From the Departments of Cardiology and Medical Physics, Westmead
Hospital, Westmead, Sydney, Australia.
Correspondence to Prof D.L. Ross, Department of Cardiology, Westmead Hospital, Westmead, NSW, 2145, Australia.
Methods and ResultsNine female patients having radiofrequency
ablation for supraventricular tachycardia were
studied. The radiation dose was determined at 41 body sites in each
patient with the use of thermoluminescent dosimeters and was correlated
with that measured simultaneously with a Diamentor
dose-area product meter. The estimated mean organ doses (mGy) per
60 minutes of fluoroscopy were: lungs 30.8; bone marrow 4.3; left
breast 5.1; right breast 3.5; and thyroid 2.4. From the average organ
doses, the estimated mean total lifetime excess risk of a fatal
malignancy was 294 per million cases (0.03%) per 60 minutes of
fluoroscopy. The risk calculation from the Diamentor dose-area
product and thermoluminescent dosimeters were similar, suggesting
that radiation dose was measured accurately. The estimated risk of
radiation-induced malignancy increased with increasing body mass index
(P=0.03).
ConclusionsProlonged fluoroscopy during radiofrequency ablation
may potentially cause a small increase in the lifetime risk of fatal
malignancy, with lung malignancy being most likely. This risk is small
only with the use of techniques and x-ray equipment optimized to keep
radiation as low as possible. The risk is increased in obese patients.
The purpose of this study was to determine more accurately the
dose of radiation delivered to patients during radiofrequency ablation
for supraventricular tachycardia and assess the
possible risk of malignancy from the exposure. We appreciated that the
amount of radiation the patient is exposed to would depend on the
fluoroscopy equipment used and how carefully the operators strove
during the procedure to reduce radiation exposure to the patient. The
use of large image intensifier fields9 and
judicious use of collimation10 would obviously
reduce radiation exposure to the patient considerably. We aimed at the
outset to use techniques and x-ray equipment optimized for the lowest
possible radiation dose commensurate with acceptable image quality.
Measurement of Radiation Dose by Thermoluminescent
Dosimetry
Radiation to the breasts was measured with 9 TLDs (Figure 2
Measurement of Radiation Dose With Dose-Area Product Meter
Radiofrequency Ablation
A Toshiba single-phase, half-wave, rectified C-arm fluoroscopy system
(model KXO-650) was used to position the catheters. The x-ray tube was
located under the table and the image intensifier above the table. The
system used continuous fluoroscopy, not pulsed fluoroscopy.
The measured x-ray beam half-value layer was 5.95 mm aluminum at
96 kV and 6.2 mm aluminum at 105 kVp. This was equivalent to
8 mm aluminum total filtration, including the table filtration
effect, calculated for a single-phase unit with 12 degree target angle.
The image intensifier had a fixed grid with grid ratio of 12:1 and a
focal length of 90 cm. The image intensifier had 3 field sizes of 23,
17, and 11 cm. The 23-cm field was used in most cases. Kilovoltage was
determined by an automatic brightness control for manually chosen tube
currents (usually 0.5 to 1.0 mA, but occasionally up to 2 mA for very
obese patients). The tube current was kept at the lowest possible
setting that gave a clinically satisfactory image. Cinefluorography was
not used in any patient. Collimation was used routinely.
Positioning of the catheters at the right atrium, right ventricle, and
His bundle was performed with fluoroscopic imaging in the
postero-anterior position. Coronary sinus
catheterization was usually performed in the left
anterior oblique position. Once the tachycardia mechanism
was confirmed, a 7F quadripolar steerable ablation catheter with a 4-mm
distal electrode (Mansfield/Webster) was introduced through the right
femoral vein or artery, depending on the location of the abnormal
pathway targeted for ablation. Some right free wall accessory pathways
required insertion of the ablation catheter from the subclavian or
internal jugular veins. For positioning of the ablation catheter in
atrioventricular junctional reentry
tachycardia cases, fluoroscopic imaging was predominantly
in the right anterior oblique view. For right free wall accessory
pathways the left anterior oblique view was mainly used. For other
types of arrhythmias the positioning of the image intensifier
was variable.
Radiofrequency energy was delivered with a Zencor MF1 bipolar
coagulator. The energy was delivered in unipolar fashion between the
4-mm catheter-tip electrode and a Valleylab Poly Hesive return
electrode.
Calculation of Radiation Risks
The ICRP population-averaged fatal cancer probability coefficient of
0.05 Sv-1 was used. However, it would be more
accurate to use age-related coefficients for estimation of risk when
the age range is known to be small. For the purposes of comparison with
other studies in which the age range is variable, we have chosen to
use the population-averaged probability coefficients. We do, however,
realize that the relatively young age of the patients will increase the
risk of breast cancer in comparison to the population average figure
and also that thyroid cancer is twice as prevalent in women as in
men.
Calculation of Excess Risk From Absorbed Dose From TLD
Calculation of Excess Risks From Diamentor Dose
Calculation of Hereditary Effects
Statistical Analysis
There was a good correlation between the risk of malignancy calculated
with TLDs and that calculated with the Diamentor (Figure 3
Biological Effects of Radiation
In this study we estimated the stochastic somatic effects as well as
the hereditary effects of radiation to the patient. The overall
increase of fatal malignancy was found to be small (0.03% for 60
minutes of fluoroscopy). The most likely malignancy is lung, the organ
exposed to the maximum amount of radiation. Lung carcinoma is 20 times
more likely and leukemia is 1.7 times more likely than breast carcinoma
from 60 minutes of fluoroscopy. The possibility of fatal thyroid
carcinoma is only 0.0002% , since this is often a survivable disease.
In the calculations of risk, there was no consideration made of a
dose/dose rate effectiveness factor because of the low dose rate and
the relatively low total doses.
Severe radiation induced skin injuries to patients resulting from
prolonged, fluoroscopically-guided, invasive procedures such as
percutaneous transluminal coronary angioplasty
and radiofrequency catheter ablation have been
reported.21 Radiation induced skin injury did not
occur in any patient in our study.
The risk of severe hereditary effects was found to be less than 1 per
million cases for 60 minutes of fluoroscopy. This is much less than the
risk of 20 per million reported by Calkins et al in female
patients.7
Accuracy of Radiation Risk Measurement
Increased Risk of Malignancy Associated With Obesity
The computer programs used to convert Diamentor and TLD data to
effective and organ dose assume a particular body size (70 kg and 174
cm height)-so-called "standard man." Smaller patients may have
doses underestimated, and similarly, larger patients may have doses
overestimated. Because the average patient mass and height in this
study were not greatly removed from standard man, dosimetry errors will
be minimal.
Reliability of Measuring Radiation to Breasts With Single
Thermoluminescent Dosimeter Over the Xiphisternum
Risks Are Dependent on Fluoroscopy Unit Used
Although an intercomparison of doses from different fluoroscopy units
was not an aim of this study, it must be pointed out that there can be
a wide range in actual doses from different equipment. For example, the
skin entranceabsorbed dose rate from another commonly used
fluoroscopy unit we have tested was 24 to 60 mGy/min for 100-cm FID and
80-cm FDD, which equates to 43 to 107 mGy/min at 60-cm FDD. The patient
skin doses and corresponding risks could therefore have been 3 to 5
times higher for the same fluoroscopy times if the other fluoroscopy
unit had been used. The output dose rate is affected by many factors
such as total filtration, programmable added filtration, generator
design and the corresponding kVp waveform, available tube current, and
pulsing techniques. The relatively low patient doses in our study are
most likely to be due to the relatively low tube currents (0.5 to 2 mA)
and the high total beam filtration.
Pregnant Patients
Study Limitations
The fluoroscopy times were relatively long. Our x-ray system did not
have the capability of last image hold. Thus continuous fluoroscopy was
needed to assess catheter positions. The study was performed in the
relatively early days of radiofrequency ablation before advances such
as special guiding sheaths, temperature-controlled radiofrequency
ablation, and new designs of ablation catheters. Nevertheless, these
long fluoroscopy times give us a chance to assess more accurately the
radiation doses to vital organs and their potential hazards. These
doses should be able to be reduced considerably by attention to
technique, faster and more effective methods of
electrophysiological studies, and
radiofrequency ablation and using new x-ray technology.
Conclusions
Clinical Implications
The risk of a fatal malignancy during 1 lifetime in the absence
of radiation exposure (other than natural background radiation) is
20%.14 Use of fluoroscopy for 1 hour during
radiofrequency ablation would make this risk 20.03% for the patient.
This small risk associated with the procedure should be looked at in
the light of the benefits associated with the procedure. The procedure
is curative in >85% of patients. It obviates the need for life-long
antiarrhythmic drug therapy with its associated side effects. It avoids
the possible proarrhythmia associated with antiarrhythmic drug
therapy, which could be fatal.24 The quality of
life is likely to be better for the patients cured of their
arrhythmia than those who are dependent on lifelong
antiarrhythmic therapy.25 It is more
cost-effective in the long term than antiarrhythmic drug
therapy.26 27 It also avoids the risks associated
with cardiac surgery, which could be as much as
5%,28 29 and is cheaper and more cost-effective
than cardiac surgery for supraventricular
tachycardia.30
Minimization of Radiation Exposure to Patients
Use of pulsed fluoroscopy is likely to decrease radiation
exposure to the patient.32 Routine use of
collimation as was done in this study is also likely to decrease
radiation exposure to the patient.10 Large
increases in exposure result from using smaller intensifier fields and
this should be kept to the minimum.9 Single-phase
x-ray generators use a higher exposure rate to achieve a given image
quality than do fluoroscopic devices with medium frequency or 3-phase
generators.9
The most effective means of checking the cumulative radiation
dose to the patient during a study is the installation of a dose-area
product meter. The operator can then not only monitor the dose at
any stage, but also (if wished) estimate the effective dose and thus
risk at the completion of a procedure. The latter becomes even more
important when very long or multiple procedures are needed on an
individual patient.
Last, the small risks of radiation-induced malignancy should be
explained to patients undergoing procedures requiring prolonged
fluoroscopy or acquisition times to ensure that they are fully informed
of the potential risks.
Received September 10, 1997;
revision received June 3, 1998;
accepted June 10, 1998.
2.
Klein LS. Radiofrequency catheter ablation: safety and
practicality. Circulation.. 1991;84:25942597.
3.
Jackman WM, Beckman KJ, McClelland JH, Wang XW, Friday
KJ, Roman CA, Moultan KP, Twidale N, Hazlitt A, Prior MI, Oren J,
Overholt ED, Lazzara R. Treatment of supraventricular
tachycardia due to atrioventricular nodal
reentry by radiofrequency catheter ablation of slow-pathway conduction.
N Engl J Med. 1992;327:313318.[Abstract]
4.
Calkins H, Langberg J, Sousa J, El-Atassi R, Leon A,
Kou W, Kalbfleisch S, Morady F. Radiofrequency catheter ablation of
accessory atrioventricular connections in 250 patients:
abbreviated therapeutic approach to Wolff-Parkinson-White syndrome.
Circulation. 1992;85:13371346.
5.
Jackman WM, Xunzhang W, Friday KJ, Roman CA, Moulton
KP, Beckman KJ, McClelland JH, Twidale N, Hazlitt A, Prior MI, Margolis
PD, Calame JD, Overholt ED, Lazzara R. Catheter ablation of accessory
atrioventricular pathways (Wolff-Parkinson-White
syndrome) by radiofrequency current. N Engl J Med. 1991;324:16051611.[Abstract]
6.
Ross DL, Johnson DC, Denniss AR, Cooper MJ, Richards
DA, Uther JB. Curative surgery for atrioventricular
junctional (AV nodal) reentrant tachycardia. J
Am Coll Cardiol. 1985;6:13831392.[Abstract]
7.
Calkins H, Niklason L, Sousa J, El-Atassi R, Langberg
J, Morady F. Radiation exposure during radiofrequency catheter ablation
of accessory atrioventricular connections.
Circulation. 1991;84:23762382.
8.
Kovoor P, Ricciardello M, Collins L, Uther JB, Ross
DL. Radiation exposure to patient and operator during radiofrequency
ablation for supraventricular tachycardia.
Aust NZ J Med. 1995;25:490495.[Medline]
[Order article via Infotrieve]
9.
NHMRC. Recommendations for minimising
radiological hazards to patients (1985). National Health and
Medical Research Council, Canberra: Australian Government Publishing
Service; 1986.
10.
Marshall NW, Faulkner K. The dependence of the
scattered radiation dose to personnel on the technique factors in
diagnostic radiology. Br J Radiol. 1992;65:4449.
11.
Shrimpton P, Wall B. An evaluation of the Diamentor
transmission ion chamber in indicating exposure-area product during
diagnostic radiological examinations. Phys Med
Biol. 1984;27:871878.
12.
ICRP60. Recommendations of the International Commission
on Radiological Protection. In: Ann ICRP. 21: 13, Oxford:
Pergamon Press; 1990:24.
13.
UNSCEAR. Sources, effects and risks of ionizing
radiation. In: Annex F. Radiation carcinogenesis in man.
United Nations Scientific Committee on the Effects of Atomic Radiation,
United Nations, New York: 1988:E.88. IX.7, 1988b.
14.
National Research Council. Health effects of
exposure to low levels of ionizing radiation, BEIR V. Washington,
DC: National Academy Press; 1990.
15.
Petersen LE, Rosenstein M. Computer Program for
Tissue Doses in Diagnostic Radiology. Food and Drug
Administration, Center for Devices and Radiological Health, Rockville,
Md. 1989.
16.
Le Heron J. Estimation of effective dose to the patient
during medical x-ray examinations from measurements of the dose-area
product. Physics Med Biol. 1992;37:21172126.[Medline]
[Order article via Infotrieve]
17.
ICRP60. Recommendations of the International Commission
on Radiological Protection. In: Ann ICRP. 21: 13, Oxford:
Pergamon Press; 1990:22.
18.
ICRP 60. Recommendation of the International Commission
on Radiological Protection. In: Ann ICRP. 21: 13, Oxford:
Pergamon Press; 1990:153.
19.
SPSS Inc. SPSS (Release 4). Chicago: SPSS
International; 1990.
20.
ICRP60. Recommendation of the International Commission
on Radiological Protection. In: Ann ICRP. 21: 13, Oxford:
Pergamon Press; 1990:45.
21.
Shope T. Avoidance of serious x-ray induced skin
injuries to patients during fluoroscopically guided procedures.
AAPM Newsletter. November-December: 1994.
22.
Faulkner K, Love HG, Sweeney JK, Bardsley RA. Radiation
doses and somatic risk to patients during cardiac radiological
procedures. Br J Radiol. 1986;59:359363.
23.
Shope T. Radiation-induced skin injuries from
fluoroscopy. Radiology (Suppl). 1995;197(P):209. Abstract.
24.
The Cardiac Arrhythmia Suppression Trial (CAST)
Investigators. Preliminary report: effect of encainide and flecainide
on mortality in a randomized trial of arrhythmia suppression
after myocardial infarction. N Engl J Med. 1989;321:406412.[Abstract]
25.
Lau C-P, Tai Y-T, Lee PWH. The effects of
radiofrequency ablation versus medical therapy on the quality-of-life
and exercise capacity in patients with accessory pathway-mediated
supraventricular tachycardia: a treatment
comparison study. PACE. 1995;18(pt l):424432.
26.
Kertes PJ, Kalman JM, Tonkin AM. Cost effectiveness of
radiofrequency catheter ablation in the treatment of
symptomatic supraventricular
tachyarrhythmias. Aust NZ J Med. 1993;23:433436.[Medline]
[Order article via Infotrieve]
27.
Kalbfleisch SJ, Calkins H, Langberg JJ. Comparison
of the cost of radiofrequency catheter modification of the
atrioventricular node and medical therapy for
drug-refractory atrioventricular node reentrant
tachycardia. J Am Coll Cardiol. 1992;19:15831587.[Abstract]
28.
Menasche P, Leclercq JF, Cauchemez B, Leenhardt A,
Coumel P, Siama R, Piwnica N. Surgery for the Wolff-Parkinson-White
syndrome in 73 consecutive patients: what have we learned from
intraoperative mapping? Eur J Cardiothoracic Surg. 1989;3:387390.[Abstract]
29.
Rowland E, Robinson K, Edmondson S, Krikler DM, Bentall
HH. Cryoablation of the accessory pathway in Wolff-Parkinson-White
syndrome: initial results and long-term follow-up. Br Heart
J. 1988;59:453457.
30.
De Buitleir M, Bove EL, Schmaltz S. Cost of catheter
versus surgical ablation in the Wolff-Parkinson-White syndrome.
Am J Cardiol. 1990;66:189192.[Medline]
[Order article via Infotrieve]
31.
Kertes P, Kalman J, Edis B, Chen J-M, Byrgiotis S,
Kelly P, Tonkin A, Wilkinson J. Radiofrequency catheter ablation of
tachyarrhythmias: adult and paediatric experience.
Aust NZ J Med. 1993;23:426432.[Medline]
[Order article via Infotrieve]
32.
Holmes DR, Bove AA, Wondrow MA, Gray JE. Video x-ray
progressive scanning: new technique for decreasing x-ray exposure
without decreasing image quality. Mayo Clin Proc. 1986;61:321326.The study aimed to determine the patient-related
radiation risks associated with radiofrequency ablation for
supraventricular tachycardia. Nine female
patients were studied. Radiation dose was determined with the use of 41
thermoluminescent dosimeters and a Diamentor dose-area product
meter. The estimated total lifetime excess risk of a fatal malignancy
was 0.03% per 60 minutes of fluoroscopy. Prolonged fluoroscopy during
radiofrequency ablation may potentially cause a small increase in the
lifetime risk of fatal malignancy. This risk is small only with the use
of techniques and x-ray equipment optimized to keep radiation as low as
possible.[Medline]
[Order article via Infotrieve]
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Risk to Patients From Radiation Associated With Radiofrequency Ablation for Supraventricular Tachycardia
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundRadiofrequency
ablation may be associated with prolonged fluoroscopy times. Previous
studies have calculated radiation risks by measuring the radiation dose
at a limited number (6) of body sites. This is an inherently inaccurate
measure. Our study aimed to quantify more precisely patient-related
radiation risks associated with radiofrequency ablation for
supraventricular tachycardia.
Key Words: radiation risks RF ablation fluoroscopy
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
Radiofrequency ablation has become the preferred method
of treatment of supraventricular
tachycardia.1 2 The success rate has
been >85% at the first attempt, with a low incidence of
complications.3 4 5 This procedure has replaced
surgery as the preferred method for cure of these
conditions.6 Prolonged fluoroscopy (66±37
minutes) is often required for these procedures, especially in patients
with multiple atrioventricular
connections.4 The majority of patients who
undergo these procedures are young, with a mean age of about 36
years.4 Predisposition to late malignancies as a
sequel to this radiation exposure is therefore a source of concern.
Calkins et al7 quantitated the radiation exposure
during radiofrequency catheter ablation of accessory
atrioventricular connections. Kovoor et
al8 assessed radiation exposure during
radiofrequency catheter ablation of atrioventricular
junctional reentry as well as accessory
atrioventricular connections. The study performed by
Calkins et al estimated the subsequent risk of a fatal malignancy for
the irradiated patient to be approximately 0.7 per 1000 patients or 1
per 1000 patients per 1 hour of fluoroscopy.7 In
that study and in the study by Kovoor et al,8
there was a large variation in the radiation doses measured by the
thermoluminescent dosimeters (TLDs) and poor correlation with
fluoroscopy times. This was most likely due to an insufficient number
of TLDs to ensure that the x-ray beam always intercepted dosimeters in
the array as the beam is moved during the study. The mean area of the
x-ray beam at the skin surface was 100±10 cm2,
whereas the mean area of the patient's back over which the beam could
possibly move was 720 cm2. Both the previously
discussed studies used TLDs at only 5 to 6 sites on the patients.
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
Radiation exposure was determined in 9 female patients having
radiofrequency ablation for supraventricular
tachycardia. Informed consent was obtained from all
patients. Only women were studied to allow accurate quantitation of
female breast irradiation. The patient characteristics are given in
Table 1
. All patients had
symptomatic supraventricular tachycardia.
View this table:
[in a new window]
Table 1. Patient and Procedure
Characteristics
Radiation dose was determined at 41 body sites in each patient
with the use of TLDs (Teledyne TLD1000 LiF-N chips, 3 mm square).
The TLDs were calibrated at the mean kVp used for each patient, with an
mdh Radcal model 2025 ionizing chamber dosimetry system. The TLDs were
positioned on the patient before commencement of the procedure.
Radiation to the thorax was measured with a grid of 30 evenly spaced (6
cm) TLDs positioned on the posterior aspect of the thorax (Figure 1
), covering a total area of 720
cm2. The positioning of these TLDs was guided by
a marked plastic sheet template so as to have uniform positions and
distribution in each patient. The C8 spinous process was used as a
reference point for placement of the template. Doses from the 20
highest-reading TLDs were averaged.

View larger version (18K):
[in a new window]
Figure 1. Median (interquartile range) skin absorbed dose of
radiation (mGy) per 60 minutes of fluoroscopy measured from 31 TLDs
positioned on posterior aspect of the patient. Grid of 30 TLDs was
positioned over posterior thorax. One TLD was positioned posterior to
thyroid at level of C8 spinous process.
). Of those 9, 4 TLDs were positioned
over each breast (lateral, medial, inferior, and just above
the nipple) and 1 TLD was placed over the xiphisternum (Figure 2
). In
addition, radiation to the thyroid was estimated with 1 TLD positioned
anterior and 1 posterior to the thyroid (Figures 1
and 2
).

View larger version (13K):
[in a new window]
Figure 2. Median (interquartile range) skin absorbed dose of
radiation (mGy) per 60 minutes of fluoroscopy measured from 10 TLDs
positioned on anterior aspect of patient. Four TLDs were positioned
over each breast (lateral, medial, inferior, and just above
the nipple) and 1 thermoluminescent dosimeter was placed over the
xiphisternum. One thermoluminescent dosimeter was positioned anterior
to the thyroid.
The dose-area product of the total output of the x-ray tube
was measured simultaneously with a PTW Diamentor-D
dose-area product meter. The ion chamber was positioned on the
x-ray tube before commencement of the procedure. The Diamentor was
calibrated in a manner previously
described.11
The procedures were performed by 2 experienced
electrophysiologists, each of whom had performed
100 radiofrequency
ablations before the study. All procedures were performed with patients
in the postabsorptive state during sedation with continuous infusion of
midazolam and fentanyl. Quadripolar catheters were inserted through the
left femoral vein and positioned in the high right atrium and right
ventricular apex. A tripolar catheter was inserted through
the left femoral vein to record the His bundle. A decapolar
catheter was inserted through the right subclavian vein into the
coronary sinus. A detailed diagnostic study was
performed in all patients before the ablation to determine the baseline
electrophysiological properties, study the
inducibility of the tachyarrhythmias, and map in detail
the locations of the accessory pathway or reentrant circuits.
Risks were calculated by 2 independent methods, using the
measured radiation absorbed dose from TLD and also the dose area
product from the Diamentor. International Council for Radiation
Protection (ICRP) 60 (1990) risk estimates12 were
applied to estimated absorbed organ doses. These risk estimates (or
probability of fatal cancer) were derived from a variety of sources,
but particularly the 1988 UNSCEAR report13 and
the BEIR V report.14 From this data, a best
estimate of risk was selected for ICRP 60. While the accuracy of the
risk estimates could be challenged, they remain the best values
available.
For the breast, the average measured skin dose for the 9 TLDs
was used. The anterior measured skin dose was used for the thyroid. For
the lung, ovaries, and bone marrow, the organ doses were calculated by
a freely available computer program15 from the
average skin absorbed dose measured for 20 sites (Figure 1
, rows 2 to
5) over an area of 24 by 18 cm, centered on the midline and PA chest
projection. The average of 20 sites was used because the x-ray
field is moved and angulated frequently during a procedure. As a
result, some TLDs could be out of the x-ray field for part of the time
or at varying distances from the x-ray tube (causing inverse square law
effects). To calculate the total risk of excess fatal cancer, the mean
absorbed dose to the critical organs in the primary beam (lungs, bone
marrow, and breasts) and to the thyroid and ovaries was multiplied by
the appropriate risk factors from ICRP 60.12 The
risk factors used were: lungs 8.5; bone marrow 5; breasts 2; and
thyroid 0.8 fatal cancers per mGy organ dose per million cases. This of
course assumes that the only significant radiation-related cancer risks
will be to these organs.
The risk of fatal cancer of all types was also determined from
the dose-area product. The dose-area product data was converted
to effective dose with the approach of LeHeron,16
with the use of a commercially available computer program (XDOSE)(1).
The effective dose was then multiplied by the probability coefficient
for fatal cancer.17 This program assumes the
patient is an adult of 70 kg mass and 174 cm height (body mass
index=23.1).
The risk of hereditary effects was obtained by multiplying the
average dose to the ovaries by the probability of severe hereditary
effects (10 per mSv per million cases).18 The
dose to the ovaries was calculated by a computer
program,15 using the average skin entrance
exposure from 20 TLD sites (Figure 1
, rows 2 to 5).
The statistical package SPSS was used.19
Spearman rank correlation coefficients were used to examine the
associations between the different variables of interest.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
The details of fluoroscopy time are summarized in Table 1
. The
median skin absorbed doses measured by the 41 TLDs are summarized in
Figures 1
and 2
. The dose of radiation to individual organs and the
estimated risk of malignancy are summarized in Table 2
. The estimated mean total lifetime risk
of fatal malignancy from 60 minutes of fluoroscopy for radiofrequency
ablation measured from the average organ doses (measured using TLDs)
was 294 excess cases of fatal malignancy per million patients or 0.03%
excess cases of fatal malignancy. From the dose-area product
measured with the Diamentor, the estimated mean total lifetime excess
risk of fatal malignancy was 317 per million cases per 60 minutes of
fluoroscopy (mean effective dose=6.34 mSv). The risk of severe
hereditary effects was less than 1 per million cases for 60 minutes of
fluoroscopy.
View this table:
[in a new window]
Table 2. Absorbed Dose (mGy) to Individual Organs per 60
Minutes of Fluoroscopy and Estimated Risk of Fatal Malignancy per 60
Minutes of Fluoroscopy
). The risk of a fatal malignancy
increased with increase of body mass index (Figure 4
). There was good correlation between
the mean absorbed dose measured by the 8 TLDs positioned directly over
the breasts and the single thermoluminescent dosimeter positioned over
the xiphisternum (Figure 5
). The
estimated number of excess cases of fatal breast malignancy was
8.6±3.8 per million patients per 60 minutes of fluoroscopy when the
radiation dose to the breast was calculated from the mean of the 8 TLDs
directly over the breasts and assuming that this figure is closely
related to the mean organ absorbed dose. Given that the TLD positions
ranged from nipple to chest wall, the approximation is reasonable.

View larger version (15K):
[in a new window]
Figure 3. Correlation between the excess risk of fatal
malignancy per million cases per 60 minutes of fluoroscopy as measured
with data from TLDs versus that measured with dose-area product
meter.

View larger version (14K):
[in a new window]
Figure 4. Correlation between excess risk of fatal
malignancy per million cases per 60 minutes of fluoroscopy and body
mass index.

View larger version (12K):
[in a new window]
Figure 5. Correlation between mean radiation (mGy) measured
from directly over breasts with 8 TLDs versus that measured with 1 TLD
over the xiphisternum.
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
This study provides a detailed assessment of the radiation
exposure to the patients and the resultant risks associated with
radiofrequency ablation for supraventricular
tachycardia. We were able to demonstrate good agreement
between 2 independent methods of radiation measurement (that is, TLD
and dose-area product methods) indicating that the results are
likely to be reliable. The study confirms that use of fluoroscopy
during catheter radiofrequency ablation is likely to result in a small
increase in the lifetime risk of a fatal malignancy and that the most
likely malignancy will be lung.
The biological effects of radiation can be both somatic (those
occurring in the exposed person) or hereditary. Somatic effects are
either stochastic (where, according to the currently accepted linear
no-threshold hypothesis any exposure carries a risk) or deterministic
(where there is a threshold dose for the effect to occur and below
which no damage occurs).20 Stochastic effects
include malignancies, especially leukemia. Deterministic effects
include cataract formation and some developmental abnormalities of
children exposed in utero. Hereditary effects are those that affect the
germ cells and may be evident in the progeny.
During electrophysiological studies, a
variety of fluoroscopy views are used. This results in considerable
variation in the areas affected by radiation. To allow for this and to
enable accurate measurement of radiation to the patient, the TLDs were
positioned at a large number (41) of body sites. The maximum radiation
to the thorax was detected by the TLDs in the center (Figure 1
),
suggesting that the radiation was well bracketed by the TLDs. The risk
estimates derived from the Diamentor dose-area product and TLDs
were similar, suggesting that the radiation dose was measured reliably
(Figure 3
). This is in contrast to the discrepancy between the 2 in
other studies in which only 7 TLDs were
used.22
As would be expected, the estimated risk of malignancy increased
with increasing body mass index of the patient (Figure 4
) because of
the increase in radiation necessary to obtain satisfactory images in
obese patients. Body mass index alone, however, is not necessarily a
good measure of body thickness in the region of the x-ray beam.
Other studies that have assessed radiation to patients associated
with cardiac catheterization have used one TLD over the
xiphisternum to determine the radiation to the
breasts.7 22 In our study, we assessed the
reliability of measuring radiation to the female breasts by using both
a single TLD over the xiphisternum and 4 TLDs over each breast. There
was a good correlation between the absorbed dose detected by the single
xiphisternal TLD and the average of multiple breast TLDs. Measurement
of radiation dose by a single TLD over the xiphisternum is an
alternative measure of breast irradiation in women with the x-ray
equipment and projections used in this study; however, it may
overestimate the mean glandular dose.
The dose rate at skin entrance from the fluoroscopy unit used in
this study was lower than those from some fluoroscopy units commonly
used for interventional cardiology. The patient dose
rate from the unit used in this study ranged from 5 to 17 mGy/min for
field sizes 15 to 23 cm and tube current settings 0.5 to 4 mA with
20-cm tissue equivalent material placed on the tabletop at 90-cm focus
to image intensifier distance (FID) and 60-cm focus to detector
distance (FDD).
It is highly likely that eventually a patient will subsequently
discover that she was pregnant at the time of the fluoroscopic
procedure. On the basis that the calculated ovary dose would be the
same as that to the uterus in very early pregnancy, the estimated doses
of <0.05 mGy indicates that there would be very little risk to a
fetus. In the case that the pregnancy is less than 3 weeks old, it is
currently believed that neither stochastic nor deterministic effects in
the live-born child due to radiation exposure are
likely.12 Beyond this period, the risks are
related to gestational age but would still be very low.
The number of patients studied was relatively small. Only women
were included in this study. This was done to enable accurate
quantitation of radiation to the female breasts. The study does not
provide information about radiation risks to men. However, it is likely
that the radiation risks are similar in men and women except to the
gonads, which are exposed to
4 times more radiation in women than
men during cardiac
catheterization.22 The radiation
to the gonads in women during our study was <0.05 mGy per 60 minutes
of fluoroscopy.
The estimated total lifetime excess risk of a fatal malignancy was
294 per million cases (0.03%) per 60 minutes of fluoroscopy associated
with radiofrequency ablation. This risk will be higher if fluoroscopy
units requiring higher output than ours are used during the procedure.
The lungs are exposed to the maximum amount of radiation and are the
most likely site of malignancy. The risk of radiation induced
malignancy increases with increase of body mass index of patients.
Radiation measured by a single thermoluminescent dosimeter positioned
over the xiphisternum gives a fair assessment of the radiation to the
breasts in women.
Prolonged fluoroscopy during radiofrequency ablation is likely to
cause a small increase in the lifetime risk of fatal malignancy. This
risk is approximately one third of the estimated risk from
radiofrequency ablation reported previously. The risk is increased in
obese patients and in laboratories with high-output fluoroscopy units.
The advent of x-ray equipment that can operate for long periods with
high tube currents necessitates caution because of the high skin doses
possible and reports of skin doses of up to 20 Gy, causing severe
burns.23 Such high-output equipment will of
course significantly increase the risks of malignancy from the use of
fluoroscopy for a minimal increase in image quality.
Every effort should be made to decrease the radiation exposure to
patients during radiofrequency ablation. The duration of fluoroscopy
should be minimized as much as possible. This is likely to occur with
increased experience and skill of operators.31
However, prolonged fluoroscopy use is still likely in complex cases
with multiple pathways.4 Operators should ensure
that high-output fluoroscopy units are not used for procedures such as
radiofrequency ablation, in which prolonged fluoroscopy usage might
occur.
![]()
Acknowledgments
(1)XDOSE is a computer program for Monte Carlo calculation of
the ICRP dose index "effective dose" from the dose-area
product. XDOSE uses data files from Software report NRPB-SR262. The
XDOSE executable file is available from the National Radiation
Laboratory, PO Box 25 to 099, Christchurch, New Zealand. The data
files, known as Software report NRPB-SR262 must be purchased directly
from the National Radiological Protection Board (NRPB), Chilton, Oxon,
OX11 0RQ, UK. The authors are indebted to John Crancher, Radiation
Safety Technician, for his technical help, Dr Karen Byth for her expert
statistical analysis, Dr John Heggie for his helpful comments,
and to the nurses and technicians of the electrophysiology laboratory
for their assistance.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
1.
Ross DL. Radiofrequency catheter ablation for
supraventricular tachycardias. Aust NZ
J Med. 1993;23:339342.[Medline]
[Order article via Infotrieve]
This article has been cited by other articles:
![]() |
R. Liew, V. Baker, L. Richmond, K. Rajappan, D. Gupta, M. Finlay, G. Thomas, M. Earley, S. Sporton, S. Harris, et al. A randomized-controlled trial comparing conventional with minimal catheter approaches for the mapping and ablation of regular supraventricular tachycardias Europace, May 2, 2009; (2009) eup108v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. E. Eckart and L. M. Epstein Interventional Therapy for Atrial and Ventricular Arrhythmias Card. Surg. Adult, January 1, 2008; 3(2008): 1357 - 1374. [Full Text] |
||||
![]() |
P. Kirchhof, A. Auricchio, J. Bax, H. Crijns, J. Camm, H.-C. Diener, A. Goette, G. Hindricks, S. Hohnloser, L. Kappenberger, et al. Outcome parameters for trials in atrial fibrillation: executive summary: Recommendations from a consensus conference organized by the German Atrial Fibrillation Competence NETwork (AFNET) and the European Heart Rhythm Association (EHRA) Eur. Heart J., November 2, 2007; 28(22): 2803 - 2817. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Kirchhof, A. Auricchio, J. Bax, H. Crijns, J. Camm, H.-C. Diener, A. Goette, G. Hindricks, S. Hohnloser, L. Kappenberger, et al. Outcome parameters for trials in atrial fibrillation: Recommendations from a consensus conference organized by the German Atrial Fibrillation Competence NETwork and the European Heart Rhythm Association Europace, November 1, 2007; 9(11): 1006 - 1023. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al. HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society. Europace, June 1, 2007; 9(6): 335 - 379. [Full Text] [PDF] |
||||
![]() |
E. P. Efstathopoulos, D. G. Katritsis, S. Kottou, N. Kalivas, E. Tzanalaridou, E. Giazitzoglou, S. Korovesis, and K. Faulkner Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures: a comprehensive analysis Europace, June 1, 2006; 8(6): 443 - 448. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Earley, R. Showkathali, M. Alzetani, P. M. Kistler, D. Gupta, D. J. Abrams, J. A. Horrocks, S. J. Harris, S. C. Sporton, and R. J. Schilling Radiofrequency ablation of arrhythmias guided by non-fluoroscopic catheter location: a prospective randomized trial Eur. Heart J., May 2, 2006; 27(10): 1223 - 1229. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Chida, H. Saito, H. Otani, M. Kohzuki, S. Takahashi, S. Yamada, K. Shirato, and M. Zuguchi Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. Am. J. Roentgenol., March 1, 2006; 186(3): 774 - 778. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Padovani and E. Quai Patient dosimetry approaches in interventional cardiology and literature dose data review Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 217 - 221. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Lickfett, M. Mahesh, C. Vasamreddy, D. Bradley, V. Jayam, Z. Eldadah, T. Dickfeld, D. Kearney, D. Dalal, B. Luderitz, et al. Radiation Exposure During Catheter Ablation of Atrial Fibrillation Circulation, November 9, 2004; 110(19): 3003 - 3010. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. D. Satti and L. M. Epstein Cardiologic Interventional Therapy for Atrial and Ventricular Arrhythmias Card. Surg. Adult, January 1, 2003; 2(2003): 1253 - 1270. [Full Text] |
||||
![]() |
J. Damilakis, N. Theocharopoulos, K. Perisinakis, E. Manios, P. Dimitriou, P. Vardas, and N. Gourtsoyiannis Conceptus Radiation Dose and Risk From Cardiac Catheter Ablation Procedures Circulation, August 21, 2001; 104(8): 893 - 897. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Perisinakis, J. Damilakis, N. Theocharopoulos, E. Manios, P. Vardas, and N. Gourtsoyiannis Accurate Assessment of Patient Effective Radiation Dose and Associated Detriment Risk From Radiofrequency Catheter Ablation Procedures Circulation, July 3, 2001; 104(1): 58 - 62. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Kottkamp, B. Hugl, B. Krauss, U. Wetzel, A. Fleck, G. Schuler, and G. Hindricks Electromagnetic Versus Fluoroscopic Mapping of the Inferior Isthmus for Ablation of Typical Atrial Flutter : A Prospective Randomized Study Circulation, October 24, 2000; 102(17): 2082 - 2086. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |