(Circulation. 2004;109:1854-1858.)
© 2004 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Third Department of Internal Medicine, Showa University, Tokyo, Japan.
Correspondence to Kaoru Tanno, MD, Third Department of Internal Medicine, Showa University, 1-5-8 Hatanodai, Shinagawa-ku, Tokyo 142-8666, Japan. E-mail k-tanno{at}p02.itscom.ne
Received October 5, 2003; revision received January 9, 2004; accepted January 14, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results Subjects were 248 consecutive patients (158 men, 90 women; mean age, 59±17 years) who underwent electrophysiological study from 1997 to 2000. TWA recording was made in sinus rhythm and at atrial pacing rates of 90, 100, 110, and 120 bpm with the Cambridge Heart CH2000 system. Alternans voltage (Valt) was measured when the alternans ratio was >3 for a period of >1 minute in VM, X, Y, Z, or 2 adjacent precordial leads. Study end point was the first appearance of VT, ventricular fibrillation (VF), appropriate implantable cardioverter-defibrillator therapy with pacing or shocks, or SCD. During the 37±12-month follow-up period, 22 patients had sustained VT, and 5 patients died of SCD. In patients with >1.9-µV Valt at rates of 90, 100, and 110 bpm, the incidence of VT/VF/SCD was 56%, 28%, and 18%, respectively. Valt of >2.9 µV at a heart rate of 90 bpm had a 70% positive predictive value for VT/VF/SCD. However, when Valt was <0.9 µV at a rate of 120 bpm, negative predictive value was 100%.
Conclusions Patients with TWA at relatively low heart rates are susceptible to ventricular tachyarrhythmias.
Key Words: death, sudden heart rate T-wave alternans tachycardia, ventricular fibrillation, ventricular
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Measurement of TWA
TWA testing was conducted for patients in the postabsorbed state in a sealed room on the day after EPS. A 5F hexapolar catheter was introduced through the left subclavian vein and positioned at the high right atrium for continuous pacing of the heart. Careful skin preparation was performed to reduce skin-electrode impedance. Special high-resolution electrodes (High-Res, Cambridge Heart, Inc) were used to minimize noise. ECG leads were placed at the standard precordial lead positions and in an orthogonal X, Y, Z configuration. TWA recordings were made with the CH2000 system (Cambridge Heart, Inc) with the patient in sinus rhythm and under high right atrial pacing at 90, 100, 110, and 120 bpm if conduction proceeded without AV block for >3 minutes at each pacing rate. A spectral method of TWA analysis was used that was designed to allow detection of alternans in the microvolt range of amplitude. Valt was measured when the alternans ratio was >3 for a period of >1 minute in VM, X, Y, Z, or 2 adjacent precordial leads without artifact. TWA was considered indeterminate if the tracing was obscured by noise, ectopic beats, or AV block during atrial pacing. Data were analyzed by 2 experienced physicians who were blinded with regard to clinical data and EPS results.
Follow-Up
Patients were followed up prospectively at our university hospital outpatient clinic every month. The primary end point of the study was the combined incidence of sustained VT, VF, appropriate implantable cardioverter-defibrillator (ICD) therapy with pacing or shocks, or SCD. Fifty patients received ICDs in this study, and stored ICD electrograms were used in the evaluation of therapy.
Statistical Analysis
Kaplan-Meier survival curves were used to estimate the cumulative percentage of patients surviving free from end-point events over time. Comparisons between survival curves were made by use of the log-rank statistic. A value of P<0.05 was considered statistically significant.
| Results |
|---|
|
|
|---|
Valt and Heart Rate
The results of TWA testing are shown in Figure 1. At each voltage criterion, the percentage of positive TWA increased as the pacing heart rate increased. However, at each voltage criterion, the percentage of negative TWA decreased as the pacing heart rate increased. In addition, as the voltage criterion increased, the percentage of positive TWA was decreased at every heart rate.
|
Follow-Up
During a mean follow-up period of 45±12 months, sustained VT or VF occurred in 22 patients, and 5 patients were diagnosed as having SCD. VT or VF was confirmed in stored ECGs of the ICD in 14 of 22 patients. Characteristics of the study population are shown in Table 1.
|
VT/VF/SCD and Valt
The incidence of VT, VF, or SCD in patients with positive TWA testing is shown in Figure 2. At a heart rate of 90 bpm, 25 patients had >0.9-µV Valt. Of these 25 patients, 10 (40%) had VT/VF/SCD during the follow-up period. Similarly, 18 patients had >1.9 µV of Valt at a rate of 90 bpm. Of these 18 patients, 10 (56%) had VT/VF/SCD. At >2.9 µV, 7 of 10 patients (70%) had VT/VF/SCD. As the Valt increased, the percentage of patients who suffered from VT/VF/SCD also increased. At pacing rates of
100 bpm, however, the percentage of patients with VT/VF/SCD decreased as the pacing heart rate increased.
|
Event-Free Survival
Event-free survival calculated by the Kaplan-Meier method during the 45 months of follow-up is shown in Figure 3. Patients with TWA whose Valt was >1.9 µV at a rate of 90 bpm suffered from VT/VF/SCD more frequently than did patients with TWA whose Valt was <1.9 µV. Similarly, VT/VF/SCD occurred more frequently in patients with >1.9-µV Valt than in patients with <1.9-µV Valt even at heart rates of 100, 110, and 120 bpm.
|
Sensitivity and Specificity
The ability of TWA to predict VT/VF/SCD statistically at each heart rate is shown in Table 2. The highest positive predictive value is 70%, when the heart rate is 90 bpm and Valt is >2.9 µV. The highest negative predictive value is 100%, at a heart rate of 120 bpm and Valt <0.9 µV.
|
Representative Case
A representative case of TWA measurement with atrial pacing is shown in Figure 4. The patient is a 46-year-old man with dilated cardiomyopathy who had polymorphic VT induced by programmed electrical stimulation. Valt was >1.9 µV at a rate of 90 bpm and increased with the pacing heart rate. Twenty-two months after TWA testing, he suffered from VF that was terminated by his ICD.
|
| Discussion |
|---|
|
|
|---|
Positivity Criteria and VT
Currently, there is only 1 machine that can measure Valt in the clinical setting. Most clinical studies on microvolt TWA use the same positivity criteria: >1.9 µV of Valt at a heart rate <110 bpm without artifact or noise. Sensitivity was found to be 78% to 93% and specificity was 45% to 61% for predicting VT/VF/SCD according to these criteria.2,3,10,14,15 The characteristics of TWA testing by exercise were found to be those of high sensitivity and moderate specificity to predict VT. The heart rate at which Valt exceeds positivity criteria is called onset heart rate. Previously, we retrospectively showed that patients with sustained VT and structural heart disease had lower onset heart rates of microvolt TWA using these criteria.5 The present prospective study showed that patients with a high Valt at a relatively low heart rate have an increased incidence of VT during follow-up periods. This finding may make it possible to stratify patients by risk. Measurement of Valt at different heart rates may help to differentiate patients who need an ICD from those who do not. The Multicenter Automatic Defibrillator Implantation Trial II (MADIT II) revealed that prophylactic ICD implantation for patients with a prior myocardial infarction and left ventricular dysfunction improved prognoses.16 In fact, it is not realistic to implant an ICD in all patients who meet the MADIT II criteria. We have to select the patients who can benefit most from ICD implantation. TWA testing might be useful in selecting such patients.17
Onset Heart Rate and VT
There is only 1 study in which onset heart rate was used as a risk factor for VT.8 That report showed that onset heart rate of patients with VT was lower than that of patients without VT. However, the problem is how to decide the onset heart rate when exercise is used to increase heart rate. Because the CH2000 system uses fast Fourier transformation as the means of alternans measurement, it is necessary to obtain 256 consecutive steady-state beats to measure Valt accurately. When exercise is used to increase heart rate for TWA testing, the heart rate at the beginning of the 256-beat cycle is definitely different from that at the end of the cycle. It is not easy to determine the onset heart rate precisely by exercise. In addition, autonomic nervous tone during atrial pacing at rest differs from that of exercise. It is possible that TWA induced by atrial pacing is different from that induced by exercise because autonomic nervous tones influence TWA. Thus, we chose to use the atrial pacing method to obtain a fixed heart rate, which made it possible to evaluate the effect of heart rate on TWA precisely.
TWA as a Predictor of VT
Although predictive factors for VT/VF/SCD exist, they are not necessarily satisfactory. TWA testing may be a promising predictor for VT. TWA testing for predicting VT is of high sensitivity, moderate specificity, low positive predictive value, and high negative predictive value. With such characteristics, TWA testing would be suitable for screening of patients with structural heart disease. However, Tapanainen et al18 reported that sustained TWA during the predischarge exercise test after acute myocardial infarction does not indicate increased risk for mortality and that an incomplete TWA test and several common risk factors provided prognostic information in this postacute myocardial infarction population. Recently, Hohnloser et al17 reported that MADIT IItype patients who test negative for microvolt TWA may not benefit from defibrillator therapy. Nonetheless, it is probably possible to stratify patients with arrhythmic risk in more detail when Valt is measured precisely at a stable heart rate as we describe here. In the present study, 7 of 10 patients with 2.9 µV of Valt at a heart rate of 90 bpm suffered from VT, whereas 0 of 64 patients with <0.9 µV of Valt at a heart rate of 120 bpm had VT during the follow-up period.
Study Limitations
Several limitations of the present study need to be addressed. First, the sample size is small. Higher numbers may provide additional insight. The second is the variety of heart disease in the patients in this study. Moreover, this study included patients with supraventricular tachyarrhythmia who were at low risk of an event. Finally, appropriate ICD therapies as the end point were used in this study. The use of this end point as a surrogate marker may overestimate the occurrence of events.
TWA is a promising new predictor for VT, but several problems must be resolved. First, the number of patients diagnosed as "indeterminate" was not small. Seven percent of the patients in this study were diagnosed as indeterminate even with the use of atrial pacing at a rate of 90 bpm. The reasons were that heart rate did not increase enough to analyze TWA, heart rate increased too rapidly to analyze TWA, there were frequent ectopic beats, or there was excessive noise. In addition, TWA testing is never applicable in patients with atrial fibrillation.
| Conclusions |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. M. Narayan, M. R. Franz, G. Lalani, J. Kim, and A. Sastry T-Wave Alternans, Restitution of Human Action Potential Duration, and Outcome J. Am. Coll. Cardiol., December 18, 2007; 50(25): 2385 - 2392. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Cantillon, K. M. Stein, S. M. Markowitz, S. Mittal, B. K. Shah, D. P. Morin, E. S. Zacks, M. Janik, S. Ageno, A. C. Mauer, et al. Predictive Value of Microvolt T-Wave Alternans in Patients With Left Ventricular Dysfunction J. Am. Coll. Cardiol., July 10, 2007; 50(2): 166 - 173. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. Narayan, D. D. Drinan, R. P. Lackey, and C. F. Edman Acute volume overload elevates T-wave alternans magnitude J Appl Physiol, April 1, 2007; 102(4): 1462 - 1468. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Krogh-Madsen and D. J. Christini Action Potential Duration Dispersion and Alternans in Simulated Heterogeneous Cardiac Tissue with a Structural Barrier Biophys. J., February 15, 2007; 92(4): 1138 - 1149. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. Narayan T-Wave Alternans and the Susceptibility to Ventricular Arrhythmias J. Am. Coll. Cardiol., January 17, 2006; 47(2): 269 - 281. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. L. Koller, S. K.G. Maier, A. R. Gelzer, W. R. Bauer, M. Meesmann, and R. F. Gilmour Jr Altered Dynamics of Action Potential Restitution and Alternans in Humans With Structural Heart Disease Circulation, September 13, 2005; 112(11): 1542 - 1548. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Gehi, R. H. Stein, L. D. Metz, and J. A. Gomes Microvolt T-Wave Alternans for the Risk Stratification of Ventricular Tachyarrhythmic Events: A Meta-Analysis J. Am. Coll. Cardiol., July 5, 2005; 46(1): 75 - 82. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|