(Circulation. 1996;94:2497-2506.)
© 1996 American Heart Association, Inc.
Articles |
the Division of Cardiology, Department of Internal Medicine, Johns Hopkins University Medical Institutions, Baltimore, Md.
Correspondence to David A. Kass, MD, Halsted 500, Division of Cardiology, The Johns Hopkins Hospital, 600 N Wolfe St, Baltimore, MD 21287. E-mail dkass@welchlink.welch.jhu.edu.
| Abstract |
|---|
|
|
|---|
Methods and Results Pressure-volume (PV) data were measured by conductance catheter in 87 patients with normal or myopathic hearts. Time-varying elastance curves were generated from 72 PV loops (52 patients) and normalized both by amplitude and time to peak amplitude. The resulting EN(tN) curves were remarkably consistent despite variations in underlying cardiac disease, contractility, loading, and heart rate, with minimal interloop variance during the first 25% to 35% of contraction. On the basis of this finding and assuming ESPVR linearity and constant volume-intercept, ESPVRs were estimated from one beat with the use of PV data measured at normalized time (tN) and end systole (tmax) to predict intercept: Vo(SB)=[EN(tN)xP(tmax)xV(tN)-P(tN)xV(tmax)]/[EN(tN)xP(tmax)-P(tN)] and slope Emax(SB)=Pes/[Ves-Vo(SB)]. Single-beat estimates were highly correlated with measured ESPVR values obtained by standard multiple-beat analysis (including data from 35 additional patients). Emax(SB) accurately reflected acute inotropic change and was influenced little by loading. The new estimation method also predicted measured ESPVRs better than prior techniques and was applicable to noninvasive analysis.
Conclusions ESPVRs can be reliably estimated in humans from single cardiac cycles by a new method that has a potential for noninvasive application.
Key Words: ventricles contractility inotropic agents patients myocardial contraction
| Introduction |
|---|
|
|
|---|
One practical implication of a consistent normalized E(t) curve in humans is that the ESPVR could be estimated from a single beat. This is clearly advantageous over complex traditional methods of generating ESPVRs from multiple variably loaded cardiac cycles.2 3 5 Methods for single-beat ESPVR estimation have been previously proposed,17 18 19 20 largely based on arbitrary mathematical curve fitting of LV pressure data measured during isovolumic contraction and relaxation of an ejecting beat. These methods have not been previously tested in abnormal hearts and require measurement of high-fidelity LV pressure. In contrast, estimates derived from the normalized E(t) curve are physiologically based and have the potential for noninvasive assessment.
Accordingly, the present investigation had four principal aims: (1) to test whether the normalized E(t) curve in humans is consistent among patients despite marked differences in underlying myocardial disease condition, contractile state, heart rate, and vascular loading; (2) to determine if a standard normalized E(t) curve could be applied to yield reliable ESPVR estimates from single-beat data; (3) to compare the new approach to existing single-beat estimation methods; and (4) to test the feasibility of noninvasive application of the new method.
| Methods |
|---|
|
|
|---|
The patient group reflected a broad range of baseline contractility (Ees, 0.5 to 4.9 mm Hg/mL), EF (16% to 87%), chamber size (EDV, 47 to 496 mL), heart rate (47 to 159 min-1), end-systolic pressures (84 to 205 mm Hg), and end-diastolic pressures (1 to 47 mm Hg). Hemodynamic characteristics for each group are provided in Table 1
. Written informed consent was obtained from all patients, and the protocols were approved by The Johns Hopkins Joint Committee on Clinical Investigation.
|
Procedures
Patients were premedicated with benzodiazepam (5 to 10 mg) and diphenhydramine (25 to 50 mg). After routine coronary angiography, left ventriculography, and right heart catheterization, PV relations were determined by conductance catheter technique21 combined with transient inferior vena caval balloon obstruction. Details of this procedure have been previously described.5 22 Briefly, a 7F volume catheter (Webster Labs, Millar, or Sentron) was introduced via a femoral artery, advanced to the LV apex under fluoroscopic guidance, and connected to a stimulator-microprocessor (Sigma-V, Leycom, or VCU, Cardiac Pacemakers Inc) to generate a volume signal. This signal was calibrated so that mean PV loop width matched stroke volume (SV) derived by thermodilution (cardiac output/heart rate) and EF matched that obtained by contrast ventriculography. A 2F micromanometer catheter (SPC-320 Millar) was placed within the distal lumen of the conductance catheter to measure chamber pressure.
Conventional measurement of the ESPVR from multiple PV loops was made with the use of a 7F balloon catheter (SP-9168, Cordis) to transiently obstruct inferior vena cava inflow. This resulted in transient preload reduction with little baroreflex, providing typically 10 or more beats at varying preloads. In several studies, heart rate was altered by atrial pacing, while in others, dobutamine (5 to 10 µg/kg per minute), verapamil (5 to 10 mg), esmolol (1 to 3 mg/kg bolus+0.1 to 0.3 mg/kg per minute continuous IV), or an experimental inotropic agent, OPC-18790 (5 to 10 µg/kg per minute),5 was infused intravenously to alter inotropic state. Data were digitalized at 200 Hz with the use of custom software and analyzed off-line.
ESPVR Determination From Multiple Loops
The reference or "gold standard" ESPVR was derived from multiple variably preloaded cardiac cycles22 shown by the example in Fig 1A
. Points of maximal pressure/(volume-Vo), where Vo was the ESPVR volume-axis intercept, were calculated for each beat in the series by the use of an iterative method to measure Vo.3 23 Data were fit by orthogonal regression24 yielding a multiple-beat ESPVR slope (Ees(MB)) and intercept (Vo(MB)). This regression minimizes the perpendicular distance between the data points and the regression line rather than assuming dependent/independent variable assignments.
|
Normalized Elastance Curve: EN(tN)
To derive the normalized time-varying elastance curve, PV data from 3 to 5 sequential steady-state beats were signal-averaged to yield a PV loop. Time-varying elastance [E(t)] was defined as the instantaneous ratio of P(t)/[V(t)-Vo(MB)]. The maximal value of E(t) [Emax(SB)] and the time to achieve Emax(SB) referenced from the R wave of the ECG (tmax) were both determined. The normalized E(t) function was then defined as
![]() | (E1) |
![]() | (E2) |
Single-Beat Estimation of ESPVR
The new method for single-beat ESPVR estimation is schematically shown in Fig 1B
. For each steady-state cardiac cycle, tmax (the time from R wave to maximal P/V ratio) was determined and time coordinates transformed to normalized time tN (by Equation 2). PV data were then measured at time tN[P(tN) and V(tN)] and tmax[P(tmax), V(tmax)], with linear interpolation as needed. The ESPVR estimate was based on two assumptions: (1) that the E(t) model was linear and (2) that the volume-axis intercept [Vo(SB)] was constant for a given cardiac cycle. Chamber elastance at time tN was
![]() | (E3) |
![]() | (E4) |
![]() | (E5) |
A modification of this method used an iterative procedure to estimate tmax. As noted above, tmax was first derived assuming Vo=0; however, once a Vo(SB) estimate was calculated, tmax could be redetermined at the time of maximal P/(V-Vo(SB)), and then Equations 3 through 5 recalculated, yielding a new Vo(SB). This process could be repeated until successive Vo(SB) estimates were
1 mL from each other. We tested whether this iterative method improved the predictive accuracy of the method.
Validation of Single-Beat ESPVR Estimation
Single-beat estimates for both Emax(SB) and Vo(SB) were compared with corresponding parameters derived from multiple-beat ESPVR analysis [Ees(MB) and Vo(MB)]. This comparison was made between the same loops used to generate the average EN(tN) curve as well as in a second independent set of data from 35 additional patients. This second group also reflected a broad range of cardiac diseases and included data from several patients before and after inotropic intervention. Thus, comparisons were made for a total of 126 conditions measured in 87 patients.
To determine if the new single-beat estimation accurately indexed acute contractility change induced by inotropic agents,
Emax(SB) was compared with
Ees(MB) in 21 patients who received dobutamine (n=5), OPC-18790 (n=5), esmolol (n=9), or verapamil (n=2). To test whether the single-beat Emax(SB) estimate was influenced by loading changes, individual cycles at varying preloads measured during IVC occlusion were subjected to single-beat analysis. Beat-to-beat variability of Emax(SB) and comparisons of these estimates to Ees(MB) were made. This analysis was performed in a random subset of 18 patients (7 normal, 3 HCM, 5 CAD, and 3 DCM).
Last, Emax(SB) estimates were compared with those obtained by the previously reported single-beat method of Sunagawa et al.17 For each steady-state beat, pressure data between EDP and dP/dtmax and between dP/dtmin and the point where pressure declined to EDP were fit by nonlinear regression to the formula
![]() | (E6) |
![]() | (E7) |
Emax(SB) Estimation From Aortic Pressure Data
By selecting tN to occur at the onset of cardiac ejection, then V(tN)=EDV, and P(tN)=AoPDIA, the aortic diastolic pressure. The end-systolic pressure, or P(tmax), was estimated from arterial systolic pressure by
![]() | (E8) |
![]() | (E9) |
![]() | (E10) |
Statistics
Data are presented as mean±SD. Single-beat estimates were compared with multiple-beat measures of resting ESPVR slope and intercept by linear regression. Similar analysis was used to test the accuracy of changes in Emax(SB) induced by inotropic intervention and for the comparison of PV loop and arterial pressure-derived estimates. Load dependence was tested with the use of a repeated-measures ANOVA and coefficient of variation.
| Results |
|---|
|
|
|---|
|
|
Single-Beat ESPVR Estimation
On the basis of the consistency of EN(tN), we then tested the single-beat ESPVR estimation method. While Equations 4 and 5 yielded Vo(SB) and Emax(SB) estimates at any time tN, it was preferable to select tN when the physiological variance of EN(tN) was small, ie, in early contraction (also see Fig 3
). Equation 5 also revealed Vo(SB) to be a hyperbolic function of EN(tN) with a vertical asymptote at EN(tN)=PN(tN). If tN was selected near this time, small errors could lead to instability of the Vo(SB) estimate. An optimal tN could be theoretically deduced to minimize both physiological and mathematical variances. As demonstrated in the "Appendix," this value was between 0.25 and 0.35.
Fig 4A
displays scatterplots of single-beat Emax(SB) and Vo(SB) estimates versus corresponding values from multiple-beat ESPVRs (Ees(MB) and Vo(MB)). The single-beat estimates were the average of results using tN=0.25, 0.30, and 0.35. Both sets of parameters were highly correlated, with regressions given by Ees(MB)=1.01xEmax(SB)+0.03, r=.903, SEE=0.42; and Vo(MB)=0.87xVo(SB)-1.4, r=.91, SEE=27.3 (both P<.001). These slopes and intercepts were not significantly different from unity and zero, respectively. The lower plots display differences between single- and multiple-beat ESPVR slope and intercept estimates. There was a tendency for the Emax(SB) to underestimate Ees(MB) (r=.46, P<.001), whereas this was not observed for Vo(SB) (r=.1, P=NS).
|
Analysis with the iterative method to update tmax and thus Vo(SB) calculations (see "Methods") yielded virtually the same results, with slightly lower slopes and greater scatter: (Ees(MB)=0.97xEmax(SB)+0.12, r=.88, SEE=0.46; and Vo(MB)=0.80xVo(SB)-1.8, r=.90, SEE=28.3 (both P<.001). The Emax(SB)-Ees(MB) difference plot also did not change, revealing a slight underestimation by the single-beat method at high elastances (slope=-0.197, versus -0.20 without iteration). Thus, despite its potential theoretic benefit, the iteration procedure did not improve ESPVR estimation, suggesting that small errors about tmax [and thus P(tmax) and V(tmax)] were not a major source of error in the estimation process.
Table 2
provides data for analyses using tN values other than those between 0.25 and 0.35. As predicted in the "Appendix," these other ranges of tN yielded greater variance, with somewhat lower correlation coefficients, and higher SEE for the single- and multiple-beat ESPVR parameter regressions.
|
Sensitivity of Single-Beat Estimate to Inotropic Change
Fig 5
compares single-beat versus multiple-beat estimates of Emax change due to inotropic interventions. Fig 5A
displays example pressure-volume data with dashed lines corresponding to multiple-beat ESPVRs. The drug intervention primarily induced a change in the relation slope. The solid lines are corresponding single-beat ESPVR estimates based on the resting (high) preload beat and were similar to the multiple-beat ESPVR. Fig 5B
shows group results for
Emax(SB) versus
Ees(MB). There was a strong direct correlation (r=.92, P<.0001), with a slope not statistically different from 1.0 and intercept not different from zero.
|
Sensitivity of Single-Beat Estimate to Preload Change
Fig 6
shows the results of tests for the influence of loading change on Emax(SB) estimation. Emax(SB) was determined from PV loops at initial (high) and reduced (after IVC obstruction) preload volume from a single multiple-beat ESPVR. Fig 6A
displays example PV loop data, with beats at high and low EDV highlighted in bold. The two sets of single-beat ESPVR predictions from these two beats are displayed by a dashed line and were very similar to the multiple-beat ESPVR (solid line). Fig 6B
displays time plots of the same data. The lower tracing shows the single-beat Emax(SB) estimate derived from each beat in the series, revealing high reproducibility despite marked changes in chamber loading. The solid line is the multiple-beat derived Ees(MB).
|
For the group data, the average coefficient of variation (100xSD/mean) of repeated single-beat estimates at varying preloads was 9.4±4%. Comparisons of Ees(MB) to Emax(SB) at high and low EDV in all 18 patients (Fig 6C
) revealed no significant differences between mean values (2.31+1.12, 2.36+0.99, and 2.28+1.09 mm Hg/mL), and values in individual subjects were also generally similar.
Comparison of New Single-Beat Estimation Technique to Prior Methods
Fig 7A
displays a comparison of Emax(SB) determined by the new method to the single-beat Emax,iso estimation based on the method of Sunagawa et al.17 Both estimates are plot versus Ees(MB). In 36 of 126 loops (28%), Equation 6 predicted Piso (isovolumic pressure) that was lower than Pes, resulting in a nonsensical (negative) ESPVR. Interestingly, this occurred more often in failing hearts (HCM, DCM, ANE) than in controls (38% versus 12.8%, P=.002 by Fisher's exact test). Data from the remaining 88 loops are shown in Fig 7A
. Emax,iso generally overestimated Ees(MB) (regression slope of 0.53, intercept of 0.73 mm Hg/mL), in comparison to a slope and offset of 1.01 and 0.008 for the new method (P<.001 by ANCOVA), and displayed a larger variance (SEE=0.57 versus 0.41 for EN(tN) method).
|
Another often-used and simpler Ees estimation is the ratio of Pes/Ves, assuming Vo=0. Fig 7B
displays this ratio versus measured Ees(MB). There were marked discrepancies between actual Ees(MB) and the Pes/Ves ratio, particularly at low values, with overestimation of Ees(MB) by Pes/Ves at higher values.
Single-Beat Estimation of Ees From Aortic Pressure
Prior analysis ("Appendix" and Table 2
) indicated that a tN of 0.25 to 0.35 was optimal for the single-beat ESPVR estimation. This timing was very close to that for aortic valve opening (tN at onset of ejection averaged 0.258±0.035, n=18), supporting the feasibility of using Equations 9 and 10 to estimate ESPVR parameters from arterial pressure and ESV, EDV data. Fig 8
displays the results of this analysis, plotting Emax(SB) calculated from Equation 10 versus the estimate measured from PV loop data (Equation 4, the latter assessed at tN=0.25). These estimates were nearly identical, with the regression given by y=0.993x-0.004, r=.982, P<.0001, SEE=0.19 mm Hg/mL.
|
| Discussion |
|---|
|
|
|---|
Assumptions and Implications of Consistent EN(tN)
The new single-beat ESPVR estimation method was based on the time-varying elastance model of cardiac contraction, assuming that there was a consistent EN(tN), independent of heart rate, loading conditions, or contractility that could be scaled by both amplitude (Emax) and time (tmax) for a given ventricle. This also implied linear elastance relations (including the ESPVR) that intersected at a common volume intercept. These are admittedly oversimplifications, as studies have revealed that ESPVRs can be nonlinear,10 11 12 can be influenced by afterload,13 14 15 16 17 18 26 27 and that Vo may change during contraction.27 Load-dependence of Emax is thought to be due to shortening deactivation,28 29 internal resistance,26 and length-dependent myofilament calcium sensitivity.30 31 Thus, it would seem somewhat surprising that EN(tN) curves from patients with widely varying cardiac diseases and structural adaptations would be so consistent or that the single-beat estimation would yield such reasonable predictions. However, by normalizing the time-varying elastance curve by time as well as amplitude, effects of ejection on tmax, such as recently analyzed by Burkhoff et al,15 were less influential. The resulting EN(tN) curves, particularly in the early phase, probably reflected myocardial properties related to actin-myosin interaction and the kinetics of stiffness generation. Differences in myofibril protein isoforms and protein kinetics among species might alter EN(tN), so one should not assume that the human EN(tN) curve directly applies to other animals.
Another potential explanation for the consistent EN(tN) curves in patients is that human vascular impedance load is more constrained than that typically used in isolated heart and muscle studies. Unlike muscle preparations in which shortening is abruptly released after a brief isometric contraction28 or in isolated hearts in which ejecting and isovolumic contractions are often contrasted,15 real impedance varies much less among humans even with various diseases, thereby limiting effects of ejection history on tmax and Emax. There was more variance in EN(tN) during relaxation, consistent with an enhanced influence of load in this period. Nonetheless, the curves were similar during the initial isovolumic contraction phase, so that during this period, a simple time-varying elastance model became an adequate descriptor of contraction. This finding is supported by prior animal data from Little and Freeman,27 who found that the normalized E(t) curve accurately predicted the early isovolumic pressure rise. Consistency of the isovolumic phase of EN(tN) also formed the basis for relations between the derivative of LV pressure and EDV32 and between chamber contraction and relaxation times.33 One caveat is that EN(tN) in the present study was derived from beats with an isovolumic contraction phase, and this curve should not be extended to hearts with significant mitral regurgitation without additional testing. These tests could not be made with the conductance catheter, since substantial mitral regurgitation can induce large signal artifacts.
Last, it should be reiterated that ESPVR linearity was implicit in the present method, whereas these relations can be curvilinear.10 11 12 Thus, Vo(SB) may have underestimated the true "Vo" (ie, volume intercept at zero pressure) had it been directly measurable, and this probably explains apparent "negative" Vo values that were often derived. Vo(SB) values did correlate well with Vo(MB), which reflected ESPVR behavior in the physiological loading range. This is significant, since Vo(MB) was derived from many beats (average >10) with a net 25 and 50 mm Hg change in Pes. The similarity between Vo(SB) and Vo(MB) is important, since it suggests that the time course of elastance for cycles within the physiological loading range can predict local "linear" ESPVR behavior in that range. This similarly also supports the notion that nonlinear physiology rather than volume-catheter signal artifacts during IVC obstruction most likely explained negative Vo(MB) values, since Vo(SB) would not be influenced by these artifacts. ESPVR nonlinearity is rarely observed clinically, and we could not test whether beats taken from a different, steeper portion of the ESPVR would yield higher Emax(SB) estimates. Nonetheless, we found that the behavior of the ventricle could be predicted over a physiologically relevant loading range from a single beat.
Comparison to Other Single-Beat Estimation Methods
Investigators have been driven by the desire to simplify the loading procedures required to measure the ESPVR and avoid the need for continuous LV volume data through the use of single-beat Emax estimation methods. The simplest, most frequently used method is to ignore Vo and report the ratio of Pes/Ves.32 33 This is inadequate (as shown in Fig 7B
), since there are marked differences in Vo, with often large nonzero intercepts in patients with infarction or dilated cardiomyopathy5 7 8 and apparent negative values in patients with high contractility and presumably nonlinear ESPVRs.
An alternative method first proposed by Igarashi and Suga18 used abrupt occlusion of the aorta during isovolumic contraction to determine peak isovolumetric pressure. The ejecting portion of the PV loop was constructed by integrating aortic flow and synchronizing the result with LV pressure. Emax was the slope of the tangent to this curve anchored at the peak isovolumic PV point. Sunagawa et al17 modified this method by showing that peak isovolumic pressure could be mathematically predicted by curve-fitting isovolumic data from an ejecting beat to a cosine function. Takeuchi et al20 tested the method in 16 patients with normal EF and found reasonable correlations to Ees derived from three variably loaded beats, with an SEE of 1.2 mm Hg/mL/m2. There was no assessment of the accuracy for indexing inotropic change, nor were hearts with significant LV dysfunction evaluated. Some of these limitations were addressed in an animal study,19 but again only in normal hearts.
We found several weaknesses in this method. First, it yielded nonsensical results in 28% of the cases (Pes<Piso, ie, negative Emax), this being three times more common in patients with LV disease. Second, the method generally overestimated directly measured values. Thus, the cosine fit would not appear to apply to all hearts, in particular those with chronic chamber dysfunction. Reliance on data during relaxation (which is more sensitive to load) as well as contraction may contribute to this limitation. The new method yielded more reliable estimates over the full range of heart conditions, used a physiologically based model [EN(tN)] rather than arbitrary curve fits, and could be used without need for left ventricular pressure data.
Noninvasive ESPVR Estimation
Noninvasive applicability of the new single-beat ESPVR estimation method would certainly be an attractive feature. In addition to predicting Vo and Emax, noninvasive assessment of ventricular-arterial interaction could be determined by coupling Emax to estimates of arterial elastance (Ea).7 8 24 34 Chamber volumes for the estimation might be obtained by two-dimensional or three-dimensional echocardiography, magnetic resonance imaging, or nuclear ventriculography, whereas arterial cuff pressures could provide diastolic and end-systolic pressures. Timing of the onset of ejection relative to end systole could be obtained noninvasively by Doppler echocardiography or from arterial pulse tracings generated by tonometry.35 36 The present study showed only that with accurate measurements at the designated time points, noninvasive data could substitute, and thus a noninvasive approach seems feasible. However, there are admittedly considerable errors in noninvasive volume and pressure measurements, and it remains to be determined whether such errors prove to be critical limitations for ESPVR estimation.
| Appendix |
|---|
|
|
|---|
![]() | (EA1) |
|
In addition to this physiological variance, there was mathematical-based variability in the sensitivity of the Vo estimate as a function of the value of EN(tN). This could be shown by differentiating Equation A1 with respect to EN(tN):
![]() | (EA2) |
Multiplying the two curves shown in Fig 9A and 9B![]()
yields a plot of dVO as a function of tN (9C), since dVO=dEN(tN)xdVO/dEN(tN), ie, this indicates the variance about a VO estimate as a function of the time point (tN) chosen from the EN(tN) curve. This plot reveals a minimal variance when tN ranges between 0.25 to 0.35. This result is consistent with the data presented in Table 2
, which shows the best correlation (ie, smallest SEE) between single-beat estimates and multiple-beat estimates for VO and Ees when tN was in this range.
Last, errors in the timing point tN could result in substantial variation in the VO estimate if the EN(tN) curve was itself rapidly changing at that particular time. This could be determined from the derivative of the EN(tN) curve, which is shown in Fig 9D
. This reveals a portion of the curve with a rapidly changing slope followed by a portion where the slope is reasonably constant. It is in this latter range (stippled area) that calculation of VO was less likely to be affected by timing measurement errors. Relevant values of EN(tN) between tN of 0.2 and 0.35 for use with Equations 3 and 4 are provided in Table 3
. In addition, a 12-term Fourier series fit to
![]() | (EA3) |
|
|
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received March 5, 1996; revision received June 11, 1996; accepted June 17, 1996.
| References |
|---|
|
|
|---|
induces a biphasic effect on myocardial contractility in conscious dogs. Circ Res. 1996;78:154-160.This article has been cited by other articles:
![]() |
F. Zanon, S. Aggio, E. Baracca, G. Pastore, G. Corbucci, G. Boaretto, G. Braggion, C. Piergentili, G. Rigatelli, and L. Roncon Ventricular-arterial coupling in patients with heart failure treated with cardiac resynchronization therapy: may we predict the long-term clinical response? Eur J Echocardiogr, June 18, 2008; (2008) jen184v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Shibata, J. L. Hastings, A. Prasad, Q. Fu, K. Okazaki, M. D. Palmer, R. Zhang, and B. D. Levine 'Dynamic' Starling mechanism: effects of ageing and physical fitness on ventricular-arterial coupling J. Physiol., April 1, 2008; 586(7): 1951 - 1962. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Zhong, R.-S. Tan, D. N. Ghista, E. Y.-K. Ng, L.-P. Chua, and G. S. Kassab Validation of a novel noninvasive cardiac index of left ventricular contractility in patients Am J Physiol Heart Circ Physiol, June 1, 2007; 292(6): H2764 - H2772. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. S.P. Lam, V. L. Roger, R. J. Rodeheffer, F. Bursi, B. A. Borlaug, S. R. Ommen, D. A. Kass, and M. M. Redfield Cardiac Structure and Ventricular-Vascular Function in Persons With Heart Failure and Preserved Ejection Fraction From Olmsted County, Minnesota Circulation, April 17, 2007; 115(15): 1982 - 1990. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. van Heusden, J. Gisolf, W. J. Stok, S. Dijkstra, and J. M. Karemaker Mathematical modeling of gravitational effects on the circulation: importance of the time course of venous pooling and blood volume changes in the lungs Am J Physiol Heart Circ Physiol, November 1, 2006; 291(5): H2152 - H2165. [Abstract] [Full Text] [PDF] |
||||
![]() |
|