(Circulation. 2004;109:443-451.)
© 2004 American Heart Association, Inc.
Review: Clinical Cardiology: New Frontiers |
From the Division of Cardiology, Mount Sinai Medical Center and Miami Heart Institute, Miami Beach, Fla (G.A.L., C.H.H., A.M.), and the Medical College of Virginia, Richmond Va (K.A.E.).
Correspondence to Gervasio A. Lamas, MD, Mount Sinai Medical CenterMiami Heart Institute, 4300 Alton Rd, Butler Building, Miami Beach, FL 33140. E-mail most95{at}aol.com
Key Words: pacemakers physiology trials
| Introduction |
|---|
|
|
|---|
The focus of development of cardiac pacing has changed over the last several decades. In the 1950s through 1970s, the emphasis was on the development of more reliable power sources and leads. By the end of the 1970s, the power source problems had been largely overcome, and efforts focused on mimicking the normal physiology of cardiac contraction.
In the 1980s, in the United States, the focus on normal physiology led to the nearly universal application of dual-chamber pacing to restore atrioventricular (AV) synchrony in patients with sinus rhythm. As sensor technology improved in the 1980s and 1990s,3,4 the same physiological focus spurred the advent and generalization of sensor-driven, rate-modulated pacing to restore chronotropic competence. By the early 1990s, there were lessons from physiology and an emerging body of evidence from observational epidemiological studies.5,6 Indeed, some pacemaker physicians believed that randomized trials in pacing might not be necessary to demonstrate incremental gains in prevention of clinical events and improvement of quality of life with atrial-based pacing.7
It is certainly true that for many hypotheses, randomized trials are neither necessary nor desirable. For example, it would have been unethical to conduct randomized trials of cardiac pacing for the treatment of Stokes-Adams-Morgagni attacks because the benefit was so large.8 As a result of the evolution of progressively more sophisticated technologies, the most plausible effect sizes have become small to moderate, or between a 15% and 30% relative reduction in risk. Such effects, however, have importance clinically as well as from a public health perspective, but they are difficult to detect reliably. This is because the amount of uncontrolled and uncontrollable confounding inherent in even well-designed and well-conducted observational epidemiological studies is as large as the most plausible effects. Thus, in such circumstances the only reliable design strategy is the large-scale randomized trial.911
In this article, we review emerging evidence on cardiac pacing, including basic research, primarily with a physiological focus, observational epidemiological studies, and randomized trials (Table 1). A glossary of pacing terms appears in Table 2.
|
|
| AV Synchrony, Cardiac Output, and Clinical Benefit |
|---|
|
|
|---|
These generally consistent improvements in cardiac output led to several apparently logical conclusionsthat AV sequential pacing would reduce the risk of heart failure, reduce mortality, and improve quality of life.
Indeed, as dual-chamber pacing was becoming the commonest form of pacing in the United States, several large-scale observational epidemiological studies with good statistical methodology appeared that seemed to suggest that demonstrating a survival improvement would only be a formality.18 For example, the largest of these studies reviewed >36 000 Medicare pacemaker implants and demonstrated that dual-chamberpaced patients had a lower mortality at 1 year and 2 years than ventricular-paced patients (1-year mortality, dual-chamber 13.7% and ventricular 18.3%, P<0.001; 2-year mortality, dual-chamber 22.3% and ventricular 28.9%, P<0.001).19 Not unexpectedly, there were many demographic and other differences between the 2 groups. However, after controlling for these differences, selection of a dual-chamber pacemaker remained as an independent predictor of survival (odds ratio for mortality with dual-chamber pacing 0.82, 95% confidence interval 0.77 to 0.87). Reviewers of these retrospective studies20 urged caution in the interpretation of these data and encouraged the performance of large-scale randomized trials.21 Eight years later, all the large-scale trials comparing dual-chamber with ventricular pacing have been completed, and all but 1 have been reported. The results of these trials have provided valuable hints as to the clinical physiology of ventricular contraction and cardiac pacing and are spawning another generation of pacing trials. Evidence-based recommendations for the use of dual-chamber and rate-modulated pacing systems in bradycardia patients are shown in Table 3.
|
In 1994, Andersen et al22 reported the results of the first randomized trial comparing single-chamber atrial (AAI) with VVI pacing. Patients were screened with an atrial-pacing test to assure normal AV conduction before randomization. Thus, 225 patients (mean age of 76 years) with sinus-node dysfunction, normal AV conduction, and a normal QRS complex were randomized. In the initial report with a mean follow-up of 3.3 years, there was no significant difference in survival. Longer follow-up of these trial participants to a mean of 5.5 years was subsequently published.23 Cardiovascular death had occurred in 19 patients from the atrial-pacing group and in 39 patients from the ventricular-pacing group (P=0.0065). In 1998, a subsequent analysis showed further long-term advantages of atrial pacing in terms of heart failure and echocardiographic changes.24 After a mean follow-up of 5.5 years, New York Heart Association (NYHA) class was significantly lower in the atrial-pacing group (NYHA class I, 84%) compared with the ventricular-pacing group (NYHA class I, 65%). Left atrial diameter increased significantly in both groups but the increase was significantly lower in the atrial-pacing group (at end of follow-up, atrial pacing left atrial diameter was 37±7 mm; ventricular pacing left atrial diameter 41±7 mm; P<0.0005). Finally, left ventricular fractional shortening decreased significantly in the ventricular-pacing group but not in the atrial-pacing group.
These results have not been reproduced by any other prospective, randomized trial comparing atrial-based to ventricular pacing. For example, an intermediate-sized trial, PAcemaker Selection in the Elderly (PASE), randomized 407 patients age 65 years or older with bradycardia requiring a pacemaker to receive a dual-chamber pacemaker programmed to DDDR (rate-modulated dual-chamber pacing) versus VVIR (rate-modulated ventricular pacing) mode.25 No significant differences were found in clinical events including death, stroke, or hospitalization for heart failure, although some of trends compatible with benefits of dual-chamber pacing were observed, particularly in the subset of patients with sinus-node disease.
The results of the 2 largest randomized trials of atrial-based pacing compared with ventricular pacing, the Canadian Trial of Physiological Pacing (CTOPP) and the Mode Selection Trial in Sinus Node Dysfunction (MOST), also have failed to uncover differences in survival between groups26,27 (Figure 1).
|
In 2000, Connolly et al26 reported CTOPP, in which 2568 patients (mean age of 73 years) with symptomatic bradycardia requiring permanent pacing, were randomized to atrial-based (AAI, AAIR, DDD, or DDDR) or ventricular pacing (VVI or VVIR) and monitored for an average of 3 years. DDD and DDDR were used in 95% of atrial-based pacingassigned patients. The combined primary endpoint of stroke or cardiovascular death was not significantly different between pacing modalities, with an annual rate of death or stroke of 4.9% for atrial-based pacing and 5.5% for ventricular pacing (P=0.33). There were no differences in the incidence of heart failure hospitalization. More recently, the CTOPP patient population underwent extended follow-up. There were no significant differences in the rate of death or stroke after a longer follow-up of almost 8 years (reported in the May 2002 Annual Scientific Meeting of the North American Society of Pacing and Electrophysiology).
MOST, reported in 2002, was a 6-year trial comparing rate-modulated ventricular with dual-chamber pacing (VVIR versus DDDR) in 2010 patients with sinus-node dysfunction. The average follow-up duration was 2.7 years. The primary endpoint was a composite of nonfatal stroke or death from any cause.
The primary endpoint occurred in 22.2% of the patients and no significant difference was found between the two treatment groups (P=0.32). There were no significant differences between groups with regard to all-cause mortality or cardiovascular mortality. However, there were significantly fewer hospitalizations for heart failure in the dual-chamberpacing group (10.3%) than in the ventricular-pacing group (12.3%, P=0.021). It is an understatement that the results of these 2 trials, randomizing nearly 5000 patients, have been a surprise to pacemaker physicians and have contributed to an ongoing fundamental rethinking of physiological pacing.28,29 This rethinking of physiology began with clinical applications in advanced heart failure, but the concepts may be applicable to bradycardia pacing as well.
AV synchrony is only a component of the physiologically normal heartbeat; ventricular activation sequence is now being appreciated as perhaps equally important. In the dog, Wiggers in 1925 reported that artificial stimulation of the right ventricle led to a less effective cardiac systole than did a normal contraction resulting from a normal cardiac depolarization.30 Other studies suggested that AV synchrony and right ventricularleft ventricular (V-V) synchrony were both independent and additive in contributing to pacing-dependent hemodynamic deterioration.3133 Although the importance of V-V synchrony in the instrumented dog is clear, its physiological and clinical importance in pacemaker patients has been overshadowed by the technical successes of providing AV synchrony and rate modulation and by the practical difficulties inherent in providing V-V synchrony.
Right ventricular (RV) apical pacing results in an asynchronous contraction whereby the electrical wavefront proceeds through slowly conducting myocardial tissue instead of through the more rapidly conducting His-Purkinje system. As such, it mimics the effect of a spontaneous left bundle-branch block, an abnormality of conduction that has been carefully studied. Grines et al34 compared patients with left bundle-branch block to patients with normal conduction. Using noninvasive techniques, they found reduced septal ejection fraction (EF) leading to a reduction in global EF, as well as unfavorable alterations in diastolic filling time. Other investigators have reported similar results.35 The relative importance of AV and V-V synchrony in pacemaker patients has been investigated by several groups of investigators. Rosenqvist et al36 measured regional and global EF during atrial pacing with a conducted QRS, AV sequential pacing, and RV pacing. The global EF was highest during atrial pacing, intermediate during AV sequential pacing, and lowest during ventricular pacing. Other studies have shown that asynchronous ventricular activation increases or leads to mitral regurgitation in some patients.37 More recently, several studies have demonstrated a wide variety of structural changes occurring related to ventricular pacing. For example, the asynchronous activation of the ventricle caused by ventricular pacing results in regional differences in myocardial blood flow, glucose uptake, oxygen consumption, and wall motion abnormalities. These regional differences may be marked by patchy or asynchronous regional differences in myocardial work and blood flow. Ventricular pacing also leads to structural abnormalities of ventricular myocytes, including changes in protein expression and ionic currents.3841 Thus, the abnormal contractile pattern has structural and cellular correlates as well.
Careful analyses of data in MOST are providing hints that, in the clinical arena, the benefits of AV synchrony may be mitigated by V-V dyssynchrony. Sweeney et al42 analyzed patients with a baseline, prepacing QRS duration of <120 ms and determined the relationship between percentage ventricular pacing, as determined by pacer event counters, and clinical events. In the DDDR-programmed patients, median cumulative percentage ventricular pacing was 90%. In the same patient group, despite maintenance of AV synchrony, percentage ventricular pacing was an independent predictor of adjudicated heart failure hospitalization (for DDDR-paced patients, hazard ratio for heart failure hospitalization of cumulative percentage ventricular pacing >40% was 2.99; 95% confidence interval, 1.15 to 7.75).42
In conclusion, although AV synchrony supports an improvement in cardiac output and ventricular pressures, these favorable hemodynamics may be attenuated by ventricular dyssynchrony from RV apical pacing. This concept may explain the discrepant results when the Andersen et al22 trial is compared with MOST and CTOPP. The atrial-pacing arm of the Andersen et al22 trial could only pace atrium because the device implanted was an atrial pacemaker. In contrast, the "physiological" device implanted in CTOPP and MOST was a dual-chamber pacemaker, which paced ventricle depending on spontaneous AV conduction and programmed AV delay.
At present, at least 2 trials are ongoing that explore the benefit of reducing ventricular pacing in patients with a baseline normal QRS. The Danish Multicenter Randomized Study on Atrial Inhibited Versus Dual-Chamber Pacing in Sick Sinus Syndrome (DANPACE) compares atrial pacing (AAI) with DDD pacing programmed to have a short AV delay and obligatory ventricular capture. The Search AF Extension for Promoting Atrioventricular Conduction (SAVE-PACE) study, an industry-sponsored trial, tests whether a programmable feature that allows the AV delay to be extended up to 350 ms will reduce percentage ventricular pacing and lead to a reduction in left ventricular remodeling and atrial fibrillation (AF). Thus, the clinical significance of reducing forced ventricular desynchronization by RV pacing is being tested by the new generation of randomized trials. In time, the venerable RV apical lead may give way to the new imperative of V-V synchrony.
| Atrial Fibrillation |
|---|
|
|
|---|
In 1994, Andersen et al22 reported the results of the first randomized trial comparing atrial (AAI) with ventricular (VVI) pacing. Perhaps as a result of the very small rates of events, there were no significant differences in incidence of AF (AF with atrial pacing 14% versus AF with ventricular pacing 23%; P=0.12). Three years later, longer follow-up of these trial participants to a mean of 5.5 years was published.23 The proportion of patients who remained free of incident or chronic AF was significantly higher in the atrial-pacing group as compared with the ventricular-pacing group (see Table 1 for details on effect size). Perhaps because of limited statistical power, a nonsignificant trend associated ventricular pacing with chronic AF (P=0.06). Interestingly, a later analysis reported in 1998 showed that left atrial diameter increased significantly less in the atrial-paced group (P<0.0005) than in the ventricular-paced group.24 The PASE trial, reviewed earlier, reported a marginal reduction in risk of AF in its initial publication.25 However, a more focused multivariable analysis found that DDDR pacing was significantly associated with a reduced incidence of AF.46
The large pacemaker trials also have found similar results (Figure 1). For example, CTOPP reported that the annual rate of AF was lower in the atrial-based pacing group (5.3%) as compared with the ventricular-paced group (6.6%) (P=0.05). This constituted a relative risk reduction of 18% and an absolute risk reduction of 3.9% for atrial-based pacing. In MOST, perhaps because only patients with sinus-node dysfunction were included, the yearly incidence of AF was higher than in CTOPP (
8%), but the relative risk reduction in the overall group quite similar (21%, P=0.008). A relevant observation that may explain varying results with pacing among different trials is that MOST patients who had never previously reported AF before pacemaker implantation had a large (50%) reduction (hazard ratio 0.50; 95% confidence interval 0.32 to 0.76; P=0.0011) in the incidence of AF after randomization. In contrast, dual-chamberpaced patients with a prior history of AF had a smaller, nonsignificant 14% reduction. This observation suggests that sinus-node dysfunction patients who have advanced further toward development of permanent AF, perhaps as a result of the effects of electrical and mechanical remodeling, may be less responsive to pacing modalities to prevent or treat AF. Indeed, this is a potential explanation for the sole failure to find a significant reduction in AF in a clinical trial of atrial-based compared with ventricular-based pacing in sinus-node dysfunction.47 Finally, the effects of progressive adverse atrial remodeling as described above in the ventricular-paced patients may be a reason that the incidence of AF continues to increase throughout follow-up in the ventricular-paced group.
An additional pacing technique has been used to specifically test the concept that a diminution in global atrial activation times and in the conduction delays may facilitate the initiation of AF. Dual-site atrial pacing involves pacing the right and left atria simultaneously.48,49 Although the technique showed much early promise, the most recent and largest clinical trial failed to demonstrate a conclusive benefit in reducing AF rates in comparison with single-site right atrial pacing.50 However, the trial had a small sample size (118 patients) and methodological problems, such as a crossover design, that limit the ability to exclude a small to moderate benefit.
| Stroke |
|---|
|
|
|---|
Nonetheless, 2 small trials of ventricular versus atrial-based pacing reported that atrial-based pacing reduced stroke risk.23,51 However, despite the plausible mechanisms listed above, the large trials have demonstrated that preventing stroke in pacemaker patients is as elusive a goal as preventing death. In CTOPP, the annual risk of stroke was 1.1% in the ventricular-paced group and 1.0% in the atrial-based pacing group. In MOST, the annualized incidence of stroke was 2.2%, with no clear difference based on pacing mode. Thus, whether there is a reduction in stroke risk of small (10% to 15%) magnitude will await planned meta-analyses of the published pacemaker trials. At the present time, dual-chamber pacing should not be selected to prevent stroke.
| Pacemaker Syndrome |
|---|
|
|
|---|
The symptoms and signs of pacemaker syndrome can be divided into the following 3 different types: congestive, hypotensive, and nonspecific. Congestive symptoms mimic heart failure, with dyspnea and orthopnea. Congestive signs include elevated neck veins (often with cannon A waves), rales, hepatomegaly, and pedal edema. Hypotensive symptoms and signs include near syncope or syncope at onset of pacing, associated with a drop in upright systolic blood pressure of at least 20 mm Hg. Nonspecific symptoms such as headache and fatigue have also been reported. The initial thoughts on the pathophysiology of pacemaker syndrome invoked only a drop in cardiac output in response to loss of AV synchrony. However, pacemaker syndrome is now recognized as a complex interaction of neurohumoral, autonomic, and vascular changes that ultimately lead to symptomatic hemodynamic consequences.
The presence of intact ventriculoatrial conduction has been thought to be central to the diagnosis of pacemaker syndrome; the hemodynamic effects of atrial systole against closed AV valves likely accounts for some of the congestive symptoms and signs12 (Figure 2). However, every pacemaker clinician has experienced the patient who develops profound hypotension with ventricular pacing, and it is in these patients that the role of the autonomic nervous system becomes most relevant.5355 Hypotension during ventricular pacing is now thought to be due to individual variation in compensatory sympathetic outflow, the effects of vasodilator drugs or vascular disease, and the activation of cardioinhibitory reflexes that are brought into play by atrial distension, and that counteract the reflex increases in adrenergic tone.
|
Clinical trial results have shed light on the incidence and manifestations of pacemaker syndrome, although conclusions drawn from US trials are quite different from those drawn from Canadian or European trials. Earlier retrospective surveys had suggested that up to 83% of ventricular-paced patients had pacemaker syndrome.56 Ellenbogen et al57 reported on patients in the PASE trial, age 65 years or older, who had been randomized to VVIR. This trial required a DDDR pacemaker in all patients, programmed at implant to VVIR or DDDR. Of 204 patients randomized to VVIR, 26% crossed over to DDDR for "intolerance to ventricular pacing," the operating definition of pacemaker syndrome. Most crossovers occurred early (Figure 3). In patients who had undergone quality-of-life assessments before and after crossover, significant improvements in quality of life were demonstrated. In MOST, pacemaker syndrome was the principal reason for crossover in 18.3% of 996 ventricular-paced patients and in 48.9% of all crossovers. Most crossovers for pacemaker syndrome occurred early (69% by 3 months and 73% by 6 months; median time to crossover 58 days). The Andersen et al22 trial and CTOPP both have reached diametrically opposite conclusions on the incidence and consequences of pacemaker syndrome. Both of these trials, in contrast to the US trials, were "hardware randomization" trials, in which the device was randomized, not the programming. Thus, crossovers required reoperation. Not unexpectedly, the rate of crossovers in both studies was low, and only 5% of the CTOPP and 1.8% of the Andersen et al22 ventricular-paced patients were crossed over to an atrial-based mode.
|
Our interpretation of these radically different data are that one series of studies made it inordinately hard to cross over, whereas the US trials perhaps made it too easy. Indeed, although pacemaker syndrome is a real and important diagnosis, the real incidence probably lies between the 2 extremes.
| Chronotropic Incompetence and Rate-Modulated Pacing |
|---|
|
|
|---|
Despite convincing physiological data, clinical investigators have found an inconsistent improvement in the exercise capacity of DDDR-paced patients when compared with DDD patients.5961 Nonetheless, unpublished data on >200 patients from the Rate Modulated Pacing and Quality of Life (RAMP) study suggest that chronotropic incompetence is an important correlate of health-related quality of life. Patients unable to reach 60% maximum predicted heart rate had worse quality of life in all dimensions than did patients who exceeded 60% maximum predicted heart rate, leaving open the question that aggressive rate-modulated pacing might improve quality of life. Furthermore, chronotropic incompetence may develop after the pacemaker has been implanted.62 Small crossover studies in highly selected patients already have suggested that DDDR pacing leads to superior quality of life.63 These small studies, however, do not provide definitive support for the concept that all or most pacemaker patients with sinus rhythm should have a DDDR pacer implanted; and indeed, it may be a realistic concern that more aggressive DDDR programming may lead to a greater frequency of ventricular pacing and more V-V dyssynchrony.
Despite the absence of convincing data showing that, in unselected populations, DDDR pacing is superior to DDD with regard to improved quality of life and reduced symptoms,
97% of all generators implanted in the United States in 2000 had rate modulation as a programmable option.64,65 The Advanced Elements of Pacing Trial (ADEPT), a well-powered trial comparing dual-sensor dual-chamber DDDR with DDD pacing will fill this important gap in knowledge.
| A Clinical Interpretation of the Randomized Data: Ventricular Versus Atrial-Based Pacing |
|---|
|
|
|---|
| Conclusions |
|---|
|
|
|---|
|
| References |
|---|
|
|
|---|
2. Furman S, Robinson G. Use of an intracardiac pacemaker in the correction of total heart block. Surg Forum. 1958; 9: 245248.[Medline] [Order article via Infotrieve]
3. Benditt DG, Milstein S, Buetikofer J, et al. Sensor-triggered rate-variable cardiac pacing. Ann Int Med. 1987; 107: 714724.
4. Leung SK, Lau CP, Camm J. An overview of sensors: ideal characteristics sensor combination, and automaticity. In: Ellenbogen KA, Kay GN, Wilkoff BL, eds. Clinical Cardiac Pacing and Defibrillation. 2nd ed. Philadelphia, Pa: WB Saunders; 2000: 219248.
5. Lamas GA, Estes NM III, Schneller S, et al. Does dual chamber or atrial pacing prevent atrial fibrillation? The need for a randomized controlled trial. Pacing Clin Electrophysiol. 1992; 15: 11091113.[CrossRef][Medline] [Order article via Infotrieve]
6. Bernstein AD, Parsonnet V. Strategies for mode selection in antibradyarrhythmia pacing. Cardiol Clin. 1992; 10: 719734.[Medline] [Order article via Infotrieve]
7. Ovsyshcher IE, Hayes DL, Furman S. Dual-chamber pacing is superior to ventricular pacing: fact or controversy? Circulation. 1998; 97: 23682370.
8. Edhag O, Swahn A. Prognosis of patients with complete heart block or arrhythmic syncope who were not treated with artificial pacemakers. Acta Med Scand. 1976; 200: 457463.[Medline] [Order article via Infotrieve]
9. Hennekens CH, Buring JE. Epidemiology in Medicine. Boston, Mass: Little, Brown, and Company; 1987.
10. Hennekens CH, Buring JE. Observational evidence. Ann N Y Acad Sci. 1994; 703: 1824.[CrossRef]
11. Montanez A, Hennekens CH, Zebede J, et al. Pacemaker mode selection: the evidence from randomized trials. Pacing Clin Electrophysiol. 2003; 26: 12701282.[CrossRef][Medline] [Order article via Infotrieve]
12. Lamas GA. Physiological consequences of normal atrioventricular conduction: applicability to modern cardiac pacing. J Card Surg. 1989; 4: 8998.[Medline] [Order article via Infotrieve]
13. Samet P, Castillo C, Bernstein WH. Hemodynamic sequelae of atrial ventricular and sequential atrioventricular pacing in cardiac patients. Am Heart J. 1966; 72: 725729.[CrossRef][Medline] [Order article via Infotrieve]
14. Rahimtoola SH, Ehsani A, Sinno MZ, et al. Left atrial transport function in myocardial infarction: Importance of its booster pump function. Am J Med. 1975; 59: 686694.[CrossRef][Medline] [Order article via Infotrieve]
15. Boucher CA, Pohost GM, Okada RD, et al. Effect of ventricular pacing on left ventricular function assessed by radionuclide angiography. Am Heart J. 1983; 106: 11051111.[CrossRef][Medline] [Order article via Infotrieve]
16. Hoeschen RJ, Reimold SC, Lee RT, et al. The effect of posture on the response to atrioventricular synchronous pacing in patients with underlying cardiovascular disease. Pacing Clin Electrophysiol. 1991; 14: 756759.[CrossRef][Medline] [Order article via Infotrieve]
17. Greenberg B, Chatterjee K, Parmley WW, et al. The influence of left ventricular filling pressure on atrial contribution to cardiac output. Am Heart J. 1979; 98: 742751.[CrossRef][Medline] [Order article via Infotrieve]
18. Hesselson AB, Parsonnet V, Bernstein AD, et al. Deleterious effects of long-term single-chamber ventricular pacing in patients with sick sinus syndrome: the hidden benefits of dual-chamber pacing. J Am Coll Cardiol. 1992; 19: 15421549.[Abstract]
19. Lamas GA, Pashos CL, Normand SLT, et al. Permanent pacemaker selection and subsequent survival in elderly Medicare pacemaker recipients. Circulation. 1995; 91: 10631069.
20. Connolly SJ, Kerr C, Gent M, et al. Dual-chamber versus ventricular pacing critical appraisal of current data. Circulation. 1996; 94: 578583.
21. Lamas GA. Pacemaker mode selection and survival: a plea to apply the principles of evidence based medicine to cardiac pacing practice. Heart. 1997; 78: 218220.
22. Andersen HR, Thuesen L, Bagger JP, et al. Prospective randomised trial of atrial versus ventricular pacing in sick-sinus syndrome. Lancet. 1994; 344: 15231528.[CrossRef][Medline] [Order article via Infotrieve]
23. Andersen HR, Nielsen JC, Thomsen PE, et al. Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome. Lancet. 1997; 350: 12101216.[CrossRef][Medline] [Order article via Infotrieve]
24. Nielsen JC, Andersen HR, Thomsen PE, et al. Heart failure and echocardiographic changes during long-term follow-up of patients with sick sinus syndrome randomized to single-chamber atrial or ventricular pacing. Circulation. 1998; 97: 987995.
25. Lamas GA, Orav EJ, Stambler BS, et al. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing. N Engl J Med. 1998; 338: 10971104.
26. Connolly SJ, Kerr CR, Gent M, et al. Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes. N Engl J Med. 2000; 342: 13851391.
27. Lamas GA, Lee KL, Sweeney MO, et al. Ventricular pacing or dual-chamber pacing for sinus-node dysfunction. N Engl J Med. 2002; 346: 18541862.
28. Andersen HR, Nielsen JC. Pacing in sick sinus syndromeneed for a prospective randomized trial comparing atrial with dual chamber pacing. Pacing Clin Electrophysiol. 1998; 21: 11751179.[CrossRef][Medline] [Order article via Infotrieve]
29. Toquero J, Geelen P, Goethals M, et al. J Cardiovasc Electrophysiol. 2001; 12: 14251428.[CrossRef][Medline] [Order article via Infotrieve]
30. Wiggers CJ. The muscular reactions of the mammalian ventricles to artificial surface stimuli. Am J Physiol. 1925; 73: 346378.
31. Grover M, Glantz SA. Endocardial pacing site affects left ventricular end diastolic volume and performance in intact anesthetized dog. Circ Res. 1983; 53: 7285.
32. Burkhoff D, Oikawa RY, Sagawa K. Influence of pacing site on canine left ventricular contraction. Am J Physiol. 1986; 251: H428H435.[Medline] [Order article via Infotrieve]
33. Zile MR, Blaustein AS, Shimizu G, et al. Right ventricular pacing reduces the rate of left ventricular relaxation and filling. J Am Coll Cardiol. 1987; 10: 702709.[Abstract]
34. Grines CL, Bashore TM, Boudoulas H, et al. Functional abnormalities in isolated left bundle branch block. The effect of interventricular asynchrony. Circulation. 1989; 79: 845853.
35. Bramlet DA, Morris KG, Coleman RE, et al. Effects of rate-dependent left bundle branch block on global and regional left ventricular function. Circulation. 1983; 67: 10591065.
36. Rosenqvist M, Isaaz K, Botvinick EH, et al. Relative importance of activation sequence compared to atrioventricular synchrony in left ventricular function. Am J Cardiol. 1991; 67: 148156.[CrossRef][Medline] [Order article via Infotrieve]
37. Mark JB, Chetham PM. Ventricular pacing can induce hemodynamically significant mitral valve regurgitation. Anesthesiology. 1991; 74: 375377.[CrossRef][Medline] [Order article via Infotrieve]
38. Prinzen F, Peschar M. Relation between the pacing induced sequence of activation and left ventricular pump function in animals. Pacing Clin Electrophysiol. 2002; 25: 484498.[CrossRef][Medline] [Order article via Infotrieve]
39. Lee MA, Dae MW, Langberg JJ, et al. Effects of long-term right ventricular apical pacing on left ventricular perfusion, innervation, function and histology. J Am Coll Cardiol. 1994; 24: 225232.[Abstract]
40. Tse H-F, Lau C-P. Long term effect of right ventricular pacing on myocardial perfusion and function. J Am Coll Cardiol. 1997; 29: 744749.[Abstract]
41. Nielsen JC, Bottcher M, Nielsen TT, et al. Regional myocardial blood flow in patients with sick sinus syndrome randomized to long-term single chamber atrial or dual chamber pacingeffect of pacing mode and rate. J Am Coll Cardiol. 2000; 35: 14531461.
42. Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003; 107: 29322937.
43. Saksena S, Mehra R, Ellenbogen KA. Pacing for prevention of tachyarrhythmias. In: Ellenbogen KA, Kay GN, Wilkoff BL, eds. Clinical Cardiac Pacing and Defibrillation, 2nd ed. Philadelphia, Pa: WB Saunders; 2000: 479507.
44. Klein LS, Miles WM, Zipes DP. Effect of the atrioventricular interval during pacing or reciprocating tachycardia on atrial size, pressure, and refractory period. Circulation. 1990; 82: 6068.
45. Calkins J, El-Atassi R, Leon A. Effect of the atrioventricular relationship on atrial refractoriness in humans. Pacing Clin Electrophysiol. 1992; 15: 771778.[Medline] [Order article via Infotrieve]
46. Stambler BS, Ellenbogen KA, Orav EJ, et al. Predictors and clinical impact of atrial fibrillation after pacemaker implantation in elderly patients treated with dual chamber versus ventricular pacing. Pacing Clin Electrophysiol. 2003; 26: 20002007.[CrossRef][Medline] [Order article via Infotrieve]
47. Wharton JM, Sorrentino RA, Campbell P, et al. Effect of pacing modality on atrial tachyarrhythmia recurrence in the tachycardia-bradycardia syndrome: preliminary results of the Pacemaker Atrial Tachycardia Trial. Circulation. 1998; 98 (suppl I): I494.Abstract.
48. Daubert C, Mabo Ph, Berder V, et al. Atrial tachyarrhythmias associated with high degree interatrial conduction block: prevention by permanent atrial resynchronization. Eur J Card Pacing Electrophysiol. 1994; 1: 3544.
49. Saksena S, Prakash A, Hill M, et al. Prevention of recurrent atrial fibrillation with chronic dual-site right atrial pacing. J Am Coll Cardiol. 1996; 28: 68794.[Abstract]
50. Saksena S, Prakash A, Ziegler P, et al. Improved suppression of recurrent atrial fibrillation with dual-site right atrial pacing and antiarrhythmic drug therapy. J Am Coll Cardiol. 2002; 40: 11401150.
51. Ausubel K, Furman S. The pacemaker syndrome. Ann Intern Med. 1985; 103: 420429.
52. Mitsui T, Hori M, Suma K, et al. The pacemaking syndrome. In: Jacobs JE, ed. Proceedings of the Eighth Annual International Conference on Medical and Biological Engineering. Chicago, Ill: Association for Advancement of Medical Instrumentation; 1969: 2933.
53. Ellenbogen KA, Thames MD, Mohanty PK. New insights into pacemaker syndrome gained from hemodynamic humoral and vascular responses during ventriculo-atrial pacing. Am J Cardiol. 1990; 65: 5359.[Medline] [Order article via Infotrieve]
54. Pehrsson SK, Hjemdahl O, Nordlander R, et al. A comparison of sympatho-adrenal activity and cardiac performance at rest and during exercise in patients with ventricular demand or atrial synchronous pacing. Br Heart J. 1988; 60: 212220.
55. Ellenbogen KA, Wood MA, Stambler B. Clinical, hemodynamic, and neurohumoral features of pacemaker syndrome. In: Barold SS, Mugia J, eds. New Perspectives in Cardiac Pacing. Mt Kisco, NY: Futura; 1992.
56. Heldman D, Mulvihill D, Nguyen H, et al. True incidence of pacemaker syndrome: an iatrogenic condition. Pacing Clin Electrophysiol. 1990; 13: 17421750.[Medline] [Order article via Infotrieve]
57. Ellenbogen KA, Stambler BS, Orav EJ, et al. Clinical characteristics of patients intolerant to VVIR pacing. Am J Cardiol. 2000; 86: 5963.[CrossRef][Medline] [Order article via Infotrieve]
58. Bjerregaard P. Mean 24 hour heart rate, minimal heart rate, and pauses in healthy subjects 4079 years of age. Eur Heart J. 1983; 4: 4451.
59. Capucci A, Boriani G, Specchia S. Evaluation of DDDR versus DDD pacing. Pacing Clin Electrophysiol. 1992; 15: 19081913.[CrossRef][Medline] [Order article via Infotrieve]
60. Jutzy RV, Florio J, Isaeff DM, et al. Comparative evaluation of rate modulated dual chamber and VVIR pacing. Pacing Clin Electrophysiol. 1990; 13: 18381846.[Medline] [Order article via Infotrieve]
61. Lazarus A, Mitchell K, et al. A prospective multicenter study demonstrating clinical benefit with a new accelerometer-based DDDR pacemaker. Pacing Clin Electrophysiol. 1996; 19 (part II): 16941697.[CrossRef][Medline] [Order article via Infotrieve]
62. Gwinn N, Lemen R, Kratz J, et al. Chronotopic incompetence: a common and progressive finding in pacemaker patients. Am Heart J. 1992; 123: 12161219.[Medline] [Order article via Infotrieve]
63. Sulke N, Chambers J, Dritsas A, et al. A randomized double-blind crossover comparison of four rate responsive pacing modes. J Am Coll Cardiol. 1991; 17: 696706.[Abstract]
64. Morgan Stanley Dean Witter. Equity Research, North America. Healthcare: Pharmaceuticals, Hospital Supplies & Medical. New York, NY: Morgan Stanley Dean Witter; July 26, 2000.
65. Lamas GA, Williams A. Have the results of randomized clinical trials of pacing altered the practice of cardiac pacing? J Cardiovasc Electrophysiol. 2003; 14 (suppl): 115.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
G K Panicker, B Desai, and Y Lokhandwala Choosing pacemakers appropriately Heart Asia, March 31, 2009; 2009(3): 11 - 15. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. E. Vardas, H. E. Mavrakis, and W. D. Toff CHAPTER 27 Bradycardia ESC Textbook of Cardiovascular Medicine, January 1, 2009; 2(1): med-9780199566990-chapter - med-9780199566990-chapter. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-F. Tse and C.-P. Lau Clinical Trials for Cardiac Pacing in Bradycardia: The End or the Beginning? Circulation, July 4, 2006; 114(1): 3 - 5. [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, E. M. Antman, D. T. Anbe, P. W. Armstrong, E. R. Bates, L. A. Green, M. Hand, J. S. Hochman, H. M. Krumholz, F. G. Kushner, et al. ACC/AHA guidelines for the management of patients with ST-Elevation myocardial infarction--executive summary: A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (writing committee to revise the 1999 guidelines for the management of patients with acute myocardial infarction) J. Am. Coll. Cardiol., August 4, 2004; 44(3): 671 - 719. [Full Text] [PDF] |
||||
![]() |
E. M. Antman, D. T. Anbe, P. W. Armstrong, E. R. Bates, L. A. Green, M. Hand, J. S. Hochman, H. M. Krumholz, F. G. Kushner, G. A. Lamas, et al. ACC/AHA Guidelines for the Management of Patients With ST-Elevation Myocardial Infarction--Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 1999 Guidelines for the Management of Patients With Acute Myocardial Infarction) Circulation, August 3, 2004; 110(5): 588 - 636. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |