From the Department of Cardiology, Skejby Sygehus, Aarhus University
Hospital, Aarhus, Denmark.
Correspondence to Henning Rud Andersen, MD, DMSc, Department of Cardiology, Skejby Sygehus, Aarhus University Hospital, 8200 Aarhus N, Denmark.
Methods and ResultsA total of 225 consecutive patients with sick
sinus syndrome and intact atrioventricular conduction
were randomized to either single-chamber atrial pacing (n=110) or
single-chamber ventricular pacing (n=115). Clinical
assessment included New York Heart Association classification,
medication, and M-mode echocardiography before
pacemaker implantation, after 3 months, and subsequently once every
year. At long-term follow-up (mean, 5.5±2.4 years), NYHA class was
higher in the ventricular group than in the atrial group
(NYHA class I/II/III/IV: 65/44/4/0 versus 84/22/2/1 patients,
P=.010). Increase in NYHA class during follow-up was
observed in 35 of 113 patients in the ventricular group
versus 10 of 109 in the atrial group (P<.0005).
Increase in dose of diuretics from randomization to last
follow-up was significantly higher in the ventricular group
than in the atrial group (21±49 versus 8±42 mg furosemide/d,
P=.033). The left ventricular fractional
shortening decreased significantly in the ventricular group
(from 0.36±0.12 to 0.31±0.08, P<.0005) but not in the
atrial group (from 0.35±0.13 to 0.33±0.09, P=.087).
The left atrial diameter increased significantly in both treatment
groups (ventricular group: from 34±7 to 41±7 mm,
P<.0005; atrial group: from 34±6 to 37±7 mm,
P=.002), but the increase was significantly higher in
the ventricular group than in the atrial group
(P<.0005).
ConclusionsDuring long-term follow-up, ventricular
pacing is associated with a higher incidence of congestive heart
failure and consumption of diuretics than atrial pacing. This
is accompanied by a decrease in left ventricular fractional
shortening and an increased dilatation of the left atrium in the
ventricular paced patients.
Therefore, a long-term prospective trial with sequential
echocardiographic evaluation was performed to study the
evolution in congestive heart failure and elucidate the left atrial and
left ventricular dimensions and the left
ventricular function (left ventricular
fractional shortening) in patients with sick sinus syndrome randomized
to atrial or ventricular pacing.
Termination of Study
Heart Failure
Mortality as the Result of Congestive Heart Failure
Atrial Fibrillation
Echocardiography
Study Population
Pacemaker Implantation
Retrograde atrioventricular conduction was assessed
only in patients randomized to ventricular pacing, as
previously described in detail.13
Ethics
Analysis
Of the 110 patients randomized to atrial pacing, 6 had
ventricular leads at primary implantation. Thirty-five
patients (32%) had rate-adaptive pacemakers. During follow-up, another
5 patients had ventricular leads, and in 5 patients the
pacemaker was upgraded to a dual-chamber system (in 4 cases due to AV
block). Ninety-four patients (85%) of the 110 patients randomized to
atrial pacing were treated as randomized during follow-up.
All patients randomized to ventricular pacing were
discharged from the hospital with ventricular pacing.
Sixteen (14%) patients had rate-adaptive pacemakers. During follow-up,
1 patient had the pacing system changed to atrial pacing; in another 3
patients, upgrading to a dual-chamber system was necessary, and 2
patients had the pacemaker system explanted. One hundred nine patients
(95%) were treated as randomized during follow-up.
During follow-up, the mean programmed pacemaker-rate varied between 66
and 70 bpm in the atrial group and between 62 and 63 bpm in the
ventricular group (except at 7 and 8 years follow-up, in
which the mean rate in the ventricular group was 67 and 69
bpm, respectively). Medical treatment, including digoxin, was similar
during follow-up in the two groups. The only exception was dosage of
diuretics, which changed significantly (see below).
Heart Failure
During follow-up, the proportion of patients suffering from dyspnea
increased from 20% to 36% in the ventricular group,
whereas it was always lower than 20% in the atrial group (Fig 2
A total of 7 patients died from congestive heart failure in the
ventricular group versus 3 in the atrial group.
Kaplan-Meier plots are shown in Fig 4
Atrial Fibrillation
Echocardiographic Findings
Left Atrial Diameter
Left Ventricular End-systolic Diameter
Left Ventricular End-diastolic Diameter
Left Ventricular Fractional Shortening
Blood Pressure
Retrograde Conduction
Heart Failure
The mechanisms leading to heart failure during ventricular
pacing are not known, but changes in myocardial blood flow causing
functional ischemia,6 9 altered left
myocardial thickness,8 and myocardial cellular
changes7 21 have been demonstrated during chronic
pacing in the right ventricular apex.
In the current study, mortality as the result of congestive heart
failure was higher in the ventricular than in the atrial
group, but the number of congestive heart failure deaths was low, and
the difference was not statistically significant. This is in accordance
with observational findings. 2 3
Left Atrial Diameter
Left Ventricular Dimensions and LVFS
In one previous observational study, no difference in LVESD was found
after 5.3 years VVI pacing as compared with AAI/DDD
pacing.11 Similarly, in two prospective
short-term follow-up studies, no difference in LVESD was found during
(1) 3 weeks of VVI pacing as compared with VDD
pacing10 and (2) 6 months of VVIR pacing
comparing preoperative and follow-up values.12
These findings are in contrast with the present results.
In the observational study by Mohan et al,11
LVEDD was higher after 5.3 years VVI pacing than after AAI/DDD pacing,
whereas in the two prospective short-term follow-up studies, LVEDD
decreased during 3 weeks of VVI pacing as compared with VDD
pacing10 and during 6 months of VVIR
pacing,12 respectively. These conflicting results
are all in contrast to the present findings and are probably caused
by the small number of patients and echocardiographic
examinations included in these studies as compared with the current
study.
Blood Pressure
Retrograde Conduction
Study Limitations
The echocardiographic measurements were done by many
different physicians otherwise not involved in the study. Although the
physicians were all experienced in
echocardiography, the many investigators might
increase the dispersion of these measurements due to a larger
interobserver variation.14 On the other hand, no
single physician (biased or unbiased) had sufficient influence to
change the overall results in any direction. Only M-mode
echocardiography was performed. Two-dimensional
echocardiography would probably have added more
information regarding left atrial and left ventricular
volumes and function. The subgroup analyses performed were not
anticipated at the time of design of the study but were decided post
hoc and therefore the study was not necessarily powered to do these
subgroup analyses. This should be taken into account when the
results of the subgroup analyses are interpreted.
In the current study, the pacemaker event counters were not checked
regularly. Thus it was not possible to correlate number of
ventricular or atrial stimulated beats versus sensed beats
to the changes in left atrial and left ventricular
dimensions, LVFS, or occurrence of heart failure.
Conclusions
Received August 4, 1997;
revision received October 21, 1997;
accepted November 14, 1997.
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Heart Failure and Echocardiographic Changes During Long-term Follow-up of Patients With Sick Sinus Syndrome Randomized to Single-Chamber Atrial or Ventricular Pacing
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundIn patients with sick
sinus syndrome, choice of pacing mode has been implicated in the
development of congestive heart failure.
Key Words: sick sinus syndrome pacing echocardiography heart failure ventricles
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
In several
observational studies of patients with sick sinus syndrome,
single-chamber ventricular pacing has been associated with
a higher incidence of congestive heart failure than single-chamber
atrial pacing.1 2 3 4 5 Furthermore, experimental
studies have demonstrated that long-term ventricular pacing
is associated with both structural and functional changes in the
ventricular myocardium, which could lead to
congestive heart failure.6 7 8 9 However,
echocardiographic studies in ventricular
paced patients have failed to demonstrate significant changes in left
ventricular systolic dimensions and
function,10 11 whereas left
ventricular diastolic dimensions have been
found to increase in one study11 and to decrease
in two other studies.10 12 In the first
prospective, randomized trial comparing atrial and
ventricular pacing in patients with sick sinus syndrome,
the left atrial diameter increased significantly in the
ventricular paced patients, whereas the trial did not show
any significant difference in heart failure between the two groups
after short-term follow-up.13 Thus conflicting
results exist about development of heart failure and
echocardiographic parameters in
ventricular paced patients with sick sinus syndrome,
possibly due to lack of long-term randomized trials including
sequential echocardiographic evaluation during
follow-up.
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
Protocol
The trial was conducted as a one-center study at Skejby
University Hospital. All patients who were referred to treatment with
their first pacemaker during the recruitment period from May 15, 1988,
to December 31, 1991, were evaluated for randomization. The patients
were asked to participate in the trial if the inclusion criteria
(symptomatic bradycardia <50 bpm or
symptomatic QRS pauses >2 seconds) and none of the
exclusion criteria (Table 1
) were
met.13 After giving informed consent, patients
were randomized to treatment with single-chamber atrial or
ventricular pacing stratified in blocks of 10-year age
groups. Medical history, physical examination, and
echocardiography were obtained before implantation.
Follow-up visits were after 3 months, 12 months, and subsequently once
per year. The follow-up visit included physical examination, 12-lead
ECG recording, pacemaker checkup, and
echocardiography. Follow-up evaluation was not
blinded.
View this table:
[in a new window]
Table 1. Study Population
In 1995 it was decided that all patients included should be
followed for at least 5 years before final data analysis, which
was determined to be December 31, 1996.
Heart failure was classified according to New York Heart
Association criteria and quantitated by the daily dose of
diuretics (converted to milligrams of furosemide per day)
before pacemaker implantation and at each follow-up visit. The
occurrence of crural edema and dyspnea (patients were asked if they had
shortness of breath during normal daily activities) was
recorded.
The cause of death was obtained by interviewing the doctors
caring for the patients and by review of hospital files and necropsy
reports. Only patients dying from progressive congestive heart failure
and without signs of acute myocardial infarction, pulmonary
embolism, or cerebral stroke were classified as dying from congestive
heart failure. Mortality as the result of congestive heart failure did
not include patients dying suddenly.
Atrial fibrillation was diagnosed by 12-lead ECG at the
follow-up visits (not from ECGs taken at any other time). Atrial
fibrillation was categorized as chronic if recorded at two
consecutive follow-up visits and no sinus rhythm was observed
subsequently.
Left atrial diameter and left ventricular
end-systolic (LVESD) and end-diastolic diameters
(LVEDD) were determined by M-mode echocardiography
in accordance with the recommendations of the American Society of
Echocardiography.14 Left
ventricular dimensions were obtained at the level of the
chordae. Measurements were done by the leading edge methodology. Left
ventricular fractional shortening (LVFS) was calculated as
(LVEDD-LVESD)/LVEDD.
During the recruitment period, 1052 patients (568 men, age
71±17 years) had their first pacemaker implanted; 827 were excluded
from randomization (Table 1
).13
Patients had an atrial pacing test during implantation at 100
and 120 bpm. Atrioventricular conduction (1:1) at 100
bpm was required for an atrial pacemaker to be implanted; otherwise,
the lead was implanted in the right ventricle. Patients randomized to
ventricular pacing always had a ventricular
lead implanted whatever the result of the atrial pacing test. Whether
to use a pacemaker with rate-adaptive modality was decided by the
physician in charge of the implantation from clinical criteria. The
programmed pacing rate after implantation and during follow-up was set
from clinical criteria.
The study was approved by the National Danish Ethical Committee
and was conducted in accordance with the rules of the Helsinki
Declaration. Informed consent was obtained from all patients before
inclusion in the study.
Power calculations were done before start of the study and have
been published previously.13 All statistical
analyses were done according to the intention-to-treat
principle. Continuous variables are expressed as mean±SD in text
and tables and as mean±SEM in the figures. Treatment groups were
compared by
2 test or Fisher's exact
test (two-tailed) for discrete variables and by two-tailed
Student's t test for continuous variables. Changes in
echocardiographic parameters within
treatment groups were analyzed by comparing mean values before
pacemaker implantation with mean values at last follow-up (last
ambulatory visit before end of study or death of patient), using the
paired t test. Intergroup statistics were done by comparing
mean change from pacemaker implantation to last follow-up between
treatment groups, using the Student's t test. To reduce the
number of statistical tests, intergroup comparisons were not done
repeatedly at several time points after implantation but only at last
follow-up or on change from implantation to last follow-up.
Kaplan-Meier plots were calculated for mortality as the result of
congestive heart failure and compared by log-rank test. Relative risk
and 95% confidence interval were calculated by univariate
Cox proportional hazards method. A value of P<.05 was
deemed significant. Unadjusted probability values are
presented.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
Patients
A total of 225 patients (142 women, 83 men; age 76±8 years;
range, 50 to 92 years) were randomized to atrial pacing (n=110) and
ventricular pacing (n=115), respectively. No patients were
lost to follow-up. One patient, who emigrated after the 1-year
follow-up visit, had subsequent follow-up by telephone interview, but
no echocardiographic measurements were accessible in
this patient. Mean follow-up was 5.5±2.4 years for the total study
population with similar follow-up in the atrial (5.7±2.3 years) and
ventricular (5.3±2.5 years) group. Baseline
characteristics of the two groups were similar except that more
patients in the atrial group were treated with digoxin (Table 2
).
View this table:
[in a new window]
Table 2. Baseline Characteristics
At randomization, there was no difference in the NYHA
classification and no difference in the use of diuretics
between the two groups (Table 2
). NYHA class at last follow-up (3
patients died before their first follow-up visit, leaving 222 patients)
was significantly higher in the ventricular group than in
the atrial group (NYHA class I/II/III/IV: 65/44/4/0 versus 84/22/2/1
patients, P=.010,
2 test).
Increase in NYHA class during follow-up was observed in 10 of 109
patients in the atrial group versus 35 of 113 patients in the
ventricular group (P<.0005,
2 test). Use of diuretics
increased in the ventricular group but not in the atrial
group during follow-up (Fig 1
). The
increase in dose of diuretics from randomization to last
follow-up was significantly higher in the ventricular group
than in the atrial group (21±49 versus 8±42 mg furosemide/d,
P=.033). At the 3- and 5-year follow-up visits, information
about use of diuretics was missing for 1 patient.

View larger version (20K):
[in a new window]
Figure 1. Mean dose of diuretics (converted to
milligrams of furosemide per day) during follow-up of patients
randomized to atrial (AAI) (n=110) or ventricular (VVI)
(n=115) pacing. Numbers below abscissa indicate number of patients in
the two treatment groups. See text for statistical comparison between
groups.
). There was no statistically
significant difference in occurrence of dyspnea at last follow-up
between the atrial group and the ventricular group (18%
versus 24%, P=.31,
2 test).
The occurrence of crural edema during follow-up was higher in the
ventricular group than in the atrial group, most markedly
after long-term follow-up (Fig 3
), but
there was no statistically significant difference in occurrence of
crural edema at last follow-up between the atrial group and the
ventricular group (14% versus 19%, P=.33,
2 test).

View larger version (19K):
[in a new window]
Figure 2. Proportion of patients with dyspnea during
follow-up of patients randomized to atrial (AAI) (n=110) or
ventricular (VVI) (n=115) pacing. Numbers below abscissa
indicate number of patients in the two treatment groups. See text for
statistical comparison between groups.

View larger version (19K):
[in a new window]
Figure 3. Proportion of patients with crural edema during
follow-up of patients randomized to atrial (AAI) (n=110) or
ventricular (VVI) (n=115) pacing. Numbers below abscissa
indicate number of patients in the two treatment groups. See text for
statistical comparison between groups.
.
The mortality as the result of congestive heart failure was not
significantly different between the two groups (relative risk for
atrial group, 0.41 [95% CI, 0.11 to 1.59], P=.18).

View larger version (17K):
[in a new window]
Figure 4. Mortality as the result of congestive heart
failure (CHF) during follow-up of patients randomized to atrial (AAI)
(n=110) or ventricular (VVI) (n=115) pacing. Numbers below
abscissa indicate number of patients at risk in the two treatment
groups. Log-rank test was used in the comparison between groups.
All patients had sinus rhythm at the time of randomization. In the
ventricular group, 40 patients (35%) had atrial
fibrillation at one or more follow-up visits versus 26 patients (24%)
in the atrial group (P=.066,
2 test). Twenty-one patients died
before the 1-year follow-up visit, leaving 204 patients (102 in each
group) at risk of developing chronic atrial fibrillation. In the
ventricular group, 22 patients (22%) developed chronic
atrial fibrillation versus 9 patients (9%) in the atrial group
(P=.011,
2 test).
All patients had echocardiographic examination
before implantation as a part of the trial. Initially,
echocardiography at follow-up was not done
routinely in all patients, but from February 1990 to December 1996, all
patients had echocardiographic examination at each
follow-up visit as a part of the trial. A total of 625
echocardiographic examinations were done in the atrial
group and 630 in the ventricular group. Only in a low
number of cases were echocardiographic measurements not
possible because of poor image quality. In some patients, left atrial
diameter but not left ventricular dimensions could be
assessed, and sometimes vice versa. Consequently, the number of
patients with measurements of atrial and ventricular
dimensions is not always the same. The numbers of patients investigated
are reported in Table 3
and below
abscissa in Figs 5 to 8![]()
![]()
![]()
.
View this table:
[in a new window]
Table 3. Echocardiographic Changes During
Follow-up

View larger version (22K):
[in a new window]
Figure 5. Left atrial diameter measured by M-mode
echocardiography during follow-up of patients
randomized to atrial (AAI) (n=110) or ventricular (VVI)
(n=115) pacing. Numbers below abscissa indicate number of patients
examined in the two treatment groups. See text for statistical
comparison between groups.

View larger version (21K):
[in a new window]
Figure 6. Left ventricular end-systolic
diameter (LVESD) measured by M-mode
echocardiography during follow-up of patients
randomized to atrial (AAI) (n=110) or ventricular (VVI)
(n=115) pacing. Numbers below abscissa indicate number of patients
examined in the two treatment groups. See text for statistical
comparison between groups.

View larger version (22K):
[in a new window]
Figure 7. Left ventricular
end-diastolic diameter (LVEDD) measured by M-mode
echocardiography during follow-up of patients
randomized to atrial (AAI) (n=110) or ventricular (VVI)
(n=115) pacing. Numbers below abscissa indicate number of patients
examined in the two treatment groups. See text for statistical
comparison between groups.

View larger version (21K):
[in a new window]
Figure 8. Left ventricular fractional shortening
(LVFS) during follow-up of patients randomized to atrial (AAI) (n=110)
or ventricular (VVI) (n=115) pacing. Numbers below abscissa
indicate number of patients examined in the two treatment groups. See
text for statistical comparison between groups.
During follow-up, there was an increase in the left atrial
diameter in both treatment groups, most markedly in the
ventricular group (Fig 5
). Statistically, the increase was
significant in both the atrial group (P=.002) and the
ventricular group (P<.0005), and the increase
was significantly higher in the ventricular group than in
the atrial group (P<.0005) (Table 3
). Because atrial
enlargement may be a consequence of atrial fibrillation, subgroup
analyses were performed in (1) patients who were without atrial
fibrillation at any follow-up visit, (2) patients without atrial
fibrillation at the time of the last echocardiographic
examination, (3) patients without chronic atrial fibrillation, and (4)
patients with chronic atrial fibrillation (Table 3
). In all these
subgroups the left atrial diameter increased more in the
ventricular group than in the atrial group, but in the
small group of patients with chronic atrial fibrillation, the
difference was not statistically significant (Table 3
). In subgroup 1
(patients without atrial fibrillation at any follow-up visit) and
subgroup 2 (patients without atrial fibrillation at the time of the
last echocardiographic examination), follow-up time was
slightly longer in the atrial group than in the ventricular
group (subgroup 1: 5.5±2.3 versus 4.6±2.5 years, P=.013,
t test; and subgroup 2: 5.5±2.3 versus 4.8±2.4 years,
P=.035, t test), whereas there was no difference
in follow-up time in subgroups 3 and 4.
Graphically, the LVESD appeared to increase more in the
ventricular group than in the atrial group (Fig 6
). Compared with preoperative values,
the increase was significant in the ventricular group
(P=.005) but not in the atrial group (P=.40). The
increase in LVESD in the ventricular group was highest in
the group of patients with chronic atrial fibrillation (Table 3
). In
patients without atrial fibrillation at last
echocardiographic measurement and in patients without
chronic atrial fibrillation, there was no statistically significant
increase in LVESD in the ventricular group (Table 3
).
Graphically, the LVEDD appeared to be higher in the
ventricular group than in the atrial group (Fig 7
), but statistically, there was no
significant change in LVEDD from pacemaker implantation to last
follow-up in either the atrial group or the ventricular
group (Table 3
).
In the plot of the LVFS during follow-up, the LVFS apparently
decreased more in the ventricular group than in the atrial
group (Fig 8
). Comparing the mean values
before pacemaker implantation with the mean values at last follow-up,
we found that the LVFS decreased significantly in the
ventricular group (P<.0005) but not in the
atrial group (P=.087). There was no statistically
significant difference in the decrease in LVFS between the two groups
(P=.097). In all subgroup analyses, the LVFS
decreased significantly in the ventricular group but not in
the atrial group-except in the group of patients without atrial
fibrillation at last follow-up, in which a minor decrease was seen also
in the atrial group (Table 3
).
At the time of randomization there was no difference in either
systolic or diastolic blood pressure between the
atrial and the ventricular group (Table 2
). Blood pressure
during follow-up is shown in Fig 9
. At
last follow-up, there was no difference in systolic or
diastolic blood pressure between the atrial and
ventricular groups (151±26 mm Hg in the
ventricular group versus 151±23 mm Hg in the atrial
group, P=.93, and 83±14 mm Hg in the
ventricular group versus 84±12 mm Hg in the atrial
group, P=.92).

View larger version (22K):
[in a new window]
Figure 9. Mean systolic blood pressure (SBP) and
diastolic blood pressure (DBP) during follow-up of patients
randomized to atrial (AAI) (n=110) or ventricular (VVI)
(n=115) pacing. Numbers below abscissa indicate number of patients
examined in the two treatment groups. See text for statistical
comparison between groups.
Retrograde conduction was present in 63 patients in the
ventricular group. Left atrial diameter increased both in
the group with retrograde conduction (from 35±7 to 41±7 mm,
P<.0005) and in the group without retrograde conduction
(from 34±7 to 40±7 mm, P<.0005), but there was no
difference comparing the two groups (P=.94). LVESD increased
significantly in the group with retrograde conduction (from 31±8 to
34±7 mm, P=.012), whereas the increase was
insignificant in the group without retrograde conduction (from 31.6±9
to 32.3±7 mm, P=.15), but there was no significant
difference when comparing the two groups (P=.53). There was
no change in LVEDD in either group. The diastolic blood
pressure at last follow-up was significantly higher in the group of
patients with retrograde conduction than in the group without
retrograde conduction (86±15 versus 80±11 mm Hg,
P=.026), whereas there was no difference in systolic
blood pressure (151±29 versus 150±21 mm Hg, P=.74).
There was no difference in NYHA class at last follow-up between the
patients with retrograde conduction (NYHA I/II/III/IV: 37/23/1/0) and
the patients without retrograde conduction (NYHA I/II/III/IV:
28/21/3/0), P=.44,
2 test.
The proportion of patients with dyspnea at last follow-up was slightly
higher in the group with retrograde conduction (28%) than in the group
without retrograde conduction (19%), but the difference was not
statistically significant (P=.28). There was no difference
in proportion of patients with crural edema at last follow-up between
the groups with and without retrograde conduction (18% versus 19%,
P=.65).
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
The present study shows that after long-term follow-up of
patients with sick sinus syndrome treated with either
ventricular or atrial pacemakers, the incidence of
congestive heart failure and the consumption of diuretics is
higher in the ventricular paced patients than in the atrial
paced patients. Furthermore, the left atrial diameter increases and the
left ventricular fractional shortening decreases more
during ventricular pacing than during atrial pacing. These
findings indicate that long-term single-chamber ventricular
pacing is associated with a harmful effect on the heart compared with
single-chamber atrial pacing and in some cases leads to excessive
congestive heart failure.
The degree of heart failure estimated by NYHA class, use of
diuretics, and occurrence of dyspnea and crural edema were
higher in the ventricular group than in the atrial group
during long-term follow-up. This is in accordance with findings in
retrospective studies.1 2 3 4 5 In our previous
short-term follow-up study (mean follow-up, 3.3 years), there was no
significant difference in degree of heart failure between the two
groups.13 This indicates that although
ventricular pacing is tolerated for some years, long-term
ventricular pacing is associated with an increased risk of
heart failure. In the present study, ventricular pacing
was associated with a decrease in LVFS, which reflects reduced left
ventricular function. Deterioration of the left
ventricular diastolic function could also be
involved as found in other studies,6 15 16 17 but
left ventricular diastolic function has not
been assessed in the current study. In a large, retrospective study
performed by Sgarbossa et al,18
ventricular pacing was not found to be associated with an
increased risk of congestive heart failure compared with
physiological pacing. However, in the study by
Sgarbossa et al, ventricular pacing (112 patients) was not
compared with atrial pacing but with physiological
pacing (374 patients with dual-chamber pacemakers and only 19 patients
with atrial pacemakers). Thus in that study, most of the patients in
the physiological group were implanted with a
ventricular lead and stimulated very frequently in the
right ventricle.19 20 The ventricular
stimulation per se is potentially harmful to cardiac
function,6 7 8 9 and even dual-chamber pacing might
be associated with a similar risk of congestive heart failure as
observed in the current study for ventricular pacing due to
stimulation of the ventricular myocardium in
both pacing modes.6 9 Only single-chamber atrial
pacing is without any artificial stimulation of the ventricles.
Therefore, the conclusion reached in the retrospective study by
Sgarbossa et al18 that there is no difference in
incidence of congestive heart failure between ventricular
and physiological pacing (dual-chamber pacemakers)
may very well be correct. However, this conclusion regarding
physiological dual-chamber pacing cannot
automatically be extended to be valid also for atrial pacing.
The left atrial diameter increased in both treatment groups during
long-term follow-up, but the increase was significantly higher in the
ventricular group compared with the atrial group. This
confirms findings in our short-term follow-up
study13 and in two minor observational
studies.11 22 Atrial enlargement was not only a
consequence of atrial fibrillation, as has been reported in patients
without a pacemaker,23 24 because atrial
dilatation was seen also in patients without atrial fibrillation. The
cause of the marked additional left atrial dilatation during
ventricular pacing is not known. Stimulation of the right
ventricular apex during ventricular pacing
results in an altered activation sequence of the ventricles and
consequently a change in the ventricular mechanical
contraction pattern.6 17 25 26 This has been
found to be associated with a decrease in the left
ventricular systolic
function17 25 and diastolic
function6 and an increase in both the right
atrial pressure and the pulmonary capillary wedge
pressure,25 27 which might be involved in the
left atrial dilatation.28 The lack of
atrioventricular synchrony in single-chamber
ventricular pacing could be associated with an increased
risk of mitral
regurgitation,12 29 which also
leads to left atrial dilatation. In the current study we found no
evidence that the presence of retrograde
atrioventricular conduction was involved in the
pathogenesis of left atrial dilatation during ventricular
pacing. The increase in the left atrial diameter observed in both the
atrial and the ventricular group might be explained partly
by increasing age30 but could also
represent a feature in the natural evolution of the sick sinus
syndrome disease, which, in part, could explain the increased risk of
arterial thromboembolism and atrial fibrillation observed
in this disease.5 13 31 32 33 34
During long-term follow-up, we observed a decrease in LVFS in the
ventricular group, whereas there was no significant change
in the atrial group. The changes were associated with an increase in
mean LVESD in the ventricular group without
simultaneous changes in mean LVEDD. A decrease in LVFS
during ventricular pacing was also described in a
prospective study of 13 patients followed for 6
months12 and is in accordance with the acute
decrease in left ventricular systolic function
observed during ventricular (VVI) pacing in several
experimental studies.17 25 26 35 36 The decrease
in mean LVFS could be explained by the asynchronous
ventricular contraction caused by the altered activation
sequence of the ventricles due to stimulation of the right
ventricular apex. The decrease in LVFS is probably the
major cause of the increased risk of heart failure observed in the
ventricular group in the present study.
In the current study, there was no difference in systolic
or diastolic blood pressure between the two groups during
long-term follow-up. Thus blood pressure during long-term follow-up of
patients with sick sinus syndrome is unaffected by mode of pacing.
The presence of perioperative retrograde
conduction in the patients treated with ventricular pacing
was not associated with an increase in left atrial diameter or left
ventricular diameters, or a decrease in LVFS as compared
with the absence of retrograde conduction. Nor was retrograde
conduction associated with a lower blood pressure, occurrence of
dyspnea, or higher degree of heart failure, which could be indicative
for presence of the pacemaker syndrome.37
Previously, retrograde conduction during ventricular pacing
was believed to be associated with the pacemaker
syndrome37 38 39 and with the development of
congestive heart failure,40 but in the current
study we found no such associations. In accordance with our results,
one previous prospective, long-term follow-up study found no difference
in the incidence of congestive heart failure during VVI pacing with and
without retrograde conduction, respectively.41
Therefore, no clear evidence exists that retrograde conduction causes
heart failure. It might instead be the ventricular
stimulation per se that is most important.
The physical examination, NYHA classification, and M-mode
echocardiography at ambulatory follow-up was not
done blinded with regard to the randomization. This introduces a
possible observer bias in these parameters. The NYHA class
was difficult to define precisely in some of our patients, many of whom
were very old and immobile, but the significant higher dose of
diuretics in the ventricular group during long-term
follow-up indicates that ventricular pacing was really
associated with a higher incidence of heart failure.
During long-term follow-up, ventricular pacing is
associated with a higher incidence of congestive heart failure and
consumption of diuretics than atrial pacing. This is
accompanied by a decrease in left ventricular fractional
shortening and an increased dilatation of the left atrium in the
ventricular paced patients. Consequently, patients with
sick sinus syndrome and intact atrioventricular
conduction should be treated with atrial pacing instead of
ventricular pacing to avoid development or worsening of
congestive heart failure, decrease in left ventricular
fractional shortening, and increase in left atrial diameter.
![]()
Acknowledgments
This study was supported by grants from the Danish Heart
Foundation and from Sygekassernes Helsefond.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
This article has been cited by other articles:
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62. [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons Circulation, May 27, 2008; 117(21): e350 - e408. [Full Text] [PDF] |
||||
![]() |
A. Quesada, G. Botto, A. Erdogan, M. Kozak, P. Lercher, J. C. Nielsen, O. Piot, R. Ricci, C. Weiss, D. Becker, et al. Managed ventricular pacing vs. conventional dual-chamber pacing for elective replacements: the PreFER MVP study: clinical background, rationale, and design Europace, March 1, 2008; 10(3): 321 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Albertsen, J. C. Nielsen, S. H. Poulsen, P. T. Mortensen, A. K. Pedersen, P. S. Hansen, H. K. Jensen, and H. Egeblad Biventricular pacing preserves left ventricular performance in patients with high-grade atrio-ventricular block: a randomized comparison with DDD(R) pacing in 50 consecutive patients Europace, March 1, 2008; 10(3): 314 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. E. Albertsen, J. C. Nielsen, S. H. Poulsen, P. T. Mortensen, A. K. Pedersen, P. S. Hansen, H. K. Jensen, and H. Egeblad DDD(R)-pacing, but not AAI(R)-pacing induces left ventricular desynchronization in patients with sick sinus syndrome: tissue-Doppler and 3D echocardiographic evaluation in a randomized controlled comparison Europace, February 1, 2008; 10(2): 127 - 133. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. O. Sweeney, A. J. Bank, E. Nsah, M. Koullick, Q. C. Zeng, D. Hettrick, T. Sheldon, G. A. Lamas, and the Search AV Extension and Managed Ventricular Pa Minimizing Ventricular Pacing to Reduce Atrial Fibrillation in Sinus-Node Disease N. Engl. J. Med., September 6, 2007; 357(10): 1000 - 1008. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. O. Sweeney and A. S. Hellkamp Heart Failure During Cardiac Pacing Circulation, May 2, 2006; 113(17): 2082 - 2088. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. D.R. Thackray, K. Witte, J. Ghosh, N. Nikitin, A. Anderson, A. Rigby, K. Goode, A. L. Clark, and J. G.F. Cleland N-terminal brain natriuretic peptide as a screening tool for heart failure in the pacemaker population Eur. Heart J., February 2, 2006; 27(4): 447 - 453. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Dilaveris, A. Pantazis, G. Giannopoulos, A. Synetos, J. Gialafos, and C. Stefanadis Upgrade to biventricular pacing in patients with pacing-induced heart failure: can resynchronization do the trick? Europace, January 1, 2006; 8(5): 352 - 357. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Rinfret, D. J. Cohen, G. A. Lamas, K. E. Fleischmann, M. C. Weinstein, J. Orav, E. Schron, K. L. Lee, and L. Goldman Cost-Effectiveness of Dual-Chamber Pacing Compared With Ventricular Pacing for Sinus Node Dysfunction Circulation, January 18, 2005; 111(2): 165 - 172. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Vernooy, X. A.A.M. Verbeek, M. Peschar, H. J.G.M. Crijns, T. Arts, R. N.M. Cornelussen, and F. W. Prinzen Left bundle branch block induces ventricular remodelling and functional septal hypoperfusion Eur. Heart J., January 1, 2005; 26(1): 91 - 98. [Abstract] |