(Circulation. 2005;112:2157-2162.)
© 2005 American Heart Association, Inc.
Imaging |
From Cardiac Imaging Research (S.L., J.D., P.C., B.B., G.R.S.), Department of Cardiology; Department of Electrical Engineering (J.D.), ESAT; and Department of Anaesthesiology (P.F.W., H.A.L.), Catholic University Leuven, Leuven, Belgium.
Correspondence to Stian Langeland, Catholic University Leuven, Department of Cardiology, Cardiac Imaging Research, Herestraat 49, B-3000 Leuven, Belgium. E-mail stian.langeland{at}uz.kuleuven.ac.be
Received November 8, 2004; de novo received April 6, 2005; revision received July 11, 2005; accepted July 12, 2005.
| Abstract |
|---|
|
|
|---|
Methods and Results In 5 open-chest sheep, ultrasound data were acquired. The new methodology was used to perform simultaneous measurements of radial and longitudinal strain in the inferolateral wall. Segment-length sonomicrometry crystals were used as the reference. After baseline acquisitions, deformation was modulated by pharmacologically changing the inotropic state of the myocardium and by inducing ischemia. Ultrasonically estimated radial and longitudinal strain were validated against sonomicrometry by means of Bland-Altman analysis and the intraclass correlation coefficient. For both strain components, good agreements were found between the ultrasound and the sonomicrometry measurements as shown by Bland-Altman statistics. The intraclass correlation coefficients were found to be 0.72 and 0.80 for the radial and longitudinal components, respectively.
Conclusions A new technique for the estimation of myocardial deformation was validated. It was shown that the current problem of angle dependency was solved and that 2 deformation components could be estimated simultaneously and accurately. Furthermore, the technique was less time-consuming, because anatomic tracking was performed automatically. This approach could potentially accelerate the clinical acceptance of ultrasound deformation imaging in cardiology.
Key Words: echocardiography strain angle dependency
| Introduction |
|---|
|
|
|---|
As a solution to these problems, a new method for deformation imaging has been developed in our laboratory, called 2D strain echocardiography. The method not only allows simultaneous analysis of radial and longitudinal myocardial deformation, independent of insonation angle, but also avoids the time-consuming step of manually tracking a myocardial segment. The technique was initially shown to be applicable to the normal heart in vivo,9 has been further optimized after a series of simulations,10 and was recently validated in tissue-mimicking phantoms.11 The aim of the present study was to further validate this new 2D approach in an in vivo setup by comparing the deformation of the sheep heart as measured by this new methodology with the deformation measured by sonomicrometry.
| Methods |
|---|
|
|
|---|
For sonomicrometry measurements, a Sonometrics Digital Ultrasonic Measurement System (Sonometrics Corporation) was used. Three segment-length ultrasound crystals (2 mm) were fixed to the myocardium in the inferolateral wall. Two crystals were sutured to the epicardium in a longitudinal direction (crystals 1 and 2 in Figure 1). One crystal (crystal 3 in Figure 1) was placed subendocardially just radially to crystal 2 through the myocardial wall. The latter was introduced in an oblique way to avoid damage to the myocardium to be studied. They enabled us to continuously measure wall thickening and longitudinal shortening at a time resolution of 1.7 ms and provided a continuous reference for the longitudinal and radial strain.
|
Data Acquisition
A Toshiba PowerVision 6000 equipped with a radiofrequency (RF) interface for research purposes was used to acquire ultrasonic RF data in a 45° sector format with a 5-MHz cardiac transducer (PSM-50AT) by placing the transducer directly on the contralateral epicardium. Parasternal long-axislike images were acquired at a frame rate of 167 Hz. The acquired RF data were stored and transferred to a personal computer for further postprocessing. Owing to overlapping frequency bands for the microcrystal system and the ultrasound system, the microcrystal data were acquired immediately before the ultrasound data and then switched off while the ultrasound data were acquired.
To change the range of strain values, myocardial strain was modulated over 4 stages with differing challenges. After the baseline acquisitions, the inotropic state was reduced by an esmolol infusion and increased by a dobutamine infusion, and finally, ischemia was induced in the region under investigation by ligating a distal branch of the circumflex coronary artery. For the esmolol and dobutamine stages, a physiological target was set to a 50% reduction in dP/dt and a 100% increase in dP/dt relative to baseline for esmolol and dobutamine, respectively. For esmolol, this was achieved at an infusion rate of
100 µg · kg1 · min1, whereas the dobutamine infusion rate was
2.5 µg · kg1 · min1. In this way, 4 data sets were obtained in each animal.
Deformation Estimation
The method is based on estimating the velocity vector rather than the velocity component along the image line alone, as is currently the case for Doppler estimation. This can be obtained by a new methodology developed in our laboratory based on 2D tracking of RF image patterns.911 Briefly, the algorithm finds the velocity vector by tracking patterns in the RF image between consecutive frames, as illustrated in Figure 2. For each pixel in the image, angle-independent velocity estimation is performed by selecting a search pattern around that pixel in 1 frame (eg, "search pattern" in Figure 2) and looking for a matching pattern in a search region around that pixel of the following frame ("search region" in Figure 2). The search pattern is placed at different positions in the search region, and the similarity is measured for the overlapping area. The position where the highest similarity is found (illustrated as a solid rectangle in the "solution" image in Figure 2) determines the in-plane frame-to-frame displacement relative to its initial position (illustrated as a dashed rectangle in the "solution" image in Figure 2). For a more detailed description of this methodology, see Langeland et al.11
|
From the velocity vector field, radial and longitudinal strain estimates were obtained in the following way: at the beginning of the heart cycle, the endocardial and epicardial borders of the region of interest were indicated on the conventional B-mode images reconstructed from the RF data. Then, a fine grid was automatically positioned within the selected region of interest with an intergrid point distance of 0.5 mm. For each frame, the position of each grid point was updated on the basis of the underlying velocity vector. In this way, the deformation of the grid followed the deformation of the underlying myocardium, and the radial and longitudinal strain could be estimated from the distance changes between grid points as a function of time. To have robust measurements, the whole width (10 mm) and length (10 mm) of the region of interest were used to calculate the radial and longitudinal strain, respectively. This gave an effective spatial resolution of 10 mm for both strain components. Because the method makes use of RF data, it is intrinsically independent of gain settings.
One heart cycle, defined as 1 RR interval on a synchronously recorded ECG, was analyzed for each data set. End systole was defined as aortic valve closure, determined as dP/dtmin20 ms.12 All postprocessing was performed with software (SPEQLE 2D) developed in our laboratory as an extension to existing custom-made software (SPEQLE).13
Statistical Analysis
For comparison of the strain curves throughout the cardiac cycle, 10 strain values were extracted during systole and diastole (20 values in total) at equidistant time points for both measurement methods. As recommended by Bland and Altman,14 the agreement between the methods was assessed by calculating the paired difference between the 2 methods for each measurement and by estimating the bias and 95% limits of agreement relative to the mean measurement of both methods. As appropriate for continuously changing variables, the mean value and limits of agreement were found with methods that accounted for such replicated measurements.15 The goal was to achieve an agreement equal to or better than the agreement found in a similar validation of the more conventional Doppler-based technique.3 The requirement of normal distribution of the differences and uniform variability was verified graphically.
In addition to the Bland-Altman statistics, the intraclass correlation was calculated as a measure of consistency between the 2 methods. The variance between sheep (
s), the variance between the inotropic stages (
i, conditional on sheep), the variance between the 20 different time points (
t, conditional on sheep and stage), and the residual error, ie, variance between the 2 methods (
e, conditional on sheep, stage, and time point) were found with repeated-measures ANOVA that included stage, time point, and their interaction as fixed effects. The intraclass correlation coefficient (ICC) was then found as: ICC=(
s2+
i2+
t2) / (
s2+
i2+
t2+
e2), where s indicates stage, i indicate inotropic stage, t indicates time point, and e indicates residual error.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Strong correlations were found between the 2 methods for both the radial and longitudinal strain components. Both components could thus be estimated with the same accuracy, which implies the angle dependency problem has been solved. This is also illustrated in Figures 3 and 4
, in which Doppler-derived strain is markedly underestimated owing to angulation (see Figure 1). Moreover, because both components are estimated simultaneously, this could open new possibilities for investigation of the different response in radial and longitudinal function. Because subendocardial myocardial fibers are more susceptible to ischemia, it might be expected that longitudinal function is altered earlier than radial function,16 because subendocardial fibers are oriented longitudinally.17 This could have interesting clinical implications in quantifying inducible ischemia during stress echocardiography. Currently, stress echocardiography is performed by visually interpreting the motion and deformation of the myocardial walls in 2D gray-scale images. This makes stress echocardiography subjective, and a high level of expertise is required to obtain correct readings.18 In particular, the timing of the deformation is difficult to assess.19 For this reason, deformation imaging has been suggested as a tool to improve the quantification of stress echocardiography.2022 A problem with the current methodology is that images from several transducer positions have to be used to assess the different deformation components. With the new methodology, only 1 transducer position is needed, because 2 deformation components can be obtained from the same data set.
As can be seen in Figures 5 and 6
, good agreement was found in the Bland-Altman analysis. The longitudinal component (eg, Figure 6) showed a better agreement than the radial component (eg, Figure 5); however, both were within clinically acceptable limits.3 From the analysis of the radial strain (eg, Figure 5), it can be seen that the values were overestimated, because the bias was found to be 2.0%. This might be due to potential errors in the reference measurements. Indeed, to keep good contact between the subendocardial microcrystal (crystal 2 in Figure 1) and the myocardium throughout the cardiac cycle, the microcrystal could not be positioned perfectly subendocardially but rather had to be placed a bit more toward the mid-myocardial layer. Given that the highest strain occurs subendocardially, the strain measured by the microcrystals might underestimate the radial strain.
The radial strain values as measured in the present study were found to be low compared with what has been found in other SRI studies.3,23 The value of radial thickening could be dependent on species and furthermore be influenced by the fact that an open-chest model was used in the present study. The aim of the present study was merely to validate the new ultrasound methodology, and thus, this was not considered a limitation, because good agreement with the microcrystal data was found to be most important.
Although interesting findings based on the current 1D SRI method have been presented in the literature, its migration into daily clinical practice has been tempered because the current method is relatively time-consuming owing to the fact that the region of interest has to be tracked manually over the cardiac cycle. It is also essential that the myocardial wall be well aligned with the ultrasonic beam, which means that a high level of expertise is required for the data acquisition and analysis. The new approach proposed here solves both problems, because the tracking is intrinsic to the method, and no alignment of the wall with the ultrasound beam is needed. We thus believe that an angle-independent method should be an important factor in accelerating the clinical acceptance of ultrasonic deformation imaging in cardiology.
The problem of temporal drifting of strain curves, as reported in earlier work,3,24 is also present in this new methodology. As in other methods, drifting is compensated for by forcing all grid points to return to their initial positions at the end of the heart cycle. This drifting could be part of the cause of the discrepancy between the 2 methods seen in Figure 4.
The ultrasound data in the present study had a frame rate of 168 Hz. This is comparable to the current SRI methodology for images at a similar sector angle. The new methodology provides a temporal resolution similar to the current SRI technique. A high temporal resolution is required to resolve the timing of the strain curves, especially for the application in stress echocardiography.
Study Limitations
In the present study, only the longitudinal and radial strain components were assessed. However, by acquiring a parasternal short-axis view, both circumferential and radial deformation can be measured. Currently, several companies are introducing 3D ultrasound imaging for cardiac applications. The method validated in the present study can easily be modified to estimate deformation in 3 dimensions, independent of angle, given an appropriate acquisition. This will allow for the estimation of radial, longitudinal, and circumferential strain from single data sets.
Because regular tissue Doppler data were not acquired in the present study, no direct comparison between Doppler-based strain imaging and this new methodology was performed. Therefore, the superiority of the new methodology over the Doppler-based method remains to be demonstrated.
Although rotation and shearing were not explicitly validated in the present study (because these fell outside the scope of this work), it would be expected that these could be measured given prior data reported in the literature25 with similar methodologies. These estimates have already been shown to have potential clinical applications.26,27
The accuracy of radial and longitudinal strain could be dependent on the relative position of the transducer to the cardiac segment under investigation. In the present study, no attempt was made to compare different transducer positions.
Conclusions
A new technique for the estimation of myocardial deformation has been validated. It was shown that the current problem of angle dependency was solved and that 2 components of the deformation can be estimated simultaneously. Furthermore, the technique is less time-consuming because anatomic tracking is performed automatically. This could potentially accelerate the clinical acceptance of deformation imaging in cardiology.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
2. Sutherland GR, Di Salvo G, Claus P, Dhooge J, Bijnens B. Strain and strain rate imaging: a new clinical approach to quantifying regional myocardial function. J Am Soc Echocardiogr. 2004; 17: 788802.[CrossRef][Medline] [Order article via Infotrieve]
3. Urheim S, Edvardsen T, Torp H, Angelsen B, Smiseth OA. Myocardial strain by Doppler echocardiography: validation of a new method to quantify regional myocardial function. Circulation. 2000; 102: 11581164.
4. Jamal F, Bergerot C, Argaud L, Loufouat J, Ovize M. Longitudinal strain quantitates regional right ventricular contractile function. Am J Physiol Heart Circ Physiol. 2003; 285: 28422847.
5. Edvardsen T, Gerber BL, Garot J, Bluemke DA, Lima JAC, Smiseth OA. Quantitative assessment of intrinsic regional myocardial deformation by Doppler strain rate echocardiography in humans. Circulation. 2002; 106: 5056.
6. Herbots L, Maes F, Dhooge J, Claus P, Dymarkowski S, Mertens P, Mortelmans L, Bijnens B, Bogaert J, Rademakers FE, Sutherland GR. Quantifying myocardial deformation throughout the cardiac cycle: a comparison of ultrasound strain rate, grey-scale M-mode and magnetic resonance imaging. Ultrasound Med Biol. 2004; 30: 591598.[CrossRef][Medline] [Order article via Infotrieve]
7. Castro PL, Greenberg NL, Drinko J, Garcia MJ, Thomas JD. Potential pitfalls of strain rate imaging: angle dependency. Biomed Sci Instrum. 2000; 36: 197202.[Medline] [Order article via Infotrieve]
8. Simmons LA, Weidemann F, Sutherland GR, Dhooge J, Bijnens B, Sergeant P, Wouters PF. Doppler tissue velocity, strain, and strain rate imaging with transesophageal echocardiography in the operating room: a feasibility study. J Am Soc Echocardiogr. 2002; 15: 768776.[CrossRef][Medline] [Order article via Infotrieve]
9. Dhooge J, Konofagou E, Jamal F, Heimdal A, Barrios L, Bijnens B, Thoen J, Van de Werf F, Sutherland GR, Suetens P. Two-dimensional myocardial strain rate measurements of the human heart in vivo. IEEE Trans Ultrason Ferroelectr Freq Control. 2002; 49: 281286.[CrossRef][Medline] [Order article via Infotrieve]
10. Langeland S, Dhooge J, Torp H, Bijnens B, Suetens P. Comparison of time-domain displacement estimators for two-dimensional RF tracking. Ultrasound Med Biol. 2003; 29: 11771186.[CrossRef][Medline] [Order article via Infotrieve]
11. Langeland S, Dhooge J, Claessens T, Claus P, Verdonck P, Suetens P, Sutherland GR, Bijnens B. RF-based two-dimensional cardiac strain estimation: a validation study in a tissue mimicking phantom. IEEE Trans on Ultrason Ferroelectr Freq Control. 2004; 51: 15371546.[CrossRef]
12. Theroux P, Ross J Jr, Franklin D, Kemper WS, Sasyama S. Regional myocardial function in the conscious dog during acute coronary occlusion and responses to morphine, propranolol, nitroglycerin, and lidocaine. Circulation. 1976; 53: 302314.
13. Claus P, Dhooge J, Langeland T, Bijnens B, Sutherland G. SPEQLE (Software Package for Echocardiographic Quantification Leuven): an integrated approach to ultrasound-based cardiac deformation quantification. IEEE Comput Cardiol. 2002; 29: 6972.
14. Altman DG, Bland JM. Measurements in medicine: the analysis of method comparison studies. Statistician. 1983; 32: 307317.[CrossRef]
15. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999; 8: 135160.
16. Jones CJ, Raposo L, Gibson DG. Functional importance of the long axis dynamics of the human left ventricle. Br Heart J. 1990; 63: 215220.
17. Greenbaum RA, Ho SY, Gibson DG, Becker AE, Anderson RH. Left ventricular fibre architecture in man. Br Heart J. 1981; 45: 248263.
18. Picano E, Lattanzi F, Orlandini A, Marini C, LAbbate A. Stress echocardiography and the human factor: the importance of being expert. J Am Coll Cardiol. 1991; 17: 666669.[Abstract]
19. Kvitting JP, Wigstrom L, Strotmann JM, Sutherland GR. How accurate is visual assessment of synchronicity in myocardial motion? An in vitro study with computer-simulated regional delay in myocardial motion: clinical implications for rest and stress echocardiography studies. J Am Soc Echocardiogr. 1999; 12: 698705.[CrossRef][Medline] [Order article via Infotrieve]
20. Jamal F, Strotmann J, Weidemann F, Kukulski T, Dhooge J, Bijnens B, Van de Werf F, De Scheerder I, Sutherland GR. Noninvasive quantification of the contractile reserve of stunned myocardium by ultrasonic strain rate and strain. Circulation. 2001; 104: 10591065.
21. Weidemann F, Dommke C, Bijnens B, Claus P, Dhooge J, Mertens P, Verbeken E, Maes A, Van de Werf F, De Scheerder I, Sutherland GR. Defining the transmurality of a chronic myocardial infarction by ultrasonic strain-rate imaging: implications for identifying intramural viability: an experimental study. Circulation. 2003; 107: 883888.
22. Voigt JU, Exner B, Schmiedehausen K, Huchzermeyer C, Reulbach U, Nixdorff U, Platsch G, Kuwert T, Daniel WG, Flachskampf FA. Strain-rate imaging during dobutamine stress echocardiography provides objective evidence of inducible ischemia. Circulation. 2003; 107: 21202126.
23. Kowalski M, Kukulski T, Jamal F, Dhooge J, Weidemann F, Rademakers F, Bijnens B, Hatle L, Sutherland GR. Can natural strain and strain rate quantify regional myocardial deformation? A study in healthy subjects. Ultrasound Med Biol. 2001; 27: 10871097.[CrossRef][Medline] [Order article via Infotrieve]
24. Dhooge J, Heimdal A, Jamal F, Kukulski T, Bijnens B, Rademakers F, Hatle L, Suetens P, Sutherland GR. Regional strain and strain rate measurements by cardiac ultrasound: principles, implementation and limitations. Eur J Echocardiogr. 2000; 1: 154170.
25. Konofagou EE, Ophir J. A new elastographic method for estimation and imaging of lateral displacements, lateral strains, corrected axial strains and Poissons ratios in tissues. Ultrasound Med Biol. 1998; 24: 11831199.[CrossRef][Medline] [Order article via Infotrieve]
26. Garot J, Pascal O, Diebold B, Derumeaux G, Gerber BL, Dubois-Rande JL, Lima JA, Gueret P. Alterations of systolic left ventricular twist after acute myocardial infarction. Am J Physiol Heart Circ Physiol. 2002; 282: 357362.
27. Separovic J, Claus P, Herbots L, Dhooge J, Rademakers FE, Herregods MC, Bijnens B, Sutherland GR. Assessment of circumferential-radial shear strains in normal and ischaemic myocardium based on ultrasonic strain/strain rate imaging. Eur Heart J. 2002; 23 (suppl): P2755. Abstract.
This article has been cited by other articles:
![]() |
D. Oxborough, A. M. Batterham, R. Shave, N. Artis, K. M. Birch, G. Whyte, P. N. Ainslie, and K. P. George Interpretation of two-dimensional and tissue Doppler-derived strain ({varepsilon}) and strain rate data: is there a need to normalize for individual variability in left ventricular morphology? Eur J Echocardiogr, July 1, 2009; 10(5): 677 - 682. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Becker, C. Ocklenburg, E. Altiok, A. Futing, J. Balzer, G. Krombach, M. Lysyansky, H. Kuhl, R. Krings, M. Kelm, et al. Impact of infarct transmurality on layer-specific impairment of myocardial function: a myocardial deformation imaging study Eur. Heart J., June 2, 2009; 30(12): 1467 - 1476. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang and S. F. Nagueh Current Perspectives on Cardiac Function in Patients With Diastolic Heart Failure Circulation, March 3, 2009; 119(8): 1146 - 1157. [Full Text] [PDF] |
||||
![]() |
B. H. Amundsen, J. Crosby, P. A. Steen, H. Torp, S. A. Slordahl, and A. Stoylen Regional myocardial long-axis strain and strain rate measured by different tissue Doppler and speckle tracking echocardiography methods: a comparison with tagged magnetic resonance imaging Eur J Echocardiogr, March 1, 2009; 10(2): 229 - 237. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Adamu, F. Schmitz, M. Becker, M. Kelm, and R. Hoffmann Advanced speckle tracking echocardiography allowing a three-myocardial layer-specific analysis of deformation parameters Eur J Echocardiogr, March 1, 2009; 10(2): 303 - 308. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. F. Nagueh, C. P. Appleton, T. C. Gillebert, P. N. Marino, J. K. Oh, O. A. Smiseth, A. D. Waggoner, F. A. Flachskampf, P. A. Pellikka, and A. Evangelisa Recommendations for the Evaluation of Left Ventricular Diastolic Function by Echocardiography Eur J Echocardiogr, March 1, 2009; 10(2): 165 - 193. [Full Text] [PDF] |
||||
![]() |
T. P. Abraham and H.-Y. Liang Stress echocardiography diastole to the rescue. J. Am. Coll. Cardiol., February 24, 2009; 53(8): 706 - 708. [Full Text] [PDF] |
||||
![]() |
B. Sjoli, S. Orn, B. Grenne, H. Ihlen, T. Edvardsen, and H. Brunvand Diagnostic capability and reproducibility of strain by Doppler and by speckle tracking in patients with acute myocardial infarction. J. Am. Coll. Cardiol. Img., January 1, 2009; 2(1): 24 - 33. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Jarnert, A. Melcher, K. Caidahl, H. Persson, L. Ryden, and M. J. Eriksson Left atrial velocity vector imaging for the detection and quantification of left ventricular diastolic function in type 2 diabetes Eur J Heart Fail, November 1, 2008; 10(11): 1080 - 1087. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Perry, C. G. De Pasquale, D. P. Chew, and M. X. Joseph Assessment of early diastolic left ventricular function by two-dimensional echocardiographic speckle tracking Eur J Echocardiogr, November 1, 2008; 9(6): 791 - 795. [Abstract] [Full Text] [PDF] |
||||
![]() |
A D'Andrea, G De Corato, R Scarafile, S Romano, L Reigler, C Mita, F Allocca, G Limongelli, G Gigantino, B Liccardo, et al. Left atrial myocardial function in either physiological or pathological left ventricular hypertrophy: a two-dimensional speckle strain study Br. J. Sports Med., August 1, 2008; 42(8): 696 - 702. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, D. S. Khoury, Y. Yue, G. Torre-Amione, and S. F. Nagueh Preserved left ventricular twist and circumferential deformation, but depressed longitudinal and radial deformation in patients with diastolic heart failure Eur. Heart J., May 2, 2008; 29(10): 1283 - 1289. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Becker, A. Lenzen, C. Ocklenburg, K. Stempel, H. Kuhl, M. Neizel, M. Katoh, R. Kramann, J. Wildberger, M. Kelm, et al. Myocardial Deformation Imaging Based on Ultrasonic Pixel Tracking to Identify Reversible Myocardial Dysfunction J. Am. Coll. Cardiol., April 15, 2008; 51(15): 1473 - 1481. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Masuda, T. Asanuma, A. Taniguchi, A. Uranishi, F. Ishikura, and S. Beppu Assessment of dyssynchronous wall motion during acute myocardial ischemia using velocity vector imaging. J. Am. Coll. Cardiol. Img., March 1, 2008; 1(2): 210 - 220. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Pirat, D. S. Khoury, C. J. Hartley, L. Tiller, L. Rao, D. G. Schulz, S. F. Nagueh, and W. A. Zoghbi A novel feature-tracking echocardiographic method for the quantitation of regional myocardial function: validation in an animal model of ischemia-reperfusion. J. Am. Coll. Cardiol., February 12, 2008; 51(6): 651 - 659. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Bordachar, L. Labrousse, J.-B. Thambo, P. Reant, S. Lafitte, M. D. O'Neill, P. Jais, M. Haissaguerre, J. Clementy, and P. Dos Santos Haemodynamic impact of the left ventricular pacing site during graded ischaemia in an open-chest pig model Europace, February 1, 2008; 10(2): 242 - 248. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Reant, L. Labrousse, S. Lafitte, P. Bordachar, X. Pillois, L. Tariosse, S. Bonoron-Adele, P. Padois, C. Deville, R. Roudaut, et al. Experimental Validation of Circumferential, Longitudinal, and Radial 2-Dimensional Strain During Dobutamine Stress Echocardiography in Ischemic Conditions J. Am. Coll. Cardiol., January 15, 2008; 51(2): 149 - 157. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Abraham and A. C. Pinheiro Speckle-Derived Strain: A Better Tool for Quantification of Stress Echocardiography? J. Am. Coll. Cardiol., January 15, 2008; 51(2): 158 - 160. [Full Text] [PDF] |
||||
![]() |
T. P. Abraham, V. L. Dimaano, and H.-Y. Liang Role of Tissue Doppler and Strain Echocardiography in Current Clinical Practice Circulation, November 27, 2007; 116(22): 2597 - 2609. [Full Text] [PDF] |
||||
![]() |
A. D'Andrea, P. Caso, S. Romano, R. Scarafile, L. Riegler, G. Salerno, G. Limongelli, G. Di Salvo, P. Calabro, L. Del Viscovo, et al. Different effects of cardiac resynchronization therapy on left atrial function in patients with either idiopathic or ischaemic dilated cardiomyopathy: a two-dimensional speckle strain study Eur. Heart J., November 2, 2007; 28(22): 2738 - 2748. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Song, W. Chang, S. Lim, H.-S. Seo, C. Y. Shim, S. Park, K.-J. Yoo, B.-S. Kim, B.-H. Min, H. Lee, et al. Tissue Transglutaminase Is Essential for Integrin-Mediated Survival of Bone Marrow-Derived Mesenchymal Stem Cells Stem Cells, June 1, 2007; 25(6): 1431 - 1438. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, D. S. Khoury, V. Thohan, G. Torre-Amione, and S. F. Nagueh Global Diastolic Strain Rate for the Assessment of Left Ventricular Relaxation and Filling Pressures Circulation, March 20, 2007; 115(11): 1376 - 1383. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Becker, R. Hoffmann, H. P. Kuhl, H. Grawe, M. Katoh, R. Kramann, A. Bucker, P. Hanrath, and N. Heussen Analysis of myocardial deformation based on ultrasonic pixel tracking to determine transmurality in chronic myocardial infarction Eur. Heart J., November 1, 2006; 27(21): 2560 - 2566. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ashraf, X. K. Li, M. T. Young, A. J. Jensen, J. Pemberton, L. Hui, P. Lysyansky, Z. Friedman, B. Park, and D. J. Sahn Delineation of cardiac twist by a sonographically based 2-dimensional strain analysis method: an in vitro validation study. J. Ultrasound Med., September 1, 2006; 25(9): 1193 - 1198. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. W.L. De Boeck, M.-J. M. Cramer, P. Loh, and P. A.F.M. Doevendans Two-Dimensional Strain Imaging to Assess the Origin and Extent of Ventricular Preexcitation Associated With an Accessory Bypass Circulation, June 6, 2006; 113(22): e835 - e839. [Full Text] [PDF] |
||||
![]() |
A. E. Weyman The Year in Echocardiography J. Am. Coll. Cardiol., February 21, 2006; 47(4): 856 - 863. [Full Text] [PDF] |
||||
![]() |
M. S. Suffoletto, K. Dohi, M. Cannesson, S. Saba, and J. Gorcsan III Novel Speckle-Tracking Radial Strain From Routine Black-and-White Echocardiographic Images to Quantify Dyssynchrony and Predict Response to Cardiac Resynchronization Therapy Circulation, February 21, 2006; 113(7): 960 - 968. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2005 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |