Donate Help Contact The AHA Sign In Home
American Heart Association
Circulation
Search: search_blue_button Advanced Search
Circulation. 2002;105:1282-1284

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lederman, R. J.
Right arrow Articles by McVeigh, E. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lederman, R. J.
Right arrow Articles by McVeigh, E. R.
Related Collections
Right arrow Catheter-based coronary and valvular interventions: other

(Circulation. 2002;105:1282.)
© 2002 American Heart Association, Inc.


Brief Rapid Communications

Catheter-Based Endomyocardial Injection With Real-Time Magnetic Resonance Imaging

Robert J. Lederman, MD; Michael A. Guttman, MSc; Dana C. Peters, PhD; Richard B. Thompson, PhD; Jonathan M. Sorger, BS; Alexander J. Dick, MD; Venkatesh K. Raman, MD; Elliot R. McVeigh, PhD

From the Cardiovascular Branch (R.J.L., A.J.D., V.K.R.) and Medical Imaging Group, Laboratory of Cardiac Energetics (M.A.G., D.C.P., R.B.T., J.M.S., E.R.M.), National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, Md.

Correspondence to Robert J. Lederman, MD, Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Building 10, Room 2c713, Bethesda, MD 20892-1061. E-mail Lederman{at}nih.gov


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background We tested the feasibility of targeted left ventricular (LV) mural injection using real-time MRI (rtMRI).

Methods and Results A 1.5T MRI scanner was customized with a fast reconstruction engine, transfemoral guiding catheter–receiver coil (GCC), MRI-compatible needle, and tableside consoles. Commercial real-time imaging software was customized to facilitate catheter navigation and visualization of injections at 4 completely refreshed frames per second. The aorta was traversed and the left ventricular cavity was entered under direct rtMRI guidance. Pigs underwent multiple injections with dilute gadolinium-DTPA. All myocardial segments were readily accessed. The active GCC and the passive Stiletto needle injector were readily visualized. More than 50 endomyocardial injections were performed with the aid of rtMRI; 81% were successful with this first-generation prototype.

Conclusion Percutaneous endomyocardial drug delivery is feasible with the aid of rtMRI, which permits precise 3-dimensional localization of injection within the LV wall.


Key Words: magnetic resonance imaging • drugs • ischemia • radiography • catheterization


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
X-ray fluoroscopy provides outstanding guidance for a transcatheter therapy, including coronary intervention, but is limited when imaging soft tissue compared with contrast-filled blood lumen or bone. MRI provides excellent tissue contrast, even without exogenous contrast agents, and is free of radiation exposure. Although fluoroscopic MRI is not a new concept,1 technology advances now permit rapid image acquisition and display. We tested the ability of in vivo real-time MRI (rtMRI) to guide catheter-based cardiac injection in swine. This technology may enable a new class of therapeutic cardiovascular procedures.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
MRI Technology
We used a General Electric Signa CV/i 1.5T scanner with a custom reconstruction engine.2 Raw data were transferred to a workstation (Onyx2; Silicon Graphics) for real-time reconstruction and display at tableside consoles.

The GE i-Drive real-time slice navigation interface was modified to add steady-state free precession (SSFP; also known as FISP) imaging3 and additional real-time controls: nonselective saturation preparation to enhance gadolinium injections; ECG gating to reduce cardiac motion; individual channel gain adjustment and color highlighting to enhance the catheter signal; and multislice acquisition and interactive volume rendering of multislice datasets as described below. An x-ray fluoroscope was available but not used.

Animals
The National Heart, Lung and Blood Institute Animal Care and Use Committee approved the protocols. Farm swine (n=6; weight, 40 to 50 kg; National Institutes of Health, Bethesda, Md) received telazol/xylazine, continuous isoflurane, and heparin. Femoral artery sheaths (9F) were placed. The explanted heart from 1 animal was fixed with unpressurized formalin.

Catheters and Injections
Custom transfemoral LV guiding catheters (Boston Scientific) were modified to serve as intravascular receiver coils (GCCs). Independently steerable coaxial 9F and 7F guiding catheters directed a needle in the LV cavity. These first-generation devices had no special adaptation to attenuate potential catheter heating. The GCC was used in parallel with a custom high-impedance phased array of 3 flexible surface coils (Nova Medical). The spring-loaded 27G injection needle (Stiletto; Boston Scientific) was modified for MRI compatibility but not as a receiver coil. The actuated Stiletto needle extends 3.5 mm.

Gadolinium-DTPA (Magnevist) was diluted with NaCl solution to achieve a concentration of 30 mmol/L and physiological osmolarity for 1-mL injections. In one animal, brilliant green dye was admixed with gadolinium injectate for postmortem inspection of injection sites. Injection depth was measured off-line (Medical Image Processing, Analysis, and Visualization; National Institutes of Health). Data are reported as mean±SD.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
Technical Details of MRI
Catheter manipulations were guided by SSFP sequences. These were highly sensitive to appropriate center-frequency and shim adjustment. At heart rates of 60 to 100 bpm, an acceptable balance of cardiac motion, signal-to-noise ratio, and spatial and temporal resolution was obtained with the following parameters: 128x96 pixel matrix, 7-mm slice thickness, 32x24-cm field of view, 60o flip angle, 125- or 62.5-kHz bandwidth, 1.8-ms echo time, 4.4-ms repetition time, and three-fourths k-space acquisition. These parameters yielded 4 completely refreshed frames per second.

Nonselective saturation preparation, with a 90o flip-angle and an effective inversion time of 120 ms, suppressed blood and myocardial signal to enhance gadolinium visualization. For most injections, intramyocardial gadolinium was clearly visible even without saturation preparation.

Catheter Tracking and Visualization
The transfemoral guiding catheter traversed the aorta entirely with rtMRI (SSFP sequences) guidance, navigating away from the cephalic vessels and across the aortic valve, using intrinsic blood and tissue contrast, and without superimposed roadmap images or subtraction masks. Coloring the catheter channel red often helped to identify catheter-related components, such as nitinol guidewires (which demonstrated inductive coupling), that were not readily visualized in some configurations with pure grayscale imaging.

The tip of the catheter appeared as a signal void and was visualized only when located within the selected imaging slice, so entry into the left ventricle was better tracked using volume-rendered multislice imaging (ie, combining multiple slices).

Scan planes were dynamically selected to show the aortic arch (for LV entry) or rotated along the long axis of the left ventricular catheter (for injection site selection). The catheter was easily manipulated into any selected scan plane; similarly, scan planes were easily manipulated to determine catheter position. The preformed coaxial catheters needed to be aligned in-plane for complete visualization within single thin slices.

The active GCC was well visualized against the white-blood SSFP sequences as a black central stripe (along the catheter lumen) surrounded by a region of enhanced local signal (Figure 1B). Catheter position could easily be demonstrated in relation to endocardial and epicardial surfaces and papillary and valvar structures. The passive Stiletto device also was well visualized as a black magnetic susceptibility–induced signal void during SSFP sequences, distinct from blood and myocardium (Figure 1C). Both devices also could be seen, weakly, to provide spatial context during T1-weighted imaging with saturation preparation intended to suppress background and enhance visualization of the myocardial gadolinium injectate (Figure 1D and 1E).



View larger version (135K):
[in this window]
[in a new window]
 
Figure 1. A representative sequence of 2 septal endomyocardial injections.

The Stiletto device could be tracked as it exited and straightened the GCC until it made contact with the targeted endocardial injection site (Figure 1C). The first-generation units tested in this study sometimes moved out of plane, although visualization of endocardial contact usually could be restored by adjusting the scan plane position. Endocardial contact usually could be maintained without significant ventricular ectopy.

Endomyocardial Injections
A total of 53 injections were attempted, 43 of which were visualized as myocardial accumulation of gadolinium (81%). A representative endomyocardial injection sequence is shown in Figure 1. Figure 1A includes an overlay to identify the 4 cardiac chambers and pulmonary artery, as well as the intracardiac GCC. Figure 1B shows the GCC directed toward the interventricular septum without making contact. In Figure 1C, the Stiletto is extended from the guiding catheter and engages the apical left ventricular septum. Figure 1D shows the same view with saturation preparation for background suppression. The first gadolinium injection appears bright, as indicated by the arrow. In Figure 1E, a second injection is made 1 minute afterward, more proximally along the septum, with the same imaging plane as before. Gadolinium from the first, more distal, injection site remains visible. After saturation pulses are stopped, in Figure 1F, both injections remain visible within the myocardial wall. Injections typically remained visible for >10 minutes.



View larger version (22K):
[in this window]
[in a new window]
 
Figure 2. Accumulation of injectate. A representative apical septum injection is shown (arrows) before (A) and after (B) saturation pulses are applied to suppress the background and enhance visualization of the gadolinium signal. For this injection, gadolinium injectate was admixed with brilliant green dye to show good concordance with dye accumulation in a formalin-fixed postmortem specimen.

The morphology of each injection was variable. Injections had an average depth of 4.0±1.4 mm, compared with the Stiletto needle length of 3.5 mm and an average wall thickness of 10.0±1.4 mm. In the long axis, injections had an average maximum diameter of 8.3±2.5 mm and an average minimum diameter of 5.7±1.5 mm. Figure 2 illustrates the close relationship between rtMRI and pathology when gadolinium was admixed with tissue-fast dye.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
We have demonstrated feasibility of rtMRI guidance of catheter-based endomyocardial injection in pigs. This technique afforded good spatial localization and detail. All myocardial segments were readily accessible, and intramuscular accumulation of injectate was imaged in vivo and in real-time.

We intend to apply rtMRI-guided catheter injection to targeted local myocardial drug delivery. Potential applications include protein or gene formulations for therapeutic angiogenesis in the treatment of myocardial ischemia, myoblast or stem cell preparations to repair failing or infarcted myocardium, and ablation of conduction tissue in rhythm disorders or of outflow tissue in hypertrophic cardiomyopathy. Treatment could be tailored, for example, according to regional MRI perfusion or functional characteristics. Successful targeting and injection can be observed directly by including dilute gadolinium or other metal chelates, as in our present experiments.

Tracking catheters is challenging in rtMRI. Passive catheters can be difficult to visualize because the signal void created by the catheter may be nonuniform, obscured by volume-averaging with surrounding bright tissues (especially in thick slabs), or invisible against surrounding dark regions. Signal-enhancing passive catheters4 may overcome some of these obstacles but do not allow independent manipulation of the catheter signal for enhanced visualization during navigation, nor do they help obtain high-resolution images of the surrounding tissue. We addressed these challenges in several ways. First, we converted the guiding catheter to an active receiver coil, which provides a bright signal.5 Second, we used thin-slab images to reduce volume-averaging. Third, we intermittently employed multislice imaging, in which multiple parallel slabs were rapidly acquired in sequence. Although multislice imaging reduced overall temporal resolution (by the number of slices—typically 3), thinner slices reduce volume-averaging for better detection of catheter signal and anatomic detail. Catheters could be tracked across multiple slices, which could be combined into a single image.

In this experience, thin slabs reduced volume-averaging sufficiently so that even the passive Stiletto needle was clearly visualized using bright-blood pulse sequences. Both single- and multislice rtMRI afforded easy tracking of active catheters. We found it convenient to rotate the imaging plane along the LV long-axis for either catheter tracking or myocardial targeting. When projection-like images were desired, as during aortic valve traversal, multislice imaging proved useful. Other teams have described rapid thick-slab (projection) myocardial MRI using spoiled gradient echo techniques4,68 or they have superimposed contextual road maps9,10 to guide catheters.

Limitations
This is a proof-of-concept experience with first-generation catheters and exploratory imaging techniques. Catheters could be improved, for example, by placing receiver coils in the injection needle, by increasing coil sensitivity, and designing guiding catheters that remain in a single imaging plane. In addition, precautions must be taken to attenuate potentially dangerous catheter heating during rtMRI. Several approaches are being tested currently.1113 Commercial MRI scanners could be improved with faster image reconstruction capability and the ability to switch rapidly between different imaging sequences. For example, it would be helpful to alternate between rapid sequences for catheter manipulation and high-resolution sequences for careful tissue inspection. Nevertheless, in our laboratory, contemporary gradient systems and a custom reconstruction engine permitted multiple completely refreshed frames to be acquired and displayed each second with satisfactory latency, signal-to-noise ratio, and spatial resolution to guide catheter manipulation, orientation, and myocardial injection from within the LV cavity.

Conclusion
Percutaneous endomyocardial drug delivery is feasible using rtMRI and permits precise 3-dimensional localization of injections within the left ventricular wall. We believe rtMRI therapeutic procedures involving direct visualization of the full thickness of the myocardium and valves will soon be possible in humans. Combined with on-line observation of functional, perfusion, and spectroscopic characteristics, rtMRI may offer a new cardiovascular therapeutic modality. In our preliminary experience, images are readily demonstrated that are not attainable even by surgical exposure, and these images are obtained without exogenous contrast agents and without regard to plane orientation or interposed air or bone.


*    Acknowledgments
 
The authors express their gratitude to Robert Balaban, Elizabeth Nabel, Scott Smith, Greg DelMain, Kenneth Larson, Joni Taylor, Diana Lancaster, Peter Kellman, Tony Faranesh, Anthony Aletras, and Annette Stine.

Received December 11, 2001; revision received January 17, 2002; accepted January 23, 2002.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 
1. Riederer SJ, Tasciyan T, Farzaneh F, et al. MR fluoroscopy: technical feasibility. Magn Reson Med. 1988; 8: 1–15.[Medline] [Order article via Infotrieve]

2. Guttman MA, McVeigh ER. Techniques for fast stereoscopic MRI. Magn Reson Med. 2001; 46: 317–23.[CrossRef][Medline] [Order article via Infotrieve]

3. Oppelt A. FISP—a new fast MRI sequence. Electromedica. 1986; 54: 15–18.

4. Omary RA, Unal O, Koscielski DS, et al. Real-time MR imaging-guided passive catheter tracking with use of gadolinium-filled catheters. J Vasc Interv Radiol. 2000; 11: 1079–1085.[Medline] [Order article via Infotrieve]

5. Atalar E, Bottomley PA, Ocali O, et al. High resolution intravascular MRI and MRS by using a catheter receiver coil. Magn Reson Med. 1996; 36: 596–605.[Medline] [Order article via Infotrieve]

6. Serfaty JM, Atalar E, Declerck J, et al. Real-time projection MR angiography: feasibility study. Radiology. 2000; 217: 290–295.[Abstract/Free Full Text]

7. Bartels LW, Bos C, van Der Weide R, et al. Placement of an inferior vena cava filter in a pig guided by high-resolution MR fluoroscopy at 1.5 T. J Magn Reson Imaging. 2000; 12: 599–605.[CrossRef][Medline] [Order article via Infotrieve]

8. Omary RA, Frayne R, Unal O, et al. MR-guided angioplasty of renal artery stenosis in a pig model: a feasibility study. J Vasc Interv Radiol. 2000; 11: 373–381.[Medline] [Order article via Infotrieve]

9. Ladd ME, Zimmermann GG, Quick HH, et al. Active MR visualization of a vascular guidewire in vivo. J Magn Reson Imaging. 1998; 8: 220–225.[Medline] [Order article via Infotrieve]

10. Atalar E, Kraitchman DL, Carkhuff B, et al. Catheter-tracking FOV MR fluoroscopy. Magn Reson Med. 1998; 40: 865–872.[Medline] [Order article via Infotrieve]

11. Liu CY, Farahani K, Lu DS, et al. Safety of MRI-guided endovascular guidewire applications. J Magn Reson Imaging. 2000; 12: 75–78.[CrossRef][Medline] [Order article via Infotrieve]

12. Ladd ME, Quick HH. Reduction of resonant RF heating in intravascular catheters using coaxial chokes. Magn Reson Med. 2000; 43: 615–619.[CrossRef][Medline] [Order article via Infotrieve]

13. Yeung CJ, Atalar E. RF transmit power limit for the barewire loopless catheter antenna. J Magn Reson Imaging. 2000; 12: 86–91.[CrossRef][Medline] [Order article via Infotrieve]




This article has been cited by other articles:


Home page
J Am Coll Cardiol ImgHome page
C. E. Saikus and R. J. Lederman
Interventional Cardiovascular Magnetic Resonance Imaging: A New Opportunity for Image-Guided Interventions
J. Am. Coll. Cardiol. Img., November 1, 2009; 2(11): 1321 - 1331.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
J. W. M. Bulte
In Vivo MRI Cell Tracking: Clinical Studies
Am. J. Roentgenol., August 1, 2009; 193(2): 314 - 325.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
S. Nazarian, A. Kolandaivelu, M. M. Zviman, G. R. Meininger, R. Kato, R. C. Susil, A. Roguin, T. L. Dickfeld, H. Ashikaga, H. Calkins, et al.
Feasibility of Real-Time Magnetic Resonance Imaging for Catheter Guidance in Electrophysiology Studies
Circulation, July 15, 2008; 118(3): 223 - 229.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
T. Geva and A. C. Marshall
Magnetic Resonance Imaging-Guided Catheter Interventions in Congenital Heart Disease
Circulation, February 28, 2006; 113(8): 1051 - 1052.
[Full Text] [PDF]


Home page
CirculationHome page
D. S. Baim and R. Y. Kwong
Is Magnetic Resonance Image Guidance the Key to Opening Chronic Total Occlusions?
Circulation, February 28, 2006; 113(8): 1053 - 1055.
[Full Text] [PDF]


Home page
CirculationHome page
R. J. Lederman
Cardiovascular Interventional Magnetic Resonance Imaging
Circulation, November 8, 2005; 112(19): 3009 - 3017.
[Full Text] [PDF]


Home page
J Am Coll CardiolHome page
V. K. Raman, P. V. Karmarkar, M. A. Guttman, A. J. Dick, D. C. Peters, C. Ozturk, B. S.S. Pessanha, R. B. Thompson, A. N. Raval, R. DeSilva, et al.
Real-Time Magnetic Resonance-Guided Endovascular Repair of Experimental Abdominal Aortic Aneurysm in Swine
J. Am. Coll. Cardiol., June 21, 2005; 45(12): 2069 - 2077.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
G. A. Krombach, J. G. Pfeffer, S. Kinzel, M. Katoh, R. W. Gunther, and A. Buecker
MR-guided Percutaneous Intramyocardial Injection with an MR-compatible Catheter: Feasibility and Changes in T1 Values after Injection of Extracellular Contrast Medium in Pigs
Radiology, May 1, 2005; 235(2): 487 - 494.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
R Corti, J Badimon, G Mizsei, F Macaluso, M Lee, P Licato, J F Viles-Gonzalez, V Fuster, and W Sherman
Real time magnetic resonance guided endomyocardial local delivery
Heart, March 1, 2005; 91(3): 348 - 353.
[Abstract] [Full Text] [PDF]


Home page
Cardiovasc ResHome page
J. E. Markkanen, T. T. Rissanen, A. Kivela, and S. Yla-Herttuala
Growth factor-induced therapeutic angiogenesis and arteriogenesis in the heart-gene therapy
Cardiovasc Res, February 15, 2005; 65(3): 656 - 664.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
M. Saeed, R. Lee, A. Martin, O. Weber, G. A. Krombach, S. Schalla, M. Lee, D. Saloner, and C. B. Higgins
Transendocardial Delivery of Extracellular Myocardial Markers by Using Combination X-ray/MR Fluoroscopic Guidance: Feasibility Study in Dogs
Radiology, June 1, 2004; 231(3): 689 - 696.
[Abstract] [Full Text] [PDF]


Home page
ANN INTERN MEDHome page
M. S. Lee and R. R. Makkar
Stem-Cell Transplantation in Myocardial Infarction: A Status Report
Ann Intern Med, May 4, 2004; 140(9): 729 - 737.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
I M Barbash, J Leor, M S Feinberg, A Tessone, S Aboulafia-Etzion, A Orenstein, J Ruiz-Cabello, J S Cohen, and Y Mardor
Interventional magnetic resonance imaging for guiding gene and cell transfer in the heart
Heart, January 1, 2004; 90(1): 87 - 91.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
A. J. Dick, M. A. Guttman, V. K. Raman, D. C. Peters, B. S.S. Pessanha, J. M. Hill, S. Smith, G. Scott, E. R. McVeigh, and R. J. Lederman
Magnetic Resonance Fluoroscopy Allows Targeted Delivery of Mesenchymal Stem Cells to Infarct Borders in Swine
Circulation, December 9, 2003; 108(23): 2899 - 2904.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
R. C. Susil, A. Krieger, J. A. Derbyshire, A. Tanacs, L. L. Whitcomb, G. Fichtinger, and E. Atalar
System for MR Image-guided Prostate Interventions: Canine Study
Radiology, September 1, 2003; 228(3): 886 - 894.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
J. M. Hill, A. J. Dick, V. K. Raman, R. B. Thompson, Z.-X. Yu, K. A. Hinds, B. S.S. Pessanha, M. A. Guttman, T. R. Varney, B. J. Martin, et al.
Serial Cardiac Magnetic Resonance Imaging of Injected Mesenchymal Stem Cells
Circulation, August 26, 2003; 108(8): 1009 - 1014.
[Abstract] [Full Text] [PDF]


Home page
Circ. Res.Home page
D. Orlic, J. M. Hill, and A. E. Arai
Stem Cells for Myocardial Regeneration
Circ. Res., December 13, 2002; 91(12): 1092 - 1102.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Data Supplement
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lederman, R. J.
Right arrow Articles by McVeigh, E. R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lederman, R. J.
Right arrow Articles by McVeigh, E. R.
Related Collections
Right arrow Catheter-based coronary and valvular interventions: other