2005 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care |
| Introduction |
|---|
| Defibrillation Plus CPR: A Critical Combination |
|---|
Several studies have documented the effects of time to defibrillation and the effects of bystander CPR on survival from SCA. For every minute that passes between collapse and defibrillation, survival rates from witnessed VF SCA decrease 7% to 10% if no CPR is provided.1 When bystander CPR is provided, the decrease in survival rates is more gradual and averages 3% to 4% per minute from collapse to defibrillation.1,2 CPR can double13 or triple4 survival from witnessed SCA at most intervals to defibrillation.
If bystanders provide immediate CPR, many adults in VF can survive with intact neurologic function, especially if defibrillation is performed within about 5 minutes after SCA.5,6 CPR prolongs VF79 (ie, the window of time during which defibrillation can occur) and provides a small amount of blood flow that may maintain some oxygen and substrate delivery to the heart and brain.10 Basic CPR alone, however, is unlikely to eliminate VF and restore a perfusing rhythm.
New Recommendations to Integrate CPR and AED Use
To treat VF SCA, rescuers must be able to rapidly integrate CPR with use of the AED. To give the victim the best chance of survival, 3 actions must occur within the first moments of a cardiac arrest: (1) activation of the emergency medical services (EMS) system or emergency medical response system, (2) provision of CPR, and (3) operation of an AED. When 2 or more rescuers are present, activation of EMS and initiation of CPR can occur simultaneously.
Delays to either start of CPR or defibrillation can reduce survival from SCA. In the 1990s some predicted that CPR could be rendered obsolete by the widespread development of community AED programs. Cobb6 noted, however, that as more Seattle first responders were equipped with AEDs, survival rates from SCA unexpectedly fell. He attributed this decline to reduced emphasis on CPR, and there is growing evidence to support this view. Part 4: "Adult Basic Life Support" summarizes the evidence on the importance of effective chest compressions and minimizing interruptions in providing compressions.
Two critical questions about integration of CPR with defibrillation were evaluated during the 2005 Consensus Conference.11 The first question concerns whether CPR should be provided before defibrillation is attempted. The second question concerns the number of shocks to be delivered in a sequence before the rescuer resumes CPR.
Shock First Versus CPR First
When any rescuer witnesses an out-of-hospital arrest and an AED is immediately available on-site, the rescuer should use the AED as soon as possible. Healthcare providers who treat cardiac arrest in hospitals and other facilities with AEDs on-site should provide immediate CPR and should use the AED/defibrillator as soon as it is available. These recommendations are designed to support early CPR and early defibrillation, particularly when an AED is available within moments of the onset of SCA.
When an out-of-hospital cardiac arrest is not witnessed by EMS personnel, they may give about 5 cycles of CPR before checking the ECG rhythm and attempting defibrillation (Class IIb). One cycle of CPR consists of 30 compressions and 2 breaths. When compressions are delivered at a rate of about 100 per minute, 5 cycles of CPR should take roughly 2 minutes (range: about 1
to 3 minutes). This recommendation regarding CPR prior to attempted defibrillation is supported by 2 clinical studies (LOE 25; LOE 36) of adult out-of-hospital VF SCA. In those studies when EMS call-to-arrival intervals were 46 to 55 minutes or longer, victims who received 1
to 3 minutes of CPR before defibrillation showed an increased rate of initial resuscitation, survival to hospital discharge,5,6 and 1-year survival5 when compared with those who received immediate defibrillation for VF SCA. One randomized study,12 however, found no benefit to CPR before defibrillation for non-paramedic-witnessed SCA.
EMS system medical directors may consider implementing a protocol that would allow EMS responders to provide about 5 cycles (about 2 minutes) of CPR before defibrillation of patients found by EMS personnel to be in VF, particularly when the EMS system call-to-response interval is >4 to 5 minutes. There is insufficient evidence to support or refute CPR before defibrillation for in-hospital cardiac arrest.
1-Shock Protocol Versus 3-Shock Sequence
At the time of the 2005 Consensus Conference, no published human or animal studies were found that compared a 1-shock protocol with a 3-stacked shock protocol for treatment of VF cardiac arrest. In animal studies, however, frequent or long interruptions in precordial chest compressions for rhythm analysis13 or rescue breathing14,15 were associated with post-resuscitation myocardial dysfunction and reduced survival rates. Secondary analyses of 2 randomized trials16,17 showed that interruption in chest compressions is associated with a decreased probability of conversion of VF to another rhythm. In 2 recent clinical observational studies (LOE 4) of out-of-hospital18 and in-hospital19 CPR by healthcare providers, chest compressions were performed only 51%18 to 76%19 of total CPR time.
In 2005 the rhythm analysis for a 3-shock sequence performed by commercially available AEDs resulted in delays of up to 37 seconds between delivery of the first shock and delivery of the first post-shock compression.13 This delay is difficult to justify in light of the first-shock efficacy of >90% reported by current biphasic defibrillators.2025 If 1 shock fails to eliminate VF, the incremental benefit of another shock is low, and resumption of CPR is likely to confer a greater value than another shock. This fact, combined with the data from animal studies documenting harmful effects from interruptions to chest compressions, suggests that a 1-shock scenario plus immediate CPR is reasonable.
When VF/pulseless ventricular tachycardia (VT) is present, the rescuer should deliver 1 shock and should then immediately resume CPR, beginning with chest compressions (Class IIa). The rescuer should not delay resumption of chest compressions to recheck the rhythm or pulse. After 5 cycles (about 2 minutes) of CPR, the AED should then analyze the cardiac rhythm and deliver another shock if indicated (Class IIb). If a nonshockable rhythm is detected, the AED should instruct the rescuer to resume CPR immediately, beginning with chest compressions (Class IIb). Concern that chest compressions might provoke recurrent VF in the presence of a post-shock organized rhythm does not appear to be warranted.25
AED voice prompts should not instruct the lay user to reassess the patient at any time. AED manufacturers should seek innovative methods to decrease the amount of time chest compressions are withheld for AED operation. Training materials for lay rescuers should emphasize the importance of continued CPR until basic or advanced life support personnel take over CPR or the victim begins to move.
First-shock efficacy for monophasic shocks is lower than first-shock efficacy for biphasic shocks.17,26,27 Although the optimal energy level for defibrillation using any of the monophasic or biphasic waveforms has not been determined, a recommendation for higher initial energy when using a monophasic waveform was weighed by expert consensus with consideration of the potential negative effects of a high first-shock energy versus the negative effects of prolonged VF. The consensus was that rescuers using monophasic AEDs should give an initial shock of 360 J; if VF persists after the first shock, second and subsequent shocks of 360 J should be given. This single dose for monophasic shocks is designed to simplify instructions to rescuers but is not a mandate to recall monophasic AEDs for reprogramming. If the monophasic AED being used is programmed to deliver a different first or subsequent dose, that dose is acceptable.
One study compared the effectiveness of 175 J versus 320 J monophasic waveform shocks for out-of-hospital VF cardiac arrest.28 Approximately 61% of patients who received shocks with either 175 J or 320 J monophasic damped sine waveform were defibrillated with the first shock, which was delivered an average of 10.6 minutes after the call to EMS. There was no significant difference in the percentage of patients who developed advanced atrioventricular (AV) block after 1 shock. AV block was more likely to develop after 2 or 3 shocks of 320 J than after 2 or 3 shocks of 175 J, but the block was transient and did not affect survival to hospital discharge.28
Healthcare providers must practice efficient coordination between CPR and defibrillation. When VF is present for more than a few minutes, the myocardium is depleted of oxygen and metabolic substrates. A brief period of chest compressions can deliver oxygen and energy substrates, increasing the likelihood that a perfusing rhythm will return after defibrillation (elimination of VF).29 Analyses of VF waveform characteristics predictive of shock success have documented that the shorter the time between a chest compression and delivery of a shock, the more likely the shock will be successful.29,30 Reduction in the interval from compression to shock delivery by even a few seconds can increase the probability of shock success.16
The rescuer providing chest compressions should minimize interruptions in chest compressions for rhythm analysis and shock delivery and should be prepared to resume CPR, beginning with chest compressions, as soon as a shock is delivered. When 2 rescuers are present, the rescuer operating the AED should be prepared to deliver a shock as soon as the compressor removes his or her hands from the victims chest and all rescuers are "clear" of contact with the victim. The lone rescuer should practice coordination of CPR with efficient AED operation.
| Defibrillation Waveforms and Energy Levels |
|---|
Shock success using the typical definition of defibrillation should not be confused with resuscitation outcomes such as restoration of a perfusing rhythm, survival to hospital admission, or survival to hospital discharge.31,33 Although resuscitation outcomes including survival may be affected by many variables in addition to shock delivery, defibrillation programs must strive to improve patient survival, not just shock success.
Modern defibrillators are classified according to 2 types of waveforms: monophasic and biphasic. Monophasic waveform defibrillators were introduced first, but biphasic waveforms are used in almost all AEDs and manual defibrillators sold today. Energy levels vary by type of device. No specific waveform (either monophasic or biphasic) is consistently associated with a higher rate of return of spontaneous circulation (ROSC) or rates of survival to hospital discharge after cardiac arrest.
Monophasic Waveform Defibrillators
Monophasic waveforms deliver current of one polarity (ie, direction of current flow). Monophasic waveforms can be further categorized by the rate at which the current pulse decreases to zero. The monophasic damped sinusoidal waveform (MDS) returns to zero gradually, whereas the monophasic truncated exponential waveform (MTE) current is abruptly returned to baseline (truncated) to zero current flow.
Few monophasic waveform defibrillators are being manufactured but many are still in use. Most of these use MDS waveforms. As noted above, no specific waveform (either monophasic or biphasic) is consistently associated with a greater incidence of ROSC or survival to hospital discharge rates after cardiac arrest than any other specific waveform. Research indicates, however, that when doses equivalent to or lower than monophasic doses are used, biphasic waveform shocks are safe and effective for termination of VF.
Biphasic Waveform Defibrillators
Researchers have collected data from both out-of-hospital3436 and in-hospital studies (electrophysiologic studies and implantable cardioverter-defibrillator [ICD] testing and evaluation).37 Overall this research indicates that lower-energy biphasic waveform shocks have equivalent or higher success for termination of VF than either damped sinusoidal or truncated exponential monophasic waveform shocks delivering escalating energy (200 J, 300 J, 360 J) with successive shocks. No direct comparison of the different biphasic waveforms has been made.
The optimal energy for first-shock biphasic waveform defibrillation yielding the highest termination rate for VF has not been determined. Several randomized (LOE 2)17,24,27 and observational studies (LOE 5)26,38 have shown that defibrillation with biphasic waveforms of relatively low energy (
200 J) is safe and has equivalent or higher efficacy for termination of VF than monophasic waveform shocks of equivalent or higher energy (Class IIa).32,3941
Compensation for patient-to-patient differences in impedance may be achieved by changes in duration and voltage of shocks or by releasing the residual membrane charge (called burping). Whether there is an optimal ratio of first-phase to second-phase duration and leading-edge amplitude is unclear. It is unknown whether a waveform more effective for immediate outcomes (defibrillation) and short-term outcomes (ROSC, survival to hospital admission) results in better long-term outcomes (survival to hospital discharge, survival for 1 year). Given the high efficacy of all biphasic waveforms, other determinants of survival (eg, interval from collapse to CPR or defibrillation) are likely to supersede the impact of specific biphasic waveforms or energies.
Fixed and Escalating Energy
Commercially available biphasic AEDs provide either fixed or escalating energy levels.
Multiple prospective human clinical studies (LOE 2)27,42 and retrospective17,24,26,38,43,44 studies have failed to identify an optimal biphasic energy level for first or subsequent shocks. Therefore, it is not possible to make a definitive recommendation for the selected energy for the first or subsequent biphasic defibrillation attempts.
Biphasic defibrillators use one of two waveforms, and each waveform has been shown to be effective in terminating VF over a specific dose range. The ideal shock dose for a biphasic device is one that falls within the range that has been documented to be effective using that specific device. Current research confirms that it is reasonable to use selected energies of 150 J to 200 J with a biphasic truncated exponential waveform or 120 J with a rectilinear biphasic waveform for the initial shock. For second and subsequent biphasic shocks, use the same or higher energy (Class IIa). In this context "selected" refers to the energy dose selected by the operator (or programmed by the AED manufacturer). With the rectilinear biphasic waveform device, selected and delivered energies usually differ; delivered energy is typically higher in the usual range of impedance. For example, in a patient with 80
impedance, a selected energy of 120 J will deliver 150 J.
None of the available evidence has shown superiority of either nonescalating or escalating energy biphasic waveform defibrillation for termination of VF. Nonescalating and escalating energy biphasic waveform shocks can be used safely and effectively to terminate short-duration and long-duration VF (Class IIa). The safety and efficacy data related to specific biphasic waveforms, the most effective initial shock, and whether to use escalating sequences require additional studies in both the in-hospital and out-of-hospital settings.
| Automated External Defibrillators |
|---|
Lay Rescuer AED Programs
Since 1995 the American Heart Association (AHA) has recommended the development of lay rescuer AED programs to improve survival rates from out-of-hospital SCA.4749 These programs are also known as public access defibrillation, or PAD, programs. The goal of these programs is to shorten the time from onset of VF until CPR and shock delivery by ensuring that AEDs and trained lay rescuers are available in public areas where SCA is likely to occur. To maximize the effectiveness of these programs, the AHA has emphasized the importance of organization, planning, training, linking with the EMS system, and establishing a process of continuous quality improvement.50,51
Studies of lay rescuer AED programs in airports52 and casinos53,54 and first-responder programs with police officers26,34,36,44,5557 have shown a survival rate of 41% to 74% from out-of-hospital witnessed VF SCA when immediate bystander CPR is provided and defibrillation occurs within about 3 to 5 minutes of collapse. These high survival rates, however, are not attained in programs that fail to reduce time to defibrillation.5860
In a large prospective randomized trial (LOE 1)61 funded by the AHA, the National Heart, Lung, and Blood Institute (NHLBI), and several AED manufacturers, lay rescuer CPR + AED programs in targeted public settings doubled the number of survivors from out-of-hospital VF SCA when compared with programs that provided early EMS call and early CPR. The programs included a planned response, lay rescuer training, and frequent retraining/practice. The following elements are recommended for community lay rescuer AED programs50,51:
More information is available on the AHA website: www.americanheart.org/cpr. Under the topic "Links on this site," select "Have a question?" and then select "AED."
Lay rescuer AED programs will have the greatest potential impact on survival from SCA if the programs are created in locations where SCA is likely to occur. In the NHLBI trial, programs were established at sites with a history of at least 1 out-of-hospital cardiac arrest every 2 years or where at least 1 out-of-hospital SCA was predicted during the study period (ie, sites having >250 adults over 50 years of age present for >16 h/d).61
To be effective, AED programs should be integrated into an overall EMS strategy for treating patients in cardiac arrest. CPR and AED use by public safety first responders (traditional and nontraditional) are recommended to increase survival rates for SCA (Class I). AED programs in public locations where there is a relatively high likelihood of witnessed cardiac arrest (eg, airports, casinos, sports facilities) are recommended (Class I). Because the improvement in survival rates in AED programs is affected by the time to CPR and to defibrillation, sites that deploy AEDs should establish a response plan, train likely responders in CPR and AED use, maintain equipment, and coordinate with local EMS systems.50,51
Approximately 80% of out-of-hospital cardiac arrests occur in private or residential settings (LOE 4).62 Reviewers found no studies that documented the effectiveness of home AED deployment, so there is no recommendation for or against personal or home deployment of AEDs (Class Indeterminate).
AEDs are of no value for arrest not caused by VF/pulseless VT, and they are not effective for treatment of nonshockable rhythms that may develop after termination of VF. Nonperfusing rhythms are present in most patients after shock delivery,25,26,28,44 and CPR is required until a perfusing rhythm returns. Therefore, the AED rescuer should be trained not only to recognize emergencies and use the AED but also to support ventilation and circulation with CPR as needed.
The mere presence of an AED does not ensure that it will be used when SCA occurs. Even in the NHLBI trial, in which almost 20 000 rescuers were trained to respond to SCA, lay rescuers attempted resuscitation before EMS arrival for only half of the victims of witnessed SCA, and the on-site AED was used for only 34% of the victims who experienced an arrest at locations with AED programs.61 These findings suggest that lay rescuers need frequent practice to optimize response to emergencies.
It is reasonable for lay rescuer AED programs to implement processes of continuous quality improvement (Class IIa). These quality improvement efforts should use both routine inspections and postevent data (from AED recordings and responder reports) to evaluate the following50,51:
Automated Rhythm Analysis
AEDs have microprocessors that analyze multiple features of the surface ECG signal, including frequency, amplitude, and some integration of frequency and amplitude, such as slope or wave morphology. Filters check for QRS-like signals, radio transmission, or 50- or 60-cycle interference as well as loose electrodes and poor electrode contact. Some devices are programmed to detect spontaneous movement by the patient or others. Prototype defibrillators were used in 2 recent clinical trials evaluating quality of CPR in the out-of-hospital and hospital settings, and they hold promise for future AEDs that may prompt rescuers to improve the quality of CPR provided.18,19
AEDs have been tested extensively, both in vitro against libraries of recorded cardiac rhythms and clinically in many field trials in adults63,64 and children.65,66 They are extremely accurate in rhythm analysis. Although AEDs are not designed to deliver synchronized shocks (ie, cardioversion for VT with pulses), AEDs will recommend a (nonsynchronized) shock for monomorphic and polymorphic VT if the rate and R-wave morphology exceed preset values.
Electrode Placement
Rescuers should place AED electrode pads on the victims bare chest in the conventional sternal-apical (anterolateral) position (Class IIa). The right (sternal) chest pad is placed on the victims right superior-anterior (infraclavicular) chest and the apical (left) pad is placed on the victims inferior-lateral left chest, lateral to the left breast (Class IIa). Other acceptable pad positions are placement on the lateral chest wall on the right and left sides (biaxillary) or the left pad in the standard apical position and the other pad on the right or left upper back (Class IIa).
When an implantable medical device is located in an area where a pad would normally be placed, position the pad at least 1 inch (2.5 cm) away from the device (Class Indeterminate). If the victim has an ICD that is delivering shocks (ie, the patients muscles contract in a manner similar to that observed during external defibrillation), allow 30 to 60 seconds for the ICD to complete the treatment cycle before attaching an AED. Occasionally the analysis and shock cycles of automatic ICDs and AEDs will conflict.67
Do not place AED electrode pads directly on top of a transdermal medication patch (eg, patch containing nitroglycerin, nicotine, analgesics, hormone replacements, antihypertensives) because the patch may block delivery of energy from the electrode pad to the heart and may cause small burns to the skin.68 Remove medication patches and wipe the area before attaching the electrode pad.
If an unresponsive victim is lying in water or if the victims chest is covered with water or the victim is extremely diaphoretic, remove the victim from water and briskly wipe the chest before attaching electrode pads and attempting defibrillation. AEDs can be used when the victim is lying on snow or ice. Most victims do not need any special preparation of the chest other than removal of the clothes from the chest. If the victim has a very hairy chest, it may be necessary to remove some hair so that the electrode pads will adhere to the chest. This may be accomplished by briskly removing an electrode pad (which will remove some hair), or it may be necessary to shave the chest in that area.
AED Use in Children
Cardiac arrest is less common in children than adults, and its causes are more diverse.6971 Although VF is not a common arrhythmia in children, it is observed in 5% to 15% of pediatric and adolescent arrests.7175 In these patients rapid defibrillation may improve outcomes.75,76
The lowest energy dose for effective defibrillation in infants and children is not known. The upper limit for safe defibrillation is also not known, but doses >4 J/kg (as high as 9 J/kg) have effectively defibrillated children77,78 and pediatric animal models79 with no significant adverse effects. Based on adult clinical data17,24 and pediatric animal models,7981 biphasic shocks appear to be at least as effective as monophasic shocks and less harmful. Recommended manual defibrillation (monophasic or biphasic) doses are 2 J/kg for the first attempt (Class IIa; LOE 582 and 679) and 4 J/kg for subsequent attempts (Class Indeterminate).
Many AEDs can accurately detect VF in children of all ages65,66 and differentiate shockable from nonshockable rhythms with a high degree of sensitivity and specificity.65,66 Some are equipped with pediatric attenuator systems (eg, pad-cable systems or a key), to reduce the delivered energy to a dose suitable for children.
For children 1 to 8 years of age the rescuer should use a pediatric dose-attenuator system if one is available.78,83,84 If the rescuer provides CPR to a child in cardiac arrest and does not have an AED with a pediatric attenuator system, the rescuer should use a standard AED.
There is insufficient data to make a recommendation for or against the use of AEDs for infants <1 year of age (Class Indeterminate). During infancy the risk of VF SCA is unknown, and most cardiac arrest is thought to be related to progression of respiratory failure or shock. As a result there is concern that repeated interruption of CPR to try to detect and treat a rhythm uncommon in that age group may introduce more risk than benefit.83
If an AED program is established in systems or institutions that routinely provide care to children, the program should be equipped with AEDs with a high specificity for pediatric shockable rhythms and with a pediatric attenuator system (eg, pediatric pad-cable system or other method of attenuating the shock dose). This statement, however, should not be interpreted as a recommendation for or against AED placement in specific locations where children are present. Ideally healthcare systems that routinely provide care to children at risk for cardiac arrest should have available manual defibrillators capable of dose adjustment.83
In-Hospital Use of AEDs
At the time of the 2005 Consensus Conference, there were no published in-hospital randomized trials of AEDs versus manual defibrillators. Evidence from 1 study of fair quality (LOE 4)85 and a case series (LOE 5)86 indicated higher rates of survival to hospital discharge when AEDs were used to treat adult VF or pulseless VT in the hospital.
Defibrillation may be delayed when patients develop SCA in unmonitored hospital beds and in outpatient and diagnostic facilities. In such areas several minutes may elapse before centralized response teams arrive with the defibrillator, attach it, and deliver shocks.87 Despite limited evidence, AEDs should be considered for the hospital setting as a way to facilitate early defibrillation (a goal of
3 minutes from collapse), especially in areas where staff have no rhythm recognition skills or defibrillators are used infrequently. An effective system for training and retraining should be in place.
When hospitals deploy AEDs, first-responding personnel should also receive authorization and training to use an AED, with the goal of providing the first shock for any SCA within 3 minutes of collapse. The objective is to make goals for in-hospital use of AEDs consistent with goals established in the out-of-hospital setting.88 Early defibrillation capability should be available in ambulatory care facilities as well as throughout hospital inpatient areas. Hospitals should monitor collapse-to-first shock intervals and resuscitation outcomes (see Part 3: "Overview of CPR").
| Manual Defibrillation |
|---|
As noted, biphasic defibrillators use one of two waveforms, and each waveform has been shown to be effective in terminating VF over a specific dose range. The ideal shock dose with a biphasic device is one that falls within the range that has been documented to be effective using that specific device. Manufacturers should display the device-specific effective waveform dose range on the face of the device, and providers should use that dose range when attempting defibrillation with that device. Providers should be aware of the range of energy levels at which the specific waveform they use has been shown to be effective for terminating VF, and they should use that device-specific dose for attempted defibrillation. At this time there is no evidence that one biphasic waveform is more effective than another.
With a biphasic defibrillator it is reasonable to use selected energies of 150 J to 200 J with a biphasic truncated exponential waveform or 120 J with a rectilinear biphasic waveform for the initial shock. For second and subsequent shocks, use the same or higher energy (Class IIa). In this context "selected" refers to the energy dose selected by the operator (or programmed by the AED manufacturer). With the rectilinear biphasic waveform device, selected and delivered energies usually differ; delivered energy is typically higher in the usual range of impedance. For example, in a patient with 80
impedance, a selected energy of 120 J will deliver 150 J.
If a provider is operating a manual biphasic defibrillator and is unaware of the effective dose range for that device to terminate VF, the rescuer may use a selected dose of 200 J for the first shock and an equal or higher dose for the second and subsequent shocks. The 200-J "default" energy level is not necessarily an optimal dose, but it was selected because it falls within the reported range of doses effective for first and subsequent biphasic shocks. In addition, this dose can be provided by every biphasic manual defibrillator available in 2005. Thus, it is a consensus default dose and not a recommended ideal dose. If devices are clearly labeled and providers are familiar with the devices they will use for clinical care, the device-specific dose will be used and there will be no need for the "default" 200-J dose.
If a monophasic defibrillator is used, select a dose of 360 J for all shocks. If VF is initially terminated by a shock but then recurs later in the arrest, deliver subsequent shocks at the previously successful energy level.
Defibrillation is achieved by generating amplitude of current flow and sustaining that flow for a time interval. Although the defibrillator operator selects the shock energy (in joules), it is the current flow (in amperes) that actually depolarizes the myocardium. Current depends in part on the selected shock dose and is affected by the thoracic pathway between the 2 defibrillator electrodes and the position of the heart in that pathway and impedance to current flow between the electrodes. The complexity of thoracic current flow has been observed experimentally.89
The most important determinant of survival in adult VF SCA is rapid defibrillation by either a monophasic or biphasic device. Thus, in the hospital it is acceptable to deliver 1 shock with a monophasic or biphasic defibrillator followed by immediate initiation of CPR, beginning with compressions. The goal is to minimize the time between chest compressions and shock delivery and between shock delivery and resumption of chest compressions. In specific settings (eg, critical care units with hemodynamic monitoring in place), this sequence may be modified at the physicians discretion (see Part 7.2: "Management of Cardiac Arrest" and Part 12: "Pediatric Advanced Life Support").
Transthoracic Impedance
The average adult human impedance is &70 to 80
.9092 When transthoracic impedance is too high, a low-energy shock will not generate sufficient current to achieve defibrillation.91,93,94 To reduce transthoracic impedance, the defibrillator operator should use conductive materials. This is accomplished with the use of gel pads or electrode paste with paddles or through the use of self-adhesive pads. No existing data suggests that one of these modalities is better than the others in decreasing impedance (Class Indeterminate).
In a male patient with a hairy chest, electrode-to-chest contact may be poor, and the hair may cause air trapping between the electrode and skin. This, as well as improper use of paddles, may result in high impedance, with occasional current arcing. Although extremely rare, in oxygen-rich environments such as critical care units, this arcing has been known to cause fires if an accelerant is present (see below). When using paddles, rescuers should apply them firmly to gel pads on the chest wall, avoiding contact with ECG leads. Use of self-adhesive pads will reduce the risk of arcing. It may be necessary to shave the area of intended pad placement.
Electrode Position
An overview of adhesive pad placement was provided in the AED section above. If electrode paddles are used instead of pads, the paddles should be well separated, and the paste or gel used to create the interface between the paddles and the skin should not be smeared on the chest between the paddles. Smearing of the paste or gel may allow current to follow a superficial pathway (arc) along the chest wall, "missing" the heart. Self-adhesive monitor/defibrillator electrode pads are as effective as gel pads or paste (LOE 39597), and they can be placed before cardiac arrest to allow for monitoring and then rapid administration of a shock when necessary.98 Consequently, self-adhesive pads should be used routinely instead of standard paddles (Class IIa; LOE 2, 4).
When providing cardioversion or defibrillation for patients with permanent pacemakers or ICDs, do not place the electrodes over or close to the device generator, because defibrillation can cause the pacemaker to malfunction. A pacemaker or ICD also may block some current to the myocardium during defibrillation attempts, resulting in suboptimal energy delivery to the heart. Because some of the defibrillation current flows down the pacemaker leads, permanent pacemakers and ICDs should be reevaluated after the patient receives a shock.99
Electrode Size
In 1993 the Association for the Advancement of Medical Instrumentation recommended a minimum electrode size of 50 cm2 for individual electrodes.100 However, advances in electrode design and chemical composition may soon require modification of this recommendation.
For adult defibrillation, both handheld paddle electrodes and self-adhesive pad electrodes 8 to 12 cm in diameter perform well, although defibrillation success may be higher with electrodes 12 cm in diameter rather than with those 8 cm in diameter.90,95 Small electrodes (4.3 cm) may be harmful and may cause myocardial necrosis.101 When using handheld paddles and gel or pads, rescuers must ensure that the paddle is in full contact with the skin. Even smaller pads have been found to be effective102 in VF of brief duration. Use of the smallest (pediatric) pads, however, can result in unacceptably high transthoracic impedance in larger children.103 It is best to use the largest pads that can fit on the chest without overlap.
| Fibrillation Waveform Analysis |
|---|
At issue is whether analysis of the VF waveform is useful in predicting therapeutic outcome and modifying therapy prospectively. Potential applications include prediction of success of cardioversion, selection of appropriate waveform type, and optimization of timing of defibrillation relative to CPR and medication delivery.
| Current-Based Defibrillation |
|---|
Clinical studies using MDS waveform shocks have tried to identify the range of current necessary to achieve defibrillation and cardioversion. The optimal current for ventricular defibrillation appears to be 30 to 40 A MDS.92 Comparable information on current dosage for biphasic waveform shocks is under investigation.
| "Occult" Versus "False" Asystole |
|---|
| Fire Hazard |
|---|
The use of self-adhesive defibrillation pads is probably the best way to minimize the risk of sparks igniting during defibrillation. If manual paddles are used, gel pads are preferable to electrode pastes and gels because the pastes and gels can spread between the 2 paddles, creating the potential for a spark (Class IIb). Do not use medical gels or pastes with poor electrical conductivity, such as ultrasound gel.
Rescuers should take precautions to minimize sparking during attempted defibrillation; try to ensure that defibrillation is not attempted in an oxygen-enriched atmosphere (Class IIa). When ventilation is interrupted for shock delivery, rescuers should try to ensure that oxygen does not flow across the patients chest during defibrillation attempts.
| Synchronized Cardioversion |
|---|
Delivery of synchronized shocks (cardioversion) is indicated to treat unstable tachyarrhythmias associated with an organized QRS complex and a perfusing rhythm (pulses). The unstable patient demonstrates signs of poor perfusion, including altered mental status, ongoing chest pain, hypotension, or other signs of shock (eg, pulmonary edema).
Synchronized cardioversion is recommended to treat unstable supraventricular tachycardia due to reentry, atrial fibrillation, and atrial flutter. These arrhythmias are all caused by reentry, an abnormal rhythm circuit that allows a wave of depolarization to travel in a circle. The delivery of a shock can stop these rhythms because it interrupts the circulating (reentry) pattern. Synchronized cardioversion is also recommended to treat unstable monomorphic VT. For additional information see Part 7.3: "Management of Symptomatic Bradycardia and Tachycardia."
Cardioversion will not be effective for treatment of junctional tachycardia or ectopic or multifocal atrial tachycardia because these rhythms have an automatic focus. Automatic rhythms are created when local cells are stimulated to spontaneously depolarize at a rapid rate. Sinus tachycardia is a good example of an automatic rhythm. It results when the cells in the sinus node are stimulated (eg, by catecholamines) to depolarize at a rapid rate. Junctional tachycardia and ectopic or multifocal atrial tachycardia also result when cells are stimulated to depolarize at a rapid rate. Delivery of a shock cannot stop these rhythms. In fact, shock delivery to a heart with a rapid automatic focus may increase the rate of the tachyarrhythmia.
Synchronized cardioversion is not used for treatment of VF, pulseless VT, or unstable polymorphic (irregular) VT. These rhythms require delivery of high-energy unsynchronized shocks (ie, defibrillation doses). Electrical therapy for VT is discussed further below. For additional information see Part 7.2: "Management of Cardiac Arrest."
Supraventricular Tachycardias (Reentry SVT)
The recommended initial monophasic energy dose for cardioversion of atrial fibrillation is 100 J to 200 J. Cardioversion of atrial flutter and other supraventricular tachycardias generally requires less energy; an initial energy of 50 J to 100 J MDS waveform is often sufficient. If the initial 50-J shock fails, providers should increase the dose in a stepwise fashion.93 These recommendations are consistent with those contained in the ECC Guidelines 2000.50 Cardioversion with biphasic waveforms is now available,132 but the optimal doses for cardioversion with biphasic waveforms have not been established with certainty. Extrapolation from published experience with elective cardioversion of atrial fibrillation using rectilinear and truncated exponential waveforms supports an initial dose of 100 J to 120 J with escalation as needed.133,134 This initial dose has been shown to be 80% to 85% effective in terminating atrial fibrillation. Until further evidence becomes available, this information can be used to extrapolate biphasic cardioversion doses to other tachyarrhythmias.135138
A recent prospective randomized study that compared the rectilinear biphasic waveform (200 J maximum selected energy) with a biphasic truncated exponential waveform (360 J maximum energy) for elective cardioversion found no significant differences in efficacy between the 2 waveforms.134
Ventricular Tachycardia
The amount of energy and timing of shocks for treatment of VT with pulses are determined by the patients condition and the morphologic characteristics of the VT.139 Pulseless VT is treated as VF (see Part 7.2: "Management of Cardiac Arrest"). Management of stable VT is summarized in Part 7.3: "Management of Symptomatic Bradycardia and Tachycardia." Unstable monomorphic (regular) VT with pulses is treated with synchronized cardioversion. Unstable polymorphic (irregular) VT with or without pulses is treated as VF using unsynchronized high-energy shocks (ie, defibrillation doses).
Monomorphic VT (regular form and rate) with a pulse responds well to monophasic waveform cardioversion (synchronized) shocks at initial energies of 100 J. If there is no response to the first shock, increase the dose in a stepwise fashion (eg, 100 J, 200 J, 300 J, 360 J). These recommendations are consistent with the recommendations in the ECC Guidelines 2000.50
Although synchronized cardioversion is preferred for treatment of an organized ventricular rhythm, for some arrhythmias synchronization is not possible. The many QRS configurations and irregular rates that comprise polymorphic ventricular tachycardia make it difficult or impossible to reliably synchronize to a QRS complex. In addition, the patient with persistent polymorphic VT will probably not maintain perfusion/pulses for very long, so any attempt to distinguish between polymorphic VT with or without pulses quickly becomes moot. A good rule of thumb is that if your eye cannot synchronize to each QRS complex, neither can the defibrillator/cardioverter. If there is any doubt whether monomorphic or polymorphic VT is present in the unstable patient, do not delay shock delivery to perform detailed rhythm analysisprovide high energy unsynchronized shocks (ie, defibrillation doses).
The recommended shock doses for high-energy, unsynchronized shocks (defibrillation) with a biphasic or monophasic device are those presented earlier in this section (see "Manual Defibrillation, Shock Energies"). After shock delivery the healthcare provider should be prepared to provide immediate CPR (beginning with chest compressions) and follow the ACLS Pulseless Arrest Algorithm if pulseless arrest develops (for further information see Part 7.2: "Management of Cardiac Arrest").
There is limited data about the treatment of polymorphic (irregular) VT. Providers should consider consultation with an expert in arrhythmia management. Treatment of the patient with polymorphic VT is presented in section 7.3: "Management of Symptomatic Bradycardia and Tachycardia."
| Pacing |
|---|
Three randomized controlled trials (LOE 2)140142 of fair quality and additional studies (LOE 3 to 7)143149 indicate no improvement in the rate of admission to hospital or survival to hospital discharge when paramedics or physicians attempted to provide pacing in asystolic patients in the prehospital or hospital (emergency department) setting. Given the recent recognition of the importance of maximizing chest compressions as well as the lack of demonstrated benefit of pacing for asystole, withholding chest compressions to attempt pacing for patients with asystole is not recommended (Class III).
Transcutaneous pacing is recommended for treatment of symptomatic bradycardia when a pulse is present. Healthcare providers should be prepared to initiate pacing in patients who do not respond to atropine (or second-line drugs if these do not delay definitive management). Immediate pacing is indicated if the patient is severely symptomatic, especially when the block is at or below the His Purkinje level. If the patient does not respond to transcutaneous pacing, transvenous pacing is needed. For further information see Part 7.3: "Management of Symptomatic Bradycardia and Tachycardia."
| Maintaining Devices in a State of Readiness |
|---|