(Circulation. 1997;96:2813-2822.)
© 1997 American Heart Association, Inc.
Articles |
From the Division of Cardiology, Department of Medicine (C.D.S.) and Department of Cardiovascular Surgery (R.M.K.), Cedars-Sinai Medical Center, and University of California Los Angeles School of Medicine. J.E. Brewer is now at SurVivalink Corp, Minneapolis, Minn; Dr Kroll is now at Pacesetter Corp, Sylmar, Calif.
Correspondence to Charles D. Swerdlow, MD, Cedars-Sinai Medical Towers, 8635 W Third St, Suite 1190 W, Los Angeles, CA 90048. E-mail swerdlow{at}ucla.edu
| Abstract |
|---|
|
|
|---|
m). We hypothesized that these models predict results
of clinical studies of ICD capacitance if human time constants are
used.
Methods and Results We studied 12 patients with epicardial
ICDs and 15 patients with transvenous ICDs. Defibrillation threshold
(DFT) was determined for 120-µF monophasic capacitive-discharge
pulses at pulse widths of 1.5, 3.0, 7.5, and 15 ms. To compare the
predictions of the average-current versus leading-edge-current methods,
we derived a new exponential average-current model. We then calculated
individual patient time parameters for each model. Model
predictions were validated by retrospective comparison with clinical
crossover studies of small-capacitor and standard-capacitor waveforms.
All three models provided a good fit to the data
(r2=.88 to .97, P<.001). Time
constants were lower for transvenous pathways (53±7
) than
epicardial pathways (36±6
) (tc, P<.001;
average-current
m, P=.002;
leading-edge-current
m, P<.06). For
epicardial pathways, optimal capacitance was greater for either
average-current model than for the leading-edge-current model
(P<.001). For transvenous pathways, optimal capacitance
differed for all three models (P<.001). All models provided
a good correlation with the effect of capacitance on DFT in previous
clinical studies: r2=.75 to .84,
P<.003. For 90-µF, 120-µF, and 150-µF capacitors,
predicted stored-energy DFTs were 3% to 8%, 8% to 16%, and 14% to
26% above that for the optimal capacitance.
Conclusions Model predictions based on measured human
cardiac-muscle time parameter have a good correlation with
clinical studies of ICD capacitance. Most of the predicted reduction in
DFT can be achieved with
90-µF capacitors.
Key Words: defibrillation waves heart-assist device
| Introduction |
|---|
|
|
|---|
pathway is 30 to 43 µF1 2 3 6 and that reduction in
capacitance will result in substantial improvement in
DFT.6 Small capacitors have performed better than
conventional capacitors8 9 10 11 12 in animals, but they have
shown modest13 or no14 15 16 17 18 19 benefit for the
majority of clinical defibrillation pathways. The objective of this
study was to test the hypothesis that these models predict results of
clinical studies of ICD capacitance if human cardiac-muscle time
parameters are used. A secondary objective was compare
optimal capacitance predicted by average-current and
leading-edge-current methods directly by use of the same data and
parallel models. | Methods |
|---|
|
|
|---|
s, which represents the exponential-decay time
constant of the defibrillation pulse.
s is the
product of pathway resistance and ICD capacitance. Theoretical
models1 2 4 define the optimal defibrillation waveform as
that which minimizes stored-energy DFT, because stored energy is a
critical determinant of ICD pulse generator size. Each model predicts
the optimum value of
s in terms of a time
parameter that characterizes the response of cardiac tissue
to the shock. Strength-duration equations are shown in the
"Appendix."
Monophasic Models
Hyperbolic average-current model. The empirical
hyperbolic strength-duration relationship20 21 may be
applied to capacitive-discharge defibrillation pulses by expressing
shock strength as average current during the pulse.22 23
This hyperbolic average-current model gives the DFT in terms of
tc and the average-current rheobase, which is the DFT at an
infinite pulse width. The waveform optimization strategy is summarized
in the "Appendix." The optimal value of
s is 0.796
tc1 2 and the optimum pulse duration is 1.045
tc.2
Leading-edge-current exponential model. A strength-duration
relationship also can be derived from Blair's resistor-capacitor model
of the cardiac cell membrane.24 25 This exponential
formulation of the strength-duration relationship was extrapolated to
describe stimulation by capacitive-discharge pulses26 and
subsequently applied to defibrillation.3 5 This model
assumes that defibrillation depends only on the peak cell response to a
shock pulse rather than the average applied current during the
pulse.4 5 6 26 The corresponding strength-duration relation
gives leading-edge-current DFT for a capacitive-discharge pulse in
terms of two time constants,
s and
Blair's24 25 "cell membrane"
m. The
third model parameter is the leading-edge-current rheobase
at an infinite value of
s. The optimum value of
s and the optimum pulse duration are both equal to
m.4 5 6 Fig 1
shows the effect of varying the relationship between
s
and
m on the predicted membrane-response and the
corresponding strength-duration curves.
|
Exponential average-current model. The average-current model
is based on the hyperbolic strength-duration relation, and the
leading-edge-current model results in an exponential one. To permit
direct comparison of predictions based on average current with those
based on leading-edge current, we derived a new exponential
average-current model. It gives the average-current DFT in terms of the
average-current
m and the average-current rheobase. The
optimal value of
s is
m, and the optimum
pulse duration is 1.337
m. Because average-current
m and leading-edge-current
m may differ,
the predicted optimal values of
s are not in general
equal for the two exponential models. Fig 1
shows that the exponential
average-current model is the limit of the leading-edge-current model
as
s
.
Biphasic Models
The optimal first phase of the biphasic waveform is assumed to
be the optimal monophasic waveform.3 4 Thus, predicted
optimal
s is equal for monophasic and biphasic models.
The optimal second phase is assumed to prevent refibrillation by
removing or "burping" the charge deposited on myocardial cells by
phase 1.3 4 12 The initial conditions for phase 2
(cell-membrane voltage and ICD-capacitor voltage) are set equal to the
corresponding values at the end of the optimal phase 1 as predicted by
each monophasic model. The leading-edge-current model is then used
to determine the optimal duration of phase 2. If phase 1 is optimized
by an average-current method, the biphasic model becomes a hybrid
model, using different methods for each phase.3
Experimental Methods
Patients
Patients undergoing ICD implantation participated in this study
after giving written informed consent according to a protocol approved
by the Human Subjects Committee. Because of the required number of
fibrillation-defibrillation episodes, patients were excluded if their
New York Heart Association class for heart failure was 3 or 4 after
optimal therapy, their left ventricular ejection fraction
was <.25, or a there was a proximal stenosis
70% in a major
coronary artery supplying viable myocardium.
Patients were excluded if they had ever received amiodarone,
and other antiarrhythmic drugs were discontinued for 5 half-lives.
However, therapy with digoxin (4 patients), ß-blockers (7 patients),
and calcium channel blockers (1 patient) was continued. Patients were
studied at new implants of ICDs with a single transvenous electrode
configuration (24 patients) or pulse generator change in patients with
a single epicardial electrode configuration (12 patients). Six patients
with transvenous ICDs were excluded from this study because the
biphasic DFT was too high (see below), and the study was aborted in 1
patient with a transvenous ICD because of hypotension. Thus, complete
data were available for 27 patients. They included 18 men and 9 women
with a mean age of 59±15 years. The mean left ventricular
ejection fraction was 0.39±0.11. Eighteen patients had
coronary artery disease, 8 patients had myocardial or
congenital disease, and 1 patient had idiopathic long-QT syndrome. The
clinical arrhythmia was sustained monomorphic
ventricular tachycardia in 16 patients and
ventricular fibrillation in 11 patients.
Surgical Technique and Electrode Configurations
Patients were studied intraoperatively as described
previously.18 The epicardial defibrillation pathway
included a large patch electrode (Medtronic model 6921L) positioned
posteriorly over the left ventricle and a large (5 patients) or medium
(Medtronic model 6921M) (7 patients) patch electrode positioned
anteriorly over the right ventricle. The left ventricular
electrode was the cathode, and the right ventricular
electrode was the anode. In patients with transvenous electrodes, a
tripolar electrode with a 5-cm defibrillation coil (Medtronic model
6966 or 6936) was positioned in the right ventricular apex.
The titanium shell of an ICD pulse generator (Active Can Emulator, 83
cm3) was positioned in a retropectoral pocket. The right
ventricular coil served as the cathode, and the titanium
shell served as the anode.
Defibrillation Waveforms
Models of defibrillation permit estimation of optimal biphasic
waveforms from monophasic strength-duration data.3 4 We
used monophasic waveforms to minimize the number of variables that
might influence calculation of the defibrillation time
parameters. Study waveforms were fixed-duration, truncated
exponential pulses delivered by an external defibrillator with a
nominal 120-µF output capacitance (model 2394, Medtronic Inc).
Clinically indicated testing was performed using biphasic pulses with
65% tilt in each phase. Polarity was reversed for phase 2 of biphasic
pulses.
Clinical Testing
First, the DFT was determined with biphasic pulses by a
step-down or step-up method. The first programmed leading-edge voltage
was 400 V, and the step size was 100 V. Patients were excluded from the
study if the biphasic DFT exceeded 500 V to avoid the possibility that
the monophasic DFT at a pulse width of 1.5 ms would exceed the maximum
output of the external defibrillator. This excluded 6 patients with
transvenous electrodes.
DFT Testing
Four monophasic pulse durations were tested in random order in
each patient: 1.5, 3, 7.5, and 15 ms. The DFT50 was
estimated by a previously described delayed three-step up-down
algorithm27 28 with 50-V steps. The method for selecting
the strength of the first defibrillation test shock is described below.
If this shock succeeded, the strength of the second defibrillation test
shock was decreased by 50 V. If it failed, the strength of the second
shock was increased by 50 V. This process was repeated until there was
a reversal of response from success to failure or from failure to
success. Then the strength of the next defibrillation test shock was
changed by 50 V in the opposite direction. The shock strength before
the first reversal of response was the first data point, and the
strengths of the subsequent two shocks were the second and third data
points. The fourth data point was predicted from the outcome of the
third defibrillation shock but not tested. The average of these four
data points was taken as the DFT50.
In a delayed up-down algorithm, the number of fibrillation-defibrillation episodes is minimized if the strength of the first test shock is near the DFT50. To estimate the strength of this first test shock, we determined the upper limit of vulnerability29 for the first of the four test waveforms. This approximates the DFT90.29 The method for determining the upper limit of vulnerability was modified to set the strength of the first monophasic upper-limit test shock at 200 V greater than the biphasic DFT and the step size to 100 V. We then set the strength of the first defibrillation test shock to 100 V below the upper limit of vulnerability. In the first 5 patients, the initial test shock for the second waveform tested was equal to the DFT50 for the first waveform. For the third and fourth waveforms, the initial test shock was equal to the average of the DFT50s for the preceding waveforms. Beginning with the sixth patient, the initial test voltage for shocks with duration of 1.5 ms was set 20% higher than that of the other three pulse durations. Defibrillation test shocks were given after 10 seconds of induced ventricular fibrillation. Overall, patients had 14.2±1.4 monophasic defibrillation shocks in this study. They also had three or four biphasic defibrillation shocks for clinical reasons.
Data Acquisition
This method has been described previously.18
Voltage and current waveforms were digitized at 100 kHz with the
Mac-Adios Board (GW Instruments) and recorded on a Macintosh
computer. A custom-modified oscilloscope emulation program (SuperScope
II, GW Instruments) was used to record voltage and current
waveforms and to detect the leading- and trailing-edge voltages and
currents.
Data Analysis
Mean resistance was determined by averaging the point-by-point
quotient of the voltage waveform divided by the current waveform. Pulse
duration was calculated as the difference in timing of the leading-edge
and trailing-edge voltages. Previously described methods were used to
calculate stored energy16 and average
current2 18 for shock pulses with a measured capacitance
of 122.2±1.6 µF.18 Fig 2
shows equivalent average-current waveforms for the truncated
exponential waveforms tested. We constructed strength-duration curves
for each patient by the method of least squares using Matlab 4.2 for
the Macintosh (The MathWorks Inc). Average-current strength-duration
data were fit to the hyperbolic average-current model of Equation 6
in
the "Appendix" to determine tc and to the exponential
average-current model of Equation 9
to determine average-current
m. To characterize these two strength-duration relations
quantitatively, we compared their rheobases and chronaxies. For a
constant-current pulse, the relationship between exponential chronaxie
and cell-membrane time constant is given by
tc=
m ln
2=0.693
m.30 Note that this relationship
does not apply to the hyperbolic chronaxie and that the chronaxie is
not an intrinsic parameter of the exponential model. Unless
specified as the exponential chronaxie, we use the term chronaxie in
its common usage, given by the hyperbolic model. Leading-edge-current
data were fit to a composite curve to determine leading-edge-current
m. The left side of this curve corresponding to
t<topt was Equation 4a
. The right side had a constant
current for t>topt. Because of the possibility that
time-dependent processes might falsely elevate the leading-edge-current
DFT for long pulses,12 31 32 these curves were fit with
(four points) and without (three points) the 15-ms data point. The
three-point fits are used unless specifically indicated. Calculated
time constants were then applied to the corresponding models to
estimate optimal waveform parameters for each patient.
|
Retrospective Validation of Models
Model predictions were validated by retrospective comparison
with results from clinical crossover studies of small-capacitor and
standard-capacitor waveforms. We first performed a literature search
for clinical crossover studies that compared DFTs for a standard 120-
to 125-µF capacitor and a smaller capacitor. Studies were excluded if
they did not include information required for model predictions, such
as pathway resistance. When clinical data were available for both
monophasic and biphasic waveforms, we used monophasic data. When only
biphasic waveforms were used, we compared their predicted
performance on the basis of the first phase. Waveforms with
equal tilt were compared for each capacitance. The ratio of
stored-energy DFTs of the experimental waveform to the stored-energy
DFT of the standard waveform±SEM was determined for each study. To
determine the corresponding predicted ratio, we first calculated the
ratio of the lowest DFT for an arbitrary capacitance to the lowest DFT
for the optimal capacitance for each model. This permitted a prediction
of the expected ratio of stored-energy DFTs for the two capacitance
values in each clinical study.
Statistical Analysis
We assessed the effect of pulse duration on DFT50
using one-factor repeated-measures ANOVA with pulse duration as the
factor. Post hoc analysis was performed by Scheffé's
test. The relationship between electrode configuration (epicardial or
transvenous) and each time constant was assessed by the unpaired
t test. The relationship between pathway resistance and each
time constant was assessed by linear regression. In this
analysis, resistance was the average value for all four pulse
widths. We used SuperANOVA 1.11 for the Macintosh (Abacus Concepts) for
ANOVA calculations. Goodness of fit was compared for different models
by the paired t test. The relationship between observed DFT
ratios in previous clinical studies and predicted ratios was assessed
by linear regression. Data are presented as mean±SD. When
multiple comparisons were performed, we required a value of
P<.05 divided by the number of comparisons. Basic
statistics were calculated with the paired two-tailed t
test.
| Results |
|---|
|
|
|---|
|
|
|
Average-Current Models
Average-current DFT50 decreased monotonically as pulse
width increased. Fig 3
shows that both average-current models provide a
good fit to the data for transvenous and epicardial pathways. For the
group as a whole, r2=.97±.02 for the hyperbolic
model and r2=.94±.04 for the exponential model.
The curves diverge at very long and very short pulse widths. The
exponential curves lie above the corresponding hyperbolic curves for
very short pulses. As pulse width increases into the clinical range,
the negative slope of the exponential curve is steeper, and it falls
below the hyperbolic curve. It then makes a sharper bend, crossing the
hyperbolic curve to reach a higher rheobase (epicardial: 3.2±0.4
versus 2.4±0.3 A, P<.001; transvenous: 4.8±0.9 versus
3.8±0.7 A, P<.001). This difference in curve shapes is
reflected in the lower exponential chronaxie for both epicardial and
transvenous pathways (epicardial: 2.7±0.3 versus 4.7±0.8 ms,
P<.001; transvenous: 2.2±0.4 versus 3.5±0.5 ms;
P<.001).
Leading-Edge-Current Model
The leading-edge-current DFT50 is highest for 1.5-ms
pulses for both transvenous and epicardial pathways. The transvenous
curve reaches a minimum at 7.5 ms and is unchanged at 15 ms. The
epicardial curve has a minimum at 3.0 ms. The epicardial
DFT50 is
15% higher at 15 ms than at 3 ms or 7.5 ms.
These differences are significant by the paired t test
(P<.01) but not by Scheffé's test (3.0 versus 15 ms:
P=.24; 7.5 versus 15 ms: P=.30). The fit to the
data (r2=.88±.14 for three points,
r2=.79±.24 for four points) is not as close as
that for the hyperbolic average-current model (P<.005) or
exponential average-current model (P<.03). When all four
data points were used for curve fitting, the value of
m
was lower for epicardial pathways (2.4±0.8 ms, P=.003) but
unchanged for transvenous pathways (2.3±0.4 ms, P=.86).
Table 2
shows that
m is lower for the
leading-edge-current model than the average-current model for both
epicardial pathways (P<.001) and transvenous pathways
(P<.001).
Correlation of Electrode Configuration, Pathway Resistance, and DFT
With Model Parameters
Table 2
shows that time parameters for all models were
greater for epicardial pathways than for transvenous pathways:
tc (P<.001), average-current
m
(P<.001), and leading-edge-current
m
(P=.06). For the group as a whole, there was a significant
inverse correlation between resistance and the average-current time
parameters (tc: r2=.48,
P<.001; average-current
m:
r2=.56, P<.001) but not
leading-edge-current
m: r2=.04,
P=.34. This correlation between resistance and the
average-current time parameters was significant for
transvenous pathways (tc=6.531-0.058xR:
r2=.57, P=.001; average-current
m=6.457-0.061xR:
r2=.72, P<.001) but not
for epicardial pathways (tc:
r2=.001, P=.94; average-current
m: r2=.01, P=.73).
There was no correlation between current, voltage, or stored-energy DFT and time parameter for any model. For example, r2 for leading-edge-current DFT varied from.004 to.12 for transvenous pathways (P=.21 to.83) and from.02 to.03 for epicardial pathways (P=.64 to.69).
Predicted Optimal Waveforms
Table 2
shows predicted optimal ICD waveforms. For epicardial
pathways, the two average-current models predict similar optimal
s, whereas the leading-edge-current model predicts a
lower value (P<.001). For transvenous pathways, optimal
s was greatest for the exponential average-current
model, intermediate for the hyperbolic average-current model, and
lowest for the leading-edge-current model. All pairwise differences
were significant at the level of P<.001. The
leading-edge-current model predicts shorter optimal durations for phase
1 than the average-current models (P<.001) and for phase 2
than the hybrid biphasic models (P<.001).
Penalty for Suboptimal Capacitance
Fig 4
shows the predicted effect on
stored-energy DFT of varying capacitance (or
s) for
transvenous pathways, provided that the best waveform is used for each
capacitance. The curves show each model's predicted "stored-energy
penalty" for suboptimal capacitance. They have a steep descending
limb for lower-than-optimal capacitance, a relatively flat valley with
a nadir at the optimal capacitance, and a gradually sloping ascending
limb for higher values. The range of penalties is 9% to 21%, 2% to
6%, 8% to 18%, and 20% to 45% for capacitance values that are 0.5,
1.5, 2, and 3 times optimal, respectively. The penalty for
underestimating optimal capacitance by 50% approximates that for
overestimating it by 100%. Expressed as multiples of optimal
capacitance, the stored-energy penalty over this range is approximately
twice as high for the exponential average-current model as either the
hyperbolic average-current or leading-edge-current model. From a
different perspective, the predicted reduction in DFT achieved by
optimizing capacitance for current ICDs ranges from 14% to 21% for a
150-µF capacitor and 5% to 10% for a 100-µF capacitor. The
predicted reduction is greatest for the leading-edge-current model,
which has the lowest optimal capacitance.
|
Comparison of Model Predictions With Results of Experimental
Studies
We identified five clinical studies that met the criteria for
retrospective analysis.13 14 15 18 19 Table 3
shows observed and predicted DFT
ratios. Most studies used biphasic waveforms. All three models provide
reasonable agreement with observed results for transvenous pathways
with experimental capacitance values of 60 to 90 µF. The
leading-edge-current model overestimates the performance of
60-µF waveforms for epicardial pathways. The correlation coefficients
are comparable for each model: hyperbolic average-current model
r2=.80, P=.0011; exponential
average-current model r2=.75,
P=.0027; leading-edge-current model
r2=.84, P=.0005. If the intercept is
required to be zero, the correlation coefficients are higher:
hyperbolic average-current model r2=.98;
exponential average-current model r2=.99;
leading-edge-current model r2=.98
(P<.001 for each model). The slope of the zero-intercept
regression line (observed DFT ratio/predicted DFT ratio) was 1.04 for
the hyperbolic average-current model, 1.04 for the exponential
average-current model, and 1.09 for the leading-edge-current model.
Thus, all three models underestimated the effect of capacitance on DFT,
and the degree of underestimation was greatest for the
leading-edge-current model.
|
| Discussion |
|---|
|
|
|---|
New Findings Regarding Models of Defibrillation
There are three secondary findings.
1. The hyperbolic and exponential average-current models predict similar optimal waveforms. The exponential strength-duration curve has a higher rheobase and lower chronaxie.
2. Average-current models predict higher values of optimal
s and capacitance than the leading-edge-current model.
Average-current and leading-edge-current models are based on different
fundamental assumptions. The leading-edge-current model explicitly
assumes that the response to a defibrillating pulse depends only on the
instantaneous peak value of membrane voltage. In contrast, the
average-current models implicitly assume that the cell membrane
responds to the average value of the applied field over the pulse. To
compare these two methods by use of parallel exponential models, we
used a new average-current model with an exponential formulation rather
than a hyperbolic one.
3. Time parameters for lower-resistance epicardial
pathways are greater than those for higher-resistance transvenous
pathways. We found an inverse correlation between pathway resistance
(or
s) and both average-current time
parameters (tc and
m) for the
group as a whole and for transvenous pathways. The minimal overlap in
resistances for epicardial and transvenous pathways precluded an
analysis to determine whether electrode configuration had an
effect on time parameters independent of the effect of
resistance. Further, because we studied acute transvenous electrodes
and chronic epicardial electrodes, we do not know whether this
difference contributed to the observed correlation between electrode
configuration and defibrillation time parameters. This
correlation was not present for leading-edge-current
m.
Previous Estimates of Defibrillation Time Parameters
Estimates of tc for the hyperbolic average-current
model have used average values from several studies in animals: 2.0
ms,1 2.0 ms,4 2.4 ms,3 and 2.7
ms.2 For transvenous defibrillation in dogs, Geddes and
Bourland7 used data from Wessale et al33 to
calculate a leading-edge-current
m of 1.2, 1.5, and 1.8
ms for trapezoidal pulses with tilt of 50% to 80%.
Cleland6 determined values for epicardial
leading-edge-current
m of 1.5 ms based on data from Tang
et al34 and 1.3, 1.4, and 1.7 ms using three different
sets of data from Walker et al.35 Walcott et
al4 used the equation
tc=0.693
m30 to calculate a
value of 2.8 ms for leading-edge-current
m from reported
values of hyperbolic average-current tc. However, this
equation relates exponential average-current tc to
exponential average-current
m, not hyperbolic
average-current tc to exponential leading-edge-current
m, as they apply it. Use of time parameters
derived by one method in a model based on another method may result in
significant error. Our results confirm the prediction of Block et
al36 that average-current defibrillation time
parameters are higher in humans than in animals. There is
only one previous estimate for tc in humans. In a
preliminary report of the effect of waveform tilt on DFT, Shorofsky et
al37 gave a value of 4.7 ms for pooled data from
epicardial, transvenous, and hybrid electrode systems. This value is
similar to tc for epicardial pathways in the present
study.
Fit of Human Defibrillation Data to Models
Both hyperbolic and exponential average-current models provided a
good fit over the range of pulse widths we tested. The better fit of
our data to the average-current models than the leading-edge-current
model may be due in part to fewer data points on the descending limb of
the leading-edge-current curve or use of a composite curve-fitting
method for this model. More data points near the anticipated value of
m might have permitted a better fit. We cannot
distinguish limitations of curve fitting from limitations of the
model.
Comparison of Model Predictions to Experimental Studies of ICD
Capacitance
All three models predict that optimal capacitance is proportional
to the model's time parameter and inversely proportional
to pathway resistance. Experimental studies generally have found a
greater beneficial effect of small capacitance on DFT in animals than
in humans. In dogs and pigs, smaller 40- to 90-µF capacitors have
performed better than conventional 120- to 140-µF
capacitors.8 9 10 11 12 In humans, however, 60- to 90-µF
capacitors have shown substantial benefit only for high-resistance
transvenous pathways.14 18 They have shown
modest13 or no14 15 16 17 18 19 benefit for the majority
of current clinical transvenous pathways, and 60-µF capacitors have
underperformed 120-µF capacitors for epicardial
pathways.18
The shape of the predicted "stored-energy DFT penalty" curves
provides a possible explanation for the differential effect of
capacitance on DFT in animals and humans. For example, the predicted
optimal capacitance in both animal and human transvenous pathways is
substantially less than values in ICDs, 32 µF1 2 versus
57 µF for the hyperbolic average-current model and 30
µF6 7 versus 45 µF for the leading-edge-current model.
However, the corresponding predicted DFT penalties for a 120-µF
capacitor are 29% and 37% in animals but only 8% and 16% in humans.
Cleland6 shows a similar DFT penalty curve based on animal
data and the leading-edge-current model in Fig 7b of that article.
Interpolation of this graph gives the penalty for a 120-µF capacitor
as
36%. Our results thus provide a conceptual basis for the
observations that defibrillation-waveform studies in animals cannot be
applied directly to humans. However, if animal time
parameters are known, the human time parameters
determined in this study may be applied to defibrillation models to
estimate corresponding results in humans.
Retrospective comparison of model predictions with results of clinical crossover trials shows that all three models provide reasonable agreement with observed results for experimental capacitance values of 60 to 90 µF. All models underestimated the observed effect of capacitance on DFT. However, model predictions were for the best waveforms with each capacitance, whereas the tested waveforms were not in general optimal for each capacitance. For biphasic waveforms, the degree to which the second phase improved defibrillation efficacy might vary for different durations of each phase.3 12
Implications for Design of ICDs
The stored-energy penalty curves have important implications for
design of ICDs because stored energy is a critical determinant of the
pulse generator size38 and current capacitor technology
limits maximum voltage to
750 V. Suppose, for example, that the
population DFT for an ICD with a 100-µF biphasic waveform is 10±4 J
for a 50-
transvenous pathway.13 39 40 Because current
technology limits the maximum voltage to
750 V, the maximum output
of 28.1 J exceeds the mean DFT+2 SD by 10 J. Using a near-optimal
60-µF waveform will decrease the mean DFT to 9.0 to 9.5±4 J.
However, the maximum 16.9-J output of the 60-µF ICD is 0.1 to 0.6 J
less than the mean DFT+2 SD. Newer bidirectional transvenous pathways
that include both a superior vena cava and active-can electrode have
resistances of 30 to 40
.17 40 41 For a 35-
pathway,
the predicted optimal capacitance is 65 to 105 µF, and the
stored-energy penalty for a 100-µF waveform is 0% to 3%. These
considerations suggest that for biphasic waveform defibrillation with
current electrode configurations, the penalty incurred by use of
100-µF capacitors in the present generation of ICDs is a small
and appropriate price to ensure reliable defibrillation of the vast
majority of patients. However, the penalty for 130- to 150-µF
capacitors used in earlier ICDs is unnecessary.
Limitations
A major limitation is that current models of defibrillation are
considered to be first-order approximations. However, the measured time
parameters for human defibrillation can be applied to
future models. A second major limitation related to curve fitting for
the leading-edge-current model has been discussed. A third major
limitation is the accuracy with which the DFT can be determined at four
different pulse widths in humans. The number of
fibrillation-defibrillation episodes in this study approaches a prudent
maximum for clinical protocols. We wish to emphasize three other
limitations: (1)
m derived from the exponential
strength-duration curves corresponds to a true cell-membrane time
constant only in a simple resistor-capacitor model of the cell
membrane.24 25 (2) The patient population may not be
representative of ICD recipients in general. Because of
the number of fibrillation-defibrillation episodes required, we
excluded the sickest patients. We also excluded transvenous patients
whose DFTs at a pulse duration of 1.5 ms had a high probability of
exceeding the maximum output of the defibrillator. Despite this,
determination of the transvenous DFT50 at 1.5 ms required
shock strengths >800 V (38 J) in 6 patients (40%) and 900 V (49 J) in
2 patients (13%). (3) This study used a single value of capacitance.
We do not know how variations in
s caused by changes in
capacitance might have affected our results.
Conclusions
When human time parameters are applied to models of
defibrillation, they provide a good estimate of the results of clinical
studies of ICD capacitance in the range 60 to 125 µF and predict the
better performance of small-capacitor waveforms in animals than
in humans. Use of 90- to100-µF capacitors realizes most of the
predicted reduction in DFT that can be achieved by optimizing
capacitance while maintaining a sufficient safety margin for current
transvenous electrode configurations.
| Selected Abbreviations and Acronyms |
|---|
|
| Acknowledgments |
|---|
| Appendix 1 |
|---|
|
|
|---|
![]() | (1) |
![]() | (2) |
Exponential Leading-Edge-Current Model
This model is derived by solving Blair's relation of Equation 1
for a capacitively discharged shock pulse described by
![]() | (3) |
![]() | (4A) |
![]() | (4B) |
s. For longer pulses,
I(t) remains constant at the value given by Equation 5
![]() | (5) |
s
. Intuitively, this can be appreciated by noting
that a truncated-exponential waveform approaches a constant-current
pulse as
s
.
Average-Current Hyperbolic Model
The empirical strength-duration relation is
![]() | (4) |
s does not appear explicitly as a term in the
average-current hyperbolic model. Instead, it is implicit in the
expression for average current of a truncated exponential pulse in
Equation 7
![]() | (6) |
Effective Current
Kroll developed the concept of effective current
(Ie) to derive the optimal pulse width and
s (or capacitance) for the hyperbolic average-current
model. For a shock pulse, Ie is defined as
"the rheobase requirement that it can
satisfy."2
![]() | (7) |
s in two
steps. First, he found the pulse width that maximized
Ie for a given
s. Then, using
this optimal pulse width, he determined the value of
s
that maximized Ie for a fixed stored energy
(Es).
Exponential Average-Current Model
This model is based on the strength-duration relation in Equation 2
if the constant current (I) is replaced by the average
current over the pulse (Iavg).
![]() | (8) |
s is implicit in the expression for average current. The
optimal value of pulse width and
s are derived by use of
the effective-current method in analogy to Kroll's approach for the
hyperbolic average-current model. We insert the expression for average
current of a truncated exponential pulse from Equation 7
![]() | (9) |
s by
Kroll's method for the hyperbolic average-current model. First, the
pulse width that maximizes Ie for a given
s is determined by differentiating Equation 9
m. Then this optimal pulse width is substituted
into Equation 9
s for a fixed
stored energy (Es) and solve for its zero. We find that
s=
m maximizes Ie
with respect to
s for a fixed value of
Es. Received April 3, 1997; revision received June 16, 1997; accepted June 19, 1997.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. Qu, L. Li, V. P. Nikolski, V. Sharma, and I. R. Efimov Mechanisms of superiority of ascending ramp waveforms: new insights into mechanisms of shock-induced vulnerability and defibrillation Am J Physiol Heart Circ Physiol, August 1, 2005; 289(2): H569 - H577. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. A. Mowrey, Y. Cheng, P. J. Tchou, and I. R. Efimov Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform Europace, January 1, 2002; 4(1): 27 - 39. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||