(Circulation. 1998;98:2210-2215.)
© 1998 American Heart Association, Inc.
Basic Science Reports |
From the Division of Cardiovascular Diseases, Department of Medicine, University of Alabama at Birmingham (G.P.W., S.B.M., W.M.S., R.E.I.); Georgetown University, Washington, DC (J.L.J.); and Physio-Control Corp (F.W.C.), Redmond, Wash.
Correspondence to Gregory P. Walcott, MD, B140 Volker Hall, 1670 University Blvd, University of Alabama at Birmingham, Birmingham, AL 35294. E-mail gpw{at}crml.uab.edu
| Abstract |
|---|
|
|
|---|
Methods and ResultsIn protocol 1, 50% success levels, ED50, were measured after 15 seconds of fibrillation for the 3 waveforms in 6 dogs. In protocol 2, defibrillation thresholds were measured for QSBW and CDSMW after 15 seconds of fibrillation and after 3 minutes of unsupported fibrillation followed by 2 minutes of fibrillation with femoral-femoral cross-circulation. In protocol 1, QSBW had a lower ED50, 16.0±4.9 J, than TEBW, 20.3±4.4 J, or CDSMW, 27.4±6.0 J. In protocol 2, QSBW had a lower defibrillation threshold after 15 seconds, 38±10 J, and after 5 minutes, 41.5±5 J, than CDSMW after 15 seconds, 54±19 J, and 5 minutes, 80±30 J, of fibrillation. The defibrillation threshold remained statistically the same for QSBW for the 2 fibrillation durations but rose significantly for CDSMW.
ConclusionsIn this animal model of sudden death and resuscitation, these 2 biphasic waveforms are more efficacious than the CDSMW at short durations of fibrillation. Furthermore, the QSBW is even more efficacious than the CDSMW at longer durations of fibrillation.
Key Words: defibrillation cardiopulmonary resuscitation arrhythmia
| Introduction |
|---|
|
|
|---|
Another type of biphasic waveform exists for external defibrillation in addition to the truncated exponential biphasic waveform, ie, the quasi-sinusoidal waveform.8 A study in humans has shown that the probability of success of a 200-J shock was higher for the biphasic quasi-sinusoidal waveform than for the critically damped monophasic waveform.9 Although both biphasic waveforms perform better than the monophasic waveform, it is not known which of these 2 biphasic waveforms has the lower defibrillation threshold for transthoracic defibrillation.
Shocks from the implantable cardioverter-defibrillator are given within seconds of the onset of fibrillation. Because most previous studies of biphasic waveforms have focused on internal defibrillation, shocks have been delivered after only 10 to 30 seconds of ventricular fibrillation. In contrast, the first transthoracic defibrillation shock in the prehospital setting is often administered >5 minutes after the patient collapses.10 11 The defibrillation efficacy of biphasic waveforms after minutes of fibrillation has not been compared with that of monophasic waveforms.
This study was designed to test the following hypotheses. First, we hypothesized that both biphasic waveforms, the truncated exponential and the quasi-sinusoidal, would defibrillate with a lower delivered energy than the critically damped monophasic waveform after short durations of ventricular fibrillation. Furthermore, we predicted that the quasi-sinusoidal biphasic waveform, with its gradual onset, would defibrillate at a lower delivered energy than the truncated exponential biphasic waveform, with its abrupt onset, and the critically damped monophasic waveform. Second, we hypothesized that the more efficacious biphasic waveform in protocol 1 would also defibrillate at a lower delivered energy than the critically damped monophasic waveform after 5 minutes of fibrillation.
| Methods |
|---|
|
|
|---|
The portions of the experimental procedure common to both protocols
will be described first. Each dog was anesthetized with sodium
pentobarbital 20 to 30 mg/kg and maintained on a constant infusion of
0.05 mg · kg-1 ·
min-1 through an intravenous
catheter in a foreleg. ECG lead II was monitored throughout the
experiment. The animal was intubated with a cuffed endotracheal tube
and ventilated with a volume-controlled respirator (Harvard
Apparatus) with an air/oxygen ratio to keep a
PO2 of
100 mm Hg. A constant
infusion (5 to 10 mL · kg-1 ·
h-1) of lactated Ringer's solution was
delivered. A femoral arterial catheter was placed for
hemodynamic monitoring and for vascular access for
blood gas and electrolyte analysis. Blood gas analysis
was performed every 30 to 60 minutes as indicated by the condition of
the animal, and corrections in PO2,
PCO2, Na+,
K+, Ca2+, and bicarbonate
were made as necessary to maintain normal values. Body temperature was
monitored with an esophageal temperature probe and maintained at 37°C
to 38°C with a heating pad placed beneath the animal.
FASTPATCH disposable defibrillation electrodes (Physio-Control Corp) were placed on the left and right chest walls. Fibrillation was induced with 60-Hz current from a quadripolar electrophysiology catheter (EP Technologies) in the right ventricle. After all test shocks were delivered, the animal was euthanized. The heart was removed from the animal, weighed, and stored in formalin.
Protocol 1
Three waveforms were tested. The monophasic waveform (Edmark)
was a critically damped sinusoidal waveform (Figure 1A
).12 One biphasic
waveform (Gurvich) was a quasi-sinusoidal waveform (Figure 1B
).8 Both sinusoidal waveforms were delivered
from a modified LIFEPAK 7 external defibrillator (Physio-Control Corp).
The second biphasic waveform was a truncated exponential waveform
delivered from a single 280-µF capacitor bank in which each phase had
a 35% tilt (Figure 1C
) delivered from an HVS-02 external defibrillator
(Ventritex Co) modified to accept an external capacitor bank of
arbitrary size and to deliver shocks of leading-edge voltage up to 1300
V. The capacitor size was made large relative to the capacitor size
used in internal defibrillators to lower the leading-edge voltage of
the shock necessary to defibrillate the animal.13
Phase 1 duration was chosen to be optimal for a 75-
impedance with a
parallel resistor-capacitor network used to represent the
heart.14 The second phase of the biphasic
waveform was chosen to be equal to the first
phase.1 Shocks were given after 15 seconds of
fibrillation. If the test shock failed, the heart was rescued with a
shock of the same shape but approximately twice the energy.
|
Probability-of-success curves were determined for each waveform in an interleaved fashion. Shocks were delivered in groups of 3, with 1 shock of each waveform shape being delivered in each group. The order of shocks within each group of 3 was randomized. The starting energy for each waveform was 25 J. An up/down protocol was followed until 15 shocks were delivered for each waveform. The 15-shock count did not start until the first reversal from success to failure or failure to success occurred. If the test shock for a particular waveform succeeded, the next shock for that waveform was decreased in energy by 10%. If the test shock failed, the next test shock was increased 10% in energy.
Probability-of-defibrillation-success curves were determined for each waveform by a Probit regression fit method.15 The 50% effective dose point (ED50) for delivered energy for each waveform was compared by a repeated-measures ANOVA. The null hypothesis that the ED50 values were not different for the different waveforms was rejected at P<0.05. The Student-Newman-Keuls test for multiple comparisons was used to determine differences among the 3 waveforms if the null hypothesis was rejected by the repeated-measures ANOVA.
Protocol 2
Two waveforms were tested, the critically damped monophasic
waveform (Figure 1A
) and the quasi-sinusoidal biphasic waveform (Figure 1B
). The defibrillation threshold was first determined for the 2
waveforms after 15 seconds of fibrillation with an interleaved up/down
protocol (Figure 2
). Shocks were
delivered in pairs, with 1 of each shape given before the next pair of
tests shocks was delivered. Shock order for each pair was randomized.
The initial shock strength was 25 J. If a test shock succeeded for a
particular waveform, the next test shock was decreased by 20% in
energy. If a test shock failed, the next test shock was increased by
20% in energy. This protocol was followed until a reversal from
success to failure or failure to success was recorded for each
shock waveform. The lowest shock strength that successfully
defibrillated the animal was defined as the defibrillation threshold
for that waveform. If the test shock failed, the animal was rescued by
a shock of the same shape but approximately twice the energy.
|
After the defibrillation threshold had been determined for short-duration fibrillation, the animal was given 3000 U heparin. Subsequent doses of 1000 U heparin were given each hour. Cannulas were placed in the right femoral artery and right femoral vein and connected to a perfusion apparatus (Sarns Inc). The perfusion apparatus was primed with 1 L normal saline, 20 mEq KCl, 50 mEq Na2CO3, and 100 mg CaCl2. Defibrillation thresholds were then determined for the 2 waveforms after 5 minutes of ventricular fibrillation. During the first 3 minutes, the animal was allowed to fibrillate without external support. Then the perfusion pump was started, and cross-circulation with unoxygenated blood was given at a flow rate of 1 L/min.
After 2 minutes of perfusion, the test shock was administered. The
initial shock strength was the defibrillation threshold measured after
15 seconds of fibrillation. Defibrillation thresholds were determined
in an interleaved fashion according to the same protocol as for 15
seconds of fibrillation (Figure 2
). When the defibrillation threshold
had been successfully determined for 1 of the waveforms, interleaving
stopped (so as to minimize the number of ventricular
fibrillation episodes), and all subsequent episodes were shocked with
the other waveform. Once defibrillated, the animal was allowed to
recover for 45 minutes before fibrillation was reinduced. If the animal
was asystolic or had pulseless electrical activity after
defibrillation, perfusion was maintained without pacing for 30 seconds.
If the animal was still asystolic after 30 seconds of
perfusion, pacing was begun from the right ventricular
catheter. Perfusion was halted once cardiac mechanical activity was
noted. If the blood pressure was <60 mm Hg after 30 seconds with
no cross-circulation, perfusion was reinstituted and the animal was
resuscitated with 1 mg epinephrine.
The defibrillation threshold was defined as the lowest shock energy that successfully converted ventricular fibrillation to a perfusing rhythm after resuscitation of the animal. A 2-level ANOVA with repeated measures was used to compare defibrillation thresholds. The 2 levels were shock waveform and duration of fibrillation. Again, the Student-Newman-Keuls multiple-comparisons test was used to compare means.
| Results |
|---|
|
|
|---|
|
Protocol 2
The defibrillation threshold for the quasi-sinusoidal biphasic
waveform was significantly lower than that for the critically damped
sinusoidal monophasic waveform for both long- and short-duration
fibrillation (Figure 4
). After 15 seconds
of fibrillation, the defibrillation threshold for the quasi-sinusoidal
biphasic waveform (38±10 J) was lower than for the monophasic
waveform (54±19 J). After 5 minutes of fibrillation, the
defibrillation threshold for the quasi-sinusoidal biphasic waveform
(41±5 J) was lower than for the monophasic waveform (80±30 J). In
addition, the defibrillation threshold for the monophasic waveform was
significantly higher at 5 minutes than at 15 seconds. The
quasi-sinusoidal biphasic waveform defibrillation threshold did not
increase significantly.
|
| Discussion |
|---|
|
|
|---|
Effect of Fibrillation Duration on Defibrillation Energy
Requirements
Only a few studies have examined the effects of the duration of
fibrillation on the defibrillation threshold. Most of these studies
deal with internal defibrillation and with fibrillation durations of
<1 minute.16 17 18 In a dog model of
defibrillation, using internal electrodes and bidirectional monophasic
defibrillation pulses delivered along 2 pathways, Echt et
al16 showed that the energy necessary to
defibrillate rose from 27±13 J at 5 seconds to 41±14 J at 30 seconds
of fibrillation. Jones et al18 showed that in a
working rabbit heart model of defibrillation, both monophasic and
biphasic waveform defibrillation thresholds increased with duration
from 5 to 15 to 30 seconds. At all durations, the biphasic threshold
was lower than the monophasic threshold. This difference increased with
fibrillation duration. In a study using sequential trapezoidal
defibrillation pulses in a pig model of defibrillation, Fujimura et
al19 concluded that a delay in defibrillation
therapy of up to 90 seconds has no significant effect on the ability to
defibrillate the heart. Bardy et al17 found no
difference between the mean defibrillation thresholds in humans when
fibrillation was allowed to continue for 10 versus 20 seconds
(11.5±5.9 versus 12.0±6.9 J, P=NS). Winkle et
al20 showed that in humans, the probability of
successful defibrillation with low-energy shocks (5.9 J) was higher for
ventricular fibrillation lasting 5 seconds than for
ventricular fibrillation lasting 15 seconds, yet there was
no significant difference between the success rates of high-energy
shocks (24.2 J) delivered at the same 2 durations. Together, these
results suggest that for ventricular fibrillation durations
up to 90 seconds, the defibrillation threshold for monophasic waveforms
increases with duration, whereas the results are inconclusive for
biphasic waveforms.
However, the time before a defibrillation shock is administered to individuals experiencing sudden cardiac arrest without an implantable cardioverter-defibrillator (the vast majority of individuals) is much longer than 90 seconds. Studies have shown that the time from initiation of ventricular fibrillation to delivery of a defibrillation shock is at best 6 to 12 minutes.11 21 Very little is known about the effect of several minutes of fibrillation on defibrillation energy requirements. This question becomes important as the concept of "public access defibrillation"22 is implemented and smaller, lighter defibrillators with new waveforms are developed. One way to make defibrillators smaller and lighter is to have their maximum shock size be smaller than the current 360-J standard.23 An important question is, how much energy need these defibrillators be capable of delivering so as to rapidly defibrillate patients after prolonged intervals of ventricular fibrillation?
This study found that the defibrillation threshold increased markedly
with time during fibrillation for the monophasic waveform tested. The
defibrillation threshold was
50% greater after 5 minutes of
fibrillation than after 15 seconds of fibrillation (Figure 4
). The
monophasic waveform tested, the critically damped sinusoidal waveform,
is the waveform used in the majority of external defibrillators that
are in use today. Conversely, the biphasic waveform tested in this
study increased only nonsignificantly between 15 seconds and 5 minutes
of fibrillation. In addition, the defibrillation threshold was much
smaller for the biphasic than the monophasic waveform at both durations
of fibrillation. After 5 minutes of fibrillation, the defibrillation
threshold for the monophasic waveform was almost twice that for the
biphasic waveform (Figure 4
).
Defibrillation Efficacy of Different Types of Biphasic
Waveforms
Two different shapes of biphasic waveforms were examined in this
study: the truncated exponential waveform, similar to that used in
implantable cardioverter-defibrillators,3 4 5 and
the quasi-sinusoidal waveform suggested by Gurvich and
Markarychev.8 After 15 seconds of fibrillation,
the ED50 for each type of biphasic waveform was
significantly lower than that for the monophasic waveform (Figure 3
).
In addition, the ED50 for the truncated
exponential biphasic waveform was
25% higher than for the
quasi-sinusoidal biphasic waveform.
One major difference between the 2 biphasic waveforms is that the truncated exponential waveform has an abrupt onset, with the leading edge of the waveform being almost a step function, whereas the quasi-sinusoidal waveform increases more slowly with time after its onset. Several experimental studies have demonstrated that waveforms with a more gradual onset, such as an ascending ramp, have lower defibrillation energy requirements than waveforms with an abrupt onset to the maximum value, such as a square wave or descending ramp.24 25 26
Modeling studies that represent stimulation or defibrillation by the achievement of a particular minimum voltage across a parallel resistor-capacitor network suggest that an ascending waveform should require less energy than a descending or square waveform if the rate of rise is within a certain range related to the time constant of the resistor-capacitor network.14 This resistor-capacitor network can also be thought of as a low-pass filter that passes the lower frequencies of the quasi-sinusoidal waveform better than the high frequencies generated by the leading and trailing edges of the truncated exponential biphasic waveform.27
The resistor-capacitor network can be used to choose an "optimal"
first-phase duration for a truncated exponential biphasic
waveform.28 If we define the optimal truncation
point for the first phase of a biphasic waveform as the time at which
the voltage across the network is at maximum, then for a 280-µF
capacitor delivered into a 75-
load, the optimal truncation point
should be
9 ms. A 35% tilt waveform using a 280-µF capacitor and
delivered into a 75-
load will have a 9-ms duration. As a first
approximation, choosing the second phase of a biphasic waveform to be
equal to the first phase of the waveform is a good
choice.1 Therefore, although it has not been
rigorously validated, the 35%/35% biphasic waveform delivered from a
280-µF capacitor is probably a reasonable truncated
exponential biphasic waveform to test.
One of the reasons to strive for a lower defibrillation energy is to allow the production of new smaller, lighter external defibrillators. A waveform that consistently defibrillates at a lower energy should allow the construction of a smaller defibrillator. With present technology, however, it is easier to make a small, light defibrillator using a truncated exponential biphasic waveform than using the quasi-sinusoidal biphasic waveform, even though defibrillation efficacy is slightly better for the quasi-sinusoidal waveform. Improvements in technology may make this difference less important in the future. These improvements might include the use of new materials that will make inductors smaller or the use of duty-cycle waveforms to shape waveforms in new ways.27
Study Limitations
Two limitations of protocol 2 are that defibrillation thresholds
instead of probability-of-success curves were measured and that 1 of
the biphasic waveforms tested in protocol 1 was not tested in protocol
2. The reason for these 2 limitations is that it is not possible to
give as many test shocks and fibrillation episodes when each
fibrillation episode lasts 5 minutes as when each fibrillation episode
lasts only 10 or 15 seconds. Because of the possible cumulative effects
of a number of 5-minute episodes of fibrillation, it is possible that
the defibrillation requirements changed during the course of the
experiments in protocol 2. For this reason, shocks for the 2 waveforms
were interleaved. However, once the threshold was determined for 1
waveform, all subsequent episodes were used to determine the threshold
for the other waveform. This was usually the monophasic waveform.
A limitation of both protocols is that the defibrillation patches were placed on the right and left sides of the thorax instead of in the precordial or anterior-posterior configurations that are more commonly used clinically. This was done because the shape of the thorax differs in dogs and humans. Whereas the chest is relatively flat and broad in humans, it is narrow and V-shaped in dogs.
Clinical Implications
These data show that the energy requirements for defibrillation do
not change significantly during the first 5 minutes of fibrillation for
the quasi-sinusoidal biphasic waveform tested in this study. These
results suggest that to be useful in the prehospital setting, a
clinical device using such a waveform may not need the ability to
deliver significantly more energy than is necessary to defibrillate
after a short period of ventricular fibrillation. Studies
are necessary to test whether these animal model results carry over for
prehospital cardiac arrest in humans before we will know whether
biphasic external defibrillators may have a maximum energy capability
below the 360-J level currently indicated in AHA
guidelines.23 Furthermore, it is unknown whether
the advantage shown for the quasi-sinusoidal biphasic waveform would
also apply to the biphasic truncated exponential waveforms.
| Acknowledgments |
|---|
Received January 23, 1998; revision received June 3, 1998; accepted June 16, 1998.
| References |
|---|
|
|
|---|
2.
Feeser SA, Tang ASL, Kavanagh KM, Rollins DL, Smith
WM, Wolf PD, Ideker RE. Strength-duration and probability of success
curves for defibrillation with biphasic waveforms.
Circulation. 1990;82:21282141.
3. Winkle RA, Mead RH, Ruder MA, Gaudiani V, Buch WS, Pless B, Sweeney M, Schmidt P. Improved low energy defibrillation efficacy in man with the use of a biphasic truncated exponential waveform. Am Heart J. 1989;117:122127.[Medline] [Order article via Infotrieve]
4. Bardy GH, Ivey TD, Allen MD, Johnson G, Mehra R, Greene L. A prospective randomized evaluation of biphasic versus monophasic waveform pulses on defibrillation efficacy in humans. J Am Coll Cardiol. 1989;14:728733.[Abstract]
5. Block M, Hammel D, Böcker D, Borggrefe M, Budde T, Isbruch F, Wietholt D, Scheld HH, Breithardt G. A prospective randomized cross-over comparison on mono- and biphasic defibrillation using nonthoracotomy lead configurations in humans. J Cardiovasc Electrophysiol. 1994;5:581590.[Medline] [Order article via Infotrieve]
6.
Gliner BE, Lyster TE, Dillion SM, Bardy GH.
Transthoracic defibrillation of swine with monophasic and
biphasic waveforms. Circulation. 1995;92:16341643.
7.
Bardy GH, Marchlinski FE, Sharma AD, Worley SJ, Luceri
RM, Yee R, Halperin BD, Fellows CL, Ahern TS, Chilson DA, Packer DL,
Wilber DJ, Mattioni TA, Reddy R, Kronmal RA, Lazzara R, for the
Transthoracic Investigators. Multicenter comparison of
truncated biphasic shocks and standard damped sine wave monophasic
shocks for transthoracic ventricular
defibrillation. Circulation. 1996;94:25072514.
8. Gurvich NL, Markarychev VA. Defibrillation of the heart with biphasic electrical impulses. Kardiologiia. 1967;7:109112.
9. Greene HL, DiMarco JP, Kudenchuk PJ, Scheinman MM, Tang ASL, Reiter MJ, Echt DS, Chapman PD, Jazayeri MR, Chapman FW, Ahmed M, Johnson JL, Niskanen RA, Investigators TBWD. Comparison of monophasic and biphasic defibrillating pulse waveforms for transthoracic cardioversion. Am J Cardiol. 1995;75:11351139.[Medline] [Order article via Infotrieve]
10. Weaver WD, Hill D, Fahrenbruch CD. Use of the automatic external defibrillator in the management of out-of-hospital cardiac arrest. N Engl J Med. 1988;319:661668.[Abstract]
11. White RD, Asplin BR, Bugliosi TF, Hankins DG. High discharge survival rate after out-of-hospital ventricular fibrillation with rapid defibrillation by police and paramedics. Ann Emerg Med. 1996;28:480485.[Medline] [Order article via Infotrieve]
12. Edmark KW, Thomas G, Jones TW. DC pulse defibrillation. J Thorac Cardiovasc Surg. 1966;3:326333.
13. Tacker WA Jr, Geddes LA. Defibrillation with waveforms other than damped sine wave. In: Electrical Defibrillation. Boca Raton, Fla: CRC Press Inc; 1980:5989.
14. Walcott GP, Walker RG, Krassowska W, Smith WM, Ideker RE. Choosing the optimum monophasic and biphasic waveforms for defibrillation. Pacing Clin Electrophysiol. 1994;17:789. Abstract.
15. Gliner BE, Murakawa Y, Thakor NV. The defibrillation success rate versus energy relationship, I: curve fitting and the most efficient defibrillation energy. Pacing Clin Electrophysiol. 1990;13:326338.[Medline] [Order article via Infotrieve]
16. Echt DS, Barbey JT, Black JN. Influence of ventricular fibrillation duration on defibrillation energy in dogs using bidirectional pulse discharges. Pacing Clin Electrophysiol. 1988;11:13151323.[Medline] [Order article via Infotrieve]
17. Bardy GH, Ivey TD, Allen M, Johnson G. A prospective, randomized evaluation of effect of ventricular fibrillation duration on defibrillation thresholds in humans. J Am Coll Cardiol. 1989;13:13621366.[Abstract]
18.
Jones JL, Swartz JF, Jones RE, Fletcher R. Increasing
fibrillation duration enhances relative asymmetrical biphasic versus
monophasic defibrillator waveform efficacy. Circ Res. 1990;67:376384.
19. Fujimura O, Jones DL, Klein GJ. Effects of time to defibrillation and subthreshold preshocks on defibrillation success in pigs. Pacing Clin Electrophysiol. 1989;12:358365.[Medline] [Order article via Infotrieve]
20.
Winkle RA, Mead H, Ruder MA, Smith NA, Buch WS,
Gaudiani VA. Effect of duration of ventricular fibrillation
on defibrillation efficacy in humans. Circulation. 1990;81:14771481.
21. Weaver WD, Cobb LA, Hallstrom AP, Fahrenbruch C, Copass MK, Ray R. Factors influencing survival after out-of-hospital cardiac arrest. J Am Coll Cardiol. 1986;7:752757.[Abstract]
22.
Weisfeldt ML, Kerber RE, McGoldrick RP, Moss AJ, Nichol
G, Ornato JP, Palmer DG, Riegel B, Smith SC Jr. Automatic External
Defibrillation Task Force. American Heart Association report on the
public access defibrillation conference December 810, 1994 [see
comments]. Circulation. 1995;92:27402747.
23.
Emergency Cardiac Care Committee and Subcommittees,
American Heart Association. Guidelines for cardiopulmonary
resuscitation and emergency cardiac care, III: adult advanced cardiac
life support. JAMA. 1992;268:21992241.
24.
Schuder JC, Rahmoeller GA, Stoeckle H.
Transthoracic ventricular defibrillation with
triangular and trapezoidal waveforms. Circ Res. 1966;19:689694.
25. Hillsley RE, Walker RG, Swanson DK, Rollins DL, Wolf PD, Smith WM, Ideker RE. Is the second phase of a biphasic defibrillation waveform the defibrillating phase? Pacing Clin Electrophysiol. 1993;16:14011411.[Medline] [Order article via Infotrieve]
26. White JB, Walcott GP, KenKnight BH, Wayland JL, Smith WM, Ideker RE. Optimal monophasic waveforms can be predicted for transthoracic defibrillation utilizing mathematical modeling. Pacing Clin Electrophysiol. 1996;19:679. Abstract.
27. Sweeney RJ, Gill RM, Jones JL, Reid PR. Defibrillation using a high-frequency series of monophasic rectangular pulses: observations and model predictions. J Cardiovasc Electrophysiol. 1996;7:134143.[Medline] [Order article via Infotrieve]
28. Walcott GP, Walker RG, Cates AW, Krassowska W, Smith WM, Ideker RE. Choosing the optimal monophasic and biphasic waveforms for ventricular defibrillation. J Cardiovasc Electrophysiol. 1995;6:737750.Recently, there has been increased interest in using biphasic waveforms for external defibrillation. Little work has been done, however, in measuring transthoracic defibrillation efficacy after long periods of ventricular fibrillation. These 2 animal protocols examined the relative efficacy of a quasi-sinusoidal biphasic waveform, a truncated exponential biphasic waveform, and a critically damped sinusoidal monophasic waveform. Defibrillation efficacy was measured after 15 seconds of fibrillation and 5 minutes of fibrillation. The results reported here show that both biphasic waveforms have a greater efficacy than the monophasic waveform after 15 seconds of fibrillation. Furthermore, the quasi-sinusoidal biphasic waveform had a greater efficacy than the monophasic waveform after 5 minutes of fibrillation.[Medline] [Order article via Infotrieve]
This article has been cited by other articles:
![]() |
F. Qu, L. Li, V. P. Nikolski, V. Sharma, and I. R. Efimov Mechanisms of superiority of ascending ramp waveforms: new insights into mechanisms of shock-induced vulnerability and defibrillation Am J Physiol Heart Circ Physiol, August 1, 2005; 289(2): H569 - H577. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. M Sado and C. D Deakin How good is your defibrillation technique? J R Soc Med, January 1, 2005; 98(1): 3 - 6. [Full Text] [PDF] |
||||
![]() |
K. A. Mowrey, Y. Cheng, P. J. Tchou, and I. R. Efimov Kinetics of defibrillation shock-induced response: design implications for the optimal defibrillation waveform Europace, January 1, 2002; 4(1): 27 - 39. [Full Text] [PDF] |
||||
![]() |
T. H. Everett IV, J. R. Moorman, L.-C. Kok, J. G. Akar, and D. E. Haines Assessment of Global Atrial Fibrillation Organization to Optimize Timing of Atrial Defibrillation Circulation, June 12, 2001; 103(23): 2857 - 2861. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Marenco, P. J. Wang, M. S. Link, M. K. Homoud, and N. A. M. Estes III Improving Survival From Sudden Cardiac Arrest: The Role of the Automated External Defibrillator JAMA, March 7, 2001; 285(9): 1193 - 1200. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. H. Tovar and J. L. Jones Electrophysiological Deterioration During Long-Duration Ventricular Fibrillation Circulation, December 5, 2000; 102(23): 2886 - 2891. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. T. Niemann, D. Burian, D. Garner, and R. J. Lewis Monophasic versus biphasic transthoracic countershock after prolonged ventricular fibrillation in a swine model J. Am. Coll. Cardiol., September 1, 2000; 36(3): 932 - 938. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. T. Leng, N. A. Paradis, H. Calkins, R. D. Berger, A. C. Lardo, K. C. Rent, and H. R. Halperin Resuscitation After Prolonged Ventricular Fibrillation With Use of Monophasic and Biphasic Waveform Pulses for External Defibrillation Circulation, June 27, 2000; 101(25): 2968 - 2974. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |