(Circulation. 1996;93:660-666.)
© 1996 American Heart Association, Inc.
Articles |
From the Departments of Cardiology and Radiology (T.D.), Free University Hospital Amsterdam, Institute for Cardiovascular Research, and Interuniversity Cardiovascular Institute, Amsterdam, the Netherlands.
Correspondence to M.A. Galjee, Department of Cardiology, Free University Hospital, De Boelelaan 1117, Amsterdam 1081 HV, Netherlands.
| Abstract |
|---|
|
|
|---|
Methods and Results Forty-seven patients with previous
histories of coronary artery bypass grafting underwent
angiography and MR SE and cine GE phase velocity imaging. SE and GE
images were evaluated by three independent observers blinded to the
angiographic results. The spatial mean velocity and volume flow were
measured and repeated for each image at consecutive 50-millisecond
intervals throughout the cardiac cycle. The 47 patients had 98 proximal
aortotomies, of which 60 were single and 38 sequential grafts.
Seventy-three grafts were patent; 25 were occluded. Eighty-four
grafts (86%) were eligible for comparison of the results of SE and GE
images. Assessment of patency was inconclusive on SE images in 7
grafts (5 occluded by angiography) and on GE images in 7 grafts (2
occluded). A comparison of the results of contrast angiography and SE
and GE MR imaging techniques showed that both techniques had a high
sensitivity (both 98%) and somewhat lower specificity (85% and 88%,
respectively) for graft patency. Combined analysis of the SE
and GE images did not improve the accuracy. The strength of the
interobserver agreement on GE images was good (
=0.66), whereas on
SE
images the agreement was moderate (
=0.51). Adequate MR phase
velocity profiles were obtained in 62 (85%) of the 73 angiographically
patent grafts. Graft flow was characterized by a balanced biphasic
forward flow pattern. The volume flow of sequential grafts to 3 regions
(136±106 mL/min) was significantly higher than in single grafts
(63±41 mL/min, P<.01).
Conclusions Considering the good interobserver agreement and the 85% success rate of quantitative flow measurements, cine GE phase velocity mapping is a promising clinical tool in the noninvasive assessment of graft patency and function.
Key Words: bypass coronary disease grafting magnetic resonance imaging
| Introduction |
|---|
|
|
|---|
Although SE and GE images give important, noninvasive, qualitative information regarding graft flow, quantitative information is not obtained. Therefore, optimal MR imaging of coronary artery bypass grafts may consist of a standard SE examination, followed by cine GE MR imaging. The latter would be performed perpendicular to the graft at selected levels that best display the graft location and allow measurement of flow by means of phase velocity mapping. The value of such a combined approach has not yet been studied. Only one limited report has been published that describes the feasibility of flow measurements in four saphenous vein coronary bypass grafts with the phase velocity mapping technique.8
Hence, the aims of the study were first to investigate whether MR cine GE images, performed in addition to standard SE images, have additional value for the assessment of graft patency and second to assess the graft function by measuring the flow pattern and flow rate with MR phase velocity imaging.
| Methods |
|---|
|
|
|---|
The number of proximal aortotomies and the sites of distal anastomoses were known for all patients from the surgical operative reports at the time of cardiac catheterization and MR imaging. Because of the clip artifact they induce on MR images, internal mammary artery grafts (n=11) were excluded from the study. All patients were in sinus rhythm.
Angiography
Selective angiography was performed by the
Judkins technique
with standard catheters. A nonionic contrast agent (10 mL) was injected
selectively over a period of 3 to 5 seconds. All bypass grafts or
stumps were visualized in at least two projections.
MR imaging
All patients underwent ECG-gated MR SE and cine GE
phase
velocity imaging to assess graft patency. We previously reported on the
application and validation of phase velocity mapping in our 0.6-T
system (Teslacon II, General Electric/CGR).9 10 In
brief,
the technique enables one to obtain simultaneously standard
MR cine GE images for visualization of anatomy and
corresponding velocity maps for calculation of velocity at multiple
frames throughout the cardiac cycle.
The patients were placed in prone
positions on a surface coil. The
grafts were identified on sagittal and transverse multilevel SE scout
views. Fig 1
illustrates the typical course of saphenous
vein coronary bypass grafts with respect to the imaging
planes.
|
Figs 2
and 3
are
representative examples of a sagittal and transverse SE
image perpendicular to grafts anastomosing with the LCx, LAD, and RCA.
SE imaging parameters were as follows: time of repetition
was equal to the length of the cardiac cycle (RR interval), time to
echo was 28 milliseconds, the average of two radiofrequency excitations
was taken, slice thickness was 5 mm, interslice gap was 0.5 mm,
acquisition matrix was 160x256 interpolated to a display matrix of
256x256, the field of view was 30x30 cm2, and
spatial resolution was 1.9x1.2x5 mm3. These images were
used to acquire cine GE images and phase-encoded velocity maps in
an orientation perpendicular to the proximal course of the grafts. Cine
GE imaging parameters were as follows: time of repetition
was 50 milliseconds, time to echo was 24 milliseconds, flip angle was
25°, and the other parameters were as in SE images.
Velocity was phase encoded in the direction of the slice-selective
gradient, thus measuring the component of flow perpendicular to the
imaging plane. Sensitivity to velocity was adjusted to measure
velocities up to 40 cm/s without aliasing.
|
|
Phase velocity imaging of
grafts to the RCA was performed in a
transverse plane at the level of the four-chamber view (Fig 4
).
Velocity imaging of grafts to the left
coronary artery was performed in a sagittal plane at the level
of the pulmonary artery trunk; in some instances, the imaging
plane was slightly rotated on the longitudinal axis (Fig 5
).
|
|
Data Analysis
Angiograms were used as the gold standard for
graft patency. The
patency of the grafts was visually evaluated by an experienced observer
with the cine angiograms. When the proximal segment of the graft, from
the origin of the graft in the ascending aorta up to and including the
first distal anastomosis, was filled by contrast, it was considered
patent.
Graft Patency
MR SE and cine GE images were evaluated from
the hard copies by
three independent observers to test interobserver agreement. They were
blinded to the angiographic results but informed about the number of
proximal anastomoses and the sites of distal anastomoses. Finally,
differences in the interpretation of graft patency between the
observers were solved by consensus. A graft was defined as patent on SE
images if a signal-free lumen was seen on at least two contiguous
images at a position consistent with the expected location for
that graft. If an intermediate signal intensity was seen in the lumen
on two contiguous SE images, which could result from slow flow or
thrombus, this was noted separately and judged as inconclusive. On the
cine GE anatomic images, a graft was defined as patent if it was seen
as a bright blood flow signal on consecutive images throughout the
cardiac cycle at the expected location. A graft was judged inconclusive
if the investigator was not sure about its patency and if no consensus
could be obtained between the three observers.
Graft Identification
The proximal segments of the grafts were
identified from the
typical sites of the aortotomy (Fig 1
). The graft with the most
cephalic origin from the aorta and coursing laterally from the distal
main pulmonary artery or the left pulmonary artery was
considered to anastomose with the LCx or marginal branches (LCx
region). The next lower graft, with a more anterior course immediately
leftward from the main pulmonary artery, was judged to
anastomose with the LAD or diagonal branches (LAD region). Fig
2
is a
typical example of such a situation. The graft with the most caudal
origin from the aorta coursing next to the right atrium or AV groove
was considered to anastomose with the RCA or posterior descending
artery (RCA region), as illustrated in Fig 3
.
Graft Flow
For graft flow analysis, delineation of the graft
had to be obtained on the consecutive cine GE images throughout the
cardiac cycle. The MR data were transferred to a computer workstation.
The anatomic image and the phase velocity image were displayed side by
side. The graft cross section was outlined on a magnified anatomic
image and copied to the corresponding velocity image. The spatial mean
velocity was then measured. These measurements were repeated for each
pair of images at consecutive 50-millisecond intervals throughout the
cardiac cycle. Volume flow was calculated by integration of the average
velocity multiplied by the cross-sectional area over the cardiac
cycle. The data were displayed in curves of flow velocity and volume
flow versus time.
To facilitate comparison of data and to allow construction of the mean saphenous vein graft flow pattern, we normalized each curve to a heart rate of 60 bpm. All systolic acquisition points were normalized to the systolic time interval of a heart rate of 60 bpm; the diastolic acquisition points, to the diastolic time interval.11
Statistical Analysis
Values are expressed as mean±SD.
Student's t test
was used for means of variables of the continuous type. Values of
P<.05 were considered statistically significant. As a
measure of interindividual agreement,
was used. This value
represents the proportion of potential agreement beyond that
expected on the basis of chance.12
| Results |
|---|
|
|
|---|
MR Imaging and Patency
The imaging procedure required 45 to
60 minutes, depending on the
heart rate and number of grafts that had to be visualized. Four
patients (8 grafts) were unable to continue the imaging protocol
because of the discomfort of lying in a prone position and were
excluded from the study. Of the remaining 90 grafts, 6 were imaged on
an incorrect plane by the cine GE technique. Therefore, 84 grafts were
eligible for comparison of the results of SE and cine GE images.
Table
1
compares the results of contrast
angiography and MR imaging. On SE images, the assessment of graft
patency by consensus was obtained in 77 of 84 grafts (92%). The
sensitivity for graft patency on SE images was 98% (63 of 64), with a
specificity of 85% (11 of 13) and a predictive accuracy of 96% (74 of
77). Of the 7 inconclusive grafts, 5 had an intermediate signal
intensity and were shown to be occluded on angiography; 1 of the 2
grafts without intermediate signal intensity was occluded.
|
On GE images, 7 grafts were defined as inconclusive, of which 2 were found to be occluded and 5 to be patent on angiography. For the remaining 77 grafts (92%), the sensitivity for graft patency was 98% (59 of 60) and the specificity was 88% (15 of 17), with a predictive accuracy of 96% (74 of 77).
Table 1
also presents the
results of a combined approach in which
the SE and GE images were evaluated together. Graft patency could be
assessed in 81 of 84 grafts (96%). The sensitivity and specificity for
graft patency of the combined approach were 98% and 76% (63 of 64 and
13 of 17), respectively. The predictive accuracy was 94% (76 of
81).
Interobserver Agreement
The interindividual agreement was
determined according to the MR
technique and imaging planes (Table 2
). The agreement on
cine GE images was good (mean
=0.66), whereas the agreement on SE
images was moderate (mean
=0.51). No major differences in
agreement
were found for the different imaging planes.
|
MR Velocity Imaging
Adequate MR phase velocity profiles were
obtained in 62 of
the 73 angiographically patent grafts (85%). Inadequate velocity
images were obtained from 3 grafts. Of the remaining 8 grafts, 2 were
considered to be occluded, and the imaging plane was not correct in 6.
The success rate of measuring graft flow did not vary between single
and sequential grafts or with the site of the first anastomosis.
Fig
6
shows the flow patterns obtained by MR phase
velocity imaging for single and sequential grafts. Graft flow was
characterized by a balanced biphasic forward flow pattern, with one
peak in early systole and one in early diastole.
|
Table 3
shows the graft flow parameters. The
volume flow of sequential grafts to three regions (136±106 mL/min) was
significantly higher than in single grafts (63±41 mL/min,
P<.01).
|
| Discussion |
|---|
|
|
|---|
Remarkably, the specificity for the combined evaluation was lower than that for the SE and GE images separately. This was due to the fact that 2 grafts classified as inconclusive by each of the techniques separately were falsely considered patent on the combined approach. Moreover, of the 5 other grafts considered inconclusive on SE images, the patency was correctly assessed on the GE images: 1 as patent and 4 as occluded. In the combined analysis, however, only 1 of the 4 occluded grafts was correctly assessed. In the other 3 cases, 1 was assessed as patent and 2 as inconclusive.
From this one can derive that in the case of an inconclusive SE image, assessment of patency should be based on GE images only. The specificity would then have been 94% (16 of 17), with an unchanged sensitivity of 98% (63 of 64).
Furthermore, we can conclude from this study that grafts with intermediate signal intensity on SE images are occluded. If this knowledge were used to predict patency on SE images, the sensitivity would remain unaltered, but the specificity would increase to 89%.
Interobserver Agreement
The interobserver agreement for graft
patency was good to very
good for cine GE images and moderate for SE images. In this respect,
cine GE imaging would be preferable to SE imaging for the assessment of
graft patency because good reproducibility of an observation is
essential for appropriate clinical decision making.
MR Velocity Imaging
In this study, quantitative information
of graft patency was
combined with functional information by MR phase velocity imaging. It
has been demonstrated that phase velocity imaging is successful in
obtaining flow profiles from patent grafts. An important contributory
factor to the successful assessment of graft flow was probably the use
of a surface coil. The spatial resolution of 1.9x1.2x5
mm3 permits accurate cross-sectional imaging of a
bypass graft because the cross-sectional area of the graft is
approximately 13-fold larger than the pixel size. Partial volume
effects are reduced by the slice thickness of 5 mm. Also, the
signal-to-noise ratio is improved for structures that are
localized superficially, as is the case with proximal parts of bypass
grafts. Moreover, the prone position on the surface coil is effective
in reducing respiratory motion artifact.
Flow Pattern
We demonstrated a biphasic forward flow pattern
with one
peak in early systole and one peak in early diastole. No
quantitative differences were found between the flow patterns of single
and sequential grafts. A biphasic velocity pattern in coronary
artery bypass grafts has also been reported by others using Doppler
ultrasound.13 14 In contrast, native coronary flow
occurs predominantly in diastole when aortic pressure
exceeds left ventricular pressure. The systolic
flow in grafts can be explained by passive capacitance of venous grafts
during the cardiac cycle, which is likely to be greater than the
passive capacitance of the native coronary artery bed. It will
influence flow pattern more in the proximal than the distal part of the
graft. During systole, the runoff from the distal part of the graft is
hampered by the high resistance of the native vessel, whereas flow is
allowed to occur in the proximal part because of the capacitance of the
graft. This hypothesis is supported by others who demonstrated a higher
systolic than diastolic velocity in the proximal
part of coronary bypass grafts.14 In accordance
with the hypothesis, the opposite pattern was demonstrated in the
distal part of the graft with a higher diastolic than
systolic velocity.13 Therefore, the biphasic
balanced forward flow pattern in the middle part of the graft,
demonstrated in this study, can be well understood.
Flow Quantification
The results of the present study are in
accordance with the
previously reported average flow rates for the coronary
arteries as measured by a nitrous oxide desaturation technique (65 to
82 mL·min-1·100
g-1) and xenon-133 clearance of 43 to 84
mL·min-1·100
g-1.15 16 17 18
Recently,
measurements by positron emission tomography with oxygen-15labeled
water revealed a basal myocardial blood flow of 1.13±0.26
mL·min-1·g-1
tissue.19 Maximum flow velocities in the present study
are lower than those found by Doppler ultrasound studies (15 to 28
cm/s).12 13 This is due to the fact that in the
present study the mean velocities were obtained as spatially
averaged values over graft cross sections. In contrast, the Doppler
ultrasound technique measures spatial peak velocity within the
graft.13 14 As mentioned, flow velocities in these
studies
were measured at different anatomic levels of the grafts, which makes
comparison even more difficult.
No differences were found in flow velocity parameters and volume flow according to the type of graft, single or sequential, or the number of coronary regions supplied, except for sequential grafts supplying three coronary artery regions. These grafts had a significantly increased volume flow, which can be explained by the fact that they have to supply a larger part of the myocardium.
Limitations and Considerations for Improvement
Internal
mammary grafts cannot be adequately assessed because of
the artifact generated by the metallic clips. A solution to this
problem would be to use nonmetallic clips, which are already
available.
GE images were performed only as a single-level multiphase cine loop, perpendicular to the plane in which the graft was expected on the basis of the SE images. We opted for this approach so that we could quantify the graft flow. This strategy provided more important additional information than graft patency alone. A multilevel GE technique would probably have increased the accuracy by allowing the investigator to assess the graft on contiguous images. Most likely, such a strategy would have reduced the number of grafts imaged at an inadequate plane.
Another important limitation is that given the relatively large voxel sizes, we cannot visualize graft stenoses. Although phase velocity imaging can be a measure of graft function, the site and morphology may also be important in the treatment strategy. Moreover, a graft stenosis or occlusion in a distal part of a sequential graft may not hamper the runoff from the proximal parts and therefore can be missed by functional assessment in the proximal part.
Improvement in imaging of the distal parts of bypass grafts can be expected from the recently introduced MR angiographic techniques developed for visualizing coronary arteries. By use of ultrafast GE techniques, two-dimensional images of coronary arteries can be obtained within a breath-hold, or three dimensional imaging can be obtained when combined with respiratory gating.20 21 22 Similarly, these approaches could be used for imaging of bypass grafts.
From the experience gained from our study, we would propose the following protocol for clinical use: first, a multislice, GE series in a transverse and sagittal plane or sequential two-dimensional breath-hold MR angiography if available, and second, phase velocity mapping at levels selected from the information obtained from the first step.
Conclusions
The present study is another step toward
functional graft
imaging, which has been proposed in recent
years.6 7 8
The good interobserver agreement and accuracy for cine GE images were important findings in the use of MR as a clinical tool in the noninvasive assessment of graft patency. Additional quantitative graft flow imaging was demonstrated to be feasible with a high success rate and revealed a biphasic forward flow pattern in saphenous vein grafts.
Future directions of MR imaging of bypass grafts might include volumetric flow studies, in conjunction with stress agents, to investigate the effects of graft or distal vessel stenosis on the flow reserve of the graft. Also, this technique might be combined with direct imaging of bypass graft stenoses by MR angiography,20 21 22 23 which might ultimately offer a comprehensive, noninvasive approach to patient stratification.
| Selected Abbreviations and Acronyms |
|---|
|
Received July 10, 1995; revision received September 27, 1995; accepted October 4, 1995.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
D. A. Bluemke, S. Achenbach, M. Budoff, T. C. Gerber, B. Gersh, L. D. Hillis, W. G. Hundley, W. J. Manning, B. F. Printz, M. Stuber, et al. Noninvasive Coronary Artery Imaging: Magnetic Resonance Angiography and Multidetector Computed Tomography Angiography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention, and the Councils on Clinical Cardiology and Cardiovascular Disease in the Young Circulation, July 29, 2008; 118(5): 586 - 606. [Full Text] [PDF] |
||||
![]() |
T. Wittlinger, O. Dzemali, I. Martinovic, and A. Moritz Assessment of coronary artery bypass grafts patency with different magnetic resonance technologies. Eur. J. Cardiothorac. Surg., September 1, 2006; 30(3): 436 - 442. [Abstract] [Full Text] [PDF] |
||||
![]() |
N I Stauder, A M Scheule, U Hahn, M Fenchel, F S Eckstein, U Kramer, C D Claussen, and S Miller Perioperative monitoring of flow and patency in native and grafted internal mammary arteries using a combined MR protocol Br. J. Radiol., April 1, 2005; 78(928): 292 - 298. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Pennell, U. P. Sechtem, C. B. Higgins, W. J. Manning, G. M. Pohost, F. E. Rademakers, A. C. van Rossum, L. J. Shaw, and E. K. Yucel Clinical indications for cardiovascular magnetic resonance (CMR): Consensus Panel report Eur. Heart J., November 1, 2004; 25(21): 1940 - 1965. [Full Text] [PDF] |
||||
![]() |
M. J. Budoff, S. Achenbach, and A. Duerinckx Clinical utility of computed tomography and magnetic resonance techniques for noninvasive coronary angiography J. Am. Coll. Cardiol., December 3, 2003; 42(11): 1867 - 1878. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Langerak, H. W. Vliegen, J. W. Jukema, A. H. Zwinderman, H. J. Lamb, A. de Roos, and E. E. van der Wall Vein Graft Function Improvement after Percutaneous Intervention: Evaluation with MR Flow Mapping Radiology, September 1, 2003; 228(3): 834 - 841. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. H. Bunce, C. H. Lorenz, A. S. John, J. R. Lesser, R. H. Mohiaddin, and D. J. Pennell Coronary Artery Bypass Graft Patency: Assessment with True Fast Imaging with Steady-State Precession versus Gadolinium-enhanced MR Angiography Radiology, May 1, 2003; 227(2): 440 - 446. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Langerak, H. W. Vliegen, J. W. Jukema, P. Kunz, A. H. Zwinderman, H. J. Lamb, E. E. van der Wall, and A. de Roos Value of Magnetic Resonance Imaging for the Noninvasive Detection of Stenosis in Coronary Artery Bypass Grafts and Recipient Coronary Arteries Circulation, March 25, 2003; 107(11): 1502 - 1508. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. L. F. Bedaux, M. B. M. Hofman, S. L. A. Vyt, J. G. F. Bronzwaer, C. A. Visser, and A. C. van Rossum Assessment of coronary artery bypass graft disease using cardiovascular magnetic resonance determination of flow reserve J. Am. Coll. Cardiol., November 20, 2002; 40(10): 1848 - 1855. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Langerak, H. W. Vliegen, A. de Roos, A. H. Zwinderman, J. W. Jukema, P. Kunz, H. J. Lamb, and E. E. van der Wall Detection of Vein Graft Disease Using High-Resolution Magnetic Resonance Angiography Circulation, January 22, 2002; 105(3): 328 - 333. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Ishida, H. Sakuma, B. P. Cruz, T. Shimono, T. Tokui, I. Yada, K. Takeda, and C. B. Higgins MR Flow Measurement in the Internal Mammary Artery-to-Coronary Artery Bypass Graft: Comparison with Graft Stenosis at Radiographic Angiography Radiology, August 1, 2001; 220(2): 441 - 447. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. O. Vetter, R. Driever, H. Mertens, U. Kempkes, and B. M. Cramer Contrast-enhanced magnetic resonance angiography of mammary artery grafts after minimally invasive coronary bypass surgery Ann. Thorac. Surg., April 1, 2001; 71(4): 1229 - 1232. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. K. von Schulthess and J. Schwitter Cardiac MR Imaging: Facts and Fiction Radiology, February 1, 2001; 218(2): 326 - 328. [Full Text] |
||||
![]() |
S. E. Langerak, P. Kunz, H. W. Vliegen, H. J. Lamb, J. W. Jukema, E. E. van der Wall, and A. de Roos Improved MR Flow Mapping in Coronary Artery Bypass Grafts during Adenosine-induced Stress Radiology, February 1, 2001; 218(2): 540 - 547. [Abstract] [Full Text] |
||||
![]() |
E. K. Yucel, C. M. Anderson, R. R. Edelman, T. M. Grist, R. A. Baum, W. J. Manning, A. Culebras, and W. Pearce Magnetic Resonance Angiography : Update on Applications for Extracranial Arteries Circulation, November 30, 1999; 100(22): 2284 - 2301. [Full Text] [PDF] |
||||
![]() |
P. Brenner, B. Wintersperger, A. von Smekal, V. Agirov, D. Bohm, E. Kreuzer, M. Reiser, and B. Reichart Detection of coronary artery bypass graft patency by contrast enhanced magnetic resonance angiography Eur. J. Cardiothorac. Surg., April 1, 1999; 15(4): 389 - 393. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Langerak, P. Kunz, H. W. Vliegen, J. W. Jukema, A. H. Zwinderman, P. Steendijk, H. J. Lamb, E. E. van der Wall, and A. de Roos MR Flow Mapping in Coronary Artery Bypass Grafts: A Validation Study with Doppler Flow Measurements Radiology, January 1, 2002; 222(1): 127 - 135. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1996 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |