(Circulation. 1995;92:1291-1299.)
© 1995 American Heart Association, Inc.
Articles |
From the Duke-North Carolina NSF/ERC in Emerging Cardiovascular Technologies, Department of Electrical Engineering, Duke University, and Departments of Pathology and Medicine, Duke University Medical Center, Durham, NC.
Correspondence to Robert A. Malkin, The University of Memphis, Herff College of Engineering, Biomedical Engineering Department, Memphis, TN 38152.
| Abstract |
|---|
|
|
|---|
Methods and Results A nonthoracotomy lead system with a biphasic waveform was used throughout. In eight dogs, the dose-response curve widths (a measure of steepness) were compared between DF data and ULV data gathered at the peak (ULVPK), middownslope (ULVDWN), midupslope (ULVUP), and all times (scanning or ULVSCN) in the T wave. In another eight dogs, ULV data (ULVRAP) were gathered by scanning the T wave after 15 rapidly paced beats (166- to 198-ms pacing interval). The rapid pacing interval was chosen to more closely mimic the hemodynamics and activation rate of early VF. ULV data (ULVSTD) at normal heart rates were gathered for all animals. In the first study, scanning significantly reduced the ULV curve width (ULVSCN, 63.5±29.7 V; ULVPK, 81.9±45.2 V; ULVDWN, 116±36.5 V; DF, 105±22.0 V; P<.03) and significantly shifted the ULV curve to the right (ULV80 SCN, 410±62.6 V; ULV80 PK, 266±35.3 V; ULV80 DWN, 355±80.4 V; DF80, 427±60.9 V; P<.001). The subscript 80 signifies that the subject was left in normal sinus rhythm 80% of the time after that stimulus strength was delivered. In the second study, the ULVRAP curve was shifted dramatically to the right, the average ULV50 RAP being greater than the average DF90. Furthermore, 92% of the ULVRAP VF inductions occurred between 10 ms before and 50 ms after the peak of the T wave, suggesting that scanning of the entire T wave may not be necessary.
Conclusions With a single rapidly paced ULV sequence with limited T-wave scanning, it may be possible to estimate highly effective defibrillation doses with few VF episodes and high-voltage stimuli.
Key Words: defibrillation electric stimulation fibrillation death, sudden
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Animal Preparation
The present study followed all applicable
institutional
guidelines in the care and treatment of laboratory animals. Sixteen
dogs (20 to 25 kg) were anesthetized with
intravenous pentobarbital (30 to 35 mg/kg initial dose,
0.05 mg · kg-1 · min-1
maintenance doses, approximately once per hour, or as
indicated). Skeletal muscle paralysis was maintained with
intravenous succinylcholine (1 mg/kg initial dose, 0.25 to
0.50 mg/kg maintenance, approximately once per hour, or as
indicated). Core body temperature was maintained at 37°C with a
hot-water blanket. The dogs were intubated with a cuffed endotracheal
tube and ventilated with room air and oxygen through a respirator
(Harvard Apparatus). A femoral artery line was inserted for
the continuous display of systemic pressure. Arterial blood
samples were drawn every 60 minutes to determine pH, partial pressure
of oxygen, partial pressure of carbon dioxide, base excess, and sodium
bicarbonate, potassium, and calcium contents. Normal saline was
continuously infused through a peripheral
intravenous line. The saline was supplemented with sodium
bicarbonate, potassium chloride, and calcium chloride as indicated to
maintain electrolytes within normal values. ECG electrodes for leads I,
II, and III were applied, and the ECG was continuously monitored.
The dogs were positioned on their backs. Both external jugular veins were exposed, and a transvenous electrode catheter (2.95 cm2 electrode area; CPI) was inserted into each vessel. One electrode was positioned in the superior vena cava (SVC) and one in the distal right ventricle under fluoroscopic guidance. The right ventricular electrode was the cathode for the first phase of a biphasic pulse described below. The anode for ULV and DF shocks was the electrode in the SVC electrically connected to a 113 cm2 cutaneous R2 patch (Darox Corp) placed on the left side of the thorax over the point of maximal impulse.
The diastolic pacing threshold and the intrinsic RR interval were measured at the beginning of the study. The pacing stimulus was a unipolar, 5-ms, monophasic pulse delivered from an electrode at the tip of a catheter in the right ventricle to the R2 patch and SVC catheter. The animal was continuously paced until the end of the study at twice the diastolic pacing current threshold at a pacing interval equal to 80% of the intrinsic RR interval. Equipment failure occasionally required resetting the stimulator. In these cases, at least 15 minutes elapsed before a new pacing current threshold was established, and an intrinsic RR interval measurement was made. At least 30 minutes elapsed after the start of continuous pacing (at the beginning of the study or due to a stimulator reset) before the protocol was continued.
All high-voltage shocks were biphasic (6-ms positive first phase, 6-ms negative second phase) truncated exponential waveforms delivered from a Ventritex HVS-02 stimulator (Ventritex Corp). The two phases were separated by 200 µs. To simulate a single capacitor waveform, the leading-edge voltage of the second phase was set equal to the trailing-edge voltage of the first phase. All voltage settings were rounded to the nearest 10 V.
Definitions
Defibrillation testing (DF) delivers the stimulus
during VF to
determine the DF dose-response curve.
Upper limit of vulnerability testing (ULV) delivers the stimulus during the T wave to determine the ULV dose-response curve. In this article, the ULV is defined only in terms of LP (see below) on the dose-response curve. No single shock strength is defined as "the ULV."
Upper limit of vulnerability testing at the middownslope (ULVDWN) delivers the stimulus during the T wave, past the peak, at a time midway between the peak and end of the T wave.
Upper limit of vulnerability testing at the midupslope (ULVUP) delivers the stimulus during the T wave, before the peak, at a time midway between the peak and beginning of the T wave.
Upper limit of vulnerability testing at the peak (ULVPK) delivers the stimulus during the T wave, at a time corresponding to the peak of the T wave.
Upper limit of vulnerability testing at all times (ULVSCN) delivers the stimulus during the T wave, at all times between the beginning and end of the T wave (see "Scanned ULV Measurements" for scanning details).
Upper limit of vulnerability testing at all times (ULVSTD) is the same as ULVSCN but is applied only to the data for protocol 2.
Upper limit of vulnerability testing at scanned times, rapidly paced (ULVRAP) delivers the stimulus during the T wave, at all times between the beginning and end of the T wave after pacing at an elevated rate.
The p percent effective dose (LP) for DF is the shock strength that defibrillates p% of the time. For ULV, it is the T-wave shock strength that does not fibrillate p% of the time.
These symbols may be used alone or in combination. For example, the symbol ULV50 PK refers to the shock strength that does not fibrillate 50% of the time when delivered at the peak of the T wave.
DF Measurements
It has been demonstrated that DF data
gathered from ULV-induced
VF are not significantly different from other DF data.11
Therefore, DF observations were made 10 seconds after a ULV test shock
initiated VF. The DF test shock strengths were selected by use of an
up-and-down
protocol12 13 14 15 with a
40-V step size and a
starting shock strength of 370 V. At least 4 minutes elapsed after
every DF. More time was permitted, if necessary, to allow the blood
pressure to return to normal. The pacing stimuli were delivered
throughout the protocol.
Scanned ULV Measurements
Both studies required creating
dose-response curves from data
gathered at multiple times throughout the T wave. These data were
called "scanned ULV" data, because the coupling interval between
the pacing stimulus and the ULV stimulus was scanned in 10-ms steps
through all possible values, ie, from the beginning to the end of the T
wave in lead III. However, unlike previous scanning
protocols,16 17 18 which started scanning
the coupling
interval from the end or beginning of the T wave, the coupling
intervals used here were scanned such that the stimuli started at the
peak of the T wave. After the initial stimulus, stimuli were delivered
alternately before and after the peak of the T wave, the coupling
intervals being selected to move the stimuli away from the peak in
10-ms steps. For example, if the initial coupling interval was 300 ms,
then the subsequent coupling intervals would be 310 ms, 290 ms, 320 ms,
280 ms, 330 ms, 270 ms, etc. The same stimulus strength was delivered
at all coupling intervals according to the following two rules.
1. If the first ULV stimulus strength induced VF at any coupling interval, then the shock strength was immediately increased 40 V and the coupling interval reset to the peak of the T wave. Thereafter, the shock strength was increased after each VF episode until a shock strength did not induce VF at any coupling interval. This concluded one scanned ULV measurement. The next scanned ULV measurement was started 40 V lower than the shock strength that did not induce VF at any coupling interval.
2. If the first ULV stimulus strength did not induce VF at any coupling interval, then the shock strength was decreased 40 V and the coupling interval reset to the peak of the T wave. Thereafter, successively weaker shocks were delivered at all coupling intervals until VF was induced. This concluded one scanned ULV measurement. The next scanned ULV measurement was started 40 V higher than the shock strength that induced VF.
ULV Measurements
Protocol 1. In eight dogs, four types of ULV
dose-response curves were measured: (1) scanned (ULVSCN),
(2) middownslope (ULVDWN), (3) midupslope
(ULVUP), and (4) peak (ULVPK) (see
"Definitions"). The standard procedure for measuring a variety of
dose-response curves in a single animal is to interleave the
observations.19 20 However, the accuracy of the
dose-response curve width (L80-L20)
measurements is improved if each dose-response curve is measured
without interruption. Interleaving the observations, on the other hand,
allows drifts in the dose-response curve caused by
physiological changes in the animal to accumulate
throughout the experiment. The accumulated drifts increase the curve
widths, obscuring the distinction between dose-response curves. If each
dose-response curve type is measured without interruption, difference
measurements, such as the curve width
(L80-L20), are affected only by the
small drifts that occur during the relatively short time it takes to
measure one dose-response curve type. The result is a more accurate
estimate of the curve widths. The order of the curve types followed a
randomized Latin square design. This design ensures that the bias in
single-point calculations is random between animals and therefore does
not affect an ANOVA.
At the beginning of the study, the times to the beginning, peak, and end of the T wave were measured in lead III. The beginning of the T wave was defined as the first dramatic change in slope in the lead III electrogram after the ST segment. Others have reported problems identifying a clear ST segment.9 In this study, the ST segment was always easily observed, perhaps because lead III was recorded on a high-resolution digitizing oscilloscope (D6100, Data Precision). The end of the T wave was defined as the dramatic change in slope in the lead III electrogram that preceded the interbeat pause. The peak of the T wave was defined as the maximum observed between the beginning and end of the T wave. Three measurements of the beginning, peak, and end of the T wave were performed and averaged. The time to the middownslope of the T wave was calculated as the sum of the time to the peak and end of the T wave divided by two. The time to the midupslope was calculated as the sum of the time to the peak and the beginning of the T wave divided by two.9 A new measurement of the T wave was taken every five DF episodes or when the stimulator was reset. With each new T-wave measurement, scanning times were recalculated.
At least 15 seconds elapsed between a ULV observation that did not induce VF and the next ULV observation. The nonscanned ULV shock strengths were selected by an up-and-down protocol with 40-V steps, starting at 500 V. A total of 15 ULV measurement sequences were made for each of the four ULV dose-response curve types, each sequence consisting of multiple observations: For the peak, middownslope, and midupslope data, a measurement sequence ended at the first reversal, ie, when (1) a ULV shock induced VF, if the first stimulus in the sequence failed to induce VF, or (2) a ULV shock failed to induce VF, if the first stimulus in the sequence induced VF. For the scanned ULV data, a measurement sequence was defined as in "Scanned ULV Measurements."
Protocol 2. In a second set of eight dogs, two ULV dose-response curve types were measured from (1) standard scanned ULV data (ULVSTD) with pacing at 80% of the intrinsic RR interval and (2) scanned ULV data with rapid pacing (ULVRAP). The standard scanned ULV data were gathered as described above ("Scanned ULV Measurements"). The ULVRAP data were gathered at a pacing interval short enough to mimic more closely the mechanical and electrical state of the heart during the initial seconds of VF without actually inducing VF. The appropriate ULVRAP pacing interval is called the target interval.
At the beginning of each experiment, before any ULV
observations were
made, the appropriate target interval was determined. Starting at 80%
of the intrinsic RR interval, the pacing interval was decreased in
10-ms steps while the surface ECG (lead III) and the
arterial pressure were continuously monitored. The pacing
interval was decreased until the arterial pressure dropped
abruptly and remained low for at least 15 paced, captured beats (2 to 3
seconds). The surface ECG was carefully monitored (Fig 1
) to
ensure that every paced beat captured and to
ensure that VF was not induced by pacing. The longest pacing interval
(slowest pacing rate) that dropped the pressure for 15 beats was taken
as the target interval. Once the target interval was found, it was not
changed throughout the experiment unless the stimulator was reset. If a
reset was required, at least 30 minutes of continuous pacing elapsed
before a new target interval was measured.
|
A preliminary study showed
that capture could not be guaranteed when
the pacing interval was abruptly decreased to the target interval.
Therefore, for the data presented here, the pacing interval was
gradually shortened to the target interval in 10-ms steps (Fig
1
). A
high-voltage ULVRAP stimulus was delivered after the 15th
paced beat at the target interval. Occasionally, a rapid pacing
sequence was unsuccessful, either because capture was lost or because
the arterial pressure did not drop. In either case, any ULV
stimuli delivered after unsuccessful rapid pacing sequences were not
included in the determination of the ULVRAP dose-response
curve.
ULV observations for protocol 2 followed the scanning rules described in "Scanned ULV Measurements" except that separate measurements of the T wave were required for rapid pacing and standard pacing intervals. A total of 15 ULV measurement sequences were made for both ULVRAP and ULVSTD. Because curve width measurements were not the primary focus of protocol 2, the measurement sequences from the two ULV dose-response curve types were randomly interleaved. At least 15 seconds elapsed after each ULV observation that did not induce VF.16
Statistical Analysis
Protocol 1. All observations from all
scan times and
measurement sequences were pooled, and dose-response curves were fitted
by PROBIT analysis.21 The
PROBIT procedure22 was used because it not
only fits the dose-response curve but also estimates shock strengths
for fixed probabilities, such as the L20 and
L80. The width of the dose-response curve was estimated as
the difference between the L80 and the
L20.15 23 After the curve width was
calculated for each dog, an ANOVA was performed to test for any
significant differences in the curve widths between the dose-response
curve types. Multiple contrasts with Bonferroni confidence
intervals24 were used to individually compare
ULVSCN curve widths with each of the other dose-response
curve types. The correlation at both the L80 and
L50 voltages was calculated between the DF and the
ULVDWN, ULVPK, and
ULVSCN, respectively. The significance levels of the
correlations were corrected for multiple simultaneous
inference based on the Bonferroni inequality.24 Unless
otherwise indicated, a lack of statistical significance
(P=NS) was considered to be P>.05. For
completeness, probability values close to .05 are noted in
parentheses.
Protocol 2. For the second study, the PROBIT procedure22 was also used to fit a dose-response curve to each animal for each dose-response curve type. An ANOVA was used to reveal any significant differences between the dose-response curve types at the L50 and the L80. Multiple contrasts with Scheffé confidence intervals25 were used to compare ULVRAP individually with ULVSTD and DF. The correlation was calculated at both the L80 and L50 voltages between the DF and the ULVDWN, ULVPK, and ULVSCN, respectively. Unless otherwise indicated, a lack of statistical significance (P=NS) was considered to be P>.05. For completeness, probability values close to .05 are noted in parentheses.
| Results |
|---|
|
|
|---|
In one dog, VF could
be induced only at the lowest voltage setting (100
V) when the stimulus was delivered at the middownslope
(ULVDWN). For this animal, it is possible to state only
that the ULVDWN 50 and ULVDWN 80 are
100 V.
Therefore, this animal was dropped from the ULVDWN curve
widths and correlation analyses. For the ANOVA tests, the
worst-case value of 100 V was taken. This is worst-case because higher
values contradict the data, lower values artificially increase the test
significance, and dropping the data eliminates known, valid data
points. Binned raw data and the fitted dose-response curves for a
typical animal are shown in Fig 2
. Table 1
summarizes the dose-response curves for all the
animals.
|
|
Scanning shifts the dose-response curve significantly back to the right (P<.001 for L80 and L50). The shift is great enough that there is no longer any significant difference between the ULV80 SCN and DF80 (17-V average difference, P<NS) or between the ULV50 SCN and DF50 (4-V average difference, P<NS). On the other hand, the average ULV80 DWN and ULV80 PK are 72 and 161 V below the average DF80 (P<.06). Not only are the averages shifted, but also, in every dog, the ULV80 DWN and the ULV80 PK are below the corresponding DF80. The ULV80 SCN is a considerably better linear predictor of the DF80 than either ULV80 DWN or ULV80 PK (r=.93, P<.05 versus r=.66, P<NS and r=.38, P<NS).
Scanning significantly reduces the dose-response curve width (P<.03). The average curve width, as estimated by the L80-L20, is 63.5 V for ULVSCN but is significantly higher (P<.03) for ULVPK (81.9 V), ULVDWN (116 V), and DF (105 V). The ULVSCN curve width is also smaller than the DF curve width on an animal-by-animal basis, often by a factor of two or more.
Protocol 2
In all eight dogs combined, the average measured
target interval
is 179±13 ms (Table 2
). Table 3
shows
the L50 and L80 voltages from the dose-response
curves for ULVSTD, ULVRAP, and
DF. To compare the dose-response curves qualitatively, Fig 3
shows composite logistic dose-response curves that
pass directly through the population-averaged L50 and
L80. The ULVRAP dose-response curve is shifted
dramatically to the right of the DF and ULVSTD
dose-response curves (P<.06 for L50). This
shift is of such great magnitude that the average ULV50 RAP
is greater than the average DF90. If we examine the mean
values in pairs (but correct for multiple comparisons), the average
ULV50 RAP is about 100 V above the average
DF50 (P<.06) or ULV50 STD
(P<.06), but there is no statistically significant
difference between the ULV50 STD and the DF50
(P<NS). The average ULV80 RAP is also about
100 V above the ULV80 STD (P<.06) and
DF80 (P<.06), but there is no statistically
significant difference between the ULV80 STD and the
DF80 (P<NS). There is also no statistically
significant difference between the ULVRAP and
ULVSTD curve widths, approximated with
2(L80-L50) (P<NS).
|
|
|
Fig
4
shows the VF induction times for the strongest
shocks that still induced VF (stepping up according to rule 1 under
"Scanned ULV Measurements") and the strongest shock that just
induced VF (stepping down according to rule 2). For
ULVSTD, 87% of these VF inductions occurred between
10 ms before and 20 ms after the peak of the T wave. In this study,
92% of these ULVRAP inductions occurred in a window
between 10 ms before and 50 ms after the peak of the T wave. The
average time at which VF was induced was 16.6 ms after the peak of the
T wave for ULVRAP and 9.20 ms after the peak for
ULVSTD. By
2 test,26
there was a highly significant difference between the distribution of
the coupling intervals for ULVSTD and ULVRAP
(P<.001) due primarily to the last three bins. The times
for ULVRAP appear to show a bimodal tendency that is absent
in the ULVSTD times.
|
| Discussion |
|---|
|
|
|---|
The ULV hypothesis5 6 extends the critical-point hypothesis to state that the same mechanism that causes some T-wave stimuli to induce VF causes weak shocks to fail to defibrillate. Shocks strong enough to raise all of the myocardium above the critical value of electric field strength cannot form a critical point and will leave a subject in normal sinus rhythm whether delivered in the T wave or during VF. In other words, a stimulus strength with a high probability of successfully defibrillating should have a high probability of success (of not inducing VF) when delivered in the T wave. Thus, the ULV and DF dose-response curves should overlap at a stimulus strength with a very high probability of success.
Since the pattern of refractoriness is more repeatable during the T wave than during VF, the critical point and ULV hypotheses also suggest that the ULV dose-response curve should be steeper (less probabilistic) than the DF dose-response curve. T-wave scanning further increases the steepness by systematically probing the repeatable pattern of the refractoriness to find the strongest shock that can still induce VF. If two lines are drawn tangent to the dose-response curves, descending from their point of intersection, the ULV line must intersect any lower probability to the right of the DF line, since the ULV dose-response curve has the greater slope. Thus, the ULV dose-response curve can be said to be to the right of the DF curve.
When the ideas presented in the last two paragraphs are combined, the ULV and critical-point hypotheses predict that the ULV dose-response curve should be steeper and to the right of the DF dose-response curve. However, laboratory data contradict this prediction.4 7 Furthermore, if the ULV and DF dose-response curves intersect at a shock strength with a very high probability of success and the ULV curve is steeper, then the ULV50 should be greater than the DF50. Although a strong correlation is seen between the DF50 and the ULV50 both in laboratory animals7 8 9 30 and in humans,10 17 18 31 32 with the exception of the studies described in References 16, 31, and 32, the data have consistently shown the ULV50 to be significantly below the DF50.
In three reported cases, the population-averaged ULV50 was not significantly below the DF50.16 31 32 However, direct comparison is difficult for any study that defines a single shock strength as "the ULV,"7 8 9 10 16 17 18 30 31 32 because the value of "the ULV" is sensitive to the subtle changes in the definition or testing protocol.32 Our use of a dose-response curve to describe ULV testing, instead of a single shock strength, should reduce the sensitivity of our results to the specific testing protocol. Comparison is also difficult because some of the previous work used monophasic ULV and DF waveforms.8 10 We used biphasic stimulation because there is increasing evidence that it will dominate future clinical practice.33
T-Wave Scanning
The first portion of this study suggests that
some investigators
may have observed the ULV50 significantly below the
DF504 7 9 10
because they did not scan the T
wave (ULVSCN) but rather followed a simplified ULV
measurement sequence, such as ULVUP or ULVPK.
The effect of scanning can be viewed in terms of the critical-point and
ULV hypotheses. If the electric field strength attains the critical
value in some region of the heart, then a point of reentry (a critical
point) can be formed during that portion of the T wave when the region
is in the critical degree of refractoriness. Therefore, the time at
which VF can be induced in the T wave by any shock strength is
influenced by two factors: (1) the strength and geometry of the
electric field, which determines which region is exposed to the
critical electric field, and (2) the myocardial recovery sequence
during the T wave, which determines the state of recovery in the
critical region. Both the geometry and the recovery sequence can change
radically between patients and subtly between beats.34
Thus, the critical-point and ULV hypotheses predict that the most
reliable way to measure the upper end of the ULV dose-response curve is
to scan the vulnerable period, thereby ensuring that shocks are given
when the region in which the electric field strength is weakest is in
the critical stage of refractoriness. Measuring at a single point in
the T wave, on the other hand, may find the critical degree of
refractoriness in a region in which the electric field is not weakest.
This will lower the ULV dose-response curve, because a lower-strength
shock will suffice, at this single point in the T wave, to exceed the
critical value of field strength in the region with the critical degree
of refractoriness. Since a lower-strength shock is needed to prevent
the induction of VF, the ULV dose-response curve is shifted to the left
of the ULV curve that would result from scanning. The data shown in
Table 1
and those given by Hwang et al32 support
this
assertion. Here, the average ULV80 measured at a single
point in the T wave was as much as 144 V to the left (below) the
average ULVSCN. Hwang et al showed that increasing the
number of scanning times in the T wave increased the average ULV by 4
J.
Chen et al9 were not able to induce VF consistently at the middownslope (ULVDWN). In the present study, we were unable to induce VF consistently at the midupslope (ULVUP). Since Chen et al9 used a different electrode configuration and pacing site, this supports the notion, presented above, that the ULV dose-response curve depends on the recovery sequence and electric field geometry and therefore on the pacing and shocking electrode configurations. Fan et al30 found that the ULV50 voltage was not dependent on the pacing site. However, Fan et al measured at only two points in the T wave. From data gathered at only two points in the T wave, it is not possible to know whether the ULVSCN dose-response curve or the ULV80 SCN voltage would have varied with the pacing site, as predicted above.
Not only do the
critical-point and ULV hypotheses predict that T-wave
scanning will shift the ULV dose-response curve to the right, but they
also predict that scanning should reduce the ULV dose-response curve
width. Slight changes in the metabolic or autonomic state
of the animal may alter the degree of refractoriness for a fixed point
in the T wave. These slight changes increase the curve width for any
ULV observations made at a single point in the T wave, eg,
ULVPK and ULVDWN. On the other hand, a
dose-response curve based on T-wave scanning should be less sensitive
to shifts in the degree of refractoriness, since scanning always finds
the time at which the region of the heart exposed to the weakest
electric field strength is in the critical degree of refractoriness.
Therefore, the ULVSCN dose-response curve should have a
smaller curve width than ULVPK or
ULVDWN, according to the critical-point and ULV
hypotheses. This prediction is supported by the data presented
in Table 1
. The average curve width of ULVSCN is 18
V
smaller than ULVPK and 53 V smaller than
ULVDWN.
Although scanning the entire T wave should yield the
smallest curve
width, knowledge of the electric field and pacing site or simple trial
and error may allow the ULVSTD or ULVRAP
dose-response curves to be measured without scanning the entire T wave.
Fig 4
suggests that it might be possible to scan through a
reduced
window around the peak of the T wave. Nearly 90% of the
ULVSTD VF inductions occurred in a window between 10 ms
before and 20 ms after the peak of the T wave. In this study, >90% of
the ULVRAP VF inductions occurred in a window between 10 ms
before and 50 ms after the peak of the T wave. Further investigation
will be required to explain why ULVRAP should require a
wider window than ULVSTD. It is not known whether a limited
scanning window exists for other pacing sites, electrode
configurations, and ULV protocols.
Rapid Pacing
In the first study, the ULVSCN
dose-response
curve is shown to be steeper and to the right of the ULVPK
and the ULVDWN dose-response curves but not enough to make
the ULV50 SCN consistently greater than the
DF50, as predicted by the ULV and critical-point
hypotheses. Rapid pacing, on the other hand, shifts the mean
ULV50 RAP above the DF50. If
differences in the electrical or mechanical state of the heart between
VF and the paced rhythm account for the lowered
ULV50 SCN, then rapid pacing may raise the
ULV50 RAP by more closely mimicking the heart in VF.
Differences have been observed between the mechanical state of the heart during the T wave, when a ULVSCN stimulus is delivered, compared with VF, when a DF stimulus is delivered. The right ventricle has been shown to enlarge early in VF.35 This change may result in a weaker electric field in some region of the heart during VF compared with the same region during pacing at normal heart rates.36 In other words, for the same stimulus, some region of the heart may be exposed to a stronger electric field when the stimulus is delivered during the T wave compared with VF. By this mechanism, standard ULV measurements at normal heart rates may shift the ULV dose-response curve to the left. It is possible that rapid ventricular pacing to the point of hemodynamic compromise (ULVRAP) places the heart in a geometric or volumetric state similar to that which occurs during VF, minimizing any electric field differences and shifting the ULV dose-response curve back to the right.
It is also possible that rapid pacing induces changes in the electrical state of the heart that can account for the increased ULV50 RAP. Pacing at the target interval, compared with 80% of the RR interval, forces the myocardial cells to activate at a rate closer to the VF activation rate. Perhaps the ULV dose-response curve is sensitive to this difference in activation rates.
Any dependence of
ULV data on the pacing rate is in direct
contradiction to the results of Chen et al.9 They reported
that there was no effect on the ULV50 when the pacing
interval was varied from 150 to 500 ms. There may be several reasons
for this contradiction. Only two (of nine) of their dogs could be
consistently paced at 150 ms.9 For the remaining
seven animals, the shortest pacing interval was 200 ms. In our study,
the average target interval was 179±13 ms (Table 2
). It
is possible,
therefore, that Chen et al did not see a significant dependence on
pacing rate because this dependence occurs only for pacing intervals
shorter than 200 ms, for which they had very few data. Another possible
explanation is that Chen et al did not scan the T wave. Perhaps
scanning is required to observe a pacing dependency in the ULV
dose-response curve. Further studies are necessary to determine the
exact cause of these contradictory results.
Clinical Implications
It has been suggested that ULV data
could be used to predict DF
efficacy.4 9 10 16 17 18 31 32
If results in patients are
shown to be similar to our results in dogs, then this study suggests
that ULVRAP may be more reliable or more accurate than
ULVSTD for DF efficacy estimation. Accuracy is defined here
as the average squared difference between the estimate and the true
value. Reliability is defined here as the probability that the estimate
is above the true value. By this definition, typical unbiased, accurate
estimates achieve only 50% reliability. A highly reliable estimate, on
the other hand, may be inaccurate, but not because of
underestimation.
Rapidly paced ULV data can be used as a reliable
estimator of the
DF80. For example, assume that several successively weaker
ULV shocks are delivered (possibly with an abbreviated scanning
sequence) during pacing at the target interval, reducing the shock
strength until VF is induced. We can assume that the lowest delivered
ULV shock strength that did not induce VF is above the
ULV80 RAP. (This assumption can always be satisfied. A
proof of universality can be directly extrapolated from Reference 37 by
setting the starting voltage high and the step size small.) Table
4
shows the differences between the
ULV80 RAP and the DF80. If the variables
in Table 4
are gaussian normal random variables with mean µ
and
standard deviation
, then it is possible to estimate the probability
that the ULV80 RAP is above the DF80:
|
![]() |
where
the right side is of the form Pr[y>(-µ/
)] and y
is
a gaussian normal random variable with mean equal to zero and
variance equal to one. From a standard table for
Pr(y>x),38 the probability that the lowest noninducing
ULVRAP shock strength is greater than the DF80
is .85. By the same logic, the probability that the last noninducing
rapid ULV shock strength is greater than the DF50 is .95.
Thus, stepping down in ULVRAP shock strengths until VF is
induced is a simple and 85% reliable estimator of the DF80
and a 95% reliable estimator of the DF50 in dogs. It is
important to determine whether equally good results can be obtained in
patients.
Even a reliable estimate may be more than is clinically necessary. In some cases, it may be sufficient to simply know that the DF80 is below some shock strength, eg, S. This knowledge could be gained with ULVRAP measurements without inducing VF. The implanting physician would deliver successively lower ULVRAP stimuli until the shock strength was equal to or just below S. Since VF has not been induced, we can again assume that this shock strength is greater than the ULV80 RAP. By the same analysis as in the previous paragraph and given that the shock strength has been stepped down to S without inducing VF, the probability that the DF80 is less than S is >.85.
Study Limitations
Several points may limit the clinical
implications
presented above. Although stepping down to S indicates that the
DF80 is probably below S, the converse does not hold:
Failure to step down to S may not indicate a high DF80.
Furthermore, the probability that the DF80 is below S is
calculated with the assumption that the starting voltage is high and
the step size small. Protocols that deviate widely from this assumption
may not attain the predicted reliability. Thus, it is possible that
clinically viable ULVRAP protocols will not attain
sufficient reliability to justify the increased morbidity associated
with rapid pacing and T-wave scanning. Finally, the dog hearts in this
study were normal. It remains to be determined whether it is possible
to rapidly pace diseased human hearts without inducing VF.
It is assumed throughout the testing for protocol 2 that rapid pacing more closely mimics the electrical or mechanical state of the heart during VF. Although some evidence supports this assertion, it is not directly proved in this study. It is possible that acute rapid pacing increases the ULV80 via a mechanism unrelated to the state of the heart in VF. However, this would not affect the results of this study, since they are based on the shifted dose-response curves, not on the mechanism responsible for the shifts.
Protocol 1 did not follow the common practice of randomly interleaving the observations to reduce bias errors due to physiological drift. Instead, each curve type was measured in sequence, and the order of the curve types followed a randomized Latin square design. This measurement sequence improves the accuracy of the estimated curve widths but reintroduces the susceptibility to physiological drifts for paired analyses. For this reason, paired analysis was not performed in this study, except in the calculation of the correlation coefficients, in which it cannot be avoided. Since any random factor, including physiological drift, will reduce the correlation, the correlation coefficients presented here may underestimate the true correlation coefficients.
Conclusions
The ULV and critical-point hypotheses predict
that the ULV
dose-response curve should be steeper and to the right of the DF
dose-response curve. Yet experimental data contradict this prediction.
The data presented here suggest that this contradiction arises
because (1) the width of the dose-response curve is increased by
experimental protocols that do not scan the T wave and (2) the ULV
measurements are made when the heart is in a different mechanical or
electrical state than the DF measurements. A significant clinical
implication of this work, which must be confirmed in patients, is that
rapidly paced ULV stimuli with limited scanning could be used to
reliably estimate the DF80.
| Acknowledgments |
|---|
Received November 7, 1994; revision received February 15, 1995; accepted February 25, 1995.
| References |
|---|
|
|
|---|
ur. 1967;60:527-544. This article has been cited by other articles:
![]() |
B. Rodriguez, L. Li, J. C. Eason, I. R. Efimov, and N. A. Trayanova Differences Between Left and Right Ventricular Chamber Geometry Affect Cardiac Vulnerability to Electric Shocks Circ. Res., July 22, 2005; 97(2): 168 - 175. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Rodriguez, B. M. Tice, J. C. Eason, F. Aguel, J. M. Ferrero Jr., and N. Trayanova Effect of acute global ischemia on the upper limit of vulnerability: a simulation study Am J Physiol Heart Circ Physiol, June 1, 2004; 286(6): H2078 - H2088. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Link, B. J. Maron, P. J. Wang, B. A. VanderBrink, W. Zhu, and N. A. M. Estes III Upper and lower limits of vulnerability to sudden arrhythmic death with chest-wall impact (commotio cordis) J. Am. Coll. Cardiol., January 1, 2003; 41(1): 99 - 104. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yashima, Y.-H. Kim, S. Armin, T.-J. Wu, Y. Miyauchi, W. J. Mandel, P.-S. Chen, and H. S. Karagueuzian On the mechanism of the probabilistic nature of ventricular defibrillation threshold Am J Physiol Heart Circ Physiol, January 1, 2003; 284(1): H249 - H255. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Chattipakorn, P. C. Fotuhi, C. M. Sreenan, J. B. White, and R. E. Ideker Pacing After Shocks Stronger Than the Upper Limit of Vulnerability : Impact on Fibrillation Induction Circulation, March 21, 2000; 101(11): 1337 - 1343. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. N. Schauerte, K. Ziegert, M. Waldmann, F. A. Schondube, F. Birkenhauer, K. Mischke, M. Grossmann, P. Hanrath, and C. Stellbrink Effect of Biphasic Shock Duration on Defibrillation Threshold With Different Electrode Configurations and Phase 2 Capacitances : Prediction by Upper-Limit-of-Vulnerability Determination Circulation, March 23, 1999; 99(11): 1516 - 1522. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Huang, B. H. KenKnight, G. P. Walcott, D. L. Rollins, W. M. Smith, and R. E. Ideker Effects of Transvenous Electrode Polarity and Waveform Duration on the Relationship Between Defibrillation Threshold and Upper Limit of Vulnerability Circulation, August 19, 1997; 96(4): 1351 - 1359. [Abstract] [Full Text] |
||||
![]() |
C. D. Swerdlow, C. T. Peter, R. M. Kass, E. S. Gang, W. J. Mandel, C. Hwang, D. J. Martin, and P.-S. Chen Programming of Implantable Cardioverter-Defibrillators on the Basis of the Upper Limit of Vulnerability Circulation, March 18, 1997; 95(6): 1497 - 1504. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||