(Circulation. 2004;110:501-507.)
© 2004 American Heart Association, Inc.
Original Articles |
From the Department of Cardiology, Childrens Hospital, Boston, Mass (A.L.K., J.E.L., S.B.P., D.B.M., K.G., M.J.L., K.J.J.); Department of Pediatrics, Childrens Hospital of Philadelphia, Philadelphia, Pa (J.J.R.); Department of Pediatrics, Columbia University College of Physicians and Surgeons, New York, NY (W.E.H.); and Department of Pediatrics, University of Illinois, Chicago (C.E.R.). Dr Perry is now at Stanford University Medical Center, Stanford, Calif. Dr Ruiz is now at the University of Illinois, Chicago.
Correspondence to Kathy J. Jenkins, MD, MPH, Department of Cardiology, Childrens Hospital, 300 Longwood Ave, Boston, MA 02115. E-mail kathy.jenkins{at}tch.harvard.edu
Received July 1, 2003; de novo received October 22, 2003; revision received April 1, 2004; accepted April 4, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results Since 1989, 170 patients (median age, 3.9 years) have undergone catheterization for closure of 1 or more congenital (n=92) or postoperative (n=78) residual VSDs using successive generations of STARFlex-type devices. Outcomes included echocardiographic assessment of residual flow and device position, assessment of VSD shunt/severity, and adverse events. Among 168 patients in whom device implantation was performed, between 1 and 7 devices were placed per patient (median, 1), with multiple devices placed in 40%. There was a significant decrease in left-to-right shunting after device implantation (P<0.001) and significant improvement in VSD size/severity, and device position proved stable. Of 332 adverse events, 39 were related to the device and 261 were related to the catheterization; all but 5 occurred in the periprocedural period. At a median follow-up of 24 months (0 to 154 months), 14 patients had died and 18 had device(s) explanted.
Conclusions Congenital and postoperative VSD closure using STARFlex-type devices resulted in stable improvement in clinical status and decreased interventricular shunting. Although periprocedural events occurred frequently, late events caused by the device were rare. Transcatheter device closure is an effective management option for patients with complex muscular VSDs that are difficult to approach surgically and for postoperative residual VSDs.
Key Words: catheterization heart septal defects ventricles
| Introduction |
|---|
|
|
|---|
Since first reporting transcatheter closure of muscular VSDs in 1988,7 we have performed device closure as primary treatment for simple defects, as an adjunct to surgery in patients with more complex disease, and as a means to reduce shunting from postoperative VSDs without further exposure to cardiopulmonary bypass.810 To date, published reports of this approach are limited to small series with relatively limited follow-up.1119 Since 1989, 168 patients with unrepaired congenital VSDs and postoperative residual VSDs have undergone transcatheter closure with successive generations of STARFlex-type devices as a part of approved regulatory trials. A subset of this experience was used to obtain FDA approval of the second-generation CardioSEAL device for closure of specific VSDs (NMT Medical Inc).
| Methods |
|---|
|
|
|---|
Patients were referred by a primary cardiologist for consideration of device closure of the VSD. For patients enrolled in cohort 2, data were reviewed by an interventional cardiologist to certify the technical feasibility of device placement and by an independent cardiologist and cardiovascular surgeon to determine whether the patient met admission criteria. Patients were selected for device closure if they had (1) 1 or more VSDs that were ascertained to result in sufficient hemodynamic derangement to warrant intervention and (2) either a type of VSD that was technically difficult to close surgically or an overall medical condition with associated surgical risks sufficient to justify the known and potential unknown risks of the device. Written informed consent for device placement and follow-up evaluation was obtained from the patient or parent/legal guardian.
Device Implantation
The Clamshell and CardioSEAL are low-profile double-umbrella devices, with hinged arms attached to Dacron fabric. The major enhancement of the STARFlex version of the CardioSEAL consists of a flexible, self-centering mechanism and a smaller delivery profile.20,21 Implantation procedures were performed as previously described.22,23
Follow-Up and Outcome Assessment
Preclosure and follow-up (immediately after, 6 months after, and most recent) assessments were performed and evaluated for all patients. Follow-up assessments were performed until at least 12 months (cohort 1) or 24 months (cohort 2), after which follow-up was collected in a registry format. Patient demographics and comorbid conditions were collected. Outcomes included survival, echocardiographic assessment of residual flow and device position, VSD assessment score, and adverse events.
Echocardiographic Assessment
Echocardiographic assessment included device arm position and integrity, contact with valve structures, residual flow through the defect, and ratio of VSD diameter to aortic annulus diameter. Flow through the VSD was characterized as none, trivial (single-color flow jet with proximal jet width <1 mm), small (single-color flow jet with a proximal jet width of 1 to 2 mm in all views in children weighing <20 kg or 1 to 3 mm in larger children and adults), or greater than small (single-color flow jet with proximal jet width >2 mm in children <20 kg or >3 mm in older children and adults; alternatively, multiple defects with small or greater than small shunting). Device position was categorized as correct, incorrect, or uncertain. Devices that were ultimately explanted were graded according to the last position documented by echocardiography. For cohort 2, a core laboratory at an unaffiliated institution was responsible for the final interpretation of the echocardiograms.
VSD Size/Severity Assessment
A VSD size/severity assessment scale was designed to assess clinical significance of the VSD in cohort 2 patients. Assessment was performed using 1 of 2 parallel yet equivalent scales: patients with native or surgically induced right ventricular outflow tract obstruction were evaluated according to the size of the VSD normalized to the aortic valve annulus diameter, and patients in whom the VSD resulted in a left-to-right shunt were evaluated according to the hemodynamic size of the shunt (Table 1). This scale allowed for unified analysis of outcomes in patients with and without obstruction to pulmonary blood flow.
|
Adverse Events
Adverse events were ascertained at each assessment point on the basis of clinical evaluation. Events were graded as not serious (minimal transient impairment), moderately serious (moderate transient impairment or requiring intervention), or serious (life-threatening). For this report, adverse events were categorized as device-related or catheterization-related. Device arm fractures, device explantations, and deaths were also recorded. For cohort 2, a Safety and Data Monitoring Committee composed of unaffiliated professionals independently reviewed all adverse events and was responsible for final attributability and severity classification.
Data Analysis
Continuous data are presented as mean±SD or median (range). Continuous and categorical variables were compared between groups using the 2-sample t test or Fisher exact test, respectively. Paired ordinal data (eg, VSD size/severity assessment) were compared by use of the Wilcoxon signed-rank test. Differences in VSD size/severity assessment from before implantation to most recent follow-up were compared between groups with the Mann-Whitney test. Freedom from device explantation or death was assessed by the Kaplan-Meier product-limit method, with comparisons between diagnostic subgroups performed by use of the log-rank test. Relationships between outcomes and patient characteristics were explored by logistic regression analysis.
| Results |
|---|
|
|
|---|
|
Device Implantation
Procedural details are summarized in Table 3. In 31 patients, 2 (n=25) or 3 (n=6) catheterization procedures were performed with the intent to close VSDs. Among these patients, the median interval between subsequent procedures was 52 days (1 day to 1.7 years), with 32% of secondary/tertiary procedures performed within 1 week of the first and 84% within 6 months.
|
A total of 296 devices were used (Table 4). The sizes of the devices implanted ranged from 17 to 40 mm, although the majority (75%) were small (17 to 23 mm).
|
Follow-Up Evaluation
Echocardiographic Assessment of Residual Flow and Device Position
Among patients in whom precatheterization VSD shunting could be assessed (ie, patients with a prospective assessment and without native or surgically induced obstruction to right ventricular outflow), 91% had "greater than small" shunting before device closure, versus 14% at most recent follow-up (P<0.001 for all postimplantation time points versus preimplantation; Figure 1, top). Of the 111 patients in whom precatheterization VSD shunting was not assessed prospectively (ie, patients in cohort 1) or could not be assessed secondary to native or surgically induced obstruction to right ventricular outflow, assessment of VSD flow at most recent follow-up was possible in 86, 87% of whom had "less than small" (n=39) or small (n=36) shunting, with "greater than small" in the remaining 13% (n=11). Figure 1, bottom shows device position at follow-up.
|
VSD Size/Severity Assessment
VSD size/severity assessment data for cohort 2 are presented in Figure 2. Baseline VSD size/severity did not differ according to demographic or diagnostic features. Patients who ultimately required multiple devices or multiple procedures for device placement had lower preimplantation VSD size/severity scores, indicating larger defects (P=0.002 and P=0.01, respectively). There were significant improvements in VSD size/severity after device implantation, which were noted at all postimplantation time points (Figure 2) and in both diagnostic subgroups. Among the entire cohort, VSD size/severity improved by
1 in 71 patients (91%) and
2 in 51 (65%). Compared with before implantation, VSD size/severity was significantly improved at most recent follow-up in both unrepaired congenital (median, 3 to 4) and postoperative residual (median, 2 to 4) subgroups (P<0.001). The improvement in VSD size/severity was more pronounced in patients with unrepaired congenital than those with postoperative residual VSDs (P=0.04). There was a trend toward greater improvement in VSD size/severity among patients who had multiple devices or multiple procedures (P=0.08), reflecting the lower preimplantation VSD size/severity in these patients. VSD size/severity at most recent follow-up did not differ according to other demographic or diagnostic features.
|
Adverse Events
A total of 332 adverse events were reported in 153 of 170 patients (90%) who underwent attempted VSD closure (Table 5). Nearly half (160 of 332) of the adverse events were graded as moderately serious or serious. There were 39 adverse events related to the device, 26 of which were moderately serious or serious, including 6 that were ongoing at the time of most recent follow-up. Only 1 device-related event was detected after hospital discharge. There were 261 events related to the catheterization procedure, of which 134 were graded as moderately serious or serious. Only 17 of these were ongoing at most recent follow-up, and 4 were detected after hospital discharge. The frequency of complications did not differ over time.
|
Device Arm Fracture, Device Explantation, and Death
Device arm fractures were observed on follow-up evaluation in 10% to 18% of devices, with a higher incidence with the Clamshell devices and larger devices (Table 4).
During the follow-up period, 18 patients had 1 or more devices explanted a median of 3.6 months (1 day to 24 months) after placement, representing 6% of the total devices placed and 11% of patients in whom device placement was attempted. The reason for device explantation was clearly related to the device in 13 of 18 patients (Table 6).
|
During the follow-up period, there were 14 deaths, for an all-cause mortality of 8.2%. The median duration from catheterization to death was 8 months (2 days to 10 years). Causes of death were cardiovascular in 7 patients: 3 had preexisting progressive multiorgan dysfunction that did not improve after VSD closure; 1 was a 3.5-month-old infant with a nearly absent ventricular septum and severe congestive heart failure who developed hypotension during the catheterization procedure and suffered progressive hemodynamic demise (at autopsy, a single ventricle was observed); 1 suffered postoperative complications after surgical closure of persistent multiple VSDs and removal of the VSD device; 1 had a presumed arrhythmia; and 1 succumbed to progressive congestive heart failure. The other 7 deaths were a result of aspiration asphyxia (n=1), respiratory syncytial virus infection (n=1), progressive pulmonary fibrosis (n=1), sepsis (n=1), and unknown (n=3).
As depicted in Figure 3, patients with a postoperative residual VSD had a significantly lower Kaplan-Meier probability of freedom from device explantation or death (P=0.01). On multivariate analysis, there were no significant differences in outcomes (survival, improvement in VSD size/severity) on the basis of patient characteristics.
|
| Discussion |
|---|
|
|
|---|
The majority of patients in this study (69%) had defects in the muscular interventricular septum, which are frequently difficult to close surgically or entail potentially deleterious approaches such as ventriculotomy.16 Accordingly, in many of these challenging patients, transcatheter device closure of unrepaired VSDs was undertaken as 1 element of a combined (surgical and interventional catheterization) management strategy. The remainder of our cohort had either intentionally fenestrated VSD patches, which we have discussed in a separate report,24 or postoperative residual defects along the margin of patches used to close VSDs in the membranous or outflow portions of the septum. Such defects are generally small, discrete, and amenable to device closure because of their size and location.
Technical and Device Considerations
Depending on the location of the VSD(s) and other cardiovascular anatomic features, several technical approaches were used. There was no discernible relationship between the location of the VSD and outcome. Frequently, multiple devices were necessary to close multiple defects, and multiple procedures were required in 18% of patients. Neither of these factors appeared to affect outcome adversely.
Device arm fractures were reported more frequently among the earlier cohort but occurred consistently in 15% of devices. The higher frequency of arm fractures observed in Clamshell devices is probably related to design issues rather than follow-up, because late fracture detection is unusual. In addition, there was a higher incidence of arm fracture in larger devices. Of note, the fractures were noted incidentally and did not appear to be clinically significant according to our measured outcomes.
Adverse Events
Adverse events were common, occurring in 90% of patients, and nearly half (48%) were categorized as moderately serious or serious. Most events occurred as a part of the hemodynamic instability associated with device positioning in these lengthy, complex procedures, which require vigilant anesthetic management.9 Events specifically related to device positioning were less common than hemodynamic instability, but important events such as embolization and valve impingement did occur and were often serious. Thirteen patients (7.7% of 168 patients who received implants) ultimately had 1 or more devices explanted because of malposition, embolization, or ineffective closure. These events highlight the importance of appropriate VSD selection and the technical expertise required to accomplish these procedures.
Fourteen patients died during the follow-up period. In 1 patient, the cause of death was directly attributable to the device or the catheterization procedure when the device was placed. Four patients were severely ill before VSD closure and never recovered, ultimately dying of multiorgan failure. Otherwise, where information was available, deaths appeared to be secondary to progression of underlying illness or unrelated events. We have previously determined that intracardiac implants are unlikely to cause sudden cardiac death.25
Limitations
There are several limitations of this study, especially for cohort 1. VSD size/severity assessment was not performed in patients in cohort 1 because this database was reconstructed in a retrospective manner. Thus, efficacy could not be assessed for patients in cohort 1 with right ventricular outflow tract obstruction. Also, we have no record of patients in cohort 1 who underwent catheterization with intended VSD closure but did not have a device implantation attempted. We chose to include cohort 1 patients to evaluate the long-term efficacy of a device closure strategy.
Conclusions
Transcatheter device closure is an effective management option in patients with unrepaired congenital muscular VSDs in locations that require ventriculotomy for surgical closure and postoperative residual VSDs. Despite the complexity of the population, only 1 patient died as a result of the procedure, and 8% required surgical device removal. The majority of patients had successful and sustained defect closure, with a significant reduction in shunting. Although adverse events were common, they were generally manageable and did not outweigh the benefits. Further analysis is necessary to determine whether devices in the ventricular septum have long-term effects on septal mechanics or ventricular function.
| References |
|---|
|
|
|---|
2. Kitagawa T, Durham LA, Mosca RS, et al. Techniques and results in the management of multiple ventricular septal defects. J Thorac Cardiovasc Surg. 1998; 115: 848855.
3. Danilowicz D, Presti S, Colvin S, et al. Results of urgent or emergency repair for symptomatic infants under one year of age with single or multiple ventricular septal defects. Am J Cardiol. 1992; 69: 699701.[CrossRef][Medline] [Order article via Infotrieve]
4. Serraf A, Lacour-Gayet F, Bruniaux J, et al. Surgical management of isolated multiple ventricular septal defects: logical approach in 130 cases. J Thorac Cardiovasc Surg. 1992; 103: 437443.[Abstract]
5. Kirklin JK, Castaneda AR, Keane JF, et al. Surgical management of multiple ventricular septal defects. J Thorac Cardiovasc Surg. 1980; 80: 485493.[Abstract]
6. Van Praagh S, Mayer JE, Berman NB, et al. Apical ventricular septal defects: follow-up concerning anatomic and surgical considerations. Ann Thorac Surg. 2002; 73: 4857.
7. Lock JE, Block PC, McKay RG, et al. Transcatheter closure of ventricular septal defects. Circulation. 1988; 78: 361368.
8. Bridges ND, Perry SB, Keane JF, et al. Preoperative transcatheter closure of congenital muscular ventricular septal defects. N Engl J Med. 1991; 324: 13121317.[Abstract]
9. Laussen PC, Hansen DD, Perry SB, et al. Transcatheter closure of ventricular septal defects: hemodynamic instability and anesthetic management. Anesth Analg. 1995; 80: 10761082.[Abstract]
10. Kumar K, Lock JE, Geva T. Apical muscular ventricular septal defects between the left ventricle and the right ventricular infundibulum: diagnostic and interventional considerations. Circulation. 1997; 95: 12071213.
11. Rigby ML, Redington AN. Primary transcatheter umbrella closure of perimembranous ventricular septal defect. Br Heart J. 1994; 23: 368371.
12. Murzi B, Bonanomi GL, Giusti S, et al. Surgical closure of muscular ventricular septal defects using double umbrella devices (intraoperative VSD device closure). Eur J Cardiothorac Surg. 1997; 12: 450454.[Abstract]
13. Sideris EB, Walsh KP, Haddad JL, et al. Occlusion of congenital ventricular septal defects by the buttoned device: "Buttoned device" clinical trials international register. Heart. 1997; 77: 276279.
14. Kalra GS, Verma PK, Dhall A, et al. Transcatheter device closure of ventricular septal defects: immediate results and intermediate-term follow-up. Am Heart J. 1999; 138: 339344.[CrossRef][Medline] [Order article via Infotrieve]
15. Janorkar S, Goh T, Wilkinson J. Transcatheter closure of ventricular septal defects using the Rashkind device: initial experience. Catheter Cardiovasc Interv. 1999; 46: 4348.[CrossRef][Medline] [Order article via Infotrieve]
16. Thanopoulos BD, Tsaousis GS, Konstadopoulou GN, et al. Transcatheter closure of muscular ventricular septal defects with the Amplatzer ventricular septal defect occluder: initial clinical applications in children. J Am Coll Cardiol. 1999; 35: 13951399.
17. Hijazi ZM, Hakim F, Al-Fadley F, et al. Transcatheter closure of single muscular ventricular septal defects using the Amplatzer muscular VSD occluder: initial results and technical considerations. Catheter Cardiovasc Interv. 2000; 49: 167172.[CrossRef][Medline] [Order article via Infotrieve]
18. Hijazi ZM, Hakim F, Hawaleh AA, et al. Catheter closure of perimembranous ventricular septal defects using the new Amplatzer membranous VSD occluder: initial clinical experience. Catheter Cardiovasc Interv. 2002; 56: 508515.[CrossRef][Medline] [Order article via Infotrieve]
19. Chessa M, Carminati M, Cao QL, et al. Transcatheter closure of congenital and acquired muscular ventricular septal defects using the Amplatzer device. J Invasive Cardiol. 2002; 14: 322327.[Medline] [Order article via Infotrieve]
20. Hausdorf G, Kaulitz R, Paul T, et al. Transcatheter closure of atrial septal defect with new flexible, self-centering device (the STARFlex Occluder). Am J Cardiol. 1999; 84: 11131116.[CrossRef][Medline] [Order article via Infotrieve]
21. Kreutzer J, Ryan CA, Wright JA, et al. Acute animal studies of the STARFlex system: a new self-centering CardioSEAL septal occluder. Catheter Cardiovasc Interv. 2000; 49: 225233.[CrossRef][Medline] [Order article via Infotrieve]
22. Lock JE, Block PC, McKay RG, et al. Transcatheter closure of ventricular septal defects. Circulation. 1988; 78: 361368.
23. Bridges ND, Perry SB, Keane JF, et al. Preoperative transcatheter closure of congenital muscular ventricular septal defects. N Engl J Med. 1991; 324: 13121317.[Abstract]
24. Marshall AC, Love BA, Lang P, et al. Staged repair of tetralogy of Fallot and diminutive pulmonary arteries using a fenestrated ventricular septal defect patch. J Thorac Cardiovasc Surg. 2003; 126: 14271433.
25. Perry YY, Triedman JK, Gauvreau K, et al. Sudden death in patients after transcatheter device implantation for congenital heart disease. Am J Cardiol. 2000; 85: 992995.[CrossRef][Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
M.-B. Gu, Y. Bai, X.-X. Zhao, X. Zheng, W.-P. Li, and Y.-W. Qin Transcatheter closure of postoperative residual perimembranous ventricular septal defects. Ann. Thorac. Surg., November 1, 2009; 88(5): 1551 - 1555. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Warnes, R. G. Williams, T. M. Bashore, J. S. Child, H. M. Connolly, J. A. Dearani, P. del Nido, J. W. Fasules, T. P. Graham Jr, Z. M. Hijazi, et al. ACC/AHA 2008 Guidelines for the Management of Adults With Congenital Heart Disease: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Develop Guidelines on the Management of Adults With Congenital Heart Disease) Developed in Collaboration With the American Society of Echocardiography, Heart Rhythm Society, International Society for Adult Congenital Heart Disease, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons J. Am. Coll. Cardiol., December 2, 2008; 52(23): e143 - e263. [Full Text] [PDF] |
||||
![]() |
C. A. Warnes, R. G. Williams, T. M. Bashore, J. S. Child, H. M. Connolly, J. A. Dearani, P. del Nido, J. W. Fasules, T. P. Graham Jr, Z. M. Hijazi, et al. ACC/AHA 2008 Guidelines for the Management of Adults With Congenital Heart Disease: Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Adults With Congenital Heart Disease) Developed in Collaboration With the American Society of Echocardiography, Heart Rhythm Society, International Society for Adult Congenital Heart Disease, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons J. Am. Coll. Cardiol., December 2, 2008; 52(23): 1890 - 1947. [Full Text] [PDF] |
||||
![]() |
C. A. Warnes, R. G. Williams, T. M. Bashore, J. S. Child, H. M. Connolly, J. A. Dearani, P. del Nido, J. W. Fasules, T. P. Graham Jr, Z. M. Hijazi, et al. ACC/AHA 2008 Guidelines for the Management of Adults With Congenital Heart Disease: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Develop Guidelines on the Management of Adults With Congenital Heart Disease): Developed in Collaboration With the American Society of Echocardiography, Heart Rhythm Society, International Society for Adult Congenital Heart Disease, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons Circulation, December 2, 2008; 118(23): e714 - e833. [Full Text] [PDF] |
||||
![]() |
C. A. Warnes, R. G. Williams, T. M. Bashore, J. S. Child, H. M. Connolly, J. A. Dearani, P. del Nido, J. W. Fasules, T. P. Graham Jr, Z. M. Hijazi, et al. ACC/AHA 2008 Guidelines for the Management of Adults With Congenital Heart Disease: Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Adults With Congenital Heart Disease): Developed in Collaboration With the American Society of Echocardiography, Heart Rhythm Society, International Society for Adult Congenital Heart Disease, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons Circulation, December 2, 2008; 118(23): 2395 - 2451. [Full Text] [PDF] |
||||
![]() |
Z. M. Hijazi and S. M. Awad Pediatric Cardiac Interventions J. Am. Coll. Cardiol. Intv., December 1, 2008; 1(6): 603 - 611. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Predescu, R. R. Chaturvedi, M. K. Friedberg, L. N. Benson, A. Ozawa, and K.-J. Lee Complete heart block associated with device closure of perimembranous ventricular septal defects. J. Thorac. Cardiovasc. Surg., November 1, 2008; 136(5): 1223 - 1228. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Khairy and A. J. Marelli Clinical Use of Electrocardiography in Adults With Congenital Heart Disease Circulation, December 4, 2007; 116(23): 2734 - 2746. [Full Text] [PDF] |
||||
![]() |
M. Carminati, G. Butera, M. Chessa, J. De Giovanni, G. Fisher, M. Gewillig, M. Peuster, J. F. Piechaud, G. Santoro, H. Sievert, et al. Transcatheter closure of congenital ventricular septal defects: results of the European Registry Eur. Heart J., October 1, 2007; 28(19): 2361 - 2368. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Butera, M. Carminati, M. Chessa, L. Piazza, A. Micheletti, D. G. Negura, R. Abella, A. Giamberti, and A. Frigiola Transcatheter Closure of Perimembranous Ventricular Septal Defects: Early and Long-Term Results J. Am. Coll. Cardiol., September 18, 2007; 50(12): 1189 - 1195. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. P. Walsh Interventional Electrophysiology in Patients With Congenital Heart Disease Circulation, June 26, 2007; 115(25): 3224 - 3234. [Full Text] [PDF] |
||||
![]() |
I. Inglessis and M. J. Landzberg Interventional Catheterization in Adult Congenital Heart Disease Circulation, March 27, 2007; 115(12): 1622 - 1633. [Full Text] [PDF] |
||||
![]() |
E. P. Walsh and F. Cecchin Arrhythmias in Adult Patients With Congenital Heart Disease Circulation, January 30, 2007; 115(4): 534 - 545. [Full Text] [PDF] |
||||
![]() |
I. A. Russell, K. Rouine-Rapp, G. Stratmann, and W. C. Miller-Hance Congenital heart disease in the adult: a review with internet-accessible transesophageal echocardiographic images. Anesth. Analg., March 1, 2006; 102(3): 694 - 723. [Full Text] [PDF] |
||||
![]() |
T. P. Graham Jr The Year in Congenital Heart Disease J. Am. Coll. Cardiol., June 7, 2005; 45(11): 1887 - 1899. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |