(Circulation. 2002;106:2930.)
© 2002 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Division of Cardiovascular Medicine (M.D.G., L.J.C., F.C, C.J.K., J.B.C., A.B.C.) and the Department of Anesthesiology (T.E.M., D.M.D.), University of Florida College of Medicine, Gainesville.
Correspondence to Mario D. Gonzalez, MD, Division of Cardiovascular Medicine, University of Florida College of Medicine, PO Box 100277, Gainesville, FL 32610-0277. E-mail gonzamd{at}medicine.ufl.edu
| Abstract |
|---|
|
|
|---|
Methods and Results We studied AV nodal conduction during right and left atrial (coronary sinus) stimulation in 46 patients (27 women and 19 men; mean age, 46±4 years) who had structurally normal hearts. At an identical cycle length (600 ms), left atrial stimulation resulted in shorter A-H intervals than right atrial stimulation (73±3 ms versus 99±3 ms; P<0.05). In addition, atrial electrograms recorded close to the His bundle changed from near to far field potentials when stimulation was shifted to the left atrium. The A-H interval prolonged as the site of pacing was progressively moved from the distal to the proximal coronary sinus. During constant pacing from the distal coronary sinus, atrial activation close to the His bundle could be advanced by late extrastimuli delivered at the anterior and posterior approaches (up to 11±2 ms and 9±1 ms, respectively), without altering His bundle activation time. In contrast, late extrastimuli delivered at the inferoparaseptal mitral annulus advanced both the A and H electrograms in 19 of 20 patients, which is consistent with a left-sided input to the AV node. Right and left atrial stimulation resulted in similar AV nodal function.
Conclusion The mitral annulus provides a left atrial input to the human AV node.
Key Words: atrioventricular node atrium conduction electrophysiology mapping
| Introduction |
|---|
|
|
|---|
|
| Methods |
|---|
|
|
|---|
Electrophysiological Study
Patients were studied in a drug-free state under sedation with intravenous fentanyl and midazolam. Multipolar electrode catheters were introduced percutaneously and positioned in the right atrial appendage, right ventricle, His bundle region, and coronary sinus. To record His bundle activation, a 7-French deflectable catheter with 4 closely spaced (2-mm) pairs of electrodes (Cordis Webster) was used. A deflectable catheter with 8 pairs of electrodes (2-mm interelectrode distance and 5 mm between pairs of electrodes; Cordis Webster) was advanced into the coronary sinus, with the most proximal pair of electrodes positioned at the coronary sinus ostium. Localization of the ostium was facilitated by observing, in the left anterior oblique projection (Figure 1), the entry point into the coronary sinus of a mapping catheter withdrawn with clockwise rotation from the isthmus. In patients who required access to the left atrium, a mapping catheter was introduced using trans-septal left heart catheterization.11
Right Atrial and Coronary Sinus Stimulation
The A-H intervals and the characteristics of the atrial electrograms recorded close to the His bundle were studied during pacing at a constant cycle length of 600 ms from the right atrial appendage and from the left atrium (distal coronary sinus). Thereafter, the site of stimulation in the coronary sinus was moved proximally in 5-mm steps until it was performed from each of the 8 pairs of electrodes. To study AV nodal function, decremental stimulation (10 ms steps) was performed from the distal coronary sinus and from the right atrial appendage until Wenckebach block proximal to the His bundle was induced. Similarly, the relative, functional, and effective refractory periods of the AV node were determined during both right and left atrial stimulation.12
Late Extrastimuli Delivered During Coronary Sinus Stimulation
To determine whether a site was part of the activation path reaching the AV node, we adapted the entrainment technique.13,14 During constant stimulation from the distal coronary sinus (S1-S1=600 ms), late atrial extrastimuli (S2) were delivered every eighth paced beat at the anterior approach (n=46), posterior approach (n=46), and inferoparaseptal mitral annulus (n=20). Stimulation of the anterior approach was obtained by positioning a pacing catheter just posterior and to the left of the catheter recording the His bundle at the level of the tendon of Todaro.15 Stimulation of the posterior approach was performed from the coronary sinus ostium. The mitral annulus was stimulated using the trans-septal approach with a catheter curved so that the tip was located at the inferoparaseptal mitral annulus. Initially, extrastimuli were introduced immediately after the local electrogram. The coupling interval was then shortened in 2-ms steps to progressively advance the local atrial electrogram, the atrial electrogram recorded close to the His bundle, and eventually His bundle activation. Stimuli were of 1-ms duration and delivered at twice the diastolic threshold.
Definitions
The A-H interval was measured from the initial deflection of the atrial electrogram (far or near field) to the initial deflection of the His bundle electrogram. A site was considered part of the wavefront reaching the AV node during coronary sinus pacing when minimal pre-excitation (2 ms) of the atrial electrogram recorded close to the AV node was followed by identical pre-excitation of the His bundle. In addition, progressively earlier extrastimuli also resulted in similar pre-excitation of atrial and His bundle activation (4 ms, 6 ms, 8 ms, etc).
Data Analysis
Data are reported as mean±SEM. Mean values obtained at each of the 8 stimulation sites along the coronary sinus and at the right atrial appendage were analyzed using repeated measures ANOVA followed by Tukey pairwise testing. The 3 entrainment sites were compared using Kruskal-Wallis ANOVA followed by pairwise multiple comparison (Dunns method). P<0.05 was accepted as statistically significant.
| Results |
|---|
|
|
|---|
|
|
The A-H Interval and Coronary Sinus Pacing Site
A progressive prolongation of the A-H interval was observed as the pacing site in the coronary sinus was shifted proximally (ANOVA, Tukey test, P<0.05). Figure 3 depicts the mean A-H interval during constant pacing at the right atrial appendage and at each site along the coronary sinus. Distal coronary sinus stimulation resulted in an A-H interval that was 11±1 ms shorter than that with proximal coronary sinus stimulation and 25±2 ms shorter than that with right atrial stimulation (P<0.05).
Entrainment During Distal Coronary Sinus Stimulation
In Figure 4, a late extrastimulus delivered at the anterior approach (Figure 1) resulted in pre-excitation and change in the characteristics of the atrial potentials recorded close to the His bundle. However, His bundle activation time remained unchanged. Similar findings were observed using extrastimuli delivered at the posterior approach. In contrast, extrastimuli delivered at the inferoparaseptal aspect of the mitral annulus (Figure 5) simultaneously advanced both atrial and His bundle electrograms without changes in the characteristics of the atrial potentials recorded close to the His bundle. Figure 6 depicts the maximal atrial pre-excitation obtained from each site without His bundle pre-excitation in the study group. Atrial electrograms could be dissociated from nearby His bundle potential in all 46 patients with extrastimuli delivered at the anterior approach (11±2 ms; median, 10 ms; range, 4 to 24 ms), in 45 of 46 patients with extrastimuli delivered at the posterior approach (9±1 ms; median, 9 ms; range, 0 to 18 ms), and in 1 of 20 patients with extrastimuli delivered at the mitral annulus (0.3±0.2 ms; median, 0 ms; range, 0 to 4 ms), respectively (P<0.001).
|
|
|
AV Nodal Function During Right Versus Left Atrial Stimulation
Right and left atrial stimulation resulted in similar AV nodal function (Wenckebach cycle length, 370±9 ms versus 369±9 ms; effective refractory period, 300±11 ms versus 307±9 ms; functional refractory period, 385±13 ms versus 386±12 ms; relative refractory period, 562±2 ms versus 564±3 ms).
| Discussion |
|---|
|
|
|---|
Previous Studies Suggesting a Left Atrial Input to the AV Node
Several published observations are consistent with the present findings. First, anatomical studies have demonstrated that the human AV node, although frequently depicted as a right-sided structure, is in fact located in the AV septum, with extensions reaching the tricuspid and mitral annuli.10 Second, in the canine heart,16 interruption of the posterior approach to the AV node can shift retrograde atrial activation to the left atrium. Third, in the swine heart, persistence of AV conduction from the left atrium has been shown in some animals after ablation of all interatrial connections.17 However, despite these findings, a left atrial input to the AV node has not been previously demonstrated in humans.
Shortening of the A-H Interval During Coronary Sinus (Left Atrial) Stimulation
Consistent with a previous study,18 we found that the A-H interval is similar when recorded simultaneously from the left and right side of the heart during coronary sinus stimulation. Therefore, the shorter A-H interval cannot be explained by poor recording from the right side of far-field potentials originating from the left side. A more likely explanation is that the left atrial wavefront propagates perpendicular and not parallel to the AV node/His bundle axis, reaching the right side of the AV septum when depolarization of the AV node is already in progress.
Two previous reports documented (although not consistently) a shorter A-H interval during coronary sinus stimulation78 that was associated in one study with a shorter Wenckebach cycle length.7 This finding was interpreted as enhanced AV nodal conduction or partial bypass of the AV node. In contrast, the present study demonstrates that the shorter A-H interval during coronary sinus stimulation is the result of activation via a different atrial input to the AV node. However, conduction via the posterior approach can also result in a shorter A-H interval.19,20 In the present study, the gradual prolongation of the A-H interval as the pacing site was moved distally to proximally along the coronary sinus and the results obtained with extrastimuli indicate that the coronary sinus and the posterior approach were not part of the activation pathway reaching the AV node in most patients. Furthermore, the progressive A-H interval prolongation when the pacing site was moved proximally along the coronary sinus probably represents a gradual shift between 2 atrial inputs (mitral annulus to posterior approach).
Use of Late Extrastimuli to Identify a Left Input to the AV Node
To our knowledge, no previous work has applied the use of late extrastimuli to identify anatomical regions that may or may not serve as inputs to the AV node during atrial stimulation. This technique is frequently used to identify sites that participate in reentrant tachycardias.13,14 In the present study, the same concept was applied to determine the wavefront path reaching the AV node during coronary sinus stimulation. Using this technique, the posterior and anterior approaches, but not the mitral annulus, could be dissociated from His bundle activation in most patients, consistent with a left-sided input to the AV node.
Limitations
The entrainment technique is typically used during reentrant tachycardias to determine if a site is part of the reentrant circuit.13 In the present study, the procedure was modified to identify the path reaching the AV node during constant stimulation and not during tachycardia. Because of potential shortcomings of the present methodology, we cannot entirely rule out the possibility that a nearby structure, and not the mitral annulus itself, provides the left atrial input to the AV node. Although very late extrastimuli were used to test potential inputs to the AV node, cycle length-dependent conduction delay in the AV node could have occurred. Under these conditions, a prolongation in the AV nodal conduction time might have prevented pre-excitation of His bundle activation, resulting in a longer than expected H-H interval for a given A-A interval. The fact that minimal pre-excitation of the mitral annulus (but not the anterior or posterior approaches) resulted in simultaneous pre-excitation of the A and H potentials in 19 of 20 patients argues against significant cycle length-dependent conduction delay.
Finally, the conduction characteristics of the left-sided input at different cycle lengths were not studied to minimize decremental conduction. Therefore, the present study did not determine if this input has fast or slow AV nodal pathway properties.
Clinical Implications
This study identifies an input to the AV node proceeding from the mitral annulus that could represent the functional counterpart of the leftward posterior extension of the AV node described by Inoue and Becker.10 A left atrial input could not only provide a greater margin of safety for AV nodal conduction, but also have a role in reentrant tachycardias. For example, ablation at the mitral annulus can successfully eliminate certain forms of AV nodal reentrant tachycardia.15
Further studies are required to elucidate the role of the left atrial input in normal and abnormal AV nodal conduction and the participation of this input in reentrant tachycardias involving the human AV node.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received July 22, 2002; revision received September 17, 2002; accepted September 19, 2002.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. Bollmann, D. Husser, L. Mainardi, F. Lombardi, P. Langley, A. Murray, J. J. Rieta, J. Millet, S. B. Olsson, M. Stridh, et al. Analysis of surface electrocardiograms in atrial fibrillation: techniques, research, and clinical applications. Europace, November 1, 2006; 8(11): 911 - 926. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Kobza, G. Hindricks, H. Tanner, and H. Kottkamp Left-septal ablation of the fast pathway in AV nodal reentrant tachycardia refractory to right septal ablation Europace, January 1, 2005; 7(2): 149 - 153. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. D. Gonzalez, L. J. Contreras, M. R.M. Jongbloed, J. Rivera, T. P. Donahue, A. B. Curtis, M. S. Bailey, J. B. Conti, G. I. Fishman, M. J. Schalij, et al. Left Atrial Tachycardia Originating From the Mitral Annulus-Aorta Junction Circulation, November 16, 2004; 110(20): 3187 - 3192. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2002 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |