From Karolinska Hospital, Stockholm, Sweden (M.R.); Ruprecht-Karls
University of Heidelberg, Heidelberg, Germany (T.B.); Westfälische
Wilhelms Universität, Münster, Germany (M.B.); Academic Hospital
Maastricht, the Netherlands (K.d.D.); and Bakken Research Center, Maastricht,
the Netherlands (J.M., F.L.).
Correspondence to Mårten Rosenqvist, MD, Department of Cardiology, Karolinska Hospital, 171.76 Stockholm, Sweden. E-mail mro{at}cardio.ks.se
Methods and ResultsAdverse events were monitored during
prospective clinical evaluation of the Medtronic model 7219 Jewel ICD
and were classified according to the definitions of the ISO 14155
standard for device clinical trials into 3 groups: severe and mild
device-related and severe nondevice-related adverse events. In
addition, events were related to the surgical procedure, treatment with
the device, or cardiac function. Seven hundred seventy-eight patients
were followed up for an average of 4.0 months after ICD implantation.
In total, 356 adverse events were observed in 259 patients. At 1, 3,
and 12 months after ICD implantation, 99%, 98%, and 97% of the
patients, respectively, survived; 95%, 93%, and 92%, respectively,
were free of surgical reintervention; and 79%, 68%, and 51%,
respectively, were free of any adverse event. Twenty patients died: 6
deaths were related to the surgical procedure, 12 deaths were
considered unrelated to ICD treatment, and 2 patients died of an
unknown cause. Of 111 nonlethal severe adverse device effects, 47
required surgical intervention, 19 times for correction of a dislodged
lead. Inappropriate delivery of therapy was observed 128 times in 111
patients, and the events were typically resolved by reprogramming or
drug adjustment. Nine of these required rehospitalization.
ConclusionsApproximately 50% of patients experience an adverse
event within the first year after ICD implantation. The observed
adverse event rate depends on the definitions and the prospective
monitoring. The incidence of inappropriate therapy emphasizes the need
for improved detection algorithms and for quality-of-life evaluations,
especially when considering ICD treatment in high-risk but
arrhythmia-free patients.
The European Community and the International Standards Organization
(ISO) have provided a standard5 6 for
strict classification of adverse events observed during trials with
implantable medical devices, defining an adverse event as any
undesirable clinical occurrence and taking into account the severity as
well as the relationship to the implanted device. It does not, however,
give any information regarding the underlying technical or clinical
cause and whether or not the event is related to the surgical
implantation or to test procedures. In 1993, a policy
statement7 on ICD patient outcome was published
by the North American Society of Pacing and Electrophysiology. This
document recommends classification of deaths into cardiac and
noncardiac. Cardiac deaths are further classified into sudden and
nonsudden. In addition, it requested that outcome reports should
include all major ICD-related morbidity.
To better understand which adverse events are caused by defibrillator
treatment in general and which are related to the surgical procedure,
we expanded the current classification system and used it prospectively
for all adverse events observed during a multicenter clinical trial of
a new ICD system implanted in 778 patients.
The aim of this study was to report the incidence of adverse events
during the initial months after abdominal or pectoral implantation of
an ICD system with a transvenous lead system. Patients undergoing ICD
replacements were excluded from this analysis.
Adverse events have thus been classified as either severe adverse
device effects or severe nondevice-related events or mild adverse
device effects. Mild, nondevice-related events have not been
considered in this study.
To give more specific information about the events, a subclassification
was applied (see Figure 1
1. The relationship of the event to the ICD therapy, eg,
procedure-related or not, ICD-therapyrelated or not.
A procedure-related event is defined as being directly or indirectly
caused by the surgical implantation procedure.
An ICD-therapyrelated event is an event related to the implanted ICD
system, including leads and nonimplantable equipment used but excluding
procedure-related events.
2. Temporal relation of the event to the surgical procedure, ie, within
(perioperative) or after
(nonperioperative) 30 days after implantation.
3. Mechanism of the event, ie, cardiac or noncardiac, for
nondevice-related events and arrhythmia-related or not.
Deaths were also classified as sudden or nonsudden. Sudden death was
defined as a death occurring without preceding symptoms or within 1
hour after the onset of or sudden change in
symptoms.8 9 An unexpected, insufficiently
documented, and unwitnessed death was conservatively also classified as
sudden.
Adverse Event Committee
Patients
Devices
Implantation, Perioperative Testing, Follow-Up, and
Reporting of Adverse Events
Patients underwent a predischarge test 3 to 7 days after implantation
with induction of ventricular fibrillation (VF). After
discharge, patients visited the outpatient clinic after 1 month and
then every 3 months or in case of any adverse event.
Adverse events were reported to the sponsor on adverse-event forms
designed for efficient reporting of the essential aspects of the event.
In addition, all data forms and ICD interrogation printouts were
screened for unnoticed adverse events by the clinical monitor.
Statistics
Ethics
In total, 356 adverse events were reported in 259 of 778 patients
(Figure 1
The percentage of adverse events reported per hospital as a function of
the number of implantations performed varied widely, but when the
centers with fewer than 5 implantations were excluded (n=29), an
average ratio of 48.8% was observed during the average follow-up
period of 4.0 months. No significant linear relationship was found
between the number of adverse events per patient and the number of
implantations in a center (P=0.59).
Deaths
The first patient was 71 years old, had sustained a myocardial
infarction, and had a left ventricular ejection fraction of
25% (NYHA functional class IV). During 6 months of follow-up, he had
several episodes of monomorphic VT, which were successfully treated,
mainly with ATP. On the day of death, 4 episodes occurred (3 were VF
with a cycle length <240 ms, and 1 was a VT); all were successfully
treated by the first therapy. The device was explanted, and
analysis disclosed no malfunction. The timing of the 4 episodes
stored in the ICD memory revealed no distinct relationship with the
timing of death.
The second patient was 70 years old and had a history of cardiac
arrest. His left ventricular ejection fraction was 10%
(NYHA class III). The patient was inducible, but during follow-up no
spontaneous episodes were observed. At the 1-month follow-up, the
device worked according to specifications. The patient died out of
hospital, unwitnessed. He was found without respiration but with his
pupils still reacting to light and with clinical signs of
electromechanical dissociation. The ICD was neither interrogated nor
explanted.
Six of the deaths were classified as ICD-procedurerelated: sepsis
(n=3; 2 were perioperative, 1 patient died 42 days
after implantation), nonperioperative
hemodynamic compromise (n=2), and
perioperative thromboembolism (n=1). Another 11
patients died of progressive heart failure (3 during the
perioperative phase), and the deaths were classified as
cardiac. One patient died of terminal cancer
(nonperioperative, classified as noncardiac).
NonProcedure-Related, Device-Related Adverse Events
Severe Events
T-wave oversensing, seen in 5 patients, and double sensing of the QRS
complex resulting in inappropriate therapy were resolved by
reprogramming the sensitivity during hospitalization. Decreased sensing
efficacy was corrected in 2 patients by RV-lead repositioning.
Undersensing with no detection of an ongoing true arrhythmia
was never reported during the study.
In 9 patients, inappropriate detection of VT or VF with subsequent
therapy delivery was reported; these events were classified as severe
because it was necessary to hospitalize the patient or to reprogram the
ICD on an urgent basis to avoid in-hospital monitoring of the
patient. In 1 patient, AV-node ablation was performed to resolve the
event.
Loss of capture/elevated pacing thresholds and failure to terminate an
appropriately detected VT/VF episode with all 4 available therapies
were reported in 5 and 6 patients, respectively. These events required
either hospitalization or emergency reprogramming of the ICD; in 1
patient, the RV lead was repositioned to ensure pacing capture.
In 8 patients, hospitalization was reported to allow adjustment of
medical regimen, to reprogram the ICD, or just to alleviate fear of
shocks.
One patient was hospitalized after the induction of atrial fibrillation
at the 3-month follow-up visit during performance of a
VF-defibrillation test.
Mild Events
The category of mild events also includes 19 events related to the
external implant support equipment. An unacceptably long capacitor
charge period was observed once before ICD implantation and atrial
fibrillation was induced 5 times because of delivery of
protocol-required test shocks.
Procedure-Related Adverse Events
Centers with more than 50 patients enrolled in the study (n=4) had an
average ratio of the number of "surgery-requiring events" to the
number of patients of 0.067, and centers with 5 to 10 implantations
(n=18) had a ratio of 0.066: no significant linear relation was found
between these parameters (P=0.29).
Twenty-seven wound problems (seroma, hematoma; infections excluded)
were the most common procedure-related adverse events.
The second most common adverse event was lead/ICD-can dislodgment,
which was observed in 23 instances; 1 patient experienced dislodgment
twice. The electrode most often dislodged was the RV lead (n=16),
observed primarily with abdominal ICD implants11;
the pectorally implanted ICD was repositioned in 2 patients.
Infection occurred in 6 patients, of whom 3 died of this condition.
General antibiotic prophylaxis was prescribed in 87% of participating
centers, whereas 4% of the centers prescribed it only to high-risk
patients. All patients who died of an infection, however, had received
antibiotic prophylaxis.
Four thromboembolic events were observed during follow-up. Aspirin
treatment was prescribed in 3 of these patients, and none received
coumarin or warfarin. Twenty-five percent of all patients
received aspirin, and 10% received warfarin/coumarin after
implantation, during at least the first month of follow-up.
Severe Adverse Events Not Related to ICD Treatment
NonDevice-Related Adverse Events of Unknown Cause
The European Community and ISO5 6 defined an
adverse event as any undesirable clinical occurrence, whether
device-related or not. Prospective monitoring in 778 patients resulted
in
Patient Deaths
Sudden cardiac death, occurring within 1 hour and after a sudden change
of symptoms, was observed in 22 patients. Procedure-related deaths
occurred in 7 instances. As in the present study, none were
considered to be related to the ICD treatment.
Pratt et al9 reported 109 deaths among 834
implant recipients during an average follow-up period of 20±13 months.
Sixty-eight were cardiac, and 17 were classified as sudden.
Classification depended on the information available in autopsy
reports. Without autopsy reports (29/109=27%), the rate of sudden
death would have increased from 2.0% to 2.9%. In the present
study, 4 of the 20 deaths were documented by an autopsy report.
Obviously, the availability of these autopsy reports did not influence
the classification of the deaths.
Therefore, the data from the present study and those from Epstein
et al8 and Pratt et al9
seem to agree that most deaths were cardiac and not related to device
treatment. Despite the availability of approximately equal rates of
autopsy reports, there was a lower incidence of sudden death but a
higher incidence of procedure-related deaths in the present study.
This discrepancy cannot be clarified completely from the available
data.
Trappe et al12 reported that 13 of 335 patients,
with an average of 22 months of follow-up, developed an infection after
ICD implantation; 7 patients died.
Even though in this study, the majority of the patients received
antibiotic prophylaxis, 3 patients died of infection. By definition,
all these events were classified as "ICD-procedurerelated"
events, although no direct link to the surgical implantation procedure
itself was demonstrated. The occurrence of these 3 fatal events, as
well as another 2 severe and 1 mild event, emphasizes that
postoperative infections causing sepsis or endocarditis are not
uncommon and carry an ominous prognosis in a defibrillator population;
every effort should be undertaken to provide adequate preventive
measures. Mounsey et al13 showed a significant
beneficial effect in patients receiving antibiotic prophylaxis before
pacemaker implantation.
Procedure-Related Adverse Events
The observed incidence of lead dislodgments (10%) is lower than
observed in other studies,14 15 possibly related
to a shorter follow-up period and/or fewer patients receiving
implantation of subcutaneous patches. As reported by Hoffmann et
al,11 the dislodgments were found to be related
primarily to abdominal rather than pectoral implants, and in the former
case primarily to RV-lead dislodgments.
Device-Related Events
Inappropriate therapy, apart from being painful, can also be
proarrhythmic. Pinski and Fahy4 reported
Ideally, the risk of experiencing a supraventricular
tachycardia should be assessed before implantation of an
ICD. However, Schmitt et al21 demonstrated that
this arrhythmia was present before implantation in only
35% of patients receiving inappropriate therapy due to atrial
fibrillation, and Page et al22 showed that most
episodes of paroxysmal atrial fibrillation tend to be
asymptomatic.
Weber et al23 reported an 18% incidence of
inappropriate therapy in 462 patients followed up for an average of 24
months. Preoperatively documented atrial fibrillation, a maximum heart
rate during exercise close to the detection interval, and a low
detection rate were found to be independent risk factors for
inappropriate therapy.
Various algorithms have been developed to avoid false detection, and
these seem to distinguish supraventricular from
ventricular arrhythmias relatively
well,24 25 although they do have certain
limitations. For example, the "sudden-onset" criterion will not
detect an exercise-induced sinus tachycardia converting
into a slow ventricular tachycardia (VT), and
it cannot distinguish a suddenly started, fast-conducting regular
atrial flutter from a VT.26 The "rate stability
criterion" will have difficulty in rejecting atrial fibrillation with
a fast and rather regular ventricular response and may
delay detection of VTs with irregular intervals. These limitations and
the lack of controlled studies evaluating the specificity, sensitivity,
and safety of these algorithms have probably contributed to their
infrequent use.
Another important reason for inappropriate detection is that
antiarrhythmic treatment in ICD patients tends to cause an overlap in
rate between VTs and the sinus rate. In a study by Paul et
al,27 only 11% of patients without
antiarrhythmic medication had sinus rates exceeding the rate of VT,
whereas this proportion increased to 35% in patients with a single
antiarrhythmic drug and to 63% of patients on combination therapy.
This emphasizes the need for improved detection algorithms in the next
generation of ICDs, especially if indications should be widened to
include prophylactic treatment in high-risk patients and
efforts are undertaken to evaluate changes in quality of life in ICD
patients.
Sensing Problems
In line with the report of Callans et al,14 most
sensing problems were overcome by reprogramming.
Limitations of the Study
Mild, nondevice-related adverse events were not collected in a
systematic way and therefore have not been included in this report.
Another limitation of this study is that the protocol registered data
only on medical and technical events. Thus, it was not possible to
assess the incidence of psychosocial problems related to ICD therapy,
which are well known to occur in this patient group and which should be
considered especially in comparisons of ICD therapy with other
therapies.
As with any classification system, some shortcomings and clinical
irrelevancies are present.
1. The hospitalization of a patient, making the adverse event by
definition "severe," does not always reflect the severity of the
clinical impact; for instance, several patients were hospitalized once
and 1 patient several times for 1 or 2 days to alleviate fear of shocks
that he had not, as yet, experienced.
2. Adverse events that could have but had not yet resulted in severe
outcomes such as an RV perforation, a failure of the last therapy to
treat a VT or a VF episode, or a lead dislodgment were classified as
"mild" because they were left untreated or were resolved by simple
reprogramming or adjustment of the antiarrhythmic medication but did
not cause hospitalization or a prolongation of the hospital stay.
Because of the eventual "severe" outcome these events could result
in, their classification as severe can be argued.
Conclusions
Received July 28, 1997;
revision received March 19, 1998;
accepted April 20, 1998.
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Adverse Events With Transvenous Implantable Cardioverter-Defibrillators
A Prospective Multicenter Study
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundA newly developed
classification system relates adverse events to the surgical procedure
or the function of the implantable defibrillator.
Key Words: events, adverse tachyarrhythmias defibrillation implantable cardioverter defibrillator (ICD)
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
It has been
suggested that indications for ICDs should be expanded to include
prophylactic treatment of patients at high risk for
ventricular arrhythmias1;
then it becomes important to assess the risk for side effects from this
treatment more strictly, especially regarding the most common problem,
ie, inappropriate therapy due to detection of
supraventricular
tachycardia,2 3 which sometimes can
be proarrhythmic and even fatal.4
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
Classification of Events
According the European standards,5 6 an
"adverse event" is "any undesirable clinical occurrence in a
subject whether it is considered to be device related or not." An
"adverse device effect" is defined as a device-related adverse
event. Adverse events can be severe or mild; severe events cause
hospitalization or undue prolongation of hospitalization because of
potential disability or danger to life, necessitate intervention, or
cause death. For this study, all other events have been labeled
"mild," and "intervention" has been interpreted as either
surgical intervention or reprogramming of the device with the specific
purpose of avoiding an otherwise required hospitalization.
) that
considered the following.

View larger version (34K):
[in a new window]
Figure 1. Classification tree. A total of 356 adverse events
were observed in 259 of 778 patients during an average follow-up period
of 4.0 months.
Four experienced cardiologists and 2
representatives of the manufacturing company
scrutinized and classified all events by consensus. The company
representatives ensured access to all clinical data and
supported technical interpretation of events but did not have any
impact on the final event classification.
Seven hundred seventy-eight patients (demographics listed in
Table 1
) received implants between March
1993 and November 1994 in 63 European centers, with either Medtronic
model 7219 C (392 patients) or model 7219 D (386 patients) Pacer
Cardioverter Defibrillator.
View this table:
[in a new window]
Table 1. Clinical Characteristics of 778 Patients Receiving a
Model 7219 ICD
The ICDs provide defibrillation and cardioversion shocks from
0.4 to 34 J, antitachycardia pacing therapies (burst, ramp
and ramp +), noninvasive
electrophysiological test procedures, and 5
seconds of electrogram storage and have a size and weight that allow
pectoral implantation (80 cm3 and 129 g for
model 7219 C). All ICDs were implanted with a tripolar, right
ventricular (RV) screw-in lead (model 6936, bipolar
pace/sense, 5-cm defibrillation coil). For model 7219 C, this was the
only lead connected because this device uses the generator can as the
second defibrillation electrode. Model 7219 D requires an additional
lead in the superior vena cava for which either model 6933 (5-cm
electrode coil) or model 6937 (8 cm coil) was used. If defibrillation
threshold was unacceptably high, model 7219 D was connected with a
third, subcutaneous (model 6939, 58 patients) or epicardial (model
6721, 6 patients) lead. Model 7219 C was implanted pectorally in the
subcutaneous (9%) or submuscular (88%) position (3% unknown),
whereas the 7219 D version was implanted either abdominally or
pectorally (40/60 ratio).
After the leads were positioned, the efficacy of pacing,
sensing, and defibrillation was assessed.
Data are given as mean±SD. To account for variable lengths
of follow-up, the probability of remaining event free was
analyzed by the Kaplan-Meier method.10
Student's t test was used for comparison of
continuous variables, and a
2 test was
used for comparison of categorical variables. A generalized linear
model analysis was performed to relate occurrence rates of
adverse events with implant rates in individual centers.
All patients received oral and written information and gave
voluntary written consent to participate in the study. The study
protocol was approved by the local ethics committee of every
participating hospital.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
Patients and Follow-Up
The average duration of follow-up was 4.0 months, with an SD of
4.6 months and a range of 0 to 21 months.
; Table 1
). Figure 2
presents the rates of freedom from death, surgical reintervention,
severe adverse effects, and any adverse events at 1, 3, 6, and 12
months after implantation as calculated with Kaplan-Meier
analysis.

View larger version (12K):
[in a new window]
Figure 2. Kaplan-Meier curves. Adverse event-free survival
estimates: overall survival, survival without additional surgery
performed, survival without occurrence of any "severe" event, and
survival without occurrence of any adverse event. 95% CIs are given at
1-year follow-up.
Twenty patients died during the study period; 6 died
perioperatively (1 pulmonary embolism, 2
sepsis, and 3 cardiac failure) (Table 2
).
All deaths except 2 were witnessed in the hospital. Two deaths were
classified as "unknown and occurring suddenly" because they were
unwitnessed, insufficient information was available, and they occurred
outside the hospital.
View this table:
[in a new window]
Table 2. Causes of Death
In all, 214 adverse events in 161 patients were related to
treatment with an ICD (Table 3
). Of
these, 45 events in 38 patients were considered severe, of which 8
required an additional surgical intervention, and 169 events in 143
patients were classified as mild.
View this table:
[in a new window]
Table 3. Severe and Mild Adverse Events Due to the ICD
System
Eight patients were hospitalized after observation of an increase
in defibrillation energy requirements, but the implanted lead system
was modified subsequently in only 4 of them.
The event with the highest incidence rate was "inappropriate
detection and subsequent therapy delivery" and was observed 119 times
in 102 patients. Of 294 patients who received ICD therapy during the
follow-up period, 116 patients had
1 inappropriate detections of
supraventricular arrhythmias (pooled mild and
severe events). In 49% of these patients, this problem was solved by
reprogramming, and in 21%, by a change in medication. In 1%, other
actions were taken, but in 29%, no specific action at all was
taken.
Altogether, 102 adverse events related to the surgical procedure
were observed in 95 patients; 6 resulted in death (Table 4
).
Kaplan-Meier analysis showed a 12-month surgery-free survival
rate of 92.4% (Figure 2
).
View this table:
[in a new window]
Table 4. Adverse Events Related to the Surgical Procedure
Altogether, 40 adverse events not related to ICD treatment were
registered (Table 5
). The majority, 31,
were cardiac in origin. Fourteen events caused the patient's death,
nearly exclusively because of congestive heart failure (n=11). Twenty
nonlethal events were registered: 14 were due to incessant VT/VF, 5 to
acute myocardial infarction or unstable angina, and 1 to the occurrence
of an unspecified type of supraventricular
arrhythmia. Six nondevice-related events were also found not
to be cardiac in origin.
View this table:
[in a new window]
Table 5. NonDevice-Related Severe Adverse Events
One patient experienced dyspnea requiring hospitalization (Table 5
). The event remained unexplained despite extensive medical
examination. The symptoms resolved spontaneously 1 month later.
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
Adverse events, prospectively monitored in patients treated with
third-generation ICDs, were described and classified for the first time
according to the new European Community and ISO standards for
clinical trials on implantable medical devices. Despite technology
enhancements, adverse events were observed in 259 of the 778 patients:
merely inappropriate detection, and procedure-related events. Efforts
should be undertaken to minimize these, especially if ICD therapy
should be advocated for prophylactic use in high-risk
patients.
0.1 event per patient per follow-up month during the initial
months after ICD implantation. Actuarial analysis showed that
adverse event occurrence rate is highest during the initial 1 to 3
months after ICD implantation. Clearly, the definitions used influence
the total number of reportable events. With the new classification
system, more details of the implantable device under evaluation are
monitored, giving better insight into the clinical and technological
aspects.
In a recent study by Epstein et al,8 a
classification system for deaths similar to our system was developed to
more specifically characterize deaths in ICD trials. This system was
used in a clinical trial comprising 1250 ICD patients: altogether, 79
deaths were observed.
Despite single defibrillation leads and smaller devices, the
incidence of wound problems and lead dislocations is still too high.
The anticipated higher incidence of wound-related problems in pectoral
ICD implantations was not observed in this study, but improvements of
surgical technique could further reduce the incidence.
Inappropriate Therapy
The high incidence of inappropriate therapy is in line with other
reports emphasizing that inappropriate detection/therapy is the most
common adverse event observed among ICD
patients.2 3 With "shock-only" devices, the
incidence of these events during long-term follow-up has ranged between
16% and 21%.16 17 Fromer et
al18 reported inappropriately detected episodes
in 10 of 102 patients with implants during an average follow-up period
of 9.4 months. In the present study, 116 of the 778 patients with
implants or of the 294 patients with a therapy delivered experienced
inappropriate episode detection followed by a therapy, despite a short
follow-up time. This higher incidence rate might be related to the
possibility of retrospectively classifying the episodes on the basis of
the stored intracardiac electrograms. Most events were mild and were
resolved by reprogramming or by changing the drug regimen.
36
patients with ICD-induced arrhythmias, of which 2 caused the
patient's death (range, 0.4% to 8%). The risk for
proarrhythmia seems to be most common in situations with fast
supraventricular rhythms with a concomitant increase in
sympathetic discharge, rate-related decrease in ventricular
refractoriness, or myocardial
ischemia.19 20
T-wave oversensing and double sensing of QRS complexes may lead to
inappropriate arrhythmia detection. The observation of 6 such
events emphasizes the need for regular, careful evaluation of
oversensing during normal sinus rhythm and during pacing.
This report relies on adequate data reporting by the individual
investigators. The adverse event rate per patient differed
substantially between centers. There was no relation, however, between
this ratio and the number of implantations at each center. The study
protocol did not require defibrillation threshold testing at follow-up,
nor were chest radiographs mandatory at each follow-up visit, which,
together with the absence of other potential tests, may have led to
underestimation of actual adverse events.
The total number of events reported in a given study clearly
depends on the definitions used. The definitions of the European
guidelines (EN 540) and of the International Organization of
Standardization (ISO 14155) provide guidance for monitoring and
classification of adverse events. Subclassification of events is
advisable to evaluate the safety and efficacy aspects of the new device
and to emphasize clinically relevant aspects of the events.
![]()
Acknowledgments
The authors would like to acknowledge the following contributors
to the study: E. Aliot, Nancy; J.M. Almendral, Madrid; J. Amlie,
Oslo; D. Andresen, Berlin; Y. Bashir, Oxford; J. Ben-David, Haifa; J.
Brachmann, Heidelberg; P. Brugada, Aalst; C. Bucknall, London; A.J.
Camm, London; R. Candinas, Zürich; P. Coumel, Paris; C. Cowan,
Leeds; J.-C. Daubert, Rennes; D.W. Davies, London; L. De Roy, Yvoir;
S. Favale, Bari; S. Fischer, Augsburg; M. Fromer, Lausanne; F.
Furlanello, Trento; G. Gadaleta, Milano; B.D. Gonska, Göttingen;
G. Grimm, Klagenfurt; W. Grimm, Marburg; B. Grosch, Essen; H. Heuer,
Dortmund; E. Hoffmann, Munich; R. Hofman, Linz; R. Höltgen,
Duisburg; L. Jordaens, Gent; W. Jung, Bonn; A. Katz, Beer Sheba; W.
Klein, Graz; V. Kühlkamp, Tübingen; B. Lemke,
Bochum; C. Menozzi, Reggio Emilia; C. Moro Serrano, Madrid;
N. Muzika, Vienna; J. Nesser, Linz; I.W.P. Obel, Johannesburg; O.-J.
Ohm, Bergen; C. Pappone, Napoli; V. Paul, Uxbridge; A. Pietersen,
Copenhagen; A. Pisapia, Marseille; A. Podczeck, Vienna; J.
Potratz, Lübeck; P. Ricard, Marseille; P. Scanu, Caen; A.
Schirdewan, Berlin; H. Schmidinger, Vienna; C. Schmitt, Munich;
K. Seidl, Ludwigshafen; J. Siebels, Hamburg; J.D. Skehan, Leicester;
C. Stellbrink, Aachen; P. Touboul, Lyon; H.J. Trappe, Hannover; J.
Tröster, Oldenburg; B. Waldecker, Giessen; A. Waleffe,
Liège; K. Walsh, Liverpool; and B. Wille, Kiel.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
This article has been cited by other articles:
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons J. Am. Coll. Cardiol., May 27, 2008; 51(21): e1 - e62. [Full Text] [PDF] |
||||
![]() |
A. E. Epstein, J. P. DiMarco, K. A. Ellenbogen, N.A. M. Estes III, R. A. Freedman, L. S. Gettes, A. M. Gillinov, G. Gregoratos, S. C. Hammill, D. L. Hayes, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): Developed in Collaboration With the American Association for Thoracic Surgery and Society of Thoracic Surgeons Circulation, May 27, 2008; 117(21): e350 - e408. [Full Text] [PDF] |
||||
![]() |
J. P. Daubert, W. Zareba, D. S. Cannom, S. McNitt, S. Z. Rosero, P. Wang, C. Schuger, J. S. Steinberg, S. L. Higgins, D. J. Wilber, et al. Inappropriate Implantable Cardioverter-Defibrillator Shocks in MADIT II: Frequency, Mechanisms, Predictors, and Survival Impact J. Am. Coll. Cardiol., April 8, 2008; 51(14): 1357 - 1365. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Salerno-Uriarte, G. M. De Ferrari, C. Klersy, R. F.E. Pedretti, M. Tritto, L. Sallusti, L. Libero, G. Pettinati, G. Molon, A. Curnis, et al. Prognostic Value of T-Wave Alternans in Patients With Heart Failure Due to Nonischemic Cardiomyopathy: Results of the ALPHA Study J. Am. Coll. Cardiol., November 6, 2007; 50(19): 1896 - 1904. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Almendral and M. E. Josephson All Patients With Hemodynamically Tolerated Postinfarction Ventricular Tachycardia Do Not Require an Implantable Cardioverter-Defibrillator Circulation, September 4, 2007; 116(10): 1204 - 1212. [Full Text] [PDF] |
||||
![]() |
S. Singh and M. M. Murawski Implantable Cardioverter Defibrillator Therapy and the Need for Concomitant Antiarrhythmic Drugs Journal of Cardiovascular Pharmacology and Therapeutics, September 1, 2007; 12(3): 175 - 180. [Abstract] [PDF] |
||||
![]() |
A. Agard, R. Lofmark, N. Edvardsson, and I. Ekman Views of patients with heart failure about their role in the decision to start implantable cardioverter defibrillator treatment: prescription rather than participation J. Med. Ethics, September 1, 2007; 33(9): 514 - 518. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Ezekowitz, B. H. Rowe, D. M. Dryden, N. Hooton, B. Vandermeer, C. Spooner, and F. A. McAlister Systematic Review: Implantable Cardioverter Defibrillators for Adults with Left Ventricular Systolic Dysfunction Ann Intern Med, August 21, 2007; 147(4): 251 - 262. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Vollmann, L. Luthje, and M. Zabel Unusual cause for an increase of the sensing integrity counter in a patient with inappropriate implantable cardioverter-defibrillator therapy Europace, May 1, 2007; 9(5): 275 - 277. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Gehi, D. Mehta, and J. A. Gomes Evaluation and Management of Patients After Implantable Cardioverter-Defibrillator Shock JAMA, December 20, 2006; 296(23): 2839 - 2847. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. S. Chan, K. Stein, T. Chow, M. Fendrick, J. T. Bigger, and S. Vijan Cost-Effectiveness of a Microvolt T-Wave Alternans Screening Strategy for Implantable Cardioverter-Defibrillator Placement in the MADIT-II-Eligible Population J. Am. Coll. Cardiol., July 4, 2006; 48(1): 112 - 121. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Reynolds, D. J. Cohen, A. D. Kugelmass, P. P. Brown, E. R. Becker, S. D. Culler, and A. W. Simon The Frequency and Incremental Cost of Major Complications Among Medicare Beneficiaries Receiving Implantable Cardioverter-Defibrillators J. Am. Coll. Cardiol., June 20, 2006; 47(12): 2493 - 2497. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. M. Bloomfield, J. T. Bigger, R. C. Steinman, P. B. Namerow, M. K. Parides, A. B. Curtis, E. S. Kaufman, J. M. Davidenko, T. S. Shinn, and J. M. Fontaine Microvolt T-Wave Alternans and the Risk of Death or Sustained Ventricular Arrhythmias in Patients With Left Ventricular Dysfunction J. Am. Coll. Cardiol., January 17, 2006; 47(2): 456 - 463. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. G. Manios, E. M. Kallergis, E. M. Kanoupakis, H. E. Mavrakis, D. C. Kambouraki, D. A. Arfanakis, and P. E. Vardas Amino-Terminal Pro-Brain Natriuretic Peptide Predicts Ventricular Arrhythmogenesis in Patients With Ischemic Cardiomyopathy and Implantable Cardioverter-Defibrillators Chest, October 1, 2005; 128(4): 2604 - 2610. [Abstract] [Full Text] [PDF] |
||||
![]() |
L.J. Jordaens and J.M. Mekel Electrical storm in the ICD era Europace, January 1, 2005; 7(2): 181 - 183. [Full Text] [PDF] |
||||
![]() |
J. C. Senges-Becker, M. Klostermann, R. Becker, A. Bauer, K. E. Siegler, H. A. Katus, and W. Schoels What is the "Optimal" follow-up schedule for ICD patients? Europace, January 1, 2005; 7(4): 319 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A.M.J. Theuns, A. P. J. Klootwijk, D. M. Goedhart, and L. J.L.M. Jordaens Prevention of inappropriate therapy in implantable cardioverter-defibrillators: Results of a prospective, randomized study of tachyarrhythmia detection algorithms J. Am. Coll. Cardiol., December 21, 2004; 44(12): 2362 - 2367. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. M. Bloomfield, R. C. Steinman, P. B. Namerow, M. Parides, J. Davidenko, E. S. Kaufman, T. Shinn, A. Curtis, J. Fontaine, D. Holmes, et al. Microvolt T-Wave Alternans Distinguishes Between Patients Likely and Patients Not Likely to Benefit From Implanted Cardiac Defibrillator Therapy: A Solution to the Multicenter Automatic Defibrillator Implantation Trial (MADIT) II Conundrum Circulation, October 5, 2004; 110(14): 1885 - 1889. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Becker, J. Ruf-Richter, J. C. Senges-Becker, A. Bauer, S. Weretka, F. Voss, H. A. Katus, and W. Schoels Patient alert in implantable cardioverter defibrillators: toy or tool? J. Am. Coll. Cardiol., July 7, 2004; 44(1): 95 - 98. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Wichter, M. Paul, C. Wollmann, T. Acil, P. Gerdes, O. Ashraf, T. D.T. Tjan, R. Soeparwata, M. Block, M. Borggrefe, et al. Implantable Cardioverter/Defibrillator Therapy in Arrhythmogenic Right Ventricular Cardiomyopathy: Single-Center Experience of Long-Term Follow-Up and Complications in 60 Patients Circulation, March 30, 2004; 109(12): 1503 - 1508. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. H. Samore, R. S. Evans, A. Lassen, P. Gould, J. Lloyd, R. M. Gardner, R. Abouzelof, C. Taylor, D. A. Woodbury, M. Willy, et al. Surveillance of Medical Device-Related Hazards and Adverse Events in Hospitalized Patients JAMA, January 21, 2004; 291(3): 325 - 334. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kostopoulou, E. Sbarouni, E. G. Livanis, G. N. Theodorakis, and D. Kremastinos Superior vena cava syndrome and syncope in an implantable cardioverter defibrillator recipient Europace, January 1, 2004; 6(3): 205 - 208. [Abstract] [Full Text] [PDF] |
||||