(Circulation. 1996;94:1690-1697.)
© 1996 American Heart Association, Inc.
Articles |
the Department of Cardiology, Thoraxcenter (Cardiovascular Research Institute COEUR), Erasmus University Rotterdam (W.J. van der G., H.M.M. van B., P.W.S.), Netherlands; the Cleveland (Ohio) Clinic and Foundation (A.M.L., S.G.E., E.J.T.); and the Mayo Clinic and Foundation (R.S.S., D.R.H.), Rochester, Minn.
Correspondence to Willem J. van der Giessen, MD, PhD, Department of Cardiology, Thoraxcenter, Bd 412, Erasmus University Rotterdam, PO Box 1738, 3000 DR Rotterdam, Netherlands.
| Abstract |
|---|
|
|
|---|
Methods and Results Five different biodegradable polymers (polyglycolic acid/polylactic acid [PGLA], polycaprolactone [PCL], polyhydroxybutyrate valerate [PHBV], polyorthoester [POE], and polyethyleneoxide/polybutylene terephthalate [PEO/PBTP]) and three nonbiodegradable polymers (polyurethane [PUR], silicone [SIL], and polyethylene terephthalate [PETP]) were tested as strips deployed longitudinally across 90° of the circumferential surface of coil wire stents. Appropriately sized polymer-loaded stents were implanted in porcine coronary arteries of 2.5- to 3.0-mm diameter. Four weeks after implantation, stent patency was assessed by angiography followed by microscopic examination of the coronary arteries. The biodegradable PCL, PHBV, and POE and the nonbiodegradable PUR and SIL evoked extensive inflammatory responses and fibrocellular proliferation (thickness of tissue response: 0.79±0.22, 1.12±0.01, 2.36±0.60, 1.24±0.36, and 1.43±0.15 mm, respectively). Less but still severe responses were observed for the biodegradable PGLA and PEO/PBTP (0.46±0.18 and 0.61±0.23 mm, respectively) and for the nonbiodegradable PETP (0.46±0.11 mm).
Conclusions An array of both biodegradable and nonbiodegradable polymers has been demonstrated to induce a marked inflammatory reaction within the coronary artery with subsequent neointimal thickening, which was not expected on the basis of in vitro tests. The observed tissue response may be attributable to a combination of parent polymer compound, biodegradation products, and possibly implant geometry.
Key Words: stents arteries angioplasty coronary disease
| Introduction |
|---|
|
|
|---|
35%; however, bleeding complications have ensued.1 The beneficial effect was shown to be sustained, as a trend toward a reduction in the need for later revascularization was also observed.2 The only approach proven to reduce the incidence of late restenosis (by 30%) is the use of coronary stents.3 4 However, despite recently promoted high-pressure deployment and antiplatelet therapy with aspirin and ticlopidine, the use of stents is not free from complications, with an incidence rate of
20% at 6 months.5 6 7 Therefore, a combination of drugs and stents has been touted as a possibility to overcome both early and late complications of PTCA.8 9 Synthetic polymers have been proposed as a solution to improve the quality of stents, to serve as a vehicle for local (high-dose and site-specific) drug delivery, or both.10 11 Therefore, efforts are under way to develop polymer compounds that can be implanted within the coronary artery.12 13 14 15 16 17 In addition, biodegradable polymers may be formulated with dispersion of drug within the polymeric preparation. Drug release would then occur by diffusion through and/or breakdown of the base polymer. Several biodegradable polymers have been screened for medical-device applications, and a few have been used for local (subcutaneous) drug delivery systems or wound healing. It is unknown, however, whether tissue compatibility data generated from in vitro systems, animal subdermal implant models, or nonvascular human application adequately reflect blood compatibility.18 19 Therefore, we studied the biocompatibility of five biodegradable polymers and three nonbiodegradable polymers after implantation within porcine coronary arteries.
| Methods |
|---|
|
|
|---|
|
The polymer specimens were processed to obtain strips 75 to 125 µm in thickness. The strips were cast longitudinally onto a balloon-expandable stent (Wiktor, Medtronic Inc) that served as the vehicle for polymer deployment. The polymer covered
90° of the stent circumference (Fig 1
). Polymer-loaded stents were mounted on standard angioplasty balloon catheters (manufacturer-specified balloon diameter, 3.0 to 3.5 mm). The implant systems were produced in a clean laboratory environment but not sterilized because of concern about changing the physicochemical properties of the polymers.
|
Animal Preparation
Experiments were performed in farm-bred pigs (weight, 20 to 30 kg) fed a normal, nonatherogenic chow. The investigations were performed according to the Guide for the Care and Use of Laboratory Animals (NIH publication 85-23, 1985), and the protocol was approved by the Experimental Animals Ethics Committee of the three participating centers. The experimental protocol of this first multicenter animal trial of restenosis was uniform for the three study sites, and each site strictly adhered to this protocol. The polymers studied at each center were as follows: polyglycolic acid/polylactic acid (PGLA), polyorthoester (POE), and polyurethane (PUR) (Cleveland Clinic); polyhydroxybutyrate valerate (PHBV) and silicone (SIL) (Mayo Clinic); and polycaprolactone (PCL), polyethyleneoxide/polybutylene terephthalate (PEO/PBTP), and polyethylene terephthalate (PETP) (Thoraxcenter).
After an overnight fast, the animals were sedated. After endotracheal intubation, the pigs were connected to a ventilator, and anesthesia was maintained with gas anesthetics. After administration of antibiotic prophylaxis, arterial access was obtained under sterile conditions by femoral or carotid artery cutdown. Thereafter, angiography was performed to select the part of the coronary tree in which to leave the implant. Heparin (5000 IU) was administered during the procedure only. Aspirin (325 mg) was given before the procedure and continued daily during the 4-week follow-up period.
Polymer-Loaded Stent Implantation
The method of implantation of the polymer-loaded stent in porcine coronary arteries was similar to that described for the conventional stent.36 Briefly, on the basis of the angiograms, at least one segment in one of the three epicardial coronary arteries (LAD, LCx, and RCA) was selected with a diameter of
2.5 to 3.0 mm. Thereafter, an angioplasty catheter with the polymer-loaded stent crimped on its deflated balloon was advanced to that site for implantation over a standard PTCA guidewire. The balloon was inflated to a maximal pressure of 8 atm for 30 seconds, deflated, and slowly withdrawn, leaving the stent in place. This procedure was eventually repeated in a second coronary artery. After repeat angiography of the stented coronary arteries to confirm patency, the arteriotomy was repaired, the skin was closed, and the animals were allowed to recover from anesthesia.
Follow-up Examination
The catheterization procedure for follow-up angiography at 4 weeks was identical to that described above. Coronary angiography was performed in the same projection as used during implantation. Thereafter, the thorax was opened by a midsternal split and a lethal dose of sodium pentobarbital was injected intravenously, immediately followed by in situ fixation of the coronary arteries according to routine procedures in the three study centers, with use of a pressure of
120 mm Hg. Subsequently, the stented vessels were dissected free and placed in 4% formaldehyde in phosphate buffer (pH 7.3) for
48 hours in preparation for microscopy.
Microscopic Examination
Serial sections over the entire length of the polymer-containing coronary segment were embedded in methacrylate or, after removal of the metal stent wires, paraffin. After routine staining (hematoxylin-azaphloxin or hematoxylin-eosin) and application of an elastin stain (resorcin-fuchsin or elasticavan Gieson), at least three representative sections of each artery were examined at each center for fibrocellular tissue response and inflammatory changes on the polymer side, after which the slides were sent to one institute (Erasmus University Rotterdam) for central review.
Morphometry
For the measurement of the constituent layers of the arterial wall, at least three elastin-stained sections from the proximal, central, and distal parts of each stented coronary segment were examined. The extent of the tissue response at the side of the polymer test sample was assessed as shown in Fig 2
. In each section, only the middle area at the polymer side was analyzed so that the potential damaging effect of the polymer edges could be excluded.
|
Gram Staining
To check for bacterial contamination of the implants, alternate histological sections of the stent-containing segments with the polymers PGLA, POE, and PUR were also stained with Gram's stain.
Statistical Analysis
All data are expressed as mean±SEM. Histological measurements were analyzed by unpaired Student's t test. Because of repeated testing, only values of P<.01 were considered statistically significant (Bonferroni correction).
| Results |
|---|
|
|
|---|
|
Follow-up and Restudy at 4 Weeks
Stent occlusion resulting in premature death of the animal occurred in three groups (PHBV, PEO/PBTP, and SIL) in the first 48 hours after implantation (Table 3
). In the groups receiving PGLA, PCL, POE, PUR, and PEO/PBTP, silent occlusion of one of the stents was angiographically demonstrated at 4 weeks. When both early and late stent occlusion were considered together, the arterial patency rate at 4 weeks varied between 70% (PEO/PBTP) and 100% (PGLA, POE, PUR, and PETP), with other groups having one or two arteries occluded (Table 3
). In most other cases, repeat angiography showed an eccentric lumen reduction at the site of the test sample implant at 4 weeks.
|
Histology
Macroscopic examination demonstrated that the eccentric lumen reduction apparent on angiography was due to a localized tissue reaction on the polymer side of the implants (Fig 3
). Light microscopy confirmed that the eccentric tissue response was located mainly on the polymer side. All polymers seemed to evoke a similar reaction; only the extent of the reaction differed (Fig 4
), ranging from a relatively benign response (Fig 4A
) to a malignant or severe inflammatory response (Fig 4D
). Thrombus remnants containing mainly fibrin but also platelet and erythrocyte remnants and hemosiderin deposits were present near the polymer strips. At the interface between polymer and tissue, multinucleated giant cells and macrophages surrounded this proteinaceous debris (Fig 5
). However, signs of acute inflammation were also observed frequently, evidenced by granulocytes (predominantly eosinophils), lymphocytes, and occasional plasma cells. This thrombotic and inflammatory reaction was seen on all sides of the polymer strips, ie, also toward the adventitial side.
|
|
|
A thick layer with a predominantly fibrocellular component was seen around this layer but was most pronounced between the polymer and the lumen. This layer contained smooth muscle cells (confirmed by immunostaining with smooth muscle cellspecific
-actin antibody) in a matrix of collagen and proteoglycans with many neocapillaries and spilled over to the bare wire sites of the specimen. Moderate to severe disruption of the architecture of the arterial wall was present in most specimens. This consisted of rupture or lysis of the elastic membranes and in some cases also of the media and was always accompanied by adventitial inflammatory infiltrates (Fig 5
). This pattern of thrombus remnants, acute and chronic inflammation, and fibrocellular hyperplasia was observed with both biodegradable and nonbiodegradable polymers.
Morphometry
Thickness and area of the tissue response per polymer test sample are summarized in Table 4
. Regardless of the type of polymer, the vessel wall reaction was more pronounced on the polymer than on the metal wire alone. The thickness or area of the tissue reaction to PHBV, POE, PUR, and SIL was significantly larger than the reaction to all other polymers. No significant differences between the other polymer groups were observed.
|
Gram Staining
Signs of bacterial contamination were not observed in any of the samples that underwent Gram staining.
| Discussion |
|---|
|
|
|---|
A unique feature of the present approach is that we chose to perform this study using identical experimental protocols at three centers experienced in the evaluation of new vascular techniques. Therefore, we were able to screen a variety of polymers within a limited period of time while at the same time allowing comparison of results between the polymers and the centers involved.
Main Findings
The main study results of this multilaboratory approach are summarized as follows: (1) after a follow-up period of only 4 weeks, all polymer implants were associated with a significant inflammatory and proliferative response; (2) this response was observed with both biodegradable and nonbiodegradable polymer implants; (3) in some groups, implants were complicated by acute thrombotic vessel occlusion, although with no more frequency than that experienced with stainless steel coronary stents.17 38 Occlusion occurred more frequently, however, than with the tantalum carrier stent alone.36
Polymer Implants in the Coronary Circulation
Implants in the cardiovascular system are more demanding than those in other parts of the human body. The requirement for blood compatibility is added to the requirements for biological performance, absence of toxic reactions (toxic, immunologic, carcinogenic), and long-term mechanical properties (fatigue life, wear resistance, kink resistance) in this dynamic environment.39 This means favorable behavior is required in an environment in which complex and integrated cellular and humoral systems (coagulation, complement, and immune systems) unite to isolate and exclude the foreign body from incorporation into the vascular wall.35 Therefore, in retrospect, it is not surprising that polymer implants in the coronary circulation elicit a more severe reaction than that predicted from subcutaneous implants. In three groups (PHBV, PEO/PBTP, and SIL), early thrombotic occlusion was observed (5 [23%] of 22 stents). This is not an exceptionally high number, because earlier studies in the same model reported even higher rates of thrombosis with stainless steel stents.17 38 Moreover, during the initial clinical experience with the Palmaz-Schatz stent, an 18% incidence of subacute closure was observed when anticoagulation treatment was withheld.40 Furthermore, it has been reported recently41 that noncoated, slotted-tube stents show a 42% thrombotic occlusion rate in the rabbit iliac model.
However, local mechanical and hemodynamic factors may influence the success or failure of a specific material.42 43 For instance, vascular grafts of PETP (Dacron) seem to perform best in larger vessels, whereas in smaller vessels, expanded polytetrafluoroethylene (Gore-Tex) yields good results. However, in vessels <4 mm in diameter, all synthetic materials fail, and the use of vein grafts offers the best solution. Our results in the coronary circulation extend this effect of recipient vessel diameter to the coronary circulation.
The results of the present study may only be applicable to the specific polymers investigated. Differences in molecular weight, polymerization catalysts, plasticizers, and fillers may all change the physicochemical behavior of the implants. Studies by others who used a PGLA stent yielded superior results,12 but the toxicity of PGLA microspheres in smooth muscle cell culture has also been reported recently.44 Furthermore, the use of PETP stents of different sources resulted in equivocal vascular responses.15 16
In the present study, a significant inflammatory response was observed with all implanted polymers. In all cases, this consisted of a chronic inflammatory reaction with an acute component and a persistent foreign-body response. In most cases, a substantial part of the overall inflammatory and proliferative response may have been aggravated by damage due to the asymmetric geometry of the implant. It is very likely that the aggressive inflammatory response may have increased the injury to the arterial architecture by the action of released proteases and elastases. This may have been influenced by by-products of the polymer. A role for greater stretch injury of the polymer side of the stent cannot be excluded, but it seems more likely that the presence of the hard polymer structure merely prevented overstretch on that side. Indeed, after in vitro expansion of some stent specimens by inflation, followed by removal of the balloon, it was evident by subsequent high-power microscopy that the polymer strip covered a smaller part of the circumference than in the unexpanded condition. In addition, the uncovered part expanded more than the part covered by the polymer. The possibility that this acute damage adds to the final outcome should be substantiated by acute experiments in future studies testing the intracoronary biocompatibility of other synthetic polymers.
In addition to the general reaction to the bulk material and the physicochemical properties of the implant surface, the surface texture could be an important determinant of the early reaction.37 A limitation of the present study is that we cannot retrospectively correlate the overall response with its several components.
It should also be emphasized that the implants in the present study were not sterilized but were manufactured in a clean laboratory environment. This may have influenced the response. Gram staining in those polymer samples that demonstrated the most vigorous responses, however, did not show bacterial contamination. Furthermore, it has been shown that the addition of steroids to one of the polymers ameliorated the inflammatory response.45 However, this does not exclude completely the possibility of bacterial or nonbacterial contamination.
Conclusions
The present study demonstrates the marked inflammatory and neointimal response to an array of biodegradable as well as nonbiodegradable polymers after implantation in the porcine coronary artery. This reaction must be fully understood biologically before we can make use of these or other polymers as implant materials in stents or drug delivery devices.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received November 13, 1995; revision received April 11, 1996; accepted April 16, 1996.
| References |
|---|
|
|
|---|
-caprolactone and its copolymers. In: Chasin M, Langer R, eds. Biodegradable Polymers as Drug Delivery Systems. New York, NY: Marcel Dekker Inc; 1990.
-caprolacton). J Biomed Mater Res. 1985;19:437-444.[Medline]
[Order article via Infotrieve]This article has been cited by other articles:
![]() |
J. Mehilli, R. A. Byrne, A. Wieczorek, R. Iijima, S. Schulz, O. Bruskina, J. Pache, R. Wessely, A. Schomig, A. Kastrati, et al. Randomized trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis Eur. Heart J., June 11, 2008; (2008) ehn253v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Fishbein, I. Alferiev, M. Bakay, S. J. Stachelek, P. Sobolewski, M. Lai, H. Choi, I. -W. Chen, and R. J. Levy Local Delivery of Gene Vectors From Bare-Metal Stents by Use of a Biodegradable Synthetic Complex Inhibits In-Stent Restenosis in Rat Carotid Arteries Circulation, April 22, 2008; 117(16): 2096 - 2103. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Aoki, A. Kirtane, M. B. Leon, and G. Dangas Coronary Artery Aneurysms After Drug-Eluting Stent Implantation J. Am. Coll. Cardiol. Intv., February 1, 2008; 1(1): 14 - 21. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. N. Helmus, D. F. Gibbons, and D. Cebon Biocompatibility: Meeting a Key Functional Requirement of Next-Generation Medical Devices Toxicol Pathol, January 1, 2008; 36(1): 70 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Egashira, K. Nakano, K. Ohtani, K. Funakoshi, G. Zhao, Y. Ihara, J.-i. Koga, S. Kimura, R. Tominaga, and K. Sunagawa Local Delivery of Anti-Monocyte Chemoattractant Protein-1 by Gene-Eluting Stents Attenuates In-Stent Stenosis in Rabbits and Monkeys Arterioscler. Thromb. Vasc. Biol., December 1, 2007; 27(12): 2563 - 2568. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Daemen and P. W. Serruys Drug-Eluting Stent Update 2007: Part I: A Survey of Current and Future Generation Drug-Eluting Stents: Meaningful Advances or More of the Same? Circulation, July 17, 2007; 116(3): 316 - 328. [Full Text] [PDF] |
||||
![]() |
B. Scheller, U. Speck, and M. Bohm Prevention of restenosis: is angioplasty the answer? Heart, May 1, 2007; 93(5): 539 - 541. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Khan, A. Agrotis, and A. Bobik Understanding the role of transforming growth factor-{beta}1 in intimal thickening after vascular injury Cardiovasc Res, May 1, 2007; 74(2): 223 - 234. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Camenzind, P. G. Steg, and W. Wijns A Cause for Concern Circulation, March 20, 2007; 115(11): 1440 - 1455. [Full Text] [PDF] |
||||
![]() |
K. Ohtani, K. Egashira, K. Nakano, G. Zhao, K. Funakoshi, Y. Ihara, S. Kimura, R. Tominaga, R. Morishita, and K. Sunagawa Stent-Based Local Delivery of Nuclear Factor-{kappa}B Decoy Attenuates In-Stent Restenosis in Hypercholesterolemic Rabbits Circulation, December 19, 2006; 114(25): 2773 - 2779. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Scheller, C. Hehrlein, W. Bocksch, W. Rutsch, D. Haghi, U. Dietz, M. Bohm, and U. Speck Treatment of Coronary In-Stent Restenosis with a Paclitaxel-Coated Balloon Catheter N. Engl. J. Med., November 16, 2006; 355(20): 2113 - 2124. [Abstract] [Full Text] [PDF] |
||||
![]() |
R R Anis and K R Karsch The future of drug eluting stents Heart, May 1, 2006; 92(5): 585 - 588. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Wessely, A. Schomig, and A. Kastrati Sirolimus and Paclitaxel on Polymer-Based Drug-Eluting Stents: Similar But Different J. Am. Coll. Cardiol., February 21, 2006; 47(4): 708 - 714. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-H. Lee, H.-C. Tan, and Y.-T. Lim Update on Drug-Eluting Stents for Prevention of Restenosis Asian Cardiovasc Thorac Ann, February 1, 2006; 14(1): 75 - 82. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. J. Salu, J. M. Bosmans, Y. Huang, M. Hendriks, M. Verhoeven, A. Levels, S. Cooper, I. K. De Scheerder, C. J. Vrints, and H. Bult Effects of cytochalasin D-eluting stents on intimal hyperplasia in a porcine coronary artery model Cardiovasc Res, February 1, 2006; 69(2): 536 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Mehilli, A. Kastrati, R. Wessely, A. Dibra, J. Hausleiter, B. Jaschke, J. Dirschinger, A. Schomig, and for the Intracoronary Stenting and Angiographic Re Randomized Trial of a Nonpolymer-Based Rapamycin-Eluting Stent Versus a Polymer-Based Paclitaxel-Eluting Stent for the Reduction of Late Lumen Loss Circulation, January 17, 2006; 113(2): 273 - 279. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Azarbal and J. W. Currier Allergic Reactions After the Implantation of Drug-Eluting Stents: Is it the Pill or the Polymer? J. Am. Coll. Cardiol., January 3, 2006; 47(1): 182 - 183. [Full Text] [PDF] |
||||
![]() |
J. R. Nebeker, R. Virmani, C. L. Bennett, J. M. Hoffman, M. H. Samore, J. Alvarez, C. J. Davidson, J. M. McKoy, D. W. Raisch, B. K. Whisenant, et al. Hypersensitivity Cases Associated With Drug-Eluting Coronary Stents: A Review of Available Cases From the Research on Adverse Drug Events and Reports (RADAR) Project J. Am. Coll. Cardiol., January 3, 2006; 47(1): 175 - 181. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Fishbein, I. S. Alferiev, O. Nyanguile, R. Gaster, J. M. Vohs, G. S. Wong, H. Felderman, I-W. Chen, H. Choi, R. L. Wilensky, et al. Bisphosphonate-mediated gene vector delivery from the metal surfaces of stents PNAS, January 3, 2006; 103(1): 159 - 164. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Q. Liu, P. K. Alkema, C. Tieche, B. J. Tefft, D. Z. Liu, Y. C. Li, B. E. Sumpio, J. A. Caprini, and M. Paniagua Negative Regulation of Monocyte Adhesion to Arterial Elastic Laminae by Signal Regulatory Protein {alpha} and Src Homology 2 Domain-containing Protein-Tyrosine Phosphatase-1 J. Biol. Chem., November 25, 2005; 280(47): 39294 - 39301. [Abstract] [Full Text] [PDF] |
||||
![]() |