(Circulation. 1995;91:445-450.)
© 1995 American Heart Association, Inc.
Articles |
From the Departments of Medicine (R.L.C., W.M.S., R.E.I.) and Pathology (D.L.R., R.E.I.), Duke University Medical Center, Durham, NC; the Engineering Research Center for Emerging Cardiovascular Technologies, the Department of Biomedical Engineering of the School of Engineering (W.M.S., R.E.I), Duke University, Durham, NC; and the I. Medical Clinic (E.U.A., E.M., M.P.C), Technische Universität Munich, Germany.
Correspondence to Raymond E. Ideker, MD, PhD, University of Alabama at Birmingham, Volker Hall Room G82A, Box 201, Birmingham, AL 35294.
| Abstract |
|---|
|
|
|---|
Methods and Results A tubular epicardial braided carbon
electrode of 7F diameter and 14-cm length applied in a
U-shape to the epicardium was compared with a standard left
ventricular epicardial 15-cm2 titanium mesh patch (CPI
Inc). As cathode, a CPI endocardial lead, a Medtronic lead, or a
carbon-platinum-iridium prototype electrode was used. Ventricular
fibrillation was induced with a 60-Hz generator and allowed to continue
for 10 seconds before a shock was given. Two different biphasic shock
waveforms (3.2/2- and 6/6-millisecond) were delivered by the six
electrode configurations. Eight dogs (weight, 24.5±1.3 kg) underwent
an up-down defibrillation protocol. The order of testing the epicardial
electrodes, the endocardial cathodes, and the waveform was randomized.
With the epicardial carbon electrode, the mean defibrillation threshold
(DFT) energy decreased 39% to 56% and the voltage decreased 24% to
35% compared with the titanium patch: from 8.3±2.5 to 4.9±3.6 J
with
the CPI lead and the 3.2/2-millisecond waveform, from 6.2±2.5 to
2.9±2.1 J with the carbon-platinum-iridium prototype, and from
6.4±3.4 J to 3.5±2.6 J with the Medtronic lead
(P
.05).
The DFT determinations with the 6/6-millisecond biphasic waveform
showed a similar trend with slightly higher values.
Conclusions Compared with a titanium patch, the new braided epicardial electrode significantly decreases the defibrillation energy requirements. This effect can be maximized by using an endocardial carbon-platinum-iridium prototype as cathode and a short duration biphasic waveform.
Key Words: electrodes defibrillation epicardial patches
| Introduction |
|---|
|
|
|---|
In this study, we tested a new epicardial prototype electrode made of braided carbon to compare its energy requirements for defibrillation with a standard epicardial titanium mesh patch in dogs.16 Additionally, we tested three different endocardial electrodes12 17 18 and two different waveforms19 to determine both the properties of this new material and the best endocardial/epicardial combination for defibrillation.
| Methods |
|---|
|
|
|---|
Animal Preparation
In eight mongrel dogs, anesthesia was
induced with intravenous
pentobarbitol (30 to 35 mg/kg body wt) and maintained with a continuous
infusion of pentobarbital at a rate of approximately 0.05 mg/kg per
minute.20 21 Succinylcholine (1 mg/kg) was also given
intravenously at the time of anesthesia induction. Supplemental doses
of succinylcholine (0.25 to 0.5 mg/kg) were given as needed to maintain
muscle relaxation. The animals were intubated with a cuffed
endotracheal tube and ventilated with room air and oxygen through a
Harvard respirator. A peripheral intravenous line was inserted, and
normal saline was continuously infused. A femoral arterial line was
placed for hemodynamic monitoring as well as for arterial blood gas
analysis and electrolyte measurements. Normal metabolic status was
maintained throughout the study by taking blood samples every 30 to 60
minutes and correcting any abnormal values. ECG leads were applied for
continuous monitoring of lead II. The chest was opened through a median
sternotomy, and the heart was suspended in a pericardial cradle. Body
temperature was measured and maintained at 35° to 37°C with a
thermal mattress and heat lamp. At the end of the study, euthanasia was
induced with a potassium chloride injection. The heart was then removed
and weighed.
Electrode Configurations
An epicardial titanium mesh patch
CPI 040 (Cardiac Pacemakers
Inc) with an active electrical surface area of 15 cm2 and a
total area of 18 cm2 was compared with a braided tubular
carbon electrode of 7F diameter and 14-cm length (Fig 1A
).
After randomization, one of these two epicardial
electrodes was positioned on the anterolateral wall of the left
ventricle. After the completion of testing of this electrode with all
endocardial electrodes and waveforms, it was removed and the other
epicardial electrode was positioned on the anterolateral wall of the
left ventricle. The patch was securely sutured with six stitches while
the U-shaped epicardial carbon wire was sutured with three
stitches, describing the outer bounds of the patch (Fig 1B
).
Three
different endocardial electrodes were used: (1) an 11F defibrillation
electrode CPI Endotak (Cardiac Pacemakers Inc) with a 3.7-cm right
ventricular coil, a 6.8-cm-long superior vena caval coil 9 cm proximal
from the distal electrode (not used for shocking in this study), and a
pacing electrode tip 5 mm distal from the right ventricular electrode;
(2) a 10.5F Medtronic 6966 Transvene electrode with a 5-cm
defibrillation electrode 25 mm proximal from the pacing tip; and (3) a
10F carbon-platinum-iridium prototype electrode with a 5-cm
defibrillation electrode 15 mm proximal to the silicon tip.
|
Only one endocardial electrode at a time was placed into the right ventricle for testing. The order in which each endocardial electrode was inserted and removed was determined by randomization. Each of the three leads was placed in the right ventricular apex through a right jugular incision under fluoroscopic guidance. Through a left jugular incision, a pacing catheter for induction of fibrillation and backup pacing was positioned in the right ventricle. The position of all endocardial leads was also verified manually.
Defibrillation Protocol and Data Acquisition
Defibrillation
success was determined by a threshold
value.22 23 Eight animals underwent defibrillation
trials.
The animal was randomized to have the epicardial patch or the carbon
lead placed first. After placement of the epicardial electrode, the
three endocardial electrodes were randomized and the first electrode
was positioned in the right ventricle. The DFT was determined with one
of the two biphasic waveforms. After determination of this threshold,
the second waveform was tested. The first endocardial lead was removed,
and this procedure was repeated with the second and then the third
endocardial lead. After all six different DFT determinations with the
first epicardial electrode were performed, it was removed. The second
epicardial electrode was placed, and the three endocardial electrodes
and the two waveforms were again tested in random order.
Defibrillation testing was performed by following a modified up-down protocol24 starting with a leading edge voltage of 400 V. The initial step size was 40 V. If the first shock failed, then incremental 40-V shocks were performed until a defibrillation success occurred. If the first shock succeeded, then decremental 40-V shocks were performed until the shock failed. After a reversal point was established, the step size was decreased to 20 V. The DFT was defined as the lowest voltage to defibrillate.
Ventricular fibrillation was
induced by 60-Hz alternating current
through the right ventricular apex pacing electrode, with the return
electrode attached to the animal's chest wall. Fibrillation was
allowed to continue for 10 seconds before attempting defibrillation. A
failed shock was followed by a rescue shock of higher voltage delivered
between the electrodes. If the rescue shock failed, it was followed by
internal defibrillation with paddles given by a Life-Pak 8
defibrillator (Physio-Control Corp). A minimum of 4 minutes elapsed
between each fibrillation-defibrillation attempt. Fibrillation was not
reinitiated until blood pressure and heart rate had returned to normal.
The defibrillation electrodes were connected to an external
defibrillator (HVS-02, Ventritex Inc). The defibrillator delivered a
dual-capacitor biphasic shock from a 150-µF capacitor bank. The
truncated exponential biphasic shock utilized a second phase of
opposite polarity to the first, with the second-phase leading edge
voltage equal to the first-phase trailing edge voltage rounded to the
nearest 10 V. For the first phase, the epicardial electrode was the
anode and the endocardial electrode was the cathode. Two biphasic
waveforms were tested, one with a first-phase duration of 3.2
milliseconds and a second-phase duration of 2 milliseconds, the other
with durations of 6/6 milliseconds. There was a 1-millisecond delay
between phases for both biphasic waveforms. The actual current and
voltage waveforms delivered to the electrodes were obtained by
isolating and recording the current across a 0.25-
resistor in
series with the electrodes and a 1:4 resistor divider in parallel with
the electrodes. These waveforms were digitized at 20 kHz and recorded
by a Data Precision 6100 waveform analyzer. Signal analysis
software within the analyzer was used to obtain impedance and energy
measurements. The data were transferred to a Sun workstation and to a
Macintosh computer for analysis.
Statistical Analysis
The data are expressed as the mean and
standard deviation for
energy, voltage, current, and impedance for all DFT determinations.
Statistical analysis was performed with ANOVA and Friedman tests,
and multiple comparisons between groups were based on rank
sums.25 Significance was defined as P
.05.
| Results |
|---|
|
|
|---|
The Table
compares the DFTs for two epicardial and the
three endocardial electrodes for the two biphasic waveforms. Expressed
are the mean and standard deviation for leading edge voltage, leading
edge current, mean impedance of the first phase, and total energy of
both phases at DFT. Statistical comparison was performed for the
epicardial patch versus the epicardial carbon wire.
|
Compared with the titanium patch, the epicardial carbon wire decreased
the leading edge voltage requirements for defibrillation with the 3.2/2
(6/6)-millisecond biphasic waveform: 27% (24%) with the CPI lead,
35% (32%) with the carbon-platinum-iridium right
ventricular-prototype, and 28% (30%) with the Medtronic lead
(P
.05 for 3.2/2 ms). Total energy also decreased 41%
(39%) with the CPI lead, 53% (56%) with the carbon prototype, and
45% (45%) with the Medtronic lead (P
.05 for 3.2/2 ms).
Mean values for current decreased between 25% and 29% (15% and 29%)
except for the Medtronic electrode. Impedance was 4% to 10% (6% to
14%) lower for the epicardial carbon configurations compared with the
respective configurations with the epicardial patch. With the exception
of the Medtronic electrode with the 6/6-millisecond waveform, voltage
and energy at DFT decreased significantly compared with the epicardial
patch irrespective of the right ventricular electrode or waveform when
the epicardial carbon electrode was used for defibrillation
(P<.05). The absolute values for the energy decreased with
the shorter waveform, whereas the mean voltage was unchanged or
increased slightly (Fig 2A
and 2B
).
|
The lowest DFT was achieved with a 3.2/2-millisecond pulse duration and the epicardial carbon wire combined with the endocardial carbon-platinum-iridium prototype (2.93±2.12 J). The highest DFT was found with the CPI right ventricular lead in combination with the titanium patch and the 6/6-millisecond waveform (8.71±3.20 J).
This beneficial effect of the epicardial carbon electrode and of the
shorter impulse duration on total energy can be seen in Fig 3
.
In this figure, the DFT determinations for the three
endocardial right ventricular electrodes were combined, and the mean
value and standard deviation were calculated. This demonstrates the
significant decrease (P
.01) in energy when defibrillating
by the epicardial carbon wire irrespective of the endocardial electrode
used, especially when using the shorter pulse duration of 3.2/2
milliseconds.
|
Independently comparing the individual data and the mean values for leading edge voltage, impedance, current, and total energy for the three different endocardial electrodes with the epicardial electrode or with the patch, we found that for the right ventricular carbon electrode there was a significantly lower voltage, current, and energy compared with the CPI electrode. The impedance was also significantly lower with the right ventricular carbon-platinum-iridium electrode compared with the CPI and Medtronic right ventricular electrode. The configuration of epicardial carbon electrode to right ventricular carbon-platinum-iridium prototype electrode resulted in six of eight dogs being defibrillated with energies below 2.4 J.
| Discussion |
|---|
|
|
|---|
In this study, we compared a newly developed tubular epicardial electrode made of braided carbon filaments to a standard epicardial patch. We also studied three different right ventricular electrodes and two different biphasic waveform durations. The reason for the study was to find a possible alternative to the epicardial patch that would defibrillate with low voltage and energy.
Our results for the endocardial/epicardial configurations gave DFTs that are comparable to those found in previous studies of our group. Dixon et al29 reported low DFTs for large contoured epicardial patches while Tang et al19 reported low energy requirements with large epicardial patches in conjunction with short biphasic waveforms. Guse et al30 demonstrated the possibility of low-energy defibrillation with four subcutaneous patches and endocardial leads. In this study, we also achieved low DFTs (2.93 J), but the endocardial/epicardial electrode configuration tested used a much smaller epicardial electrode.
A possible limitation of our study was the use of an open chest model. However, Kallok et al31 showed no difference in the DFT for a closed chest compared with an open chest determination using epicardial patches.
Both epicardial electrodes were positioned over the same region of the anterolateral wall of the left ventricle. The patch was sutured with several stitches to ensure good tissue contact. The carbon wire was sutured with three stitches in a U-shaped position describing the outer bounds of the patch. The geometrical area described by the two anodes was thus the same, but due to the outer insulation of the patch, the area within the U-shaped area with the carbon wire was slightly larger than the area covered by the titanium mesh of the patch. The direct electrode-tissue interface area for the patch is 15 cm2, compared with 3 cm2 calculated for the tubular carbon electrode (14-cm length, 7F diameter).
When the study was designed, our primary focus was to investigate an epicardial electrode that should be more easily applied on the epicardium, and we hoped to achieve similar or only slightly higher DFTs compared with standard patch placement. To our surprise, the voltage, energy, and current for defibrillation at DFT significantly decreased with the carbon electrode, yet impedance remained almost the same. A possible explanation for the lower defibrillation requirements despite primarily constant impedance values is a larger minimum potential gradient32 33 34 35 created by the carbon wire. Another possible explanation is a better electrode-tissue interface, as is seen with the carbon tips of bradycardia pacing electrodes,27 causing less polarization at the interface with more impedance contributed by the tissue because of the slightly different location of the carbon wire.
We used three different endocardial defibrillation catheters and two waveforms, as described in our previous study.18 We aimed to repeat our previous comparison of the three endocardial leads and to find the best of the six endocardial/epicardial electrode and waveform combinations for defibrillation. As in the previous study, the configurations with the carbon-platinum-iridium prototype electrode demonstrated the lowest DFTs.18 By combining the epicardial carbon wire and the endocardial carbon-platinum-iridium prototype and by using a short waveform, the defibrillation energy was significantly decreased by more than 50% compared with the least beneficial combination of electrodes and waveform duration.
The reason for the statistically significant lower impedance for the epicardial carbon compared with the patch when used in combination with the right ventricular carbon electrode remains unclear. A possible explanation is the high electrical surface of the carbon due to the microstructure of the multiple carbon fiber filaments, which may allow delivery of more current and voltage in a shorter time.18 28
The basic mechanism for the lower defibrillation energy requirements with a tubular electrode describing only the outer bounds of a patch is currently not fully understood. Some part of the mechanism may be attributed to carbon. A great deal of the lower DFT with this new concept may be attributed to the U-shape and location of the electrode. The extent of the effect attributable to the material and the extent to the shape and location needs to be evaluated in further studies.
In our experience, no other electrode material used has such high flexibility and pliabilitytherefore allowing good contact with the beating heartas the carbon-based electrode. Our results were obtained by placing the carbon electrode onto the heart via thoracotomy. Depending on further trials, in the future this may allow a small epicardial carbon electrode to be placed by a minimally invasive procedure.35 36 We conclude that the carbon electrode placed on the epicardium to describe the outer bounds of the region normally encompassed by a patch electrode is a promising possible alternative for the standard patch electrode and deserves further investigation.
Conclusions
Within the limitations of our model, these
results
demonstrate a significant decrease in voltage, energy, and current
requirements for defibrillation with an epicardial carbon wire
electrode compared with a standard epicardial patch. The defibrillation
energy can be additionally lowered by using an endocardial
carbon-platinum-iridium electrode and a short biphasic waveform. Carbon
may be a possible alternative material for defibrillation electrodes.
The flexibility and pliability of the epicardial carbon wire may be
beneficial in several respects. The mechanisms of replacing the patch
with a single electrode describing its outer bounds need further
evaluation, as do possible future means of noninvasive placement of the
epicardial carbon electrode.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. Gradaus, D. Bocker, A. Dorszewski, B. Lamp, D. Hammel, G. Breithardt, and M. Block Fractally coated defibrillation electrodes: Is an improvement in defibrillation threshold possible? Europace, January 1, 2000; 2(2): 154 - 159. [Abstract] [PDF] |
||||
![]() |
S. Windecker, R. E. Ideker, V. J. Plumb, G. N. Kay, G. P. Walcott, and A. E. Epstein The influence of ventricular fibrillation duration on defibrillation efficacy using biphasic waveforms in humans J. Am. Coll. Cardiol., January 1, 1999; 33(1): 33 - 38. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. U. Alt, P. C. Fotuhi, R. L. Callihan, E. Mestre, W. M. Smith, and R. E. Ideker Endocardial Carbon-Braid Electrodes : A New Concept for Lower Defibrillation Thresholds Circulation, September 15, 1995; 92(6): 1627 - 1633. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |