(Circulation. 2004;110:810-814.)
© 2004 American Heart Association, Inc.
Original Articles |
From the Department of Internal Medicine IIICardiology/Angiology (B.S., M.B., G.N.), University of Saarland, Homburg/Saar, Germany; the Department of Radiology (U.S., C.A.), Charité University Hospital, Berlin, Germany; and Anakat GmbH (U.B.), Berlin, Germany.
Correspondence to Bruno Scheller, MD, Internal Medicine III (Cardiology/Angiology), University of Saarland, D 66421 Homburg/Saar, Germany. E-mail bruno.scheller{at}uniklinik-saarland.de
Received January 29, 2004; revision received May 4, 2004; accepted May 6, 2004.
| Abstract |
|---|
|
|
|---|
Methods and Results Stainless steel stents (n=40; diameter, 3.0 to 3.5 mm; length, 18 mm) were implanted in the left anterior descending and circumflex coronary arteries of domestic pigs. Both conventional uncoated and 3 different types of paclitaxel-coated, percutaneous transluminal coronary angioplasty balloons (contact with vessel wall for 1 minute) were used. No difference in short-term tolerance between coated and uncoated balloons and no signs of thrombotic events were observed. Quantitative angiography and histomorphometry of the stented arteries asserted the statistical equality of the baseline parameters between the control and the 3 treatment groups. Paclitaxel balloon coating led to a marked, dose-dependent reduction of parameters characterizing in-stent restenosis (reduction of neointimal area up to 63%). Despite the marked reduction in neointimal proliferation, endothelialization of stent struts was present in all samples. There was no evidence of a significant inflammatory response in the neighborhood of the stent struts.
Conclusions Paclitaxel balloon coating is safe, and it effectively inhibits restenosis after coronary angioplasty with stent implantation in the porcine model. The degree of reduction in neointimal formation was comparable to that achieved with drug-eluting stents.
Key Words: restenosis angioplasty paclitaxel
| Introduction |
|---|
|
|
|---|
Drug-eluting stents (DES) were shown to be safe and feasible in reducing restenosis.37 However, their efficacy and safety have not been confirmed in all clinical settings, especially with regard to treating in-stent restenosis. Concerns have been raised that the polymeric matrix on the stent in which the antiproliferative drug is embedded might induce inflammation and thrombosis.8 Another important limitation of DES is the fact that the drug concentration is highest at the stent struts, where healing is most important. On the other hand, incomplete suppression of neointimal hyperplasia at the stent margins or between the struts may limit the efficacy of DES.4
Nonstent-based local delivery of antiproliferative drugs may offer additional flexibility and efficacy in the entire range of applications. It may also deliver drugs to vessel areas not directly covered by the stent, which could be of special interest for small and tortuous vessels. Furthermore, healing and reendothelialization of stent struts that do not carry antiproliferative agents may be facilitated.
Paclitaxel has already been investigated in previous studies that included a variety of catheter-based, local drug-delivery approaches. The "double-balloon" catheter,9 the "porous balloon,"10 and even intrapericardial administration11 have been used. Although all of these approaches showed efficacy in preclinical trials, they require special and sometimes cumbersome devices, involve blockage of coronary blood flow, or induce additional vascular injury.
The aim of the present study was to test a novel method of intracoronary local drug delivery in the porcine coronary overstretch model: paclitaxel coating of conventional percutaneous transluminal coronary angioplasty (PTCA) balloon catheters with a new coating technique that allows for immediate drug release on inflation.
| Methods |
|---|
|
|
|---|
2 µg paclitaxel per square millimeter of balloon surface, and procedure Ac used acetone as the solvent and was completed with 2 concentrations of paclitaxel: a low dose (AcL), resulting in 1.3 µg/mm2, and a regular dose (AcR), resulting in 2.5 µg/mm2. The paclitaxel dose in EEE and AcR catheters was the maximum achievable at the time of the experiments. After coating, uncoated bare stents (MeoFlex, MeoMedical) were crimped onto all balloons. No decomposition products of paclitaxel were detected by high-performance liquid chromatography (HPLC) after ethylene oxide sterilization. Pressure resistance of the balloons and stent delivery were not affected by the coating. Adhesion of paclitaxel to the surface of standard balloons was fully satisfactory, with no loss due to normal handling, significantly <1% of the dose adhering to the close-fitting, balloon-protecting tube, and very little loss in blood unless the balloon was expanded (see Results). The coating resulted in a very slight increase in profile but no recognizable change in flexibility. All experiments were conducted in accordance with the guidelines for animal experiments set forth by the local animal protection committee. Twenty-six domestic pigs were presedated by intramuscular injection of 6 mL ketamine and 4 mL xylazine. A venous access was provided, and anesthesia was induced by intravenous injection of propofol, followed by orotracheal intubation, which was maintained with 1.0% to 2.0% isoflurane, 70% N2O, and 30% oxygen (all vol/vol). All animals received 5000 IU heparin, 250 mg aspirin, and 200 µg intracoronary nitroglycerin. The coronary arteries were imaged by a standard angiographic technique via the left carotid artery. Target segments were selected in the left anterior descending (LAD) and circumflex (CX) arteries, and vessel diameters were estimated by comparison with the diameter of the angiographic catheter.
Paclitaxel Loss in Blood and Coronary Tissue Concentration After Inflation
To determine the uptake of paclitaxel into the vessel wall and various tissues, 4 domestic pigs (25 to 30 kg body weight) received paclitaxel-coated and uncoated balloon catheters without or with premounted stents. Coated balloon catheters without and with stents were advanced into the coronary arteries, where they were left uninflated in the bloodstream for
5 minutes and then withdrawn. The paclitaxel content of the balloons was determined by HPLC after extraction and compared with the paclitaxel content of unused catheters of the same batch. In the same animals, 3 different treatments were performed in randomized order: (1) implantation of a bare stent premounted on a paclitaxel-coated balloon catheter (AcR); (2) angioplasty with a paclitaxel-coated balloon catheter (AcR) without stent implantation (to "label" the site of dilatation, a short stent was implanted distal to it before PTCA was performed); and (3) bare-stent implantation with an uncoated balloon and then postdilatation of the stented segment with a paclitaxel-coated balloon catheter (AcR).
After 40 to 60 minutes, treated coronary artery segments from the LAD, CX, and right coronary arteries were dissected. Paclitaxel was extracted from homogenized tissue samples with dimethyl sulfoxide and from plasma samples with ethyl acetate, with docetaxel as an internal standard. Samples were analyzed by HPLC on a Waters Symmetry C-18 column (5 µm, 25 cm x 4.6 mm) and an acetonitrile/0.005 mol/L potassium phosphate buffer, pH 3.5. With 40:60 (vol/vol) acetonitrile/potassium phosphate, paclitaxel retention time was 35 to 45 minutes; with 55:45, paclitaxel retention time was
7.5 minutes. Detection was performed by ultraviolet photometry at 230 nm.
Restenosis Study
Stents were implanted with an oversize ratio of
1.2 (implantation pressure, 12 atm). According to the randomization list, the 20 pigs (body weight, 30.9±2.3 kg) received a total of 40 identical stents into LAD and CX arteries. The stents were premounted on identical balloon catheters, which were either uncoated or coated in 3 different ways: (1) control, consisting of bare stents crimped onto conventional, uncoated PTCA balloon catheters n=13); (2) consisting of bare stents crimped onto paclitaxel-coated PTCA balloon catheters with use of the EEE coating technique and a regular paclitaxel dose (n=10); (3) AcL, consisting of bare stents crimped onto paclitaxel-coated, PTCA balloon catheters with use of the Ac coating technique and low-dose paclitaxel (n=10); and (4) AcR, consisting of bare stents crimped onto paclitaxel-coated, PTCA balloon catheters with use of the Ac coating technique and regular paclitaxel dose (n=11).
After stent implantation, prophylactic intramuscular streptomycin and penicillin were given; pigs were allowed to recover and were monitored for 5 weeks. Oral aspirin 100 mg and ticlopidine 250 mg per day were administered starting 3 days before the procedure and continuing until sacrifice. After 5 weeks, the animals were euthanized with 20 mL pentobarbital while under deep anesthesia. Hearts were rapidly excised, the coronary system was flushed with 0.9% saline, and the arteries were fixed by perfusion with 4% buffered formalin under physiological pressure and overnight immersion. The target segments were then dissected and samples for histology obtained.
Quantitative Coronary Analysis
Coronary artery imaging was performed with a Philips PolyArc fluoroscope connected to a digitizer and an Apple Macintosh Power PC. The CAAS II System (Pie Medical) was used for quantitative coronary analysis by 2 experienced cardiologists blinded to the treatment groups. Discrepancies were resolved by mutual consensus.
Histology
Stented coronary arteries were dissected from the formalin-fixed hearts and immersed in methylmethacrylate (Merck). Three representative cross sections per stent were cut from the blocks with a coping saw, polished, and glued onto acrylic plastic slides. Final specimens were stained with hematoxylin eosin. After digitizing, histomorphometric measurements were taken with the NIH Image program (PC version Scion Image, Scion Corp). The evaluated parameters were luminal diameter, external elastic lamina (EEL) diameter, maximal neointimal thickness, EEL area, luminal area, and neointimal area. Injury scores were assigned as previously described by Schwartz et al,12 and the inflammation score for each individual strut was graded as described by Kornowski et al.13
Statistical Analysis
Histomorphometric variables of the 3 cross-sectional planes were averaged to obtain a mean value per stent. Continuous variables of quantitative coronary angiography were compared by 1-way, repeated-measures ANOVA (3 or 4 factors) and Students t test after a normalized distribution was assured by a statistical program (SPSS 11.5 for Windows, SPSS Inc). Data are presented as mean±SD.
| Results |
|---|
|
|
|---|
|
Restenosis Study
There were no acute or subacute thrombotic complications and no significant adverse events in terms of ECG parameters, blood pressure, or contractility during or after the coronary interventions. After inflation, stent deposition, and withdrawal through the guiding catheter, the paclitaxel content of the balloons was measured. The postprocedural content was 13±2% of the initial dose on the EEER balloons (n=10), 10±2% on the AcL balloons (n=10), and 7±2% on the AcR balloons (n=11).
Quantitative angiography and histomorphometry of the stented arteries ascertained the statistical equality of baseline parameters between the control and treatment groups (Table 2 and Figure 1). After 5 weeks, histological evaluation showed that all stents were sufficiently expanded (Figure 2). All groups had similar coronary artery sizes, as reflected by the EEL diameter and area. There were no significant angiographic and histomorphometric differences between control and EEER-coated balloons (Table 2 and Figure 1). The Ac coating procedure resulted in a marked, dose-dependent, and statistically significant reduction of late lumen loss and an equally impressive, statistically significant increase in minimal lumen diameter compared with controls (Table 2). Quantitative coronary angiography revealed no edge effects. Histomorphometry showed a statistically significant increase in lumen diameter and lumen area and a corresponding decrease in maximal neointimal thickness and neointimal area in the vessels treated with AcL- and AcR-coated balloons. The AcL paclitaxel coating led to a reduction of the neointimal area by 42% and the AcR coating, by 63%. Despite the marked reduction in neointimal proliferation, endothelialization of stent struts was present in all samples (Figure 2). There was no evidence of a significant inflammatory response in the neighborhood of the stent struts (Table 2).
|
|
|
| Discussion |
|---|
|
|
|---|
6% of the dose on the way to the coronary arteries and back. About 80% of the drug is released during inflation. We tested 2 different types of coating. According to the histomorphometry results, the EEE procedure was ineffective in the coronary arteries of our animal model. The use of paclitaxel-coated balloons prepared with the Ac procedure was associated with a marked reduction of neointimal proliferation. Endothelialization of stent struts was present in all samples. The difference in efficacy of the 2 coating procedures may be explained by the presence of a hydrophilic x-ray contrast-medium substance in the coating preparation in the case of the Ac version. The results of this study confirm that short-term exposure of injured arteries is sufficient to inhibit restenosis during a critical period of time after angioplasty in a recognized animal model. This model used a normal coronary artery with a very thin intima and moderately sized media, which is different from an atherosclerotic human coronary artery. Despite the differences in clinical application, this new approach may offer several advantages: There is no need for radiation, polymers, or other sustained-release techniques. This new technique allows for drug delivery to the total injured vessel area. Potential clinical indications are the prevention of restenosis after angioplasty, with or without stent implantation, and the treatment of in-stent restenosis (as an alternative to brachytherapy or stent-in-stent).
According to current concepts and the results of clinical trials, effective inhibition of restenosis requires local drug administration by implantation of DES. The use of stents, however, is not always possible or desirable. Restenosis inhibition is required in vessel areas beyond the immediate stent coverage, between stents, in small and tortuous vessels, and in restenotic, stented vessels. Furthermore, healing and reendothelialization of stent struts not carrying antiproliferative agents may be facilitated if the drug supply does not have its origin near the stent struts.
On the other hand, drug-releasing stents are unique in allowing sustained release for days and weeks after a single, short intervention. Restenosis due to neointimal hyperplasia is a slow process, suggesting the need for prolonged or repeated drug administration. Sustained drug release is considered to be essential for preventing restenosis by local drug delivery.14,15 However, in the TAXUS II trial, both slow- and moderate-release paclitaxel-eluting stents prevented neointimal formation to the same degree.16
Our concept of nonstent-based, local paclitaxel delivery was stimulated by the surprising observation that the short period of exposure during the passage of contrast media through the coronary arteries allows for taxane uptake sufficient to inhibit restenosis.17 A short incubation time (3 minutes) with paclitaxel added to the contrast agent iopromide (Ultravist) almost completely inhibited vascular smooth muscle cell proliferation for up to 12 days.17 Creel et al18 reported that uptake of the lipophilic paclitaxel into the arterial wall was increased 20-fold over heparin after 15 minutes of exposure. Although a process that requires 15 minutes is far too long to be applicable during PTCA, the experiment indicates that lipophilic drugs such as paclitaxel or sirolimus are rapidly taken up by the tissue. In 2 animal studies, a dose-dependent reduction of neointimal area and of restenosis by intracoronary iopromide taxane administration was demonstrated.17,19
If sustained release of antiproliferative drugs should indeed prove to be unnecessary for inhibiting biological processes that result in restenosis, alternative new methods of drug supply during the interventional procedure are conceivable. Application of drugs on the surface of balloon catheters is one possible option. The antiproliferative drug stays on the balloon until it is inflated. During inflation it is pressed against the stenotic vessel wall, and a sufficiently large proportion of the drug is dissolved and penetrates the vessel wall.
Sustained drug release from stent coatings results in long-lasting, low tissue concentrations over a longer period until the healing process is well advanced. In the case of coated balloons, higher tissue concentrations may initially be expected. In the present study, paclitaxel tissue concentration in the treated coronary arteries was measured 40 to 60 minutes after balloon insufflation. In combination with stents,
16% of the total paclitaxel dose was transferred to the injured vessel area. In a prior animal trial that used contrast mediapaclitaxel formulations for restenosis prevention, we found that the tissue concentration half-life of paclitaxel in the coronary artery wall ranged from 45 minutes (LAD) to 85 minutes (CX). Twenty-four hours later, paclitaxel was still detected in the LAD of 2 of 3 pigs. Paclitaxel concentrations were almost identical in the vessel wall proximal and distal to the stent and in the stented segment.17 Because of the half-life of
1 hour, prolonged drug action due to persisting high tissue concentrations seems unlikely.
Therefore, short-term exposure to paclitaxel obviously blocks early growth-initiating events that are crucial for subsequent neointimal formation. Importantly, exclusive abrogation of this premature proliferation is sufficient to prevent restenosis. Our data indicate that the paclitaxel balloon coating is safe in the porcine overstretch stent model. The degree of reduction in neointimal formation is at least comparable to that achieved with sustained drug release from stents.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2. Kuntz RE, Baim DS. Prevention of coronary restenosis: the evolving evidence base for radiation therapy. Circulation. 2000; 101: 21302133.
3. Morice MC, Serruys PW, Sousa JE, et al. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. N Engl J Med. 2002; 346: 17731780.
4. Moses JW, Leon MB, Popma JJ, et al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med. 2003; 349: 13151323.
5. Colombo A, Drzewiecki J, Banning A, et al. Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions. Circulation. 2003; 108: 788794.
6. Grube E, Silber S, Hauptmann KE, et al. TAXUS I: six- and 12-month results from a randomized, double-blind trial on a slow-release paclitaxel-eluting stent for de novo coronary lesions. Circulation. 2003; 107: 3842.
7. Park SJ, Shim WH, Ho DS, et al. A paclitaxel-eluting stent for the prevention of coronary restenosis. N Engl J Med. 2003; 348: 15371545.
8. van der Giessen WJ, Lincoff AM, Schwartz RS, et al. Marked inflammatory sequelae to implantation of biodegradable and nonbiodegradable polymers in porcine coronary arteries. Circulation. 1996; 94: 16901697.
9. Herdeg C, Oberhoff M, Baumbach A, et al. Local paclitaxel delivery for the prevention of restenosis: biological effects and efficacy in vivo. J Am Coll Cardiol. 2000; 35: 19697196.
10. Axel DI, Kunert W, Goggelmann C, et al. Paclitaxel inhibits arterial smooth muscle cell proliferation and migration in vitro and in vivo using local drug delivery. Circulation. 1997; 96: 636645.
11. Hou D, Rogers PI, Toleikis PM, et al. Intrapericardial paclitaxel delivery inhibits neointimal proliferation and promotes arterial enlargement after porcine coronary overstretch. Circulation. 2000; 102: 15751581.
12. Schwartz RS, Huber KC, Murphy JG, et al. Restenosis and the proportional neointimal response to coronary artery injury: results in a porcine model. J Am Coll Cardiol. 1992; 19: 267274.[Abstract]
13. Kornowski R, Hong MK, Tio FO, et al. In-stent restenosis contributions of inflammatory responses and arterial injury to neointimal hyperplasia. J Am Coll Cardiol. 1998; 31: 224230.
14. Gershlick AH. Treating atherosclerosis: local drug delivery from laboratory studies to clinical trials. Atherosclerosis. 2002; 160: 259271.[CrossRef][Medline] [Order article via Infotrieve]
15. Moses JW, Kipshidze N, Leon MB. Perspectives of drug-eluting stents: the next revolution. Am J Cardiovasc Drugs. 2002; 2: 165172.
16. Tanabe K, Serruys PW, Degertekin M, et al. Chronic arterial responses to polymer-controlled paclitaxel-eluting stents: comparison with bare metal stents by serial intravascular ultrasound analyses: data from the randomized TAXUS-II Trial. Circulation. 2004; 109: 196200.
17. Scheller B, Speck U, Schmitt A, et al. Addition of paclitaxel to contrast media prevents restenosis after coronary stent implantation. J Am Coll Cardiol. 2003; 42: 14151420.
18. Creel CJ, Lovich MA, Edelman ER. Arterial paclitaxel distribution and deposition. Circ Res. 2000; 86: 879884.
19. Scheller B, Speck U, Romeike B, et al. Contrast media as a carrier for local drug delivery: successful inhibition of neointimal proliferation in the porcine coronary stent model. Eur Heart J. 2003; 24: 14621467.
This article has been cited by other articles:
![]() |
M. Unverdorben, C. Vallbracht, B. Cremers, H. Heuer, C. Hengstenberg, C. Maikowski, G. S. Werner, D. Antoni, F. X. Kleber, W. Bocksch, et al. Paclitaxel-Coated Balloon Catheter Versus Paclitaxel-Coated Stent for the Treatment of Coronary In-Stent Restenosis Circulation, June 16, 2009; 119(23): 2986 - 2994. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Werk, S. Langner, B. Reinkensmeier, H.-F. Boettcher, G. Tepe, U. Dietz, N. Hosten, B. Hamm, U. Speck, and J. Ricke Inhibition of Restenosis in Femoropopliteal Arteries: Paclitaxel-Coated Versus Uncoated Balloon: Femoral Paclitaxel Randomized Pilot Trial Circulation, September 23, 2008; 118(13): 1358 - 1365. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Tepe, T. Zeller, T. Albrecht, S. Heller, U. Schwarzwalder, J.-P. Beregi, M.D. C. D. Claussen, A. Oldenburg, B. Scheller, and U. Speck Local Delivery of Paclitaxel to Inhibit Restenosis during Angioplasty of the Leg N. Engl. J. Med., February 14, 2008; 358(7): 689 - 699. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Scheller, U. Speck, and M. Bohm Prevention of restenosis: is angioplasty the answer? Heart, May 1, 2007; 93(5): 539 - 541. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Scheller, C. Hehrlein, W. Bocksch, W. Rutsch, D. Haghi, U. Dietz, M. Bohm, and U. Speck Treatment of Coronary In-Stent Restenosis with a Paclitaxel-Coated Balloon Catheter N. Engl. J. Med., November 16, 2006; 355(20): 2113 - 2124. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Speck, B. Scheller, C. Abramjuk, C. Breitwieser, J. Dobberstein, M. Boehm, and B. Hamm Neointima Inhibition: Comparison of Effectiveness of Non-Stent-based Local Drug Delivery and a Drug-eluting Stent in Porcine Coronary Arteries. Radiology, August 1, 2006; 240(2): 411 - 418. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2004 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |