(Circulation. 1995;91:2627-2634.)
© 1995 American Heart Association, Inc.
Articles |
From the Departments of Radiology (G.D.C., R.E., C.C., D.S., J.D., H.F.L., R.W.P.) and Internal Medicine (W.G.H., R.M.P.), University of Texas Southwestern Medical Center at Dallas; and the Magnetic Resonance Clinical Research Division (M.N.), Picker International, Inc, Highland Heights, Ohio. Present address for Dr NessAiver is Radiology Department, University of Maryland College of Medicine (Baltimore).
Correspondence to Geoffrey D. Clarke, PhD, Radiology Department/Medical Physics Division, University of Texas Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX 75235-9071.
| Abstract |
|---|
|
|
|---|
Methods and Results In eight mongrel dogs, a closed chest model of partial left anterior descending coronary artery (LAD) occlusion was created. Coronary flows in the left circumflex artery (LCx) and LAD were measured at rest and during adenosine infusion using velocity-encoded, breath-hold MRI and perivascular ultrasound (US) flowmeters. MRI measurements of absolute coronary flow and coronary flow reserve were highly correlated with US (r=.96 and .94, respectively). Flow reserve measured in the constricted LAD was significantly lower than that in the unconstricted LCx by both US (P=.002) and MRI (P=.011).
Conclusions MRI measurements of coronary flow and flow reserve were in good agreement with US measurements. In addition, MRI measurements of coronary flow reserve successfully discriminated stenotic from normal vessels. These results indicate that MRI is a useful method for the noninvasive assessment of coronary flow and stenosis.
Key Words: blood flow magnetic resonance imaging adenosine
| Introduction |
|---|
|
|
|---|
Rapid and accurate MRI measurements of pulsatile flow using the PEG approach combined with quantitative phase-contrast velocity mapping have been demonstrated in test objects.4 This method was used to measure coronary flow velocities in human volunteers during episodes of breath-holding5 6 ; however, the spatial resolution was not adequate for accurate flow measurements. High-performance gradient coils have recently become available, which allow the acquisition of breath-hold cine MRI velocity images with submillimeter pixel sizes.
In the present study, the following questions were addressed. (1) Can adequate spatial and temporal resolution be obtained to perform phasic breath-hold MRI flow measurements of coronary arterial flow in vivo? (2) How well do MRI measurements of absolute coronary arterial flow and coronary flow reserve agree with those obtained from perivascular ultrasound (US) flowmeters? (3) Are measurements of absolute coronary flow and flow reserve adequate to distinguish flow in normal vessels from those with a subcritical stenosis?
| Methods |
|---|
|
|
|---|
At the magnet, pressure and flow measurements were calibrated and recorded. Each animal was placed on its right side on the table of the MRI system and ventilated with a mixture of isoflurane (1% to 2%) and oxygen (2.5 L, 12 breaths per minute). US flow, aortic pressure, and ECG signals were recorded on a strip-chart. US flow measurements and aortic pressure were recorded continuously except during the MRI acquisition as these electronic instruments were found to produce unacceptable radiofrequency noise in the MR image. MRI baseline flow measurements were performed with the US flow data and aortic pressures recorded before the first adenosine bolus of 6 mg was administered intravenously. After a wait of 1 to 2 minutes, a second bolus of 12 mg was given intravenously. An intravenous adenosine infusion of 0.14 mg · kg-1 · min-1 was started, and pressure and US flow measurements were recorded until a consistent, high coronary flow rate was established in the LCx.
US and MRI flows were recorded during maximum coronary dilation. The adenosine infusion was stopped. Pressure and flows were monitored as they returned to baseline levels, and MRI flow measurements were repeated. The animal was then removed from the magnet and killed by a lethal dose of sodium pentobarbitol, and the heart was excised.
Magnetic Resonance Imaging
MRI signal acquisition was
performed using a method of
k-segmentation that allows the acquisition of several phase-encoding
steps for each cine frame in a single heartbeat.3 In this
method, data are acquired in PEG, and short repetition times are used
to acquire multiple phase-encoding steps for each cardiac frame for
each RR interval. A complete cardiac cine with four to six frames can
be obtained during a single breath-hold. The scan time is reduced by a
factor equal to the PEG size without reduction in spatial resolution or
special modifications to the system hardware.
The specific type of k-space segmentation used in the present study involved symmetrical, centrally ordered PEG (SCOPEG).4 10 With this method, heartbeat-to-heartbeat misregistration of the data is avoided by acquiring the data in the central portion of k-space (low spatial frequencies) at the same time in the cardiac cycle for successive heartbeats. The effects of eddy currents, produced by rapid pulsing of the gradients, are minimized by designing the phase-encoding gradient pulses so that a positive pulse is always followed by a negative pulse, resulting in smoother transitions. To preserve uniform modulation of the gradient pulses, the positive and negative regions of k-space are sampled separately, resulting in a data set that is symmetrically ordered about the zeroth phase-encoding step. Thus, this segmentation scheme reduces the incidence of ghosting and eddy current effects while also minimizing the blurring due to motion between low spatial frequency phase-encode steps.
Phase-contrast cine MRI velocity measurement requires that at least two independent images are acquired.11 By manipulation of magnetic field gradient pulses, one image is "motion compensated" so that minimal dephasing of the echo signal due to first-order motion occurs. The second acquisition is designed so that the phase of the signal is "velocity encoded" for a predetermined range of velocities. The difference in the phases of the two complex MRI signals from moving spins is directly proportional to the velocity. To produce quantitative phase-contrast MRI velocity images, an image of the spatial distribution of the MRI signal's phase, as distinct from the more commonly used signal amplitude, must be produced. The pixel-by-pixel difference between the motion-compensated and velocity-encoded phase images produces the velocity image. By interleaving the "motion-compensated" and "velocity-encoded" acquisitions, view-to-view misregistration artifacts are reduced.12
The present study was conducted with a 1.5-T Vista HPQ MRI system (Picker International) that was fitted with prototype high-performance gradient coils designed using methods described by Morich.13 These gradient coils consisted of a distributed winding with a peak gradient of 15 mT/m and a 500-µs minimum ramping time to 10 mT/m. A receive-only quadrature surface coil (20 cmx25 cm), designed for spine imaging, was placed on the left thorax of the animal over the incision. Multiphasic pilot images (TR=19.1 milliseconds, TE=11 milliseconds) were obtained using the breath-hold PEG method.
Starting with a sagittal view and continuing with oblique orientations, a series of high-resolution cine images with the LAD and LCx in the image plane were obtained. Breath-hold MRI pilot images were obtained with the ventilator stopped during the scan to determine the position of the LCx and LAD during a breath-hold. A perpendicular view of the vessel was acquired to confirm that the section was straight and adequate for the slice thickness of the anticipated flow measurement and that there was no excessive through-plane motion. Through-plane motion was evaluated by measuring the frame-to-frame displacement of landmarks on the myocardium near the arteries that were visible on the cine scout images.
MRI images could not be positioned at the same position as the US flow probes since a significant magnetic susceptibility gradient virtually eliminated signal in these regions. Placement of the velocity image acquisition was generally along the arterial segment, proximal to the US flow probe. In two animals, no sufficiently straight arterial segment proximal to US flow probe on the LAD artery was found. In these cases, the LAD flow velocity images were obtained distal to the US flow probe.
Flow was determined using interleaved velocity-encoded and motion-compensated MRI scans. The velocity encoding was calibrated for the range of flow velocities from -138.4 cm/s to +138.4 cm/s in the slice select direction. MRI image parameters were TR, 19.1 milliseconds; TE, 11 milliseconds; 40-degree nutation; and slice thickness, 7 mm. The following parameters varied with the specific image orientation and the animal's heart rate: FOV=18 to 24 cm corresponding to a pixel size range of 0.7 to 0.94 mm2, with phase-encode aspect ratios ranging from 160:256 to 220:256, and PEG sizes of two or three. Each cine set of four to six images was acquired during a 30- to 40-second breath-hold. A total of 8 to 12 flow measurements were made in each of the animals using quantitative MRI velocity imaging while measuring flow by US just before and after MR image acquisition.
Flow measurements in the LAD were in the presence of a simulated subcritical stenosis. In the LCx, all measurements were made with no constrictor present, both before and during adenosine administration. In the course of a typical experiment, at least two MRI multiphasic velocity image sets under baseline and adenosine infusion conditions were acquired from each vessel. Seventy-five MRI coronary flow values were compared with the concurrent flow probe measurements, providing a database adequate for meaningful statistical analysis.
Data Analysis
The coronary artery flow was estimated from the
MRI data in the
following manner. The magnitude images were displayed in a cine loop,
and the position and shape of the artery were noted. To mitigate
identification problems due to in-flow signal enhancement, the lumen
cross-sectional area was determined from the magnitude image of the
cine set in which the vessel appeared to be largest. Previous studies
have demonstrated a mean variation in coronary vessel size of 3% from
beat to beat and of 6% from systole to diastole,14 15
suggesting that beat-to-beat and phasic changes in coronary arterial
diameter could not be measured at the MRI spatial resolutions used in
the present study. Thus, the region of interest (ROI) size was kept
constant and the frame-to-frame location of the vessel was determined
using the magnitude image set. By placing the ROI at the identical
position on the associated velocity image, the mean velocity in the
vessel was measured. The flow during each phase of the cardiac cycle
was calculated from the product of the mean velocity in the ROI and the
ROI area. The MRI estimated coronary flow (in mL/min) was the mean of
the calculated flow values for the entire set of images.
Coronary flow reserve in each artery was defined as the ratio between the average of the vasodilated coronary flow values and the mean resting coronary flow value. The predicted and US-measured flow reserves in each artery were calculated by taking the mean coronary flow measured during adenosine infusion and dividing it by the mean baseline coronary flow.
The correlation coefficients and the slope and intercept of the regression line were calculated by linear regression of the MRI-predicted and US-measured coronary flow and coronary flow reserve data.16 The MRI-predicted flow values were compared for agreement with the US flow data and analyzed using the method of Bland and Altman.17 The reproducibility of the MRI-predicted and US flow measurements was assessed using the data from the pairs of flow measurements performed under stable hemodynamic conditions. The standard deviation of the mean difference between two successive measurements multiplied by 2 yields the reproducibility coefficient.17
Relative coronary flow reserve was assessed
by comparing the maximal
coronary flow reserve in the LAD (with subcritical stenosis) with the
maximal coronary flow reserve in the LCx (no stenosis) during the same
measurement interval. An unpaired Student's t test with
equal variances12 was used to determine the probability of
the null hypothesis that the coronary flow reserve in the two vessels
would be measured as equal. A paired t test12
was used to determine the statistical significance of differences
between vessel sizes, mean velocities, and mean flow values measured
with MRI before and during adenosine infusion. Except where noted, all
data are expressed as mean±SD. Statistical significance is defined as
a value of P
.05.
| Results |
|---|
|
|
|---|
Physiological Responses
Hemodynamic data are listed in Table
1
. There
was an increase in the baseline coronary artery flow recorded at the
beginning of MRI measurements compared with those recorded during
surgery. This difference was presumed to be due to the vasodilatory
effects of the isoflurane anesthesia used during the MRI
scan.18
|
No problems were encountered with changing heart
rate. There was a
small, but not statistically significant, increase from baseline heart
rate during adenosine infusion (Table 1
). The mean heart rate
was 112.1
beats per minute before a breath-hold and 113.0 beats per minute after
a breath-hold. The mean percent difference between heart rate before
and after breath-holding was <1.0%.
LAD and LCx Coronary Flow
The MRI acquisition took
approximately 90 minutes, including
set-up and velocity imaging. During the course of each experiment,
coronary arterial flow increased with both MRI and US flowmeter
measurements (Fig 1
). The vessel mean diameters measured
by MRI before and during adenosine infusion are listed in Table
2
. Both magnitude and velocity images depicted changes
related to increases in coronary flow during adenosine infusion. The
magnitude images were observed to have increased signal in the vessels.
Although coronary flow velocities were observed to increase in the LAD
velocity images (Fig 1
), a more noticeable effect was
visualized in the velocity images of the LCx during adenosine infusion
(Fig 2
).
|
|
|
The mean velocity between frames was
1.8±2.24 cm/s due to bulk motion
in the slice select direction of the velocity image. The mean number of
pixels per vessel was 12.5±7.2. The frequency distribution of the
number of pixels per vessel was positively skewed. In the LCx, both
vessel size and flow velocity increased significantly during adenosine
infusion, whereas in the LAD, only flow velocity increased
significantly. MRI-estimated flow values in both the LAD and LCx were
similar to US-measured flows, during both baseline and adenosine
infusion conditions (Table 2
).
Linear regression
demonstrated the correlation of MRI and US flow
measurements (Fig 4
) made at the same time. The mean
difference between the coronary arterial flow values obtained using US
and MRI was small, -1.2±12 mL/min. The limits of agreement,
defined
as ±2 SD from the mean difference, were +22.7 and -25.1
mL/min. If a
normal distribution is assumed, 95% of the differences between the two
measurement methods would fall within this range (Fig 5
).
|
|
Measured coronary arterial flow ranges were 15 to 178 mL/min for MRI and 17 to 180 mL/min for US. The reproducibility coefficients, calculated from successive measurements, were 16.8 mL/min for US and 26.4 mL/min for MRI.
Coronary Flow Reserve Measurements
The mean predicted flow
reserve in the LAD arteries was 1.38±0.31
compared with 1.42±0.31 as measured by US. The mean predicted flow
reserve in the LCx was 2.57±0.92, and the US-measured LCx flow reserve
was 2.55±0.77. Regression of the MRI and US flow reserve values
produced a correlation coefficient (r=.94,
intercept=0.10,
and slope=1.04; Fig 6
).
|
The mean difference between the two methods of coronary flow reserve measurement was 0.01. The limits of agreement between the US- and MRI-predicted flow reserve measurements were 0.64 to -0.61. Both measurement methods indicated a significant difference between the flow reserves of the LAD and LCx (US, P=.0017; MRI, P=.0107).
| Discussion |
|---|
|
|
|---|
The results of the present study demonstrate that despite the small size of the coronary vessels and the gross heart motion, it is possible to measure absolute flow in the coronary arteries using breath-hold MRI phase-contrast images. MRI flow measurements were in good agreement with perivascular US flowmeter measurements. Absolute coronary flow and flow reserve measurements obtained using MRI were sufficiently accurate to discriminate between normal arteries and arteries with simulated subcritical stenoses.
The data presented offer the first direct, quantitatively validated
measurements of coronary flow reserve in individual coronary vessels
using MRI phase-difference velocity imaging. Improved MRI velocity
image spatial resolution has allowed measurement of coronary blood flow
in a specific experimental setting. Tang et al20
demonstrated in phantoms that the error in measured volume flow rate
will be <10% if the ratio between in-plane pixel dimension and vessel
radius is <0.5. The spatial resolutions of the images obtained in the
present study are at the margin of the above criterion. Diameter
measurements, as reported in Table 2
, were determined from the
number
of pixels across a vessel. These pixel sizes, as stated in
"Methods," ranged from 0.7 to 0.94 mm and are known to have good
accuracy. Thus, we report a vessel measured to be three 0.7-mm pixels
in diameter as 2.1 mm. The large dogs in the present study had mean
coronary artery cross-sectional areas of 9.0 mm2 for the
LAD and 6.6 mm2 for the LCx. These measurements correspond
to the range of the mean sizes of proximal coronary vessels in men 55
to 74 years old.21
Tang et al20 also noted that the error observed in conditions of laminar flow is much less than that seen in plug flow. Laser and high-frequency US Doppler velocimetric studies of normal coronary arteries in dogs22 and human patients23 have demonstrated that the velocity profile in the proximal coronary arteries, while not strictly laminar, has many features resembling laminar flow. Poststenotic flow, however, has been demonstrated to exhibit flow reversal and irregularities in the flow velocity profile.24 Previous MRI phantom studies in open and stenotic tubes have demonstrated a decrease in MRI signal25 and disruption of the phase/velocity relationship26 ; however, these effects have been observed only at Reynolds numbers much higher than those of the coronary arteries studied in the present study.
Negative flow velocities were measured in the MRI velocity images of the constricted LAD during the first phase of the cardiac cycle (early systole) in four of the eight animals in the present study. The negative excursions were not large in magnitude and were applied uniformly with the other measurements in determining the total coronary flows. The present study was not designed to evaluate the phasic variation of coronary flow over the cardiac cycle using MRI methods. However, phasic variation in the MRI flow measurement was routinely observed and was found to be similar to those recorded in flowmeter measurements.
Other approaches have been taken to perform measurements of absolute coronary flow and coronary flow reserve with MRI. Global coronary flow reserve in healthy subjects and in patients has been measured indirectly by phase-contrast MRI velocity imaging using a model that considers the antegrade and retrograde flows in the aorta during systole and diastole.27 This method has not been evaluated for absolute accuracy, and the global flow reserve is of limited usefulness.
Echo-planar MR images (EPI), which can be produced in
100
milliseconds, have spatial resolutions too poor for measuring arterial
lumen cross-sectional areas. Recently, investigators have reported a
breath-hold time-of-flight method using EPI to measure coronary flow
velocities in healthy subjectsts.28 However, the pixel
size reported was 1.5x3.0 mm, and the model used in this method relied
on the assumptions that the vessel was straight and that flow in the
coronary artery was laminar. Nevertheless, EPI methods are still under
development and may be capable of producing accurate measurements of
absolute coronary flow in the future. Other methods, especially
radiographic angiography, intra-arterial Doppler US measurements, and
nuclear medicine, have been used to assess coronary flow reserve. The
catheterization methods offer the present gold standard but are
invasive. Our coronary flow reserve results compare favorably with
those obtained in a similar animal model using quantitative
radiographic angiography correlated with perivascular Doppler
US.29
Doppler wires are now being used to measure flow velocities in vessels <2.5 mm in diameter, and they offer the potential for evaluating coronary flow reserve in arterial branches. This capability is of interest for determining flow reserve in small vessels with diffuse and multiple stenoses and in helping to understand the details of the role of collateral flow in coronary artery disease. Improvements in MRI gradient coil performance and methods for enhancing signal-to-noise ratio, which are currently being proposed by MRI equipment vendors, will lead to improved temporal and spatial resolution of MRI phase-contrast velocity imaging, allowing evaluation of the distal coronary vasculature.
The present study had four methodological limitations. First, the conditions under which the MRI flow velocity measurements were performed were not the most favorable for demonstrating the ability of the method to measure coronary flow reserve. The baseline flow measured by MRI was relatively high due to the vasodilatory effects of the isoflurane anesthesia. It is also unlikely that the degree of adenosine-induced vasodilation was maximum in some cases.
Second, the coronary anatomy of the dog differs significantly from that of humans, in that the LAD and LCx supply the vast majority of blood to the canine heart. These vessels were isolated at surgery, and perivascular US flowmeters, which produced signal voids in the MRI images, were placed around the vessels. Thus, the coronary arteries were relatively easy to locate due to the presence of the US probes. Also, the MRI measurements were not performed at the most optimal locations on the arteries because the US flow probes were placed at these sites.
The third limitation of the present study was the length of time required to carefully position each arterial flow measurementapproximately 30 minutes. Holding still in the magnet for this length of time may be prohibitive for some patients. The length of the coronary artery in which the flow measurement is to take place is studied from at least two oblique orthogonal cine views during the pilot scan and set-up phase of the study, as described in "Methods." The flow image slice is then positioned so that the coronary artery will appear in the center of the velocity image field of view. Careful positioning of the flow velocity image slice was found to be the greatest technical challenge to a high-quality breath-hold PEG study. It is extremely important to measure flow in a straight segment of the artery and to ensure that there is limited through-plane motion of the artery.
The coronary arteries in general exhibit complex curvature and are moving; therefore, one would have difficulty viewing them in other than multiple, parallel segments regardless of the method of flow measurement being used. Acquisition of phase-contrast velocity measurements with the vessel in the plane of the image has been studied by Kraft et al,30 who demonstrated that partial volume effects due to static tissue on either side of the vessel cause severe phase errors. They recommended that small vessels be imaged transversely, as was done in the present study.
The fourth limitation is the length of the breath-hold period. In healthy, sedated dogs, a breath-hold of 30 to 40 seconds presents no problem. In a previous study, patients with known myocardial infarction have been able to hold their breaths repeatedly for as long as 30 seconds.31 However, many patients do not have this capability, especially those with pulmonary complications. Two possible approaches to reducing the length of the breath-hold include decreasing TR using a very high-performance gradient system and reducing the number of frames obtained per cardiac cycle. However, reducing the number of frames significantly may create other problems because these vessels are much easier to perceive in a cine-loop presentation. The magnitude and velocity images can be projected side-by-side, and the cyclic motion of the vessels is better understood. The flow character, in terms of both the gray scale in the velocity image and the in-flow effects in the magnitude images, can also be more easily evaluated.
Investigators have previously reported problems with inadequate breath-holding or patient movement during a breath-hold in real-time cardiac imaging,32 and this was observed in the present study in one inadequately anesthetized animal. Patient motion causes ghosting in the phase-encode direction of MRI images and is a concern for the clinical application of MRI flow measurement methods in conscious patients undertaking voluntary breath-holds. This is another argument for improving the method to make the breath-holding durations as short as possible.
The results of the present study suggest that cine breath-hold MRI phase-contrast velocity images can be used for the accurate, noninvasive measurement of absolute flow and flow reserve in the major epicardial coronary arteries. Also, these data suggest that MRI can be used to assess the severity of a known stenosis.
|
| Acknowledgments |
|---|
Received August 2, 1994; revision received November 22, 1994; accepted November 26, 1994.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. J. Pennell, U. P. Sechtem, C. B. Higgins, W. J. Manning, G. M. Pohost, F. E. Rademakers, A. C. van Rossum, L. J. Shaw, and E. K. Yucel Clinical indications for cardiovascular magnetic resonance (CMR): Consensus Panel report Eur. Heart J., November 1, 2004; 25(21): 1940 - 1965. [Full Text] [PDF] |
||||
![]() |
K. Shan, G. Constantine, M. Sivananthan, and S. D. Flamm Role of Cardiac Magnetic Resonance Imaging in the Assessment of Myocardial Viability Circulation, March 23, 2004; 109(11): 1328 - 1334. [Full Text] [PDF] |
||||
![]() |
S. E. Langerak, H. W. Vliegen, J. W. Jukema, A. H. Zwinderman, H. J. Lamb, A. de Roos, and E. E. van der Wall Vein Graft Function Improvement after Percutaneous Intervention: Evaluation with MR Flow Mapping Radiology, September 1, 2003; 228(3): 834 - 841. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Nagel, T. Thouet, C. Klein, S. Schalla, A. Bornstedt, B. Schnackenburg, J. Hug, E. Wellnhofer, and E. Fleck Noninvasive Determination of Coronary Blood Flow Velocity With Cardiovascular Magnetic Resonance in Patients After Stent Deployment Circulation, April 8, 2003; 107(13): 1738 - 1743. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. A. Silber, D. A. Bluemke, P. Ouyang, Y. P. Du, W. S. Post, and J. A. C. Lima The relationship between vascular wall shear stress and flow-mediated dilation: endothelial function assessed by phase-contrast magnetic resonance angiography J. Am. Coll. Cardiol., December 1, 2001; 38(7): 1859 - 1865. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Koskenvuo, J. J. Hartiala, J. Knuuti, H. Sakuma, J. O. Toikka, M. Komu, M. Saraste, and P. Niemi Assessing Coronary Sinus Blood Flow in Patients with Coronary Artery Disease: A Comparison of Phase-Contrast MR Imaging with Positron Emission Tomography Am. J. Roentgenol., November 1, 2001; 177(5): 1161 - 1166. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Ishida, H. Sakuma, B. P. Cruz, T. Shimono, T. Tokui, I. Yada, K. Takeda, and C. B. Higgins MR Flow Measurement in the Internal Mammary Artery-to-Coronary Artery Bypass Graft: Comparison with Graft Stenosis at Radiographic Angiography Radiology, August 1, 2001; 220(2): 441 - 447. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. K. Lund, M. F. Wendland, A. Shimakawa, H. Arheden, F. Ståhlberg, C. B. Higgins, and M. Saeed Coronary Sinus Flow Measurement by Means of Velocity-encoded Cine MR Imaging: Validation by Using Flow Probes in Dogs Radiology, November 1, 2000; 217(2): 487 - 493. [Abstract] [Full Text] |
||||
![]() |
H. Sakuma, J. W. Koskenvuo, P. Niemi, N. Kawada, J. O. Toikka, J. Knuuti, H. Laine, M. Saraste, M. Kormano, and J. J. Hartiala Assessment of Coronary Flow Reserve Using Fast Velocity-Encoded Cine MR Imaging: Validation Study Using Positron Emission Tomography Am. J. Roentgenol., October 1, 2000; 175(4): 1029 - 1033. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Hundley, L. D. Hillis, C. A. Hamilton, R. J. Applegate, D. M. Herrington, G. D. Clarke, G. A. Braden, M. S. Thomas, R. A. Lange, R. M. Peshock, et al. Assessment of Coronary Arterial Restenosis With Phase-Contrast Magnetic Resonance Imaging Measurements of Coronary Flow Reserve Circulation, May 23, 2000; 101(20): 2375 - 2381. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Hundley, C. A. Hamilton, G. D. Clarke, L. D. Hillis, D. M. Herrington, R. A. Lange, R. J. Applegate, M. S. Thomas, J. Payne, K. M. Link, et al. Visualization and Functional Assessment of Proximal and Middle Left Anterior Descending Coronary Stenoses in Humans With Magnetic Resonance Imaging Circulation, June 29, 1999; 99(25): 3248 - 3254. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hansen, D. Sayad, G. D. Thomas, G. D. Clarke, R. M. Peshock, and R. G. Victor Exercise-induced attenuation of alpha-adrenoceptor mediated vasoconstriction in humans: evidence from phase-contrast MRI Cardiovasc Res, January 1, 1999; 41(1): 220 - 228. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. C. Wu, E. A. Zerhouni, R. M. Judd, C. H. Lugo-Olivieri, L. A. Barouch, S. P. Schulman, R. S. Blumenthal, and J. A. C. Lima Prognostic Significance of Microvascular Obstruction by Magnetic Resonance Imaging in Patients With Acute Myocardial Infarction Circulation, March 3, 1998; 97(8): 765 - 772. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. C. L. Correia, E. Atalar, M. D. Kelemen, O. Ocali, G. M. Hutchins, J. L. Fleg, G. Gerstenblith, E. A. Zerhouni, and J. A. C. Lima Intravascular Magnetic Resonance Imaging of Aortic Atherosclerotic Plaque Composition Arterioscler. Thromb. Vasc. Biol., December 1, 1997; 17(12): 3626 - 3632. [Abstract] [Full Text] |
||||
![]() |
W. G. Hundley, R. A. Lange, G. D. Clarke, B. M. Meshack, J. Payne, C. Landau, R. McColl, D. E. Sayad, D. L. Willett, J. E. Willard, et al. Assessment of Coronary Arterial Flow and Flow Reserve in Humans With Magnetic Resonance Imaging Circulation, April 15, 1996; 93(8): 1502 - 1508. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||