(Circulation. 1996;94:1036-1040.)
© 1996 American Heart Association, Inc.
Articles |
the Division of Cardiology, Department of Internal Medicine and Department of Electrical Engineering and Computer Science (B.P.), University of Michigan, Ann Arbor.
Correspondence to Fred Morady, MD, University of Michigan Hospital, Division of Cardiology, B1-F245, 1500 E Medical Center Dr, Ann Arbor, MI 48109-0022.
| Abstract |
|---|
|
|
|---|
Methods and Results Pacing was performed at the high right atrium during type I AF in 24 patients in the electrophysiology laboratory. The response to pacing was assessed at cycle lengths 10, 20, 30, 40, and 50 ms less than the mean baseline atrial cycle length. Digitized tracings of the baseline tachycardia and the response to pacing were recorded from the high right atrium and from the distal coronary sinus. Computer analysis of these signals was used to calculate a left atrial electrogram density before, during, and after pacing. Two hundred eighty-eight segments of AF with a duration of 3.9±0.5 seconds (mean±SD) were analyzed. Local capture of the right atrium during AF was demonstrated for at least one pacing cycle length in each patient. The left atrial electrogram density was significantly greater than baseline at each pacing cycle length that resulted in local capture, except when pacing at 50 ms less than the mean AF cycle length. There was no significant change in the baseline left atrial electrogram density compared with baseline when pacing did not result in local capture of AF.
Conclusions Local right atrial capture is often possible by pacing during type I AF and consistently influences the left atrial electrograms recorded in the coronary sinus. These results confirm the presence of excitable tissue in the right and left atria in type I AF.
Key Words: fibrillation pacing excitation
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
Electrophysiological Testing
After informed consent was obtained, three 7F venous sheaths (Daig Corp) were placed in a femoral vein, and three quadripolar electrode catheters (Mansfield EP) with an interelectrode spacing of 2-5-2 mm were positioned as clinically needed. Patients were sedated with intravenous midazolam and fentanyl and received 3000 U heparin IV. Leads V1, I, II, and III and three intracardiac electrograms were recorded (Mingograph 7; Siemens-Elema AB). Pacing was performed with a programmable stimulator (Bloom Associates, Ltd).
Study Protocol
The study protocol was approved by the Human Research Committee and was performed on completion of the clinically indicated portion of the electrophysiology procedure. Quadripolar electrode catheters were positioned in the high right atrium and the distal coronary sinus. So as to have as much overlap as possible between the pacing and recording sites in the right atrium, electrodes 1 and 3 were used for pacing and electrodes 2 and 4 were used to record the local right atrial electrogram. Type I AF was induced by bursts of rapid pacing at the high right atrium. A sustained rhythm was defined as being >30 seconds in duration. Type I AF was defined as atrial electrograms with discrete complexes with varied morphology and cycle length separated by an isoelectric baseline and with a mean atrial cycle length <180 ms (Fig 1
).3
|
Type I AF was induced by pacing in 23 patients and was the baseline rhythm in 1 patient. The mean baseline atrial cycle length and the mean maximum variability in cycle length during AF were 163±22 and 56±23 ms, respectively.
During sustained type I AF, pacing was performed at a current strength of 10 mA and a pulse width of 2 ms. Local capture was assessed by inspection of the local atrial electrogram and by the presence of a postpacing pause longer than the longest atrial cycle length during AF (Fig 2A and 2B![]()
). The initial pacing cycle length was 10 ms less than the mean atrial cycle length. The response to pacing was then assessed at pacing cycle lengths 20, 30, 40, and 50 ms less than the mean atrial cycle length. Bipolar electrograms filtered at 30 to 500 Hz were simultaneously recorded via electrodes 2 and 4 of the right atrial catheter and electrodes 1 and 2 of the coronary sinus catheter at a gain of 20 mV/mm with a TEAC portable data recorder (model R-71, TEAC Corp of America). Recordings were continuous and began with the induction of the sustained arrhythmia.
|
After completion of the study protocol, 18 patients required cardioversion. Three patients converted to sinus rhythm after infusion of procainamide, and 15 patients required electrical cardioversion. There were no complications related to the study protocol or to cardioversion.
Analysis of Data
The analog recordings of AF were digitized at 1000 Hz with commercially available software (CODAS, Dataq Instruments, Inc). For data analysis, digitized recordings from the right atrium were reviewed and divided into four segments: baseline AF, pacing with local capture, pacing without local capture, and segments recorded after pacing with local capture. Segments recorded after pacing with local capture began 2 seconds after cessation of pacing. For each right atrial segment, the simultaneously recorded digitized atrial signal from the distal coronary sinus catheter was analyzed and a left atrial electrogram density was calculated. The electrogram density was calculated by use of a custom-designed computer program that sampled the digitized signal at 1000 Hz and determined the period of time the atrial signal was isoelectric (SPARCstation, Sun Microsystems, Inc). The time period that the signal was not isoelectric was then divided by the total time of analysis to calculate the left atrial electrogram density (Fig 3
). The computer-calculated isoelectric window was defined as any point with a value between zero and one half of the mean of the rectified atrial signal.
|
Two hundred eighty-eight segments were analyzed. One hundred twenty-seven recordings were of pacing with local capture, 127 were of a segment recorded after pacing with local capture, and 34 were of pacing without local capture. The mean duration of the analyzed segments of AF was 3.9±0.5 seconds.
Statistical Analysis
Continuous variables are expressed as mean±1 SD. Differences between continuous variables were tested by Student's t test, and differences between categorical variables were tested by
2 analysis. Comparisons of atrial electrogram density were performed by ANOVA with repeated measures. A value of P<.05 was considered significant.
| Results |
|---|
|
|
|---|
Left Atrial Electrogram Density
The baseline left atrial electrogram density during type I AF was 39±7%. During pacing with local capture of the high right atrium at cycle lengths 10, 20, 30, and 40 ms less than the mean baseline AF cycle length, the left atrial electrogram density increased to 44±9%, 44±7%, 44±8% (all P<.001 versus baseline), and 42±10% (P<.002 versus baseline), respectively (Fig 4
). During pacing at a cycle length 50 ms less than the mean AF cycle length, the left atrial electrogram density was 40±11% (P=NS versus baseline, Fig 4
). An example of the response of the left atrium to pacing with local capture at the high right atrium during type I AF is shown in Fig 5
. With local capture of the right atrium, the left atrial electrogram becomes more disorganized and the electrogram density increases.
|
|
The left atrial electrogram density 2 seconds after cessation of pacing that resulted in local capture during pacing at 10, 20, 30, 40, and 50 ms less than the mean baseline AF cycle length was 41±8% (P=.1 versus baseline), 41±7% (P=.2 versus baseline), 42±6% (P=.4 versus baseline), 42±8% (P=.1 versus baseline), and 40±13% (P=.4 versus baseline), respectively (Fig 6
).
|
During pacing without local capture of the high right atrium at 10 ms less than the mean baseline AF cycle length, the left atrial electrogram density was 40±4% (P=.8 versus baseline).
Postpacing Electrograms
There were 127 recordings of a segment of AF after pacing with local capture of the right atrium. Immediately upon cessation of pacing, there was transient fragmentation and disorganization of the right atrial electrogram in 14 cases (11%) and of the left atrial electrograms in 111 cases (87%). The mean duration of transient disorganization after pacing was 33±85 ms in the right atrium and 497±336 ms in the left atrium (P<.005). In 112 recordings (88%), the left atrium required more time than the right atrium to reorganize to the original electrogram morphology (Fig 7
), and in the remaining 15 recordings (12%), the left and right atria reorganized simultaneously. The left atrial electrograms never became reorganized before the right atrial electrograms.
|
| Discussion |
|---|
|
|
|---|
Transient Disorganization of AF After Pacing
Pacing during AF at times resulted in transient disorganization of the right atrial electrograms immediately after cessation of pacing. This observation is consistent with the results of Kirchhof et al,2 who reported pacing-induced acceleration of AF in dogs attributable to transient, small leading circle reentry circuits induced near the pacing site.
In the present study, pacing during AF often resulted in transient disorganization of the left atrial electrograms without an accompanying disorganization of the right atrial electrograms. Furthermore, when electrograms in both atria became transiently disorganized or more fragmented as a result of pacing, the electrograms recorded in the right atrium often regained their original morphology before the left atrial electrograms. These results demonstrate that the effects of atrial pacing during AF may be more pronounced at sites distant from the pacing site than at the pacing site itself. The explanation for this phenomenon remains unclear, but it is possible that functional or anatomic differences between the two atria result in a greater susceptibility of the left atrium to an acceleration in rate during AF.
Analysis of Pacing Without Capture
When there was not local capture, pacing during AF did not alter left atrial activity. This observation provides validation for the computer-based assessment of left atrial activity used in this study by demonstrating that the computer analysis was not artifactually influenced by the pacing stimuli. To minimize the potential confounding effects of the pacing stimuli, the pacing stimuli were bipolar, the coronary sinus catheter was advanced to the most lateral aspect of the mitral annulus, and coronary sinus recordings were obtained with a bipole distance of 2 mm. The fact that there was no significant increase in the electrogram density when pacing was not accompanied by local capture indicates that the pacing stimuli did not artificially elevate the electrogram density.
Previous Experimental Studies
The presence of an excitable gap during AF has been demonstrated previously in a dog model of AF.1 2 The area of local capture had a mean diameter of
6 cm in one study,1 whereas the other study demonstrated entrainment of
4 cm of the left atrium by local pacing during AF.2 Review of the electrogram tracings included in these experimental studies indicates that the pattern of AF was not consistent with type I AF. The results of the present study suggest that pacing from the high right atrium during type I AF affects a larger portion of the atria and extends to the left atrium. This difference in results between the present study and the previous experimental studies may be attributable to a difference in the type of AF induced in these studies or to other differences between AF induced in dogs and humans.
Previous Clinical Studies
One of the earliest clinical studies to assess pacing during type I AF recorded bipole atrial signals from epicardial electrodes
1 cm apart in humans after open-heart surgery.3 During rapid atrial pacing, there was no evidence of atrial capture or of a change in the AF morphology. In contrast to the present study, atrial pacing was epicardial, which may explain the lack of response of the atria to pacing during type I AF.
Two studies in humans have reported the results of high-density mapping during pacing-induced AF.6 7 These studies described the presence of broad and uniform wave fronts propagating across the right atrium during type I AF but did not assess the response to pacing. Cox et al6 described a right atrial reentrant loop around an area of functional block associated with the sulcus terminalis. That study proposed that one type of AF may be due to a large reentrant circuit in the right atrium and that this circuit may have either a stable or a variable cycle length. Konings et al7 described a similar pattern of a broad uniform wave front activating the right atrium and suggested that the right atrium may be part of a large reentrant circuit. The response to pacing during type I AF in the present study is compatible with a single reentrant circuit arising in the right atrium but does not rule out other possible mechanisms of AF, such as multiple asynchronous sources or multiple wandering reentry wavelets.
In a recent study by Capucci et al,4 atrial overdrive pacing at cycle lengths of 48 to 80 ms during AF resulted in a postpacing mean decrease in atrial cycle length of
9% at the pacing site in 10 patients with idiopathic AF. In contrast to the results of the present study, no changes were observed at other recording sites. This discrepancy between the results of the two studies may be attributable to differences in the techniques used for analysis of electrograms; for example, in the previous study, the only parameter measured during AF was the atrial cycle length, as opposed to electrogram density, which was the parameter measured in the present study. In addition, electrogram analyses in the previous study were performed only after the cessation of overdrive pacing, whereas in the present study, electrograms were analyzed during overdrive pacing as well as after cessation of overdrive pacing.
Pace Termination of AF
Although rapid pacing during AF in the present study often resulted in local capture and distant effects, implying the presence of excitable tissue in both the right and left atria, termination of AF by pacing was never observed. In contrast, in a preliminary study by Haffajee et al,8 right atrial pacing was effective in converting AF to sinus rhythm in three of eight patients with paroxysmal AF. However, a pacing cycle length of 20 ms was used in the previous study, whereas the pacing cycle length in the present study was never <110 ms. It is possible that this difference in pacing cycle lengths explains the different responses to pacing between the two studies.
Study Limitations
The findings of this study are specific to pacing-induced type I AF and may not apply to patients exhibiting spontaneous chronic or paroxysmal AF. Also, only left atrial electrograms recorded from the distal coronary sinus were analyzed, and therefore the response of other areas of the left atrium to right atrial pacing and capture is unknown.
Conclusions
In conclusion, this study demonstrates for the first time that rapid right atrial pacing during AF in humans often results not only in local capture but also in acceleration or increased disorganization of the atrial electrograms at distant left atrial sites recorded from within the coronary sinus. These findings demonstrate that both atria are excitable during type I AF in humans. The ability of right atrial pacing to influence both atria during AF suggests a mechanism by which rapid pacing may terminate AF.
Received October 11, 1995; revision received March 6, 1996; accepted March 13, 1996.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Gulizia, S. Mangiameli, S. Orazi, G. Chiaranda, G. Boriani, G. Piccione, N. DiGiovanni, A. Colletti, C. Puntrello, G. Butera, et al. Randomized comparison between Ramp and Burst+ atrial antitachycardia pacing therapies in patients suffering from sinus node disease and atrial fibrillation and implanted with a DDDRP device Europace, July 1, 2006; 8(7): 465 - 473. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Ashihara, T. Namba, T. Ikeda, M. Ito, K. Nakazawa, and N. Trayanova Mechanisms of Myocardial Capture and Temporal Excitable Gap During Spiral Wave Reentry in a Bidomain Model Circulation, February 24, 2004; 109(7): 920 - 925. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kadish, D. Johnson, W. Choe, J. Goldberger, and G. Horvath Characterization of fibrillatory rhythms by ensemble vector directional analysis Am J Physiol Heart Circ Physiol, October 1, 2003; 285(4): H1705 - H1719. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. H. Everett IV, J. G. Akar, L.-C. Kok, J. R. Moorman, and D. E. Haines Use of global atrial fibrillation organization to optimize the success of burst pace termination J. Am. Coll. Cardiol., November 20, 2002; 40(10): 1831 - 1840. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillis, C. Unterberg-Buchwald, H. Schmidinger, S. Massimo, K. Wolfe, D. J. Kavaney, M. F. Otterness, S. H. Hohnloser, and GEM III AT Worldwide Investigators Safety and efficacy of advanced atrial pacing therapies for atrial tachyarrhythmias in patients with a new implantable dual chamber cardioverter-defibrillator J. Am. Coll. Cardiol., November 6, 2002; 40(9): 1653 - 1659. [Abstract] [Full Text] [PDF] |
||||
![]() |
A R J Mitchell, P A R Spurrell, L Cheatle, and N Sulke Effect of atrial antitachycardia pacing treatments in patients with an atrial defibrillator: randomised study comparing subthreshold and nominal pacing outputs Heart, May 1, 2002; 87(5): 433 - 437. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. W. Adler II, C. Wolpert, E. N. Warman, S. K. Musley, J. L. Koehler, and D. E. Euler Efficacy of Pacing Therapies for Treating Atrial Tachyarrhythmias in Patients With Ventricular Arrhythmias Receiving a Dual-Chamber Implantable Cardioverter Defibrillator Circulation, August 21, 2001; 104(8): 887 - 892. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Duytschaever, F. Mast, M. Killian, Y. Blaauw, M. Wijffels, and M. Allessie Methods for Determining the Refractory Period and Excitable Gap During Persistent Atrial Fibrillation in the Goat Circulation, August 21, 2001; 104(8): 957 - 962. [Abstract] [Full Text] [PDF] |
||||
![]() |
R.J Schilling, A.H Kadish, N.S Peters, J Goldberger, and D.W. Davies Endocardial mapping of atrial fibrillation in the human right atrium using a non-contact catheter Eur. Heart J., April 1, 2000; 21(7): 550 - 564. [Abstract] [PDF] |
||||
![]() |
K. Shinagawa, H. Mitamura, A. Takeshita, T. Sato, H. Kanki, S. Takatsuki, and S. Ogawa Determination of refractory periods and conduction velocity during atrial fibrillation using atrial capture in dogs: Direct assessment of the wavelength and its modulation by a sodium channel blocker, pilsicainide J. Am. Coll. Cardiol., January 1, 2000; 35(1): 246 - 253. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Pandozi, L. Bianconi, M. Villani, A. Castro, G. Altamura, S. Toscano, A. P. Jesi, G. Gentilucci, F. Ammirati, F. L. Bianco, et al. Local Capture by Atrial Pacing in Spontaneous Chronic Atrial Fibrillation Circulation, May 20, 1997; 95(10): 2416 - 2422. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|