From St Mary's Hospital and Imperial College School of Medicine,
London, UK.
Correspondence to D. Wyn Davies, MD, Waller Department of Cardiology, St Mary's Hospital, Praed St, London W2 1NY, United Kingdom.
Methods and ResultsThe MEA was deployed in the left ventricles
of 13 patients (end-diastolic diameters, 61.7±8.4 mm
[mean±SD]). We recorded contact electrograms at 76 points
equatorial and 32 points nonequatorial to the MEA during sinus rhythm
using a catheter-locator signal to record direction and distance
from the MEA. Morphology (cross-correlation) and timing of maximum
-dV/dt of contact and reconstructed electrograms were compared
at different distances from the MEA center to endocardium (M-E) and
from the MEA equatorial plane. For equatorial data, the M-E was
32.12±12.12 mm. The timing of reconstructed with respect to
contact electrograms was -1.94±7.12 ms for M-E <34 mm and
-14.16±19.29 ms at M-E >34 mm (P<0.001).
Cross-correlation of electrograms was 0.87±0.12 (95% CI, 0.84 to
0.91) and 0.76±0.18 (95% CI, 0.69 to 0.83) for M-E <34 mm and
>34 mm, respectively. Nonequatorial points were 32.33±10.81
mm (range, 16.9 to 55.6 mm) from the MEA equatorial plane;
electrogram timing difference was -8.97±15.75 ms and was unrelated to
this distance from the equator.
ConclusionsThis noncontact mapping system accurately
reconstructs endocardial unipolar electrograms from the human left
ventricle. At M-E distances >34 mm, timing accuracy of
reconstruction decreases.
Simultaneous endocardial activation mapping with basket
electrodes has been achieved with limited
resolution.8 More detailed
simultaneous activation mapping has been performed during
surgery by applying arrays of multiple electrodes to epicardial or
endocardial surfaces.9 10 11 This method of mapping
requires a thoracotomy and general anesthesia, and many VTs
may be nonsustained or noninducible under these
conditions.12 There has therefore been no means
of producing high-resolution activation maps of the entire intact left
ventricle (LV) that would enable adequate mapping of the
tachycardia circuit in just a few beats and thereby
potentially guide ablative therapy for the large proportion of patients
with poorly tolerated VT.
We report the first human studies using a noncontact mapping system to
detect far-field endocardial potentials from within a cardiac chamber
and, from these potentials, reconstruct unipolar endocardial
electrograms at 3360 points, producing instantaneous endocardial
isopotential maps on a computer-generated "virtual" endocardium.
The aim of this study was to validate the system during sinus rhythm
(SR) by comparing the timing and morphology of reconstructed unipolar
electrograms with contact unipolar electrograms recorded from the
same LV endocardial location.
Mapping Procedure
Noncontact Mapping System
The MEA (a woven braid of 64 0.003-in-diameter wires) is mounted on a
7.6-mL balloon on a 9F catheter (Figure 1
Before deployment of the MEA, patients were given 10 000 IU heparin
with later boluses to maintain activated clotting time at 300
to 400 seconds. The MEA catheter was deployed via the retrograde
transaortic route over a 0.032-in J-tipped guidewire advanced to the LV
apex. With the pigtail of the MEA in the LV apex, the guidewire was
withdrawn and the balloon inflated with a contrast-saline mixture
(Figure 2
Catheter Location
This locator signal serves 2 purposes. First, it is used to provide
measured samples for the geometry matrix of the inverse solution by
constructing a 3-dimensional computer model of the endocardium (virtual
endocardium). This was achieved in the present study either by use
of an elliptical model of the LV fitted to 6 endocardial points
equatorial to the balloon or by dragging a mapping catheter around the
LV chamber, building up a series of coordinates for the endocardium
(contour geometry), generating an anatomically more accurate
endocardial model. Geometric points were sampled at the beginning of
the study during SR and were taken by gating 6 ms before the R wave of
the ECG. The locator signal is also used to display and log the
position of the mapping catheter on the virtual endocardium during a
study (Figure 3
Reconstruction of Electrograms
The potential distribution on the MEA created by potentials at
the blood-endocardial boundary is described by Laplace's equation. The
potential field at any 1 electrode is influenced to a degree by the
potentials from the entire endocardium, the degree of influence being
inversely proportional to the distance between the electrode and each
endocardial point. The potential field created on the MEA surface is
therefore related to the MEAendocardial geometry matrix. When this is
known, it is possible to compute endocardial electrograms from the MEA
potentials by inverse solution of Laplace's equation.
The inverse solution is inherently ill posed, meaning that noise from
the MEA electrodes or inaccuracy in the MEAendocardial geometry
matrix results in large errors in reconstruction of electrograms. To
minimize this, stability is provided by application of a mathematical
constraint, with physiological basis, to the
solution by use of a technique called
regularization,13 14 for which a custom-designed
algorithm based on methods described by Tikhonov and
Arsenin14 is used. Accuracy of
reconstruction of endocardial electrograms is therefore dependent on
the solution to Laplace's equa-tion, the regularization technique
used, and the accuracy of the geometry matrix. Errors in geometry will
still occur and may be related to the number of endocardial points
sampled and the complexity of the geometry of the chamber. These errors
will be reflected in the results shown later.
Several modifications have been applied to previously described
techniques for inverse solution15 to improve the
accuracy of the system. The inverse solution is based on Green's
second formula and is executed by use of higher-order algebraic
expressions:
where D is a domain,
The geometry matrix defines the relationship between the location
of the 64 electrodes on the MEA and 3360 points on the endocardium
where the reconstruction is computed. Theoretically, the use of a model
based on linear splines may cause significant errors in the geometry
estimation because the sharp pyramidal points of the spline
are a poor estimate of the curvature of the endocardial surface.
Therefore, a model based on bicubic splines, which fits the sampled
endocardial points with curves rather than lines, was used.
With these techniques, the system can, from the MEA potentials,
reconstruct and interpolate >3300 electrograms
simultaneously over the entire virtual LV endocardium
(Figure 3
Signal Recording
Validation of Electrogram Reconstruction
The effects of catheter pressure on local electrogram timing and
morphology have been demonstrated.17 Therefore,
unipolar electrograms from the ring electrode, 2 mm from the
mapping catheter tip, were also compared with the reconstructed
electrogram to explore any differences resulting from pressure.
All locations were marked by use of the locator signal. Software
filtering produced a bandwidth of 4 to 300 Hz for both contact and
reconstructed electrograms.
Data Analysis
1. Electrogram timings were compared at the point of maximum -dV/dt in
the reconstructed and contact electrograms.
2. Differences in electrogram morphology, with reference to polarity,
relative amplitude, and frequency of electrogram components, were
compared after the gains had been adjusted to produce an optimal visual
match. Electrogram morphologies were examined by overlaying
reconstructed and contact electrograms recorded from the same site,
and a visual morphology score (1 to 5) was used wherein 5 indicates an
exact match (no difference could be seen in overlying electrograms), 4
indicates features are exact with phase shift (no visible difference in
electrogram morphology, with an overall shift in timing), 3 indicates
many features match with or without phase shift (the overlying
electrograms match closely, but differences in electrogram features can
be seen), 2 indicates few features match with or without phase shift
(similar features can be identified in the electrograms but with a poor
overall match), and 1 indicates no match (<2 clearly identifiable
features can be matched).
3. Equatorial electrogram morphologies were compared by use of a
well-described template comparison algorithm to calculate the
cross-correlation.18 19 20
4. Off-line cross-correlation can be used to measure differences in
timing between reconstructed and contact catheter electrograms. A
computer shifts the reconstructed against the contact electrograms in
time to produce the maximum correlation between the 2 sets of amplitude
samples, and this timing shift is called correlation timing.
Statistics
Equatorial Data
Timing. Perfect matches in timing were obtained as far as 52 mm
from the MEA center, but differences in timing of maximum -dV/dt for
reconstructed electrograms with respect to contact counterparts
increased gradually with distance and significantly at data points
>34 mm from the MEA center (Figure 5A
Morphology. The subjective morphology score (Figure 6A
Nonequatorial Data
Timing. Perfect electrogram reconstruction was demonstrated at distances of up
to 44.1 mm perpendicular to the MEA equator. No relationship was
demonstrated between perpendicular distance from the equator and timing
differences (Figure 8A
Morphology. The mean morphology score of nonequatorial reconstructed electrograms
(Figure 6B
Distal Ring Electrode Data Compared With Reconstructed
Electrograms
The efficacy of catheter ablation of VT has been disappointing.
Although 71% to 90%5 7 23 24 25 26 of VTs are
rendered noninducible, long-term recurrence rates are high.
Attempts to improve these results have been directed both at increasing
the size of ablation lesions and at improving mapping techniques to
increase the precision of delivery of lesions to the complex substrate
causing reentry in ischemic VT.
Detailed simultaneous endocardial and epicardial activation
mapping of VT has been performed at surgery by application of arrays of
multiple electrodes directly to the epicardium or
endocardium.10 However, mapping under anesthetic
with an open chest and possible ventriculotomy has an associated
morbidity and mortality.27 28
Simultaneous mapping of multiple points can also be
performed by use of endocardial basket arrays deployed
percutaneously. Resolution remains limited to the
proportion of electrodes in contact with the endocardium and by unequal
deployment and spacing of the splines.8
Three-dimensional electroanatomic reconstruction of sequentially
acquired contact catheter data has also recently been described and
validated.20 Although this is a significant
advance in mapping technology, like all sequential systems, resolution
is limited by the time available to acquire separate data points, and
its use remains restricted in nonsustained or
hemodynamically unstable arrhythmias.
Noncontact mapping has several potential advantages over these
techniques because high-resolution maps of the entire intact cardiac
chamber are created simultaneously, producing complete maps
of each beat of tachycardia. Noncontact endocardial mapping
was first described by Taccardi et al29 when
olive-shaped and cylindrical probes were used to recorded
noncontact electrograms of ventricular-paced beats in dogs.
Such recordings resulted in low-frequency, low-amplitude cavity
potentials.30 To enhance this technique,
principles that had previously been applied to reconstruction of
epicardial maps from skin-surface electrograms31
were applied to reconstruct endocardial
electrograms.15 32 33 Endocardial reconstruction
has been applied by use of cylindrical probes in an open chest dog with
epicardial echocardiography used to provide a
geometry matrix, with good correlation between contact and
reconstructed electrograms.15
The system reported herein combines inverse-solution mathematics with a
catheter-location device and thus has potential advantages over
previous noncontact systems. Cavity geometry is defined with limited
need for an alternative imaging system, and the progress of a mapping
catheter within the cavity can be monitored continuously and
recorded, potentially reducing exposure to x-rays.
This system was first validated in vitro34
when accurate catheter location and good correlation were seen between
reconstructed and contact potentials by use of a lower-order inverse
solution than presented in this article. With the presently
used inverse solution, reconstruction of electrograms and
catheter-location accuracy decreased at distances >40 mm from the
MEA center.16 In normal dogs, cross-correlation
between reconstructed and contact unipolar electrograms was 0.98, and a
mapping catheter was positioned to within 4.0±3.2 mm of a target
pacing plunge electrode.16 The LVs of those dogs
were smaller (3.98±0.22 cm LV end-diastolic dimension
[mean±SD]) than the LVs of the patients in the present study
(6.17±0.84 cm), which were typical of patients with ischemic
heart disease and VT.
The results of the present study show that this system can
accurately reconstruct electrograms at distances of at least 50 mm
from the center of the MEA but that accuracy decreases with distance,
significantly so >34 mm from the MEA center. There is no
significant difference in timing of maximum -dV/dt between contact
catheter and reconstructed electrograms <34 mm from the MEA
center, but the timing of reconstructed electrogram maximum -dV/dt
becomes earlier than that of contact electrograms >34 mm from the
MEA center. The trend for the reconstructed electrograms to be
displayed earlier than contact electrograms at distances >34 mm
from the MEA center probably results from a combination of errors.
These may be introduced by distance of the endocardium from the MEA,
signal-to-noise ratio, complexity of anatomy, motion of
different regions of the LV, and the presence of scar. These possible
sources of error may be further compounded by regularization, in which
subtle features of the endocardial potentials may be rejected because
of rapidly changing geometry or electrogram amplitude. Thus, the
complex relationship between reconstruction errors and MEA-endocardial
distance can be easily demonstrated but not easily defined. If this
complex, reproducible error in the reconstruction algorithm is defined,
it may be possible to modify the mathematical solution to adjust for
this.
The computerized correlation timing data also demonstrated that
differences in electrogram timing increased with increasing
endocardium-to-MEA distances, but there was not a consistent
shift in timing of reconstructed electrograms to precede contact
electrograms, and differences in timing were less than those measured
by comparison of maximum -dV/dt. Computerized correlation
timing provides a more global measurement of electrogram timing but may
underestimate errors in activation timing of reconstructed electrograms
and requires further validation.
Previous data have suggested that accuracy of electrogram
reconstruction decreases at points toward the poles of the
MEA.16 Our data show no such trend, and although
the morphological accuracy of electrogram reconstruction decreased with
increasing distance from the center of the MEA, the morphology did not
alter significantly with points at increasing perpendicular distances
from the MEA equator. This may be explained by the ellipsoid shape of
the balloon, because although points at increasing angles from the MEA
equator are farther from the MEA center, they may be closer to the
polar surface of the array.
There was no difference between the results of tip- and ring-electrode
electrogram comparisons, which suggests that the pressure effect was
not significant in this study.
Limitations of Study
Conclusions
Received December 2, 1997;
revision received April 1, 1998;
accepted April 20, 1998.
2.
Karagueuzian HS, Fenoglio JJ, Weiss MB, Wit AL.
Protracted ventricular tachycardia induced by
premature stimulation in the canine heart after coronary artery
occlusion and reperfusion. Circ Res. 1979;44:833846.
3.
de Bakker JMT, Coronel R, Tasseron S, Wilde AAM,
Opthof T, Janse MJ, van Capelle FJL, Becker AE, Jambroes G.
Ventricular tachycardia in the infarcted,
Langendorff-perfused human heart: role of the arrangement of surviving
cardiac fibers. J Am Coll Cardiol. 1990;15:15941607.[Abstract]
4.
Wellens HJJ, Schuilenberg RM, Durrer D. Electrical
stimulation of the heart in patients with ventricular
tachycardia. Circulation. 1972;46:216226.
5.
Stevenson WG, Khan H, Sager P, Saxon LA, Middlekauff
HR, Natterson PD, Wiener I. Identification of reentry circuit sites
during catheter mapping and radiofrequency ablation of
ventricular tachycardia late after myocardial
infarction. Circulation. 1993;88:16471670.
6.
El-Sherif N, Mehra R, Gough WB, Zeiler RH. Reentrant
ventricular arrhythmias in the late myocardial
infarction period: interruption of reentrant circuits by cryothermal
techniques. Circulation. 1983;68:644656.
7.
Morady F, Harvey M, Kalbfleisch SJ, El-Atassi R,
Calkins H, Lanberg JJ. Radiofrequency ablation of
ventricular tachycardia in patients with
coronary artery disease. Circulation. 1993;87:363372.
8.
Johnson SB, Packer DL. Intracardiac ultrasound
guidance of multipolar atrial and ventricular mapping
basket applications. J Am Coll Cardiol. 1997;29:202A.
Abstract.
9.
Kimber SK, Downar E, Harris L, Langer G, Mickleborough
LL, Masse S, Sevaptsidis E, Chen TC. Mechanisms of spontaneous shift of
surface electrocardiographic configuration during
ventricular tachycardia. J Am Coll
Cardiol. 1992;20:13971404.[Abstract]
10.
Downar E, Saito J, Doig JC, Chen TC, Sevaptsidis E,
Masse S, Kimber S, Mickleborough L, Harris L. Endocardial mapping of
ventricular tachycardia in the intact human
ventricle, III: evidence of multiuse reentry with spontaneous and
induced block in portions of reentrant path complex. J Am
Coll Cardiol. 1995;25:15911600.[Abstract]
11.
Masse S, Sevaptsidis E, Parson ID, Downar E. A
three-dimensional display for cardiac activation mapping. Pacing
Clin Electrophysiol. 1991;14:538545.[Medline]
[Order article via Infotrieve]
12.
Josephson ME, Horowitz LN, Speilman SR, Waxman HL,
Greenspan AM. Role of catheter mapping in the preoperative evaluation
of ventricular tachycardia. Am J
Cardiol. 1982;49:207220.[Medline]
[Order article via Infotrieve]
13.
Messinger-Rapport BJ, Rudy Y. Regularization of the
inverse problem in electrocardiography: a model
study. Math Biosci. 1988;89:79118.
14.
Tikhonov AN, Arsenin VY. Solutions of Ill-Posed
Problems. Washington, DC: VH Winston & Sons; 1977:2794.
15.
Khoury DS, Taccardi B, Lux RL, Ershler PR, Rudy Y.
Reconstruction of endocardial potentials and activation sequences from
intracavity probe measurements: localization of pacing sites and
effects of myocardial structure. Circulation. 1995;91:845863.
16.
Gornick CC, Adler SW, Pederson B, Hauck J, Budd
J, Schweitzer J. Validation of a new noncontact catheter system for
electroanatomic mapping of the left ventricular
endocardium. Circulation. In press.
17.
Pappone C, Stabile G, DeSimone A, Senatore G,
Turco P, Damiano M, Ioro D, Spaminato N, Chiarello M. Role of
catheter-induced mechanical trauma in localization of target sites of
radiofrequency ablation in automatic atrial tachycardia.
J Am Coll Cardiol. 1996;27:10901097.[Abstract]
18.
Chiang CM, Jenkins JM, DiCarlo LA. Digital signal
processing chip for detection and analysis of intracardiac
electrograms. Pacing Clin Electrophysiol. 1994;17:13731379.[Medline]
[Order article via Infotrieve]
19.
Greenhut SE. Identification of ventricular
tachycardia using intracardiac electrograms: a comparison
of unipolar versus bipolar waveform analysis. Pacing Clin
Electrophysiol. 1991;14:427433.[Medline]
[Order article via Infotrieve]
20.
Gepstein L, Hayam G, Ben-Haim S. A novel method for
nonfluoroscopic catheter-based electroanatomical mapping of the heart.
Circulation. 1997;95:16111622.
21.
Myerburg RJ, Kessler KM, Kimura S, Castellanos A.
Sudden cardiac death: future approaches based on identification and
control of transient risk factors. J Cardiovasc
Electrophysiol. 1992;3:626640.
22.
Kadish A, Morady F. Mechanism-specific therapy of
ventricular tachycardia. J
Cardiovasc Electrophysiol. 1990;1:6869.
23.
Morady F, Kadish A, Rosenheck S, Calkins H, Kou WH,
De-Buitleir M, Sousa J. Concealed entrainment as a guide for catheter
ablation of ventricular tachycardia in patients
with prior myocardial infarction. J Am Coll Cardiol. 1991;17:678689.[Abstract]
24.
Rothman SA, Hsia HH, Cossú SF, Chmielewski IL,
Buxton AE, Miller JM. Radiofrequency catheter ablation of
postinfarction ventricular tachycardia:
long-term success and the significance of inducible nonclinical
arrhythmias. Circulation. 1997;96:34993508.
25.
Scheinman MM. Patterns of catheter ablation practice in
the United States: results of the 1992 NASPE survey. Pacing Clin
Electrophysiol. 1994;17:873875. Editorial.[Medline]
[Order article via Infotrieve]
26.
Borgreffe M, Chen X, Hindricks G, Haverkamp W, Willems
S, Martinez-Rubio A, Shenasa M, Rotman B, Breithardt G. Catheter
ablation of ventricular tachycardia in patients
with heart disease. In: Zipes DP, ed. Catheter Ablation of
Arrhythmias. Armonk, NY: Futura Publishing Co Inc;
1994:277308.
27.
Mickleborough LL, Mizuno S, Downar E, Gray GC.
Late results of operation for ventricular
tachycardia. Ann Thorac Surg. 1992;54:832838.[Abstract]
28.
Bockner D, Breithardt G, Block M, Borggrefe M.
Management of patients with ventricular
tachycardia: does an optimal therapy exist? Pacing
Clin Electrophysiol. 1994;17:559570.[Medline]
[Order article via Infotrieve]
29.
Taccardi B, Arisi G, Macchi E, Baruffi S, Spaggiari S.
A new intracavitary probe for detecting the site of origin of ectopic
ventricular beats during one cardiac cycle.
Circulation. 1987;75:272281.
30.
Khoury DS, Rudy Y. A model study of volume
conductor effects on endocardial and intracavitary
potentials. Circ Res. 1992;71: 511525.
31.
Rudy Y, Messinger-Rapport BJ. The inverse problem in
electrocardiography: solutions in terms of
epicardial potentials. Crit Rev Biomed Eng. 1988;16:214268.
32.
Khoury DS, Rudy Y. Reconstruction of endocardial
potentials from intracavitary probe potentials: a model study.
Proceedings of Computers in Cardiology, Durham, NC, Oct 1114, 1992.
Piscataway, NJ: The Institute of Electrical and Electronics Engineers
Inc Service Center; 1992.
33.
Macchi E, Arisi G, Colli-Franzone P, Guerri L, Olivetti
G, Taccardi B. Localization of ventricular ectopic beats
from intracavitary potential distributions: an inverse model in terms
of sources. Proceedings of the 11th Institute of Electrical and
Electronics Engineers Engineering in Medicine and Biology Society,
Piscataway, NJ: Institute of Electrical and Electronics Engineers Inc
Service Center; 1989.
34.
Beatty GE, Remole SC, Hansen R, Johnston MK, Holte JE,
Benditt DG. Non-contact electrical extrapolation technique to
reconstruct endocardial potentials. Pacing Clin
Electrophysiol. 1994;17:765. Abstract.
© 1998 American Heart Association, Inc.
Clinical Investigation and Reports
Simultaneous Endocardial Mapping in the Human Left Ventricle Using a Noncontact Catheter
Comparison of Contact and Reconstructed Electrograms During Sinus Rhythm
![]()
Abstract
Top
Abstract
Introduction
Methods
Results
Discussion
References
BackgroundCatheter ablation of
ventricular tachycardia is limited in part by
difficulty in identifying suitable sites for ablation. A noncontact
multielectrode array (MEA) has been developed that allows
reconstruction of 3360 electrograms, using inverse-solution
mathematics, that are superimposed onto a computer-simulated model of
the endocardium. This study assesses the accuracy of timing and
morphology of reconstructed unipolar electrograms compared with contact
unipolar electrograms from the same endocardial site.
Key Words: mapping tachyarrhythmia electrophysiology
![]()
Introduction
Top
Abstract
Introduction
Methods
Results
Discussion
References
Reentry is the
mechanism responsible for the majority of cases of
ventricular tachycardia
(VT).1 2 3 4 5 Ablation of VT is dependent on locating
the diastolic activity critical for maintenance of
the reentrant circuit.5 6 Only 10% of patients
with structural heart disease and reentrant VT are suitable for
catheter ablation by use of conventional
techniques7 ; the principal reason for this
limitation is hemodynamic intolerance of the
tachycardia for a sufficient time to allow conventional
sequential endocardial mapping.
![]()
Methods
Top
Abstract
Introduction
Methods
Results
Discussion
References
Patients
Thirteen consecutive patients (mean age, 60 years; range, 32 to
75 years; 1 woman) were studied while undergoing endocardial LV mapping
for catheter ablation of well-tolerated VT (Table 1
). All patients were being treated with
amiodarone therapy, and other antiarrhythmic medications were
continued for the study.
View this table:
[in a new window]
Table 1. Patient Demographics
(n=13)
The study was approved by the local ethics committee, whose
guidelines were followed. All patients were studied in the
postabsorptive state and had given written informed consent. A standard
quadripolar catheter was sited in the right ventricular
apex, and 2 standard, deflectable, mapping/ablation, 4-mm-tip catheters
were passed to the LV, 1 by a retrograde transaortic route and the
other via a transeptal puncture. Pulmonary and systemic
arterial blood pressures were monitored continuously.
The noncontact mapping system (EnSite 3000; Endocardial
Solutions) consists of a catheter-mounted multielectrode array
(MEA), a custom-built amplifier system, and a Silicon Graphics
workstation to run specially designed system software.
). Each wire has a 0.025-in break in
insulation, producing a noncontact unipolar electrode. The raw
far-field electrographic data from the MEA are acquired and fed into a
multichannel recorder and amplifier system, sampled at 1.2 kHz, and
filtered with a bandwidth of 0.1 to 300 Hz. The amplifier also has 16
channels for contact catheters and 12 for the surface ECG. A ring
electrode located on the proximal shaft of the MEA catheter in the
descending aorta is used as a reference for both noncontact and contact
unipolar electrogram recordings.

View larger version (78K):
[in a new window]
Figure 1. Top, Noncontact mapping catheter. Central lumen
allows passage of a 0.032-in guidewire while the side lumen is used to
inflate the 7.6-mL balloon (inset) used to deploy the MEA. Bottom,
Diagrammatic representation of balloon showing positions of
electrodes around the balloon. Exact positions may vary with each
balloon; therefore, electrode positions on each MEA are determined at
manufacture and recorded on a microchip, which is used to
calibrate and orient the computer.
).

View larger version (140K):
[in a new window]
Figure 2. Posteroanterior radiograph showing noncontact
mapping catheter with contrast medium/saline in the balloon, deployed
in the LV (A). Also seen in the LV is a transeptal mapping catheter (B)
and a retrograde transaortic mapping catheter (C) in an equatorial
position to the MEA. Other catheters are positioned in the high right
atrium, right ventricular apex, and right
ventricular outflow tract. This patient also has an
implantable defibrillator lead (D).
The system is able to locate any conventional catheter in space
with respect to the MEA by passing a 5.68-kHz, low-current
"locator" signal between the catheter being located and
alternately between ring electrodes proximal and distal to the MEA on
the noncontact catheter. The MEA detects and determines the locator
signal angles and thus positions the source.
).

View larger version (65K):
[in a new window]
Figure 3. Detail from screen of noncontact-system
workstation. Within the geometric contour of the computer-generated
virtual LV endocardium (top), position of MEA is
represented as a yellow frame. Lateral wall of virtual
endocardium has been removed to reveal endocardial surface. Some
anatomic locations, as identified on fluoroscopy, are labeled in green,
having been marked on the map by use of the catheter-location system.
Inf indicates inferior; Infsept, inferoseptal; Ant,
anterior; and Ant Sept, anteroseptal. A reconstructed isopotential map
is superimposed onto the virtual LV endocardium. Color scale for the
isopotential map, shown as a horizontal line below the virtual
endocardium, has been set so that white represents endocardial
regions where the potential is <-9 mV and purple represents
endocardial regions where the potential is >-8 mV (thus producing an
activation map). Green locator signal line emerging from MEA marks the
position of the mapping catheter tip on the endocardium. Below the
virtual endocardium are waveforms showing surface ECG lead I, contact
electrogram from the mapping catheter tip, and reconstructed
electrogram at the same location.
The electrical activity detected by the MEA is generated
primarily by the electrograms on the endocardial surface and is of
lower amplitude and frequency than the source on the endocardium. A
technique to enhance and resolve these far-field potentials has been
devised based on an inverse solution to Laplace's equation by use of a
boundary element method (BEM). The inverse solution considers how a
signal detected at a remote point will have appeared at the source, and
the BEM is a method for applying the inverse solution to resolve a
matrix of such signals from a source at a known boundary (eg, the
blood-endocardial boundary).

D is the boundary
of D,
/
n represents the outward normal on
D,
2 is the Laplacian,
dA is the surface area differential, dD is the
volume differential, w is the potential field created by a
unit charge in free space, and v is a solution of the
Laplace equation.
). The use of a BEM by the inverse solution means that the
3-dimensional myocardium is treated as a 2-dimensional
endocardial surface as it is by contact endocardial catheters.
Also, reconstructed electrograms are subject to the same electrical
principles as contact electrograms, so that both contain contributions
from surrounding endocardium as well as underlying
myocardium. The contribution of each component to the
electrogram is weighted by amplitude and distance from the site of
measurement. In generating the geometry matrix, data from structures
protruding into the chamber cavity are likely to be ignored in favor of
data from more distant endocardial points on the same vector.
Structures such as papillary muscles, therefore, are not likely to
cause significant distortion of the virtual endocardium but, as with
local contact electrograms, may make a small contribution to the
far-field electrograms.
Contact catheter data and the surface ECG were recorded
simultaneously on a conventional electrophysiology system
(Bard Labsystem or Prucka Cardiolab) and the noncontact mapping system.
Both the contact electrograms and the position of the contact catheter
as determined by the locator signal were recorded
simultaneously. Reconstructed electrograms could be
selected later at any location on the virtual endocardium and displayed
individually, allowing comparison of a contact electrogram and a
reconstructed electrogram from the same endocardial location.
Contact unipolar recordings from a mapping catheter were
made in all patients during SR from
6 separate endocardial locations
equatorial to the balloon. The accuracy of reconstruction was compared
with the distance from the MEA center as defined by the
catheter-location system, which has been validated by in vitro and
animal studies.16 The perpendicular equatorial
distance of the contact catheter from the true equatorial plane of the
balloon was calculated by the catheter-location system and displayed in
real time. Equatorial positions were defined as endocardial sites
within 15 mm perpendicular to the true equatorial plane of the
MEA. Recordings were also made at points >15 mm from the
MEA equator to assess the accuracy of reproduction toward the
poles of the MEA. Contact catheter sites were taken from as varied an
anatomic distribution as possible for each patient (Figure 4
).

View larger version (126K):
[in a new window]
Figure 4. Virtual endocardium showing positions at which
electrograms were recorded in 1 study patient. Endocardium is shown
with base at top and apex at bottom and has been opened along the
anterior septum. MEA is shown as a yellow frame. Data points (yellow
dots) are seen outside (n=6) and within the zone of equatorial
positions (n=6), which is shown as a shaded blue area.
Data were analyzed on the Silicon Graphics workstation
as follows:
Continuous data, including differences in timing,
correlation, and endocardial distance, are presented as means,
SDs, and 95% CIs. Means of continuous data were compared with
Student's t test. Negative values for comparison of timing
indicate that the reconstructed electrograms had earlier maximum
-dV/dt than their contact counterparts. Results of the
noncontinuous subjective morphology score were compared with
endocardial distances by use of general linear model ANOVA and linear
regression analysis.
![]()
Results
Top
Abstract
Introduction
Methods
Results
Discussion
References
Results are summarized in Table 2
.
View this table:
[in a new window]
Table 2. Results
Data were collected from 76 equatorial LV points (mean, 5.9
points/patient; range, 5 to 7).
). Thus, data were handled in 2 sets of
points <34 mm and >34 mm from the center of the MEA.
Computer timing analysis did not demonstrate this predominantly
negative shift of reconstructed electrograms with respect to contact
electrograms (Figure 5B
).

View larger version (13K):
[in a new window]
Figure 5. Difference between the timing (ms) of
reconstructed and contact unipolar electrograms versus distance (mm)
from MEA center for equatorial points. A, Timing difference between
reconstructed and mapping catheter tip contact electrograms versus
distance from MEA center. A negative timing value indicates that the
reconstructed electrogram timing precedes the contact electrogram. B,
Correlation timing shift required by a computer algorithm to produce
best fit between reconstructed and catheter tip electrograms versus
distance. C, Timing shift required by a computer algorithm to produce
best fit between reconstructed and catheter ring electrograms versus
distance.
) showed no apparent threshold distance
beyond which the morphology of reconstructed electrograms worsened
(Figure 7A
). Cross-correlation worsened
at greater endocardial distances, but a threshold distance for this was
not clear (Figure 7B
).

View larger version (93K):
[in a new window]
Figure 6. Comparison of corresponding contact (A) and
reconstructed (B) unipolar electrograms taken at 12 different positions
in 1 patient. Calipers have been placed on the maximum -dV/dt for
contact (blue) and reconstructed (red) electrograms so that a negative
time value means that reconstructed electrogram -dV/dt occurs before
contact electrogram -dV/dt. Calipers are positioned with a facility
that displays the dV/dt of each electrogram. Distance (D) of the
catheter from MEA center and morphology score (Score) are recorded
for each electrogram comparison. A, Equatorial points; B,
nonequatorial points.

View larger version (14K):
[in a new window]
Figure 7. Graphs comparing reconstructed electrogram and
contact electrogram morphology versus distance from MEA center for
equatorial points. A, Histogram of mean and SD of subjective morphology
scores comparing reconstructed and mapping catheter tip electrograms at
equatorial points plotted versus 10-mm ranges of endocardial distance.
*P<0.05, **P<0.005 compared with
morphology scores at endocardium-MEA distances of 10 to 20 mm. B,
Cross-correlation of reconstructed and mapping catheter tip electrogram
morphologies versus endocardial distance (mm) for equatorial
endocardial points. C, Cross-correlation of reconstructed and mapping
catheter ring electrogram morphologies versus endocardial distance (mm)
for equatorial endocardial points.
Thirty-two nonequatorial points were recorded (mean, 2.5
points/patient; range, 1 to 6). The mean perpendicular distance of
points from the equatorial plane of the MEA for nonequatorial data was
32.33±10.81 mm (range, 16.9 to 55.6 mm).
).

View larger version (18K):
[in a new window]
Figure 8. Graphs comparing timing (ms) and morphology
of reconstructed and contact electrograms versus distance (mm) from
equatorial plane of MEA. A, Difference in timing between reconstructed
and mapping catheter tip electrograms versus distance from equatorial
plane of MEA for nonequatorial points. B, Histogram of mean and SD of
subjective morphology scores comparing reconstructed and mapping
catheter tip electrograms versus 10-mm ranges of distance from
equatorial plane of MEA (mm) for nonequatorial points.
P=NS.
) showed no statistically significant trend to alter with
increasing perpendicular distance from the equator (Figure 8B
).
Comparison of reconstructed electrograms with the distal ring
electrograms produced results similar to those above (Figures 5C
and 7C
).
![]()
Discussion
Top
Abstract
Introduction
Methods
Results
Discussion
References
The commonest cause for VT in coronary heart disease is
reentry.21 22 The ability to ablate such VT with
catheters has been limited by the time taken to adequately map
endocardial activation with conventional sequential techniques,
restricting this therapy to a subset of patients with
hemodynamically stable VT.
The design of this study in humans meant that the display of
reconstructed electrograms for comparison with contact electrograms
depended on both the reconstruction algorithm and the accuracy of the
catheter-location system. This study is limited by its inability to
determine whether errors seen were due to the reconstruction process or
the location system. However, although previous in vitro data have
supported the accuracy of the locator system and construction of the
virtual endocardium,16 no animal study has yet
compared the virtual endocardium with the true anatomy of the
cardiac chamber. Because this study was performed on humans, it was not
possible to precisely locate the catheter by other techniques and thus
independently determine the accuracy of the locator system. This
mapping system integrates the locator and electrogram reconstruction
technologies, and this study addresses the performance of the
system as a whole. Because the principal purpose of the mapping system
described herein is to map human VT, further study is required to
confirm that data obtained during VT, particularly small-amplitude
diastolic potentials, can also be accurately reconstructed.
A theoretical limitation is that geometry data are acquired during SR
but applied to mapping of VT, when the LV dimensions may be
different.
The noncontact mapping system described in this study can
faithfully reconstruct endocardial electrograms in the human LV during
SR. The accuracy of reconstruction depends on the distance between the
endocardium and the center of the MEA. The consistent nature of
the timing error suggests that the reconstruction of electrograms at
MEA-endocardial distances >34 mm may be improved with refinements
in the mathematical solution and hardware used. Further studies will
determine whether this system can assist in the mapping and treatment
of human arrhythmias.
![]()
Acknowledgments
Richard Schilling is a British Heart Foundation Junior Research
Fellow. The authors would like to thank Dr Gerry C. Kaye of Castle Hill
Hospital, Hull, UK, for his assistance with this study.
![]()
References
Top
Abstract
Introduction
Methods
Results
Discussion
References
1.
Josephson ME, Horowitz LN, Farshidi A,
Kastor JA. Recurrent sustained ventricular
tachycardia, I: mechanisms. Circulation. 1978;57:431439.
This article has been cited by other articles:
![]() |
G. Sivagangabalan, J. Pouliopoulos, K. Huang, M. A. Barry, J. Lu, S. P. Thomas, D. L. Ross, A. Thiagalingam, and P. Kovoor Simultaneous Biventricular Noncontact Mapping and Ablation of Septal Ventricular Tachycardia in a Chronic Ovine Infarct Model Circ Arrhythm Electrophysiol, August 1, 2009; 2(4): 441 - 449. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Aliot, W. G. Stevenson, J. M. Almendral-Garrote, F. Bogun, C. H. Calkins, E. Delacretaz, P. D. Bella, G. Hindricks, P. Jais, M. E. Josephson, et al. EHRA/HRS Expert Consensus on Catheter Ablation of Ventricular Arrhythmias: Developed in a partnership with the European Heart Rhythm Association (EHRA), a Registered Branch of the European Society of Cardiology (ESC), and the Heart Rhythm Society (HRS); in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA) Europace, June 1, 2009; 11(6): 771 - 817. [Full Text] [PDF] |
||||
![]() |
C. Knackstedt, P. Schauerte, and P. Kirchhof Electro-anatomic mapping systems in arrhythmias Europace, November 1, 2008; 10(suppl_3): iii28 - iii34. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Rao, Y. Ling, R. He, A. L. Gilbert, N. G. Frangogiannis, J. Wang, S. F. Nagueh, and D. S. Khoury Integrated multimodal-catheter imaging unveils principal relationships among ventricular electrical activity, anatomy, and function Am J Physiol Heart Circ Physiol, February 1, 2008; 294(2): H1002 - H1009. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Ahlberg, A. M. Yue, N. D. Skadsberg, P. R. Roberts, P. A. Iaizzo, and J. M. Morgan Investigation of pacing site-related changes in global restitution dynamics by non-contact mapping Europace, January 1, 2008; 10(1): 40 - 45. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. W. Kimmel, N. D. Skadsberg, C. L. Byrd, D. J. Wright, T. G. Laske, and P. A. Iaizzo Single-site ventricular and biventricular pacing: investigation of latest depolarization strategy Europace, December 1, 2007; 9(12): 1163 - 1170. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. U. Klemm, R. Ventura, D. Steven, C. Johnsen, T. Rostock, B. Lutomsky, T. Risius, T. Meinertz, and S. Willems Catheter Ablation of Multiple Ventricular Tachycardias After Myocardial Infarction Guided by Combined Contact and Noncontact Mapping Circulation, May 29, 2007; 115(21): 2697 - 2704. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Kanter The Fontan Right Atrium--In Context Circulation, April 3, 2007; 115(13): 1698 - 1700. [Full Text] [PDF] |
||||
![]() |
D. J. Abrams, M. J. Earley, S. C. Sporton, P. M. Kistler, M. A. Gatzoulis, M. J. Mullen, J. A. Till, S. Cullen, F. Walker, M. D. Lowe, et al. Comparison of Noncontact and Electroanatomic Mapping to Identify Scar and Arrhythmia Late After the Fontan Procedure Circulation, April 3, 2007; 115(13): 1738 - 1746. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, M. A. Creager, D. R. Holmes Jr, G. Merli, G. P. Rodgers, C. M. Tracy, et al. American College of Cardiology/American Heart Association 2006 Update of the Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion: A Report of the American College of Cardiology/American Heart Association/American College of Physicians Task Force on Clinical Competence and Training Developed in Collaboration With the Heart Rhythm Society J. Am. Coll. Cardiol., October 3, 2006; 48(7): 1503 - 1517. [Full Text] [PDF] |
||||
![]() |
M. J. Earley, D. J.R. Abrams, S. C. Sporton, and R. J. Schilling Validation of the Noncontact Mapping System in the Left Atrium During Permanent Atrial Fibrillation and Sinus Rhythm J. Am. Coll. Cardiol., August 1, 2006; 48(3): 485 - 491. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. De Ponti and J. A. Salerno-Uriarte Non-fluoroscopic mapping systems for electrophysiology: the 'tool or toy' dilemma after 10 years Eur. Heart J., May 2, 2006; 27(10): 1134 - 1136. [Full Text] [PDF] |
||||
![]() |
H. C. Coghlan, A. R. Coghlan, G. D. Buckberg, and J. L. Cox 'The electrical spiral of the heart': its role in the helical continuum.: The hypothesis of the anisotropic conducting matrix Eur. J. Cardiothorac. Surg., April 1, 2006; 29(Suppl_1): S178 - S187. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Yue, T. R. Betts, P. R. Roberts, and J. M. Morgan Global Dynamic Coupling of Activation and Repolarization in the Human Ventricle Circulation, October 25, 2005; 112(17): 2592 - 2601. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Yue, M. R. Franz, P. R. Roberts, and J. M. Morgan Global Endocardial Electrical Restitution in Human Right and Left Ventricles Determined by Noncontact Mapping J. Am. Coll. Cardiol., September 20, 2005; 46(6): 1067 - 1075. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Thiagalingam, E. M. Wallace, C. R. Campbell, A. C. Boyd, V. E. Eipper, K. Byth, D. L. Ross, and P. Kovoor Value of Noncontact Mapping for Identifying Left Ventricular Scar in an Ovine Model Circulation, November 16, 2004; 110(20): 3175 - 3180. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Lemery, L. Soucie, B. Martin, A. S.L. Tang, M. Green, and J. Healey Human Study of Biatrial Electrical Coupling: Determinants of Endocardial Septal Activation and Conduction Over Interatrial Connections Circulation, October 12, 2004; 110(15): 2083 - 2089. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. W.C. Chow, O. R. Segal, D. W. Davies, and N. S. Peters Mechanism of Pacing-Induced Ventricular Fibrillation in the Infarcted Human Heart Circulation, September 28, 2004; 110(13): 1725 - 1730. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Yue, J. R. Paisey, S. Robinson, T. R. Betts, P. R. Roberts, and J. M. Morgan Determination of Human Ventricular Repolarization by Noncontact Mapping: Validation With Monophasic Action Potential Recordings Circulation, September 14, 2004; 110(11): 1343 - 1350. [Abstract] [Full Text] [PDF] |
||||
![]() |
T.-Y. Liu, C.-T. Tai, B.-H. Huang, S. Higa, Y.-J. Lin, J.-L. Huang, Y. Yuniadi, P.-C. Lee, Y.-A. Ding, and S.-A. Chen Functional characterization of the crista terminalis in patients with atrial flutter: implications for radiofrequency ablation J. Am. Coll. Cardiol., May 5, 2004; 43(9): 1639 - 1645. [Abstract] [Full Text] [PDF] |
||||
![]() |
P D. Bella and C Carbucicchio Non-contact left ventricular endocardial mapping for cardiac resynchronisation therapy: a "slow conduction" towards the fast solution Heart, May 1, 2004; 90(5): 483 - 484. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Auricchio, C. Fantoni, F. Regoli, C. Carbucicchio, A. Goette, C. Geller, M. Kloss, and H. Klein Characterization of Left Ventricular Activation in Patients With Heart Failure and Left Bundle-Branch Block Circulation, March 9, 2004; 109(9): 1133 - 1139. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Higa, C.-T. Tai, Y.-J. Lin, T.-Y. Liu, P.-C. Lee, J.-L. Huang, M.-H. Hsieh, Y. Yuniadi, B.-H. Huang, S.-H. Lee, et al. Focal Atrial Tachycardia: New Insight From Noncontact Mapping and Catheter Ablation Circulation, January 6, 2004; 109(1): 84 - 91. [Abstract] [Full Text] [PDF] |
||||
![]() |
P D Lambiase and M S Marber Myocardial gene and cell delivery Heart, January 1, 2004; 90(1): 1 - 2. [Abstract] [Full Text] [PDF] |
||||
![]() |
R J Schilling Non-contact mapping of the left ventricle and new insights into the mechanisms for success of biventricular pacing Heart, January 1, 2004; 90(1): 3 - 4. [Abstract] [Full Text] [PDF] |
||||
![]() |
P D Lambiase, A Rinaldi, J Hauck, M Mobb, D Elliott, S Mohammad, J S Gill, and C A Bucknall Non-contact left ventricular endocardial mapping in cardiac resynchronisation therapy Heart, January 1, 2004; 90(1): 44 - 51. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Markides, R. J. Schilling, S. Yen Ho, A. W.C. Chow, D. W. Davies, and N. S. Peters Characterization of Left Atrial Activation in the Intact Human Heart Circulation, February 11, 2003; 107(5): 733 - 739. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Wetzel, G. Hindricks, P. Schirdewahn, A. Dorszewski, A. Fleck, F. Heinke, and H. Kottkamp A stepwise mapping approach for localization and ablation of ectopic right, left, and septal atrial foci using electroanatomic mapping Eur. Heart J., September 1, 2002; 23(17): 1387 - 1393. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-T. Tai, J.-L. Huang, Y.-K. Lin, M.-H. Hsieh, P.-C. Lee, Y.-A. Ding, M.-S. Chang, and S.-A. Chen Noncontact three-dimensional mapping and ablation of upper loop re-entry originating in the right atrium J. Am. Coll. Cardiol., August 21, 2002; 40(4): 746 - 753. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Della Bella, A. Pappalardo, S. Riva, C. Tondo, G. Fassini, and N. Trevisi Non-contact mapping to guide catheter ablation of untolerated ventricular tachycardia Eur. Heart J., May 1, 2002; 23(9): 742 - 752. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Kress, D. Krum, V. Chekanov, J. Hare, N. Michaud, M. Akhtar, and J. Sra Validation of a left atrial lesion pattern for intraoperative ablation of atrial fibrillation Ann. Thorac. Surg., April 1, 2002; 73(4): 1160 - 1168. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Schilling, N. S. Peters, J. Goldberger, A. H. Kadish, and D. W. Davies Characterization of the anatomy and conduction velocities of the human right atrial flutter circuit determined by noncontact mapping J. Am. Coll. Cardiol., August 1, 2001; 38(2): 385 - 393. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Hindricks and H. Kottkamp Simultaneous Noncontact Mapping of Left Atrium in Patients With Paroxysmal Atrial Fibrillation Circulation, July 17, 2001; 104(3): 297 - 303. [Abstract] [Full Text] [PDF] |
||||
![]() |
N Saoudi, F Cosio, A Waldo, S.A Chen, Y Iesaka, M Lesh, S Saksena, J Salerno, and W Schoels A classification of atrial flutter and regular atrial tachycardia according to electrophysiological mechanisms and anatomical bases. A Statement from a Joint Expert Group from the Working Group of Arrhythmias of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology Eur. Heart J., July 2, 2001; 22(14): 1162 - 1182. [PDF] |
||||
![]() |
T. Paul, B. Windhagen-Mahnert, T. Kriebel, H. Bertram, R. Kaulitz, T. Korte, M. Niehaus, and J. Tebbenjohanns Atrial Reentrant Tachycardia After Surgery for Congenital Heart Disease : Endocardial Mapping and Radiofrequency Catheter Ablation Using a Novel, Noncontact Mapping System Circulation, May 8, 2001; 103(18): 2266 - 2271. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Anselme, A. Savoure, A. Cribier, and N. Saoudi Catheter Ablation of Typical Atrial Flutter : A Randomized Comparison of 2 Methods for Determining Complete Bidirectional Isthmus Block Circulation, March 13, 2001; 103(10): 1434 - 1439. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al. American College of Cardiology/American Heart Association Clinical Competence Statement on invasive electrophysiology studies, catheter ablation, and cardioversion: A report of the american college of cardiology/american heart association/american college of physicians-american society of internal medicine task force on clinical competence J. Am. Coll. Cardiol., November 1, 2000; 36(5): 1725 - 1736. [Full Text] [PDF] |
||||
![]() |
C. M. Tracy, M. Akhtar, J. P. DiMarco, D. L. Packer, H. H. Weitz, W. L. Winters, J. L. Achord, A. W. Boone, J. W. Hirshfeld Jr, B. H. Lorell, et al. American College of Cardiology/American Heart Association Clinical Competence Statement on Invasive Electrophysiology Studies, Catheter Ablation, and Cardioversion : A Report of the American College of Cardiology/American Heart Association/American College of Physicians-American Society of Internal Medicine Task Force on Clinical Competence Circulation, October 31, 2000; 102(18): 2309 - 2320. [Full Text] [PDF] |
||||
![]() |
G. L. Pierpont, S. S. Chugh, J. A. Hauck, and C. C. Gornick Endocardial activation during ventricular fibrillation in normal and failing canine hearts Am J Physiol Heart Circ Physiol, October 1, 2000; 279(4): H1737 - H1747. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. R. Betts, P. R. Roberts, S. A. Allen, A. P. Salmon, B. R. Keeton, M. P. Haw, and J. M. Morgan Electrophysiological Mapping and Ablation of Intra-Atrial Reentry Tachycardia After Fontan Surgery With the Use of a Noncontact Mapping System Circulation, July 25, 2000; 102(4): 419 - 425. [Abstract] [Full Text] [PDF] |
||||
![]() |
R.J. Schilling, D.W. Davies, and N.S. Peters Clinical developments in cardiac activation mapping Eur. Heart J., May 2, 2000; 21(10): 801 - 807. [PDF] |
||||
![]() |
R.J Schilling, A.H Kadish, N.S Peters, J Goldberger, and D.W. Davies Endocardial mapping of atrial fibrillation in the human right atrium using a non-contact catheter Eur. Heart J., April 1, 2000; 21(7): 550 - 564. [Abstract] [PDF] |
||||
![]() |
S. A. Strickberger, B. P. Knight, G. F. Michaud, F. Pelosi, and F. Morady Mapping and ablation of ventricular tachycardia guided by virtual electrograms using a noncontact, computerized mapping system J. Am. Coll. Cardiol., February 1, 2000; 35(2): 414 - 421. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Y. Ho and R. H Anderson Invited Commentary Asian Cardiovasc Thorac Ann, December 1, 1999; 7(4): 349 - 352. [Full Text] [PDF] |
||||
![]() |
S S FURNISS and R W F CAMPBELL Ventricular tachycardia: is it a burning issue? Heart, June 1, 1999; 81(6): 567 - 569. [Full Text] |
||||
![]() |
R J Schilling, N S Peters, and D W Davies Mapping and ablation of ventricular tachycardia with the aid of a non-contact mapping system Heart, June 1, 1999; 81(6): 570 - 575. [Abstract] [Full Text] |
||||
![]() |
R. J. Schilling, N. S. Peters, and D. W. Davies Feasibility of a Noncontact Catheter for Endocardial Mapping of Human Ventricular Tachycardia Circulation, May 18, 1999; 99(19): 2543 - 2552. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kadish, J. Hauck, B. Pederson, G. Beatty, and C. Gornick Mapping of Atrial Activation With a Noncontact, Multielectrode Catheter in Dogs Circulation, April 13, 1999; 99(14): 1906 - 1913. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. W.C. Chow, R. J. Schilling, D. W. Davies, and N. S. Peters Characteristics of Wavefront Propagation in Reentrant Circuits Causing Human Ventricular Tachycardia Circulation, May 7, 2002; 105(18): 2172 - 2178. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1998 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |