(Circulation. 1997;95:1497-1504.)
© 1997 American Heart Association, Inc.
Articles |
From the Division of Cardiology (C.D.S., C.T.P., E.S.G., W.J.M., C.H., D.J.M., P.-S.C.) and the Department of Cardiovascular Surgery (R.M.K.), Cedars-Sinai Medical Center, Los Angeles, Calif.
Correspondence to Charles D. Swerdlow, MD, 8635 W Third St, Suite 975 W, Los Angeles, CA 90048. E-mail swerdlow{at}ucla.edu.
| Abstract |
|---|
|
|
|---|
Methods and Results We studied 100 consecutive patients at ICD implantation and during follow-up of 20±7 months. At implantation, the ULV and DFT were determined, and the ICD system was tested at a shock strength equal to the ULV+3 J. During follow-up, the strength of the first shock was programmed to the ULV+5 J for arrhythmias detected in the VF zone (cycle length <292±17 ms). We reviewed stored detection intervals and electrograms from spontaneous episodes of ICD-detected VF to determine the success rate for appropriate first shocks. The programmed first-shock strength was 17.5±5.2 J. During follow-up, there were 120 appropriate first shocks in 37 patients. The arrhythmia was rapid monomorphic ventricular tachycardia (VT) in 70% of episodes (31 patients), VF in 11% (13 patients), polymorphic VT in 1%, and unclassified in 17% (15 patients). The first shock was successful in 119 of 120 episodes (99%; 95% CI, 93% to 100%). One unclassified episode required two shocks. No patient had syncope associated with an ICD shock or arrhythmic death.
Conclusions ICD shocks can be programmed on the basis of the ULV, a measurement made in regular rhythm, without a direct measure of defibrillation efficacy.
Key Words: defibrillation heart-assist device upper limit of vulnerability
| Introduction |
|---|
|
|
|---|
The ULV is the weakest shock strength at or above which VF is not induced when a stimulus is delivered during the vulnerable phase of the cardiac cycle. The ULV hypothesis of defibrillation links the ULV, a measurement made in regular rhythm, to the shock strength that defibrillates reliably.19 20 21 22 23 The ULV has been proposed as a patient-specific measure of defibrillation efficacy because it can be determined with a single episode of VF. Recently, we described a clinical method for determining the ULV24 that provides a good estimate of the shock strength that defibrillates with a 90% probability of success.25 A shock strength equal to this ULV+3 J provides an adequate defibrillation safety margin for VF induced at ICD implantation25 and 3 months later.26 The purpose of this prospective study was to test the hypothesis that patient-specific shock strengths programmed on the basis of the ULV, without regard to the DFT, convert spontaneous ICD-detected VF reliably.
| Methods |
|---|
|
|
|---|
Implantation Procedure
Surgical technique. Implantations were performed
under general anesthesia and after endotracheal intubation as described
previously.25 27 A transvenous ICD (Medtronic Jewel model
7202D, 7218C, 7219C, or 7219D) and electrodes were inserted through a
single left anterior axillary incision28 in all 100
patients.
Electrode configuration. A tripolar lead with a 5-cm defibrillation coil and bipolar pace-sense electrodes was positioned in the right ventricular apex via the cephalic or subclavian vein. It served as the cathode for the first phase of biphasic shocks. The pulse generator was positioned in a retropectoral pocket. In the first 70 patients who received model 7202D or 7219D pulse generators, the first-phase anode was a patch electrode positioned deep to the pulse generator (34 patients) or the patch combined with a coil electrode positioned at the junction of the innominate vein and superior vena cava (34 patients)27 or in the coronary sinus (2 patients). In the last 30 patients who received a pulse generator with an "active can" (Medtronic model 7218C or 7219C), the titanium shell of the ICD served as the first-phase anode.
Determination of the ULV. The main features of this method
have been reported.24 25 29 Briefly, pacing was performed
via the implanted bipolar electrodes at the right ventricular apex. The
baseline drive (S1) was delivered at a cycle length of 500
ms. All six surface ECG limb leads were recorded simultaneously on a
physiological recorder or computer screen. The T waves from all
recorded leads were inspected to select the lead with the
latest-peaking monophasic T wave. The interval from S1 to
the peak of this T wave was measured initially and after every other
change in shock strength. The first T-wave shock was delivered after
eight S1s to coincide with the peak of this T wave. If VF
was not induced, subsequent shocks were delivered 20 and 40 ms before
the peak. If any of the three shocks induced VF, the strength of the
next shock was increased; if none of the three shocks induced VF, the
strength of the next shock was decreased. After VF was induced in
patients 53 through 100, an additional shock with a coupling interval
of 20 ms after the peak was given at the weakest shock strength that
did not induce VF at 0, 20, and 40 ms before the peak. The ULV was
defined as the weakest measured shock strength that did not induce
VF.3 There was a 60-second recovery period between shocks
that failed to induce VF for strengths of
15 J and a 30-second
interval for weaker shocks.
The strength of the first test shock was 15 J in patients 1 through 22 and 73 through 100. In patients 23 through 72, it was 20 J. The strength of the shocks was changed by 5-J steps for energies >15 J. For energies between 15 and 5 J, the step-down increment was 5 J, followed by a step-up increment of 2.5 J. At 5 J, the step-down increment was 2.5 J. If VF was not induced at a programmed shock strength of 2.5 J, the ULV was the corresponding measured shock strength. This procedure resulted in a resolution of 5 J above the 15-J level and 2.5 J below it. All shocks were biphasic pulses from 120-µF capacitors with 65% tilt in each phase delivered by an external defibrillator (Medtronic model 2394).
Determination of the DFT. To permit direct comparison of the
DFT and the ULV, the sequence of shock strengths for DFT testing was
identical to the sequence for ULV testing.25 29 VF was
induced by T-wave shocks during ULV testing. If the ULV was determined
before the DFT was determined, subsequent episodes of VF were induced
by T-wave shocks of
2.0 J. The DFT was defined as the lowest measured
shock strength that terminated VF after 10 seconds. There was a
3-minute recovery period after each defibrillation shock. If the
defibrillation test shock failed, a 200-J rescue shock from an external
defibrillator was delivered 3 to 4 seconds later. Eighty-five patients
also participated in an additional investigational
protocol.25 27 29 30
ICD testing. In all patients, defibrillation was tested successfully at the programmable shock strength closest to the ULV+3 J25 by use of the implanted ICD system.
Data Recording
Voltage and current waveforms were recorded as described
previously.31 For each phase of the biphasic pulse, we
recorded the leading-edge voltage, trailing-edge voltage, leading-edge
current, and trailing-edge current. Resistance and stored energy were
calculated.25 31
Perioperative Complications
One patient developed hypotension requiring vasopressors after
induction of anesthesia and before ICD implantation; a second patient
developed pulmonary edema postoperatively. Both had severe left
ventricular dysfunction. One patient required reintubation for 18 hours
because of laryngeal edema. Two patients who were treated
postoperatively with heparin because of mechanical heart valves
developed wound hematomas requiring surgical drainage. The remaining
patients had uncomplicated clinical courses.
ICD Programming
The Medtronic Jewel is a tiered-therapy, biphasic-waveform
ICD.32 33 Its three tachyarrhythmia detection zones (VT,
fast VT, and VF) were programmed on the basis of the cycle length of
spontaneous and induced arrhythmias. We programmed a single detection
zone (VF) in 48 patients, two zones (VF and VT) in 38 patients, and
three zones (VF, VT, and fast VT) in 14 patients. For simplicity, in
this report fast VT is not distinguished from VT.
VF detection. The mean programmed cycle length below which arrhythmias were classified in the VF zone was 292±17 ms. It was 340 ms in 4 patients, 320 ms in 36 patients, 300 ms in 28 patients, 280 ms in 17 patients, and <280 ms in 15 patients. In 89 patients, detection of VF required that 18 of 24 intervals in a rolling detection window be in the VF-detection zone. The number of required intervals was 12 of 16 in 8 patients and 24 of 30 in 3 patients.
VF therapy. The first shock was set to the programmable shock strength34 closest to the ULV+5 J. This shock strength was chosen because analysis of retrospective data indicated that it defibrillated reliably,24 and preliminary analysis of prospective data indicated that a shock strength equal to the ULV+3 J defibrillated reliably.25 The strengths of the second and third shocks were programmed 6 to 10 J greater than the immediately preceding shock. Subsequent shocks were programmed to the maximum value of 34 J.
ICD memory. The Medtronic Jewel models 7202 and 7219 store RR intervals preceding the last five VT or VF episodes and after the last therapy delivered in each episode. An arrhythmia episode begins with detection of VT or VF and ends with redetection of "sinus rhythm," which is defined as eight consecutive intervals with cycle lengths longer than the longest cycle length detected as VT or VF. These ICDs store a total of 10 seconds of bipolar electrograms recorded at the right ventricular apex. They were programmed to store all 10 seconds for the last detected arrhythmia (VT or VF). Thus, electrograms were not available for episodes of ICD-detected VF if a subsequent episode of VT or VF was detected before device interrogation. Neither electrograms nor intervals were available if antitachycardia pacing or cardioversion given for an arrhythmia detected in the VT zone accelerated the rhythm to the VF zone; the only recorded data corresponded to the initial detection of VT. The Medtronic Jewel model 7218 had sufficient memory to record intervals or global, coil-to-can electrograms for all episodes of ICD-detected VF.35 All ICDs stored the capacitor-charge time for the last delivered VT or VF shock.
Follow-up
Patients were followed up for a total of 1694 patient-months
from the date of postoperative electrophysiological study until the
date of study closure or the patient's death (2 patients). The mean
duration of follow-up was 20±7 months (median, 20 months; range, 5 to
35 months). One patient died of heart failure and 1 of stroke. ICDs
were interrogated at 1 month and 3 months postoperatively, every 3
months thereafter, and whenever patients complained of lightheadedness,
palpitations, or shocks. Patients were specifically questioned about
syncope and presyncope whenever a shock occurred.
Antiarrhythmic drugs. During follow-up, amiodarone was discontinued in 3 of the 6 patients who were receiving it at the time of ICD implantation. Overall, 12 patients received antiarrhythmic drugs during follow-up: 8 to diminish the frequency of VT, 1 to prevent atrial flutter, and 3 to prevent symptomatic, paroxysmal atrial fibrillation. Eight of these patients received amiodarone and 4 received sotalol.
ICD reprogramming. Electrophysiological testing of the chronically implanted ICD was recommended if any of the following events occurred: cardiac surgery, acute myocardial infarction, change in heart failure status of more than one New York Heart Association functional class, or a 25% increase in the dose of antiarrhythmic drugs other than ß-blockers, digoxin, or calcium antagonists.26 The chronic ULV and DFT were determined as reported previously.26 The strength of the first shock was then reprogrammed to 5 J above the chronic ULV.
Classification of Rhythms Detected as VF
The cardiac rhythms associated with ICD-detected VF were
classified on the basis of data stored in the ICD in conjunction with
clinical data. Previously reported, prospective criteria for stored
detection intervals36 37 and stored
electrograms38 39 40 were used to diagnose monomorphic VT,
polymorphic VT, VF, atrial fibrillation, sinus tachycardia, and T-wave
oversensing during sinus tachycardia. ECG recordings were used for
classification whenever they were available. First shocks were
considered appropriate if the rhythm was classified as monomorphic VT,
polymorphic VT, or VF. They were also considered appropriate if the
detected rhythm occurred as a result of VT therapy unless ECG
recordings resulted in a different classification. First shocks were
considered inappropriate if they did not conform to these
classifications. Secondary arrhythmias were defined as arrhythmias that
occurred when therapy for an arrhythmia detected in the VT zone
accelerated the rhythm to the VF zone. These arrhythmias could be
classified for patients with the Medtronic Jewel model 7218 but not for
patients with models 7202 and 7219 because of the limited memory in
these devices.
First-Shock Efficacy
The Medtronic Jewel records the outcome of each delivered VF
therapy as effective, ineffective, undetermined, or conversion to VT.
The outcome of each appropriate first shock was analyzed by use of
stored postshock intervals, postshock electrograms (whenever
available), and clinical data. The efficacy of first VF shocks was
classified according to the recorded outcome unless this additional
analysis produced specific, unequivocal evidence resulting in a
different classification. The efficacy of VT shocks was not
analyzed.
Statistical Analysis
Data are presented as mean±SD. Basic comparative statistics
were calculated by the two-tailed, paired, or unpaired t
test. The Pearson correlation coefficient was used to assess the
correlation between DFT and ULV. We estimated 95% CIs for the
distribution of unclassified secondary arrhythmias from the
distribution for classified secondary arrhythmias by either the
2 test or Fisher's exact test. The first-shock
strength of the ULV+5 J was compared with the commonly used strengths
of the DFT+10 J and twice the DFT. The number of patients whose
first-shock strengths would be <25 J was determined for each of these
programming strategies and compared by the
2
test. This value of 25 J was selected because it approximates the
maximum output of the next generation of smaller ICDs. Patients who
received appropriate VF shocks were compared with the remaining
patients with respect to age, sex, type of structural heart disease,
left ventricular ejection fraction, and presenting arrhythmia. A value
of P<.05 was used to reject the null hypothesis for single
comparisons.
| Results |
|---|
|
|
|---|
ULV Versus DFT
Table 1
summarizes values for DFT and ULV in all
100 patients. The ULV was 3.0±3.0 J greater than the DFT, with a range
of -3.0 to 15.3 J. The correlation coefficient between DFT and ULV was
.83 for stored energy (P<.001) and .82 for voltage
(P<.001). Determination of the ULV required more total
shocks than determination of the DFT (9.8±3.0 versus 4.1±0.8,
P<.001) but fewer fibrillation-defibrillation episodes
(1.5±0.6 versus 4.1±0.8, P<.001).
|
First Shocks
The mean strength of first shocks was 17.5±5.2 J. During
follow-up, there were 120 appropriate first shocks in 37 patients.
These 37 patients did not differ from the remaining 63 patients with
respect to any of the clinical variables analyzed, ULV (12.4±5.8
versus 12.7±5.1 J, P=.82), or DFT (9.2±5.2 versus 9.8±4.9
J, P=.52). The number of appropriate first shocks was 1 in
14 patients, 2 in 8 patients, 3 in 4 patients, 4 in 4 patients, 5 in 2
patients, 6 to 10 in 2 patients, and
11 in 2 patients.
Arrhythmias Treated by First Shocks
Table 2
summarizes the appropriately treated
arrhythmias in the VF zone. Fig 1
shows representative
stored electrograms. Overall, 76% were primary arrhythmias and 24%
were secondary arrhythmias. Seventy percent were documented monomorphic
VT with cycle length shorter than the programmed cycle length for
detection of VF (82 primary arrhythmias in 31 patients and 2 secondary
arrhythmias in 1 patient). Eleven percent were documented VF (8 primary
arrhythmias in 6 patients and 7 secondary arrhythmias in 7 patients).
Nine of 29 secondary arrhythmias were classified as VF (7 arrhythmias
in 7 patients) or monomorphic VT (2 arrhythmias in 1 patient). Twenty
secondary arrhythmias (17% of all arrhythmias) in 15 patients could
not be classified because only data regarding the initial detection of
VT were available. The likely distribution of unclassified secondary
arrhythmias was estimated from the distribution for classified
secondary arrhythmias. Based on 95% CIs, 7 to 20 of the unclassified
arrhythmias were VF and 0 to 6 of them were VT; on a per-patient basis,
VF occurred in 5 to 15 of the patients who had unclassified arrhythmias
and VT occurred in 0 to 4 of them. Table 2
shows that there were no
significant differences among the strengths of first shocks delivered
for different arrhythmias. Overall, stored detection intervals were
available for 98 episodes (82%), electrograms for 78 episodes (65%),
and ECG recordings for 9 episodes (8%).
|
|
First-Shock Success
The ICD recorded 116 of the 120 appropriate first shocks as
successful, 1 as ineffective, and 3 as undetermined because the ICD was
interrogated before it classified 8 consecutive intervals as sinus.
These 3 shocks occurred in 1 patient during a clinical "storm" of
monomorphic VT. ECG recordings and postshock RR intervals documented
that each shock terminated the arrhythmia. The only ineffective shock
was delivered to treat an unclassified secondary arrhythmia. Before
this shock, the patient received four trials of ineffective
antitachycardia pacing followed by a 2-J cardioversion that accelerated
the rhythm. The strength of this first shock was 20 J, corresponding to
the DFT+12 J or 2.5 times the DFT. The second shock at 26 J was
effective. The cycle length for detection of VF was 320 ms.
Overall, the first shock was successful in 119 of 120 episodes (99%; 95% CIs, 93% to 100%). The only unsuccessful shock was delivered for an unclassified arrhythmia that could have been VT, VF, or atrial fibrillation induced by low-energy cardioversion. If we use 95% CIs to estimate the distribution of unclassified shocks, first shocks were successful in at least 84 of 85 VT episodes (99%) and at least 21 of 22 VF episodes (95%). On a per-patient basis, 95% CIs indicate that first shocks were successful in at least 30 of 31 patients with VT (97%) and at least 17 of 18 patients with VF (94%).
Syncope
No patient had syncope associated with an ICD shock. The patient
who had the ineffective shock was the only patient who experienced
presyncope. The charge time was recorded for 69 first-VF shocks with a
shock strength of 18.1±5.7 J. It was 3.3±0.7 seconds.
Relationship of First-Shock Strength to DFT
Fig 2
shows the difference between the first-shock
strength and the DFT in all 100 patients. The shock was 8.0±3.0 J
greater than the DFT (range, 2 to 20 J). It was within 5 J of the
DFT in 28 patients and >10 J above the DFT in 17 patients. The ratio
of the first shock to the DFT was 2.1±0.8 (range, 1.2 to 4.9). The
mean first-shock strength was less than the strength that would have
been programmed if a guideline of the DFT+10 J (20.6±5.0 J,
P<.001) or twice the DFT (19.1±9.7 J, P=.01)
had been used. Furthermore, a shock strength of twice the DFT exceeded
the 34-J maximum output of the ICD in 7 patients. Fig 3
shows a percentile plot of the programmed stored energy for the first
shock in all 100 patients in comparison with values that would have
been programmed if the DFT+10 J or twice the DFT had been used. The
number of patients whose programmed shock strength was
25 J was 93
for the ULV+5-J strategy. It was lower (78 patients, P=.008)
for the twice-DFT strategy and not significantly different (90
patients, P=.62) for the DFT+10-J strategy.
|
|
On a per-shock basis, the 120 delivered appropriate first shocks exceeded the corresponding DFTs by 7.4±2.4 J and were 2.0±0.6 times as strong as the DFT. The 37 patients who received appropriate VF shocks did not differ from the remaining 63 patients with respect to the difference between the strength of the first shock and the DFT (8.3±2.7 versus 7.9±3.2 J, P=.52) or the ratio of the first shock to the DFT (2.2±0.7 versus 2.1±0.8, P=.67).
ICD Reprogramming
Nine patients underwent chronic electrophysiological testing
and subsequent reprogramming of the first shock. The chronic ULV
increased by
10 J in 4 patients and by 5 to 10 J in 1 patient. It
differed from the acute ULV by <5 J in 4 patients. Only 1 patient
received VF therapy after the shock strength was increased at
reprogramming.
| Discussion |
|---|
|
|
|---|
Programming Strategies for ICDs
Programming based on the ULV may be compared with conventional
programming strategies.
Programming by DFT. The principal advantage is that patient-specific programming may limit excessive shock strength during follow-up. The principal disadvantage is that multiple fibrillation-defibrillation episodes are required at implantation. Although this testing generally is considered safe, complications include cerebral hypoperfusion,8 10 electroencephalographic changes10 associated with postoperative cognitive deficits,47 myocardial ischemia,11 systemic hypoperfusion,13 diminished left ventricular function,12 14 15 prolonged circulatory arrest due to intractable VF,15 16 17 and rarely, death.11 12
Programming by defibrillation safety margin. This strategy limits fibrillation-defibrillation testing to the minimum number of episodes necessary to determine whether there is a sufficient safety margin between the maximum output of the ICD and the shock strength required for consistent defibrillation.7 8 9 48 49 The first shock is then programmed to maximum output. The disadvantages apply to patients who do not require maximum-strength shocks during long-term follow-up. Maximum-output shocks may increase the risk of syncope during capacitor charging (see below). They may result in rapid battery depletion if the ICD capacitors charge frequently because of appropriate shocks, inappropriate shocks, or aborted shocks.50 In in vitro and animal experiments, excessively strong shocks cause myocardial depression1 2 and conduction block.4 5 6 In humans, multiple high-energy shocks have been associated with a poor prognosis, but it is not known whether they cause adverse clinical consequences or merely identify patients at high risk.51 52
Programming by ULV. This strategy both provides patient-specific programming and minimizes the number of fibrillation-defibrillation episodes. It thus combines the advantages of programming by DFT and programming by ICD safety margin. The principal disadvantage is the need for more total shocks during paced rhythm. Although clinical data from the present study and clinical24 and creatine kinase isoenzyme data25 from previous studies have identified no complications from ULV testing, subclinical effects have not been studied. The ULV method could be adapted to a "vulnerability" safety-margin approach. For example, if VF is not induced by a T-wave shock at 15 J, the ICD could be programmed to 20 J. This method would minimize both the number of shocks and the number of fibrillation-defibrillation episodes.
First-Shock Success: Conventional Programming
For biphasic-waveform, transvenous ICDs, several studies report
success rates for appropriate first shocks as assessed by stored
electrograms. Trappe et al53 reported that a shock
strength equal to the DFT+10 J resulted in a success rate of 75%.
Vijgen et al54 reported a success rate of 85% for a shock
equal to or greater than the DFT+9 J. For patients whose DFTs were
15
J, Heisel et al55 reported that the success rate for a
shock strength of twice the DFT was 98%. With the maximum-output,
safety-margin strategy for the same ICDs as in the present study, the
success rate has been reported to be 90% for the active-can
configuration and 95% for systems without the active
can.32
First-Shock Success: Programming by ULV
In the present study, a shock strength equal to the ULV+5 J
resulted in a 99% success rate despite a lower mean shock strength
than programming based on the DFT. There are at least two plausible
explanations for this apparent paradox. (1) The ULV permits an accurate
estimate of the minimum shock strength that results in near 100%
defibrillation because it is highly reproducible56 and
corresponds to a shock strength with a high (90%) and highly
predictable probability of defibrillation.25 This permits
selection of a patient-specific shock strength that has a high success
rate but not an excessive safety margin. In contrast, the DFT does not
permit an accurate estimate of the minimum shock strength that results
in uniformly successful defibrillation because it is less
reproducible,56 and the probability of successful
defibrillation at the DFT is lower and less predictable than at the
ULV.8 57 To select a patient-specific shock strength that
has a high success rate, an excessive safety margin must be programmed
in some patients. However, in any individual patient, the shock
strength programmed on the basis of the ULV may be higher than that
programmed on the basis of the DFT, as shown in Fig 2
. (2) Programming
based on the DFT might have performed as well as programming based on
the ULV in our patients. However, because the only unsuccessful shock
in this study was programmed to the DFT+12 J or 2.5 times the DFT, it
is unlikely that programming to the DFT+10 J or twice the DFT would
have performed better.
Syncope
Because the time required to charge an ICD capacitor is longer for
stronger shocks, the duration of the arrhythmia before the shock is
also longer. Thus, the risk of syncope with possible resultant trauma
may be higher when shocks are programmed to maximum output. For a
2-year-old Medtronic Jewel, the minimum charge time is 4.5 seconds for
the 18-J mean first-shock strength in the present study versus 11
seconds for the maximum shock strength of 34 J. Syncope or presyncope
has been reported to occur in 17% to 42% of appropriate shock
episodes when the strength of the first shock is
maximum.58 59 In the present study, no patient had syncope
and only 1 had presyncope. This may be due in part to short charge
times.
Strengths and Limitations of Conventional Programming
Strategies
These strategies result in operative mortality of
1% for
biphasic-waveform, transvenous ICDs.32 60 However, as
noted previously, DFT testing still results in rare mortality and
finite morbidity. Programming to the DFT+10 J is associated with a
significant first-shock failure rate,53 54 61 and
programming to twice the DFT would not have been applicable to the 10%
of patients in the present study whose DFTs exceeded 15 J. Furthermore,
this latter strategy may limit the fraction of patients who are
candidates for smaller, lower-output ICDs. Programming first shocks to
maximum output results in occasional but important problems, such as
rapid battery depletion due to frequent capacitor charging or syncope
associated with long capacitor charge times. Patients who have frequent
self-terminating arrhythmias in the VF zone (eg, torsade de pointes)
present a particularly difficult problem for this strategy: A long
detection time reduces the number of aborted shocks but places patients
at risk for syncope because of both long detection times and charge
times; a short detection time results in frequent aborted shocks and
rapid battery depletion. ICDs have reduced the annual sudden death rate
to
2%.32 60 However, among patients who have
arrhythmias in the VF zone, it may be as high as 8.7%,60
and ICDs have been less successful at decreasing total
mortality.51 52 It is not known whether excessive shock
strength contributes to morbidity and mortality from heart failure in
some ICD recipients.
Limitations of the Present Study
1. The number of arrhythmias in the VF zone is limited, and the
corresponding number of patients is even smaller. This is a consequence
both of the relative infrequency of rapid VT and primary
VF36 60 and of programming of an intermediate zone for
fast VT in some patients.
2. In this study, most arrhythmias in the VF zone were rapid VT rather than VF. This finding is consistent with the observations of Raitt et al36 and previous studies of first-shock success rate in which stored electrograms or intervals were recorded.32 53 54 55 61 Thus, the first-shock success rate in this study applies to arrhythmias in the VF zone of ICDs as they are programmed in clinical practice. This differs from previous studies that correlated the ULV with the shock strength required to terminate induced VF. We have limited first-shock success data for true defibrillation; this limitation also applies to previous studies of first-shock success rates.
3. We did not compare programming based on the ULV with a conventional programming strategy. A prospective study would be needed for direct comparison of perioperative morbidity, first-shock success rates, syncope during capacitor charging, shock-induced morbidity, and pulse-generator longevity.
4. The vulnerability-safety-margin method has not been validated or compared with the defibrillation-safety-margin method.
5. Limitations of the ULV method have been described.25 29 We do not know how well it applies to patients taking antiarrhythmic drugs other than amiodarone.24
Clinical Implications
Because current programming methods prevent sudden death
with acceptable morbidity, independent confirmation of our method is
desirable before widespread clinical application. However, programming
based on vulnerability testing may be recommended in patients at higher
than usual risk for fibrillation-defibrillation testing. Examples may
include patients in whom external-rescue defibrillation is unreliable
or those with high-grade coronary stenoses or recent coronary
interventions. Depending on the benefits of minimizing shock strength
in a given patient, vulnerability testing may be performed with either
a threshold or safety-margin strategy. A threshold strategy is
preferred for patients who have frequent self-terminating arrhythmias
in the VF zone (to prevent rapid battery depletion) and for those in
whom it is important to minimize the risk of syncope during capacitor
charging. Our present recommendations apply only to the strength of the
first shock and assume that subsequent shocks will be programmed to a
higher energy.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
Received August 21, 1996; revision received October 31, 1996; accepted November 18, 1996.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. Lemke, T. Lawo, M. Zarse, A. Lubinski, U. Kreutzer, J. Mueller, A. Schuchert, S. Mitzenheim, D. Danilovic, T. Deneke, et al. Patient-tailored implantable cardioverter defibrillator testing using the upper limit of vulnerability: the TULIP protocol Europace, May 30, 2008; (2008) eun136v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. B. Curtis Defibrillator Implantation Without Induction of Ventricular Fibrillation: Good Enough? Circulation, May 8, 2007; 115(18): 2370 - 2372. [Full Text] [PDF] |
||||
![]() |
J. D. Day, R. N. Doshi, P. Belott, U. Birgersdotter-Green, M. Behboodikhah, P. Ott, K. A. Glatter, S. Tobias, H. Frumin, B. K. Lee, et al. Inductionless or Limited Shock Testing Is Possible in Most Patients With Implantable Cardioverter- Defibrillators/Cardiac Resynchronization Therapy Defibrillators: Results of the Multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations) Circulation, May 8, 2007; 115(18): 2382 - 2389. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Li, V. Nikolski, D. W. Wallick, I. R. Efimov, and Y. Cheng Mechanisms of enhanced shock-induced arrhythmogenesis in the rabbit heart with healed myocardial infarction Am J Physiol Heart Circ Physiol, September 1, 2005; 289(3): H1054 - H1068. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Swerdlow, K. Shivkumar, and J. Zhang Determination of the Upper Limit of Vulnerability Using Implantable Cardioverter-Defibrillator Electrograms Circulation, June 24, 2003; 107(24): 3028 - 3033. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Wathen, M. O. Sweeney, P. J. DeGroot, A. J. Stark, J. L. Koehler, M. B. Chisner, C. Machado, and W. O. Adkisson Shock Reduction Using Antitachycardia Pacing for Spontaneous Rapid Ventricular Tachycardia in Patients With Coronary Artery Disease Circulation, August 14, 2001; 104(7): 796 - 801. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. D. Swerdlow Implantation of Cardioverter Defibrillators Without Induction of Ventricular Fibrillation Circulation, May 1, 2001; 103(17): 2159 - 2164. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Glatter and L. B. Liem Implantable Cardioverter Defibrillator: Current Progress and Management Seminars in Cardiothoracic and Vascular Anesthesia, September 1, 2000; 4(3): 162 - 179. [Abstract] [PDF] |
||||
![]() |
P. N. Schauerte, K. Ziegert, M. Waldmann, F. A. Schondube, F. Birkenhauer, K. Mischke, M. Grossmann, P. Hanrath, and C. Stellbrink Effect of Biphasic Shock Duration on Defibrillation Threshold With Different Electrode Configurations and Phase 2 Capacitances : Prediction by Upper-Limit-of-Vulnerability Determination Circulation, March 23, 1999; 99(11): 1516 - 1522. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Windecker, R. E. Ideker, V. J. Plumb, G. N. Kay, G. P. Walcott, and A. E. Epstein The influence of ventricular fibrillation duration on defibrillation efficacy usin |