(Circulation. 2004;110:2591-2596.)
© 2004 American Heart Association, Inc.
Arrhythmia/Electrophysiology |
From the Vanderbilt University Medical Center, Nashville, Tenn (M.S.W.); Medtronic, Inc, Minneapolis, Minn (P.J.D., A.J.S., M.F.O.); Brigham and Womens Hospital, Boston, Mass (M.O.S.); Portsmouth Cardiology, Portsmouth, Va (W.O.A.); Texas Cardiac Arrhythmia Foundation, Austin, Tex (R.C.C.); Easton Hospital, Easton, Pa (K.K.); Providence Hospital, Southfield, Mich (C.M.); Arrhythmia Consultants, Greenville, SC (D.S.R.); and University of Pennsylvania, Philadelphia (K.J.V.).
Correspondence to Mark Wathen, MD, Vanderbilt Page-Campbell Heart Institute, Vanderbilt University Medical Center, 2220 Pierce, Nashville, TN 37232. E-mail mark.wathen{at}vanderbilt.edu
Received March 25, 2004; revision received June 15, 2004; accepted June 18, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results We randomized 634 ICD patients to 2 armsstandardized empirical ATP (n=313) or shock (n=321)for initial therapy of spontaneous FVT. ICDs were programmed to detect FVT when 18 of 24 intervals were 188 to 250 bpm and 0 of the last 8 intervals were >250 bpm. Initial FVT therapy was ATP (8 pulses, 88% of FVT cycle length) or shock at 10 J above the defibrillation threshold. Syncope and arrhythmic symptoms were collected through patient diaries and interviews. In 11±3 months of follow-up, 431 episodes of FVT occurred in 98 patients, representing 32% of ventricular tachyarrhythmias and 76% of those that would be detected as ventricular fibrillation and shocked with traditional ICD programming. ATP was effective in 229 of 284 episodes in the ATP arm (81%, 72% adjusted). Acceleration, episode duration, syncope, and sudden death were similar between arms. Quality of life, measured with the SF-36, improved in patients with FVT in both arms but more so in the ATP arm.
Conclusions Compared with shocks, empirical ATP for FVT is highly effective, is equally safe, and improves quality of life. ATP may be the preferred FVT therapy in most ICD patients.
Key Words: defibrillators, implantable pacing tachyarrhythmias
| Introduction |
|---|
|
|
|---|
Numerous studies have demonstrated that ATP terminates 78% to 94% of ventricular tachycardias (VTs) <188 to 200 bpm with a 2% to 4% risk of acceleration.69 Accordingly, ATP has conventionally been applied to only slower, presumably hemodynamically tolerated VTs. Fast VT (FVT, 188 to 250 bpm) is typically programmed to receive painful shock therapy even though ATP might be successful. More recently, high success and low acceleration rates for FVT were demonstrated in coronary artery disease patients10,11 but in a nonrandomized fashion, so a comparison with shocks for safety and QoL was not performed. This trial examined the efficacy, safety, and QoL associated with treatment of FVT with ATP compared with shocks in a broad ICD population.
| Methods |
|---|
|
|
|---|
Patient Selection
Enrollment followed institutional review board acceptance of the protocol and informed consent in patients with standard ICD indications from January 2001 through March 2002 at 42 US centers. Inclusion criteria were broad, excluding only ICD patients believed unlikely to have substrate for stable monomorphic VT susceptible to ATP (hypertrophic cardiomyopathy, long-QT syndrome, or Brugada syndrome). Transvenous defibrillation leads were positioned at the right ventricular apex. Defibrillation thresholds (DFTs) were determined by use of a binary search protocol with a minimum 10-J DFT safety margin required for enrollment. Patients who met enrollment and implantation criteria were randomized to either ATP or shock as the first therapy for FVT. All patients were seen every 3 months for 1 year. Patient diaries and interviews were used to capture symptoms possibly related to spontaneous ventricular arrhythmias (eg, palpitations, lightheadedness, dizziness, near syncope, or syncope). Syncope required complete loss of consciousness with loss of postural tone, and near syncope was defined as severe dizziness or lightheadedness without complete loss of consciousness. Symptoms were correlated with ICD-stored episode data of spontaneous ventricular arrhythmias. Data were transferred to a central database for analysis.
Device Description and Programming
Device programming was standardized except for the initial randomized therapy for FVT. Detection in the ventricular fibrillation (VF) zone required that 18 of the last 24 R-R intervals had a cycle length (CL) <320 ms (>188 bpm). Only ICDs capable of programming a FVT detection zone defined within the VF zone (FVT via VF) for a CL of 240 to 320 ms (250 to 188 bpm) were used (Medtronic models 7227, 7229, 7231, 7271, 7273, and 7275). When any one of the final 8 R-R intervals preceding the moment of detection was <240 ms (>250 bpm), the episode was classified as VF and received a DFT+10-J shock followed by maximal shocks if necessary. When all of the last 8 R-R intervals were
240 ms (
250 bpm), the episode was detected as FVT. In the ATP arm, the first therapy in the FVT zone was a single ATP sequence (8-pulse burst pacing train at 88% of the VT CL). Failed ATP was followed by shock at DFT+10 J and then maximal shocks as necessary. In the shock arm, first therapy was DFT+10 J followed by maximal shocks. A VT zone with a CL of 320 to
360 ms (
167 to 188 bpm) was programmed in all patients. ATP in the VT zone was programmed to 3 sequences of 8 pulses at 88% of the VT CL with 20-ms decrement between sequences. Supraventricular tachycardia (SVT) discrimination was programmed "on" in the VT zone of all dual-chamber ICDs with the SVT limit of 320 ms. All devices were programmed to store the far-field electrogram (EGM) before the onset of detected episodes to aid in rhythm classification.
Rhythm Classification and Definitions
The principal investigator at each site reviewed and classified the detected rhythm of all spontaneous episodes of VT with stored EGM. Each episode was also reviewed by one of 5 electrophysiologists on an episode review committee. When both reviewers agreed on both the detected rhythm and final rhythm after therapy, the episode received no further review. On any disagreement, the episode was reviewed by a different committee member. Classification was taken on agreement by any 2 physicians. All other episodes were classified by a joint committee review with majority rule applied.
Arrhythmia CL was reported as the median of the final 12 intervals preceding detection. Episode duration was measured from the first tachycardia beat in the detection window to the first nontachycardic return beat. Acceleration was defined as
10% CL reduction after therapy.
QoL Assessment
Self-reported health-related QoL was measured at baseline and at 12 months with the Medical Outcomes Study 36-item Short-Form General Health Survey (SF-36) that included 8 subscales and 2 summary measurements.12 Scores ranged from 0 to 100, with higher scores representing better QoL. The change in score from baseline to 12 months was compared between the ATP and shock arms for each scale.
Statistical Analysis
Data were analyzed on an intention-to-treat basis. To adjust for multiple episodes per patient, the generalized estimating equations method was used in calculations of therapy delivery, syncope, acceleration, and episode duration estimates and comparisons unless otherwise noted.13 Baseline QoL and change in QoL over time were compared between randomization groups by use of a Wilcoxon test. Paired nonparametric exact methods were used to compare each patients change in QoL over time. Mortality rate was determined by Kaplan-Meier estimation. Summary statistics for episode measures such as CL were adjusted for multiple episodes per patient by calculating the median for each patient and then calculating summary statistics on the basis of each patients median. Statistical analyses were performed with SAS (version 8).
| Results |
|---|
|
|
|---|
|
Spontaneous Episodes Detected
Among 4230 spontaneous episodes retrieved from ICD counters that the device detected as ventricular tachyarrhythmia, 1837 had complete electrogram data and were included in the analysis. Missing data were due to ICD memory limitations, incomplete interrogation, or cleared memory. The proportion of patients with missing EGMs for spontaneous episodes was similar between treatment arms. All reported episodes of syncope and death are included in the analysis.
Of the 1837 spontaneous episodes with stored EGMs, 491 (27%) were determined to be inappropriately detected SVT, and 4 (0.2%) were nonphysiological artifact. Thus, 1342 of 1837 detected episodes (73%) were true ventricular arrhythmia and were subjected to analysis. Of these, 777 (58%) were detected as VT, 431 (32%) as FVT, and 134 (10%) as VF. Figure 1 shows the CL distribution of rhythms according to detection zone.
|
The ATP arm had 284 FVT episodes in 47 patients compared with 147 episodes in 51 patients in the shock arm. Two ATP patients had 56 and 75 episodes, accounting for the large difference in episode numbers between arms. There was no difference in median FVT CL between treatment groups (290 ms for ATP versus 300 ms for shock; P=0.10). A total of 13 polymorphic VT episodes were detected as FVT (ATP, n=11; shock, n=2). Four episodes in the shock arm were terminated by ATP in crossover patients.
Therapy Efficacy
The shock arm had 147 detected FVT episodes, yet only 99 episodes (64%) were treated by 110 shocks (Figure 2). The reduction in shocked episodes was due to spontaneous termination during the 3.3 seconds (median) of capacitor charging (44 episodes, 34%) and 4 episodes in crossover patients that were terminated by ATP.
|
In the ATP arm, the first ATP attempt terminated 229 of 284 FVT episodes (81% unadjusted, 72% adjusted; 95% CI, 60 to 81), including 8 in which VT reinitiated so quickly that the device assumed that the ATP had failed and advanced therapy to a shock. ATP was not attempted in 1 episode because of crossover. Of the 54 episodes in which ATP failed, 49 episodes were shocked, whereas 5 were terminated by additional ATP. In all, 57 episodes (31%) received 62 shocks. Within individuals, ATP success was 100% in 26 patients and 0% in 7 patients and ranged from 25% to 89% (unadjusted) in the remaining 14 patients. Thus, 40 of 47 patients (85%) presumably had at least 1 FVT shock prevented by ATP. In patients who had ATP failure of the first FVT episode, the efficacy of ATP overall was 45% (95% CI, 22 to 70).
Figure 3 shows ATP efficacy for FVT by detected CL. The ATP success rate was similar for FVT with CL between >280 to 320 ms and 240 to 280 ms (76% versus 62%; P=0.38).
|
Clinical Safety of ATP for FVT
Relative safety of ATP versus shocks was assessed by comparing the FVT episode duration, incidence of acceleration, syncope, first shock efficacy, and sudden death between treatment arms (Table 2).
|
The median FVT episode duration was 10.0 seconds in the ATP arm (mean±SE, 12.7±0.8 seconds) and 9.7 seconds in the shock arm (mean±SE, 10.7±0.7 seconds). This included failed therapies and spontaneously terminating shock episodes. Thus, the primary objective of the trial was satisfied (P<0.001).
Acceleration of FVT was similarly low between treatment groups. Acceleration occurred in 4 of 273 monomorphic VT episodes (2%) in the ATP arm versus 2 of 145 (1%) in the shock arm. No FVT acceleration required >1 shock to terminate.
Syncope during FVT was also similarly low between treatment groups. There were 3 episodes of syncope during treatment for FVT (ATP, n=2; shock, n=1). The FVT episodes associated with syncope were short (15, 10, and 9 seconds) and rare, so the effect of episode duration on risk of syncope could not be measured. Presyncope or milder symptoms with FVT were reported for 23 of 284 (13%) in the ATP arm and 15 of 147 (10%) in the shock arm. The proportion of FVT episodes with symptoms was 38% in the ATP arm and 66% in the shock arm (P<0.01). Other syncope included 4 VF, 1 SVT, and 26 nonarrhythmic syncopal episodes, none associated with missing ICD data.
First shock success was identical (92%) whether it was administered after a failed ATP attempt (ATP arm) or de novo (shock arm).
Fifty-six patients (9%) died during the study. There were 32 deaths (10%) in the ATP arm and 24 (7%) in the shock arm (P=0.22). The cumulative 12-month survival probability for all-cause mortality was 91% (95% CI, 89 to 93). An independent committee classified the cause of death, with nonsudden cardiac death occurring in 17 patients (5%) in the ATP arm and 15 (5%) in the shock arm. Sudden cardiac death occurred in 1 patient (0.3%) in the ATP arm and 2 patients (0.6%) in the shock arm and was too infrequent for statistical analysis. The 2 shock arm patients died of congestive heart failure and electromechanical dissociation unrelated to shocks. The cause of the ATP arm sudden cardiac death was unknown. This patients death in an intensive care unit was immediately preceded by sinus rhythm. However, 16 to 30 minutes before death, the patient had a successful shock for FVT accelerated by ATP with 2 additional VT episodes terminated by ATP.
Quality of Life
Of 131 patients with episodes detected in the FVT zone (including inappropriate detections), 98 completed SF-36 measurements at baseline and 12 months (ATP, 43 of 63; shock, 55 of 68). Baseline values were similar (P
0.05) between arms for each subscale and summary score. The shock arm had significantly improved bodily pain scores at 12 months without improvement in the other 9 measures. The ATP arm had significant improvement in 5 subscalesphysical functioning, role physical, bodily pain, social functioning, and role emotionaland the 2 summaries, mental and physical. None of the subscale measurements were significantly reduced at 12 months in either arm. Figure 4 compares the change in scores at 12 months. The ATP group had statistically greater improvement in physical functioning (P=0.02), role physical (P=0.04), and mental summary (P=0.03), whereas social functioning was of borderline significance (P=0.053).
|
Factors Predictive of ATP Efficacy
Univariate analyses were performed on selected patient variables to determine their correlation with ATP success. Only left ventricular ejection fraction provided marginally significant prediction; with every 5% increase in left ventricular ejection fraction, the odds of successful ATP for FVT are 18% higher (P=0.06). Other variables that were not predictive of ATP success include prior history of non-sustained VT (NSVT), presence of coronary artery disease, and indication for implant. ATP success was 67% (185 of 232 episodes) in patients with coronary artery disease and 83% (44/52 episodes) in patients without coronary artery disease (P=0.16). Patients with primary and secondary implant indications had comparable ATP success, 74% (107 of 130 episodes) and 71% (122 of 154 episodes), respectively (P=0.77).
| Discussion |
|---|
|
|
|---|
Early studies testing ATP for FVT induced in the electrophysiology laboratory had low success, perhaps in part because they were treating induced rather than spontaneous VT.1417 Not only was the success rate low (41% to 68%), but acceleration rates were high (5% to 55%). Other investigations tested ATP for spontaneous FVT in ICDs and found 47% to 79% success with acceleration of 3% to 10%.911 However, none of these trials compared ATP with shock therapy. Thus, the safety of ATP programming was not established until this trial. Concern over safety has rightfully been the limiting factor for using ATP for FVT. However, we found no safety differences between shock and ATP as initial therapy for FVT.
Our study used an ATP regimen of 88% of VT CL, fairly nonaggressive ATP supported by Peinado et al9 that perhaps led to our high efficacy and low acceleration. Given that ATP was programmed empirically, the results also support prior observations that "tailoring" based on electrophysiology study of induced VTs is not necessary in most situations.10
No clinical predictors of ATP success could be found. Diminished left ventricular ejection fraction slightly reduced efficacy but reached only marginal statistical significance. Although VTs with CL of 240 to 280 ms trended toward lower ATP success, they still had nearly two thirds of their FVT safely terminated by ATP. ATP was equally effective in patients without coronary artery disease, a group excluded from the pilot study, PainFREE Rx.11
The greatest predictor of ATP success in PainFREE Rx was NSVT.11 One may infer that the device was claiming ATP success for treating nonsustained FVT episodes that were destined to terminate regardless of therapy, a reasonable conclusion given that VF plus FVT detection used 12 of 16 intervals. Consequently, this trial used 18 of 24 intervals for VF plus FVT detection and found that NSVT was not a predictor of ATP success. The fact that 34% of detected FVT episodes in the shock arm terminated before shock suggests that a longer delay will further reduce unnecessary device detections. It is intuitive that increased duration of tachycardia might increase syncope. However, delaying detection to 18 of 24 beats proved safe because arrhythmic syncope (8 of 1837, 0.5%) did not increase compared with PainFREE Rx (13 of 1248, 2.0%).11 Multiple factors determine the hemodynamic response to VT, likely explaining the heterogeneous response observed among patients in clinical studies.18 In this study, the 3 syncopal FVT episodes lasted 15, 10, and 9 seconds, whereas episodes in other patients were tolerated up to
30 seconds without syncope. Most importantly, the additional time required for capacitor charging and delivery of definitive shock therapy did not appear to increase the probability of syncope after failed ATP for FVT. Furthermore, the fast charge time of the newly implanted ICDs in this study may have created a bias favoring shocks that would not be present in other ICDs later in their lifecycles, implying that ATP may have an advantage not shown by these results. This may be especially true for todays ICDs with short capacitor charge times and the ability to deliver ATP while charging.
ICDs have become a mainstay of therapy because they deliver immediate and nearly absolute therapy of the leading cause of death in cardiovascular patients. Shocks are associated with significant morbidity contributing to anxiety syndrome in 15% to 38% of ICD patients.15 Despite differences in the literature between study designs and clinical outcomes, the negative impact of ICD shocks on QoL was consistent between studies and was related to the increased number of shocks received.3,4 However, whether reduction of shocks by applying ATP therapy improves QoL has not been tested previously. This is the first trial to demonstrate that ATP does improve QoL compared with shock delivery in patients who experience FVT. In the ATP arm, 5 of 8 subscales showed improvement, whereas only 1 subscale improved in shock arm patients. In agreement with previous reports,14 the summary scales reflect that by virtue of reducing shocks, the main contribution of ATP to improved QoL was mental.
Study Limitations
As with other trials, bias may have been introduced because the analysis could not be performed on episodes of tachycardia without EGM. Analysis was based on 43% of detected arrhythmias, although importantly every episode of syncope and death was included. Reporting only on properly adjudicated episodes increases confidence in data accuracy. Because most ATP episodes are asymptomatic, ATP episodes are potentially underreported in this trial. Speculatively, underreporting of successful ATP episodes may bias the results toward reduced ATP success.
Like other trials assessing syncope, this trial relied on patient diaries to document syncopal episodes. Although a concerted effort was made to collect unbiased syncope data, the occurrence of syncopal episodes may still be underestimated. However, missing syncope data should be independent of FVT therapy, so safety comparisons should be unaffected.
Conclusions
Empirical ATP is highly effective in treating FVT. It prevents painful shocks without any clinical difference in episode duration, arrhythmic syncope, acceleration, or sudden death while yielding improvement in QoL. The investigators of the PainFREE Rx II trial recommend ATP as the preferred therapy for FVT in most ICD patients.
| Appendix |
|---|
|
|
|---|
| Acknowledgments |
|---|
Disclosure
Mark Wathen has received research support from Medtronic, Guidant, and St. Jude; has received honoraria for speaking engagements for Medtronic, Guidant, and St. Jude; and has received consulting fees or other remuneration from Medtronic and Guidant. Paul J. DeGroot, Alice J. Stark, and Mary F. Otterness are employed by Medtronic. Michael O. Sweeney has received consulting fees or other remuneration from Medtronic. Robert Canby has received research support and consulting fees or other remuneration from Medtronic, Guidant, St. Jude, Endocardial Solutions, Boston Scientific, Impulse Dynamics, the National Institutes of Health, and Stereotaxis. Christian Machado has received consulting fees or other remuneration from Medtronic and Guidant and has received honoraria for speaking engagements for Biotronik. Kent J. Volosin has received research support from Medtronic, Guidant, and St. Jude.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
G. Boriani, E. Occhetta, S. Cesario, S. Grossi, M. Marconi, G. Speca, P. Silvestri, M. Biffi, M. Bortnik, C. Martignani, et al. Contribution of morphology discrimination algorithm for improving rhythm discrimination in slow and fast ventricular tachycardia zones in dual-chamber implantable cardioverter-defibrillators Europace, June 9, 2008; (2008) eun146v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62. [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons Circulation, May 27, 2008; 117(21): e350 - e408. [Full Text] [PDF] |
||||
![]() |
M. H. Raitt Implantable Cardioverter-Defibrillator Shocks: A Double-Edged Sword? J. Am. Coll. Cardiol., April 8, 2008; 51(14): 1366 - 1368. [Full Text] [PDF] |
||||
![]() |
R. Passman, H. Subacius, B. Ruo, A. Schaechter, A. Howard, S. F. Sears, and A. Kadish Implantable Cardioverter Defibrillators and Quality of Life: Results From the Defibrillators in Nonischemic Cardiomyopathy Treatment Evaluation Study Arch Intern Med, November 12, 2007; 167(20): 2226 - 2232. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-C. Yap, J. W. Roos-Hesselink, E. S. Hoendermis, W. Budts, H. W. Vliegen, B. J.M. Mulder, A. P.J. van Dijk, M. J. Schalij, and W. Drenthen Outcome of implantable cardioverter defibrillators in adults with congenital heart disease: a multi-centre study Eur. Heart J., August 1, 2007; 28(15): 1854 - 1861. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Passman and A. Kadish Sudden Death Prevention With Implantable Devices Circulation, July 31, 2007; 116(5): 561 - 571. [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, C. A. Baessler, A. B. Curtis, N.A. M. Estes III, B. J. Gersh, B. Grubb, and L. B. Mitchell Addendum to "Personal and Public Safety Issues Related to Arrhythmias That May Affect Consciousness: Implications for Regulation and Physician Recommendations: A Medical/ Scientific Statement From the American Heart Association and the North American Society of Pacing and Electrophysiology" Public Safety Issues in Patients With Implantable Defibrillators A Scientific Statement From the American Heart Association and the Heart Rhythm Society Circulation, March 6, 2007; 115(9): 1170 - 1176. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Gehi, D. Mehta, and J. A. Gomes Evaluation and Management of Patients After Implantable Cardioverter-Defibrillator Shock JAMA, December 20, 2006; 296(23): 2839 - 2847. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. L. Wilkoff, K. T. Ousdigian, L. D. Sterns, Z. J. Wang, R. D. Wilson, J. M. Morgan, and for the EMPIRIC Trial Investigators A Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators: Results From the Prospective Randomized Multicenter EMPIRIC Trial J. Am. Coll. Cardiol., July 18, 2006; 48(2): 330 - 339. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Ripplinger, V. I. Krinsky, V. P. Nikolski, and I. R. Efimov Mechanisms of unpinning and termination of ventricular tachycardia Am J Physiol Heart Circ Physiol, July 1, 2006; 291(1): H184 - H192. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. A. Friedman, R. L. McClelland, W. R. Bamlet, H. Acosta, D. Kessler, T. M. Munger, N. G. Kavesh, M. Wood, E. Daoud, A. Massumi, et al. Dual-Chamber Versus Single-Chamber Detection Enhancements for Implantable Defibrillator Rhythm Diagnosis: The Detect Supraventricular Tachycardia Study Circulation, June 27, 2006; 113(25): 2871 - 2879. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Goldberger and R. Lampert Implantable Cardioverter-Defibrillators: Expanding Indications and Technologies JAMA, February 15, 2006; 295(7): 809 - 818. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Connolly Use and Misuse of Surrogate Outcomes in Arrhythmia Trials Circulation, February 14, 2006; 113(6): 764 - 766. [Full Text] [PDF] |
||||
![]() |
K. A. Ellenbogen, J. H. Levine, R. D. Berger, J. P. Daubert, S. L. Winters, E. Greenstein, A. Shalaby, A. Schaechter, H. Subacius, A. Kadish, et al. Are Implantable Cardioverter Defibrillator Shocks a Surrogate for Sudden Cardiac Death in Patients With Nonischemic Cardiomyopathy? Circulation, February 14, 2006; 113(6): 776 - 782. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Connolly, P. Dorian, R. S. Roberts, M. Gent, S. Bailin, E. S. Fain, K. Thorpe, J. Champagne, M. Talajic, B. Coutu, et al. Comparison of {beta}-Blockers, Amiodarone Plus {beta}-Blockers, or Sotalol for Prevention of Shocks From Implantable Cardioverter Defibrillators: The OPTIC Study: A Randomized Trial JAMA, January 11, 2006; 295(2): 165 - 171. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Sadoul, R. Mletzko, F. Anselme, R. Bowes, W. Schols, C. Kouakam, G. Casteigneau, R. Luise, N. Iscolo, E. Aliot, et al. Incidence and Clinical Relevance of Slow Ventricular Tachycardia in Implantable Cardioverter-Defibrillator Recipients: An International Multicenter Prospective Study Circulation, August 16, 2005; 112(7): 946 - 953. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Buxton, M. O. Sweeney, M. S. Wathen, M. E. Josephson, M. F. Otterness, E. Hogan-Miller, A. J. Stark, P. J. DeGroot, and for the PainFREE Rx II Investigators QRS Duration Does Not Predict Occurrence of Ventricular Tachyarrhythmias in Patients With Implanted Cardioverter-Defibrillators J. Am. Coll. Cardiol., July 19, 2005; 46(2): 310 - 316. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Olshansky Wide QRS, Narrow QRS: What's the Difference? J. Am. Coll. Cardiol., July 19, 2005; 46(2): 317 - 319. [Full Text] [PDF] |
||||
![]() |
S. F. Sears Jr, J. B. Shea, and J. B. Conti How to Respond to an Implantable Cardioverter-Defibrillator Shock Circulation, June 14, 2005; 111(23): e380 - e382. [Full Text] [PDF] |
||||
![]() |
Switch on Antitachycardia Pacing for ICD Patients Journal Watch Cardiology, January 28, 2005; 2005(128): 4 - 4. [Full Text] |
||||
![]() |
W. G. Stevenson, B. R. Chaitman, K. A. Ellenbogen, A. E. Epstein, W. L. Gross, D. L. Hayes, S. A. Strickberger, M. O. Sweeney, and for the Subcommittee on Electrocardiography and Ar Clinical Assessment and Management of Patients With Implanted Cardioverter-Defibrillators Presenting to Nonelectrophysiologists Circulation, December 21, 2004; 110(25): 3866 - 3869. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. R. Hanna and J. J. Langberg The Shocking Story of Azimilide Circulation, December 14, 2004; 110(24): 3624 - 3626. [Full Text] |