(Circulation. 1997;96:2124-2127.)
© 1997 American Heart Association, Inc.
Articles |
From the Departments of Diagnostic Radiology (A.C., H.R., M.G.), Cardiology (D.H.), Thoracic and Cardiovascular Surgery (C.D., A.H., H.G.B.), and Anesthesiology (K.J.), Hannover Medical School, Germany.
Correspondence to Ajay Chavan, MD, DMRD, Department of Diagnostic Radiology, Hannover Medical School, Carl Neuberg Str 1, 30625 Hannover, Germany.
| Abstract |
|---|
|
|
|---|
Methods and Results Nine percutaneous fenestrations (one with additional stenting of the infrarenal true aortic lumen) were performed under local anesthesia in seven patients with aortic dissection. The presenting symptoms were abdominal angina or claudication. By the transfemoral approach, the intimal flap was initially punctured with a needle-catheter combination through which a guidewire was placed across the dissection flap. The fenestration was carried out with a balloon catheter introduced over the guidewire. The procedure was performed under on-line guidance with intravascular ultrasound imaging. The procedure was performed successfully and without complications in all patients. After intervention, symptoms resolved in all seven patients.
Conclusions Intravascular ultrasoundguided percutaneous fenestration of the intimal flap in symptomatic aortic dissections with distal vessel involvement is a technically feasible and safe procedure that can effectively relieve the patient's symptoms.
Key Words: dissection, aortic fenestration ultrasonics
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
|
After diagnostic angiograms of the two lumina and the
branch vessels (Fig 3a
and 3b
) had been obtained, a 6.2F, 12.5-MHz IVUS
catheter (Sonicath, Boston Scientific Corp) was introduced
transfemorally into the aorta over a 0.025-in guidewire (Fig 1a
). From the opposite femoral artery (9F
sheath), a Ross transjugular liver-access needle-catheter combination
(William Cook Europe) was introduced over a 0.035-in stiff guidewire
(Amplatz Super Stiff, Boston Scientific Corp), and its tip was
positioned at the level of the transducer on the IVUS catheter (Fig 1a
). While the needle was advanced, its tip was always kept covered and
protected by the outer catheter (Fig 1a
). Under IVUS guidance, the
needle was rotated and its tip positioned against the center of the
dissection flap so that it did not point toward the free aortic wall
(Fig 2a
). After the guidewire was removed
and the outer catheter retracted, the flap was perforated with the
needle by a quick, short thrust. The intraluminal position of the
needle tip now in the second lumen was confirmed with IVUS (Fig 2b
),
and the outer catheter was advanced over the needle from one lumen into
the other. After an angiogram was performed with the catheter tip in
the second lumen, the stiff guidewire was reintroduced and the
needle-catheter combination removed, leaving the guidewire placed
across the intimal flap. A balloon catheter with a balloon diameter of
15 mm (Vas-cath Inc) was advanced over this guidewire and
positioned across the intimal flap. Inflation of the balloon created
the desired fenestration. A "waisting" of the balloon during
inflation indicated satisfactory balloon position (Fig 1b
). The outcome
of the procedure was controlled by a flush aortogram (Fig 3c
and 3d
) and with IVUS imaging after an
agitated mixture containing 19 mL normal saline and 1 mL air was
injected at the level of the fenestration (Fig 2c
and 2d
). True and
false lumen manometry was carried out in two patients (Table
, patients
W.H. and S.W.). In two patients, small preexisting reentries above the
aortic bifurcation were widened with a balloon catheter. One patient
(Table
, W.H.) underwent stenting of the true lumen of the infrarenal
aorta 4 months after fenestration with a 5-cm-long stent that was
expanded to a diameter of 18 mm (Palmaz stent, Johnson & Johnson).
The previous fenestration was also widened with a 20-mm balloon.
|
|
|
| Results |
|---|
|
|
|---|
Clinical Results
In one patient, abdominal angina resolved after the fenestration.
In four of the other six patients suffering from claudication,
numbness, or leg pain, the symptoms disappeared after just one
fenestration (Table
). In one patient (Table
, M.G.), no significant
relief of symptoms was observed after the first fenestration at the
level of the celiac trunk. After a second fenestration above the aortic
bifurcation, his claudication resolved. In the last patient (Table
,
W.H.), the walking distance improved only moderately, from 1000 to
1500 m after fenestration at the level of the renal arteries.
After infrarenal stenting of the narrowed true lumen, his claudication
disappeared.
| Discussion |
|---|
|
|
|---|
Four of the seven patients responded favorably to just one fenestration
performed between the levels of the celiac trunk and the renal
arteries. In two further patients with claudication (Table
, M.G. and
W.H.), the symptoms of claudication did not improve significantly after
an initial high fenestration. In one of them (Table
, M.G., and Fig 3
),
who had left leg claudication, an initial fenestration at the level of
the celiac trunk did not result in clinical improvement. In retrospect,
only a high fenestration with the intention of decompressing the false
lumen and improving the perfusion of the true lumen of the left common
iliac artery could not have been expected to cause clinical
improvement, because the left common iliac artery was not adequately
opacified from either of the two lumina after the first procedure. A
second fenestration above the aortic bifurcation that improved local
blood flow across the dissection flap into the left common iliac artery
resulted in resolution of symptoms. The second patient (Table
, W.H.)
improved clinically only after stenting of the narrowed true lumen of
the infrarenal aorta and widening of the fenestration with a 20-mm
balloon. The seventh patient with claudication (Table
, S.W.), who
underwent one fenestration at the level of the renal arteries and one
above the aortic bifurcation, also responded well clinically. Thus, it
appears that just one fenestration (with a 15-mm balloon) to decompress
the false lumen may not be adequate in relieving symptoms; it may be
necessary to resort to a second fenestration further downstream to
provide local blood flow across the intimal flap15 or to
measures such as stenting of the true lumen to keep it
patent.12
To the best of our knowledge, 20 cases of percutaneous fenestration in patients with aortic dissection have been reported.8 9 10 11 12 13 16 17 18 19 20 21 Six of these patients underwent stenting of the true aortic lumen in addition to the fenestration.11 12 13 No major complications occurred. Three of these 20 patients died of causes not related to the procedure.9 10 18 The remaining patients responded well clinically, as did our seven patients. All our patients had chronic dissections. However, this procedure has also been performed successfully in the acute and subacute setting, with immediate clinical improvement in most cases.8 9 10 11 16 17 18 21 On the basis of our data and the data in the literature, percutaneous fenestration in symptomatic aortic dissections with distal vessel obstruction appears to be a technically feasible and safe procedure associated with a low complication rate. Most of the patients respond well symptomatically, thus obviating the need for surgical relief of the obstruction. Additional measures such as stent implantation in the true lumen may be necessary for complete relief of symptoms in some cases.
Received May 27, 1997; revision received July 22, 1997; accepted July 25, 1997.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
H. E. Garrett Jr and B. A. Wolf Management of Acute Infrarenal Aortic Occlusion Secondary to Type A Dissection Ann. Thorac. Surg., April 1, 2006; 81(4): 1500 - 1502. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. H. Koschyk, C. A. Nienaber, M. Knap, T. Hofmann, Y. V. Kodolitsch, V. Skriabina, M. Ismail, O. Franzen, T. C. Rehders, C. Dieckmann, et al. How to Guide Stent-Graft Implantation in Type B Aortic Dissection?: Comparison of Angiography, Transesophageal Echocardiography, and Intravascular Ultrasound Circulation, August 30, 2005; 112(9_suppl): I-260 - I-264. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. G. Hartnell and J. Gates Aortic Fenestration: A Why, When, and How-to Guide RadioGraphics, January 1, 2005; 25(1): 175 - 189. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. T. O'Gara, A. J. Greenfield, N. A. Afridi, and S. L. Houser Case 12-2004 - A 38-Year-Old Woman with Acute Onset of Pain in the Chest N. Engl. J. Med., April 15, 2004; 350(16): 1666 - 1674. [Full Text] [PDF] |
||||
![]() |
S. D. Moffatt and R. S. Mitchell Endovascular Stent Management of Thoracic Aneurysms and Dissections Card. Surg. Adult, January 1, 2003; 2(2003): 1191 - 1204. [Full Text] |
||||
![]() |
R. H. Mehta, P. T. O'Gara, E. Bossone, C. A. Nienaber, T. Myrmel, J. V. Cooper, D. E. Smith, W. F. Armstrong, E. M. Isselbacher, L. A. Pape, et al. Acute type A aortic dissection in the elderly: clinical characteristics, management, and outcomes in the current era J. Am. Coll. Cardiol., August 21, 2002; 40(4): 685 - 692. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L. Packer, C. L. Stevens, M. G. Curley, C. J. Bruce, F. A. Miller, B. K. Khandheria, J. K. Oh, L. J. Sinak, and J. B. Seward Intracardiac phased-array imaging: methods and initial clinical experience with high resolution, under blood visualization: Initial experience with intracardiac phased-array ultrasound J. Am. Coll. Cardiol., February 6, 2002; 39(3): 509 - 516. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. H. Mehta, T. Suzuki, P. G. Hagan, E. Bossone, D. Gilon, A. Llovet, L. C. Maroto, J. V. Cooper, D. E. Smith, W. F. Armstrong, et al. Predicting Death in Patients With Acute Type A Aortic Dissection Circulation, January 15, 2002; 105(2): 200 - 206. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Sailer, P. Peloschek, T. Rand, M. Grabenwoger, S. Thurnher, and J. Lammer Endovascular Treatment of Aortic Type B Dissection and Penetrating Ulcer Using Commercially Available Stent-Grafts Am. J. Roentgenol., December 1, 2001; 177(6): 1365 - 1369. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Passman, W. A. Marston, M. A. Farber, L. V. Owens, F. S. Sandhu, P. F. Jaques, and B. A. Keagy Percutaneous Balloon Fenestration and Endovascular Stent Placement for Type B Acute Aortic Dissection with Renal Ischemia: A Case Report Vascular and Endovascular Surgery, September 1, 2000; 34(5): 459 - 465. [Abstract] [PDF] |
||||
![]() |
P. G. Hagan, C. A. Nienaber, E. M. Isselbacher, D. Bruckman, D. J. Karavite, P. L. Russman, A. Evangelista, R. Fattori, T. Suzuki, J. K. Oh, et al. The International Registry of Acute Aortic Dissection (IRAD): New Insights Into an Old Disease JAMA, February 16, 2000; 283(7): 897 - 903. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Nienaber, R. Fattori, G. Lund, C. Dieckmann, W. Wolf, Y. von Kodolitsch, V. Nicolas, and A. Pierangeli Nonsurgical Reconstruction of Thoracic Aortic Dissection by Stent-Graft Placement N. Engl. J. Med., May 20, 1999; 340(20): 1539 - 1545. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1997 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |