Circulation. 2004;110:3430-3434
Published online before print November 22, 2004, doi: 10.1161/01.CIR.0000148371.53174.05
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
110/22/3430    most recent
01.CIR.0000148371.53174.05v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Takebayashi, H.
Right arrow Articles by Moses, J. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Takebayashi, H.
Right arrow Articles by Moses, J. W.
Right arrowPubmed/NCBI databases
*Compound via MeSH
*Substance via MeSH
Related Collections
Right arrow Restenosis

(Circulation. 2004;110:3430-3434.)
© 2004 American Heart Association, Inc.


Coronary Heart Disease

Nonuniform Strut Distribution Correlates With More Neointimal Hyperplasia After Sirolimus-Eluting Stent Implantation

Hideo Takebayashi, MD; Gary S. Mintz, MD; Stéphane G. Carlier, MD, PhD; Yoshio Kobayashi, MD; Kenichi Fujii, MD; Takenori Yasuda, MD; Ricardo A. Costa, MD; Issam Moussa, MD; George D. Dangas, MD, PhD; Roxana Mehran, MD; Alexandra J. Lansky, MD; Edward Kreps, MD; Michael B. Collins, MD; Antonio Colombo, MD; Gregg W. Stone, MD; Martin B. Leon, MD; Jeffrey W. Moses, MD

From the Cardiovascular Research Foundation, Lenox Hill Heart and Vascular Institute, New York, NY.

Correspondence to Stéphane G. Carlier, MD, PhD, Cardiovascular Research Foundation, 55 E 59th St, 5th Floor, New York, NY 10022. E-mail scarlier{at}crf.org

Received June 2, 2004; revision received July 28, 2004; accepted September 9, 2004.


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Background— Little is known about causes of intimal hyperplasia (IH) after sirolimus-eluting stent (SES) implantation.

Methods and Results— Intravascular ultrasound was performed in 24 lesions with intra-SES restenosis and a comparison group of 25 nonrestenotic SESs. To assess stent strut distribution, the maximum interstrut angle was measured with a protractor centered on the stent, and the visible struts were counted and normalized for the number of stent cells. In SES restenosis patients, minimum lumen site was compared with image slices 2.5, 5.0, 7.5, and 10.0 mm proximal and distal to this site. The minimum lumen site had a smaller IVUS lumen area at follow-up (2.7±0.9 versus 6.2±1.9 mm2; P<0.01), larger maximum interstrut angle (135±39° versus 72±23°; P<0.01), larger IH area (3.4±1.5 versus 0.6±1.1 mm2; P<0.01) and thickness (0.7±0.3 versus 0.1±0.2 mm; P<0.01) at maximum interstrut angle, and fewer stent struts (4.9±1.0 versus 6.0±0.5; P<0.01) even when normalized for the number of stent cells (0.78±0.15 versus 0.97±0.07; P<0.01). Compared with nonrestenotic SES, the restenosis lesions also had a smaller minimal lumen area, larger IH area, thicker IH at maximum interstrut angle, fewer stent struts, and larger maximum interstrut angle. Multivariate analysis identified the number of visualized stent struts normalized for the number of stent cells and maximum interstrut angle as the only independent IVUS predictor of IH cross-sectional area (P<0.01 and P<0.01), minimum lumen area (P<0.01 and P<0.01), and IH thickness (P<0.01 and P<0.01).

Conclusions— The number and distribution of stent struts affect the amount of neointima after SES implantation.


Key Words: restenosis • stents • ultrasonics


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Although sirolimus-eluting stents (SESs) strongly suppress neointimal hyperplasia (IH),1,2 in-stent restenosis after SES implantation still occurs.3–5 One in vitro study has shown that nonuniform circumferential stent strut distribution affects local drug concentration.6 It was our hypothesis that the number and distribution of the stent struts might also affect the magnitude of IH after SES implantation in human coronary arteries.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
Study Patients
From the Lenox Hill Hospital clinical and core intravascular ultrasound (IVUS) laboratory databases, we identified 24 nonostial intrastent restenoses (>50% angiographic diameter stenosis) after de novo SES implantation. A consecutive series of 25 SES-treated lesions in asymptomatic patients without restenosis or late malapposition served as a comparison group. Prespecified clinical and laboratory data were obtained from hospital charts. This study was approved by the Institutional Review Board, and written informed consent was obtained from all patients.

Quantitative Coronary Angiography
Angiographic analysis was performed with a computer-assisted, automated edge-detection algorithm (CMS, MEDIS) by an independent observer unaware of the clinical and IVUS findings. Vessel angulation was measured in a nonforeshortened view at end diastole as the angle formed by the centerline through the lumen proximal to the stenosis and opposite a second centerline in the straight portion of the artery distal to the stenosis. The accuracy of this method has been reported in detail.7

IVUS Imaging and Analysis
IVUS imaging was performed after intracoronary administration of 0.1 to 0.2 mg nitroglycerin with motorized transducer pullback and a commercially available scanner (Boston Scientific) incorporating a 40-MHz single-element beveled transducer rotating at 1800 rpm. The ultrasound catheter was advanced >10 mm beyond the stent into the distal vessel. The transducer was withdrawn at a pullback speed of 0.5 mm/s to a point >10 mm proximal to the stent. Images were recorded onto 0.5-in high-resolution super VHS for offline analysis.

Quantitative IVUS analysis was performed with computerized planimetry (TapeMeasure, Indec Systems) by an independent experienced observer who was unaware of the clinical data. Measurements were performed at 9 cross sections: the minimum lumen site and 8 remote sites 2.5, 5.0, 7.5, and 10.0 mm proximal and distal to the minimum lumen site. (The 2.5- and 5.0-mm slices and the 7.5- and 10.0-mm slices were combined for analysis.) For the nonrestenotic (comparison) group, measurements were performed only at the minimum lumen site because of the lack of neointima in these patients. At each cross section, the following were obtained: (1) maximum angle between adjacent stent struts; (2) number of stent struts; (3) stent, lumen, external elastic membrane (EEM), plaque and media (P&M) behind stent (peri-stent P&M equals EEM minus stent), and intimal hyperplasia cross-sectional area (CSA) [IH equals stent minus minimum lumen area (MLA)]; (4) maximum IH and peri-stent P&M thickness; and (5) maximum IH and P&M thickness within the maximum interstrut angle (Figure 1). Stent eccentricity was calculated as minimum divided by maximum stent diameter. The number of visualized stent struts was normalized for (divided by) the number of stent cells: 2.5- and 3.0-mm SESs have 6 cells, and 3.5-mm SESs have 7 cells. To exclude strut fracture, the number of stent struts in 13 patients with postintervention and follow-up IVUS was counted millimeter by millimeter along the stent length at both time points.



View larger version (24K):
[in this window]
[in a new window]
 
Figure 1. Schematic of circumferential distribution of P&M, IH, and struts at follow-up. Schema demonstrates maximum angle between stent struts and measurements of maximum P&M thickness (a) and P&M (b) and IH (c) thickness within largest interstrut angle.

Statistical Analysis
Statistical analysis was performed with StatView 5.0 (SAS Institute). Data are presented as frequencies or mean±SD. Comparison was performed with {chi}2 statistics, unpaired Student’s t test, and ANOVA with post hoc comparison using the Bonferroni correction. Values of P<0.05 were considered statistically significant. Univariate variables with P<0.2 were entered into the multivariate model to find independent predictors of IH CSA, MLA, and IH thickness.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowConclusions
down arrowReferences
 
The interval from SES implantation to follow-up was 6.1±1.8 months (range, 2.0 to 10.6 months). In the restenosis group, 19 of 24 patients had recurrent angina or a positive stress test, and 5 of 24 asymptomatic patients underwent scheduled follow-up. All 25 patients in the comparison group underwent scheduled follow-up.

SES Restenosis
SES restenosis was all focal (≤10 mm). There was no significant correlation between the follow-up interval and IH length (r=0.206, P=0.3), IH CSA (r=0.049, P=0.7), MLA (r=0.044, P=0.8), and IH thickness (r=0.014, P=0.9).

In the 24 SES restenosis lesions, the minimum lumen site was compared with sections 2.5/5.0 and 7.5/10.0 mm proximal and distal to the minimum lumen site (Table 1). The minimum lumen site had a smaller IVUS follow-up MLA, a larger maximum interstrut angle, and a larger follow-up IH CSA and thickness at the maximum interstrut angle.


View this table:
[in this window]
[in a new window]
 
TABLE 1. IVUS Data in 24 Patients With SES Restenosis Comparing the Minimum Lumen Site, Sites 2.5 and 5.0 mm Proximal and Distal to the Minimum Lumen Site, and Sites 7.5 and 10.0 mm Proximal and Distal to the Minimum Lumen Site

To exclude strut fracture (defined as a decrease in number of stent struts between implantation and follow-up in a single cross section), we analyzed 13 SES restenosis patients with IVUS at implantation and follow-up. There was a decrease in the number of struts at the minimum lumen site in only 2 patients. In 1 patient, the number of stent struts decreased from 6 to 3, the maximum interstrut angle increased from 57° to 146°, but the stent CSA did not change (8.4 to 8.6 mm2) (Figure 2). In the second patient, the number of stent struts decreased from 6 to 3, the maximum interstrut angle increased from 62° to 157°, but there was also no change in stent CSA (8.6 mm2 at implantation and follow-up). In the other 11 patients with SES restenosis and both postimplantation and follow-up IVUS studies, the minimum lumen site (compared with the 2.5/5.0- and 7.5/10.0-mm sites) had a smaller follow-up IVUS MLA and a larger IH CSA and thickness at the maximum interstrut angle.



View larger version (139K):
[in this window]
[in a new window]
 
Figure 2. Intrastent restenosis caused by stent fracture after deployment of SES. A, Coronary angiography showed 90% stenosis in middle right coronary artery (RCA). B, SES (3.0x33 mm) was deployed at 18 atm to cover mid RCA (dashed line). C, Follow-up angiography demonstrates 90% stenosis in mid RCA. IVUS images are from 2.5 mm proximally (a, d), minimal lumen site (b, e), and 2.5 mm distally (c, f) at postprocedure (a, b, c) and at follow-up (d, e, f). At minimum lumen site, number of stent struts decreased from 6 (b) to only 3 (e), and maximum interstrut angle increased from 57° to 146°. IH measured 5.6 mm2, but stent CSA did not change (8.4 to 8.6 mm2). On the other hand, number of stent struts (6 to 6), maximum interstrut angle (54° to 52°), and stent CSA (10.1 to 9.8 mm2) did not change proximally, and IH CSA was less (1.4 mm2).

Comparison With Non-SES Restenosis
Table 2 compares the 24 patients with SES restenosis and the group without SES restenosis. The SES restenosis group had longer SESs, higher maximum inflation pressures at implantation, smaller IVUS minimal lumen CSA at follow-up, smaller EEM CSA, less peri-stent P&M CSA, smaller minimal stent CSA, larger IH CSA, fewer visualized stent struts even when normalized for the number of stent cells, and a larger maximum interstrut angle at the minimum lumen site. We also compared implantation and follow-up IVUS studies in 15 non-SES restenosis patients; none had a decrease in the number of struts at the minimum lumen site suggestive of strut fracture.


View this table:
[in this window]
[in a new window]
 
TABLE 2. Comparison of 24 Patients With and 25 Patients Without SES Restenosis

IVUS Predictors of In-Stent Neointimal Proliferation
Combining both the restenotic and nonrestenotic groups, we found no significant correlation between vessel angulation and normalized number of stent struts at the minimum lumen site (r=0.219, P=NS).

Of the 368 cross sections selected for analysis, detectable IH was present in 119. Maximum IH thickness was observed within the maximal interstrut angle in 98 of these 119 sections (82%), and the maximum peri-stent P&M thickness was observed within the maximal interstrut angle in 41 of these 119 sections (34%).

Univariate IVUS predictors of IH CSA, MLA, and IH thickness at follow-up are shown in Table 3. Multivariate analysis identified the number of visualized stent struts normalized for the number of stent cells and the maximum interstrut angle as the only independent predictor of IH CSA (coefficient, –4.154, P<0.01; and coefficient, 0.013, P<0.01) and thickness (coefficient, –0.642, P<0.01; and coefficient, 0.003, P<0.01). Stent CSA, number of visualized stent struts normalized for the number of stent cells, and maximum interstrut angle were the only independent IVUS predictors of MLA (coefficient, 0.788, P<0.05; coefficient, 4.377, P<0.01; and coefficient, –0.013, P<0.01).


View this table:
[in this window]
[in a new window]
 
TABLE 3. Univariate Predictors of IH Area and Thickness and MLA at Follow-Up


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowConclusions
down arrowReferences
 
IVUS analysis of SES restenoses demonstrated fewer stent struts, especially when normalized for the number of stent cells, and a larger maximum interstrut angle at the minimum lumen site compared with remote cross sections in restenotic SESs and with nonrestenotic SES lesions. This indicates that nonuniform strut distribution contributes to IH after SES implantation in de novo lesions.

In the Study of the Sirolimus-Eluting Balloon-Expandable Stent in the Treatment of Patients With De Novo Native Coronary Artery Lesions (SIRIUS) and TAXUS-II trials, restenotic lesions frequently were located near the stent margins or at the site of a gap between stents (64.5% in SIRIUS, 50.0% in TAXUS-II).3,8 Recently, SES fractures have been reported in cases of very focal intrastent restenosis while there was complete abolition of IH in the rest of the stented segment.9,10 A gap between stents, stent strut fracture, or the stent edge may be associated with a decrease in local drug delivery that may then contribute to development of restenosis. Similarly, an in vitro study demonstrated that local drug underdosing occurred at sites of nonuniform circumferential stent strut distribution.6 The present report found that maximum IH thickness occurred at the site of the maximal interstrut angle in 82% of SESs, along with a significant positive correlation between IH CSA and IH thickness and the maximum interstrut angle. In addition, 16 of 24 intrastent restenoses occurred in segments with <6 struts; there was a negative correlation between IH CSA and IH thickness and number of struts (having fewer struts was associated with more IH); and the number of visualized struts normalized for the number of stent cells was an independent predictor of IH CSA and thickness. In 13 patients with IVUS at implantation and follow-up, there was a decrease in stent strut number in only 2 patients, suggesting stent strut fracture in these patients. (Evidence of strut fracture was not seen in any nonrestenotic lesion.) In the remaining 11 patients with IVUS at implantation and follow-up, the number of stent struts remained constant. This suggests that nonuniform stent strut distribution may more common than stent strut fracture. Because SES effectiveness depends on correct dosing, nonuniform stent strut distribution may be a more important cause of IH proliferation than stent strut fracture. It may be necessary to optimize stent expansion or to implant another drug-eluting stent to counteract underdosing from nonuniform stent strut distribution.

Study Limitations
The number of lesions was small. Longitudinal distribution of stent struts was not considered. The angle of the ultrasound beam relative to the stent may affect the number of struts visualized; however, there was no correlation between the number of struts visualized and the maximum vessel bend in the lesion (determined angiographically). Postimplantation and follow-up IVUS was performed in only 13 of 24 restenotic SESs; therefore, it is not possible to exclude strut fracture in the other 11 patients. However, when only the 11 patients with postimplantation and follow-up IVUS and no change in the number of stent struts were analyzed, the conclusions were unchanged. The correlation coefficients are not strong, indicating that other factors may be important.


*    Conclusions
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*Conclusions
down arrowReferences
 
Nonuniform stent strut distribution and fewer struts may result in more IH after SES.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
up arrowConclusions
*References
 
1. Sousa JE, Costa MA, Sousa AG, Abizaid AC, Seixas AC, Abizaid AS, Feres F, Mattos LA, Falotico R, Jaeger J, Popma JJ, Serruys PW. Two-year angiographic and intravascular ultrasound follow-up after implantation of sirolimus-eluting stents in human coronary arteries. Circulation. 2003; 107: 381–383.[Abstract/Free Full Text]

2. Morice MC, Serruys PW, Sousa JE, Fajadet J, Ban Hayashi E, Perin M, Colombo A, Schuler G, Barragan P, Guagliumi G, Molnar F, Falotico R, for the RAVEL Study Group. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002; 346: 1773–1780.[Abstract/Free Full Text]

3. Moses JW, Leon MB, Popma JJ, Fitzgerald PJ, Holmes DR, O’Shaughnessy C, Caputo RP, Kereiakes DJ, Williams DO, Teirstein PS, Jaeger JL, Kuntz RE, for the SIRIUS Investigators. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med. 2003; 349: 1315–1323.[Abstract/Free Full Text]

4. Colombo A, Orlic D, Stankovic G, Corvaja N, Spanos V, Montorfano M, Liistro F, Carlino M, Airoldi F, Chieffo A, Di Mario C. Preliminary observations regarding angiographic pattern of restenosis after rapamycin-eluting stent implantation. Circulation. 2003; 107: 2178–2180.[Abstract/Free Full Text]

5. Lemos PA, Saia F, Ligthart JM, Arampatzis CA, Sianos G, Tanabe K, Hoye A, Degertekin M, Daemen J, McFadden E, Hofma S, Smits PC, de Feyter P, van der Giessen WJ, van Domburg RT, Serruys PW. Coronary restenosis after sirolimus-eluting stent implantation: morphological description and mechanical analysis from a consecutive series of cases. Circulation. 2003; 108: 257–260.[Abstract/Free Full Text]

6. Hwang CW, Wu D, Edelman ER. Physiological transport forces govern drug distribution for stent-based delivery. Circulation. 2001; 104: 600–605.[Abstract/Free Full Text]

7. Popma JJ, Gibson MC. Qualitative and quantitative angiography. In: Topol EJ, ed. Textbook of Interventional Cardiology. 4th ed. Philadelphia, Pa: WB Saunders; 2003: 827–846.

8. Colombo A, Drzewiecki J, Banning A, Grube E, Hauptmann K, Silber S, Dudek D, Fort S, Schiele F, Zmudka K, Guagliumi G, Russell ME, for the TAXUS II Study Group. Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions. Circulation. 2003; 108: 788–794.[Abstract/Free Full Text]

9. Sianos G, Hofma S, Ligthart JM, Saia F, Hoye A, Lemos PA, Serruys PW. Stent fracture and restenosis in the drug-eluting stent era. Cathet Cardiovasc Interv. 2004; 61: 111–116.[CrossRef][Medline] [Order article via Infotrieve]

10. Halkin A, Carlier S, Leon MB. Late incomplete lesion coverage following Cypher stent deployment for diffuse right coronary artery stenosis. Heart. 2004; 90: e45.[Abstract/Free Full Text]




This article has been cited by other articles:


Home page
ESC Textbook of Cardiovascular MedicineHome page
C. Di Mario, G. R. Heyndrickx, F. Prati, and N. H.J. Pijls
CHAPTER 8 Invasive Imaging and Haemodynamics
ESC Textbook of Cardiovascular Medicine, January 1, 2009; 2(1): med-9780199566990-chapter - med-9780199566990-chapter.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol IntvHome page
E. Romagnoli, G. M. Sangiorgi, J. Cosgrave, E. Guillet, and A. Colombo
Drug-eluting stenting the case for post-dilation.
J. Am. Coll. Cardiol. Intv., February 1, 2008; 1(1): 22 - 31.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
S. J. Corbett, J. Cosgrave, G. Melzi, R. Babic, G. G.L. Biondi-Zoccai, C. Godino, N. Morici, F. Airoldi, I. Michev, M. Montorfano, et al.
Patterns of restenosis after drug-eluting stent implantation: insights from a contemporary and comparative analysis of sirolimus- and paclitaxel-eluting stents
Eur. Heart J., October 1, 2006; 27(19): 2330 - 2337.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
G. S. Mintz and N. J. Weissman
Intravascular Ultrasound in the Drug-Eluting Stent Era
J. Am. Coll. Cardiol., August 1, 2006; 48(3): 421 - 429.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
M.-K. Hong, G. S. Mintz, C. W. Lee, D.-W. Park, B.-R. Choi, K.-H. Park, Y.-H. Kim, S.-S. Cheong, J.-K. Song, J.-J. Kim, et al.
Intravascular ultrasound predictors of angiographic restenosis after sirolimus-eluting stent implantation
Eur. Heart J., June 1, 2006; 27(11): 1305 - 1310.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
D. J. Kereiakes and D. P. Faxon
Left Main Coronary Revascularization at the Crossroads
Circulation, May 30, 2006; 113(21): 2480 - 2484.
[Full Text] [PDF]


Home page
PERSPECT VASC SURG ENDOVASC THERHome page
K. T. Delis and P. Gloviczki
Middle Aortic Syndrome: From Presentation to Contemporary Open Surgical and Endovascular Treatment
Perspectives in Vascular Surgery and Endovascular Therapy, September 1, 2005; 17(3): 187 - 203.
[Abstract] [PDF]


Home page
PERSPECT VASC SURG ENDOVASC THERHome page
K. T. Dells and P. Gloviczki
Last Word
Perspectives in Vascular Surgery and Endovascular Therapy, September 1, 2005; 17(3): 205-1 - 206.
[PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
110/22/3430    most recent
01.CIR.0000148371.53174.05v1
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Takebayashi, H.
Right arrow Articles by Moses, J. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Takebayashi, H.
Right arrow Articles by Moses, J. W.
Right arrowPubmed/NCBI databases
*Compound via MeSH
*Substance via MeSH
Related Collections
Right arrow Restenosis