(Circulation. 1995;91:845-863.)
© 1995 American Heart Association, Inc.
Articles |
From the Department of Biomedical Engineering, Cardiac Bioelectricity Research and Training Center (D.S.K., Y.R.), Case Western Reserve University, Cleveland, Ohio; and the Nora Eccles Harrison Cardiovascular Research and Training Institute (B.T., R.L.L., P.R.E.), The University of Utah (Salt Lake City).
| Abstract |
|---|
|
|
|---|
Methods and Results A boundary element based mathematical method, combined with a numeric regularization technique, was developed for computing the inverse solution. Endocardial potentials were computed from intracavitary potentials measured with a multielectrode probe placed in the cavity of an isolated, perfused canine left ventricle. Data were acquired during rhythms induced by electrical stimuli applied at different locations and varying depths within the myocardium. Endocardial potentials were measured using intramural needles to evaluate the accuracy of the inverse solutions by direct comparison. Inversely computed endocardial potentials, from measured probe potentials, reconstruct with good accuracy the major features (potential maxima and minima, regions of negative and positive potentials) compared with the measured endocardial potentials. During early activation, the computed endocardial potentials exhibit a potential minimum in close proximity to the pacing site, determining the location of the stimulus with good accuracy (within 10-mm error). Multiple stimuli, as close as 10 to 20 mm to each other, can be distinguished and localized to their sites of origin by the inverse reconstruction. Similar to the measured endocardial potentials, the spatial distribution of the computed endocardial potentials reflects the underlying cardiac fiber direction, and dynamic changes of the computed endocardial potentials reflect the rotation of fibers with intramural depth. Maps of isochrones show good correspondence between the isochrones determined from the computed endocardial potentials and those determined directly from the measured endocardial potentials.
Conclusions Compared with actual, measured endocardial potentials and activation sequences, endocardial potential patterns and activation sequences can be reconstructed on a beat-by-beat basis from cavitary potentials measured with a multielectrode, noncontact probe. The approach presented here is shown to reconstruct, with 10-mm accuracy and resolution of 10 to 20 mm, local events of cardiac excitation (eg, pacing sites). In addition, the reconstructed endocardial potentials correctly reflect the underlying fibrous structure of the myocardium. These results demonstrate the feasibility of the approach. In the experiments, the probe position and endocardial geometry were determined invasively. To be clinically applicable, the reconstruction method should be combined with a noninvasive method for determining the probe-cavity geometry in the catheterization laboratory. It could then be developed into a catheter-based technique for locating arrhythmogenic sites and for studying and diagnosing conduction abnormalities, reentrant activity, and the effects of drugs and other interventions on cardiac activation and arrhythmias.
Key Words: endocardium mapping arrhythmia electrophysiology
| Introduction |
|---|
|
|
|---|
Current techniques of mapping the potentials directly from the endocardium present certain difficulties. At present, intravascular electrode-catheter mapping1 is limited in the number of recording sites, and the procedure is very time consuming. Moreover, mapping from multiple sites is carried out sequentially over several cardiac cycles without accounting for possible beat-to-beat variability in the activation pattern. A more complicated mapping procedure involves the use of an endocardial multielectrode balloon or sponge.2 3 4 The procedure requires open heart surgery, heart-lung bypass, emptying the blood from the heart, and inflating the balloon so that the electrodes are in direct contact with the endocardium. In addition to the difficult procedure and risk involved, very often the arrhythmia to be mapped cannot be induced during surgery. An alternative indirect mapping approach was introduced earlier by Taccardi and his colleagues5 through the development of an intracavitary multielectrode catheter-probe (olive shaped or cylindrical). Unlike the balloon, the probe can be introduced into the blood-filled ventricular cavity without occluding it. The probe permits the simultaneous recording of intracavitary potentials from multiple directions. However, unlike the balloon, the probe is not necessarily in direct contact with the endocardium. Probes based on the same principle, eg Foley-like inflatable catheters,6 carrying the electrodes on the surface of a small balloon, could be used during routine catheterization studies. It was demonstrated recently by Khoury and Rudy,7 in an idealized model of the probe-heart-torso volume conductor, that the spatial pattern of cardiac excitation is accurately reflected in the endocardial potentials but not in the associated cavitary probe potentials, which exhibit smoothed-out, low-amplitude distributions.
The overall objective of the present study was to develop and test a mathematical method to reconstruct the endocardial potential distribution from cavitary potentials measured with a multielectrode probe that is not in direct contact with the endocardium (the "inverse problem"). Endocardial potentials are computed during paced rhythms in the isolated, perfused canine left ventricle (LV). Specifically, the reconstruction procedure is assessed for its ability to locate the site of origin of single and multiple pacing stimuli, reflect the effect of cardiac fiber direction on the spatial pattern of the potentials, and reconstruct the major events during the spread of excitation.
The procedure of reconstructing endocardial potentials and, subsequently, the activation sequences from intracavitary, noncontact probe data can be divided into two subprocedures. The first consists of determining the probe-cavity geometry, which is required for determining the transfer relation between the probe and endocardial surfaces. The second consists of the actual computation of the endocardial potentials, assuming that the probe-cavity geometry is known. In this article, we focus on the second subprocedure of mathematically reconstructing the endocardial potentials from noncontact probe data. This essential step is difficult and crucial to the success of the entire approach. It must be carefully evaluated, with minimal effects from other complicating factors to assess feasibility of the technique. We therefore develop and test the reconstruction method, using a geometric relation that is determined invasively. Using specific protocols, we assess quantitatively the effects of errors in geometry (ie, errors in probe position and orientation) on the reconstructed endocardial potentials. Of course, to be clinically applicable the geometry must be determined during catheterization. Combining the reconstruction procedure with a noninvasive method for determining the geometry will be the next step in the development of this approach as a clinical tool.
The work presented here is a first step in the development of a catheter-based method for the simultaneous mapping of endocardial potentials and endocardial activation patterns (isochrones). If successful, such a method will permit detailed examination of global as well as regional cardiac electrical events in the clinical cardiac electrophysiology laboratory and in the intact experimental animal. With this approach, information can be obtained on the nature of conduction abnormalities, site of origin and type of arrhythmia, pathway of reentrant activity (including its spatial organization and its important components such as the area of slow conduction), and beat-to-beat dynamic changes during the arrhythmia (eg, initiation and termination). Moreover, the method will permit localization of the arrhythmogenic site before catheter ablation.
| Methods |
|---|
|
|
|---|
The cavity potential distribution is described
mathematically by
Laplace's equation. Computation of endocardial potentials from
measured intracavitary probe potentials requires solving Laplace's
equation in the cavity volume (
) bounded by the probe surface
(Sp) and the endocardial surface (Se), as
illustrated in Fig 1
. The potential (v) is obtained by
solving Laplace's equation:
|
![]() | (1) |
subject to the following Cauchy boundary conditions:
![]() | (2) |
That is, the potential is known on the probe surface, and
![]() | (3) |
That is, current cannot enter the probe, which is assumed to be nonconducting.
A standard boundary element method technique11 was used to discretize the probe and endocardial surfaces and to solve for the potentials in a realistic-geometry probe-cavity system. The method is the same as we previously applied to computing epicardial potentials from body surface potentials ("the electrocardiographic inverse problem").8 9 10 12 13 14 Following the discretization of the surfaces, the ensuing relation is obtained:
![]() | (4) |
where
Vp is a vector of probe potentials of order
Np (number of probe nodes or electrodes), Ve is
a vector of endocardial potentials of order Ne (number of
endocardial nodes or electrodes), and A is a matrix
(NpxNe) of influence coefficients that
represents the geometric relation between the two (endocardial
and probe) realistic surfaces. The problem of solving the endocardial
potentials (Ve) for a given set of measured probe
potentials (Vp) in Equation 4
is ill-posed in the
sense
that small perturbations in the data (Vp) due to
measurement noise or to systematic errors in determining the geometry
result in large variation in the solution (Ve). The
solution of endocardial potentials (Ve) was therefore
stabilized by applying zero-order Tikhonov regularization
technique15 in conjunction with the CRESO a
posteriori method16 for determining the
regularization parameter (further explanation is provided in References
8 through 10). In this approach, the solution of endocardial potentials
(Ve) is obtained as an optimal estimate of these potentials
that is also stable. Stability is achieved by imposing a physiological
constraint on the solution, which, for Tikhonov zero order, implies
endocardial potentials of bounded amplitudes. This technique was
previously applied, with good results, to the reconstruction of
epicardial potentials from body surface potentials using a setup of an
isolated dog heart in a human geometry electrolytic tank
torso.13 14 The numeric algorithms used in the study
presented here were initially tested by reconstructing endocardial
potentials from intracavitary probe potentials in an idealized
eccentric spheres model of the probe-heart-torso volume
conductor.7 17
Experimental Model: Isolated Heart Preparation
The inverse
method was applied to the reconstruction of LV
endocardial potentials from actual, measured LV intracavitary probe
potentials in the isolated canine heart. Two experiments were
performed, with two dogs used in each experiment. The dogs were
anesthetized with pentobarbital (30 mg/kg). Mechanical ventilation was
maintained from an external respirator through an endotracheal tube. In
each experiment, a large dog was used to support an isolated heart
obtained from a smaller dog. The support dog was used to supply
oxygenated blood, with blood flowing into the aorta of the isolated
heart to perfuse the coronary arteries (similar to Langendorff
perfusion; Fig 2
). The heart was suspended in a warm
chamber and contracted freely. Heat exchangers maintained normal blood
temperature. Arterial pressure was monitored for the support dog by
cannulating the left femoral artery. Once the preparation was stable,
the right ventricular (RV) wall was surgically removed to expose the
septum, thereby providing an access for inserting needle electrodes
from the right side of the heart through the septum to record septal
potentials from the LV side. The preparation was stable for 4 hours on
average. We implemented at least six preparations identical to the one
described here (with the RV wall removed), for different purposes, and
the preparation was stable for many hours. We also used dozens of
"entire" heart preparations, and the duration was even
longer.
|
Intracavitary and Endocardial Potential Measurements
Intracavitary potentials were measured using two different
probes (one for each experiment). Recorded probe potentials served as
the input data for the inverse reconstruction of LV endocardial
potentials. In the first experiment (case 1; Fig 2
), a
65-electrode
cylindrical probe was inserted through a purse string in the left
atrial appendage of the isolated heart and positioned along the
blood-filled LV cavity. The probe electrodes were distributed along 8
circumferential rings on the surface and a tip electrode at a tapered
end. The probe was 11.5 mm wide, and the distance between the tip
electrode and the proximal ring was 28 mm. In the second experiment
(case 2; Fig 2
), an 89-electrode cylindrical probe was inserted
through
a purse string in the LV apex of the isolated LV preparation. The probe
electrodes were distributed along 11 circumferential rings on the
surface and a tip electrode at a tapered end. The probe was 12.5 mm
wide, and the distance between the tip electrode and the proximal ring
was 40.1 mm. A photograph of the actual probe used in case 2 is shown
in Fig 3A
. In both probes, the electrodes in each ring
were spaced 45 degrees from each other.
|
Endocardial potentials were
measured by inserting 52 or 94 intramural
multielectrode needles into the ventricular walls of the isolated LV
from all directions (Fig 3B
; photograph of actual heart
preparation
used in case 2). The needles protruded slightly into the blood cavity,
so that the tip electrodes measured potentials on the cavity side of
the endocardium (the tip electrodes formed an "endocardial
envelope" or an effective "endocardial balloon"). Recorded
endocardial potentials served as the gold standard for evaluating the
accuracy of the inversely computed potentials. Other intramurally
located electrodes on the needles were used for recording and pacing at
different depths in the myocardium. Two different types of electrode
needles were used. A 13-mm-long needle, with 3 electrodes 1.6 mm apart
at the endocardial end, was used for recording. A 14.4-mm-long needle,
with 10 electrodes 1.6 mm apart, the 10th electrode being epicardial,
was used for pacing and recording.
Pacing/Recording Protocol
Potentials were recorded during
rhythms induced by LV
stimulation. Current pulses of just suprathreshold intensity, 2 ms in
duration, were applied. The heart was paced at a cycle length of 360 ms
in case 1 and at 350 ms in case 2. Pacing was performed from 12 (case
1) or 14 (case 2) pacing needles (at different intramural depths). The
stimuli were applied between electrodes 1 and 2 (distal end of pacing
needle) for subendocardial pacing, between electrodes 5 and 6 for
midwall pacing, and between electrodes 9 and 10 (proximal end of pacing
needle) for subepicardial pacing. The heart was paced from one needle
or from two needles simultaneously. A total of 52 (pacing and
recording) intramural needles were inserted throughout the LV wall and
the septum in case 1, and 94 needles were inserted in case 2.
Unipolar electrographic signals from the probe electrodes and the intramural needle electrodes were simultaneously amplified, sampled at 1 kHz per channel, digitized (12 bits), and stored in a Microvax II computer. The common reference electrode was placed on the lump of noncontractile tissue close to the base of the isolated heart.
Volume Conductor Geometry Measurements
At the end of the
experiment and after the completion of
different pacing protocols, the LV cavity was filled with gelatin to
preserve the geometry of the endocardial surface. The needles were then
replaced by metal rods that were 101.6 mm long, and the heart was
stored in a formalin solution. A few days later, the geometry of the
endocardial envelope formed by the needle tips (approximating the
endocardial surface) was digitized. This was accomplished by
determining the cylindrical coordinates of the needle tips (see
schematic illustration of the measurement setup in Fig 4
). The
coordinates were determined for two points on
the exterior portion of the metal rod (external end point and entry
point on the epicardium); the coordinates of the tip (electrode 1) were
subsequently computed by linear extrapolation (ie, using the equation
of a straight line in three-dimensional space). Similarly, the probe
surface was digitized by determining the cylindrical coordinates of the
electrodes on its surface using the same setup shown in Fig 4
.
|
The probe and endocardial surfaces were reconstructed by connecting the
electrodes (nodes) with triangles (the electrodes being the vertices of
the triangles). The probe position and orientation relative to the
endocardial surface (including horizontal and vertical shifts,
rotation, and tilt with respect to the vertical axis of the LV) were
mathematically determined using measured endocardial and probe
potential data. This was accomplished by minimizing the
root-mean-square error between the actual measured probe potentials and
the forward-computed probe potentials (using the forward relation
between endocardial potentials and probe potentials in Equation
4
in
"Mathematical Formulation"). The root-mean-square error was
minimized over five consecutive time frames early in the activation
process. This approach is similar in concept to an earlier method by
Macchi et al18 19 where the site of origin of ectopic
events was localized by minimizing the difference between measured
probe potentials and potentials computed for an oblique dipole layer
model that approximated cardiac sources during early ectopic
activation.20 Fig 5
is an anterior view of
the discretized (triangulated) probe and LV endocardial surfaces with
the probe positioned within the cavity (anterior wall is removed in Fig
5
to show the probe). The two surfaces describe the geometry
required
for the boundary value problem illustrated in Fig 1
. A separate
probe
position was determined for each pacing protocol, as the position
changed over the period of the experiment due to ventricular
contraction, variation in the location of the pacing site, and changes
in ventricular blood filling. This served as a best-case scenario for
each protocol for computing endocardial potentials from probe
potentials while minimizing error due to inaccuracy in probe position.
Using this best-case scenario as a baseline, study of the effects of
error in probe position and orientation on the accuracy of
reconstruction was conducted by displacing or reorienting the probe
from its original position and/or orientation. Note that the above
method of determining the probe position and orientation within the
cavity cannot be extrapolated to clinical application, as it requires a
priori knowledge of the actual endocardial potentials. Clinically, a
noninvasive method for determining probe position and orientation (eg,
echocardiography) will have to replace the invasive approach used in
the present experiment.
|
Data Processing and Display
After the experiment, recorded
unipolar electrographic signals
were baseline adjusted, calibrated, and signal averaged (6 cycles). Few
electrode needles were discarded due to malfunction during the
experiment. In case 1, a total of 50 actual recording needles were
used, and the endocardial surface was therefore represented
by 50 nodes. In case 2, a maximum of 76 actual recording needles were
used (some pacing protocols had more malfunctioning needles than
others). An additional 6 nodes were used to completely close off the
endocardial surface at the base and apex, resulting in a total of 82
nodes to represent the endocardial surface. The probe surface
was closed off by adding a node at the proximal end of the probe. In
both cases, potentials at nodes with missing data were interpolated by
minimizing the laplacian at all nodes of the surface
considered.21 During subendocardial pacing (stimulus
applied between electrodes 1 and 2 of the pacing needle), the tip
electrode of the pacing needle was assigned the value of the potential
measured by the third electrode. In case 2, the 8 electrodes in the
proximal circumferential ring of the probe were removed, as they were
touching the myocardium in the apical region. This, in effect, resulted
in an 81-electrode probe. Therefore, the matrix A in Equation 4
was of
order 66x50 for case 1 and 82x82 for case 2. All the needle tips
were
in the cavity, some very close to the real endocardium, and some a few
millimeters into the blood. Thus, the reconstructed surface was not the
real endocardial surface but rather an envelope that closely lined the
endocardium.
The instantaneous potential distribution on the unrolled
surface of the
probe is displayed in the form of equipotential contour maps. The probe
and endocardial potentials are presented as seen by an observer
looking at the probe surface and endocardial surface from within the
probe. The endocardial potential maps are displayed on the actual
endocardial surface; cut at the left lateral LV, and projected on a
two-dimensional surface with a view of the anterior LV, septum, and
posterior LV as illustrated in Fig 6
. Equipotential
contour maps are presented for every millisecond. Regions bounded
by the most positive or least negative equipotential contour are
shaded. Endocardial isochrone (activation) maps were obtained by
determining the time of occurrence of the negative peak of the first
derivative of QRS at every electrode site on the endocardial surface.
Maps of isochrones determined on the probe surface are called
"pseudoisochrones,"5 as the probe electrodes were
not in direct contact with the endocardium. The end of the stimulus
pulse was selected as the reference point. The sites where the stimuli
were delivered are indicated by asterisks on all the endocardial
maps.
|
| Results |
|---|
|
|
|---|
|
|
Probe and
endocardial potentials generated by a subepicardial
stimulus applied at an anteroseptal location (same pacing needle as in
Fig 7
) are shown in Fig 9
. Initially, at 17 ms,
a
potential minimum -1.97 mV in magnitude appears in the measured
endocardial potential distribution (Fig 9A
-I) that is 8 mm away
from
the endocardial tip of the pacing needle. As the wave front moves from
the epicardium to the endocardium, the potential minimum progressively
increases in magnitude, becoming -18.18 mV at 28 ms (Fig
9B
-I) and
-40.11 mV at 33 ms, when it appears at the endocardial site of the
pacing needle (Fig 9C
-I). Similarly, the simultaneously
measured probe
potentials exhibit a potential minimum facing the pacing region that
progressively increases in magnitude: -0.52 mV at 17 ms (Fig
9A
-II),
-2.78 mV at 28 ms (Fig 9B
-II), and -38.39 mV at 33
ms (Fig 9C
-II).
The potential minimum, in the corresponding inversely computed
endocardial potentials, appears from the start at the endocardial
position of the pacing needle. It increases in magnitude with time
progression: -0.67 mV at 17 ms (Fig 9A
-III), -10.01
mV at 28 ms (Fig 9B
-III), and -41.97 mV at 33 ms (Fig
9C
-III).
|
The alignment of the peak positive region with
respect to the negative
region, early in the activation process, is similar in the measured and
inversely computed endocardial potentials. However, the alignment is
different for a subepicardial stimulus (Fig 9A
-I and Fig
9A
-III) from
that for a subendocardial stimulus (Fig 7A
-I and Fig
7A
-III). The
initial pattern of activation due to a stimulus at a certain depth
within the myocardium is generally a reflection of the local fiber
direction at that level. In Fig 9A
-I and 9A-III, the alignment
of the
positive and negative regions on the endocardium is more
circumferential than expected based on the direction of the epicardial
fibers themselves (see fiber directions in Fig 8
). This
difference
probably results from the fact that by the time distinct negative and
positive potential regions can be detected on the endocardial surface,
the wave front has penetrated into deeper subepicardial layers that are
rotated clockwise and are more circumferential than the epicardial
fibers. As time progresses, a clockwise expansion of the positive
region with respect to the negative region is observed in both the
measured and the computed endocardial potentials as well as in the
measured probe potentials. This is a consequence of the clockwise
rotation of the fibers with depth from the epicardium to the
endocardium (as viewed from the endocardial surface), as illustrated in
Fig 8
. Later in the activation process (33 ms), the positive
potential
region expands to form a circular pattern, with the potential minimum
located at the endocardial node of the pacing needle in both the
measured and the computed endocardial potentials. The formation of a
circular pattern with high degree of symmetry and no obvious
positive-negative axis probably results from the cumulative (averaging)
effect of the rotating fibers, as activation spreads from epicardium to
endocardium through many layers of different fiber
directions.35 36 A similar effect, as seen from the
epicardium, was described by Watabe et al31 and recently
by Taccardi and colleagues.32 33
A map of
isochrones determined from the measured endocardial potentials
for the pacing protocol of Fig 7
(subendocardial stimulus) is
shown in
Fig 10A
. The earliest activation time is 2 ms, as
determined from the electrogram of the third distal electrode on the
pacing needle (see "Methods"). The corresponding inversely
reconstructed isochrone map, determined from the inversely computed
endocardial potentials, is shown in Fig 10B
; the earliest
activation
time is 23 ms. Both isochrone maps indicate that activation starts in
the anteroseptal region (in the vicinity of the pacing needle) and
progresses to the posterior region toward the base of the LV. A map of
pseudoisochrones (see "Methods") determined on the probe surface
is shown in Fig 11A
. The earliest activation time
determined directly from the measured probe potentials is 22 ms. For
subepicardial pacing (Fig 9
), a map of isochrones determined
from the
measured endocardial potentials is shown in Fig 12A
.
The earliest activation time is 29 ms, later than that for
subendocardial stimulation (Fig 10A
). The corresponding
isochrone map,
determined from the inversely computed endocardial potentials, is shown
in Fig 12B
; the earliest activation time is 32 ms. A map of
pseudoisochrones determined on the probe surface is shown in Fig
11B
for subepicardial pacing. The earliest activation time determined
directly from the measured probe potentials is 32 ms.
|
|
|
Endocardial and
probe potential distributions during the initial phase
of activation due to stimuli applied at varying depths along the same
transmural needle are presented in Fig 13
. The
pacing needle, indicated by the asterisk, is located in the
anterolateral LV region. The potential distributions generated 11 ms
after a subendocardial pacing stimulus are shown in Fig 13A
.
The
measured endocardial potentials (Fig 13A
-I) exhibit a
potential minimum
-19.77 mV in magnitude located at the pacing site (potential measured
by the third distal electrode on the pacing needle). A -0.32 mV
minimum appears in the measured probe potential distribution (Fig
13A
-II). The inversely computed endocardial potentials display
a -2.63
mV potential minimum that is 8.2 mm away from the pacing site (Fig
13A
-III). A sequence of successive histological sections of
the
ventricular wall in the anterolateral region (containing the pacing
needle) showing fiber directions (viewed from the endocardial side) are
provided in Fig 14
. Note that in the endocardial
potential maps of Fig 13A
-I and 13A-III, the peak positive
regions are
aligned with respect to the negative region in a direction that is
parallel, in a qualitative sense, to the subendocardial fiber
direction. A single positive area appears in the computed endocardial
potentials as opposed to two positive regions in the measured
potentials. The potential distributions, generated by a midwall
stimulus at 10 ms after stimulation, are shown in Fig 13B
. The
magnitude of the measured endocardial potential minimum is -7.27 mV,
and the relative minimum in the probe potential is 0 mV. A potential
minimum -1.11 mV in magnitude is reconstructed in the computed
endocardial potential distribution and is located at the endocardial
node of the pacing needle (3.6 mm away from the minimum of the
corresponding measured potentials). Notice that the positive and the
negative regions reflected in the measured and computed endocardial
potential maps (in Fig 13B
-I and 13B-III, respectively) are
oriented
parallel to the circumferential (horizontal) fiber direction at the
midwall level, as can be deduced from Fig 14
. The inversely
computed
potentials are generally noisy at low levels of measured probe
potentials. For a subepicardial stimulus (Fig 13C
), the
measured and
computed endocardial potential distributions at 14 ms reflect an
alignment of the positive and negative regions that is indicative of
the fiber direction at the subepicardial level (see Fig 14
).
The
measured and reconstructed potential minima, -1.56 mV in Fig
13C
-I and
-0.67 mV in Fig 13C
-III, are at the same node, 4.1 mm
from the pacing
needle. The corresponding probe potential minimum is -0.06 mV.
|
|
Results of subendocardial stimulation in the apical region are shown in
Fig 15
. The measured endocardial potentials at 10 ms
for a pacing stimulus (asterisk) applied in the posteroseptal portion
of the apex (Fig 15A
-I) exhibit a potential minimum
-22.25 mV in
magnitude that is located at the pacing node (potential measured by the
third distal electrode on the pacing needle). A corresponding broad
potential minimum -3.43 mV in magnitude appears in the measured probe
potentials (Fig 15A
-II). The inversely computed endocardial
potentials
(Fig 15A
-III) reveal a potential minimum -7.40 mV in
magnitude located
4.2 mm away from the pacing site. For a stimulus applied in the
posterolateral portion of the apex (Fig 15B
-I), the measured
endocardial potentials at 14 ms reveal a potential minimum -1.55 mV in
magnitude located at the pacing node. Measured probe potentials (Fig
15B
-II) reflect a minimum -1.31 mV in magnitude. The
inversely
computed endocardial potentials (Fig 15B
-III) reconstruct a
potential
minimum -2.27 mV in magnitude that is exactly at the same pacing node.
Note that the probe position and orientation were the same for both
pacing protocols in Fig 15
.
|
Thus far, the results of
the pacing protocols of Figs 7
, 9
,
13
, and 15
demonstrate that measured endocardial potentials exhibit a potential
minimum, early in activation, at or in close proximity to the pacing
site. This is in keeping with the observations of Harada et
al.4 Measured probe potentials exhibit minima that appear
on its surface at electrodes facing the pacing sites. More important,
the location of the pacing site can be reconstructed with good accuracy
(less than 10 mm error) on the endocardial surface from the inverse
solution. Early during activation, the magnitudes of the measured probe
potentials are small. This results in low magnitude and low
signal-to-noise ratio of the input data for the inverse reconstruction.
As a result, the inversely computed potential maps are generally noisy
and do not depict the activation pattern at this early stage. As the
cardiac sources become more extensive with the progression of
activation, probe potentials and, consequently, the inversely computed
endocardial potentials increase in magnitude and reflect the activation
pattern. In effect, an apparent delay appears in the computed
endocardial potential maps compared with the measured ones (see
"Discussion"). This phenomenon is a manifestation of the volume
conductor geometric and conductive effects that attenuate the probe
potentials.7
Early in the activation process, the relative orientation of the positive and negative regions on the endocardial surface is generally a reflection of the fiber direction at the intramural depth of the stimulating electrode. In addition, for subsequent time frames, regions of positive potentials expand around the negative regions in a rotary fashion, reflecting the rotation of the fibers as the activation penetrates increasing depths within the myocardium. These characteristics are clearly reflected in the inversely reconstructed endocardial potentials. For the relatively large probe sizes used in these experiments, a broad pattern of rotational expansion can be deduced from the measured probe potentials as well.
Distinguishing Multiple Cardiac Events
The ability to resolve
and localize multiple ectopic events was
investigated by pacing the LV from two sites simultaneously and
recording the instantaneous potentials generated by pacing stimuli at
varying distances from each other. This provides a measure of the
spatial resolution capability of the reconstruction procedure. In Fig
16
, two pacing stimuli separated by 29.5 mm were
applied subendocardially: one stimulus in the anterobasal region, and
the other in the posterobasal region (asterisks). At 15 ms, the
measured endocardial potentials reveal two minima at the two pacing
sites: a -46.44 mV minimum on the anterior surface and a -28.82 mV
minimum on the posterior surface (potentials measured by the third
distal electrode on the pacing needle). Two potential minima also
appear in the simultaneously measured probe potentials (Fig
16C
): a
-2.66 mV minimum facing the anterior wall and -2.36 mV minimum
facing
the posterior wall. The inversely computed endocardial potentials
reconstruct the two potential minima exactly at the location of the
pacing nodes (Fig 16B
): a -15.23 mV minimum on the
anterior wall and a
-7.71 mV minimum on the posterior wall. Single pacing stimuli were
then applied separately at the same pacing sites. Measured probe
potentials generated at 15 ms by the single anterior stimulus are shown
in Fig 16D
and for the single posterior stimulus in Fig
16E
. The
potential minimum is -3.00 mV in Fig 16D
and -3.71
mV in Fig 16E
. It
can be observed that probe potentials generated by the two simultaneous
stimuli of Fig 16C
approximate a superposition of the
potentials
generated by the application of the stimuli separately (Fig
16D
plus
16E).
|
A map of isochrones determined from the measured endocardial
potentials
for the pacing protocol of Fig 16
is shown in Fig
17A
.
The map indicates two sites of early activation: the earliest
activation time is 9 ms in the anterior region and 6 ms in the
posterior region. The corresponding inversely reconstructed isochrone
map, determined from the inversely computed endocardial potentials, is
shown in Fig 17B
. Two distinct regions of early activation can
be
observed in Fig 17B
: the earliest activation time is 20 ms in
the
anterior region and 14 ms in the posterior region. A map of
pseudoisochrones determined on the probe surface from the measured
probe potentials themselves is shown in Fig 18
for the
double stimuli protocol of Fig 16C
. Notice that a single broad
region
of early activation is reflected on the probe surface; the earliest
activation time is 14 ms. This suggests that for multiple cardiac
events, maps of pseudoisochrones constructed directly from the measured
probe potentials do not always reflect the complexity of the underlying
cardiac sources and can be misleading in determining the cardiac
activation pattern.
|
|
In the pacing protocol of Fig 19
,
the two pacing sites
were moved closer to each other (separated by 22.4 mm); the posterior
pacing site (in Fig 14
) was moved toward the lateral LV while
maintaining the anterior site in the same location as before. Probe
position was the same as in the previous protocol. At 11 ms, measured
endocardial potentials display two minima at the anterior and lateral
sites (Fig 19A
-I), -44.16 mV and -21.84 mV in
magnitude,
respectively. Two potential minima initially appear on the probe
surface (Fig 19A
-II): -0.76 mV and -0.08 mV in
magnitude for the
respective anterior and lateral sites (at 11 ms). The inversely
computed endocardial potentials (Fig 19A
-III) accurately
depict the
pacing sites by the location of the potential minima: -4.79 mV for the
anterior and -2.88 mV for the lateral. At 16 ms, measured endocardial
potentials reveal two minima in Fig 19B
-I: -44.96 mV for
the anterior
site and -23.30 mV for the lateral site. Notice that the
simultaneously measured probe potentials reveal only a single broad
minimum -5.2 mV in magnitude (Fig 19B
-II). This
demonstrates the
smoothing effect of the volume conductor on probe potentials; an
observation that was presented earlier in an idealized model of the
probe-heart-torso volume conductor for different configurations of
equivalent myocardial sources.7 However, when the probe
potential data of Fig 19B
-II are used to compute the inverse
solution,
the reconstructed endocardial potentials (Fig 19B
-III) exhibit
two
potential minima at the locations of the two pacing nodes. The anterior
potential minimum is -19.26 mV and the lateral minimum is -9.05 mV
in
magnitude.
|
Probe Position and Orientation
In the clinical environment,
the probe position and orientation
will have to be determined using noninvasive imaging methods. This will
introduce an uncertainty in the geometry that will translate to an
error in the matrix A of Equation 4
(in "Methods")
and, in turn,
in the computed endocardial potentials. It is important, therefore, to
study the sensitivity of the approach to uncertainty in the geometry.
The effect of error in determining probe position on the inversely
recovered endocardial potentials was assessed by adding a 5-mm shift
(error) to the probe position determined in the experiment (see
"Methods"). The measured probe potentials of Fig 7
were used to
inversely compute the endocardial potentials in the presence of a 5-mm
shift of the probe, toward the anterior LV (Fig 20
).
The potentials in Fig 20
were computed at the same time frames
of Fig 7
. The results of Fig 7
provided a
baseline for comparison in the
absence of the imposed geometric error. In Fig 20A
(at 13 ms),
a
potential minimum -3.84 mV in magnitude appears at the pacing node;
the same location as in Fig 7A
-III. Furthermore, note the
similarity in
the orientation of the positive region with respect to the negative
region in the two figures. At a later time frame (18 ms), the positive
region expands in a counterclockwise fashion with respect to the
negative region (Fig 20B
) in a similar fashion as in Fig
7B
-III. The
potential minimum is -4.08 mV and is located at the pacing node. At 23
ms (Fig 20C
), the negative region becomes more extensive with
a
potential minimum of -24.25 mV, accompanied by further
counterclockwise expansion of the positive region.
|
The effect of error
in probe position due to 5-mm shift toward the
septum (instead of toward anterior LV, the situation explored in Fig
20
) was studied in Fig 21
for the same pacing
protocol
of Fig 7
. The inversely computed potentials in Fig
21
exhibit a similar
potential distribution to that of Fig 7
(in the absence of
position
error) in terms of the orientation of the positive and negative
regions, as well as the counterclockwise expansion. The potential
minimum appears 9 mm away (at the closest septal node) from the pacing
site in both Fig 21A
(13 ms) and Fig 21B
(18
ms) (-3.21 mV and -4.58
mV in magnitude, respectively). The potential minimum in Fig
21C
(23
ms) is -36.68 mV and appears 1.2 mm away from the pacing site.
|
The effect of error in determining the probe orientation (rotation
along its axis) on the inverse solution was tested by rotating the
probe by a certain angle relative to the orientation as determined in
the experiment and using the measured probe potential data to compute
the inverse solution. The data of the pacing protocol of Fig
16
were
used in evaluating the effect of error in probe orientation. For an
error of 5 degrees (Fig 22A
), the potential minima
(anterior, -14.57 mV; posterior, -5.85 mV) do not change in
location.
For an error of 10 degrees (Fig 22B
), the anterior potential
minimum
(-10.91 mV) remains in the same location, whereas the posterior
minimum (-3.66 mV) shifts toward the septum (direction of the
rotational error) by 11 mm, to the node closest to the pacing
site.
|
| Discussion |
|---|
|
|
|---|
The endocardial surface is easily accessible with intravascular electrode catheters during routine electrophysiology studies. However, current electrode catheter mapping techniques are limited in the number of recording sites, are time consuming, and do not account for possible beat-to-beat variations in the activation pattern or changes in the mechanism of the arrhythmia. In addition, brief arrhythmias may not be adequately mapped. The endocardial balloon or sponge mapping technique provides direct measurements of the endocardial surface potentials from multiple sites simultaneously. However, the balloon (or sponge) mapping procedure is limited in use, as it requires a complicated surgery. It is not always possible to induce sustained tachycardias in the operating room, and, furthermore, arrhythmias induced during surgery may be of different morphology or mechanism than those observed preoperatively. The approach presented in this study, for the inverse reconstruction of endocardial potentials from intracavitary potential data, involves the use of multielectrode catheter probes that can be percutaneously introduced into the cavity in a way that is similar to electrode catheters used in electrophysiology studies. Although the probes used in the present study were relatively large, the same methodology used here for the inverse reconstruction could be applied to a smaller size cylindrical catheter probe, such as the 9F (3 mm) multielectrode catheter used earlier by Taccardi and his colleagues.5 This "noncontact" mapping approach is aimed at providing isopotential and isochronal maps of the endocardial surface in a way that is similar to the endocardial balloon approach but without using the balloon or without the need for surgery. In addition, similar to the balloon, the inverse problem mapping technique is carried out over a single cardiac cycle. That is, mapping would be conducted on a beat-by-beat basis and could essentially allow for mapping nonsustained arrhythmias or infrequent ectopic events during routine electrophysiology studies. Moreover, with the advent of catheter-based ablation techniques,37 38 the inverse mapping approach would be ideal for localizing arrhythmogenic sites and specific components in these sites (eg, the area of slow conduction in a reentry pathway) before ablation, without the need for surgery.
As mentioned above, the same mathematical methods used in this study would be applied to a smaller catheter probe. Although a smaller probe would result in greater smoothing of its potentials,39 certain computational procedures would improve. For example, the surface elements of the probe (triangles) will be smaller, and error due to interpolation over the probe surface elements will be reduced in the mathematical formulation of the boundary element method.11 Qualitatively, the inverse solutions obtained in case 1 (65-electrode probe) were as good as the solutions obtained in case 2 (81-electrode probe, effectively). In both cases, the pacing sites could be localized with good accuracy. The effect of rotational anisotropy of the fibers was reflected in the inverse solutions in both cases equally well (in terms of clockwise and counterclockwise rotation). Further experiments are needed to compare the inverse solution obtained with different probe diameters and designs, as well as to determine the minimum number of electrodes on the probe surface required for obtaining sufficient resolution.
The computational methods used in this work were very efficient. All
computations were run on a desk-top workstation. The total time to
construct the geometry matrix A (in Equation 4
) and to compute
the
inverse solution for all the equipotential maps throughout the cardiac
cycle (sampled at 1 frame per millisecond) takes less than 60 seconds.
This makes the inverse reconstruction method practical for mapping in
the clinical as well as the experimental laboratory.
The specific objective of the present study was to develop and test a mathematical method to reconstruct the endocardial potential distribution from intracavitary potentials measured with a noncontact, multielectrode probe. Intracavitary probe potentials were measured, in an isolated and perfused canine LV, during rhythms induced by electrical stimuli applied at different locations and varying depths within the myocardium. These measured probe potentials, together with a knowledge of the geometry of the probe-cavity volume conductor, were used to reconstruct the potentials on the endocardial surface. The actual endocardial potentials were also measured simultaneously and served as the gold standard for evaluating the accuracy of the reconstruction. The reconstructed endocardial potentials were assessed for their ability to locate the site of origin of electrical stimuli; to distinguish between multiple, simultaneous cardiac events at different sites; to reflect the effect of the underlying cardiac fiber direction; and to recover the sequence of endocardial activation.
The results confirm that LV wall pacing gives rise to a potential minimum on the probe surface, facing the region of stimulation. Moreover, the results demonstrate that the inversely computed endocardial potentials, from the measured probe potentials, reconstruct with good accuracy the major features (potential maxima and minima, regions of negative and positive potentials) compared with the actual measured endocardial potentials. Note that the reconstructed potential magnitudes are reduced compared with the measured amplitudes. This is a property of the regularization scheme that constrains the magnitudes of the reconstructed potential in a least-squares sense. However, the spatial pattern of the potential distribution and the sequence of activation (isochrones) are preserved by the regularization approach. During early activation, the computed endocardial potentials exhibit a potential minimum on the endocardial surface in close proximity to the pacing site, determining the location of the stimulus with good accuracy (within 10-mm error) in a similar way to the measured endocardial potentials. Simultaneously applied stimuli from multiple locations, as close as 10 to 20 mm apart, can be separated (distinguished) from each other and localized to their corresponding sites of origin by the reconstruction procedure. The ability to reconstruct local cardiac events would be valuable in locating the site of origin of ventricular arrhythmias. This includes tachycardias resulting from ectopic activity or from circus movement reentry.40 Furthermore, the reconstruction method could detect changes in the mechanism of the tachycardia or a shift in the ectopic focus to a new site. The resolution of the reconstructed maps suggests that a reentrant circuit, with a radius as small as 10 to 20 mm, could be detected. Mapping the Wolff-Parkinson-White syndrome with one or, more important, two Kent bundles and two preexcitation sites is another possible clinical application, particularly when one of the sites is septal and therefore is more difficult to detect with the usual procedures.
Previously, Macchi and colleagues18 19 attempted to localize the site of origin of paced ventricular beats in the canine heart using a similar intracavitary probe. The approach was to represent the cardiac ectopic source by two equivalent dipoles of equal strength and opposite polarity, with the ectopic focus located at the midpoint of the line connecting the dipoles. The position and moment of the equivalent source were then obtained by minimizing the difference between the computed (forward problem) and the measured potentials on the surface of the intracavitary probe in an infinite conducting medium. This approach is limited to locating a single site of ectopic activity at an early time frame (QRS onset); unlike the reconstruction method presented here, it does not provide a spatial distribution of the potentials on the endocardial surface and cannot be used to locate multiple sites of ectopic activity. Furthermore, the method is not suited for studying the sequence of activation or progression of excitation.
The results of this study demonstrate that the spatial distribution of the reconstructed endocardial potentials (positions and relative orientation of maxima and minima and of the negative and positive regions) correlates well, in a qualitative sense, with the measured potentials and reflects the underlying cardiac fiber direction. The measured and computed endocardial potentials show that the initial pattern of activation due to a stimulus at a certain level (depth) within the myocardium reflects the fiber direction at that level. Earlier investigations23 24 25 26 27 28 29 30 31 32 33 showed that myocardial tissue anisotropy affects the potentials generated during cardiac excitation by determining the shape of the activation wave front (the isochrone), influencing the strength and distribution of the cardiac sources and modifying the potential fields generated by these sources. Studies showed that as a consequence of anisotropy, ectopic excitation spreads faster in a direction parallel to the cardiac fibers than perpendicular to them (resulting in ellipsoidal-like isochrones). The axial component of the cardiac sources due to point (ectopic) stimulation plays a dominant role in determining the associated potential field compared with the much weaker transverse component. As a result, the orientation of the region of positive endocardial potentials with respect to the negative region reflects the direction of the fibers at the level of the stimulus within the myocardium. This property is captured in the reconstructed endocardial potential distribution. In addition, the progression of depolarization is reflected well in the development of both the measured and computed endocardial potentials. As depolarization progresses, the negative potentials increase in magnitude and, at the same time, regions of positive potentials expand in a clockwise or counterclockwise fashion around the negative regions, as a consequence of the influence of the fiber rotation on the spread of the excitation wave fronts. The influence of fiber rotation through deep myocardial layers ("rotational anisotropy") on wave front propagation was demonstrated in the canine myocardium28 29 30 31 32 33 and in a macroscopic model of the ventricular tissue that incorporates the anisotropic properties of the myocardium.35 36 In the present study, the measured and the computed endocardial potential distributions show that for a subendocardial stimulus, the region of positive potentials expands in a counterclockwise fashion around the negative region. This is due to the influence of counterclockwise rotation of the fibers, from endocardium to epicardium, on the propagating wave front. Conversely, the potential distributions exhibit a clockwise rotation for stimuli originating from subepicardial sites. Stimuli applied at midwall tend to result in rotational expansion of the positive regions in both directions. These results demonstrate that not only do the reconstructed endocardial potentials localize the pacing site on the endocardial surface but they can also detect the depth of the site of origin as revealed by the relative orientation of positive and negative regions and by the progression of the potential pattern during depolarization.
The limited spatial resolution we achieved in our experiments,
particularly in the experiment of Fig 7
, where only 52 needles
were
distributed through the entire wall plus septum, made it difficult to
define the potential distributions in the positive areas in great
detail. Despite this, correlation between endocardial potential
patterns and fiber direction can be inferred from the data
presented as discussed above and demonstrated in "Results" by
direct correlation with fiber direction determined histologically.
Theoretically, according to the oblique dipole layer theory,10 20 one expects a pacing stimulus to produce a potential pattern with one minimum and two maxima. However, this typical pattern was not necessarily observed in all of the measured and reconstructed potential maps for one or more of the following reasons: (1) low resolution due to the limited number of recording needles or probe electrodes may have obscured one of the maxima. (2) Due to the convoluted shape of the real endocardial surface, the endocardial fibers were not necessarily parallel to the "endocardial" surface defined by the tips of the needles. Also, the direction of the intramural fibers is not perfectly parallel to the endocardium (ie, they form an angle of attack at the endocardial surface). (3) Purkinje involvement during endocardial pacing may indeed have complicated the potential patterns because of additional wave fronts. However, previously recorded epicardial maps32 33 showed that Purkinje involvement did not suppress the rotation-expansion of the epicardial potential pattern. (4) Blood attenuates the potentials, a property that was demonstrated earlier by the authors in an idealized model of the probe-heart-torso geometry.
Results of isochrone maps show a good correspondence between the isochrones determined from the reconstructed endocardial potentials and those determined from the actual measured endocardial potentials. Pattern of activation (propagation of the wave front) and regions of earliest and latest activation can be observed from the isochrones determined from both the measured and the computed potentials. Regions of earliest activation are located in the vicinity of the pacing site. Earliest activation times (time of endocardial breakthrough) for pacing stimuli that originate deep in the myocardium are longer than activation times for subendocardial stimuli. However, pseudoactivation times determined directly from the probe surface, as well as activation times determined from the reconstructed endocardial potentials, appear delayed compared with the activation times determined directly from the measured endocardial potentials. Similar behavior was observed in the potential distributions, where probe potentials and, consequently, the computed endocardial potentials increase sufficiently in magnitude when the cardiac sources become sufficiently developed. The probe-cavity volume conductor geometric and conductive effects result in smoothed-out and low-amplitude probe potentials. As depolarization progresses, the signal-to-noise ratio in the reconstructed endocardial potentials improves, and primary features can then be realized, after an "apparent delay." In this study, the regularization method is formulated under the assumption of time-independent quasistationary conditions, that is, steady-state conditions apply at any instant of time. Taking advantage of the fact that the process of cardiac excitation is continuous in time, Oster and Rudy41 incorporated information from the time progression of excitation in the regularization procedure. Results of using the temporal information in the inverse reconstruction of epicardial potentials from body surface potential data demonstrated a marked improvement in the inverse solution. In particular, physiological events not detected early enough by the quasistatic approach could be detected at the time of their occurrence using the temporal approach. A similar approach that incorporates temporal information in the inverse solution can also be adopted to the probe-endocardium inverse problem discussed here.
Although attempts were made to obtain as accurate a discretization of
the endocardial surface as possible, determining the geometry of the
endocardium and the probe position within the cavity were major sources
of error. Ideally, the exact endocardial geometry and probe position
should be determined at the time of the potential measurements. In our
experiments, the endocardial geometry was obtained at a later time.
After the completion of the experiment, the cavity was filled with
gelatin wax to preserve its geometry. However, the myocardium changed
its properties on termination of perfusion, resulting in alteration of
the position of the intramural needles (distances between neighboring
needle tip electrodes were in the order of 0.5 to 1.0 cm). Furthermore,
all of the computations were made under the assumption of a single
cavity geometry (end-diastolic volume) throughout the
experiment, whereas the amount of blood filling varied from one pacing
protocol to another. The probe-endocardium geometric relation was
determined by finding the probe position that minimized the differences
between the measured probe potentials and those computed from the
measured endocardial potentials (the latter constitutes a solution to
the forward problem). The position so determined approximates the
actual position of the probe during the experiments. However, an error
is introduced since the endocardial geometry used in the computation
differs somewhat from the endocardial geometry at the time of the
potential measurement, as explained above. Despite this error, and
considering the results of simulating additional error in probe
position and orientation (Figs 20
, 21
, and
22
), the
results demonstrate that the inverse solution is robust in the presence
of geometric errors. This property is very encouraging in terms of the
clinical application of the approach. In fact, using existing
noninvasive techniques in the clinical electrophysiology laboratory,
the probe position can be determined at the time of the potential
measurement. Consequently, the geometry determination is likely to be
more accurate in patients than we could achieve in the experiments.
Recent reports by Derfus et al42 43 to assess the
effect
of uncertainty in probe position on estimating endocardial potentials
in an idealized geometry and, subsequently, in an exposed canine heart
have presented concerns in regard to the feasibility of the inverse
reconstruction of endocardial potentials from the measured probe
potentials. Our results, from the present study as well as previous
simulations,17 suggest that actual potential magnitudes
computed by the inverse solution are sensitive to error in geometry and
probe position. Nevertheless, the potential distribution itself and its
spatial characteristics (eg, locations of maxima and minima) are not as
sensitive to error in geometry. Even in the presence of geometric
uncertainties, the reconstructed endocardial potential patterns
determine, with good accuracy, the location of pacing sites at
different positions and intramural depths. The potential distributions
also reflect correctly the fibrous structure of the myocardium
("rotational anisotropy") and its effect on the activation wave
front and the associated myocardial sources. In addition, isochronal
maps constructed from the inversely computed endocardial potentials
provide accurate information on the sequence and pattern of endocardial
activation.
In summary, in the present study we developed methodology, tested the accuracy, and demonstrated the feasibility of reconstructing endocardial potential maps from intracavitary potentials measured by a noncontact multielectrode probe. The reconstruction procedure is the key step in the development of a catheter-based technique that can be applied in the clinical electrophysiology laboratory. It must be emphasized that, in the present study, the probe position/orientation and cavity geometry were determined invasively in a way that cannot be applied clinically. However, the demonstrated feasibility and good performance of the reconstruction procedure and its robustness in the presence of geometric errors indicate that it can be combined with noninvasive imaging methods for determining the geometry. In fact, noninvasive methods used clinically in patients may yield probe-cavity geometry measurements that are more accurate than our experimental method. One such technique that can determine the endocardial geometry and probe position with high measurement accuracy is transesophageal echocardiography.44 45 Other imaging techniques that are available in the clinical electrophysiology laboratory should also be examined. The successful performance of the potential reconstruction procedure developed here suggests that combining it with a clinically applicable method for determining the geometry should be the next step in its development as a clinical tool.
| Acknowledgments |
|---|
| Footnotes |
|---|
Received June 10, 1994; accepted August 19, 1994.
| References |
|---|
|
|
|---|
2.
Spielman SR, Michelson EL, Horowitz LN, Spear JF, Moore EN.
The limitation of epicardial mapping as a guide to the surgical therapy
of ventricular tachycardia. Circulation. 1978;57:666-670.
3. De Bakker JMT, Janse MJ, Van Cappelle FJL, Durrer D. Endocardial mapping by simultaneous recording of endocardial electrograms during cardiac surgery for ventricular aneurysm. J Am Coll Cardiol. 1983;2:947-953. [Abstract]
4. Harada A, D'Agostino HJ Jr, Schuessler RB, Boineau JP, Cox JL. Potential distribution mapping: new method for precise localization of intramural septal origin of ventricular tachycardia. Circulation. 1988;78(suppl III):III-137-III-147.
5. Taccardi B, Arisi G, Macchi E, Baruffi S, Spaggiari S. A new intracavitary probe for detecting the site of origin of ectopic ventricular beats during one cardiac cycle. Circulation. 1987;75: 272-281.
6. Black JJ, Allan A, Williams BT. The use of Foley balloon catheters in cardiac surgery. Eur J Cardiothorac Surg. 1993;7:617-619. [Abstract]
7.
Khoury DS, Rudy Y. A model study of volume conductor effects
on endocardial and intracavitary potentials. Circ
Res. 1992;71:511-525.
8. Rudy Y, Messinger-Rapprt BJ. The inverse problem in electrocardiography: solutions in terms of epicardial potentials. CRC Crit Rev Biomed Eng. 1988;16:215-268. [Medline] [Order article via Infotrieve]
9. Rudy Y, Oster HS. The electrocardiographic inverse problem. CRC Crit Rev Biomed Eng. 1992;20:25-46. [Medline] [Order article via Infotrieve]
10. Rudy Y. The electrocardiogram and its relationship to excitation of the heart. In: Sperelakis N, ed. Physiology and Pathophysiology of the Heart. Kluwer Academic Publishers. 1994:201-239.
11. Brebbia CA, Dominguez J. Boundary Elements: An Introductory Course. Southampton and Boston, Mass: Computational Mechanics Publications; 1989:45-132.
12. Messinger-Rapport BJ, Rudy Y. Regularization of the inverse problem in electrocardiography: a model study. Math Biosci. 1988;89:79-118.
13. Messinger-Rapport BJ, Rudy Y. Computational issues of importance to the inverse recovery of epicardial potentials in a realistic heart-torso geometry. Math Biosci. 1989;97:85-120. [Medline] [Order article via Infotrieve]
14.
Messinger-Rapport BJ, Rudy Y. Noninvasive recovery of
epicardial potentials in a realistic heart-torso geometry: normal sinus
rhythm. Circ Res. 1990;66:1023-1039.
15. Tikhonov AN, Arsenin VY. Solutions of Ill-Posed Problems. Washington, DC: VH Winston & Sons; 1977:27-94.
16. Colli-Franzone P, Guerri L, Taccardi B, Viganotti C. Finite element approximation of regularized solutions of the inverse potential problem of electrocardiography and applications to experimental data. Calcolo. 1985;22:91-186.
17. Khoury DS, Rudy Y. Reconstruction of endocardial potentials from intracavitary probe potentials: a model study. Proc Comput Cardiol. Durham NC, Oct 11-14, 1992;9-12.
18. Macchi E, Arisi G, Taccardi B. Intracavitary mapping: an improved method for locating the site of origin of ectopic ventricular beats by means of a mathematical model. Proc 10th IEEE Engineering in Medicine and Biology Society. New Orleans, Nov 4-7, 1988:187-188.
19. Macchi E, Arisi G, Colli-Franzone P, Guerri L, Olivetti G, Taccardi B. Localization of ventricular ectopic beats from intracavitary potential distributions: an inverse model in terms of sources. Proc 11th IEEE/EMBS. 1989:191-192.
20.
Colli-Franzone P, Guerri L, Viganotti C, Macchi E, Baruffi S,
Spaggiari S, Taccardi B. Potential fields generated by oblique dipole
layers modeling excitation wavefronts in the anisotropic myocardium:
comparison with potential fields elicited by paced dog hearts in a
volume conductor. Circ Res. 1982;51:330-346.
21. Oostendorp TF, Van Oosterom A, Huiskamp G. Interpolation on a triangulated 3D surface. J Comput Phys. 1989;80:331-343.
22. Tweddell JS, Branham BH, Harada A, Stone CM, Rokkas CK, Schuessler RB, Boineau JP, Cox JL. Potential mapping in septal tachycardia: evaluation of a new intraoperative mapping technique. Circulation. 1989;80(suppl I):I-97-I-108.
23.
Corbin LV II, Scher AM. The canine heart as an
electrocardio-graphic generator: dependence on cardiac cell
orientation. Circ Res. 1977;41:58-67.
24. Baruffi S, Spaggiari S, Stilli D, Musso E, Taccardi B. The importance of fiber orientation in determining the features of the cardiac electrical field. In: Antaloczy Z, ed. Modern Electro-cardiology. Amsterdam: Excerpta Medica; 1978:89-92.
25.
Roberts DE, Hersh LT, Scher AM. Influence of cardiac fiber
orientation on wavefront voltage, conduction velocity, and tissue
resistivity in the dog. Circ Res. 1979;44:701-712.
26.
Spach MS, Miller WT III, Miller-Jones E, Warren RB, Barr RC.
Extracellular potentials related to intracellular action potentials
during impulse conduction in anisotropic canine cardiac muscle.
Circ Res. 1979;45:188-204.
27. Roberts DE, Scher AM. Effects of tissue anisotropy on extracellular potential fields in canine myocardium in situ. Circ Res. 1982;50: 342-351.
28. Spaggiari S, Baruffi S, Arisi G, Macchi E, Taccardi B. Effect of intramural fiber direction on epicardial isochrone and potential maps. Circulation. 1987;76(suppl IV):IV-241. Abstract.
29. Baruffi S, Spaggiari S, Arisi G, Malanca A, Macchi E, Taccardi B. The effect of myocardial anisotropy on epicardial potential distributions in the dog heart in situ. In: Yamada K, Harumi K, Musha T, eds. Advances in Body Surface Potential Mapping. Nagoya, Japan: University of Nagoya Press; 1983:151-155.
30.
Frazier DW, Krassowska W, Chen PS, Wolf PD, Danieley ND, Smith
WM, Ideker RE. Transmural activations and stimulus potentials in
three-dimensional anisotropic canine myocardium. Circ
Res. 1988;63:135-146.
31.
Watabe S, Taccardi B, Lux RL, Ershler PR. Effect of
nontransmural necrosis on epicardial potential fields: correlation with
fiber direction. Circulation. 1990;82:2115-2127.
32. Spaggiari S, Baruffi S, Arisi G, Macchi E, Taccardi B. Effect of intramural fiber direction on epicardial isochrone and potential maps. Circulation. 1987;87(suppl IV):IV-241. Abstract.
33.
Taccardi B, Macchi E, Lux RL, Ershler PE, Spaggiari S, Baruffi
S, Vyhmeister Y. Effect of myocardial fiber direction on epicardial
potentials. Circulation. 1994;90:3076-3090.
34.
Streeter DD Jr, Spotnitz HM, Patel DP, Ross J Jr, Sonnenblick
EH. Fiber orientation in the canine LV during diastole and systole.
Circ Res. 1969;24:339-347.
35. Colli Franzone P, Guerri L, Tentoni S. Mathematical modeling of the excitation process in myocardial tissue: influence of fiber rotation on wavefront propagation and potential field. Math Biosci. 1990;101:155-235. [Medline] [Order article via Infotrieve]
36. Colli Franzone P, Guerri L, Taccardi B. Potential distributions generated by point stimulation in a myocardial volume: simulation studies in a model of anisotropic ventricular muscle. J Cardiovasc Electrophysiol. 1993;4:438-458. [Medline] [Order article via Infotrieve]
37. Onufer JR, Cain ME. Impact of mapping and ablation of ventricular tachycardia on management strategies for the 1990s. J Cardiovasc Electrophysiol. 1991;2:77-91.
38. DiMarco JP. Management of sudden cardiac death survivors: role of surgical and catheter ablation. Circulation. 1992;85(suppl I):I-125-I-130.
39. Arisi G, Macchi E, Passolunghi P, Corradi C. Localization of ectopic ventricular focuses by means of a diameter multielectrode catheter. In: Antaloczy Z, ed. Advances in Electrocardiology. Elsevier Science Publishers BV; 1990:67-70.
40.
Pogwizd SM, Hoyt RH, Saffitz JE, Corr PB, Cox JL, Cain ME.
Reentrant and focal mechanisms underlying ventricular tachycardia in
the human heart. Circulation. 1992;86:1872-1887.
41. Oster HS, Rudy Y. The use of temporal information in the regularization of the inverse problem of electrocardiography. IEEE Trans Biomed Eng. 1992;39:65-75. [Medline] [Order article via Infotrieve]
42. Derfus DL, Pilkington TC. Assessing the effect of uncertainty in intracavitary electrode position on endocardial potential estimates. IEEE Trans Biomed Eng. 1992;39:676-681. [Medline] [Order article via Infotrieve]
43. Derfus DL, Pilkington TC, Simpson EW, Ideker RE. A comparison of measured and calculated intracavitary potentials for electrical stimuli in the exposed dog heart. IEEE Trans Biomed Eng. 1992;39:1192-1206. [Medline] [Order article via Infotrieve]
44. de Bruijn NP, Clements FM. Transesophageal Echocardiography. Boston-Dordrecht-Lancaster: Martinus Nijhoff Publishing; 1987:29-54.
45. King DL, Gopal AS, Sapin PM, Schroder KM, Demaria AN. Three-dimensional echocardiography. Am J Card Imaging. 1993;7: 209-220.
This article has been cited by other articles:
![]() |
B. Taccardi, B. B. Punske, E. Macchi, R. S. MacLeod, and P. R. Ershler Epicardial and intramural excitation during ventricular pacing: effect of myocardial structure Am J Physiol Heart Circ Physiol, April 1, 2008; 294(4): H1753 - H1766. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Rao, Y. Ling, R. He, A. L. Gilbert, N. G. Frangogiannis, J. Wang, S. F. Nagueh, and D. S. Khoury Integrated multimodal-catheter imaging unveils principal relationships among ventricular electrical activity, anatomy, and function Am J Physiol Heart Circ Physiol, February 1, 2008; 294(2): H1002 - H1009. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Jeyaraj, L. D. Wilson, J. Zhong, C. Flask, J. E. Saffitz, I. Deschenes, X. Yu, and D. S. Rosenbaum Mechanoelectrical Feedback as Novel Mechanism of Cardiac Electrical Remodeling Circulation, June 26, 2007; 115(25): 3145 - 3155. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Burnes, B. Taccardi, and Y. Rudy A Noninvasive Imaging Modality for Cardiac Arrhythmias Circulation, October 24, 2000; 102(17): 2152 - 2158. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. L. Pierpont, S. S. Chugh, J. A. Hauck, and C. C. Gornick Endocardial activation during ventricular fibrillation in normal and failing canine hearts Am J Physiol Heart Circ Physiol, October 1, 2000; 279(4): H1737 - H1747. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Strickberger, B. P. Knight, G. F. Michaud, F. Pelosi, and F. Morady Mapping and ablation of ventricular tachycardia guided by virtual electrograms using a noncontact, computerized mapping system J. Am. Coll. Cardiol., February 1, 2000; 35(2): 414 - 421. [Abstract] [Full Text] [PDF] |
||||
![]() |
R J Schilling, N S Peters, and D W Davies Mapping and ablation of ventricular tachycardia with the aid of a non-contact mapping system Heart, June 1, 1999; 81(6): 570 - 575. [Abstract] [Full Text] |
||||
![]() |
R. J. Schilling, N. S. Peters, and D. W. Davies Feasibility of a Noncontact Catheter for Endocardial Mapping of Human Ventricular Tachycardia Circulation, May 18, 1999; 99(19): 2543 - 2552. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Schmitt, B. Zrenner, M. Schneider, M. Karch, G. Ndrepepa, I. Deisenhofer, S. Weyerbrock, J. Schreieck, and A. Schomig Clinical Experience With a Novel Multielectrode Basket Catheter in Right Atrial Tachycardias Circulation, May 11, 1999; 99(18): 2414 - 2422. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. C. Gornick, S. W. Adler, B. Pederson, J. Hauck, J. Budd, and J. Schweitzer Validation of a New Noncontact Catheter System for Electroanatomic Mapping of Left Ventricular Endocardium Circulation, February 16, 1999; 99(6): 829 - 835. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Schilling, N. S. Peters, and D. W. Davies Simultaneous Endocardial Mapping in the Human Left Ventricle Using a Noncontact Catheter : Comparison of Contact and Reconstructed Electrograms During Sinus Rhythm Circulation, September 1, 1998; 98(9): 887 - 898. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. S. Oster, B. Taccardi, R. L. Lux, P. R. Ershler, and Y. Rudy Electrocardiographic Imaging : Noninvasive Characterization of Intramural Myocardial Activation From Inverse-Reconstructed Epicardial Potentials and Electrograms Circulation, April 21, 1998; 97(15): 1496 - 1507. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Khoury, K. L. Berrier, S. M. Badruddin, and W. A. Zoghbi Three-Dimensional Electrophysiological Imaging of the Intact Canine Left Ventricle Using a Noncontact Multielectrode Cavitary Probe: Study of Sinus, Paced, and Spontaneous Premature Beats Circulation, February 3, 1998; 97(4): 399 - 409. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. M. Pogwizd, M. K. Chung, and M. E. Cain Termination of Ventricular Tachycardia in the Human Heart: Insights From Three-dimensional Mapping of Nonsustained and Sustained Ventricular Tachycardias Circulation, June 3, 1997; 95(11): 2528 - 2540. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1995 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |