity in a rotating tube exposure system compared to the mechanical index. J Acoust Soc Am. 1993; 93: 3475–3480.[Medline] [Order article via Infotrieve]
  14. Carstensen EL, Kelly P, Church CC, et al. Lysis of erythrocytes by exposure to CW ultrasound. Ultrasound Med Biol. 1993; 9: 147–165.
  15. Gillebert TC, De Hert SG, Andries LJ, et al. Intracavitary ultrasound impairs left ventricular performance: presumed role of endocardial endothelium. Am J Physiol. 1992; 263: H857–H865.[Abstract]
  16. De Hert SG, Gillebert TC, Brutsaert DL. Alteration of left ventricular endocardial function by intracavitary high power ultrasound: interaction with volume, inotropic state, and alpha 1-adrenergic stimulation. Circulation. 1993; 87: 1275–1285.[Abstract/Free Full Text]
  17. Ramaciotti C, McClellan G, Sharkey A, et al. Cardiac endothelial cells modulate contractility of rat heart in response to oxygen tension and coronary flow. Circ Res. 1993; 72: 1044–1064.[Abstract/Free Full Text]
  18. Li K, Rouleau JL, Andries LJ, et al. Effect of dysfunctional vascular endothelium on myocardial performance in isolated papillary muscles. Circ Res. 1993; 72: 768–777.[Abstract/Free Full Text]
  19. Van der Wouw PA, Brauns AC, Bailey SE, et al. Premature ventricular contractions during triggered imaging with ultrasound contrast. J Am Soc Echo. 2000; 13: 288–294.[Medline] [Order article via Infotrieve]
  20. Marwick TM, Brunken R, Meland N, et al. Accuracy and feasibility of contrast echocardiography for detection of perfusion defects in routine practice: comparison with wall motion and technetium-99m sestamibi single-photon emission computed tomography: the Nycomed NC100100 Investigators. J Am Coll Cardiol. 1998; 32: 1260–1269.[Abstract/Free Full Text]
  21. Binder T, Assayag P, Baer FM, et al. NC100100, a new echo contrast agent for the assessment of myocardial perfusion: safety and comparison with MIBI- SPECT in a randomized multicenter trial. Clin Cardiol. 1999; 22: 273–282.[Medline] [Order article via Infotrieve]
  22. Keller MW, Glasheen W, Kaul S. Albunex: a safe and effective commercially produced agent for myocardial contrast echocardiography. J Am Soc Echo. 1989; 2: 48–52.[Medline] [Order article via Infotrieve]
  23. Main ML, Escobar JF, Hall SA, et al. Safety and efficacy of QW7437, a new fluorocarbon-based echocardiographic contrast agent. J Am Soc Echo. 1997; 10: 798–804.[Medline] [Order article via Infotrieve]
  24. Shohet RV, Chen S, Zhou Y-T, et al. Echocardiographic destruction of albumin microbubbles directs gene delivery to the myocardium. Circulation. 2000; 101: 2554–2556.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
J Ultrasound MedHome page
D. L. Miller, M. A. Averkiou, A. A. Brayman, E. C. Everbach, C. K. Holland, J. H. Wible Jr, and J. Wu
Bioeffects Considerations for Diagnostic Ultrasound Contrast Agents
J. Ultrasound Med., April 1, 2008; 27(4): 611 - 632.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
H. A.J. Struijker-Boudier, A. E. Rosei, P. Bruneval, P. G. Camici, F. Christ, D. Henrion, B. I. Levy, A. Pries, and J.-L. Vanoverschelde
Evaluation of the microcirculation in hypertension and cardiovascular disease
Eur. Heart J., December 1, 2007; 28(23): 2834 - 2840.
[Abstract] [Full Text] [PDF]


Home page
J Ultrasound MedHome page
L. Greenbaum, P. Burns, J. Copel, D. Cosgrove, J. B. Fowlkes, B. Goldberg, R. Mattrey, D. Merton, M. Robbin, and S. Wilson
American Institute of Ultrasound in Medicine Recommendations for Contrast-Enhanced Liver Ultrasound Imaging Clinical Trials
J. Ultrasound Med., June 1, 2007; 26(6): 705 - 716.
[Full Text] [PDF]


Home page
Eur Heart JHome page
D. Vancraeynest, J. Kefer, C. Hanet, C. Fillee, C. Beauloye, A. Pasquet, B. L. Gerber, M. Philippe, and J.-L. J. Vanoverschelde
Release of cardiac bio-markers during high mechanical index contrast-enhanced echocardiography in humans
Eur. Heart J., May 2, 2007; 28(10): 1236 - 1241.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
G. Van Camp, S. Droogmans, and B. Cosyns
Bio-effects of ultrasound contrast agents in daily clinical practice: fact or fiction?
Eur. Heart J., May 2, 2007; 28(10): 1190 - 1192.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. V. Shohet and P. A. Grayburn
Potential Bioeffects of Ultrasonic Destruction of Microbubble Contrast Agents
J. Am. Coll. Cardiol., April 4, 2006; 47(7): 1469 - 1470.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
D. L. Miller, E. M. Driscoll, C. Dou, W. F. Armstrong, and B. R. Lucchesi
Microvascular Permeabilization and Cardiomyocyte Injury Provoked by Myocardial Contrast Echocardiography in a Canine Model
J. Am. Coll. Cardiol., April 4, 2006; 47(7): 1464 - 1468.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
D. Vancraeynest, X. Havaux, A.-C. Pouleur, A. Pasquet, B. Gerber, C. Beauloye, P. Rafter, L. Bertrand, and J.-L. J. Vanoverschelde
Myocardial delivery of colloid nanoparticles using ultrasound-targeted microbubble destruction
Eur. Heart J., January 2, 2006; 27(2): 237 - 245.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
D. L. Miller, P. Li, C. Dou, D. Gordon, C. A. Edwards, and W. F. Armstrong
Influence of Contrast Agent Dose and Ultrasound Exposure on Cardiomyocyte Injury Induced by Myocardial Contrast Echocardiography in Rats
Radiology, October 1, 2005; 237(1): 137 - 143.
[Abstract] [Full Text] [PDF]


Home page
Stem CellsHome page
M. Iwasaki, Y. Adachi, T. Nishiue, K. Minamino, Y. Suzuki, Y. Zhang, K. Nakano, Y. Koike, J. Wang, H. Mukaide, et al.
Hepatocyte Growth Factor Delivered by Ultrasound-Mediated Destruction of Microbubbles Induces Proliferation of Cardiomyocytes and Amelioration of Left Ventricular Contractile Function in Doxorubicin-Induced Cardiomyopathy
Stem Cells, October 1, 2005; 23(10): 1589 - 1597.
[Abstract] [Full Text] [PDF]


Home page
J Ultrasound MedHome page
S. Chapman, J. Windle, F. Xie, A. McGrain, and T. R. Porter
Incidence of Cardiac Arrhythmias With Therapeutic Versus Diagnostic Ultrasound and Intravenous Microbubbles
J. Ultrasound Med., August 1, 2005; 24(8): 1099 - 1107.
[Abstract] [Full Text] [PDF]


Home page
Eur J EchocardiogrHome page
B. Cosyns, C. Weytjens, M. Vanderhoogstrate, C. Daniels, D. Schoors, and G. Van Camp
Tissue Doppler imaging does not show infraclinical alteration of myocardial function after contrast echocardiography
Eur J Echocardiogr, August 1, 2005; 6(4): 238 - 242.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
J. M. Tsutsui, A. Elhendy, F. Xie, E. L. O'Leary, A. C. McGrain, and T. R. Porter
Safety of dobutamine stress real-time myocardial contrast echocardiography
J. Am. Coll. Cardiol., April 19, 2005; 45(8): 1235 - 1242.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
R. Bekeredjian, P. A. Grayburn, and R. V. Shohet
Use of ultrasound contrast agents for gene or drug delivery in cardiovascular medicine
J. Am. Coll. Cardiol., February 1, 2005; 45(3): 329 - 335.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Physiol. Heart Circ. Physiol.Home page
J. Song, P. S. Cottler, A. L. Klibanov, S. Kaul, and R. J. Price
Microvascular remodeling and accelerated hyperemia blood flow restoration in arterially occluded skeletal muscle exposed to ultrasonic microbubble destruction
Am J Physiol Heart Circ Physiol, December 1, 2004; 287(6): H2754 - H2761.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
I. Kondo, K. Ohmori, A. Oshita, H. Takeuchi, S. Fuke, K. Shinomiya, T. Noma, T. Namba, and M. Kohno
Treatment of acute myocardial infarction by hepatocyte growth factor gene transfer: The first demonstration of myocardial transfer of a "functional" gene using ultrasonic microbubble destruction
J. Am. Coll. Cardiol., August 4, 2004; 44(3): 644 - 653.
[Abstract] [Full Text] [PDF]


Home page
StrokeHome page
T. Shiogai, N. Takayasu, T. Mizuno, M. Nakagawa, and H. Furuhata
Comparison of Transcranial Brain Tissue Perfusion Images Between Ultraharmonic, Second Harmonic, and Power Harmonic Imaging
Stroke, March 1, 2004; 35(3): 687 - 693.
[Abstract] [Full Text] [PDF]


Home page
Cardiovasc ResHome page
G. Basta, L. Venneri, G. Lazzerini, E. Pasanisi, M. Pianelli, N. Vesentini, S. Del Turco, C. Kusmic, and E. Picano
In vitro modulation of intracellular oxidative stress of endothelial cells by diagnostic cardiac ultrasound
Cardiovasc Res, April 1, 2003; 58(1): 156 - 161.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
R. Beeri, J. L. Guerrero, G. Supple, S. Sullivan, R. A. Levine, and R. J. Hajjar
New Efficient Catheter-Based System for Myocardial Gene Delivery
Circulation, October 1, 2002; 106(14): 1756 - 1759.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
J. Song, M. Qi, S. Kaul, and R. J. Price
Stimulation of Arteriogenesis in Skeletal Muscle by Microbubble Destruction With Ultrasound
Circulation, September 17, 2002; 106(12): 1550 - 1555.
[Abstract] [Full Text] [PDF]


Home page
J CARDIOVASC PHARMACOL THERHome page
R. J. Price and S. Kaul
Contrast Ultrasound Targeted Drug and Gene Delivery: An Update on a New Therapeutic Modality
Journal of Cardiovascular Pharmacology and Therapeutics, September 1, 2002; 7(3): 171 - 180.
[Abstract] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Ay, T.
Right arrow Articles by Vanoverschelde, J.-L. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Ay, T.
Right arrow Articles by Vanoverschelde, J.-L. J.
Related Collect