(Circulation. 2004;109:701-705.)
© 2004 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Department of Cardiovascular Pathology, Armed Forces Institute of Pathology, Washington, DC (R.V., A.F., F.D.K.), and the Cardiovascular (G.G., G.M., N.G., L.M., M.T., O.V.) and Pathology (T.M.) Departments, Ospedali Riuniti di Bergamo, Bergamo, Italy.
Correspondence to Renu Virmani, MD, Chair, Department of Cardiovascular Pathology, Armed Forces Institute of Pathology, 6825 16th St NW, Washington, DC. E-mail virmani{at}afip.osd.mil
Received December 1, 2003; revision received December 17, 2003; accepted December 29, 2003.
| Abstract |
|---|
|
|
|---|
Methods and Results We present findings of a 58-year-old man who died of late stent thrombosis 18 months after receiving 2 Cypher stents for unstable angina. Although angiographic and intravascular ultrasound results at 8 months demonstrated the absence of neointimal formation, vessel enlargement was present. An autopsy showed aneurysmal dilation of the stented arterial segments with a severe localized hypersensitivity reaction consisting predominantly of T lymphocytes and eosinophils.
Conclusions The known pharmacokinetic elution profile of Cypher stents and the presence of polymer fragments surrounded by giant cells and eosinophils suggest that a reaction to the polymer may have caused late stent thrombosis. Careful long-term follow-up of patients with vessel enlargement after Cypher stent placement is recommended.
Key Words: sirolimus stents thrombosis death, sudden
| Introduction |
|---|
|
|
|---|
This report describes the first case of fatal acute myocardial infarction and cardiac rupture as a result of late thrombosis of a Cypher stent deployed 18 months previously. The patient, enrolled in the European Sirolimus Eluting Stent in De Novo Native Coronary Lesions (E-SIRIUS) trial, developed a hypersensitivity reaction limited to the coronary artery wall surrounding the stent.
| Case Report |
|---|
|
|
|---|
|
Three weeks later, the patient presented to the clinic with a skin rash on the trunk, neck, ankle, and wrist with itching and irritation; there was no previous history of allergy. The reaction was interpreted as an allergic response to ticlopidine, which was discontinued and replaced by clopidogrel 75 mg/d for 2 months. At this time, the total leukocyte count was normal, without eosinophilia. The rash resolved within a few days.
At 8 months after stenting, per protocol, the patient underwent follow-up angiography and IVUS. Angiographic quantitative coronary arteriography and IVUS imaging demonstrated an absence of in-stent restenosis or neointimal proliferation (Figures 1 and 2
). Laboratory blood tests were normal, including a total eosinophil count of 140 µL. The patient remained asymptomatic at 1-year clinical follow-up, with a negative isotope stress test.
|
Eighteen months after stenting, the patient developed an episode of epigastric and retrosternal chest pain with syncope followed by prompt recovery. Over the next few days, the individual experienced recurrent episodes of intermittent chest pain and was admitted to the coronary care unit with a diagnosis of a recent nonQ-wave myocardial infarction (peak creatine kinase of 423 U/L and a troponin I of 34 ng/mL). The patient was now asymptomatic and was treated with heparin, ß-blockers, aspirin, and intravenous nitroglycerin. There was no fever, and the leukocyte count was normal, without evidence of eosinophilia. Coronary angiography 8 days after the onset of chest pain showed occlusion of the LCx at the entrance of the proximal Cypher stent, with TIMI grade 1 flow. In addition, there was progressive stenosis of lesions in the mid LAD and proximal RCA. A careful attempt to cross the LCx stenosis with a floppy guidewire was unsuccessful (Figure 1). Soon thereafter, the patient experienced a sudden profound hypotension immediately followed by cardiorespiratory arrest and pulseless electrical activity; resuscitation attempts were unsuccessful.
An autopsy limited to the heart and brain was performed. A 300-mL hemorrhagic pericardial effusion was present associated with a ruptured acute transmural basal lateral myocardial infarction. The coronary arteries were dissected free from the heart and radiographed (Figure 3, A and B). The 2 sequential Cypher stents were observed in close apposition but without overlap (Figure 3B). The adventitial surface of the LCx artery in the region of the distal stent was hemorrhagic, and the proximal stented segment was dilated. The stented region of the LCx was processed for methylmethacrylate embedding and sectioned at 3-mm intervals for a total of 12 sections. The remainder of the epicardial coronary arteries were sectioned at 3- to 4-mm intervals and embedded in paraffin.
|
The stented LCx showed an occlusive luminal thrombus, starting at the mouth of the proximal stent and partially obstructing the distal stent (Figure 3, C through F). The wall of the stented artery was aneurysmally dilated with an extensive inflammatory infiltrate involving the intima, media, and adventitia consisting of lymphocytes, plasma cells, macrophages, and eosinophils. The luminal surface of the proximal stent was surrounded by fibrin-rich thrombus with sparse smooth muscle cells, whereas the distal stent showed an extensive inflammatory infiltrate consisting predominantly of lymphocytes and eosinophils and occasional giant cells (Figure 3, G through J). The abluminal surfaces of the proximal and distal stent were focally malapposed, with thick layers of fibrin thrombus separating the stent from the underlying plaque and arterial wall (Figure 3, D and F). The proximal stented artery also showed focal giant cell reaction surrounding a few polymer remnants that had become separated from the stent struts (Figure 3K). Although no medial necrosis or neutrophil infiltrates were observed in any of the sections, the inflammatory cells diffusely infiltrated the media, causing medial disruption and destruction.
Immunostains for inflammatory cells in the intima, media, and adventitia revealed dominant T-lymphocyte infiltration (CD45Ro) with scattered B lymphocytes (CD20) and less numerous macrophages (CD68) (Figure 3, M through O). Luna stains showed an extensive eosinophilic infiltration, especially prominent in the distal stent in the adventitia, media, and intima around stent struts (Figure 3, K and L). Stains for bacteria and fungi were negative. Collectively, these pathological changes are consistent with a localized hypersensitivity reaction.
Severe arterial stenoses were found in the nonstented lesions of the LAD and RCA; however, no significant inflammation or eosinophilic infiltrates were present.
| Discussion |
|---|
|
|
|---|
Hypersensitivity to metals such as molybdenum, nickel, and chromium has been reported in 10% of patients undergoing stenting.10 To the best of our knowledge, hypersensitivity to bare stainless steel stents has been associated with restenosis and not thrombosis, and a late eosinophil-rich infiltrate has not been reported in human stented arteries. We have examined >400 stainless steel stents placed in human coronary arteries, and of these,
150 have been in place >3 months; to date, none of the latter group show an eosinophilic reaction near the stent.11 In bare metal stents, inflammation around stent struts typically consists of macrophages and T lymphocytes, with a few B lymphocytes and giant cells. Moreover, the extent of inflammation correlates with the presence of restenosis, whereas in the present case, there was little to no underlying neointima.11 Remarkably, adventitial eosinophilic infiltrates are found primarily in patients with spontaneous coronary dissection. It is usually an acute phenomenon occurring soon after dissection of the artery; healed spontaneous dissections typically do not show inflammation.
We recently reported pathological findings of late stent thrombosis
30 days with bare metal stents.12 The mechanism of late thrombosis was stenting across a major branch ostium with either single or bifurcating stents, brachytherapy, plaque rupture just proximal or distal to the stent, extensive necrotic core prolapse, and in-stent restenosis with thrombosis. In these cases, there was no evidence of excessive inflammation of the arterial wall or infiltration of eosinophils. Incomplete neointimal healing was present in 12 of 13 cases with late thrombosis. In the present case, the marked hypersensitivity reaction clearly prevented arterial healing and was probably not exaggerated during the interval after the onset of chest pain and death, because this is not observed in the arterial wall in fatal cases of plaque rupture.
The hypersensitivity in the present patient was unlikely to have been caused by sirolimus, because pharmacokinetic studies performed in dogs and rabbits show that the drug is undetectable in the arterial wall by 60 days.2,13 Despite widespread use of oral sirolimus for transplant rejection (kidney and heart), we know of only 1 case report of leukoclastic vasculitis to sirolimus.14 Conversely, sirolimus has been shown to suppress eosinophilic infiltration in an animal model of bronchial hypersensitivity.15 Adverse side effects to sirolimus are limited primarily to bone marrow suppression and hypercholesterolemia and hypertriglyceridemia.16 Other reported side effects include hypokalemia, hyperglycemia, diarrhea, and abnormal liver function.17
Polymers like those in latex and vinyl gloves, methylmethacrylate in dentistry, and polyurethane, among others, have been shown to produce hypersensitivity reactions in some individuals.1820 However, many nonbioerodable polymers [polyurethane, poly(dimethyl)siloxane (silicone), and polyethylene terephthalate (Dacron)] are known to promote inflammation when implanted in swine coronary arteries.2123 Most inflammatory reactions to polymers involve macrophages and multinucleated giant cells and lymphocytes without eosinophils. In studies unrelated to stents, poly-n-butyl methacrylate, a component of bone cement and the same polymer coating as used in Cypher stents, when implanted subcutaneously promotes a macrophage and giant cell reaction accompanied by tissue damage and fibrosis.24 In addition, the polyethylenevinyl acetate component of the Cypher copolymer, when used as an antigen-delivery matrix, has been shown to elicit inflammation in 25% of rabbits.25
Although there are no reports of excessive inflammation with Cypher stents in animals, our laboratory has observed inflammatory reactions in the arterial walls of pig coronary arteries with 28- and 90-day stainless steel26 and DESs (including Cypher, R.V., unpublished data, October 2003). Granulomas occur in 10% to 20% of implanted porcine coronary stents and involve either a few or multiple stent struts. Inflammatory infiltrates typically consist of lymphocytes, macrophages, giant cells, neutrophils, and eosinophils. In stainless steel stents, inflammation is usually much less pronounced at 90 days than at 28 days, whereas the reciprocal effect is found in DESs with nonerodable polymers. In DESs, there are often greater numbers of eosinophils, localized primarily to stent struts, which may extend into the lumen and adjoining media and adventitia. The inflammation is usually accompanied by excessive neointimal thickening, and thrombosis is infrequent.
The hypersensitivity to the Cypher stents was most likely present at 8 months after deployment, because significant vessel enlargement and persistent positive remodeling was found by IVUS. Furthermore, there was progressive enlargement of the artery, with aneurysm formation between 8 and 18 months. The significant malapposition noted in both proximal and distal stents probably occurred in the last 10 months of life; only focal late malapposition was observed at 8 months. In the RAVEL trial, late stent malapposition was found in 21% of Cypher stent implants, compared with 4% in control BX Velocity stents27; 1 patient with a Cypher stent developed an aneurysm.28 Lower rates of malapposition were reported in the SIRIUS trial (9% in Cypher versus 0% in bare metal stents) without any clinical events present at 18 months.29 A regional increase in external elastic membrane remodeling may represent the main cause of late stent malapposition.30 Although the precise mechanism(s) of late stent malapposition and thrombosis in our case are unknown, inflammatory destruction of the medial wall may have caused further dilation accompanied by the accumulation of fibrin between stent struts and native plaque, suggesting a relationship between malapposition, inflammation, and thrombosis.
In October of 2003, a US Food and Drug Administration Public Health Web Health Alert notified physicians of possible generalized hypersensitivity reactions in 50 patients receiving Cypher stents.3 The symptoms included pain, rash, respiratory alterations, hives, itching, fever, and blood pressure changes. None of these findings were documented in our patient at the time of death, and the hypersensitivity reaction was localized to the stented arterial segment at 18 months, when the drug is completely eluted off the stent. In this case, the polymer may serve as an antigenic stimulus resulting in a T-lymphocyte and/or foreign body response accompanied by eosinophils. Activated T lymphocytes, specifically CD4 helper cells, may secrete interleukins 4 and 13, causing an allergic response with eosinophilic infiltration.31
There is a likely spectrum of allergic responses to DESs in sensitive patients, varying from benign reactions to excessive inflammation with medial destruction, stent malapposition, and aneurysm formation with late in-stent thrombosis. The enhanced surveillance of patients with DES for late complications along with the development of tests to prescreen individuals with potential reactions to polymers may help avoid some of the late-term complications with DES. Continued efforts to improve polymer biocompatibility are warranted.
| Footnotes |
|---|
The opinions or assertions contained herein are the private views of the authors and are not to be construed as official or as reflecting the views of the Department of the Army or the Department of Defense.
This article originally appeared Online on January 26, 2004 (Circulation. 2004;109:r38r42).
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
L. T. Newsome, M. A. Kutcher, and R. L. Royster Coronary Artery Stents: Part I. Evolution of Percutaneous Coronary Intervention Anesth. Analg., August 1, 2008; 107(2): 552 - 569. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. T. Newsome, R. S. Weller, J. C. Gerancher, M. A. Kutcher, and R. L. Royster Coronary Artery Stents: II. Perioperative Considerations and Management Anesth. Analg., August 1, 2008; 107(2): 570 - 590. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Sullivan and N. Amarshi Dual antiplatelet therapy with clopidogrel and aspirin Am. J. Health Syst. Pharm., June 15, 2008; 65(12): 1134 - 1143. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Mehilli, R. A. Byrne, A. Wieczorek, R. Iijima, S. Schulz, O. Bruskina, J. Pache, R. Wessely, A. Schomig, A. Kastrati, et al. Randomized trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis Eur. Heart J., June 11, 2008; (2008) ehn253v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. I. Hamilos, M. Ostojic, B. Beleslin, D. Sagic, L. Mangovski, S. Stojkovic, M. Nedeljkovic, D. Orlic, B. Milosavljevic, D. Topic, et al. Differential effects of drug-eluting stents on local endothelium-dependent coronary vasomotion. J. Am. Coll. Cardiol., June 3, 2008; 51(22): 2123 - 2129. [Abstract] [Full Text] [PDF] |
||||
![]() |
B X Chen, F Y Ma, W Luo, J H Ruan, W L Xie, X Z Zhao, S H Sun, X M Guo, F Wang, T Tian, et al. Neointimal coverage of bare-metal and sirolimus-eluting stents evaluated with optical coherence tomography Heart, May 1, 2008; 94(5): 566 - 570. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Suzuki, F. Ikeno, T. Koizumi, F. Tio, A. C. Yeung, P. G. Yock, P. J. Fitzgerald, and W. F. Fearon In Vivo Comparison Between Optical Coherence Tomography and Intravascular Ultrasound for Detecting Small Degrees of In-Stent Neointima After Stent Implantation J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 168 - 173. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L. van der Hoeven, S.-S. Liem, J. Dijkstra, S. C. Bergheanu, H. Putter, M. L. Antoni, D. E. Atsma, M. Bootsma, K. Zeppenfeld, J. W. Jukema, et al. Stent Malapposition After Sirolimus-Eluting and Bare-Metal Stent Implantation in Patients with ST-Segment Elevation Myocardial Infarction: Acute and 9-Month Intravascular Ultrasound Results of the MISSION! Intervention Study J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 192 - 201. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Saia, A. Marzocchi, and A. Branzi Review: The safety of drug-eluting stents Therapeutic Advances in Cardiovascular Disease, February 1, 2008; 2(1): 43 - 52. [Abstract] [PDF] |
||||
![]() |
J. Aoki, A. Kirtane, M. B. Leon, and G. Dangas Coronary Artery Aneurysms After Drug-Eluting Stent Implantation J. Am. Coll. Cardiol. Intv., February 1, 2008; 1(1): 14 - 21. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Jabara, N. Chronos, D. Conway, W. Molema, and K. Robinson Evaluation of a Novel Slow-Release Paclitaxel-Eluting Stent With a Bioabsorbable Polymeric Surface Coating J. Am. Coll. Cardiol. Intv., February 1, 2008; 1(1): 81 - 87. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Wilson and J. T. Willerson Myocardial Revascularization with Percutaneous Devices Card. Surg. Adult, January 1, 2008; 3(2008): 573 - 598. [Full Text] |
||||
![]() |
J. Cosgrave, A. Qasim, A. Latib, T. C. Aranzulla, and A. Colombo Very Late Restenosis after Paclitaxel-Eluting Stent Implantation Ann Intern Med, December 18, 2007; 147(12): 885 - 887. [Full Text] [PDF] |
||||
![]() |
K. Egashira, K. Nakano, K. Ohtani, K. Funakoshi, G. Zhao, Y. Ihara, J.-i. Koga, S. Kimura, R. Tominaga, and K. Sunagawa Local Delivery of Anti-Monocyte Chemoattractant Protein-1 by Gene-Eluting Stents Attenuates In-Stent Stenosis in Rabbits and Monkeys Arterioscler. Thromb. Vasc. Biol., December 1, 2007; 27(12): 2563 - 2568. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Doyle, C. S. Rihal, C. J. O'Sullivan, R. J. Lennon, H. J. Wiste, M. Bell, J. Bresnahan, and D. R. Holmes Jr Outcomes of Stent Thrombosis and Restenosis During Extended Follow-Up of Patients Treated With Bare-Metal Coronary Stents Circulation, November 20, 2007; 116(21): 2391 - 2398. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. Wilson, J. E. Polovick, B. A. Huibregtse, and B. C. Poff Overlapping paclitaxel-eluting stents: Long-term effects in a porcine coronary artery model Cardiovasc Res, November 1, 2007; 76(2): 361 - 372. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. Fleisher, J. A. Beckman, K. A. Brown, H. Calkins, E. L. Chaikof, K. E. Fleischmann, W. K. Freeman, J. B. Froehlich, E. K. Kasper, J. R. Kersten, et al. ACC/AHA 2007 Guidelines on Perioperative Cardiovascular Evaluation and Care for Noncardiac Surgery: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 2002 Guidelines on Perioperative Cardiovascular Evaluation for Noncardiac Surgery) Developed in Collaboration With the American Society of Echocardiography, American Society of Nuclear Cardiology, Heart Rhythm Society, Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Interventions, Society for Vascular Medicine and Biology, and Society for Vascular Surgery J. Am. Coll. Cardiol., October 23, 2007; 50(17): e159 - e242. [Full Text] [PDF] |
||||
![]() |
L. A. Fleisher, J. A. Beckman, K. A. Brown, H. Calkins, E. L. Chaikof, K. E. Fleischmann, W. K. Freeman, J. B. Froehlich, E. K. Kasper, J. R. Kersten, et al. ACC/AHA 2007 Guidelines on Perioperative Cardiovascular Evaluation and Care for Noncardiac Surgery: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 2002 Guidelines on Perioperative Cardiovascular Evaluation for Noncardiac Surgery) Circulation, October 23, 2007; 116(17): e418 - e500. [Full Text] [PDF] |
||||
![]() |
A. E. Rodriguez, A. O. Maree, J. Mieres, D. Berrocal, L. Grinfeld, C. Fernandez-Pereira, V. Curotto, A. Rodriguez-Granillo, W. O'Neill, and I. F. Palacios Late loss of early benefit from drug-eluting stents when compared with bare-metal stents and coronary artery bypass surgery: 3 years follow-up of the ERACI III registry Eur. Heart J., September 1, 2007; 28(17): 2118 - 2125. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Daemen and P. W. Serruys Drug-Eluting Stent Update 2007: Part II: Unsettled Issues Circulation, August 21, 2007; 116(8): 961 - 968. [Full Text] [PDF] |
||||
![]() |
P. Jimenez-Quevedo, M. Sabate, D. J. Angiolillo, F. Alfonso, R. Hernandez-Antolin, M. SanMartin, J. A. Gomez-Hospital, C. Banuelos, J. Escaned, R. Moreno, et al. Long-term clinical benefit of sirolimus-eluting stent implantation in diabetic patients with de novo coronary stenoses: long-term results of the DIABETES trial Eur. Heart J., August 2, 2007; 28(16): 1946 - 1952. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R Bennett Vascular pathology as a result of drug-eluting stents Heart, August 1, 2007; 93(8): 895 - 896. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. M M Pires, D. Eefting, M. R de Vries, P. H A Quax, and J W. Jukema Sirolimus and paclitaxel provoke different vascular pathological responses after local delivery in a murine model for restenosis on underlying atherosclerotic arteries Heart, August 1, 2007; 93(8): 922 - 927. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S. Santiago, H. Ishii, S. Shafi, R. Khurana, P. Kanellakis, R. Bhindi, M. J. Ramirez, A. Bobik, J. F. Martin, C. N. Chesterman, et al. Yin Yang-1 Inhibits Vascular Smooth Muscle Cell Growth and Intimal Thickening by Repressing p21WAF1/Cip1 Transcription and p21WAF1/Cip1-Cdk4-Cyclin D1 Assembly Circ. Res., July 20, 2007; 101(2): 146 - 155. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Daemen and P. W. Serruys Drug-Eluting Stent Update 2007: Part I: A Survey of Current and Future Generation Drug-Eluting Stents: Meaningful Advances or More of the Same? Circulation, July 17, 2007; 116(3): 316 - 328. [Full Text] [PDF] |
||||
![]() |
R. Jaffe and B. H. Strauss Late and Very Late Thrombosis of Drug-Eluting Stents: Evolving Concepts and Perspectives J. Am. Coll. Cardiol., July 10, 2007; 50(2): 119 - 127. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Holmes Jr, D. J. Kereiakes, W. K. Laskey, A. Colombo, S. G. Ellis, T. D. Henry, J. J. Popma, P. W.J.C. Serruys, T. Kimura, D. O. Williams, et al. Thrombosis and Drug-Eluting Stents: An Objective Appraisal J. Am. Coll. Cardiol., July 10, 2007; 50(2): 109 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. V. Finn, G. Nakazawa, M. Joner, F. D. Kolodgie, E. K. Mont, H. K. Gold, and R. Virmani Vascular Responses to Drug Eluting Stents: Importance of Delayed Healing Arterioscler. Thromb. Vasc. Biol., July 1, 2007; 27(7): 1500 - 1510. [Abstract] [Full Text] [PDF] |
||||
![]() |