From the Noninvasive Cardiac Imaging Laboratory (R.H.M., J.B., S.L.,
J.A., R.M.L.), University of Chicago Medical Center, Chicago, Ill;
Cardiovascular Fluid Mechanics Laboratory (R.S.H., A.P.Y.), Georgia Institute
of Technology, Atlanta, Ga; Departments of Pathology (R.B.) and Cardiovascular
Surgery (W.P.), Rush-PresbyterianSt. Lukes' Medical Center,
Chicago, Ill; and Echocardiography Laboratory (B.V., R.E.K.), University of
Iowa Hospital, Iowa City.
Correspondence to Richard H. Marcus, Cardiovascular Ultrasound Laboratories, Iowa Heart Center, 411 Laurel St, Suite 1250, Des Moines, IA 50314-3046.
Methods and ResultsRelations between the orifice
transprosthetic gradient (equivalent to Doppler), the
downstream gradient in the zone of recovered pressure (equivalent to
catheter), and fluid mechanical energy losses were examined in vitro.
Pressure-flow relations across the 2 prostheses were evaluated by
Doppler echocardiography in vivo. For both
types of prosthesis in vitro, the orifice was higher than the
downstream gradient (P<0.001), and fluid mechanical
energy losses were as strongly correlated with orifice as with
downstream pressure gradients (r2=0.99 for
both). Orifice and downstream gradients were higher and fluid
mechanical energy losses were larger for the St Jude than the Medtronic
Hall valve (all P<0.001). Whereas estimated effective
orifice areas for the 2 valves in vivo were not significantly
different, model-independent dynamic analysis of pressure-flow
relations revealed higher gradients for the St Jude than the Medtronic
Hall valve at a given flow rate (P<0.05).
ConclusionsEven in the presence of significant pressure
recovery, the Doppler-derived gradient across small-diameter aortic
mechanical prostheses does have hemodynamic relevance
insofar as it reflects myocardial energy expenditure. Small differences
in function between stenotic aortic mechanical prostheses,
undetectable by conventional orifice area estimations, can be
identified by dynamic Doppler echocardiographic
analysis of pressure-flow relations.
This study was designed to examine these issues. Two types of
small-diameter mechanical aortic prosthesis of different design
(namely, tilting-disk 20-mm Medtronic Hall prosthesis and
bileaflet 19-mm St Jude valve) were used as the vehicles for this
hemodynamic assessment. These valves, both of which are
functionally stenotic even when disk/leaflet excursion is
normal,7 were chosen for their high relative
prevalence among mechanical aortic valves inserted in the United
States8 and their similar
hemodynamic profiles.9
First, an in vitro model was constructed that would simulate in vivo
pressure-flow relations across the 2 prostheses in the aortic position.
The relation between the maximal transprosthetic gradient
measured just distal to the valve orifice (analogous to the Doppler
gradient)3 and the gradient in the zone of
recovered pressure measured further downstream of the orifice
(analogous to the in vivo catheter gradient) was studied. The
physiological relevance of each of these gradients
(ie, orifice and downstream) was evaluated in terms of the associated
fluid mechanical energy losses incurred during transprosthetic
fluid transit for both the St Jude and Medtronic Hall valves. Next, the
physiologically validated Doppler-derived
gradients were determined in vivo and applied to the estimation of
valve orifice areas for the 2 prostheses to ascertain whether or not
the known small hemodynamic differences between the 2
prostheses could be identified. Finally, the 2 prostheses were compared
by a novel model-independent analysis of pressure and flow
data, performed over a range of incremental flows. This method
circumvents the limitations inherent to conventional model-based
orifice area estimations and avoids potentially confounding effects of
low flows.
Energy Loss Measurements
Data Acquisition
In Vivo Study
Doppler Echocardiography
Calculations
Cardiac output (CO) was calculated as the product of stroke volume
and heart rate. The mean transprosthetic gradient was
determined by the extended Bernoulli equation
(MPGB)
The effective prosthetic valve orifice area
(EOAG) was calculated by the Gorlin formula as
The product of CC and
CV has been termed the empiric
constant,13 which approximates a value of 1.
Statistical Analysis
Baseline hemodynamic variables for the 2 mechanical
prostheses in vivo were compared with nonparametric
statistics (Mann-Whitney). A multiplicative model was used to examine
pressure-flow relations. These relations for the 2 prostheses were
compared by ANCOVA, with pressure as the covariate, to identify if
either conferred a hemodynamic advantage over the other
in respect of the transvalvular flow that would result from a
common transvalvular driving pressure.
At all stroke volumes, the orifice gradient was higher than the
downstream gradient (P<0.001), reflecting significant
pressure recovery during fluid transit across both types of
prosthesis. The downstream (recovered) pressure gradient for
the Medtronic Hall valve ranged from 3.7 mm Hg (at a stroke
volume of 40 mL) to 20.5 mm Hg (at a stroke volume of 120 mL). At
the same values for stroke volume, downstream gradients across the St
Jude valve were 4.5 and 38.8 mm Hg, respectively. For all stroke
volumes >40 mL, downstream gradients were significantly higher for the
St Jude than for the Medtronic Hall valve.
Downstream and orifice gradients were strongly correlated for both
prostheses (r2=0.99 for both). For a given
orifice gradient, the downstream gradient tended to be lower for the
Medtronic Hall than the St Jude valve (Figure 3
Fluid Mechanical Energy Losses
Relations Between Transprosthetic Pressure Gradients and
Fluid Mechanical Energy Losses
In Vivo Study
Hemodynamic Comparison of Medtronic Hall and
St Jude Prostheses
For both valves, the power function y=bxa was the
best-fit model for analysis of pressure-flow relations. The
slope (a) of the derived linear logarithmic relation (log x versus log
y) gives the exponentiality of the pressure-flow relation for each
prosthesis. Accordingly, the linear relation between log
cardiac output and log mean pressure gradient was used for quantitative
analytical comparison of pressure-flow relations across the 2 valves
over a wide range of hemodynamic conditions induced by
dobutamine (Figure 7
Physiological Relevance of the Orifice
Gradient
Hemodynamic Assessment of Prosthetic
Valve Function
Aris et al21 recently reported significant
differences in valve orifice area between 19-mm St Jude and 20-mm
Medtronic Hall prostheses. Although valve orifice area values in the
present study were similar to those of Aris et al, differences
between the 2 valves did not reach statistical significance, either at
rest or at peak exercise. Because the value for effective valve orifice
area is derived indirectly from an analysis of pressure-flow
relations, the accuracy of this parameter is entirely
dependent on the accuracy of the pressure and flow measurements.
Differentiation of 1 orifice area from another requires greater
accuracy in gradient (and hence, velocity) measurements when values for
flow are low, irrespective of the model used (Figure 8
A multiplicative model was therefore used to examine pressure-flow
relations, making none of the assumptions that are inherent to valve
orifice area estimation, either with respect to the relation between
pressure and flow or in regard to the coefficients of orifice
contraction and velocity. This dynamic analysis, conducted over
a range of values for flow and pressure gradient, demonstrated that the
head of pressure required to drive a column of blood across the St Jude
valve was significantly greater than that necessary to generate the
same flow across the Medtronic Hall valve. This type of
hemodynamic evaluation appears to detect minor
hemodynamic differences more sensitively than
conventional orifice area estimations that are based on data
traditionally acquired under only a single condition of pressure and
flow. The importance of generating a range of incremental flow values
is highlighted further by the in vitro observation that even under the
most carefully controlled hemodynamic conditions, the
differences in fluid mechanical energy losses across the 2 types of
prosthetic valve were difficult to detect at lower stroke
volume values. The noninvasive imaging laboratory is an ideal
environment in which to perform this type of
physiological evaluation. The time constraints that
are inherent to invasive procedures performed in the cardiac
catheterization laboratory would limit the application
of such studies. Moreover, the noninvasive nature of this approach
lends itself to serial evaluations.
Our data suggest that serial comparisons of families of pressure-flow
relations generated for each hemodynamic condition (ie,
at the time of each Doppler echocardiographic
assessment) will facilitate detection of minor differences or changes
in valve function that may not be detected by estimation of valve
orifice area alone.
Summary
2. The higher proximal gradient, which can be measured either by
continuous-wave Doppler or by a properly positioned manometer,
correlates as well with fluid mechanical energy losses across
small-caliber mechanical prosthetic valves as the lower distal
gradient (in the zone of recovered pressure) measured by conventional
manometric techniques in the catheterization
laboratory. Both measurements therefore reflect the
hemodynamic burden imposed by the valve.
3. The pressure gradient required to drive a given flow volume and the
fluid mechanical energy losses incurred during flow transit are
slightly greater for the 19-mm St Jude than for the 20-mm Medtronic
Hall prosthesis in the aortic position.
4. Small changes in the hemodynamic function of
stenotic aortic valves can be more sensitively detected by
direct analysis of pressure-flow relations or calculation of
fluid mechanical energy losses over a range of incremental flow values
than by conventional valve orifice area estimations.
Received December 12, 1997;
revision received April 1, 1998;
accepted April 20, 1998.
2.
Levine RA, Jimoh A, Cape G, McMillan S, Yoganathan AP,
Weyman AE. Pressure recovery distal to a stenosis: potential
cause of gradient "overestimation" by Doppler
echocardiography. J Am Coll
Cardiol. 1989;13:706715.[Abstract]
3.
Baumgartner H, Khan S, DeRobertis M, Czer L, Maurer G.
Discrepancies between Doppler and catheter gradient in aortic
prosthetic valves in vitro: a manifestation of localized
pressure gradient and pressure recovery. Circulation. 1990;82:14671475.
4.
Vandervoort PM, Greenberg NL, Powell KA, Cosgrove DM,
Thomas JD. Pressure recovery in bileaflet heart valve prostheses:
localized high velocities and gradients in central and side orifices
with implications for Doppler-catheter gradient relation in aortic
and mitral position. Circulation. 1995;92:34643472.
5.
Khan SS. Pitfalls in the interpretation of Doppler
gradients in prosthetic valves. Clin Cardiol. 1992;18:2328.
6.
Cannon SR, Richard KL, Crawford M. Hydraulic
estimation of stenotic orifice area: a correction of the Gorlin
formula. Circulation. 1985;71:11701178.
7.
Walker PG, Yoganathan AP. In vitro pulsatile flow
hemodynamics of five mechanical aortic heart valve
prostheses. Eur J Cardiothorac Surg. 1992;6(suppl
I):I-113I-123.
8.
Mendenhall S. Heart valves: a mature market?
Cardiovascular Network News. 1995;1:110.
9.
Tatineni S, Barner HB, Pearson AC, Halbe D, Woodruff
R, Labovitz AJ. Rest and exercise evaluation of St Jude Medical and
Medtronic Hall prostheses: influence of primary lesion,
valvular type, valvular size, and left
ventricular function. Circulation. 1989;80(suppl
I):I-16I-23.
10.
Heinrich RS, Fontaine AA, Grimes RY, Sidhaye A, Yang S,
Moore KE, Levine RA, Yoganathan AP. Experimental analysis of
fluid mechanical energy losses in aortic valve stenosis:
importance of pressure recovery. Ann Biomed Eng. 1996;24:685694.[Medline]
[Order article via Infotrieve]
11.
Munson B, Young D, Okiishi T. Fundamentals of
Fluid Mechanics. New York, NY: John Wiley & Sons; 1990.
12.
Leefe S, Gentle C. Theoretical evaluation of energy
loss methods in the analysis of prosthetic heart
valves. J Biomed Eng. 1987;9:121127.[Medline]
[Order article via Infotrieve]
13.
Gorlin R, Gorlin SG. Hydraulic formula for calculation
of the area of the stenotic mitral valve, other cardiac valves,
and central circulatory shunts. Am Heart J. 1951;41:129.[Medline]
[Order article via Infotrieve]
14.
Holen J, Aaslid R, Landmark K, Simonsen S.
Determination of pressure gradient in mitral stenosis with a
non-invasive ultrasound Doppler technique. Acta Med
Scand. 1976;199:455460.[Medline]
[Order article via Infotrieve]
15.
Hatle L, Angelsen B. Doppler Ultrasound in
Cardiology. 2nd ed. Philadelphia, Pa: Lea &
Febiger; 1985:24.
16.
Fox RW, McDonald AT. Introduction to Fluid
Mechanics. 2nd ed. New York, NY: John Wiley & Sons;
1978;149:273296, 417423.
17.
Rothbart RM, Smucker ML, Gibson RS. Overestimation by
Doppler echocardiography of pressure gradients
across Starr-Edwards prosthetic valves in the aortic position.
Am J Cardiol. 1988;61:475476.[Medline]
[Order article via Infotrieve]
18.
Arabia FA, Talbot TL, Stewart SF, Nast EP, Clark RE. A
computerized physiologic pulse duplicator for in-vitro hydrodynamic and
ultrasonic studies of prosthetic heart valves. Biomed
Instrum Technol. 1989;23:205215.[Medline]
[Order article via Infotrieve]
19.
Yoganathan AP, Jones M, Sahn DJ, Ridgeway A, Jimoh A,
Tamura T. Bernoulli gradient calculations for mechanical aortic valves:
in vitro Doppler study. Circulation. 1986;74(suppl
II):II-391. Abstract.
20.
Ridgway AJ. Ultrasound Doppler Evaluation of
Mechanical Aortic Heart Valves [PhD thesis]. Atlanta, Ga:
Georgia Institute of Technology; 1986.
21.
Aris A, Ramirez I, Camara ML, Carreras F, Borras X,
Pons-Llado G. The 20 mm Medtronic Hall prosthesis in the
small aortic root. J Heart Valve Dis. 1996;5:459462.[Medline]
[Order article via Infotrieve]
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Assessment of Small-Diameter Aortic Mechanical Prostheses
Physiological Relevance of the Doppler Gradient, Utility of Flow Augmentation, and Limitations of Orifice Area Estimation
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundNoninvasive assessment of
functionally stenotic small-diameter aortic mechanical
prostheses is complicated by theoretical constraints relating to the
hemodynamic relevance of Doppler-derived
transprosthetic gradients. To establish the utility of
Doppler echocardiography for evaluation of
these valves, 20-mm Medtronic Hall and 19-mm St Jude prostheses were
studied in vitro and in vivo.
Key Words: prosthesis echocardiography hemodynamics
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
More than two thirds
of mechanical aortic prostheses inserted annually in the United States
are small and functionally stenotic (valve size
23 mm,
effective orifice area <2.0
cm2).1 Assessment of these
prostheses is complicated by physical constraints that preclude or
inhibit direct measurement of transprosthetic pressure
gradients as well as theoretical constraints relating to the
application of the Bernoulli model to flow velocities that frequently
are not uniformly distributed across the plane of the
prosthetic valve orifice.2 3 4
Specifically, the hemodynamic relevance of the
Doppler-derived transprosthetic gradient (which frequently
exceeds the catheter-derived manometric gradient) has been
challenged,5 6 and invasive verification by
catheter in patients with elevated Doppler gradients is hampered by
the requirement for transprosthetic catheter placement or
septal puncture. Moreover, the sensitivity of conventional valve
orifice area estimates for detection of small changes or differences in
hemodynamics, such as might occur in the early stages
of pannus formation (tissue ingrowth) and/or thrombotic occlusion, has
not been meaningfully evaluated.
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
In Vitro Study
Model
Hemodynamic measurements were made in vitro with
a physiological pulsatile flow system driven by a
piston pump (Superpump SPS3891 Vivitro, Inc). To facilitate
prosthetic valve insertion, a straight model was used with
circular cross-sectional inlet and outlet areas of 5.07
cm2 (Figure 1
). The
piston was controlled with a PC-based analog motion controller (R4005,
Rapid Systems, Inc). Peripheral resistance and compliance
were simulated by PVC ball valves and flexible tubing, respectively.
Compliance and resistance values remained constant throughout the
experiments. All experiments were performed at a heart rate of 60 bpm
with systolic duration equal to one third of the total cycle
time. Data were acquired at incremental stroke volumes between 40 and
120 mL. For purposes of acoustic reflection, a saline solution of
kinematic viscosity (1x10-6
m2/s) with 1% cornstarch particles (by volume)
was used for all experiments. Measurements were made for the 20-mm
Medtronic-Hall tilting-disk valve and the 19-mm St Jude bileaflet
prosthesis (internal orifice areas of 1.74 and 1.21
cm2, respectively).7

View larger version (20K):
[in a new window]
Figure 1. Pulsatile flow system designed to simulate
physiological transaortic flow.
Fluid mechanical energy loss was defined as energy lost to
viscous dissipation (heat) that was no longer able to cause fluid
motion. Fluid mechanical energy losses were assessed by use of control
volume analysis to calculate the energy flux entering and
leaving a volume that spanned the entire prosthesis model from
inlet to outlet.10 The control volume extended
from a pressure tap 5.8 cm upstream of the valve to a pressure tap 17.9
cm downstream of the valve. Pump work (WI) was
defined as energy entering the control volume and energy loss as the
difference between energy entering and leaving the control volume
during 1 cardiac cycle:
The energy leaving the control volume (WO)
was defined as the energy available to drive blood through the systemic
circulation. The energy crossing each boundary of the control volume
was calculated by integrating flow rate (Q) and the total pressure of
the fluid at the control volume boundary over the cardiac
cycle:

The total pressure (PT) is equal to the
static pressure (P) plus the dynamic pressure or kinetic energy (1/2

v2):
All calculations were based on a control volume form of
conservation of fluid mechanical energy.11 The
derivation of equations 1 through 3 involves the following
assumptions12 : (1) The fluid in the control
volume is incompressible. (2) The control volume is coincident with the
internal surfaces of the model and perpendicular to the cross section
at the inlet and outlet. (3) Cycle-to-cycle increases in stored energy
within the control volume are negligible, so that flow through and
within the control volume does not change from 1 cycle to the next over
the measurement period (this was verified experimentally by repeating
measurements over 10 consecutive cardiac cycles).

Pressures were measured simultaneously at 3
wall-mounted pressure taps, sited respectively at the inlet (5.8 cm
upstream of the valve), just distal to the prosthetic valve
orifice (0.76 cm from the downstream edge of the valve), and at the
outlet (17.9 cm downstream of the orifice), with strain gauge pressure
transducers (Uniflow, Baxter Healthcare). Flow rate was measured with
an in-line electromagnetic flow probe (EP680, Carolina Medical)
attached to a flowmeter (FM501, Carolina Medical). The pressure and
flow rate signals were filtered at 100 Hz to remove high-frequency
noise and displayed by a physiological signal
recorder (Cardiomed 4008, Medi-Stim, Inc). Fluid velocity was
measured with a custom-made 10-MHz Doppler ultrasound needle probe,
operating in high-pulse-repetition frequency mode and interfaced with
an ultrasound signal processor (SD-100, Vingmed A/S). Doppler
ultrasound velocity was automatically tracked on-line and outputted as
a digital signal. Pressure and flow rate signals from the Cardiomed
4008 were outputted as analog signals and were digitized at 1 kHz. Ten
consecutive simulated heart cycles were collected and averaged at each
pressure-tap location. All data were collected with a 12-bit A/D board
(DAQPad-1200, National Instruments, Inc) connected through the parallel
port to a PC (Latitude Xpi, Dell Computers Corp) with the use of custom
data collection software (LabVIEW 3.0, National Instruments, Inc). Data
were analyzed off-line. Mean systolic pressure
differences between the upstream and orifice pressure taps and between
the upstream and downstream pressure taps were calculated numerically
by use of FORTRAN programs. Work and fluid mechanical energy losses
were calculated through numerical integration of the flow rate,
pressure, and Doppler ultrasound velocity signals over the cycle
duration by use of the trapezoidal rule.
Patients
Nineteen patients were studied 6 to 96 months after insertion of
19-mm St Jude (n=11) or 20-mm Medtronic Hall (n=8) aortic prostheses
for management of critical aortic stenosis. Patients ranged in
age from 42 to 86 years (mean, 72±10); 18 (95%) were female. All were
in NYHA functional class I or II without a history of angina. All had
normal left ventricular systolic
performance as assessed by 2-dimensionaltargeted M-mode
echocardiography (ie, shortening fraction >30%).
Normal prosthetic leaflet or disk excursion was confirmed in
every patient by cineangiography or 2-dimensional
echocardiography.
Two-dimensional echocardiographic and
Doppler data (pulsed wave and continuous wave) were acquired in the
left lateral decubitus position with a Hewlett Packard Sonos 1500
echocardiographic imaging system using 2.5-MHz and
Pedoff transducers. The internal diameter of the
prosthetic valve ring was measured from 2-dimensional images in
the parasternal long-axis view. Flow velocities at the level of the
prosthetic valve ring and just distal to the prosthetic
valve leaflets or disk were determined in the apical 5-chamber view by
pulsed and continuous-wave Doppler, respectively. All data were
recorded as the average value determined from 3 consecutive
Doppler or 2-dimensional echocardiographic images.
Data were acquired at baseline and during incremental
dobutamine infusion (2.5, 5.0, 7.5, and 10.0 µg ·
kg-1 · min-1). A
5-minute equilibration period was allowed before data acquisition after
each dobutamine dose adjustment.
Stroke volume (SV) was determined from Doppler
echocardiographic data as
where VTIPA is the velocity-time integral
across the prosthetic valve annulus and
DPA is the prosthetic valve annular
diameter.

where T is the duration of ejection, t0 is
the onset of ejection, and V2t and
V1t are the instantaneous velocities at time t
across the valve leaflets or disk and the valve annulus,
respectively.

where Q is the transprosthetic flow rate
(cm3 · s-1)
calculated as stroke volume divided by ejection time,
CC is the coefficient of orifice contraction
across the prosthetic valve disk, CV is
the velocity coefficient, and g is the acceleration due to gravity (980
cm · s-2).

Differences in energy losses, orifice pressure gradients, and
recovered pressure gradients between the 2 prostheses were assessed by
2-sample t tests with 95% CIs. A P value <0.05
was considered statistically significant. Relations between downstream
and orifice gradients were assessed by linear regression
analysis. Relations between pressure gradients and energy
losses were assessed by best-fit (second-order polynomial) regression
analysis.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
In Vitro Study
Orifice and Downstream Pressure Gradients
Mean systolic orifice and downstream pressure gradients
for both prostheses over a wide range of stroke volumes are shown in
Figure 2
(upper and middle panels). For
both valves, mean transprosthetic gradients (orifice and
downstream) increased with increasing stroke volume. Orifice gradients
for the Medtronic Hall valve ranged from 6.9 mm Hg (at a stroke
volume of 40 mL) to 41.8 mm Hg (at a stroke volume of 120 mL). At
the same values for stroke volume, orifice gradients across the St Jude
valve were 8.2 and 58.6 mm Hg, respectively. Over the entire
range of stroke volume values, orifice gradients were higher for the St
Jude bileaflet 19-mm valve than for the Medtronic Hall tilting-disk
20-mm valve.

View larger version (20K):
[in a new window]
Figure 2. Effect of increasing stroke volume on
orifice (top) and downstream (middle) mean pressure gradients and on
fluid mechanical energy losses (bottom) across 20-mm Medtronic Hall and
19-mm St Jude valves during simulated transaortic flow.
).

View larger version (15K):
[in a new window]
Figure 3. Relations between orifice and downstream mean
pressure gradients for 20-mm Medtronic Hall (dashed lines) and 19-mm St
Jude (solid lines) valves in the aortic position.
Figure 2
(lower panel) shows the fluid mechanical energy losses
over the range of stroke volume values for the 2 prosthetic
valves. Predictably, the energy losses, which define the
physiological workload of each valve, increased
with increasing stroke volume. At each stroke volume value, the energy
lost during fluid transit across the St Jude valve was higher than that
dissipated during transit across the Medtronic Hall valve
(P<0.001).
Fluid mechanical energy losses were as strongly correlated with
orifice as with downstream transprosthetic pressure gradients
for both valves (r2=0.99 for both the
Medtronic Hall and St Jude valves) (Figure 4
). Relations between orifice gradient
and fluid mechanical energy losses were similar for the Medtronic Hall
and St Jude valves at low stroke volumes. At higher stroke volumes, the
energy losses incurred for a given driving pressure (ie, orifice
gradient) tended to be larger for the St Jude than for the Medtronic
Hall prosthesis (Figure 4
, upper panel).

View larger version (21K):
[in a new window]
Figure 4. Relation between fluid mechanical energy losses
and transprosthetic gradient measured just distal to the valve
orifice (orifice gradient, top) and 17.9 cm downstream of the valve
(downstream gradient, bottom) for the 20-mm Medtronic Hall (dashed
lines) and 19-mm St Jude (solid lines) valves.
Hemodynamic Effects of Flow Augmentation
Representative continuous-wave Doppler data at
increasing dobutamine doses are shown in Figure 5
. In all cases, incremental
dobutamine caused a stepwise increase in cardiac output
(4.1±0.8 versus 6.5±1.2 L/min, baseline versus peak), accompanied by
a corresponding progressive increase in mean transprosthetic
gradient (MPGB) from 13±6 mm Hg at
baseline to 26±10 mm Hg at peak dobutamine.

View larger version (49K):
[in a new window]
Figure 5. Transprosthetic continuous-wave
Doppler flow velocities during incremental dobutamine
infusion in a representative patient.
Baseline heart rates and stroke volumes were similar for patients
with Medtronic Hall and St Jude valves (73±13 versus 75±10 bpm and
58±10 versus 55±17 cm3/beat, respectively).
Because no significant differences were detectable in baseline mean
transprosthetic gradient values for the 2 valves (12±4
mm Hg for Medtronic Hall and 15±8 mm Hg for St Jude), effective
orifice areas estimated by the Gorlin formula were also similar
(1.25±0.27 and 1.07±0.17 cm2 for Medtronic Hall
and St Jude, respectively). Incremental flow augmentation did not
influence effective orifice area estimates (Figure 6
). Even at peak dobutamine
doses, differences between estimated effective orifice areas for the 2
valves were not significant.

View larger version (50K):
[in a new window]
Figure 6. Effect of incremental dobutamine doses
on estimated effective orifice area for 20-mm Medtronic Hall and 19-mm
St Jude valves.
). The
slopes of the 2 lines were similar, but the y intercepts were
different. For any given transprosthetic flow rate, the
transprosthetic gradient was higher across the 19-mm St Jude
than the 20-mm Medtronic Hall valve (P<0.05).

View larger version (25K):
[in a new window]
Figure 7. Logarithmic plot of cardiac output versus mean
transprosthetic gradient, determined from flow velocities by
use of modified Bernoulli equation (MPGB), for 20-mm
Medtronic Hall (dashed lines) and 19-mm St Jude (solid lines) valves.
Slopes are similar, implying similarity of best-fit power functions
describing pressure-flow relations across the 2 prostheses. y
Intercepts are different, implying differences in driving pressure
required to effect similar flow across the 2 prostheses.
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
Determination of Transprosthetic Pressure Gradients From
Flow Velocity Data
Doppler-based estimation of pressure gradients across
stenotic lesions invokes the fundamental physical principal of
energy conservation.14 15 16 As blood accelerates
across the constricted valve orifice, energy that is being expressed as
hydrostatic pressure is converted to kinetic energy, manifested as
velocity. Previous studies have demonstrated that as the
poststenotic jet expands, some degree of pressure recovery may
occur.3 17 18 19 Therefore, if standard manometric
techniques are used, the maximal pressure gradient across a
stenotic orifice will be identified only if the distal
manometer is sited exactly in the region of minimal pressure/maximal
velocity (ie, at the vena contracta, where the streamlines
are maximally constricted). Continuous-wave Doppler, on the other
hand, routinely detects the maximal velocity over a range of sample
sites. Derivation of the pressure gradient by application of the
Bernoulli equation to continuous-wave Doppler flow velocity data,
therefore, generally provides the maximal pressure gradient in the
region of flow velocity interrogation. It follows that if there is
significant pressure recovery, pressure gradients derived
from continuous-wave Doppler flow velocity data are likely to
exceed manometric measurements.2 3 Our in vitro
data confirm previous reports of significant pressure recovery during
physiological systolic ejection across
small-caliber St Jude valves.1 3 17 18 19 However,
this phenomenon was also observed with 20-mm Medtronic Hall valves, a
finding that has not been emphasized previously.2
For both types of prosthetic valve, pressure gradients detected
by manometers positioned 17.9 cm distal to the prosthetic valve
orifice were systematically lower than those detected by manometers
positioned just distal to the orifice, as shown in Figure 2
. Previous
studies have shown excellent agreement between the Doppler-derived
gradient and the orifice gradient detected by manometers positioned
just distal to the prosthetic valve orifice for both the 19-mm
St Jude and the 20-mm Medtronic Hall valves.3 20
That continuous-wave Doppler provides an accurate assessment of the
orifice gradient is not in doubt; it is the
physiological relevance of this Doppler
gradient that has been challenged.5
The in vitro portion of this study showed an excellent correlation
between the orifice pressure gradient and fluid mechanical energy
losses incurred during ejection for both prostheses, implying an
association between the Doppler gradient and myocardial energy
expenditure. Because this relationship was as powerful as that between
the more distally determined transprosthetic gradient (measured
in the zone of recovered pressure) and fluid mechanical energy losses,
our data suggest that the Doppler-derived gradient across these
small-caliber prostheses has similar hemodynamic
relevance to the catheter gradient that is conventionally measured in
the zone of recovered pressure. Both these gradients are related to
energy losses and accurately reflect the hemodynamic
workload placed on the left ventricle by the valve. Indeed, because the
gradient value is so highly dependent on the position of the distal
catheter, it might be argued that Doppler, by consistently
providing access to the maximal pressure gradient, is the method of
choice for serial assessment of the function of small-caliber
mechanical prostheses.
By clinical convention, the estimated valve orifice area, which
incorporates simultaneous measurements of
transvalvular pressure gradient and flow, is considered the
"gold standard" for in vivo hemodynamic evaluation
of stenotic cardiac valves.
). At the low-flow rates that frequently
prevail in elderly aortic stenosis patients with small aortic
roots, small hemodynamic changes that might occur with
early pannus or thrombus formation may not be detected by serial
estimates of valve orifice area determined at the baseline
hemodynamic state.

View larger version (21K):
[in a new window]
Figure 8. Representative examples of
pressure-flow relations across the aortic prostheses of 2 patients with
orifice areas of 1.3 cm2 (solid line) and 0.9
cm2 (interrupted line), respectively. Expected gradient
differentials for the 2 orifice areas at flow rates of 5.0 and 3.0
L/min are shown, highlighting the narrower margin for error at low
flows. VOA indicates valve orifice area.
1. The transprosthetic pressure gradient measured just
distal to the orifice of small-caliber mechanical prostheses tends to
be higher than that measured further downstream in the ascending aorta.
This phenomenon, which pertains to both bileaflet and tilting-disk
prostheses, is due to initial hydrostatic pressure loss and then
pressure recovery of blood transiting the proximal aorta.
![]()
Acknowledgments
This work was supported in part by an educational grant from
Medtronic Inc. We appreciate the insights and advice of Dr Stephen
Khan, who collaborated generously during the course of this
project.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
1.
Baumgartner H, Khan SS, DeRobertis M, Czer LS,
Maurer G. Doppler assessment of prosthetic valve orifice
area: an in vitro study. Circulation. 1992;85:22752283.
This article has been cited by other articles:
![]() |
R J Edwards, M Al-Bustami, and C D J Ilsley A novel method of assessing Starr-Edwards aortic valve stenosis Heart, December 1, 2004; 90(12): 1479 - 1480. [Full Text] [PDF] |
||||
![]() |
M. De Carlo, A. D. Milano, C. Nardi, G. Mecozzi, and U. Bortolotti Serial Doppler echocardiographic evaluation of small-sized sorin bicarbon prostheses J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 337 - 343. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Siniawski, H. Lehmkuhl, Y. Weng, M. Pasic, C. Yankah, M. Hoffmann, I. Behnke, and R. Hetzer Stentless aortic valves as an alternative to homografts for valve replacement in active infective endocarditis complicated by ring abscess Ann. Thorac. Surg., March 1, 2003; 75(3): 803 - 808. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. A. McCullough, E. F. Philbin, J. A. Spertus, S. Kaatz, K. R. Sandberg, and W. D. Weaver Confirmation of a heart failure epidemic: findings from the Resource Utilization Among Congestive Heart Failure (REACH) study J. Am. Coll. Cardiol., January 2, 2002; 39(1): 60 - 69. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. G. Butchart, H.-H. Li, N. Payne, K. Buchan, and G. L. Grunkemeier Twenty years' experience with the Medtronic Hall valve J. Thorac. Cardiovasc. Surg., June 1, 2001; 121(6): 1090 - 1100. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |