(Circulation. 1995;91:821-830.)
© 1995 American Heart Association, Inc.
Articles |
From the Cardiovascular Division, University of Virginia School of Medicine, Charlottesville.
Correspondence to Sanjiv Kaul, MD, Cardiovascular Division, Box 158, University of Virginia, Charlottesville, VA 22908.
| Abstract |
|---|
|
|
|---|
Methods and Results In 12 open-chest dogs, MCE was performed and myocardial blood flow (MBF) was measured by use of radiolabeled microspheres at baseline and during phenylephrine-induced coronary hyperemia. In the presence of this drug, stenoses were placed during different stages on the left anterior descending (LAD) and left circumflex (LCx) coronary arteries, and MCE and MBF assessments were performed. LA injections of 2 mL of 0.5 billion/mL microbubbles (mean diameter, 4.3 µm) were performed at each stage in all 12 dogs, and RA injections of 10 mL of 6 billion/mL microbubbles (mean diameter, 3.7 to 5.3 µm) were administered in 7 dogs. MCE images in which the contrast disparity between the LAD and LCx beds was maximal were digitally subtracted from precontrast images, and mean videointensities in these beds were measured after the dynamic range of gray-scale intensities was increased in the subtracted image and the image was color coded. The region showing hypoperfusion during LAD stenosis was planimetered and expressed as a percentage of the myocardial area in the short-axis slice. There was an excellent correlation between the LAD/LCx bed videointensity ratio and LAD/LCx bed MBF ratio (y=0.5x+0.44, r=.91, P<.001) during 57 LA injections. There was also an excellent correlation between the hypoperfused bed size on MCE during LA injection of contrast in the presence of LAD stenosis and the hypoperfused myocardium as determined by radiolabeled microspheres (y=0.8x+4.2, r=.90, P<.001, SEE=2.4, n=11). The anterior myocardium was opacified in 6 dogs receiving RA injections of contrast, and the hypoperfused area during LAD stenosis correlated closely with that determined by radiolabeled microspheres (y=0.86x+3.4, r=.93, P<.01).
Conclusions Coronary stenoses, which are not flow limiting at rest, can be detected and the degree and spatial extent of blood flow mismatch during pharmacologically induced coronary hyperemia can be quantified with MCE using LA and RA injections of contrast. Thus, it is possible that the severity of coronary stenoses and the quantum of myocardium in jeopardy could be quantified in the future with MCE using venous injection of contrast.
Key Words: stenosis blood flow echocardiography
| Introduction |
|---|
|
|
|---|
The present study was performed to test the following hypotheses: (1) coronary stenoses that are not flow-limiting at rest can be detected with MCE during pharmacologically induced hyperemia using LA and RA injections of contrast, and (2) the degree and spatial extent of blood flow mismatch in the myocardium supplied by these stenoses can be quantified by this approach.
| Methods |
|---|
|
|
|---|
A left lateral thoracotomy was performed, and the heart was suspended in a pericardial cradle. The proximal portions of the left anterior descending (LAD) and the left circumflex (LCx) coronary arteries were dissected free from the surrounding tissues, and custom-designed reversible snares were placed loosely around them. Doppler flow probes (2 mm, series SB, Transonics) were placed on these vessels proximal to the snare and were connected to a digital flowmeter (model T206, Transonics). A 7F catheter was placed in the LA for injection of microbubbles and radiolabeled microspheres. The RA was also cannulated with a similar catheter for microbubble injections.
Myocardial Contrast Echocardiography
MCE was performed with a
phased-array system (RT5000, General
Electric Medical Systems) with a 5-MHz transducer. A saline bath served
as an acoustic interface between the heart and the transducer. Gain
settings were optimized initially and were held constant throughout
each experiment. A maximal dynamic range of 72 dB was used. Imaging was
performed at the midpapillary muscle short-axis level at a depth
setting of 8 cm, and the data were recorded on 1.25-cm VHS videotape
with a high-fidelity video recorder (Panasonic AG6200, Matsushita
Electrical Co).
Sonicated albumin microbubbles (Albunex, Molecular Biosystems, Inc) were used for the LA injections.4 These bubbles have a mean diameter of 4.3±0.3 µm and a concentration of 0.5 billion/mL. Two mL of this agent and 3 mL of saline flush (total volume of injectate, 5 mL) were administered as a bolus over 1 second with a power injector (model 3000, Liebel-Flarsheim). We have previously shown that 2 mL of Albunex is the optimal LA dose for myocardial opacification.3 At this dose, a maximal signal-to-noise ratio is achieved without significant attenuation in the anterior myocardium. For RA injections, 10 mL of 3.7- to 5.3-µm bubbles with a concentration of 6 billion/mL was used. These bubbles were custom-designed for us by Molecular Biosystems, Inc, and were hand-injected as a bolus as previously described.1 2
MCE images were analyzed with an off-line computer (Mipron, Kontron Electronics)5 and were transferred from videotape to the image memory of the computer in a 244x244x8 bit format. For LA injections, end-systolic frames were used for analysis. These frames were selected because the left ventricular cavity is the smallest in these frames and the amount of contrast in the left ventricular cavity is the least, resulting in the least posterior wall attenuation. A precontrast frame and a similar frame with maximal disparity in the contrast effect between the LAD and LCx beds were selected and aligned by computer cross-correlation as previously described.5 Each pixel in the digitally subtracted image showing an increase in intensity was assigned a gray-scale value from 0 to 255 and was then assigned colors whereby gradations of red to orange, to yellow, to white, represented increasing gray-level values.1 2 The color coding of each pixel was performed in reference to the brightest pixel in each image, which was assigned a gray-scale value of 255. Pixels showing either no change or a decrease in gray level were not color-coded, and the left ventricular cavity was masked out.
The LAD and LCx beds were defined on MCE as regions demonstrating relative color deficiency during LAD and LCx stenoses, respectively, and transmural regions of interest (ROI) were placed over them. An attempt was made to include as much of the LCx bed as possible without including areas that were subject to attenuation from the presence of contrast in the left ventricular cavity. Care was also taken not to include the speckled epicardial and endocardial borders. The average size of the ROI over the LAD bed was 1766±488 pixels, and that over the LCx bed was 1392±371 pixels.
The LAD perfusion defect size varied with the severity of the stenosis; it was planimetered in each dog only during placement of the more severe stenosis and was expressed as a percentage of the myocardial area in the short-axis slice. Because of attenuation of the posterior wall, the posterior border of the LCx bed could not be defined; thus, the LCx perfusion defect size was not measured.
Unlike LA injections, end-systolic frames could not be used for RA injections because the myocardial contrast effect was significantly less in systole than in diastole. Furthermore, because of poor signal-to-noise ratio, subtraction of single precontrast and postcontrast frames did not result in optimal images. Therefore, as previously described,1 2 three consecutive end-diastolic precontrast frames were averaged to improve the signal-to-noise ratio, and three consecutive contrast-enhanced end-diastolic frames were similarly averaged. The averaged precontrast frame was digitally subtracted from the averaged postcontrast frame, and the subtracted image was assigned colors as described above. Since the posterior myocardium was attenuated during RA injections, ROI were placed only over the LAD bed and the hypoperfused zone was measured only during LAD stenosis. Because of posterior wall attenuation, a precontrast short axis frame was selected for the measurement of the myocardial area.
Myocardial Blood Flow Measurement
Approximately
2x106 11-µm radiolabeled
microspheres (Dupont Medical Products) suspended in 4 mL 0.9% saline
solution/0.01% Tween 80 were injected into the LA at each stage.
Reference samples were withdrawn from the femoral artery over 130
seconds with a constant-rate withdrawal pump (model 944, Harvard
Apparatus). At the end of the experiment, the short-axis slice of the
left ventricle corresponding to the MCE image was cut into 16
wedge-shaped pieces, and each piece was divided into endocardial,
midwall, and epicardial portions. The papillary muscles were included
in the endocardial portions. All samples were counted in a well counter
with a multichannel analyzer (model 1282, LKB Wallac). Corrections were
made for activity spilling from one window to the next with a
custom-designed computer program.6
Flow to each myocardial sample was calculated from the equation Qm=(CmxQr)/Cr, where Qm is myocardial flow (mL/min), Cm is tissue counts, Qr is rate of arterial sample withdrawal (mL/min), and Cr is counts in the arterial reference sample.7 Transmural blood flow (mL · min-1 · g-1) was derived by dividing the sum of flows to the individual segments by their combined weight. Myocardial blood flow (MBF) to the LAD and LCx beds was calculated by averaging transmural flows in the segments that demonstrated reduced flows during LAD and LCx stenosis, respectively. Segments interspersed between the two beds and demonstrating intermediate levels of flow during stenosis were not included in the analysis. For the stage with severe LAD stenosis, the combined weight of the hypoperfused segments was expressed as a percent of the weight of the entire short-axis slice.
Experimental Protocol
Each dog was studied at baseline and
during coronary hyperemia
induced by an intravenous infusion of 0.4 to 0.8
mg · kg-1 · min-1 of
phenylephrine
HCl (Schein Pharmaceuticals). We have previously shown that
phenylephrine HCl increases MBF by increasing MBV and that these two
are closely coupled.3 Stenoses of different severities
were created on the LAD during continuous infusion of phenylephrine,
and the degree of severity was judged by reduction in hyperemic LAD
flow recorded on the Doppler flow probe. The LAD stenoses were then
reversed, and a stenosis was created on the LCx during the continuous
infusion of phenylephrine HCl. The degree of stenosis was again judged
with the Doppler flow probe. Radiolabeled microspheres were injected
into the LA, and the aortic pressures were recorded at each stage. MCE
was performed with LA injection at each stage but with RA injection
only during phenylephrine-induced hyperemia before and after an LAD
stenosis was placed. The dog was then killed, the heart was excised,
and the slice corresponding to the image on MCE was processed to
determine MBF.
Statistical Methods
All data were analyzed with RS/1 (Bolt,
Beranek, and Newman)
resident on a minicomputer (VAX4000, Digital Equipment Corp) and were
expressed as mean±SD.8 Comparisons between MCE and other
measurements were made by linear regression analysis. Statistical
significance was defined as P<.05 (two-sided).
| Results |
|---|
|
|
|---|
Tables 1
and 2
depict the hemodynamic,
MBF, and MCE data for LA injections during baseline and LAD and LCx
stenoses. As would be expected, MBF increase during hyperemia was
associated with an increase in blood pressure induced by phenylephrine.
Videointensity in the LAD and LCx beds changed during
phenylephrine infusion in relation to changes in MBF. At baseline,
although MBF was lower than during stenoses placed in the presence of
phenylephrine, videointensities in the myocardial beds supplied by
these vessels were higher during stenoses. This phenomenon occurred
because videointensities in the subtracted images were adjusted
relative to the brightest pixel in each image. At baseline, the
brightest pixel and the least bright pixel were not very different;
during stenosis placed in the setting of coronary hyperemia, the
difference between the two was significantly greater, which resulted in
a lower pixel intensity in the stenotic bed even when absolute flow in
that bed was higher than at baseline.
|
|
Fig 1
illustrates
color-coded images from four
stages in a dog after LA injection of contrast. During baseline (panel
A), MBF to the LAD bed was 0.93
mL · min-1 · g-1, and that to
the LCx bed was 1.34
mL · min-1 · g-1; the LAD/LCx
bed MBF
ratio was 0.73. A greater preponderance of yellow is seen in the
lateral than in the anterior wall. The background-subtracted
videointensity in the anterior wall was 163, whereas that in the
lateral wall was 173, with an LAD/LCx bed videointensity ratio of 0.94.
During phenylephrine HClinduced hyperemia (panel B), MBF to the two
beds was nearly identical (2.7 and 2.6
mL · min-1 · g-1,
respectively), with an LAD/LCx bed flow ratio of 1.03. Both beds show
increased preponderance of whites and oranges. The
background-subtracted videointensity in the anterior wall was 203,
whereas that in the lateral wall was 207, with an LAD/LCx bed
videointensity ratio of 0.98. In the presence of an LAD stenosis (panel
C), MBF to the LAD bed was reduced to 0.51
mL · min-1 · g-1, and that to
the LCx bed was 1.34
mL · min-1 · g-1. The LAD/LCx
bed
flow ratio was 0.38. A lack of oranges and whites is noted in the LAD
(arrow) compared with the LCx bed. The background-subtracted
videointensity in the anterior wall was 91, whereas that in the lateral
wall was 153, with an LAD/LCx bed videointensity ratio of 0.60. During
LCx stenosis (panel D), MBF to the LAD bed was 2.7
mL · min-1 · g-1 and that to
the LCx
bed was 1.35 mL · min-1 · g-1,
with an LAD/LCx bed flow ratio of 2.0. This time, a lack of oranges,
yellows, and whites is noted in the LCx (arrow) compared with the LAD
bed. The background-subtracted videointensity in the anterior wall was
174, and that in the lateral wall was 111, with an LAD/LCx bed
videointensity ratio of 1.6.
|
Fig 2
illustrates the
relation between LAD/LCx MBF ratio
and LAD/LCx bed background-subtracted videointensity ratio during the
57 LA injections in all the 12 dogs. The relation is linear over a wide
range of values (0 to 4), and the correlation is highly significant.
Fig 3
depicts the relation between perfusion defect size
as a percent of myocardial area in the short-axis slice versus the
weight of the hypoperfused segments as a percent of the total
myocardial weight of that slice in the 11 stages with moderate to
severe LAD stenosis. The relation is linear, with an excellent
correlation coefficient, indicating that the perfusion defect size
during LA injection of contrast in the presence of a stenosis
corresponds to the amount of myocardium receiving relatively lower
flow during pharmacologically induced coronary vasodilation.
|
|
Right Atrial Injections
We were able to perform RA injections
in only 7 dogs because of
unavailability of the custom-designed bubbles for all animals.
Myocardial opacification was noted during phenylephrine-induced
hyperemia in 6 of the 7 dogs, a success rate in accordance with our
previous experience.1 2 Table 3
depicts MBF
and background-subtracted videointensity in the LAD bed before and
after LAD stenosis in the presence of phenylephrine HCl in the 6 dogs
with successful myocardial opacification. It is apparent that the
videointensity is decreased in the presence of reduced MBF to the LAD
bed. In only 1 dog was the LAD flow during stenosis below the resting
level.
|
Fig 4
illustrates end-diastolic color-coded
images from two stages from a dog after RA injection of contrast. Both
stages are during phenylephrine HClinduced hyperemia. The first stage
(panel A) is before the placement of LAD stenosis, when MBF to the LAD
bed was 5.1 mL · min-1 · g-1
and the
background-subtracted videointensity was 100 relative units. The second
stage (panel B) is after placement of the LAD stenosis, when MBF to the
LAD bed was 1.3
mL · min-1 · g-1 and
the background-subtracted videointensity was 37 relative units. Both
images demonstrate the significant posterior wall attenuation that is
noted with end-diastolic frames, since the number of
microbubbles in the left ventricular cavity is much more during end
diastole than during end systole (see Fig 2
).
|
Fig
5
illustrates the relation between perfusion
defect size as a percentage of myocardial area in the short-axis slice
versus the weight of the hypoperfused segments as a percentage of the
total myocardial weight of that slice during moderate to severe LAD
stenosis. The relation is close, indicating that perfusion defects
during RA injection of contrast show the extent of myocardium receiving
relatively lower flow in the presence of moderate stenosis during
pharmacologically induced coronary hyperemia.
|
| Discussion |
|---|
|
|
|---|
In the present study, we show that in a model in which MBF and MBV are closely coupled (such as during the administration of catecholamines), the ratio of background-subtracted videointensity from two myocardial beds correlates with the ratio of MBF to those beds during LA injection of contrast. In "Appendix A," we attempt to explain mathematically the occurrence of a linear relation between background-subtracted videointensity from two myocardial beds and relative MBV to those beds. We assume that the relation between background-subtracted videointensity and microbubble concentration was linear in our study. We used just enough microbubbles to cause a mild change (10 to 50 relative units in the original 2DE image) in videointensity, which would relate linearly with changes in microbubble concentration within tissue.3
As long as autoregulation remains intact, there are two major
mechanisms by which MBF increases via an increase in MBV (Fig
6
). The first mechanism is recruitment of
microvessels,15 and the second is microvessel
dilatation.16 17 During any particular condition, one
mechanism may be predominant or both may play significant roles in
increasing MBF. In "Appendix A" and Figs 7
and
8
, we show that, in either case, the
relation between videointensity ratio between two myocardial beds is
linearly related to the ratio of MBV of the two beds.
|
|
|
An interesting observation from our study is that myocardial videointensity decreases in the presence of a coronary stenosis. This decrease in videointensity could be explained by two possible mechanisms. The first mechanism implies that MBV actually diminishes during pharmacologically induced hyperemia in a bed supplied by a stenosis that limits an increase in flow to that bed compared with a normal bed. In previous experiments, we calculated MBV by dividing MBF measured with radiolabeled microspheres by microbubble transit rate (volume=flow/transit rate).3 18 We found that the calculated MBV of the bed supplied by the stenotic vessel actually decreased in the presence of pharmacologically induced hyperemia.3 18 Recent studies using direct16 19 20 21 22 and indirect11 12 observations of the coronary microvasculature have demonstrated that coronary microvessel size is both flow- and pressure-dependent. The pressure gradient across a stenosis that is not flow-limiting at rest is minimal, and the microvessels actively dilate to maintain MBF.16 19 In the presence of coronary hyperemia, however, the same stenosis becomes flow-limiting and results in a greater pressure gradient,16 resulting in a decrease in both flow and pressure, which can cause a decrease in microvessel size and hence, MBV.20 21 Although earlier studies using direct observations of the microvasculature had demonstrated the effect of flow and pressure on arteriolar size alone,20 recent studies have shown the same effect even in venules,21 which may be an important autoregulatory site and may be responsible for a significant fraction of the MBV during pharmacologically induced coronary hyperemia.22
Since in our study we did not measure microvessel size and MBV directly, we have to entertain another possible explanation for the linear relation between the background-subtracted videointensity and MBF ratios in the two beds. In "Appendix B," we attempt to explain mathematically the basis of such a relation in the absence of a decrease in MBV in the bed supplied by the stenotic vessel. It is apparent from our discussion that for MBF ratios <4.0, a linear relation is possible between MBF and videointensity ratios even when the MBVs of the two beds are equal. It should be noted, however, that such a relation is possible only when videointensity is measured, as was done in our study, at the moment of maximal disparity in the contrast effect between the two beds. It could be argued that since MBVs of the two beds are equal in this model, the transit rate ratios should theoretically provide a linear relation over a wider range of MBF ratios compared with videointensity ratios.14 Because of the great variability in the input function after venous injection, however, the derivation of myocardial transit rates may not be accurate and thus may not be a reliable indicator of MBF.3
Critique of Our Methods
To correlate background-subtracted
videointensity to MBV, it is
imperative that the relation between microbubble concentration and
videointensity be linear. In the 2DE system we used, this relation is
almost linear at videointensity levels <50 units; above this level,
the relation becomes flat.3 We have previously shown that
when 2 mL of Albunex is injected into the LA, the relation between
myocardial videointensity and microbubble concentration is linear and
attenuation does not occur in the anterior
myocardium.3
The videointensity values reported in this study were not obtained from the original 2DE data but rather from the digitally subtracted images, in which these values were readjusted over a wider dynamic range to accentuate minor differences in videointensities. Consequently, the videointensity values reported in this study are several times higher than the original values and reflect only relative changes. Our results are also influenced by the computer algorithms used for color-coding, which have been optimized to enhance subtle differences in videointensities so as to clearly define the borders between high and low flows. In this study, we were able to separate zones with as little as 15% difference in hyperemic flow, which is significantly superior to other imaging techniques.
Pharmacological increases in MBF could also have been achieved with vasodilators such as adenosine and dipyridamole. In the presence of these drugs, although maximal increase in MBV can occur, MBF becomes dependent primarily on the coronary driving pressure.23 Nevertheless, even if absolute MBF is altered in the presence of coronary vasodilators because of changes in systemic pressure, relative MBV and hence relative MBF to two beds need not be affected.24
Clinical Implications
The aim of this study was not to
demonstrate that myocardial
opacification is easily possible using RA injections of sonicated
albumin microbubbles. In fact, the bubbles used in this study for RA
injection are highly concentrated, and their use is not advisable in
the clinical setting. The aim of the study was to define the principles
based on which LA and RA (and hence, presumably venous) injections of
contrast could be used for the detection of coronary stenoses that are
not flow-limiting at rest and for the quantification of the degree and
spatial extent of flow mismatch that occurs in the presence of these
stenoses during pharmacologically induced coronary hyperemia. It is
very likely that such an approach will be used when venous contrast
agents become available. Newer contrast agents have recently been shown
to result in reproducible myocardial opacification from venous
injection when smaller doses and concentrations are
used.25 26 27 The principles advocated in
this article could
also be applied to the use of these agents.
MCE in the cardiac catheterization laboratory has usually been performed with direct intracoronary injections of microbubbles. The results of our present study suggest that injection of microbubbles into a chamber such as the left ventricle could be used to determine the physiological significance of coronary stenoses. Preliminary data from our laboratory demonstrate that injection of microbubbles into the aortic root also provides similar results.18 24 Unlike the situation with LA, venous, and left ventricular injection of contrast, attenuation is not a problem when contrast is injected into the aortic root.
Although the newer contrast agents25 26 27 result in less posterior wall attenuation, this phenomenon will always occur as long as there are microbubbles in the left ventricular cavity. Different imaging planes (such as apical views) could be used to better define the posterior wall. In case of difficulties in adequately imaging the posterior wall or in acquiring overall high-quality images from the transthoracic approach, the transesophageal approach could prove useful. Further experimental and clinical studies are necessary to define the clinical utility of MCE for the detection of coronary disease and estimation of the ischemic burden by use of venous injection of contrast.
|
| Acknowledgments |
|---|
| Appendix A |
|---|
|
|
|---|
Microvessel Recruitment
If the time-intensity curve obtained from the passage of
microbubbles through a single microvessel within an ROI (Fig
6A
) during
MCE has a gamma-variate form,28
![]() | (1) |
where
/2 is the mean transit rate of the microbubbles through
the ROI, then the time-intensity curve from n number of vessels lying
within the same ROI will have the form
y=nAte-
t. If two ROI have n1
and
n2 number of vessels, respectively, each of the same radius
(Fig 6B
), then the corresponding time-intensity curves would be
y1=n1Ate-
t and
y2=n2Ate-
t,
respectively.
Dividing the time-intensity curve from one bed by that
from the other,
we get
y1/y2=n1/n2.
Since in this model of microvessel recruitment, MBF (f) is proportional
to MBV (v) and hence
is the same in both ROI, the ratio of the
videointensities from the two beds at any given time will relate to the
ratio of number of microvessels and thus MBV in the two beds (Fig
7A
).
In practical terms, the measurement of peak videointensities is easy,
and because
is constant in this model, peak intensities occur at
the same time in both beds (Fig 7B
).
Microvessel Dilatation
Unlike recruitment,
mathematical modeling of microvascular
dilatation is more complex. Not only does f change because of change in
v, but
also changes. Thus, the value A in the gamma-variate
function also changes. According to Poiseuille's law,29
flow (f) through a tubing is proportional to the fourth power of the
radius of the tubing and, therefore, the second power of volume (v) of
the tubing, provided that there is a constant pressure gradient across
the tubing and the flow through it is laminar. Then,
![]() | (2) |
where k1 is a constant.
Thus, unlike recruitment, in which an increase in the number of microvessels by twofold results in an increase in MBV by twofold, in the case of microvascular dilatation, an increase in microvascular radius by twofold will result in an increase in MBV by fourfold.
To
mathematically describe A in Equation 1
, one must relate the
microbubble concentration to f. If w is the total number of bubbles
injected into the LA, c is the cardiac output, and f is flow through an
ROI (Fig 6A
), then the number of microbubbles, m, entering the
ROI is
given by the equation:
![]() | (3) |
According to the Stewart-Hamilton equation,30
![]() | (4) |
where a is the area under the time-intensity curve obtained from a myocardial bed and k2 is a constant.
![]() | (5) |
and
![]() | (6) |
Substituting
Equations 2 through
6![]()
![]()
![]()
![]()
into Equation 1
and
simplifying, we get
![]() | (7) |
where B=k12k2(w/c).
Applying Equation
7
to two different myocardial beds of volumes
v1 and v2 (Fig 6C
) and corresponding
background-subtracted videointensities of i1 and
i2, respectively, we get
![]() | (8) |
and
![]() | (9) |
The
ratio of videointensities in the two beds is independent of
B but is dependent on v1/v2,
k1, and t. Since the moment of the greatest
disparity in contrast effect between the two beds is the easiest and
most practical to determine (Fig 8A
), we simulated
time-intensity
curves for i1 and i2 for various values of
v1 and v2 at this time.
For simulation of data,
k1 was derived from previously
published data by Wu and colleagues.11 In this article,
the relation between f and v was determined to be
v=9.5f1/2. Therefore,
f=(1/9.52)v2, or
f=0.011v2, or k1=0.011. Substitution of
this value of k1 in Equations 8 and 9 allowed generation of
time-intensity curves for i1 and i2 for various
values of v1 and v2. Plotting the ratio of
i1/i2 against
v1/v2, we find that the two are
closely related (Fig 8B
). Although theoretically the relation
is
quadratic, in practical terms, because of the value of
k1, the relation is almost linear.
| Appendix B |
|---|
|
|
|---|
Using
Equations 3
, 4
, and 6
above, we
get
![]() | (10) |
The
background-subtracted videointensities (i1 and
i2) from the two myocardial beds can be described by
i1=A1te-
1t
and i2=A2te-
2t.
![]() | (11) |
The difference in videointensities between the two beds, z, is
![]() | (12) |
The
time T at which the difference in the videointensities in
the two beds is maximal is obtained by differentiating Equation
12
with
respect to time and equating to zero:
![]() | (13) |
from which we get
![]() | (14) |
The ratio of videointensities from the two beds when the difference in videointensities is maximal is given by
![]() | (15) |
Substituting
for the right side of Equation 14
, we get
![]() | (16) |
Also
substituting for A1 and A2 from
Equation 11
in Equation 13
, we get
![]() | (17) |
We
numerically obtained values for T from Equation 17
for
various values of
1 and
2 and substituted
these in Equation 16
to obtain the ratio of videointensities
(i1/i2) at the time of maximal disparity
in the videointensities. A plot of i1/i2
versus
1/
2 is shown in Fig 9C
(since volume is maximal in both beds,
equals f). It can be noted
that the relation between videointensity and
ratios from the two
beds is linear up to an
ratio of approximately 4.0, which may
explain the good correlation between MBF ratio and videointensity ratio
in Fig 2
even if MBV did not decrease with a decrease in MBF.
When the
flow ratios are >4, one would not expect a linear relation between
these ratios.
Received May 26, 1994; accepted August 19, 1994.
| References |
|---|
|
|
|---|
2.
Villanueva FS, Glasheen WP, Sklenar J, Kaul S. Assessment of
risk area during coronary occlusion and infarct size after reperfusion
with myocardial contrast echocardiography using left and right atrial
injections of contrast. Circulation. 1993;88:596-604.
3.
Skyba D, Jayaweera AR, Goodman NC, Ismail S, Camarano G, Kaul
S. Quantification of myocardial perfusion with myocardial contrast
echocardiography during left atrial injection of contrast: implications
for venous injection. Circulation. 1994;90:1513-1521.
4. Keller MW, Glasheen WP, Kaul S. AlbunexR: a safe and effective commercially produced agent for myocardial contrast echocardiography. J Am Soc Echocardiogr. 1989;2:48-52. [Medline] [Order article via Infotrieve]
5. Jayaweera AR, Matthew TL, Sklenar J, Spotnitz WD, Watson DD, Kaul S. Method for the quantitation of myocardial perfusion during myocardial contrast two-dimensional echocardiography. J Am Soc Echocardiogr. 1990;3:91-98. [Medline] [Order article via Infotrieve]
6. Kaul S, Kelly P, Oliner JD, Glasheen WP, Keller MW, Watson DD. Assessment of regional myocardial blood flow with myocardial contrast two-dimensional echocardiography. J Am Coll Cardiol. 1989;13:468-482. [Abstract]
7. Heyman MA, Payne BD, Hoffman JI, Rudolf AM. Blood flow measurements with radionuclide-labeled particles. Prog Cardiovasc Dis. 1977;20:52-79.
8. BBN Software Products. RS/1 User's Guide. Cambridge, Mass: Bolt Beranek & Newman Inc; 1988.
9. Kaul S. A look at 15 years of planar thallium-201 imaging. Am Heart J. 1989;118:581-601. [Medline] [Order article via Infotrieve]
10. Maddahi J, Kiat H, Friedman JD, Berman DS, Van Train KF, Garcia EV. Technetium-99m-sestamibi myocardial perfusion imaging for evaluation of coronary artery disease. In: Nuclear Cardiology. State-of-the-Art and Future Directions. St Louis, Mo: CV Mosby Co; 1993:191-200.
11.
Wu X, Ewert DL, Liu Y, Ritman EL. In vivo relation of
intra-myocardial blood volume to myocardial perfusion: evidence
supporting microvascular site for autoregulation.
Circulation. 1992;85:730-737.
12.
Eigler NL, Schuhlen H, Whiting JS, Plaff JM, Zeiher A, Gu S.
Digital angiographic impulse response analysis of regional
myocardial perfusion: estimation of coronary flow, flow reserve, and
distribution volume by compartmental transit time measurement in a
canine model. Circ Res. 1991;68:870-880.
13.
Keller MW, Segal SS, Kaul S, Duling BR. The behavior of
sonicated albumin microbubbles within the microcirculation: a basis for
their use during myocardial contrast echocardiography.
Circ Res. 1989;65:458-467.
14.
Jayaweera AR, Edwards N, Glasheen WP, Villanueva FS, Abbott
RD, Kaul S. In-vivo myocardial kinetics of air-filled albumin
microbubbles during myocardial contrast echocardiography: comparison
with radiolabeled red cells. Circ Res. 1994;74:1157-1165.
15. Tillmanns H, Leinberger H, Neumann FJ, Steinhausen M, Parekh N, Zimmerman R, Dussel R, Kuebler W. Myocardial microcirculation in the beating heart: in vivo microscopic studies. In: Spaan JAE, Bruschke AVG, Gittenberger-de Groot AC, eds. Coronary Circulation. Dordrecht, The Netherlands: Martinus Nijhoff Publishers; 1987:88-94.
16.
Kanatsuka H, Sekiguchi N, Aaki K, Wang Y, Komaru T, Ashikawa
K, Takishima T. Microvascular sites and mechanisms responsible for
reactive hyperemia in the coronary circulation of the beating canine
heart. Circ Res. 1992;71:912-922.
17. Fedor JM, McIntosh DM, Rembert JC, Greenfield JC Jr. Coronary transmural myocardial blood flow responses in awake domestic pigs. Am J Physiol. 1978;235:H435-H444.
18. Camarano G, Jayaweera AR, Ismail S, Goodman NC, Skyba D, Kaul S. Estimation of myocardial blood volume by contrast echocardiography can provide an estimation of myocardial blood flow. J Am Coll Cardiol. 1994;23:450A. Abstract.
19.
Kanatsuka H, Lamping KG, Eastham CL, Marcus ML. Heterogenous
changes in epimyocardial microvascular size during graded coronary
stenosis: evidence for the microvascular site for autoregulation.
Circ Res. 1989;66:389-396.
20.
Kuo L, Chilian WM, Davis MJ. Interaction of pressure- and
flow-induced responses in porcine coronary resistance vessels. Am
J Physiol. 1991;261:H1706-H1715.
21.
Kuo L, Arko F, Chilian WM, Davis MJ. Coronary venular
responses to flow and pressure. Circ Res. 1993;72:607-615.
22.
Chilian WM, Layne SM, Klausner EC, Eastham CL, Marcus ML.
Redistribution of coronary microvascular resistance produced by
dipyridamole. Am J Physiol. 1989;256:H383-H390.
23. Marcus ML. Autoregulation in the coronary circulation. In: The Coronary Circulation in Health and Disease. New York, NY: McGraw-Hill Book Co; 1983:93-112.
24. Camarano G, Jayaweera AR, Ismail S, Goodman NC, Kaul S. Measurement of myocardial blood volume is a preferable indicator of microvascular reserve than measurement of coronary blood flow: a study utilizing myocardial contrast echocardiography. J Am Soc Echocardiogr. 1994;7:S16. Abstract.
25. Dittrich HC, Bales GL, Kuvelas T, D'Sa A, Greener Y. Myocardial perfusion using new intravenous ultrasound contrast agents: comparison between video and integrated backscatter images with a prototype software program. J Am Coll Cardiol. 1994;23:393A. Abstract.
26. Cotter B, Kwan OL, Cha YM, Dittrich H, Bhargava V, DeMaria AN. Dose-response characteristics, time-course, and hemodynamic responses to QW3600, an ultrasonic contrast agent capable of myocardial opacification by intravenous injection. J Am Coll Cardiol. 1994;23:393A. Abstract.
27. Arai A, Kenny A, Shiota T, Unger E, Yellowhair D, Passafini A, Matsunaga T, Grauer SE, Pantley G, Sahn DJ. Transpulmonary passage of AerosomesTM, a pressure stable, lipid-based echocardiographic contrast agent: studies in pigs. J Am Coll Cardiol. 1994;23:25A. Abstract.
28.
Thompson MK, Starmer CF, Whorten RE, McIntosh MD.
Indi- cator transit time considered as a gamma variate.
Circ Res. 1964;14:502-514.
29. Burton AC. Physiology and Biophysics of the Circulation. 2nd ed. Chicago, Ill: Year Book Medical Publishers; 1972:40.
30. Kinsman JM, Moore JW, Hamilton WF. Studies on the circulation, I: injection method: physical and mathematical considerations. Am J Physiol. 1929;89:322-330.
This article has been cited by other articles:
![]() |
S. Kaul Myocardial Contrast Echocardiography: A 25-Year Retrospective Circulation, July 15, 2008; 118(3): 291 - 308. [Full Text] [PDF] |
||||
![]() |
P. A. Dijkmans, R. Senior, H. Becher, T. R. Porter, K. Wei, C. A. Visser, and O. Kamp Myocardial Contrast Echocardiography Evolving as a Clinically Feasible Technique for Accurate, Rapid, and Safe Assessment of Myocardial Perfusion: The Evidence So Far J. Am. Coll. Cardiol., November 8, 2006; (2006) j.jacc.2006.05.079v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. R. Porter, F. Xie, M. Silver, D. Kricsfeld, and E. O'Leary Real-time perfusion imaging with low mechanical index pulse inversion Doppler imaging J. Am. Coll. Cardiol., March 1, 2001; 37(3): 748 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Leistad, K. Ohmori, T. A. Peterson, G. Christensen, and A. N. DeMaria Quantitative assessment of myocardial perfusion during graded coronary artery stenoses by intravenous myocardial contrast echocardiography J. Am. Coll. Cardiol., February 1, 2001; 37(2): 624 - 631. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Scherrer-Crosbie, W. Steudel, R. Ullrich, P. R. Hunziker, N. Liel-Cohen, J. Newell, J. Zaroff, W. M. Zapol, and M. H. Picard Echocardiographic determination of risk area size in a murine model of myocardial ischemia Am J Physiol Heart Circ Physiol, September 1, 1999; 277(3): H986 - H992. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Sakuma, Y. Hayashi, K. Sumii, M. Imazu, and M. Yamakido Prediction of short- and intermediate-term prognoses of patients with acute myocardial infarction using myocardial contrast echocardiography one day after recanalization J. Am. Coll. Cardiol., October 1, 1998; 32(4): 890 - 897. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Kaul Myocardial Contrast Echocardiography : 15 Years of Research and Development Circulation, November 18, 1997; 96(10): 3745 - 3760. [Full Text] |
||||
![]() |
J. Sklenar, G. Camarano, N. C. Goodman, S. Ismail, A. R. Jayaweera, and S. Kaul Contractile Versus Microvascular Reserve for the Determination of the Extent of Myocardial Salvage After Reperfusion: The Effect of Residual Coronary Stenosis Circulation, September 15, 1996; 94(6): 1430 - 1440. [Abstract] [Full Text] |
||||
![]() |
T. R. Porter and F. Xie Transient Myocardial Contrast After Initial Exposure to Diagnostic Ultrasound Pressures With Minute Doses of Intravenously Injected Microbubbles : Demonstration and Potential Mechanisms Circulation, November 1, 1995; 92(9): 2391 - 2395. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |