(Circulation. 1999;99:1458-1463.)
© 1999 American Heart Association, Inc.
Clinical Investigation and Reports |
From the Department of Medical Informatics, Faculty of Medicine and Health Sciences, Erasmus University, Rotterdam, the Netherlands.
Correspondence to J.A. Kors, PhD, Department of Medical Informatics, Faculty of Medicine and Health Sciences, Erasmus University, PO Box 1738, 3000 DR Rotterdam, Netherlands. E-mail kors{at}mi.fgg.eur.nl
| Abstract |
|---|
|
|
|---|
Methods and ResultsThe T loop is characterized by its amplitude and width (defined as the spatial angle between the mean vectors of the first and second halves of the loop). We reasoned that small, wide ("pathological") T loops produce larger QTD than large, narrow ("normal") loops. To quantify the relationship between QTD and T-loop morphology, we used a program for automated analysis of ECGs and a database of 1220 standard simultaneous 12-lead ECGs. For each ECG, QT durations, QTD, and T-loop parameters were computed. T-loop amplitude and width were dichotomized, with 250 µV (small versus large amplitudes) and 30° (narrow versus wide loops) taken as thresholds. Over all 1220 ECGs, QTDs were smallest for large, narrow T loops (54.2±27.1 ms) and largest for small, wide loops (69.5±33.5 ms; P<0.001).
ConclusionsQTD is an attribute of T-loop morphology, as expressed by T-loop amplitude and width.
Key Words: electrocardiography computers potentials
| Introduction |
|---|
|
|
|---|
It is erroneous to think, however, that the end of the T wave in a given lead is directly related to the action potential durations in some corresponding region of the heart. At the end of the cardiac cycle, when the last myocardial cells finish their repolarization, the electric field generated by these last active sources will extend throughout the trunk and be picked up by every single electrode one might choose to put in or on the body. At all electrode sites, repolarization potentials must therefore have principally the same duration. This is a consequence of simple electric field theory.9 One cannot measure, however, the potential of individual electrodes but only the potential difference between 2 lead electrodes. If the electric signal in a lead falls to zero, the electrode potentials have become equal. (It must be remembered that the precordial leads are also essentially bipolar because the central terminal by no means constitutes a zero potential.10 ) There is only 1 end of repolarization, which is the common end for all leads together, when the electric field dissolves and all potential differences vanish.
Still, QTD cannot be dealt with simply as a measurement problem because it has been shown that QTD has a certain diagnostic capability.3 5 11 12 13 14 15 16 17 18 In this article, we will present evidence that QTD can be regarded as a manifestation of spatial T-loop morphology; ie, we will explain the phenomenon of QTD in terms of 3 interacting factors: the amplitude of the T loop, its width, and the number of leads in which the end of the T wave could not be determined because of too low T-wave amplitudes. This will be worked out quantitatively with a large database of ECGs and a computer program for automatic measurement of QTD and T-loop parameters.
| Methods |
|---|
|
|
|---|
Measurements
For data processing, the Modular ECG Analysis System
(MEANS), our ECG computer program,20 was used. The
operation of the waveform recognition algorithms has been described and
validated extensively.20 21 MEANS determines common QRS
onset and offset and T offset for all 12 leads together on 1
representative averaged beat by use of template
matching techniques.22
For QTD measurement, the location of the overall end of T is taken as a starting point. The program then determines the end of the T wave in each separate lead by use of a threshold algorithm that is dependent on noise level. QTD is then computed as the difference between the maximum and minimum QT intervals. The T wave in a lead may be so flat that measuring its end point is impossible. In common QTD measurement practice, the lead will then be excluded from analysis. In our experiments, we excluded leads with peak-to-peak ST-T amplitudes of <50 µV (1/2 mm), at which level the wave is considered flat for all practical purposes. The performance of the program in determining QTD was shown to be comparable with that of human observers.23
To derive T-loop parameters, vectorcardiographic leads X,
Y, and Z were reconstructed from the standard ECG
leads.24 25 The following parameters were
taken to characterize T-loop morphology: initial and terminal axes,
width, and maximal amplitude. Each parameter can be
determined for the spatial T loop and its projection on the frontal
(XY), horizontal (XZ), and sagittal (YZ) planes. The initial axis,
T1, is obtained by vectorially adding the
instantaneous heart vectors during the first half of the T loop (half
is defined as half the geometrical circumference of the loop). The
terminal axis, T2, is obtained similarly for the
second half of the T loop. The greater the width of the loop, the more
divergent the initial and terminal parts. We thus defined T width as
the angle between T1 and
T2. To quantify the effect of T-loop width and
amplitude on QTD, we dichotomized these parameters, taking
250 µV (small versus large amplitudes) and 30° (narrow versus wide
loops) as threshold values. The amplitude threshold was based on
previous reports on normal values of T-loop amplitude; the width
threshold was, after some geometrical manipulations, derived from the
rule of thumb that a normal T loop should have a length
2.5 times
greater than its width.26
Statistical Analysis
Statistical analysis was performed by use of Student's
t test for unpaired samples. Data are presented as
mean±SD.
| Results |
|---|
|
|
|---|
Figure 1
shows the mean of the 1220
differences for each lead as a function of the angle between lead axis
and T2 for the frontal and horizontal planes
separately. For leads parallel to T2 (angle
between lead axis and T2 of 0° or 180°), the
mean difference between QT in the lead and QTmax
is smallest. (In fact, QT in that lead will often be
QTmax.) The larger the angle is between
T2 and the lead, the shorter QT tends to be and
the larger the difference is until, when a lead is perpendicular to
T2 (at -90° or 90°), the mean difference is
largest, ie, QT is shortest. For each lead, we tested whether the mean
difference at 0° or 180° was equal to the difference at -90° or
90°. All differences between these means proved highly significant
(P<0.001) for all leads.
|
Lead Exclusion
In terms of T-loop parameters, we will examine the
situations in which flat ST-T waves may occur in a lead. If a T loop
has a narrow, elongated ("spindlelike") shape, its projection
on a given lead axis will result in a well-discernible T wave as long
as the angle with the lead axis is narrow enough (Figure 2A
). The more perpendicular to a lead the
loop becomes, the smaller its projection is and the lower the ST-T
amplitude is in the lead until it may decrease to <50 µV, which was
used as an exclusion criterion (Figure 2B
). This will be the
case in 1 lead only unless the T loop is perpendicular to 1 of the
planes, in which situation it will be perpendicular to all leads of the
plane. The smaller the T loop, the less perpendicular to a lead it has
to be before the ST-T amplitude is so low that it results in exclusion
of the lead. Moreover, a small-enough T loop will produce low ST-T
waves in more leads than the 1 more or less perpendicular to it.
|
A wide, round T loop with its terminal axis perpendicular to a lead
will tend to have an initial axis at an oblique angle of incidence to
the lead, so its projection will not become zero and the ST-T wave
will stay large enough not to be excluded (Figure 2C
). Exclusion
will occur only if the loop is overall very small.
Of the 1220 ECGs, leads were excluded in 429 ECGs (35.2%), of which
326 had 1, 79 had 2, and 24 had >2 excluded leads. To illustrate the
relationship between lead exclusion and T-loop parameters,
Figure 3
shows a scatterplot of frontal
T2 versus frontal amplitude for those ECGs in
which lead III was excluded, with narrow loops marked as "o" and
wide loops as "x." As expected, lead III is excluded only if the
ECG has a tall, narrow T loop with a T2 that
clusters at
30° and -150° (which is perpendicular to the lead
axis of III, assumed to lie at 120°) or if the amplitude of its T
loop is small.
|
QTD and T Loop
Several factors come into play to explain the relation between QTD
and T-loop morphology. We hypothesize that spatial T amplitude, T
width, and number of excluded leads (which in turn depends on axis,
amplitude, and width, as explained above) affect QTD. If the angle
between the axis of the terminal part of the T loop
(T2) and the lead axis deviates sufficiently from
90°, the end of the T wave in that lead will coincide approximately
with the end of the spatial T loop, ie, with the cessation of all
repolarization activity (Figure 2A
). If T2
is perpendicular to the lead, however, the projection on the lead
axis will be zero, and the T wave in that lead will end before the
overall end of repolarization (Figure 2C
). QT duration will then
be shorter in that lead than in a less perpendicular lead, and QTD
results.
There is a difference between narrow and wide loops, as discussed
above. In a narrow loop, when T2 is perpendicular
to a lead, T1 also is approximately
perpendicular, and the projection of the whole loop on the lead
axis is small, so the ST-T wave may become subthreshold and be excluded
(Figure 2B
). If in a round loop the terminal part is
perpendicular to a lead, QT duration is likewise shortened. The initial
part, however, because of the roundness of the loop, will generally be
less perpendicular to the lead, and the ST-T wave retains sufficient
amplitude not to be excluded. QTD might therefore be measured in round
loops, while being absent in narrow loops by exclusion of the very
leads by which it would become manifest.
Overall decreased spatial amplitude will result in decreased ST-T amplitudes. This also leads to increased QTD because of the increased uncertainty in determining the end of low T waves,23 28 29 as long as the measurement is not excluded because of too low ST-T amplitude.
The Table
shows the mean QTD of
ECGs in subgroups by spatial T amplitude and width, distinguishing
between ECGs in which all 12 leads or in which <12 leads could be
measured. Mean QTDs are smallest for narrow, high-amplitude T loops
(54.2±27.2 ms) and largest for small, wide loops (69.5±33.5 ms,
P<0.001). The percentage of ECGs with
1 excluded leads is
lowest for large, wide T loops (22%) and highest for small, narrow
loops (79%). The mean QTD of ECGs with
1 leads excluded is between
3.7 and 14.8 ms shorter than the QTD of ECGs with no excluded
leads.
|
For the presentation of these results, correlation coefficients that presuppose linear relations between variables are not the appropriate means. The relationships between our parameters and QTD are highly nonlinear: The exclusion of leads is an all-or-none decision; the increases in amplitude and width do not lead to a proportional increase in QTD; and the interplay between the 3 factors also creates unforeseeable nonlinearities.
| Discussion |
|---|
|
|
|---|
We also demonstrated that the T axis is associated with QT duration: The more perpendicular the terminal T axis is to the lead axis, the shorter the QT duration is. Conversely, the more parallel the T axis is with the lead axis, the longer the QT duration is. One may wonder whether this fact alone would not be sufficient to explain the differences in QTD between patient and control groups that have been reported in many studies. However, this is not the case because the lead axes of the extremity leads and the precordial leads are periodically distributed over the frontal and transversal planes, respectively. The direction of the terminal T axis determines which lead will have the shortest QT duration (the lead perpendicular to the T axis) and which will have the longest (the lead parallel with it). If the direction of the terminal T axis changes, the leads with the shortest and longest QT will change, but the difference between the longest and shortest QT duration, ie, QTD, will remain the same, independent of the terminal T axis. To explain differences in QTD, T-loop morphology has to be taken into account.
We chose a simple model for the lead axes, and more sophisticated
lead models would have been possible. However, more realistic sets of
lead axes also have limitations because each was derived from only 1,
often not even heterogeneous, mathematical or physical
torso model. Moreover, we use the lead model only to illustrate the
relationship between QT duration and terminal T axis (Figure 1
),
not to demonstrate that QTD is an attribute of T-loop morphology (the
Table
).
Mean differences between maximum QT and QT durations in individual leads were studied before by Cowan et al,11 but they did not relate their findings to the T axis. Several investigators11 28 30 have suggested that QTD might be due to the different projections of the heart vector on the different lead axes. To the best of our knowledge, an explanation in terms of T-loop morphology has not been given before.
Priori et al31 have related QT duration to T-wave morphology as expressed in the principal components of the 8 independent ECG signals. They defined an index of complexity of repolarization (CR) as the ratio between the first and second eigenvalues and showed that CR discriminates between long-QT syndrome patients and control subjects. Their index, however, was not significantly correlated with QTD, and they did not try to explain the phenomenon of QTD in terms of the index. The approach of these authors is mathematical and is not concerned with T-loop shape. We expect the first eigenvalue to be related to the maximum amplitude of the T loop and CR to its width.
In a study by Badilini et al,32 myocardial infarction patients and individuals with long-QT syndrome were shown to have increased QTD compared with normal subjects. Badilini et al also assessed T-loop roundness, similar to CR, and planarity. T-loop amplitude and number of excluded leads were not taken into account. Only moderate correlations between QTD and T-loop parameters were found, which should not come as a surprise in view of the nonlinearity of the relationships, as indicated above. Again, an explanation of the phenomenon of QTD, the major objective of our study, is not offered.
The relation between epicardial action potential durations and QTD has been investigated by Zabel et al.8 Action potential durations were measured on a rabbit heart suspended in a tank and were varied by administration of D-sotalol. The dispersion of action potential durations correlated with the QTDs measured at the surface of the tank. The authors do not explain how the heterogeneity of repolarization in the heart is connected to QTD. Changes in the course and duration of action potential durations in the heart will certainly cause changes in the surface ECG, but this does not mean that areas with longer or shorter action potential durations in the heart are mapped onto discrete areas of increased or decreased repolarization potential duration on the body surface, resulting in a QTD increase. In our thinking, heterogeneity of repolarization leads to greater variability in T-loop morphology. A characteristic of T-loop morphology is width, and we demonstrated that larger T-loop width increases QTD; low T-loop amplitude is a second factor. This does not contradict our principle that local variations in repolarization potential durations on the body surface cannot exist, considering that all repolarization potentials must end at the same moment. The determination of the end of T in the ECG is the measurement of a potential difference, and this measurement yields zero simply when the lead electrodes have equal potential, which can occur in any lead at which the repolarization vector becomes perpendicular to the lead axis.
Our results would also explain why increased QTD is associated
with a variety of pathologies, as has been reported in many previous
studies.3 5 11 12 13 14 15 16 17 18 In clinical vectorcardiography, it is a
well-known fact that normal T loops usually have elongated, narrow
shapes with spatial T amplitudes of
500 µV.33 34 Wide
T loops, on the other hand, are considered a sign of various forms of
pathology, as are small amplitudes of the T loop.35 In the
present study, the difference in mean QTD between these normal
(long and narrow) and abnormal (small and wide) T loops was 15.3 ms. In
previous studies that compared QTDs of myocardial infarction patients
and control subjects (2 groups that constitute about two thirds of our
study population), differences between mean QTDs ranged from 15 to 26
ms.4 11 13 29 Our results are in accordance with these
findings. The thresholds of 250 µV for amplitude and 30° for width
are not very critical. Changes of 15% to 20% up or down did not give
essentially different results.
As has been said, the shape of the T loop somehow reflects the distribution and course of action potentials in the ventricular myocardium. Unfortunately, although certain T-loop characteristics are helpful in recognizing pathological conditions, our understanding of the relationships between T-loop morphology and the pathophysiology of specific repolarization abnormalities is limited. In that respect, one might object that we are not better off with T-loop morphology than with QTD. But as we have argued, QTD is not a physically sound concept. Its existence is due to a measuring problem that can be understood in terms of T-loop morphology. This suggests that T-loop parameters may have a discriminative and prognostic value that is at least as good as that of QTD. Moreover, they can be measured more easily and less ambiguously than QTD.
| Acknowledgments |
|---|
Received August 10, 1998; revision received November 18, 1998; accepted December 7, 1998.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C.-Y. Lin, L.-Y. Lin, and P.-C. Chen Analysis of T-wave morphology from the 12-lead electrocardiogram for prediction of long-term prognosis in patients initiating haemodialysis Nephrol. Dial. Transplant., September 1, 2007; 22(9): 2645 - 2652. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Rubulis, J. Jensen, G. Lundahl, J. Tapanainen, and L. Bergfeldt Ischemia induces aggravation of baseline repolarization abnormalities in left ventricular hypertrophy: a deleterious interaction J Appl Physiol, July 1, 2006; 101(1): 102 - 110. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. H. van Huysduynen, A. van Straten, C. A. Swenne, A. C. Maan, H. J. Ritsema van Eck, M. J. Schalij, E. E. van der Wall, A. de Roos, M. G. Hazekamp, and H. W. Vliegen Reduction of QRS duration after pulmonary valve replacement in adult Fallot patients is related to reduction of right ventricular volume Eur. Heart J., May 1, 2005; 26(9): 928 - 932. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Chen and F. M. Kusumoto QT Dispersion: Much Ado About Something? Chest, June 1, 2004; 125(6): 1974 - 1977. [Full Text] [PDF] |
||||
![]() |
P. M. Okin QT interval prolongation and prognosis: further validation of the quantitative approach to electrocardiography J. Am. Coll. Cardiol., February 18, 2004; 43(4): 572 - 575. [Full Text] [PDF] |
||||
![]() |
P. M. Okin, R. B. Devereux, E. T. Lee, J. M. Galloway, and B. V. Howard Electrocardiographic Repolarization Complexity and Abnormality Predict All-Cause and Cardiovascular Mortality in Diabetes: The Strong Heart Study Diabetes, February 1, 2004; 53(2): 434 - 440. [Abstract] [Full Text] |
||||
![]() |
G. Salles, S. Xavier, A. Sousa, A. Hasslocher-Moreno, and C. Cardoso Prognostic Value of QT Interval Parameters for Mortality Risk Stratification in Chagas' Disease: Results of a Long-Term Follow-Up Study Circulation, July 22, 2003; 108(3): 305 - 312. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Kardys, J. A Kors, I. M van der Meer, A. Hofman, D. A.M van der Kuip, and J. C.M. Witteman Spatial QRS-T angle predicts cardiac death in a general population Eur. Heart J., July 2, 2003; 24(14): 1357 - 1364. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Geelen, I. A. Brouwer, P. L. Zock, J. A. Kors, C. A. Swenne, M. B. Katan, and E. G. Schouten (n-3) Fatty Acids Do Not Affect Electrocardiographic Characteristics of Healthy Men and Women J. Nutr., October 1, 2002; 132(10): 3051 - 3054. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Pietila, H. Fodstad, E. Niskasaari, P.a. J. Laitinen, H. Swan, M. Savolainen, Y. A. Kesaniemi, K. Kontula, and H. V. Huikuri association between HERG K897T polymorphism and QT interval in middle-aged finnish women J. Am. Coll. Cardiol., August 7, 2002; 40(3): 511 - 514. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. D. Gbadebo, R. W. Trimble, M. S.C. Khoo, J. Temple, D. M. Roden, and M. E. Anderson Calmodulin Inhibitor W-7 Unmasks a Novel Electrocardiographic Parameter That Predicts Initiation of Torsade de Pointes Circulation, February 12, 2002; 105(6): 770 - 774. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Brendorp, H. Elming, L. Jun, L. Kober, M. Malik, G. B. Jensen, and C. Torp-Pedersen QT Dispersion Has No Prognostic Information for Patients With Advanced Congestive Heart Failure and Reduced Left Ventricular Systolic Function Circulation, February 13, 2001; 103(6): 831 - 835. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Malik and V. N. Batchvarov Measurement, interpretation and clinical potential of QT dispersion J. Am. Coll. Cardiol., November 15, 2000; 36(6): 1749 - 1766. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Zabel, B. Acar, T. Klingenheben, M. R. Franz, S. H. Hohnloser, and M. Malik Analysis of 12-Lead T-Wave Morphology for Risk Stratification After Myocardial Infarction Circulation, September 12, 2000; 102(11): 1252 - 1257. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Zabel, M. R. Franz, J. A. Kors, G. van Herpen, and J. H. van Bemmel The Electrophysiological Basis of QT Dispersion: Global or Local Repolarization? Response Circulation, June 27, 2000; 101 (25): e235 - e236. [Full Text] [PDF] |
||||
![]() |
M. Malik QT dispersion: time for an obituary? Eur. Heart J., June 2, 2000; 21(12): 955 - 957. [PDF] |
||||
![]() |
Is QT dispersion a reliable index of heterogeneity of ventricular repolarization and a pro-arrhythmic marker? Eur. Heart J., January 1, 2000; 21(1): 79 - 79. [PDF] |
||||
![]() |
P. M. Okin, R. B. Devereux, R. R. Fabsitz, E. T. Lee, J. M. Galloway, and B. V. Howard Principal Component Analysis of the T Wave and Prediction of Cardiovascular Mortality in American Indians: The Strong Heart Study Circulation, February 12, 2002; 105(6): 714 - 719. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Zabel, M. Malik, K. Hnatkova, V. Papademetriou, A. Pittaras, R. D. Fletcher, and M. R. Franz Analysis of T-Wave Morphology From the 12-Lead Electrocardiogram for Prediction of Long-Term Prognosis in Male US Veterans Circulation, March 5, 2002; 105(9): 1066 - 1070. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Circulation Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 1999 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |